Clinical Chemistry 47: 1845-1847, 2001;
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (9)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rolinski, B.
Right arrow Articles by Horn, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rolinski, B.
Right arrow Articles by Horn, K.
Related Collections
Right arrow Pediatric Clinical Chemistry
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2001;47:1845-1847.)
© 2001 American Association for Clinical Chemistry, Inc.


Technical Briefs

Total Bilirubin Measurement by Photometry on a Blood Gas Analyzer: Potential for Use in Neonatal Testing at the Point of Care,

Boris Rolinski1a, Helmut Küster2, Bernhard Ugele1, Rudolf Gruber1 and Klaus Horn1

1 Department of Clinical Chemistry, Ludwig-Maximilians-University, Ziemssenstrasse 1, D-80336 Munich, Germany

2 Department of Neonatology, Children’s University Hospital, Ludwig-Maximilians-University, Maistrasse 11, D-80337 Munich, Germany

aaddress correspondence to this author at: Zentralalbor Innenstadt, Klinikum der Universität, Medizinische Klinik Innenstadt, Ziemssenstrasse 1, D-80336 Munich, Germany; fax 49-89-5160-3901, e-mail Rolinski{at}kk-i.med.uni-muenchen.de

Neonatal jaundice is a common finding in neonates and has considerable impact on healthcare costs (1)(2)(3). Bilirubin at higher concentrations may cause kernicterus, a severe and often debilitating disease (4). Thus, jaundice may require phototherapy or even exchange transfusion to reduce plasma bilirubin concentrations (5). A decision-making process predominantly based on plasma bilirubin concentration is required in clinically jaundiced infants. Therefore, measurement of total bilirubin is one of the most frequently performed laboratory tests in newborns.

For monitoring hyperbilirubinemia in the newborn, methods are desirable that provide fast but reliable results from very small volumes of whole blood, enabling bilirubin measurements at the point of care. We evaluated a new method that uses a blood gas analyzer with multiple-wavelength photometry to determine total bilirubin concentrations in whole blood from jaundiced and nonjaundiced neonates and compared the results with a standard clinical chemistry procedure for total bilirubin measurement in plasma.

Samples were collected between April and June 2000 in a 40-bed nursery situated within the obstetrics department of the University Hospital. A total of 142 blood samples from 79 neonates (ages, 0–22 days) were investigated. Blood samples were obtained by heel prick and collected simultaneously into heparin-containing 55-µL capillaries (Radiometer Medical) and heparin-containing 200-µL microtubes (Sarstedt) used routinely for clinical chemistry determinations in our nursery. Oral informed consent was obtained from the parents.

The sample in the 55-µL capillary was measured immediately on an ABL 735 blood gas analyzer (Radiometer Medical) for pH, PCO2, and PO2 and by multiple-wavelength photometry for bilirubin. Whole blood was hemolyzed in an ultrasound cell and then measured simultaneously at 128 wavelengths between 478 to 672 nm. From these photometry data, bilirubin was calculated together with oximetry values (hemoglobin, fetal hemoglobin, oxygen saturation, and others). The algorithm (software version SW 3.20) was designed to correct for the presence of hemoglobin, fetal hemoglobin, and other matrix components.

The microtubes were transferred to the clinical chemistry laboratory within 2 h and centrifuged; total bilirubin was measured by the 2,5-dichlorophenyldiazonium (DPD) method (6). Additionally, direct bilirubin was determined by the method of Jendrassik and Grof (7) in 56 samples, and hemolysis, lipemia, and icterus were determined by multiple-wavelength photometry in 63 samples. All measurements were done on a Hitachi 917 automated analyzer (Roche Diagnostics) with reagents from the same supplier.

The Hitachi 917 automated analyzer was calibrated with routine multicalibrator solutions. Quality control analyses were performed at least twice daily at two concentrations with commercially available control samples based on human serum (Multiqual Levels 1 and 3; Bio-Rad). Four additional bilirubin controls (final concentrations, 328, 164, 33, and 0 µmol/L) were measured on each day of the study. These controls were prepared from the certified reference material SRM 916a (99.0% ± 0.1% unconjugated bilirubin) obtained from NIST and diluted in an albumin solution (40 g/L bovine serum albumin fraction V in 0.1 mol/L Tris, pH 7.4).

Statistical evaluation of the data was performed using SPSS 10 and Graph Pad Prism as well as the Freeware Method Validator.

For the NIST standards, on average, the values measured on the Hitachi 917 exceeded the true concentrations. The median slope of the regression lines from the 41 study days was 1.098 (range, 1.010–1.182), and the median y-intercept was -0.2 µmol/L (range, -3.1 to 12.3 µmol/L). Regression lines were extremely linear with a median goodness of fit (r2) of 0.9999 (range, 0.9996–1.0000). The intraassay CVs (n = 10) were 1.8%, 1.1%, and 1.1% and the interassay CVs (n = 41) were 8.3%, 3.5%, and 3.3% for 33, 164, and 328 µmol/L, respectively. Routine internal quality control during the study period (n = 41) yielded a mean of 14.5 µmol/L (expected concentration, 15.4 µmol/L) for the Level 1 and 106 µmol/L (expected concentration, 103 µmol/L) for the Level 3 controls with interassay CVs of 5.9% and 2.4%, respectively.

In the patient samples, total bilirubin concentrations covered the physiologically relevant span and were 13.7–412.2 µmol/L when measured photometrically in whole blood on the ABL 735, compared with 15.4–372.9 µmol/L when measured in plasma by the DPD method on the Hitachi 917. As shown in Fig. 1A , regression analysis according to Deming (8) yielded a slope of 1.002 (95% confidence interval, 0.914–1.090) and a y-intercept of -4 µmol/L (95% confidence interval, -22 to 14 µmol/L). The correlation coefficient (r2) was 0.76 (P <0.0001). A Bland–Altman (9) difference plot (Fig. 1B ) revealed a mean difference ± SD of -3.5 ± 36.2 µmol/L and only little dependence of the disagreement between both methods on the mean total bilirubin concentration. However, the difference between methods was >35 µmol/L in 30% of cases, and the maximum difference reached 116 µmol/L.



View larger version (16K):
[in this window]
[in a new window]
 
Figure 1. Correlation between total bilirubin concentrations measured in plasma by the DPD method (Hitachi 917; Ref method) and those measured in whole blood by multiple-wavelength photometry (ABL 735; Field method).

(A), regression according to Deming. The dashed line represents the line of identity. (B), Bland–Altman difference plot. Mean difference (solid line) ± 2 SD (dashed lines) is shown.

The presence of lipemia or hemolysis in the plasma samples, as measured by spectral analysis, did not interfere with total bilirubin determination by the DPD method. Neither the amount of direct bilirubin measured in the patient samples nor the age of the patients affected the goodness or the slope of the correlation between total bilirubin concentrations measured on the Hitachi 917 and the ABL 735 (data not shown). In addition, the correlation between both methods was not significantly influenced by hemoglobin content measured by oximetry (slope, 0.477 ± 1.288; r2 = 0.001) or by the pH of the samples (slope, -0.063 ± 0.022; r2 = 0.055).

Bilirubin is generally considered one of the most problematic analytes in clinical chemistry (10), and no accepted reference method for total bilirubin determination is currently available. Thus, the widely used DPD procedure was chosen as the comparison method in this study. Quality specifications in terms of precision and accuracy met CLIA (11) as well as European regulatory requirements (12).

In the patient samples, the measured bilirubin values covered the physiologically relevant range (85–400 µmol/L) within which decisions on phototherapy or exchange transfusion have to be made (2)(5). In general, there was a good agreement between bilirubin concentrations measured in plasma by the DPD method and those obtained from whole blood by multiple-wavelength photometry. The slope of the regression line was close to 1, and no significant systematic bias between the two methods was observed. There was, however, a rather wide spread of data around the regression line. Thus, correlation was considerably weaker than desirable for equivalent methods. On the basis of the within- and between-subject biologic variability of bilirubin [CVwithin = 25.6% and CVbetween = 30.5%, respectively (13)], the desirable imprecision (CV) and bias can be calculated to be <13% and <10%, respectively. Thus, the desirable total error should be <31.1% (13). In our study, these quality requirements were met in 86% of cases.

In conclusion, determination of total bilirubin from whole blood by multiple-wavelength photometry on a blood gas analyzer is a promising method for point-of-care testing. It combines the advantages of small sample volume and fast turnaround time. The manufacturer is currently working on an update of the measurement algorithm.


Acknowledgments

Addendum: Since completion of this study, the manufacturer has optimized the algorithm because of a spectral difference between the laboratory blood samples supplemented with "synthetic" bilirubin used to calibrate the ABL and "real" blood samples. Radiometer Medical has implemented this algorithm in the software version 3.61, released in January 2001.


References

  1. Newman TB, Easterling MJ, Goldman ES, Stevenson DK. Laboratory evaluation of jaundice in newborns. Frequency, cost and yield. Am J Dis Child 1990;144:364-368.[Abstract/Free Full Text]
  2. Maisels MJ, Gifford K. Normal serum bilirubin levels in the newborn and the effect of breast feeding. Pediatrics 1986;78:837-843.[Abstract/Free Full Text]
  3. Saigal S, Lunyk O, Bennett KJ, Patterson MC. Serum bilirubin levels in breast- and formula-fed infants in the first 5 days of life. Can Med Assoc J 1982;127:985-989.[Abstract]
  4. Schmorl C. Zur Kenntnis des icterus neonatorum, insbesondere der dabei auftretenden Gehirnveränderungen. Verh Dtsch Ges Pathol 1904;6:109-115.
  5. . American Academy of Pediatrics, Provisional Committee for Quality Improvement and Subcommittee on Hyperbilirubinemia. Practice parameter: management of hyperbilirubinemia in the healthy term newborn. Pediatrics 1994;94:558-565.[Abstract/Free Full Text]
  6. Wahlefeld AW, Herz G, Bernt E. Modification of the Malloy-Evelyn method for a simple, reliable determination of total bilirubin in serum [Abstract]. Scand J Clin Lab Invest 1972;29(Suppl 126):11-12.
  7. Jendrassik L, Grof P. Vereinfachte photometrische Methoden zur Bestimmung des Bilirubins. Biochem Z 1938;297:81-89.[Web of Science]
  8. Cornbleet PJ, Gochman N. Incorrect least-squares regression coefficients in method-comparison analysis. Clin Chem 1979;25:432-438.[Abstract/Free Full Text]
  9. Bland JW, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurements. Lancet 1986;1:307-310.[Web of Science][Medline] [Order article via Infotrieve]
  10. Doumas BT, Eckfeld JH. Errors of measurement of total bilirubin: a perennial problem. Clin Chem 1996;42:845-848.[Free Full Text]
  11. . US Department of Health and Human Services. Medicare, Medicaid and CLIA programs: regulations implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA). Final rule. Fed Regist 1992;57:7002-7186.[Medline] [Order article via Infotrieve]
  12. Fraser CG, Petersen PH, Ricos C, Haeckel R. Proposed quality specifications for the imprecision and inaccuracy of analytical systems in clinical chemistry. Eur J Clin Chem Clin Biochem 1992;30:311-317.[Web of Science][Medline] [Order article via Infotrieve]
  13. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jiminez CV, et al. Current database on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491-500.[Web of Science][Medline] [Order article via Infotrieve]



The following articles in journals at HighWire Press have cited this article:


Home page
Ann Clin BiochemHome page
J. M Kirk
Neonatal jaundice: a critical review of the role and practice of bilirubin analysis
Ann Clin Biochem, September 1, 2008; 45(5): 452 - 462.
[Abstract] [Full Text] [PDF]


Home page
PediatricsHome page
K. Grohmann, M. Roser, B. Rolinski, I. Kadow, C. Muller, A. Goerlach-Graw, M. Nauck, and H. Kuster
Bilirubin Measurement for Neonates: Comparison of 9 Frequently Used Methods
Pediatrics, April 1, 2006; 117(4): 1174 - 1183.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
S. C. Kazmierczak, A. F. Robertson, P. G. Catrou, K. P. Briley, B. L. Kreamer, and G. R. Gourley
Direct Spectrophotometric Method for Measurement of Bilirubin in Newborns: Comparison with HPLC and an Automated Diazo Method
Clin. Chem., July 1, 2002; 48(7): 1096 - 1097.
[Full Text] [PDF]


This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (9)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rolinski, B.
Right arrow Articles by Horn, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rolinski, B.
Right arrow Articles by Horn, K.
Related Collections
Right arrow Pediatric Clinical Chemistry
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS