Abstracts of Oak Ridge Posters |
1
2123 Dentistry/Pharmacy Center, Faculty of Pharmacy and Pharmaceutical Sciences, University of Alberta, Edmonton, Alberta, T6G 2N8 Canada;
2
Chenomx Inc., 2007, 8308-114 Street, Edmonton, Alberta, T6G 2E1 Canada;
3
122 Athabasca Hall, Artificial Intelligence Group, Department of Computing Science, University of Alberta, Edmonton, Alberta, T6G 2E8 Canada
aauthor for correspondence: fax 780-432-3388, e-mail lquerengesser{at}chenomx.com
Magnetic resonance diagnostics (MRD) uses automated, high-throughput nuclear magnetic resonance (NMR) spectroscopy for the rapid identification and quantification of small-molecule metabolites in biofluid mixtures (blood, urine, saliva, cerebrospinal fluid, and others). Specifically, MRD involves using a high-field (400 MHz) NMR instrument equipped with a small-volume flow probe and robotic sample handler to rapidly load biofluid samples and to collect their 1H NMR spectra. Spectral deconvolution software automatically assigns individual peaks to particular compounds and calculates concentrations from peak areas. MRD uses the principle of chemical shift separation to physically separate and identify individual compounds directly from 1H NMR spectra, thus avoiding chromatographic separation steps (e.g., HPLC, gas chromatography, and capillary electrophoresis). MRD is useful for rapid (<2 min per sample) qualitative and quantitative assessment of small-molecule metabolites.
NMR spectroscopy is not new to the field of clinical chemistry. Indeed several important applications have already been demonstrated in the area of diagnosis and therapeutic monitoring of metabolic disorders (1)(2)(3)(4), in toxicologic and renal testing (5)(6), and in the profiling of blood lipoproteins and cholesterol (7). An emerging approach to enable high-throughput in vivo toxicology is called metabonomics, which uses high-resolution NMR to rapidly evaluate the metabolic status of an animal (8)(9).
A key limitation to all of these NMR approaches is that they depend on manual sample handling and/or manual (i.e., expert) spectral analysis. This has made most NMR approaches to clinical analyses far too slow or too costly for routine chemical profiling or high-throughput screening. Because MRD is fully automated (sample handling, spectral collection, and spectral analysis are all handled by robots or computers), this technique offers the potential for high-throughput, comprehensive, and inexpensive chemical analysis of a wide range of biofluid samples.
To demonstrate the potential of MRD for high-throughput clinical screening and metabolic profiling, we constructed a simulated test-run of 1000 urine samples processed by a prototype MRD instrument developed jointly by our laboratory and Varian Inc. (Palo Alto, CA). Our intent was to investigate the performance of the MRD instrument and software under the demands of a high-throughput clinical testing laboratory. The instrument was assessed on sample-processing speed, robustness of sample handling, and accuracy of identifying samples and compounds.
We followed protocols and conditions approved by the University of Albertas Health Research Ethics Board to collect 1000 anonymous urine samples. A total of 925 samples were obtained from healthy adult volunteers who had completed consent forms. Seventy-five samples (63 children, 12 adults) were obtained as anonymous "discards" from several hospitals and clinics across Canada and were from patients with a wide variety of inborn errors of metabolism, neuroblastoma, and alcohol poisoning. All of the abnormal samples used in this test had been identified previously as such through conventional clinical screens. Among the abnormal samples were the following: 14 with propionic acidemia; 11 with methylmalonic aciduria; 11 with cystinuria; 6 with alkaptonuria; 4 with glutaric aciduria I; 3 each with pyruvate decarboxylase deficiency, ketosis, Hartnup disorder, cystinosis, neuroblastoma, phenylketonuria, ethanol toxicity, glycerol kinase deficiency, and hydroxymethylglutaryl-CoA-lyase deficiency; and 2 with carbamoylphosphate synthetase deficiency.
For MRD analysis, 990-µL portions from each urine sample were transferred to 1.8-mL autosampler vials, to which 0.5 mmol/L (10 µL of a 50 mmol/L solution) 3-(trimethylsilyl)-1-propane-sulfonic acid, sodium salt (Sigma-Aldrich) was added. We manually adjusted the samples to pH 6.5 using HCl(aq) or NaOH(aq), as necessary. Manual pH adjustment was necessary because the software module required to automatically measure the pH and adjust the spectral deconvolution process was not completed in time for this study.
The prototype MRD instrument consisted of a 400 MHz Varian NMR spectrometer equipped with a 60-µL triple resonance-flow probe with an interchangeable flow cell and a modified Varian VAST (Versatile Automatic Sample Transport) system. The VAST system uses a robotic liquid handler (Gilson Model 215) and three computer-controlled switching valves, which direct sample flow to and from the flow probe through small-diameter Teflon tubing. Each urine sample (250 µL) was automatically loaded into the NMR spectrometer and a one-dimensional 1H NMR spectrum collected (12 scans, 1.998 s; acquisition time, 0.5 s; acquisition delay, 6000 Hz sweepwidth) at ambient temperature (21.5 °C ± 0.5 °C). After data collection, the urine sample was ejected and the flow probe extensively rinsed with distilled water before the next sample was loaded. Sample carryover was <1%.
NMR spectra were autoprocessed (e.g., transformed, phased, and referenced) and deconvolved with a suite of specially developed software applications and databases. The deconvolution process allows for the automated identification and quantification of components in biofluid mixtures through spectral database comparisons. We tested for 149 compounds currently contained in our spectral database.
The instrument automatically loaded and analyzed all 1000 samples in 35.2 h (1 sample every 2.1 min) with minimal human supervision. The mean sample loading and rinsing time was 93 s, whereas spectral acquisition had a mean of 32 s. The mean times for spectral processing and deconvolution (which can be performed in parallel with sample loading and data acquisition) were 7.1 and 72 s, respectively. During the test run, one sample-loading failure occurred, but did not lead to instrument downtime.
The deconvolution software was tested for accuracy for the following: (a) identification and quantification of urinary metabolites, (b) identification of nonpathologic and abnormal urine samples, and (c) identification of specific disease states or conditions. We evaluated compound-identification and/or -quantification accuracy primarily through detailed analysis of the abnormal samples. Specifically, the 15 disorders found in the abnormal urine samples were characterized by 34 unique or abnormally abundant metabolites (e.g., homogentisic acid, glycerol, glutaric acid, and others). Our results indicate that the MRD software succeeded in correctly identifying all 34 abnormal metabolites in all 75 abnormal samples. All 34 metabolites had been identified previously through conventional HPLC, gas chromatographymass spectrometry, or amino acid or organic acid analysis. In some cases, these metabolites were also detected in nonpathologic urine samples, but at concentrations too low to be of significance or at concentrations well within the reference interval (10).
Although it was not possible to verify the MRD-measured concentrations for all metabolites (most conventional tests provide only qualitative results), our results for those metabolites that could be reliably quantified indicated a correlation coefficient between conventionally measured concentrations and MRD-measured concentrations of 0.99 (spanning a concentration range of 400 µmol/L to 550 mmol/L). An example of compound identification and quantification achieved by this MRD software is shown in Fig. 1
, which shows an abnormal urine NMR spectrum (specifically, methylmalonic aciduria), along with the MRD-calculated NMR spectrum. The calculated spectrum was generated from individual NMR spectra of MRD-identified and -quantified compounds.
|
Among nonpathologic urine samples, 87 ± 11 compounds were routinely identified and quantified by the MRD software on average. The identification and concentration of these metabolites were partially verified through manual spectral analysis of three randomly chosen samples and their subsequent chemical analysis. It was not practical to attempt to verify the identity of all compounds in the nonpathologic urine samples via conventional assays. The MRD software failed to consistently identify citrate, glycine, and histidine, probably because of the extreme sensitivity of their chemical shifts to pH and/or calcium ion concentrations.
A simple concentration-threshold algorithm achieved 96% sensitivity and 100% specificity, with 72 of 75 abnormal samples being detected and all 925 nonpathologic samples being correctly classified as nonpathologic. The three missed abnormal samples were cystinuria, glutaric aciduria, and carbamoylphosphate synthetase deficiency. This misclassification arose from the fact that these three samples exhibited relatively modest concentrations of abnormal metabolites. This particular algorithm compared MRD-quantified metabolite data with those values reported by Tietz (10), a database of "normal" and "abnormal" metabolite concentrations (both absolute and relative to creatinine) tabulated from several literature sources, as well as metabolite data collected from our own assays. Age- and gender-related adjustments were not included in the classification scheme.
Tests of the disease classification accuracy (for 1 healthy state and 15 diseases or conditions) yielded 95.5% and 92.4% for sensitivity and specificity, respectively. Interestingly, all false positives were confined to those urine samples already classified as abnormal. In particular, four false positives were identified for propionic aciduria, three false positives for phenylketonuria, two false positives were identified for cystinuria, and one false positive for ethanol toxicity. These classification problems likely arose from the age-independent, all-or-none classification scheme used by our algorithm. A probabilistic assessment that included age- or gender-related values would likely have improved the results or revealed multiple disease possibilities.
Overall, this prototype instrument was rapid, robust, and accurate. It rapidly identified and quantified key metabolic markers of both common conditions (ethanol toxicity) and rare disorders (neuroblastoma). It appears to offer a high-throughput, inexpensive approach to metabolic profiling, with possible applications in medical diagnostics, drug compliance testing, toxicology, and food testing.
Acknowledgments
We wish to thank the University of Alberta Hospital, Toronto Hospital for Sick Children, Childrens & Womens Health Centre of British Columbia, and Health Sciences Centre (Winnipeg, Manitoba) for contributing samples for this study. We also thank Dr. Fiona Bamforth (Department of Laboratory Medicine and Pathology, University of Alberta) for invaluable advice.