Dried blood spots (DBSs) are used to screen newborns for phenylketonuria and other aminoacidopathies. The calibrators for this testing are usually DBSs with values for Phe. Two DBS reference materials have been prepared, the European Working Standard for Phe (EWS-Phe-01) (1) and the amino acid reference material (AARM) from the CDC (2). The two reference materials are not interchangeable because they differ in blood hematocrit, blood-spot size, and filter paper, each of which (3)(4)(5)(6) affects analyte recovery. We measured quantitatively the effects of these differences on analyte recovery from DBSs and used results from our measurements to predict expected Phe recoveries from tandem analyses of the two sets of materials.
In EWS-Phe-01 (1)(7), human blood with a 50.5% hematocrit and intact red cells was divided into five portions for enrichment with 0, 20, 40, 80, and 120 mg Phe/L blood (0, 120, 240, 480, and 720 µmol/L blood). The liquid added during enrichment (7) was sufficient to reduce the hematocrit to 50.1%. The Phe-enriched blood portions were dispensed in 35-µL aliquots (7) onto Schleicher & Schuell (S&S) Grade 2992 (lot no. 121576) filter paper (1).
The AARM was prepared (2) by dividing human blood with a 57% hematocrit and intact red cells into six portions for enrichment with pure amino acids to cover the usual analytic ranges of Phe, Tyr, Leu, Met, and Val. The Phe enrichments were 0, 40, 80, 120, 160, and 200 mg Phe/L blood (0, 240, 480, 720, 960, and 1200 µmol/L blood). The liquid added during enrichment was sufficient to reduce the hematocrit to 53%. The whole-blood pools were dispensed in 100-µL portions onto S&S Grade 903 (lot no. W941) filter paper with dashed-line 13-mm printed circles (2).
To examine the effect of blood hematocrit on Phe recovery, we prepared whole-blood portions with 40%, 45%, 50%, 55%, 60%, 65%, and 70% hematocrits from a single batch of packed erythrocytes and a single batch of clarified serum. We enriched the hematocrit-adjusted blood portions with 125I-labeled thyroxine, dispensed 25 100-µL aliquots of each portion onto S&S Grade 903 (lot no. W961) filter paper, and punched ~3 mm (1/8-inch) disks from the north, east, south, west, and center of each dried spot to determine the mean blood absorption volume per disk (8) at each hematocrit. We performed a simple linear regression analysis of blood volumes per disk vs hematocrit and used the resulting regression line slope to predict the blood volume of a ~3-mm disk from blood with hematocrits of 53% (the calculated hematocrit of the AARM) and 50.1% (the calculated hematocrit of the EWS-Phe-01).
The EWS-Phe-01 materials were dispensed in 35-µL spots (spot size range, 3340 µL); the AARMs were dispensed in 100-µL (± 0.24%) spots. To compare Phe recoveries from spot volumes representative of the two sets of materials, we used a single batch of whole blood, adjusted to 55% hematocrit and enriched with 80 mg Phe/L blood, to dispense 35- and 100-µL blood volumes onto S&S Grade 903 (lot no. W941) filter paper. We punched an ~6 mm (1/4-inch) disk from the center of 20 dried spots of each blood volume and measured the Phe concentrations of all of the punched disks in a single HPLC run that was performed according to previously described protocols (2)(9)(10).
To compare the blood absorption characteristics of the S&S filter
papers that were used to prepare the EWS-Phe-01 and AARM, we dispensed
100-µL spots of 125I-thyroxine-enriched blood
with a 55% hematocrit onto clean, unprinted areas of 10 cards taken
from each set of materials. We punched ~3-mm disks from the north,
east, south, west, and center of one spot per card for gamma counting
in a single analytic run and used a standardized method (8)
to equate
counts to the serum volume contained in each DBS disk. We
used statistical analyses of the counting data to determine the mean
serum-absorption volume per disk for each paper.
We used analyte recoveries computed from our examinations of hematocrit effects, blood-volume-per-spot effects, and filter-paper serum-absorption volumes to predict expected differences between Phe recoveries from the EWS-Phe-01 and the AARM. To evaluate the reliability of the Phe concentration values that were predicted from measurements of matrix variables, we analyzed the EWS-Phe-01 and the AARM in duplicate in each of five HPLC runs and compared the mean values of the measured Phe concentrations with the predicted Phe concentrations. We also compared regression slopes, derived from measured vs enriched Phe concentrations of each set of DBS materials, to show the ratio of their measured Phe recoveries.
Studies of hematocrit effects showed that the blood volume per ~3-mm disk was positively correlated with the hematocrit. The slope of the regression line, derived from measured blood volume per disk vs hematocrit, predicted that the EWS-Phe-01 materials, with a calculated hematocrit of 50.1%, should have a blood volume of 3.1 µL per ~3-mm disk, whereas the AARM, with a 53% calculated hematocrit, should have a blood volume of 3.2 µL per disk.
In studies of the effect of blood volume per spot on analyte recovery, the total Phe concentration of the blood used was equal to the endogenous Phe, which was not measured, plus the Phe enrichment. The mean of recovered Phe concentrations from the 35-µL spots (85.1 ± 4.7 mg/L blood) was lower than that from the 100-µL spots (95.5 ± 9.1 mg/L blood; P <0.01, Student t-test).
The mean serum-absorption volume per ~3-mm disk punched from the S&S Grade 2992 paper used to prepare the EWS-Phe-01 was 1.250 ± 0.104 µL, whereas that from the S&S Grade 903 paper used to prepare the AARM was 1.502 ± 0.188 µL (P <0.01).
The controlled comparisons of hematocrits, blood volumes per spot, and
filter-paper serum-absorption volumes showed, in all cases, that
analyte recovery per DBS disk was lower for the conditions used in the
preparation of EWS-Phe-01 than from those used in the preparation of
AARM. By summing these observed differences, we projected (Table 1
) that analyte recovery from EWS-Phe-01 and AARM could be
expected to differ by 30.8% when the two sets of materials were
analyzed in tandem. By comparing the regression slopes derived from
HPLC-measured vs enriched Phe concentrations of each set of DBS
materials, we found that observed Phe recoveries from EWS-Phe-01
(regression slope, 0.698) were 31.6% lower than those from AARM
(regression slope, 1.020) when the two sets of reference materials were
analyzed in tandem against the same set of calibrators. Our HPLC
calibration and data-reduction protocol was that provided by the
manufacturer to newborn-screening laboratories in the United States.
This protocol yields ~100% Phe recovery from 100-µL spots of
intact-cell blood dispensed onto S&S Grade 903 filter paper; therefore,
Phe recoveries from the AARM were expected to equal the AARM target
values (2). Conversely, in an analytic system with
calibration and data-reduction protocols appropriate for European
newborn-screening samples, Phe recoveries from the EWS-Phe-01 would be
expected to equal the EWS-Phe-01 target values (1).
|
The effects of the differences in the blood hematocrit, blood volume per spot, and filter-paper sources used to prepare the EWS-Phe-01 and AARM yielded analytic recovery differences of 3.1%, 10.9%, and 16.8%, respectively. These relationships between blood-spot preparation variables and analytic recovery illustrate the importance of (a) preparing reference materials for newborn-screening tests from blood with a hematocrit typical of newborns and in spot sizes similar to those of the newborn-screening specimens and on the same filter paper grade and lot number used for collecting the newborn-screening samples, and (b) considering the blood-spot preparation variables when comparing different sets of blood-spot reference materials.
The EWS-Phe-01 and AARM were prepared from blood with hematocrits typical of newborns and with filter papers and blood-spot sizes that reflect newborn-screening practices in the regions in which they are used. Because of the observed matrix-related differences in analyte recoveries, we predicted that Phe recoveries from tandem analyses of the EWS-Phe-01 and AARM would be 30.8% higher from the AARM. In fact, measured Phe concentrations from the AARM were 31.6% higher than those from the EWS-Phe-01 when the materials were analyzed in tandem. We conclude that controlled measurements of blood-spot preparation variables can be used to reliably predict analyte recoveries from DBS materials, and we have shown that the Phe contents of the EWS-Phe-01 and AARM are concordant when the effects of their preparation variables are normalized; thus we verified the suitability of the materials as calibrators for their respective regions.
The effects of blood-spot preparation variables are not limited to phenylketonuria tests or to newborn screening. Investigators who use DBSs must be aware of the variables that affect blood-spot test results. Knowledge of the relationship between reference materials and DBS test samples and awareness of the relationships among different sets of reference materials are essential for evaluating test results, comparing data among laboratories, and evaluating laboratory performance in different areas of the world.
References
The following articles in journals at HighWire Press have cited this article:
![]() |
D. H. Chace, T. A. Kalas, and E. W. Naylor Use of Tandem Mass Spectrometry for Multianalyte Screening of Dried Blood Specimens from Newborns Clin. Chem., November 1, 2003; 49(11): 1797 - 1817. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. W. McDade and B. Shell-Duncan Whole Blood Collected on Filter Paper Provides a Minimally Invasive Method for Assessing Human Transferrin Receptor Level J. Nutr., December 1, 2002; 132(12): 3760 - 3763. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. H. Chace, J. C. DiPerna, T. A. Kalas, R. W. Johnson, and E. W. Naylor Rapid Diagnosis of Methylmalonic and Propionic Acidemias: Quantitative Tandem Mass Spectrometric Analysis of Propionylcarnitine in Filter-Paper Blood Specimens Obtained from Newborns Clin. Chem., November 1, 2001; 47(11): 2040 - 2044. [Full Text] [PDF] |
||||
![]() |
D. H. Chace, J. C. DiPerna, B. L. Mitchell, B. Sgroi, L. F. Hofman, and E. W. Naylor Electrospray Tandem Mass Spectrometry for Analysis of Acylcarnitines in Dried Postmortem Blood Specimens Collected at Autopsy from Infants with Unexplained Cause of Death Clin. Chem., July 1, 2001; 47(7): 1166 - 1182. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. V. Mei, J. R. Alexander, B. W. Adam, and W. H. Hannon Use of Filter Paper for the Collection and Analysis of Human Whole Blood Specimens J. Nutr., May 1, 2001; 131(5): 1631S - 1636. [Abstract] [Full Text] |
||||
![]() |
D. H. Chace, J. C. DiPerna, B. W. Adam, and W. H. Hannon Errors Caused by the Use of D,L-Octanoylcarnitine for Blood-Spot Calibrators Clin. Chem., April 1, 2001; 47(4): 758 - 760. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |