|
|
||||||||
Articles |
1
Amersham Pharmacia Biotech, SE-751 84 Uppsala, Sweden.
2
Department of Surgery, Uddevalla Hospital, SE-541 80
Uddevalla, Sweden.
3
Department of Surgery, University Hospital of Northern
Sweden, SE-901 85 Umeå, Sweden.
a Author for correspondence. Fax 46-18-6121826; e-mail sara.byding{at}eu.apbiotech.com
| Abstract |
|---|
|
|
|---|
Methods: The sensitivity, specificity, and positional accuracy of mutation detection by EMD with the PASSPORTTM Mutation Scanning Kit were assessed in a blind fashion for three analytical platforms (radioactive detection and automated laser sequencers ALFexpress and ABI PRISM 377). PCR products of 703 bp covering codons 188393 of the P53 gene were prepared from colorectal tumor samples and analyzed by EMD; the results were compared to data from cDNA sequencing. A 1362-bp PCR product prepared from IL4r gene was used to test detection of multiple base changes in long PCR products.
Results: The sensitivity for detection of mutations using EMD exceeded 90%, and the specificity exceeded 80% on all analysis platforms. The method localized 90% of mutations to within two codons and four codons for automated laser sequencers and detection by radioactivity, respectively. The method detected at least five mismatches in heteroduplexes >1 kb.
Conclusions: The EMD system facilitates efficient detection of genetic variation in fragments exceeding 1 kb irrespective of location and type. The technology is particularly well suited to the detection of mutations in genes frequently mutated at unpredictable locations.
| Introduction |
|---|
|
|
|---|
Sequencing has been the gold standard for detection of genetic variability. The method has been used, for example, to detect mutations, insertions, and deletions in the P53 gene (the model gene used in this study) in the form of solid-phase sequencing of DNA and cDNA (1)(2)(3) and cycle sequencing (4). Automated sequencing equipment can also be used to specifically detect insertions and deletions (5). Although very accurate in most situations, sequencing generates large amounts of redundant information at considerable cost when used to detect genetic variability at nucleotide frequencies of 1 in every 1001000 bp. Efficient methods for rapid scanning of genes or exons of genes for the presence of genetic variation, combined with sequencing to generate exact information of the genetic variation, could drastically reduce the cost of analysis.
The number of methods that may be used for detection of genetic variability is growing steadily. Most methods are applicable only in situations where the location and type of variability have been defined. These include the allele-specific oligonucleotide assay (6), the oligo-ligation assay (7), minisequencing (8)(9), TaqMan® (10), InvaderTM (11), pyrosequencing (12), and PCR-restriction fragment length polymorphism analysis (13).
Other methods that can detect genetic variability irrespective of the location and type of aberration are less common and, in general, less reliable. These methods include heteroduplex analysis (14), single strand conformation polymorphism (SSCP) (15), denaturing gradient gel electrophoresis (16), chemical cleavage (17), dideoxy fingerprinting (18), and denaturing HPLC (dHPLC) (19). Some of these methods have been reviewed recently (20).
A new method for detection of genetic variability, enzyme mutation detection (EMD), has recently become available under the trade name PASSPORTTM Mutation Scanning (Amersham Pharmacia Biotech, Uppsala, Sweden). EMD is based on enzymatic cleavage of mismatch-containing heteroduplexes formed by hybridizing PCR products generated from an unknown sample and a wild-type reference. Heteroduplexes are cleaved by the resolvase T4 endonuclease VII, usually in both strands within 36 bp 3' from the mismatch (21). Cleavage products are then detected by electrophoretic separation. The lengths of the fragment(s) generated indicate the approximate location of the mismatch, which can be confirmed by sequencing.
EMD can be used for analyzing fragments up to several kilobases in size, depending on the resolution of gel electrophoresis. At least five mismatches can be detected in a single heteroduplex. The method can also detect insertions and deletions involving one or more bases. Furthermore, the high sensitivity of EMD vs many other techniques can be used for detection of low abundance mutations in clinical specimens or to reduce the cost for analysis of SNPs by mixing samples while maintaining the probability of detecting genetic variation (22).
The functional and biochemical properties of T4 endonuclease VII have been studied in great detail (23)(24). Recently, the x-ray structure was determined (25), contributing to the understanding of the structure-function relationship of the enzyme. T4 endonuclease VII has been used for mutation detection (26)(27)(28), but its limited availability has hampered its utilization for routine application.
EMD may serve as a scanning procedure to identify aberrant samples and to get an indication of the approximate location of the aberration. This information can be used to simplify confirmatory sequencing in a limited region of the sample sequence. The use of EMD as a scanning method can therefore substantially reduce the amount of sequencing work, depending on mutation prevalence and the length of the fragment that is analyzed. The success of this strategy, however, depends on the ability of EMD to detect aberrant samples with high sensitivity and specificity.
This study was undertaken to determine whether PASSPORT has the sensitivity, specificity, and positional accuracy required of a scanning procedure to be used before confirmatory sequencing of samples containing mismatches caused by mutations. As a model system, tumor samples from patients with colorectal cancer were analyzed using PASSPORT on three analytical platforms (ABI PRISMTM 377, ALFexpressTM, and radioactive detection) for the presence of aberrations in a fragment of the P53 gene comprising codons 188393. These data were compared with data generated previously by cDNA sequencing (2), and the accuracy in terms of sensitivity, specificity, and positional accuracy was determined. All comparisons were performed in a blind fashion. This approach also allowed an assessment of the accuracy of cDNA sequencing data generated by independent PASSPORT analysis on three analytical platforms.
| Materials and Methods |
|---|
|
|
|---|
Briefly, RNA was prepared from frozen tumor samples using the
QuickPrepTM Total RNA Extraction Kit (Amersham Pharmacia
Biotech) according to manufacturers instructions, under
stringent conditions to avoid degradation and contamination. This
procedure was followed by an enzymatic synthesis of cDNA, using RNA as
template as described previously (3). Codons 188393,
extending into the flanking nontranslated 3' region, were amplified
from the cDNA by PCR with two overlapping primer pairs (Table 1
), to give fragments 3 (primers PNT3-1 and BK-1; codons
188348; 482 bp) and 4 (primers PNT4-3 and PB4-8; codons
300393; 445 bp). One of the primers in each primer pair was
modified with biotin. Thus, biotin-labeled PCR products were generated
to enable solid-phase sequencing (29). A manifold-based
version of solid-phase sequencing was used
(AutoLoadTM; Amersham Pharmacia Biotech)
essentially as described elsewhere (30). The sequencing
products generated using the primers PC3-6 (fragment 3) and PC4-S1
(fragment 4) were analyzed using an automated laser fluorescence
sequencer (ALFexpress; Amersham Pharmacia Biotech). The sequence
was finally compared with the P53 wild-type sequence by a
prototype software program, p53 SBDecipher, Ver. 1.00 (Amersham
Pharmacia Biotech). Nucleotide changes that had an impact on the
protein were considered as mutations. Each mutation was verified by
sequencing an entirely new PCR product from the corresponding cDNA.
Additional confirmation was obtained by analysis of the neighboring PCR
fragment when the mutation was located in an overlapping segment.
|
selection of clinical specimens for comparative study using emd
The study included 178 tumor samples from patients with colorectal
cancer (2) from which there were cDNA remaining from
previous cDNA sequencing. The study was approved by the Ethical
Committee, University of Uppsala. To maintain the blind nature of the
comparative study, samples were given a new identity before the first
round of PCR. A new unique identity was also given for each sample and
each analytical platform before the second round of PCR to generate
fragments intended for PASSPORT reactions. The integrity of the sample
code remained unbroken until results on all samples for each platform
were reported. This is referred to as the "blind phase of the
study" as opposed to when the sample codes had been broken.
pcr
To save cDNA material, amplification was done in a nested fashion,
generating the full-length open-reading frame of the
P53 gene in the first PCR round. In the second round of
amplification, the region covering codons 188393 and the flanking
nontranslated 3' region was amplified to give a PCR product of 703 bp,
called fragment 34.
Amplification of "full-length" product from cDNA.
The PCR
product covering the complete open-reading frame of P53 was
generated using the following mixture in 100 µL: 1x GeneAmp PCR
buffer II (PE Biosystems), 15 pmol each of primers PN8 and PN9, 1.5
mmol/L MgCl2, 200 µmol/L dNTPs (Amersham
Pharmacia Biotech), 4 U of AmpliTaq Gold (PE Biosystems), and 2 µL of
cDNA template. Thermocycling using touchdown was performed on a GeneAmp
9600 (PE Biosystems) as follows: 95 °C for 9 min; 10 cycles of
95 °C for 30 s, 72 °C for 30 s, and 72 °C for
45 s; 10 cycles of 95 °C for 30 s, cooling from 72 °C
to 67 °C at 0.5 °C/cycle for 30 s, followed by 72 °C for
45 s; 20 cycles of 95 °C for 30 s, 66 °C for 30 s,
and 72 °C for 45 s; 72 °C for 5 min; and cooling to
4 °C.
Amplification of fragment 34.
The PCR product was generated
using the following mixture in 100 µL: 1x GeneAmp PCR buffer II, 15
pmol each of primers 34-1 and 34-2, 1.5 mmol/L
MgCl2, 200 µmol/L dNTPs, 4 U of AmpliTaq DNA
Polymerase, and 2 µL of the long PCR product obtained from the
first round of amplification. The primers were modified for use on
automated sequencers as follows: for the ABI 377,
hexachloro-6-carboxyfluorescein (HEX)-labeled 34-1 and
6-carboxy-fluorescein (FAM)-labeled 34-2 were used; for the ALFexpress,
both primers labeled with Cy5. AmpliWax Gem 100 (PE Biosystems) was
used to separate reaction components, thereby enabling "hot start"
PCR. Cycling conditions were as follows: 40 cycles of 95 °C for
15 s, 66 °C for 30s, 72 °C for 45 s; followed by
72 °C for 5 min, and cooling to 4 °C.
evaluation of pcr products by silver staining
PCR products (4 µL) were separated on GeneGel 24/12.5 (Amersham
Pharmacia Biotech) according to the manufacturers instructions and
using GenePhor (Amersham Pharmacia Biotech). The gels were stained in a
Hoefer Autostainer using DNA silver stain (Amersham Pharmacia Biotech).
emd analysis
PASSPORT reactions were prepared for analysis of fragment 34 (703
bp) of the P53 gene covering codons 188393 on the ABI
PRISM 377, using 32P and ALFexpress II,
respectively, for the presence of mutations, according to
manufacturers instructions. Codons 205393 were considered
informative in PASSPORT analysis.
mutation databases
BRCA1 gene-related data and the mutation spectrum
were retrieved from the National Human Genome Research Institute,
Division of Intramural Research, Breast Cancer Information at
http://www.nhgri.nih.gov/intramural_research/lab_transfer/bic/member/brca1_mutation_database.html.
| Results |
|---|
|
|
|---|
|
samples available for analysis by passport
Depending on the success rate in the PCR, various numbers of
samples fulfilled the criteria set to qualify for PASSPORT analysis on
individual platforms (Table 2
). For assessment of the cDNA sequence data, the number of
samples fulfilling inclusion criteria on all analytical platforms was
further reduced to 98.
|
mutation detection using passport mutation scanning on three
different analytical platforms
Preliminary analysis.
Initially, we attempted to amplify the
desired PCR fragment on 178 samples. After PCR products not fulfilling
quality and quantity requirements for further analysis were eliminated,
143 remained for analysis on the ABI PRISM 377, 155 remained for
analysis using 32P, and 114 remained for analysis
on the ALFexpress.
The results shown in Table 3
indicate that there is a good agreement between PASSPORT
analysis on different analytical platforms and cDNA sequencing in terms
of sensitivity and specificity for mutation detection. If it is assumed
that cDNA sequencing data represent the true mutation spectrum, then
the sensitivity for detection of mutations, i.e., the ability to
accurately detect mutations, was 90.5100%. Similarly, the
specificity, i.e., the ability to accurately detect wild-type sequence,
was 78.686.4%.
|
Four samples that were interpreted as wild-type in the PASSPORT assay for ABI PRISM 377 did in fact contain mutations according to sequencing. When reexamined, three of the samples did contain weak cleavage products corresponding to the expected fragment sizes when compared with cDNA sequence data but were not registered as mutated during the blind phase of the study. All samples containing mutations according to sequencing were correctly identified as mutated when PASSPORT for 32P was used.
Three samples that did contain mutations according to cDNA sequencing were interpreted as wild-type by PASSPORT assay for ALFexpress.
Typical results from PASSPORT analysis of the three analytical
platforms are shown in
Figs. 24
.
|
|
|
Detailed comparison of cDNA sequence data.
In all comparisons,
several samples were initially classified as mutant in PASSPORT but
wild-type by cDNA sequencing. This could be attributable to either
false-positive results in the PASSPORT analysis or false-negative
results in the cDNA sequencing. To determine which was the case, the
following analysis was undertaken.
In total, 98 samples generated results on all PASSPORT platforms. The
mutational status of those samples found to be mutated with respect to
cDNA sequencing and EMD is shown in Table 4
. Among these, samples fulfilling the following criteria were
identified:
|
Six samples fulfilled these selection criteria on all three
analytical platforms: samples 21C10, 17H7, 17J4, 17J5, 17E9, and 17I7.
When the cDNA sequence data on the six samples were reassessed, two of
the six samples were confirmed to have mutations close to the locations
indicated by PASSPORT analysis. One sample was incorrectly reported as
wild-type despite the presence of a clear mutation in the sequence
data. The other sample contained a mutation at the borderline of
detection for cDNA sequencing, and when reevaluated was found to be
mutated. Considering these results, we decided that these two samples
(21C10 and 17H7) could be reclassified as mutated. The resulting
specificities and sensitivities are shown in Table 3
(values in
parentheses). The four remaining samples might represent false
negatives in sequencing (see the Discussion), although this
was not confirmed in the present study. This would give sensitivities
and specificities, respectively, of 91.6% and 85.3% for ABI PRISM
377, 100% and 91.3% for 32P, and 91.7% and
84.6% for ALFexpress (detailed analysis not shown).
positional accuracy of passport mutation scanning
The positional accuracy of PASSPORT Mutation Scanning was
evaluated by comparing mismatch locations generated from PASSPORT
analysis with data from cDNA sequencing on samples displaying single
point mutations; this was a subset of the "true positives" shown in
Table 3
. The approximate locations of detected mismatches were
calculated from the sizes of the complementary cleavage fragments in
PASSPORT and were expressed as the codon (ABI PRISM 377) and the
alternative codons (ALFexpress and 32P)
identified to contain a mismatch. Finally, the deviation in location
between cDNA sequencing and PASSPORT analysis, respectively, was
calculated. The results are shown in Fig. 5
.
|
potential of passport mutation scanning
To illustrate the potential advantages of using PASSPORT for
detection of genetic variability, an experiment was performed on a
relatively long PCR product containing several mismatches. Furthermore,
in situations where few samples are to be analyzed for genetic
variability, alternative labeling procedures may be used, e.g., by
incorporating labeled dUTPs during PCR. To illustrate this, a 1362-bp
fragment corresponding to a part of the IL4r gene from an
individual heterozygous at six different positions was prepared by
incorporation of R110-dUTP during PCR (31).
The fragment was analyzed in a blind fashion using PASSPORT for the
presence of mismatches according to manufacturers instructions for
heterozygous samples. The results are shown in Fig. 6
and Table 5
.
|
|
| Discussion |
|---|
|
|
|---|
To eliminate any possible operator influence, the study was performed in a blind fashion. Thus, samples were recoded before PCR amplification, and none of the operators and evaluators of EMD results knew the results from any other analytical platforms or previously performed cDNA sequencing.
The accuracy of PASSPORT for detection of samples harboring mutations was clearly acceptable. The sensitivity and specificity exceeded 90% and 80%, respectively, on all platforms. We addressed two essential questions when we assessed the accuracy of two methods: (a) Are the samples homogeneous regarding P53 mutation status? (b) If not, are the two methods equally sensitive in detecting low-abundance P53 mutations in tumors?
DNA sequencing is considered the gold standard for mutation analysis. At first sight it seems quite simple to implement large-scale DNA sequencing to identify genetic variation of any type and at any possible location. However, DNA sequencing is not free from weaknesses and limitations. For example, the analytical sensitivity is limited to 2530% at best (32), which hampers its utility when somatic mutations are studied in cases where a mutation may be diluted below the threshold of detection of sequencing in heterogeneous tumors. Another limitation is the maximum length of template that can be reliably sequenced for the detection of heterozygotes, which seldom extends beyond 500 bases. Furthermore, the primary application of DNA sequencing technology is in de novo sequencing, with some 9899% accuracy. Even in the very best case, i.e., when genomic DNA is used as sample source, this means that in a reasonably large gene of 1000 bases, a significant number (1020) of false-positive base calls are generated, suggesting presence of mutations that require reanalysis to be resolved. When somatic mutations are analyzed in heterogeneous sample material, this problem will increase further.
One attempt to address the question of cDNA sequencing accuracy
involved independent analysis of samples on three different PASSPORT
platforms. An identical preparation of cDNA for each individual tumor
sample was used for cDNA sequencing as well as for all PASSPORT
platforms. The PCR products analyzed on PASSPORT were prepared by
nested PCR. Any artifact that could occur during this procedure that
introduced mutations in the full-length P53 product would
affect subsequent analysis on the different PASSPORT platforms and the
outcome of the study. Guided by the data on localization of mutations,
however, we found very good agreement between PASSPORT on the three
analytical platforms, respectively, and cDNA sequencing on samples
carrying mutations. Furthermore, data generated on the three
PASSPORT platforms suggested the presence of mutations in six
samples, of which two were confirmed by reevaluating cDNA sequence data
(Table 4
). Taken together, we found no evidence indicating that the PCR
procedure in this set of samples contributed to the observed
discrepancies between PASSPORT and cDNA sequence data.
In a detailed study, Finkelstein et al. (33) investigated the expression of p53 protein as a tool to characterize the potential heterogeneity of colorectal tumors. They found several typical expression patterns of p53 protein in a minor proportion of tumors, as determined by p53 immunohistochemical analysis and subsequent DNA sequencing of exons 58, suggesting in several cases that there were two expanding cell populations with different p53 mutational status. If tumors were heterogeneous, the detection of mutations in tumors would depend on the analytical sensitivity of the technologies used. Recent data suggest that the analytical sensitivity of EMD technology may extend below 5% of mutated vs wild-type PCR products (22), which is superior to the 2530% that may be possible with DNA sequencing (32).
Given the experimental data available, we cannot exclude the
possibility that mutations are present at low abundance in samples.
These might be detectable in PASSPORT but not by cDNA sequencing, thus
affecting the specificity slightly, as suggested by the results
presented here. Thus, it is quite possible that four samples (17J4,
17J5, 17E9, and 17I7 in Table 4
), for which EMD on all three platforms
had given comparable positive results, might be false negatives in
sequencing because of low abundance of mutated cells in the sample.
Confirmation of this interpretation would demand analysis by an
independent detection method. We conclude, however, that the
specificity of P53 mutations determined by PASSPORT on
samples originating from colorectal tumors might be underestimated
compared with a similar study performed on breast tumors
(22).
There are several circumstances that decide the choice of technology for the detection of genetic variation. In the first situation, the analytical target has been predefined regarding location and type, which permits the design of tests to classify unknown samples in two or more categories.
The other situation is when genes or parts of genes contain variations at unpredictable locations or in a region that is poorly characterized (e.g., scanning for SNPs) or that contains sporadic mutations, such as the P53 gene. We advocate the use of PASSPORT in this situation. In the case of mutations, a single nucleotide base change at a frequency of 1:1000 or 1:10 000 and located in the coding region of a gene could be sufficient to alter, and possibly completely abolish, the normal function of a protein. It is important to identify any such mutation irrespective of its location and type. Omission of any part of a gene in the analysis can be justified only if the mutation spectrum has been well characterized after comprehensive analysis of many samples and the clinical implications have been evaluated thoroughly. Thus, in most cases, the complete coding region and possible splice sites of a gene should be analyzed to identify any type of aberrations located at any nucleotide position.
When new disease-causing genes are discovered, the initial mutation spectrum often is rather simple. That is sometimes true even after extensive studies, as in the case of the ras gene and colorectal cancer (34). However, it is common that the number of mutations increases as more affected patients are investigated, often producing a rather complex spectrum. The P53 (10 exons, 393 codons) and BRCA1 (22 exons, 1893 codons) genes, for example, harbor mutations at unpredictable locations. More than 700 sequence variants (as of September 1999) have been reported in the BRCA1 gene, of which most have been classified as mutations. Henceforth, all detected mutations in these genes should be regarded as potentially pathogenic because their biological implications are difficult to establish until large clinical studies have been conducted.
A commonly used technique for the detection of mutations is SSCP (15), which is based on aberrant electrophoretic behavior of single-stranded DNA when mutated is compared with wild-type. SSCP has attracted much attention because of its simplicity and low cost. SSCP has been claimed to reveal 7095% of mutations in a fragment of 200 bp (35)(36). Beyond this size of fragment, the sensitivity of mutation detection decreases rapidly (37). Furthermore, SSCP does not reveal the location of a mutation; it only has the potential to identify aberrant fragments from normal. Very few studies have been conducted to independently compare the performance of SSCP with DNA sequencing. Recently, however, such a study revealed that only 62% of mutations were detected by SSCP when analyzed exon-wise compared with DNA sequencing. The authors concluded that SSCP analysis could not be used reliably to screen for P53 mutations (38). Another weakness of SSCP is its inability to correctly classify fragments harboring infrequent mutations and, concomitantly, an SNP at reasonable frequency. The presence of SNPs will confound correct identification of samples containing mutations.
A more recent technology for the detection of DNA variability is dHPLC (14), which utilizes the aberrant behavior of mismatch-containing heteroduplexes compared with wild-type homoduplexes when separated by ion-pair reversed-phase HPLC under partially denaturing conditions. dHPLC has been claimed to efficiently detect mutated samples, using a highly automated system with limited resource requirement (39). Most blind studies published to date have used moderately sized fragments (<600 bp), possibly because of the ready availability of existing material for evaluation. In one recent study, only 8 of 42 different amplicons exceeded 350 bp in size (40), and in another study the median size of amplicons tested was 280 bp (41). Although these studies indicate high accuracy, the performance of dHPLC should be interpreted relative to the methods overall productivity, including the amplicon size.
In contrast, in another study PCR fragments averaging 300 bp covering
106 genes were analyzed for genetic variability by dHPLC
(42). The technique gave
40% false-positive reactions
and identified only 87% of those SNPs previously detected by
sequencing
PASSPORT, on the other hand, is capable of analyzing substantially larger fragments and has been shown to maintain high accuracy even when 2.2-kb amplicons are analyzed (43). This permits full exploitation of gene structure in terms of exon/intron composition. For example, some exons in BRCA1 have a size of 34 kb. If cDNA is prepared from samples, exons can be analyzed in a contiguous segment of several kilobases. PASSPORT has other advantages over dHPLC: it localizes the variation and can detect at least five mismatches in a fragment. This is particularly important when it is used for mutation analysis in fragments containing several SNPs, a situation that might be difficult to resolve accurately by dHPLC.
In conclusion, PASSPORT has properties that fulfill several
requirements for methods to detect genetic variation. The more
prominent features are (a) it has high accuracy for
detection of genetic variation; (b) it can analyze large
DNA fragments (
2.2 kb); (c) it gives the approximate
location of the genetic variation (90% within 2 codons);
(d) it resolves genetic variation at several locations
in a fragment; (e) it has high analytical sensitivity,
detecting mutations in low abundance; and (f) one
protocol is used for all fragments. PASSPORT also displays some
weaknesses : (a) it requires PCR products of reasonable
quality and quantity; (b) the cleavage efficiency varies
with mismatch type and sequence context and requires some experience to
interpret easily; and (c) the enzyme generates
constitutive cleavage background patterns specific for each DNA
fragment that may obscure weak, specific cleavage products. Used
correctly, however, PASSPORT has the potential to substantially reduce
costs for the analysis of mutations and SNPs while maintaining a high
degree of accuracy.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
-thiotriphosphate nucleotides. Clin Chem 1995;41:1461-1466.The following articles in journals at HighWire Press have cited this article:
![]() |
T. Norberg, S. Klaar, L. Lindqvist, T. Lindahl, J. Ahlgren, and J. Bergh Enzymatic Mutation Detection Method Evaluated for Detection of p53 Mutations in cDNA from Breast Cancers Clin. Chem., May 1, 2001; 47(5): 821 - 828. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. P. Diamandis Sequencing with Microarray Technology--A Powerful New Tool For Molecular Diagnostics Clin. Chem., October 1, 2000; 46(10): 1523 - 1525. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |