| Introduction |
|---|
|
|
|---|
HPLC assays for PAA and PBA in plasma have limited sensitivity (4). A gas chromatographicmass spectrometric method had improved sensitivity (13), but the method did not use stable isotopes as internal standards and the clinical utility of the method was not documented. Here, we describe a rapid, precise, and convenient method to measure blood concentrations of these drugs using dried filter-paper blood spots.
BA, PAA, and PBA were purchased from Sigma-Aldrich Co. [ring-d5]-BA, [ring-d5]-PAA, and [2,2,3,3-d4]-4-PBA were purchased from Cambridge Isotope Laboratories. Regisil RC-3 [N,O-bis-(trimethylsilyl)trifluoroacetamide containing trimethylchlorosilane (10:1 by volume; BSTFA-TMCS)] was purchased from Regis Chemical Co. Blood collection cards (cat. no. 10538414) were purchased from Schleicher & Schuell.
Blood was blotted onto absorbent filter paper and dried at room
temperature. The analytical samples consisted of four punched-out discs
(6.3 mm in diameter), each of which contained
10.6 µL of blood.
When less blood was available, smaller discs were used. The method was
sensitive enough to measure peak drug concentrations in a single 3-mm
disc (2.4 µL of blood). The blood was eluted from the paper discs by
sonication for 5 min in 1 mL of deionized water containing 100 µg of
d5-BA,
d5-PAA, and
d4-PBA, respectively. The blood-spot
extract was filtered through an Amicon Centrifree YM-30 filter
(Millipore), yielding a clear filtrate. The filtrate was saturated with
1 g of NaCl and acidified to pH <1 with 0.2 mL of 5 mol/L HCl.
The acids were extracted twice with 3 mL of ethyl acetate for 5 min on
a mechanical shaker. The organic layers were combined and dried under a
stream of nitrogen at 50 °C. Trimethylsilyl (TMS) derivatives of the
organic acids were made by incubation with 100 µL of BSTFA-TMCS (10:1
by volume) for 20 min at 60 °C. The derivatized extract (2 µL) was
injected into the gas chromatograph (Hewlett Packard Model
5890 II; Agilent Technologies) by an automated injector. The
acids were separated on a cross-linked 5% diphenyl-95%
dimethylsiloxane capillary column [Ultra 2; 30 m x 0.2 mm
(o.d.); film thickness, 0.33 µm; Agilent Technologies].
The column temperature began at 100 °C and was ramped to 270 °C
at 5 °C/min. The injection port and transfer line temperatures were
250 and 280 °C, respectively. The split injection ratio was 1:40.
The flow rate of the helium carrier gas was
1 mL/min, and the linear
velocity was 35 cm/s. Column-head pressure was 22.5 psi. The ions were
detected by a quadrupole mass selective detector (Model 5971A; Hewlett
Packard ) equipped with an electron impact ion source that was
autotuned and run in high-resolution mode with selected-ion monitoring.
The following ions were determined within the 15-min runtime:
m/z 179 for native BA-TMS, m/z 184 for
d5-BA-TMS, m/z 164 for
native PAA-TMS, m/z 169 for
d5-PAA-TMS, m/z 221 for
native PBA-TMS, and m/z 225 for
d4-PBA-TMS. The elution times were
7.37 min for native BA-TMS, 7.34 min for
d5-BA-TMS, 8.50 min for native
PAA-TMS, 8.47 for d5-PAA-TMS, 13.85
min for native PBA-TMS, and 13.80 for
d4-PBA-TMS. Dwell time for each
ion was 50 ms, and the resolution was 0.5 atomic mass units.
Quantification was performed with the calibration curves of
m/z 179/184, m/z 164/169, and m/z
221/225 for BA, PAA, and PBA, respectively, representing the ratios of
known amounts of native acids (010 µg) to isotopic acids (100
µg).
The extraction efficiency was calculated from six 100-µg aliquots of BA, PAA, and PBA spotted on filter paper. Reproducibility was determined by analysis of blood spots with 4.0 µg of added BA, PAA, and PBA. The stability of BA, PAA, and PBA on filter-paper blood samples over time was tested by analyzing nine blood spots, each containing 4.0 µg of the three acids, stored at room temperature and subsequently analyzed over a 1-month period. The heat stability of BA, PAA, and PBA in blood spots at 80 °C was examined by placing the samples in an oven for 4 and 24 h.
Fifteen patients with urea cycle disorders and 1 with propionic acidemia treated with one or more of these medications were tested using this method.
The prominent m/z 179 and 184 ions (native and isotope,
respectively) for BA, m/z 164 and 169 ions for PAA, and
m/z 221 and 225 ions for PBA (Fig. 1
) were selected for quantification. The calibration curves were
linear from 0 to 100 µg. The equations of the response curve were:
y = 0.0131x + 0.0003 for BA;
y = 0.0135x + 0.0002 for PAA; and
y = 0.0121x + 0.0006 for PBA.
|
To determine the detection limits for BA, PAA, and PBA from filter-paper blood spots, samples with 0.0110 µg of the acids added were prepared, followed by analysis as described above. The detection limits (signal-to-noise ratio >10) for the blood-spot samples were 0.01, 0.06, and 0.04 µg for BA, PAA, and PBA, respectively. When four 6.3-mm (diameter) discs (42.4 µL of blood) were used, the acids could be detected at concentrations as low as 1.9, 10.4, and 5.7 µmol/L for BA, PAA, and PBA, respectively. This method was sensitive enough to monitor the concentrations of the three drugs in the blood of the patients. To monitor the peaks of PAA and PBA (typically 200500 µmol/L), a single 3-mm disc (2.4 µL of blood) was sufficient.
An analysis of the mass spectra of BA, PAA, and PBA at various concentrations showed spillover to the stable isotopes of 0.07%, 0.06%, and 0.25% from native BA (m/z 184), PAA (m/z 169), and PBA (m/z 225), respectively. Even at peak blood concentrations, these spillovers were inconsequential. To increase the sensitivity and overcome the potential problems of spillover in measuring these acids, the column temperature could be ramped at a slower rate (e.g., 3 °C/min) to better separate the native and isotopic peaks.
The extraction efficiencies for BA, PAA, and PBA were 74% ± 2%, 76% ± 3%, and 74% ± 3%, respectively (mean ± SD; n = 6). Repeated analysis of blood supplemented with 4.0 µg of BA, PAA, and PBA gave results of 3.9 ± 0.2, 4.0 ± 0.1, and 4.0 ± 0.1 µg (mean ± SD; n = 6), respectively. Dried filter-paper blood samples that were supplemented with 4.0 µg of BA, PAA, and PBA and stored up to 1 month at room temperature before analysis gave results of 4.0 ± 0.2, 4.0 ± 0.1 and 3.9 ± 0.1 µg (mean ± SD; n = 9), respectively. The results for blood spots heated at 80 °C for 24 h were within 2.6% (BA), 2.5% (PAA), and 5.2% (PBA) of those not heated.
We determined the peak (1.5 h postdose) and trough (predose)
concentrations of BA, PAA, and PBA in 16 treated patients (Table 1
). In six patients, PBA and PAA were 220551 µmol/L
and 307473 µmol/L, respectively, 1.5 h after 100125 mg/kg
doses of PBA. In several patients taking similar or higher doses
(e.g., patients 10 and 11), however, PBA and PAA were much lower. In
patient 13, PBA was 1542.4 µmol/L, and PAA was 405.3
µmol/L 1.5 h after 4.6 g of sodium
phenylbutyrate (71 mg/kg) given by gastrostomy. In all patients,
the predose concentration of PBA was <26 µmol/L, and the predose
concentration of PAA was <170 µmol/L. The highest postdose
PBA concentration measured was 1542.4 µmol/L, and
the highest postdose PAA concentration was 664.7 µmol/L. The BA
concentration in patients not taking BA was 1262 µmol/L,
which likely reflects values in healthy individuals.
|
The method described here is sensitive, specific, and accurate. It
allows easy collection and delivery to the testing site and should
promote good monitoring of therapy. We also documented that the three
analytes in the dried blood were stable to heat, alleviating concerns
of heat stability if the sample was placed in an outside mailbox.
Although extraction efficiency was
75%, the high reproducibility
and the use of stable isotopes as internal standards compensated for
incomplete extraction. In addition, the use of selected-ion monitoring
and the highly reproducible retention times of a capillary column
allowed this method to detect the three acids with virtually no
interference from other compounds.