| Abstract |
|---|
|
|
|---|
Methods: A method was developed to allow routine analysis of whole blood samples for molybdenum, cobalt, chromium, and nickel. Sample preparation consisted of a simple 1:10 dilution of whole blood with a solution of 10 mL/L Triton X-100, 0.0002 mol/L EDTA, and 0.01 mol/L ammonium hydroxide. Final determination was performed by a double-focusing magnetic sector inductively coupled plasma mass spectrometer operated in medium-resolution mode (resolution, 3400). Online addition of rhodium was used for internal standardization.
Results: Detection limits in whole blood were 0.06 µg/L for chromium, cobalt, and molybdenum and 0.30 µg/L for nickel. Base concentrations of 0.22, 0.17, 0.62, and 0.99 µg/L for chromium, cobalt, molybdenum, and nickel, respectively, in whole blood have been found. Polyatomic interferences on all four elements have been shown to be resolved from the analyte masses by use of a resolution of >3000.
Conclusions: The simple, rapid method of sample preparation is effective in minimizing potential contamination and enables 60 samples (run time, 8 h) to be analyzed before cleaning the instrument is necessary. A resolution >3000 was sufficient to separate polyatomic interferences from the masses of interest. The method was used to analyze a large number of blood samples taken from primary patients awaiting total hip arthroplasty. The method is sensitive enough to provide base concentrations for chromium, cobalt, and molybdenum in whole blood. The results for nickel were compromised by high signals for blank samples.
| Introduction |
|---|
|
|
|---|
Two main types of prostheses are used in current practice: a metal-on-plastic prosthesis, in which a metal stem articulates on a plastic cup; and a metal-on-metal prosthesis, in which both components are metallic. Three main metal alloys are in use: stainless steel; cobalt chrome; and titanium, vanadium, and aluminum.
Recently, there has been a clinical trend for reintroduction of metal-on-metal bearing surfaces, especially for young orthoplast patients, and an increasing popularity of modular porous-coated devices with large surface areas of exposed metal (4). A recent study found a ninefold increase in serum chromium and a threefold increase in serum cobalt in patients with long-term, well-functioning metal-on-metal implants (5). This and other studies have led to concern about the effect of long-term exposure to cobalt, chromium, nickel, molybdenum, and other metals and have highlighted the need for accurate biological monitoring of metal exposure. Effective monitoring can also provide insights into the mechanism of metal release. Corrosion is an important source of metal release, and it has been suggested that accurate monitoring could be a useful tool in the diagnosis of patients who are symptomatic after a total joint replacement (4) because the wear of metal-on-metal bearings cannot be measured on standard radiographs (1).
The main analytical problem in determining these metals in blood is they are present at extremely low (sub-µg/L) concentrations in a very complicated matrix. Electrothermal atomic absorption spectroscopy has been used successfully for such analyses (4)(5), but one of its disadvantages is that the analytical procedures are relatively time-consuming. This is because of the cycle time needed for electrothermal atomic absorption spectroscopic analysis and because the cuvette must be cleaned regularly to remove carbon, which is present in high quantities in the sample matrix. Most analyses are performed on serum samples because the matrix is less problematic than whole blood. There is evidence to suggest, however, that chromium, for example, accumulates in the red blood cells (6)(7), thus increasing the need for reliable methods of analyzing whole blood. Inductively coupled plasma mass spectrometry (ICP-MS) has sufficient sensitivity for such analysis, but quadrupole instruments suffer greatly from interferences in the mass range of interest (8)(9). A few methods have been reported that overcome these interferences, but the procedures are quite complicated, making routine analysis difficult (10)(11)(12). Newer magnetic sector ICP-MS (HR-ICP-MS) instruments are potentially very powerful tools for the analysis of samples with complex matrices. The strength of HR-ICP-MS lies in its ability to produce sufficient resolving power to separate polyatomic interferences from the analytes, which is of particular importance in the mass range of 1880, where interferences from plasma gases are prevalent. In addition to resolving many of the common polyatomic interferences, HR-ICP-MS instruments routinely have detection limits one order of magnitude lower than quadrupole ICP-MS instruments.
We recently published a report (13) demonstrating the ability of HR-ICP-MS to determine chromium, nickel, and cobalt in whole human blood based on the method originally developed by Delves and Sieniawska (14). Our method, however, was only used for a very limited number of samples. The aim of this present work is twofold: (a) to assess the effectiveness of the method as a routine, cost-effective technique for the analysis of a large number of samples; and (b) to determine baseline concentrations of metals associated with metal implants with particular reference to how metal concentrations might change as a result of an orthopedic surgery.
| Materials and Methods |
|---|
|
|
|---|
|
preparation of samples and calibrants
Blood was withdrawn from a vein in the forearm using a standard
Venflon cannula at the time of induction of anesthesia before
joint-replacement surgery. The blood was drawn into standard plastic
syringes, and the first 5 mL was discarded to avoid contamination from
the trocar. The second 5 mL was then discharged into a lithium Teflon
plastic container specifically designed for trace metal analysis
(Teklab).
Blood samples (1 mL) were diluted 10-fold with a diluent consisting of 0.01 mol/L ammonium hydroxide solution (specific gravity, 0.88; BDH), 0.0002 mol/L EDTA (Aldrich), and 10 mL/L Triton X-100 (Scintran grade; BDH) in deionized water (18 Ohm; Elga Maxima).
A multielement, 10 mg/L stock solution was prepared in 0.7 mol/L ultrapure HNO3 (Ultrex grade; J.T. Baker) from commercial single-element 1000 ± 3 mg/L calibration solutions (Spec CertiPrep Inc) from which daily calibrators were prepared by serial dilution in 0.14 mol/L HNO3. Final calibrators were prepared in the diluent. An internal standard (rhodium; Spec CertiPrep Inc), was prepared in the diluent described above. The internal standard was added on-line to all calibrators and sample solutions via a T-joint at a concentration of 1 µg/L to correct for instrument drift and matrix effects in a continuous flow system. Calibration was achieved by external calibration using 0.1, 0.2, 0.5, 0.75, and 1.0 µg/L calibration solutions with the calibration curves calculated using the Linear Through Zero option. Samples were analyzed after dilution, and a dilution factor of 10 was applied to the results after analysis.
Method recovery was checked by adding to a diluted blood control sample the required amount of multielement solution to give a final concentration of 0.45 µg/L. After dilution, all samples were kept at 5 °C and analyzed within 48 h of dilution. The recovery was calculated by subtracting the concentration measured in the native blood sample from the concentration measured in the blood supplemented with the multielement solution and comparing that difference with the known amount added. The reference whole blood Seronorm Level 2 (Nycomed AS) was analyzed as an additional quality-control sample.
| Results and Discussion |
|---|
|
|
|---|
|
isotopic interferences
Polyatomic interferences are known to occur in analysis by ICP-MS.
However, when an analysis is directed to trace or ultratrace
concentrations of metals, possible interferences that appeared to be
insignificant or nonexistent can have an important impact on the
validity of the analysis. A list of known and possible interferences
for chromium, cobalt, nickel, and molybdenum in the whole blood matrix
are given in Table 3
, along with the resolution required to spectrally resolve them
from the analyte isotope.
|
To evaluate possible interferences, scans of control blood samples from
healthy preoperative patients were made under standard measurement
conditions, covering the masses required. These scans are shown in Fig. 1
. As can be seen, all of the isotopes used in this
study exhibited substantial polyatomic interferences at the
trace concentrations in whole blood and required a resolution >3000 to
resolve. If a resolution >3000 had not been used (i.e., a quadrupole
low-resolution ICP-MS was used), the interfering species shown in Fig. 1
could have caused substantial errors in the analysis.
|
The unidentified interference on 95Mo (Fig. 1C
) can cause problems if an instrument is not set up and
optimized very carefully. The advantage of 95Mo
over 98Mo is that the
95Mo isotope does not suffer from direct
isobaric interference (Table 3
). However, we decided to use
98Mo in future work despite the
98Ru interference because we believe that the
amount of ruthenium in whole blood is negligible. In addition,
the abundance of 98Ru is only 1.88%, so
substantial amounts (with respect to molybdenum) would have to be
present to cause substantial enhancement in the
98Mo signal. For 98Mo, the
Pt2+ and possibly Hg2+
peaks are well resolved from the 98Mo peak at
medium resolution (resolution, 3400; Fig. 1D
).
method performance characteristics
Calibrations were linear over the calibration range with
regression coefficients (R) >0.99 for all elements.
Additional typical calibration parameters are given in Table 4
. Instrumental mass calibration was performed daily using a 1
µg/L multielement calibrator and monitored during each analysis. If
substantial mass calibration drift was detected, the instrument was
recalibrated for mass calibration and the calibrations were run again
before proceeding. Recoveries for the different trace elements
(concentration of each element, 0.45 µg/L) in diluted blood
samples are given in Table 4
. Supplemented blood samples were
analyzed within every batch along with replicates from previous batches
to provide continuity data. An additional check on accuracy involved
the analysis of the reference blood Seronorm Level 2. This had
reference values for chromium, cobalt, and nickel with an indicative
value for molybdenum. These data are also given in Table 4
, in which
good agreement can be observed for chromium, cobalt, and nickel, the
three elements for which there was a reference value. For molybdenum,
however, the ICP-MS method seemed to produce higher results, possibly
because the value indicated for this Seronorm sample does not take into
consideration the presence of endogenous molybdenum in the
blood.
|
Whole-blood detection limits are given in Table 5
. Detection limits were calculated as 3 SD above the value for a
blank sample (n = 6) plus the dilution factor, which in this case
was 10. The nickel detection limit was nearly always compromised by a
relatively high signal for the blank sample due to Ni contamination.
Care was taken to limit this nickel blank by use of a platinum
sample and skimmer cones, ultrapure HNO3, 18 Ohm
deionized water, and Scintran-grade Triton X-100, and soaking the
quartz spray chamber and injector overnight in 0.7 mol/L
HNO3. In this way, the results for the nickel
blank were brought down to manageable concentrations.
|
Drift was corrected by analysis of an intermediate-concentration
calibrator every six samples. If the drift was >10%, a simple linear
drift-correction procedure was performed. This consisted of calculating
the average drift for sequential drift-correction in-house calibrators.
This average drift factor was then applied as a percentage correction
to the six samples that were analyzed between the two calibrators. Fig. 2
shows the typical drift encountered for a
95Mo calibration solution over a 7-h run of 60
blood samples.
|
results of preoperative patients
The blood from 47 preoperative patients was analyzed using the
described HR-ICP-MS method. The results are summarized in Table 5
. The
detection limits obtainable with the method are acceptable for the
evaluation of the base concentrations of chromium, cobalt, and
molybdenum in preoperative blood. This is illustrated by the fact that
only four samples for chromium, two for cobalt, and none for molybdenum
were below the detection limit; mean blood concentrations were
0.22 ± 0.26 µg/L for chromium, 0.17 ± 0.17 µg/L for
cobalt, and 0.62 ± 029 µg/L for molybdenum. In this respect,
the multielement HR-ICP-MS procedure seemed to produce better data than
reported by previous authors using the slower, single-element technique
of electrothermal atomic absorption spectroscopy, as shown in Table 5
.
The baseline value for nickel in blood (0.99 ± 1.15 µg/L)
obtained in this study must be regarded as an overestimation because
only 57% of the samples had concentrations above the detection limit.
This would concur with the conclusion of Templeton et al.
(15), who stated in that the concentration of
nickel in serum and blood is
0.1 µg L. The more abundant
58Ni was not used in this work because it suffers
from isobaric interference from 58Fe. Although
the abundance of 58Fe is only 0.28%, the low
concentrations of nickel in the samples would have been severely
affected by the relatively high concentrations of iron in the
whole-blood matrix.
In conclusion, we have developed a simple, rapid, and cost-effective method for the analysis of certain trace metals in whole blood. The method relies on the simple 10-fold dilution of whole blood with an ammonium/EDTA/Triton X-100 diluent with analysis by HR-ICP-MS using medium resolution (>3000). Substantial interferences for all the trace metals under study were found, indicating the limitations of low-resolution (quadrupole) ICP-MS for this type of analysis. The procedure has been applied in the study of trace metal concentrations in whole blood of preoperative arthroplasty patients. Baseline blood concentrations for chromium, cobalt, and molybdenum have been established. At the present time, baseline data for nickel are compromised by high signals for the blank. A large study is currently being performed to determine, in detail, what changes take place in these metals in whole blood, in particular with metal-on-metal prostheses at different postoperative intervals.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
D. J. Langton, S. S. Jameson, T. J. Joyce, J. Webb, and A. V. F. Nargol The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip J Bone Joint Surg Br, September 1, 2008; 90-B(9): 1143 - 1151. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Antoniou, D. J. Zukor, F. Mwale, W. Minarik, A. Petit, and O. L. Huk Metal Ion Levels in the Blood of Patients After Hip Resurfacing: A Comparison Between Twenty-eight and Thirty-six-Millimeter-Head Metal-on-Metal Prostheses J. Bone Joint Surg. Am., August 1, 2008; 90(Supplement_3): 142 - 148. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Lidgren Chronic inflammation, joint replacement and malignant lymphoma J Bone Joint Surg Br, January 1, 2008; 90-B(1): 7 - 10. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Daniel, H. Ziaee, P. B. Pynsent, and D. J. W. McMinn The validity of serum levels as a surrogate measure of systemic exposure to metal ions in hip replacement J Bone Joint Surg Br, June 1, 2007; 89-B(6): 736 - 741. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Ziaee, J. Daniel, A. K. Datta, S. Blunt, and D. J. W. McMinn Transplacental transfer of cobalt and chromium in patients with metal-on-metal hip arthroplasty: A CONTROLLED STUDY J Bone Joint Surg Br, March 1, 2007; 89-B(3): 301 - 305. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Daniel, H. Ziaee, A. Salama, C. Pradhan, and D. J. W. McMinn The effect of the diameter of metal-on-metal bearings on systemic exposure to cobalt and chromium J Bone Joint Surg Br, April 1, 2006; 88-B(4): 443 - 448. [Abstract] [Full Text] [PDF] |
||||
![]() |