| Abstract |
|---|
|
|
|---|
Methods: This study included 14 patients who underwent ICD implantation and intraoperative testing. We measured cTnT, cTnI, creatine kinase MB (CK-MB) mass, CK activity, and myoglobin before and at definite times after intraoperative shock application.
Results: Depending on the effectiveness of shocks and the energy applied, the cardiac-specific markers cTnT and cTnI, as well as CK-MB mass, showed a significant increase compared with the baseline value before testing and peaked for the most part 4 h after shock application. In contrast, the increases in CK activity and myoglobin were predominantly detectable in patients who received additional external shocks.
Conclusions: ICD implantation and testing leads to a short release of cardiac markers into the circulation. This release seems to be of cytoplasmic origin and depends on the number and effectiveness of the shocks applied.
| Introduction |
|---|
|
|
|---|
The cardiac isoforms of troponin T (cTnT) and troponin I (cTnI) are highly specific markers for the detection of myocardial damage. After myocardial ischemia but also as a result of other myocardial stress, such as myocarditis, cTnI and cTnT are released into the circulation (4)(5) where they can be detected with sensitive immunoassays (6)(7).
This study assesses the extent of myocardial injury after ICD implantation and testing by the release of the cardiospecific markers cTnT and cTnI in relation to other, not fully cardiospecific, markers of cardiac ischemia: creatine kinase (CK), CK-MB mass concentration, and myoglobin.
| Materials and Methods |
|---|
|
|
|---|
|
After clotting and centrifugation, the serum samples were stored at -70 °C until analysis
methods
Implantation and ICD used.
ICD systems from Guidant
Corporation, Cardiac Pacemakers (Models 1749, 1782, 1783, and 1849) and
from Medtronic, Inc. (Models 7250 and 7227) were used. Tined,
steroid-eluting electrodes were used in 10 patients (Medtronic
"Sprint" 6942/6932 and Cardiac Pacemakers Endotak Endurance: 145),
and tined, non-steroid-eluting electrodes were used in 4 patients
(Cardiac Pacemakers Endotak Endurance: 135). In five patients,
additional steroid-eluting "screw-in" leads (Medtronic 6940) for
atrial sensing were implanted.
Each patient received a 1-J test shock before every testing episode. VF
was induced with a shock-on-T induction with 1.2 J. For determining the
defibrillation threshold, the verification technique was used.
Intermediate energies that successfully defibrillate the patient and
provide an adequate energy margin are determined in the verification
procedure. This technique is widely accepted because it does not
require the subject to be tested with energies that do not defibrillate
as the threshold method does (8). If the first shock was
effective, the physician could repeat the testing with the same energy
or stop further testing. In case of failure, the system
automatically applied a second shock with maximum energy. Implantation
criteria were met if induced VF could be terminated one time with a
shock of
1015 J to maximum output or if two subsequently induced
VF episodes could be defibrillated successfully with 1015 J to
maximum output.
In case of missing implantation criteria, repositioning of the ventricular lead was performed before retesting. In one patient (patient 5), the physician decided not to reposition the electrode although the second shock after the first repositioning (shock 4) was ineffective and VF had to be terminated with an effective 31-J shock. Instead of repositioning, the physician tested again and defibrillated successfully with 23 J. In another patient (patient 18), the electrode had to be repositioned twice, and the patient received two external 360-J shocks. Three patients received atrial shocks through a additional lead for terminating atrial fibrillation.
To guarantee myocardium recovery between the ischemic episodes produced by VF, care was taken not to induce VF within 5 min after the previous testing episode.
Analytical procedures.
cTnT and CK-MB were measured with an
electrochemiluminescence immunoassay on an
Elecsys® 2010 analyzer (Roche
Diagnostics). The cTnT assay uses two monoclonal antibodies
specific for the cardiac isoform of TnT and has <0.01%
cross-reactivity with skeletal muscle troponin T. The CV is
5.1% at the cutoff of 0.1 µg/L for cTnT, and for CK-MB, the CV is
3.9% at 5.98 µg/L; the cutoff is 5 µg/L (9). cTnI was
measured on a Dimension® RxL analyzer (Dade
Behring) using the TROP FlexTM reagent cassette.
This assay uses two monoclonal antibodies specific for two different
epitopes on the cTnI molecule. The cross-reactivity with skeletal
troponin I is <0.1%. The CV is 9.4% at 0.5 µg/L, and the cutoff is
0.4 µg/L. Myoglobin was measured on a BN 2 analyzer (Dade Behring)
using a immunonephelometric assay (N-Latex Myoglobin). The CV is 4.8%
at 85 µg/L, and the cutoff is 70 µg/L. Total CK activity was
measured on a Vitros® 950 IRC (Ortho Clinical
Diagnostics) at 25 °C. The CV is 3% at 70 U/L, and the cutoff is 80
U/L for men and 70 U/L for women. All assays were performed according
the recommendations of the manufacturers.
Statistics.
Analyses were performed using a commercially
distributed software package (Astute, Statistics Add-in for Microsoft
Excel; DDU Software). To test for the differences in serum markers at
baseline and after ICD implantation, the Wilcoxon matched-pairs
signed-ranks test was used. Significance was estimated as P
<0.05.
| Results |
|---|
|
|
|---|
|
The individual markers and the biochemical findings of all patients are
shown in Table 3
. In nine patients, cTnT was detectable. cTnI was detectable in
eight patients, all of whom also had detectable cTnT values. cTnI peak
concentrations above the cutoff were seen in three patients, two of
whom also exceeded the cTnT cutoff. In both patients, we also found
CK-MB mass increases over the cutoff. All of the patients with
cTnT/cTnI peak concentration above the cutoff received two ventricular
shocks; two patients had additional atrial screw-in electrodes, and one
of them received three additional atrial shocks with lower energy. No
repositioning had to be performed in all of these patients. There was a
weak correlation between troponin increase and cumulative energy in all
of the patients who received two effective ventricular shocks
(r = 0.647 for cTnI and r = 0.636 for
cTnT).
|
CK-MB mass concentrations above the cutoff were seen in four patients. One of these patients received three atrial and seven ventricular shocks (cumulative energy, 194 J). Additionally, the physician applied two external 360-J shocks, and the electrode had to be repositioned twice.
The CK activity increased to >80 U/L in five patients. Patient 18 showed the largest increase (164 U/L); this patient received the largest cumulative energy (194 J) and two external 360-J shocks. In two of the five patients showing CK activity >80 U/L, additional atrial leads were implanted (patients 7 and 18). In all of these five patients, CK showed no peak within 24 h.
Myoglobin concentrations peaked at >70 µg/L in nine patients. In three patients, additional atrial leads were implanted. The very high myoglobin peak concentration for patient 18 (291 µg/L) correlated with the high CK in this patient.
No patient who received only one shock showed detectable cTnT or cTnI values. When we compared the number of shocks and the increases in cTnT and cTnI, only patients receiving two effective shocks showed concentrations above the cutoff. Median peak values postoperatively were 0.08 µg/L (range, 02.16 µg/L) for cTnI, 0.023 µg/L (range, 00.26 µg/L) for cTnT, and 2.5 µg/L (range, 014.93 µg/L) for CK-MB.
| Discussion |
|---|
|
|
|---|
We found increased serum concentrations of all markers, cTnT, cTnI, CK-MB, CK, and myoglobin, after ICD implantation. Patients who received only one shock showed no increases in cTnT, cTnI, and CK-MB mass above the cutoff values. All patients with cTnT and cTnI peak concentrations above the cutoff had received two ventricular shocks.
It is remarkable that patients who received more than two shocks, very high cumulative energy, and repositioning of the electrode (patients 5 and 18) had no (patient 5) or only marginal (patient 18) increases in cTnT and cTnI. Wrong positioning of the electrode leads to a higher number of ineffective shocks; these ineffective shocks therefore do not lead to a release of cTnT and cTnI from the myocardium. One patient received two additional external 360-J shocks. This patient showed only marginal increases in the specific markers cTnT, cTnI, and CK-MB mass, and large increases in the nonspecific markers CK and myoglobin. External shocks may lead only to a release from the thoracic skeletal muscles (11)(12)(13)(14), which explains the large increases in CK activity and myoglobin in this patient.
Our results correlate well with those of Hurst et al. (15)
and Runsiö (10), who also found increased serum
concentrations of cTnT and cTnI or cTnT and CK-MB after ICD
implantation (Table 4
). In contrast to these studies, we found a clearly visible
lower postoperative peak value for cTnT, but also lower peak values for
cTnI and CK-MB. This discrepancies are attributable to the lower
cumulative energy used in our patients, but the assays used for
determining the cardiac-specific markers, especially for cTnT, are also
important influencing factors. For measuring cTnT, we used a highly
specific second-generation assay (9) with negligible
cross-reactivity with skeletal muscle troponin T (6). The
cTnT assay used by Hurst et al. (15) and Runsiö et al.
(10) shows 2% cross-reactivity with skeletal muscle
troponin T (16), leading to possible false-positive results.
|
Troponins T and I are structurally bound regulator proteins of the
contractile apparatus of the skeletal muscle and myocardium. Only a
small portion of both can be found as a soluble portion in the
cytoplasm. For cTnT, this is
68%; for cTnI, it is 2.5%
(17). The release into the circulation is therefore
characterized by two steps. In the first step, the cytoplasmic pool is
released simultaneously with other cytoplasmic proteins, such as CK,
CK-MB, and myoglobin. After loss of integrity of the contractile
apparatus, cTnT and cTnI are released continuously until the infarcted
area is healed. The first release can be measured in the blood
4 h
after damage, whereas the second release will occur into the serum
after 24 h (18). Additionally, the half-time of both
cTnT and cTnI is
2 h (10). As shown in the
Results, cTnT, cTnI, and CK-MB show a peak value after
4 h. Therefore, we suggest that the release of cTnT and cTnI is
only from the small amount located in the cytoplasm. Cell necrosis
would cause a second, additional release of disintegrated structurally
bound proteins into circulation, which did not occur in our patients.
CK showed additional increases even after 24 h. Each testing
episode caused a short skeletal muscle spasm that could explain the
further increase of CK after 24 h.
In conclusion, ICD implantation and testing often leads to a short release of cardiac markers into the circulation. The effectiveness of the individual shock, depending on the positioning of the electrode and the position of the heart in the electrical field, seems to influence the release of the cardiac-specific markers. Because the cardiac-specific markers cTnT and cTnI peak after 4 h, theoretically no cell necrosis can occur. It seems to be rather from cytoplasmic origin as a result of a reversible change of permeability of the cellular membrane.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
D Blendea, M Blendea, J Banker, and C A McPherson Troponin T elevation after implanted defibrillator discharge predicts survival Heart, July 15, 2009; 95(14): 1153 - 1158. [Abstract] [Full Text] [PDF] |
||||
![]() |