| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
In 1997, it was reported (1) that there were more than 120 million diabetics world-wide; this number is predicted to increase to more than 220 million by the year 2010. Diabetes mellitus is a complicated and serious condition and is one of the most prevalent noncommunicable diseases in the world. Successful management of diabetes involves knowledge of the current blood glucose concentration to allow the diabetic to compensate by diet, oral medication, or insulin injections. Without knowledge of the blood glucose concentration, the correct treatment is not possible and serious complications affecting internal organs, circulation, and eyesight may occur.
The current methods of measuring blood glucose concentrations require the diabetic to obtain a blood sample for analysis by a test strip. According to a recent 9-year study, The Diabetes Control and Complications Trial (2), optimum treatment for a diabetic requires regular blood glucose measurements; however, the discomfort experienced by users means that present methods are not being used by diabetics with sufficient regularity. There are indications (3) that the use of spot check measurements does not provide adequate information for optimal glucose control, and there is a strong case for more frequent or "continuous" measurements. This type of measurement would also yield the rate of change of glucose, which would allow planning of the appropriate insulin dose or dietary content.
The measurement of blood glucose by any technique is inherently complex because of the wide range of potentially interfering components. For a noninvasive technique, not only are there many analytes within human blood that could interfere with the measurement, but there are also other problems such as the variability and inhomogeneity of human skin and the constantly changing human physiology.
Research into new, less painful methods of measurement has been carried out for several decades, and several reviews on the subject have been published (4)(5)(6). The two most popular techniques being attempted are measurement of the change in optical transmission or polarization rotation attributable to the presence of glucose. In transmission configurations, the intensity of light transmitted at several wavelengths is used to determine glucose concentrations. Transmission spectroscopy is susceptible to light scattering in tissue and is technically limited by the requirement of fixed, diametrically opposed, positioning of light source and detector. Transmission measurements have been carried out on the skin (7) and on animal and human eyes (8), making use of the correlation between blood glucose concentration and the glucose concentration of the aqueous humor in the eye. Optical rotation of polarized light in the aqueous humor of the eye (9)(10)(11)(12) has also been investigated. Other researchers have carried out in vitro tests using Raman spectroscopy (13)(14)(15)(16), fluorescence studies (17) and Fourier-transform infrared spectroscopy (18)(19)(20). The human lip has been used as a location for glucose measurements using techniques including diffuse reflectance (21) and attenuated total reflectance (22). The latter technique has also been used for tests on the finger (23).
Other noninvasive approaches attempt to use a correlation between glucose content in interstitial fluid and capillary blood (24). The use of a patch or reservoir on the skin surface to collect glucose has also been explored (25).
In this report, we discuss the alternative technique of pulsed laser photoacoustic spectroscopy. In this technique, short pulses of laser light are directed into the tissue. The resulting acoustic signal depends on the optical and physical characteristics of the sample and may be the basis of determining blood glucose concentrations noninvasively.
Recent developments in optoelectronic technologies give the photoacoustic method of measurement the potential for portability and long component life with the use of piezoelectrics and compact electronics and diode lasers with levels of optical radiation that are several orders of magnitude below pain or tissue damage thresholds.
| Background |
|---|
|
|
|---|
The photoacoustic pulse consists of an isolated pressure wave
comprising a compressive pulse followed by a rarefaction. Generally,
the peak-to-peak amplitude of the detected photoacoustic pulse is used
for analysis and is directly proportional to the absorbed energy.
Accordingly, an energy measurement is used to normalize the
peak-to-peak measurement. A typical photoacoustic signal is shown in
Fig. 1
. The lower trace is the optical signal, and the upper trace is
the photoacoustic signal. A time delay between the optical pulse and
the photoacoustic pulse is evident. This time delay is attributable to
the propagation of the photoacoustic pulse from the acoustic source to
the detector and can be used for velocity of sound measurements or to
determine the location of the acoustic source in imaging techniques.
|
The generated pressure in the photoacoustic process can be described by
the following wave equation, where electrostriction (26) is
ignored:
![]() | (1) |
where I is the intensity of the laser light,
v is the sound velocity in the medium,
is the optical
absorption coefficient, ß is the coefficient of volumetric
thermal expansion, Cp is the specific
heat capacity, and p(r,t) is the acoustic pressure.
For a weakly absorbing sample, the peak pressure P can be
described (27)(28) by the following equation:
![]() | (2) |
In some cases, the combination of physical parameters gives an enhancement of the photoacoustic response compared with the response in conventional spectroscopy, and this can be exploited, for example, in the detection of hydrocarbons in water (29)(30). The response to glucose, however, does not give substantial photoacoustic signal enhancement.
Photoacoustics have been used to obtain information about the concentrations of species, including the monitoring of trace gas concentrations (31)(32)(33)(34), medical applications such as the detection of cancer (35)(36), and investigations into skin structure (37) and chemical penetration in the skin (38). Photoacoustics has also been used for lateral and depth imaging (35)(36)(39)(40)(41)(42)(43)(44) and has applications in other medical and biological areas, including the determination of melanin in human hair (45) and investigations into the diffusion of chromophores in human skin (46).
| Experimental Studies |
|---|
|
|
|---|
Although previous work in the mid-infrared region demonstrated the potential of photoacoustics as a method of measuring glucose concentrations, this wavelength region is not regarded as viable for human tissue studies because of the high water absorption that reduces penetration depths to microns. This penetration may not be sufficient to investigate blood constituents within human tissue, although interaction with interstitial fluid yields measurements that correlate with blood glucose concentrations but with a time shift (48).
The spectral region that shows the most promise for absorption by the analytes within blood is within the "tissue window", around the 12 µm region (49). Although measurements within this region are advantageous for tissue studies, they coincide with a region of lower glucose absorption. Initial results with pulsed photoacoustics achieved glucose concentration measurements within this region in aqueous solutions (50), gelatin phantoms (51), and blood samples (52).
Aqueous glucose solutions in the concentration range 1.733 mmol/L (30600 mg/dL) have been used for in vitro measurements in this tissue window spectral region (49). The percentage of change in the photoacoustic response of the glucose solutions was compared with that of water at a wavelength of 1700 nm. A linear relationship, y = 0.21x - 0.02 with a correlation coefficient of 0.99 was obtained, where x is the glucose concentration and y is the percentage of change of the photoacoustic response from that of water.
Photoacoustic measurements in a gelatin-based tissue phantom (51) have also shown a linear correlation when a wavelength of 1.064 µm is used to investigate glucose concentrations between 1 and 100 mmol/L (180 and 18 000 mg/dL). At this wavelength, there was a 71% change in the photoacoustic signal over the above concentration range.
In vitro photoacoustic investigations of blood plasma and human whole blood samples within the glucose concentration range 0.5610 mmol/L (101800 mg/dL) have also been investigated (47)(53) and gave a linear correlation between the glucose concentration and the change in the photoacoustic response at 1180 and 1700 nm.
Preliminary in vivo studies have also been undertaken to assess the feasibility of photoacoustics as a noninvasive technique on humans (49).
Eight consenting volunteers were studied under a protocol, which was approved by the local ethics committee. Four nondiabetic subjects, two type 2 diabetics (non-insulin-dependent), and two type 1 diabetics (insulin-dependent) were studied.
The non-diabetic and type 2 volunteers fasted from the previous evening, and the type 1 diabetic volunteers omitted their morning insulin. Non-diabetic and type 2 diabetic volunteers received a standard 75-g D-glucose load orally. The blood glucose concentration of the type 1 diabetics was controlled by administration of insulin, and all subjects with higher than normal blood glucose concentrations were monitored until their blood glucose concentration fell to normal values. The subject was seated throughout the procedure with the right index fingers immobilized in a mount. The laser pulse was incident on the side of the finger through an optical fiber, and the photoacoustic signal was detected from the pad of the finger, which was in contact with the piezoelectric transducer.
A near-infrared pulsed laser source (30) was utilized in the in vivo studies. A piezoelectric transducer detected the photoacoustic pulses, and the resulting signals were acquired with a similar set-up as described in the in vitro studies discussed later. The mean photoacoustic signals were acquired over a 5-s period at intervals of 5 min.
Reference blood samples were taken at 10-min intervals, and blood glucose concentrations were measured using a Yellow Springs Instruments analyzer (YSI 2300G). In addition, blood glucose concentrations from the diabetic subjects were monitored periodically during the tests using an Exactech Companion 2 meter and test strips.
The photoacoustic data presented in Fig. 2
are based on the measurement
of the peak-to-peak amplitude of the photoacoustic signal, with a
correction for linear system drift, which may be attributable to
physiological factors as a result of the subject being immobilized
throughout the observations. The CV in the reference measurement
of the blood glucose concentration was 3%, and that in the
photoacoustic response varied between 1% and 5%, depending on the
initial instrument gain settings.
|
| In Vivo Results |
|---|
|
|
|---|
The results for all of the measurements, at 10-min intervals, when both
clinical and photoacoustic measurements were available for all eight
subjects are shown in Fig. 3
. This shows a linear correlation of 0.96 between the predicted
and the clinical glucose concentration.
|
We recognize that this is a favorable presentation of the results, and
a more realistic representation of the data can be seen through
analysis of the percentage of change of the predicted glucose
concentration from the clinical measurement. The results of this
analysis are presented in Fig. 4
, which shows that ~91% of the data points lie within ±
20% of the clinical glucose concentration, corresponding to zone A in
a Clarke (54) error grid analysis.
|
| Current In Vitro Studies |
|---|
|
|
|---|
A tunable MOPO-710 laser system (Spectra Physics), pumped by the third harmonic of a Nd:YAG laser, was used as the optical source. This provided nearly continuously tunable pulsed optical energy from 400 to beyond 2000 nm, with a pulse duration of approximately 7 ns at a repetition rate of 10 Hz. The optical output was launched into a 2-m length of optical fiber (diameter, 1000 µm), using a 6-cm focal length lens. The fiber output was directed through a beam splitter at an angle of 45° to create an energy monitoring channel, and each optical channel was focused into a photoacoustic cell using a 10-cm focal length lens. The sample to be analyzed was contained within the first photoacoustic cell, whereas the presence of a second reference photoacoustic cell provided energy monitoring. For the experiments under consideration here, the energy monitor was provided by the second photoacoustic reference cell containing distilled water, which allowed compensation for variations in incident optical energy and the photoacoustic response to the water spectra.
The voltage pulses from the piezoelectric detector were amplified with
a precision AC 9452 amplifier (1MHz bandwidth; 100 M
input
impedance; 1 k
output impedance; Brookdeal/EG & G). The resulting
signals were digitized by a 125 million samples/s, dual channel
CompuScope CS2125 PC plug in card (GaGe Applied Sciences), where
the signals were recorded and analyzed. Typically, one photoacoustic
measurement resulted from 100 pulses averaged in the software to reduce
random noise.
Glucose solutions were prepared by dilution of a 1.67 mol/L (30 g/dL) D-glucose stock solution. The glucose concentrations were also cross-referenced with an AbbottVisionTM clinical chemistry analyzer (Abbott Laboratories).
The spectral region of 800-1200 nm was chosen for these investigations after initial work showed the largest changes with glucose concentration occurred at ~1040 nm.
Photoacoustic responses from the sample and baseline water samples were
measured alternately to separate long-term system drifts from changes
attributable to the sample concentration. The spectrum of distilled
water and a 500 mmol/L (9 g/dL) glucose solution can be seen in Fig. 5
. To investigate the changes from the water spectra, the
percentages of difference of all the spectra from the baseline water
spectrum were calculated. In each of the following studies, the
letters a, b, c, d, and e indicate the sequence in which the spectra
were obtained.
|
The results of this procedure for the glucose spectrum of Fig. 5
can be
seen in Fig. 6
. The absolute values of the peak wavelengths are inexact in
this type of low-resolution experiment, but spectral features may be
noted that relate to established spectroscopy of glucose in solution
(4). The greatest percentage of change in the photoacoustic
response from the glucose sample is in the region of the C
H second
overtone at 1126 nm, with a further peak in the region of the second
O
H overtone at 939 nm. These assignments are tentative at this stage
and require further investigation. Similarly, in Fig. 6
, the reduction
in photoacoustic response at 825 nm is within the resolution of the
experiment and may be attributable to a displacement effect, but this
also requires further investigation.
|
To investigate the possible contributions of other blood analytes to
the total photoacoustic response from blood, solutions of sodium
chloride, cholesterol, and BSA were investigated. In this preliminary
study, the percentages of change of the photoacoustic signals of the
analytes compared with a baseline from water were investigated within
the same spectral region as the glucose study. As can be seen from the
results, sodium chloride (Fig. 7
), cholesterol (Fig. 8
), and BSA (Fig. 9
) each exhibited a different photoacoustic response.
|
|
|
The photoacoustic response of the sodium chloride solution was
distinctly different from the corresponding response of the glucose
solution, as shown in Fig. 6
. The sodium chloride solution gave a
significant increase in the photoacoustic response from 800 to 975 nm.
The peak of this response coincides with an overtone band in the water
spectrum at 970 nm (55), and the dissolution of NaCl is
known to cause a shift of such water bands to longer wavelengths
(56). It may also be that the changed physical parameters
related to the presence of sodium chloride enhance such changes.
The difference in the cholesterol photoacoustic response showed the least spectral features of all four samples, with a basically flat response. However, the sensitivity and resolution in this experiment may be insufficient to show the weak features that have been observed in previous work (57).
The photoacoustic response from the BSA solution showed a small feature at 1050 nm and a significant increase in the photoacoustic response at wavelengths below 900 nm that had not been observed in previous studies (58) and may be attributable to coloration or other contaminants in the sample.
This initial study is a precursor to the detailed investigation that will be required before selecting a set of wavelengths on a clear spectroscopic basis that, with appropriate algorithms or other analysis to compensate for the background response from other analytes, may allow the unique determination of glucose concentrations.
| Discussion |
|---|
|
|
|---|
|
Because of the complicated nature of both the photoacoustic generation process and the structure and composition of human tissue, it is inevitable that further analysis techniques will be required to eliminate these influences. At present, the analysis of data is carried out using the peak-to-peak value of the photoacoustic response, although the entire signal is recorded. Other methods are under consideration that use the recorded temporal signal, such as Fourier and wavelet analysis, as well as general data analysis techniques including chemometrics. The absolute noninvasive determination of blood glucose concentrations may also require the combination of the photoacoustic technique with other optical techniques or physical measurements.
| Conclusion |
|---|
|
|
|---|
Current work is focusing on improving the repeatability and sensitivity of photoacoustic measurement to that required by the diabetic community as well as investigating possible detection sites on the human body and sensor design geometry for such body sites.
The overall conclusion from this initial study is that although the photoacoustic method has several promising aspects, it is possible that additional data may be required from the fund of interesting work in other areas that is being pursued in the challenging quest for the noninvasive measurement of blood glucose.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
K. V. Larin, M. S. Eledrisi, M. Motamedi, and R. O. Esenaliev Noninvasive Blood Glucose Monitoring With Optical Coherence Tomography: A pilot study in human subjects Diabetes Care, December 1, 2002; 25(12): 2263 - 2267. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. B. Sacks, D. E. Bruns, D. E. Goldstein, N. K. Maclaren, J. M. McDonald, and M. Parrott Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus Clin. Chem., March 1, 2002; 48(3): 436 - 472. [Abstract] [Full Text] [PDF] |
||||
![]() |
R Swaminathan and M Wheeler Robotics into the millennium J. Clin. Pathol., January 1, 2000; 53(1): 22 - 26. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |