| Abstract |
|---|
|
|
|---|
Methods: Busulfan concentrations were quantified using 200 µL of plasma and liquidliquid extraction with diethyl ether after the addition of [2H8]busulfan as the internal standard. Separation and detection of busulfan and [2H8]busulfan were achieved with a LUNA C8 column (5 µm; 150 x 2 mm i.d.) at 30 °C, a HP 1100 liquid chromatography system, and a HP 1100 single-quadrupole mass spectrometer. Busulfan and [2H8]busulfan were detected as ammonium adducts in selected-ion monitoring mode at m/z 264.2 and 272.2, respectively.
Results: The calibration curve was linear at 52000 µg/L busulfan. Intra- and interassay imprecision (CV) and bias were both <11%. The limits of detection and quantification were 2 and 5 µg/L, respectively. Extraction recovery of busulfan was >87%. Analysis of pharmacokinetics in four patients receiving high-dose busulfan indicated that minimum busulfan concentrations before the next dose were 405603 µg/L, with no interference observed.
Conclusions: The new rapid and sensitive liquid chromatographicmass spectrometric assay is an appropriate method for quantification of busulfan in human plasma, making therapeutic drug monitoring of busulfan faster and easier in clinical practice.
| Introduction |
|---|
|
|
|---|
Different HPLC and gas chromatographic methods for the determination of busulfan in plasma and other biologic fluids have been described previously (10)(11)(12)(13)(14)(15)(16)(17), but these have several disadvantages. For example, because of the limited absorbance of busulfan over the ultraviolet wavelength spectrum, derivatization of busulfan and the internal standard before assay is necessary (10)(11)(12)(14). In addition, there are some limitations in respect to the sensitivity of previously described methods. For example, HPLC analyses have limits of quantification between 50 and 60 µg/L (11)(12)(14)(15), whereas two gas chromatographic methods (13)(16) have limits of quantification of 20 and 40 µg/L, respectively. Furthermore, with the exception of only two methods (12)(15), 1 mL of plasma is required as a minimum sample volume, which seems to be unsuitable for clinical TDM in small children. However, gas chromatographic analysis is time-consuming because sample preparation takes 34 h. We describe here a relatively simple, rapid, and sensitive liquid chromatographymass spectrometry (LC-MS) assay for quantification of busulfan in human plasma.
| Materials and Methods |
|---|
|
|
|---|
solutions
Stock solutions of busulfan and [2H8]busulfan were prepared in dimethyl formamide at a concentration of 10 g/L. Working solutions were diluted in acetonitrile to obtain final concentrations of 1 mg/L, 10 µg/L, and 1 µg/L busulfan. A solution of [2H8]busulfan in acetonitrile (20 mg/L) was used as the internal standard. All stock solutions were stored at -20 °C for up to 3 months. Calibrators at concentrations of 52000 µg/L were prepared by supplementing busulfan working solutions with 200 µL of busulfan-free plasma. Quality controls were obtained by adding busulfan to 10 mL of blank plasma to obtain final concentrations of 5, 10, 100, 250, and 2000 µg/L. Aliquots were frozen at -20 °C and analyzed together with patient samples. Two mobile phases, designated A and B, were as follows: 10 mmol/L ammonium acetate and 10 mL/L acetic acid in water (A) and acetonitrile (B).
sample preparation
Internal standard (10 µL; 20 mg/L [2H8]busulfan) and 300 µL of a saturated solution of sodium chloride were added to 200 µL of plasma in a 10-mL glass tube. The samples were extracted with 4 mL of diethyl ether by shaking for 15 min and centrifuging for 5 min at 1200g. The organic layer was evaporated to dryness under a stream of nitrogen at 37 °C. The residue was dissolved in 120 µL of a mixture of mobile phase Amobile phase B (85:15 by volume).
lc-ms analysis
Chromatographic separation of busulfan and [2H8]busulfan was performed using a LUNA C8 analytical column (5 µm particle size; 150 x 2 mm i.d.; Phenomenex) at a column temperature of 30 °C, and a HP 1100 LC system (Agilent Technologies) with a binary pump. Gradients were programmed as follows (flow rate, 0.5 mL/min): 15% mobile phase B for 2 min, increased to 50% B for 6 min, decreased to 15% B for 1 min, and then reequilibrated with 15% B for 2 min, until the next sample was injected. A column-switching valve was used to switch the liquid flow into the detector 48 min after injection.
A HP1100 single-quadrupole mass spectrometer (Agilent Technologies) equipped with an electrospray source was used. Electrospray settings were as follows: spray voltage, 2000V; fragmentor, 60V; gas temperature, 345 °C; nebulizer pressure, 50 psi; drying gas, 11 L/min. Positive ionization with selected-ion monitoring mode was used, and the multiplier gain was set on 1. Busulfan and [2H8]busulfan were detected as ammonium adducts at m/z 264.2 and 272.2, respectively. Peak areas of busulfan and [2H8]busulfan were analyzed and subsequent calculation of calibration curves and quantification of busulfan was performed with HP Chemstation software (Ver. 98; Agilent Technologies).
assay validation
Calibration curve samples with final concentrations of 10, 20, 100, 500, 1000, and 2000 µg/L busulfan were prepared as described above. Calibration curves were obtained by plotting the peak area ratio m/z 264.2/272.2 of busulfan and [2H8]busulfan, respectively, against the initial busulfan concentrations. Intraday accuracy and precision were assessed by assaying five replicates each of 5, 10, 25, and 250 µg/L busulfan quality-control samples on a single assay day. Interday accuracy and precision were determined by analysis of 10, 100, and 2000 µg/L busulfan quality-control samples on 6 different assay days. Calibration curves were obtained by plotting the peak area ratio m/z 264.2/272.2 of busulfan and [2H8]busulfan, respectively, against the initial busulfan concentrations. The limit of detection was defined as the concentration that gave a signal-to-noise ratio of 3, and the limit of quantification was determined as the lowest concentration with a CV of <20% (n = 5).
Recovery of busulfan was investigated by adding known concentrations (10, 100, and 2000 µg/L) of busulfan to blank plasma samples and extracting the busulfan. Peak area ratios were compared with those obtained by direct injection of the corresponding busulfan concentrations.
Long-term stability (3 months) of busulfan in plasma at -20 °C was investigated via repeated analysis of patient samples. In addition, the stability of busulfan in plasma at 4 °C and at room temperature for up to 6 days was assessed, as well as the stability of busulfan in extracted samples at room temperature and at -20 °C.
clinical samples
Four patients with different kinds of malignancies [high-grade non-Hodgkin lymphoma (n = 2), Hodgkin disease (n = 1), and plasmocytoma (n = 1)] were treated with BusulfexTM [6 g/L solution of busulfan dissolved in N,N-dimethylacetamidepolyethylene glycol (33:67 by volume); Orphan Medical] in conditioning regimes for HSCT. Three patients were male and one was female (mean age, 37.8 years; age range, 1651 years; refer to Table 1
for demographic data), and all patients gave written informed consent before the study. Plasma samples were collected before administration of the eighth dose of busulfan, given as an infusion over 2 h and at 0, 0.5, 1, 2, 3, 4, 6, 12, 18, and 26 h after the end of the infusion. Blank human plasma was obtained from healthy volunteers. Samples were stored at -20 °C until further processing.
|
data analysis
Stability of busulfan after 3 months storage was analyzed with non-least-squares linear regression (GraphPad PRISM, Ver. 2.0; GraphPad Software Inc.). Pharmacokinetic data were calculated according to the standard noncompartmental model with the TopFit 2.1 program (18).
| Results |
|---|
|
|
|---|
chromatographic separation
Chromatograms resulting from the extractions of a blank plasma sample, a quality-control sample at the limit of quantification (5 µg/L), and plasma samples from one patient are shown in Fig. 1
. Under the LC-MS conditions described, the retention times of busulfan and [2H8]busulfan were 5.8 and 5.7 min, respectively. Because of the highly selective detection method, there were no interfering peaks present in 10 blank samples investigated in the time window of 46.5 min when the analytes were eluted from the column. The gradient program was used to ensure that all contaminants were washed from the column within a run time of <10 min, ensuring a prolonged life of the analytical column. No deterioration of the column was observed in >350 injections of the extracted plasma samples.
|
assay validation
Calibration curves were linear over the concentration range used for busulfan, with correlation coefficients of 0.99920.9999 (n = 4). The mean (± SD) y-intercept was 0.000975 (±0.00692), and the mean (± SD) slope was 0.00093 (±0.000007) over 4 independent assay days.
The accuracy and precision values are shown in Table 2
. The limit of detection and limit of quantification were 2 and 5 µg/L, respectively. Recoveries of 10, 100, and 2000 µg/L busulfan after extraction were >87% (Table 3
).
|
|
Busulfan is known to react with nucleophilic groups, such as amines, in a substitution reaction. This reaction may occur during storage; therefore, the stability of busulfan in plasma samples under different conditions was investigated. After 3 months of storage at -20 °C, the busulfan concentration differed by a mean of 2.9% (range, -7.7% to 12.6%) from that originally quantified (concentration range, 91895 µg/L), indicating long-term stability. Furthermore, because there is a possibility that samples may be left accidentally at room temperature or in a typical refrigerator for some time during acquisition or processing, stability at 4 °C and at room temperature was investigated. The degradation of 105 µg/L busulfan in plasma under these conditions is shown in Fig. 2
. Although busulfan is stable for up to 24 h at 4 °C, the busulfan concentration decreased to
75% within 24 h at room temperature. After 6 days at room temperature, only 20% of the initial concentration of busulfan remained. Extracted samples redissolved in mobile phase were stable for 12 h at room temperature and for up to 1 month at -20 °C.
|
clinical samples
Steady-state plasma concentrations (CSSs) of busulfan at and after the administration of the eighth busulfan dose are shown in Fig. 3
. The mean minimum busulfan plasma concentration before the eighth dose in the four patients was 497 µg/L (range, 405603 µg/L). The mean area under the plasma concentrationtime curve [(AUC); 026 h] for busulfan was 1839 µmol/L · min. Plasma concentrations of busulfan displayed linear pharmacokinetic behavior with a mean elimination half-life of 3.6 h. Complete pharmacokinetic data of all patients are given in Table 4
.
|
|
| Discussion |
|---|
|
|
|---|
Busulfan has been used since the 1950s, and currently it is one of the most frequently used chemotherapeutic agents in high-dose regimes for patients undergoing HSCT (1). The majority of clinical studies investigating pharmacokinetics of busulfan have demonstrated a pharmacodynamic relationship between busulfan CSS and outcome in patients with busulfancyclophosphamid preparative regimens in HSCT, as well as busulfan-related toxicity (for example, venoocclusive disease or intestinal pneumonia) (9). Thus, several investigators have recommended TDM of busulfan to maximize the likelihood of engraftment and reduce busulfan-related toxicity (9)(22).
This report describes a new, highly sensitive and specific LC-MS assay for the quantification of busulfan in small volumes of patient plasma. This method requires only 200 µL of human plasma for accurate determination of busulfan concentrations because of a low limit of quantification (5 µg/L), which is 10-fold lower than the majority of previous reports (11)(12)(14)(15). Furthermore, this volume is more appropriate when attempting TDM of plasma busulfan concentrations in children when conducting AUC-guided administration.
A further advantage of this method is that it does not require a derivatization step in contrast to the majority of methods described previously (10)(11)(12)(13)(14)(15)(16)(23). In addition, a simple liquidliquid extraction was sufficient to produce chromatography free from interference of endogenous compounds in plasma. Although most previous methods have also used liquidliquid extraction, the present method used diethyl ether as the solvent in comparison with diethyl ether-dichloromethane (10), ethyl-acetate (11)(12)(13)(16), toluene (14), or dichloromethane (23). Of all of these solvents, diethyl ether is the quickest to evaporate under N2 at a lower temperature and, compared with chlorinated solvents, has a lower environmental hazard potential. Taken together, the use of diethyl ether for the extraction of busulfan from plasma and the absence of a derivatization step markedly reduce assay time. Zuccaro et al. (17) described a LC-MS method using particle beam evaporation together with electron-impact ionization. In addition to a comparatively low sensitivity with a limit of quantification of 100 µg/L, this kind of LC-MS system is too temperamental for routine laboratory use.
Apart from assay duration, accuracy and precision as well as extraction recovery must also be considered when assessing the quality of any particular method. Intra- and interassay validation data demonstrate that even at the limit of quantification of the present assay, accuracy and precision are retained (see Tables 2
and 3
). Recovery of the internal standard, [2H8]busulfan, was not assessed because of the high degree of similarity with busulfan; however, recoveries of busulfan were 8792% over a concentration range of 102000 µg/L. Previous methods have reported recoveries of 90102% (10)(12)(13)(14)(15)(16). Therefore, it can be concluded that the method of extraction in the present assay, although quick and simple, is just as effective as previous methods used. Unless blood samples are left at room temperature or in a typical refrigerator for more than a few hours or up to 1 day, respectively, the described method is well suited for the rapid and accurate TDM of busulfan during conditioning before HSCT. Additionally, the new formulations of intravenous busulfan, such as Busulfex (dissolved in dimethylacetamide) and a liposomal form, require clinical trials that include an evaluation of pharmacokinetic parameters, especially in children. Thus, the present method may also be helpful for clinical research.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
A. J. L. Cooper, I. R. Younis, Z. V. Niatsetskaya, B. F. Krasnikov, J. T. Pinto, W. P. Petros, and P. S. Callery Metabolism of the Cysteine S-Conjugate of Busulfan Involves a {beta}-Lyase Reaction Drug Metab. Dispos., August 1, 2008; 36(8): 1546 - 1552. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Hempel, D. Oechtering, C. Lanvers-Kaminsky, T. Klingebiel, J. Vormoor, B. Gruhn, and J. Boos Cytotoxicity of Dimethylacetamide and Pharmacokinetics in Children Receiving Intravenous Busulfan J. Clin. Oncol., May 1, 2007; 25(13): 1772 - 1778. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |