Clinical Chemistry 47: 1437-1442, 2001;
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (10)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mürdter, T. E.
Right arrow Articles by Schwab, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mürdter, T. E.
Right arrow Articles by Schwab, M.
Related Collections
Right arrow Drug Monitoring and Toxicology
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2001;47:1437-1442.)
© 2001 American Association for Clinical Chemistry, Inc.


Articles

Sensitive and Rapid Quantification of Busulfan in Small Plasma Volumes by Liquid Chromatography–Electrospray Mass Spectrometry

Thomas E. Mürdter1a, Janet Coller1, Alexander Claviez2, Frank Schönberger1, Ute Hofmann1, Peter Dreger2 and Matthias Schwab1

1 Dr. Margarete Fischer-Bosch-Institute of Clinical Pharmacology, Auerbachstrasse 112, D-70376 Stuttgart, Germany.
2 Bone Marrow Transplant Unit, University of Kiel, Schwanenweg 20, D-24105 Kiel, Germany

aAuthor for correspondence. Fax 49-0711-859295; e-mail thomas.muerdter{at}ikp-stuttgart.de.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Background: High-dose busulfan is widely used in conditioning regimens before hematopoietic stem cell transplantation in both adults and children. Large interindividual variability in pharmacokinetics after oral administration has been reported; therefore, therapeutic drug monitoring of busulfan may decrease the incidence of drug-related toxicity (for example, hepatic venoocclusive disease) and may also improve therapeutic efficacy.

Methods: Busulfan concentrations were quantified using 200 µL of plasma and liquid–liquid extraction with diethyl ether after the addition of [2H8]busulfan as the internal standard. Separation and detection of busulfan and [2H8]busulfan were achieved with a LUNA C8 column (5 µm; 150 x 2 mm i.d.) at 30 °C, a HP 1100 liquid chromatography system, and a HP 1100 single-quadrupole mass spectrometer. Busulfan and [2H8]busulfan were detected as ammonium adducts in selected-ion monitoring mode at m/z 264.2 and 272.2, respectively.

Results: The calibration curve was linear at 5–2000 µg/L busulfan. Intra- and interassay imprecision (CV) and bias were both <11%. The limits of detection and quantification were 2 and 5 µg/L, respectively. Extraction recovery of busulfan was >87%. Analysis of pharmacokinetics in four patients receiving high-dose busulfan indicated that minimum busulfan concentrations before the next dose were 405–603 µg/L, with no interference observed.

Conclusions: The new rapid and sensitive liquid chromatographic–mass spectrometric assay is an appropriate method for quantification of busulfan in human plasma, making therapeutic drug monitoring of busulfan faster and easier in clinical practice.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
High-dose busulfan is widely used as a conditioning treatment before hematopoietic stem cell transplantation (HSCT) 1 in both adults and children (1). Because of highly variable bioavailability after oral administration, plasma concentrations of busulfan display high inter- and intraindividual variability (2). Consequently, using a standard dosing regimen of busulfan of 4 mg · kg-1 · day, patients with lower plasma concentrations and lower systemic exposure have been reported to have an increased risk of engraftment failure and late relapse after bone marrow transplantation or HSCT (3)(4). In contrast, high systemic exposure of busulfan has been correlated with an increased risk of drug-related toxicity. For example, the incidence of hepatic venoocclusive disease, a hepatic lesion involving obstruction of small intrahepatic venules and subsequent damage to the surrounding centrilobular hepatocytes, was observed in up to 20–40% of children and adults who underwent bone marrow transplantation (5)(6). Therefore, therapeutic drug monitoring (TDM) has been considered of benefit for individual optimization of busulfan therapy (7)(8)(9).

Different HPLC and gas chromatographic methods for the determination of busulfan in plasma and other biologic fluids have been described previously (10)(11)(12)(13)(14)(15)(16)(17), but these have several disadvantages. For example, because of the limited absorbance of busulfan over the ultraviolet wavelength spectrum, derivatization of busulfan and the internal standard before assay is necessary (10)(11)(12)(14). In addition, there are some limitations in respect to the sensitivity of previously described methods. For example, HPLC analyses have limits of quantification between 50 and 60 µg/L (11)(12)(14)(15), whereas two gas chromatographic methods (13)(16) have limits of quantification of 20 and 40 µg/L, respectively. Furthermore, with the exception of only two methods (12)(15), 1 mL of plasma is required as a minimum sample volume, which seems to be unsuitable for clinical TDM in small children. However, gas chromatographic analysis is time-consuming because sample preparation takes 3–4 h. We describe here a relatively simple, rapid, and sensitive liquid chromatography–mass spectrometry (LC-MS) assay for quantification of busulfan in human plasma.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
materials
Busulfan was purchased from Sigma. The internal standard, [2H8]busulfan, was obtained from Euriso-Top. Other materials were obtained from the following sources: JT Baker (acetonitrile), Merck (acetic acid, ammonium acetate, and sodium chloride); Promochem (diethyl ether); and Fluka Chemie AG (dimethyl formamide).

solutions
Stock solutions of busulfan and [2H8]busulfan were prepared in dimethyl formamide at a concentration of 10 g/L. Working solutions were diluted in acetonitrile to obtain final concentrations of 1 mg/L, 10 µg/L, and 1 µg/L busulfan. A solution of [2H8]busulfan in acetonitrile (20 mg/L) was used as the internal standard. All stock solutions were stored at -20 °C for up to 3 months. Calibrators at concentrations of 5–2000 µg/L were prepared by supplementing busulfan working solutions with 200 µL of busulfan-free plasma. Quality controls were obtained by adding busulfan to 10 mL of blank plasma to obtain final concentrations of 5, 10, 100, 250, and 2000 µg/L. Aliquots were frozen at -20 °C and analyzed together with patient samples. Two mobile phases, designated A and B, were as follows: 10 mmol/L ammonium acetate and 10 mL/L acetic acid in water (A) and acetonitrile (B).

sample preparation
Internal standard (10 µL; 20 mg/L [2H8]busulfan) and 300 µL of a saturated solution of sodium chloride were added to 200 µL of plasma in a 10-mL glass tube. The samples were extracted with 4 mL of diethyl ether by shaking for 15 min and centrifuging for 5 min at 1200g. The organic layer was evaporated to dryness under a stream of nitrogen at 37 °C. The residue was dissolved in 120 µL of a mixture of mobile phase A–mobile phase B (85:15 by volume).

lc-ms analysis
Chromatographic separation of busulfan and [2H8]busulfan was performed using a LUNA C8 analytical column (5 µm particle size; 150 x 2 mm i.d.; Phenomenex) at a column temperature of 30 °C, and a HP 1100 LC system (Agilent Technologies) with a binary pump. Gradients were programmed as follows (flow rate, 0.5 mL/min): 15% mobile phase B for 2 min, increased to 50% B for 6 min, decreased to 15% B for 1 min, and then reequilibrated with 15% B for 2 min, until the next sample was injected. A column-switching valve was used to switch the liquid flow into the detector 4–8 min after injection.

A HP1100 single-quadrupole mass spectrometer (Agilent Technologies) equipped with an electrospray source was used. Electrospray settings were as follows: spray voltage, 2000V; fragmentor, 60V; gas temperature, 345 °C; nebulizer pressure, 50 psi; drying gas, 11 L/min. Positive ionization with selected-ion monitoring mode was used, and the multiplier gain was set on 1. Busulfan and [2H8]busulfan were detected as ammonium adducts at m/z 264.2 and 272.2, respectively. Peak areas of busulfan and [2H8]busulfan were analyzed and subsequent calculation of calibration curves and quantification of busulfan was performed with HP Chemstation software (Ver. 98; Agilent Technologies).

assay validation
Calibration curve samples with final concentrations of 10, 20, 100, 500, 1000, and 2000 µg/L busulfan were prepared as described above. Calibration curves were obtained by plotting the peak area ratio m/z 264.2/272.2 of busulfan and [2H8]busulfan, respectively, against the initial busulfan concentrations. Intraday accuracy and precision were assessed by assaying five replicates each of 5, 10, 25, and 250 µg/L busulfan quality-control samples on a single assay day. Interday accuracy and precision were determined by analysis of 10, 100, and 2000 µg/L busulfan quality-control samples on 6 different assay days. Calibration curves were obtained by plotting the peak area ratio m/z 264.2/272.2 of busulfan and [2H8]busulfan, respectively, against the initial busulfan concentrations. The limit of detection was defined as the concentration that gave a signal-to-noise ratio of 3, and the limit of quantification was determined as the lowest concentration with a CV of <20% (n = 5).

Recovery of busulfan was investigated by adding known concentrations (10, 100, and 2000 µg/L) of busulfan to blank plasma samples and extracting the busulfan. Peak area ratios were compared with those obtained by direct injection of the corresponding busulfan concentrations.

Long-term stability (3 months) of busulfan in plasma at -20 °C was investigated via repeated analysis of patient samples. In addition, the stability of busulfan in plasma at 4 °C and at room temperature for up to 6 days was assessed, as well as the stability of busulfan in extracted samples at room temperature and at -20 °C.

clinical samples
Four patients with different kinds of malignancies [high-grade non-Hodgkin lymphoma (n = 2), Hodgkin disease (n = 1), and plasmocytoma (n = 1)] were treated with BusulfexTM [6 g/L solution of busulfan dissolved in N,N-dimethylacetamide–polyethylene glycol (33:67 by volume); Orphan Medical] in conditioning regimes for HSCT. Three patients were male and one was female (mean age, 37.8 years; age range, 16–51 years; refer to Table 1 for demographic data), and all patients gave written informed consent before the study. Plasma samples were collected before administration of the eighth dose of busulfan, given as an infusion over 2 h and at 0, 0.5, 1, 2, 3, 4, 6, 12, 18, and 26 h after the end of the infusion. Blank human plasma was obtained from healthy volunteers. Samples were stored at -20 °C until further processing.


View this table:
[in this window]
[in a new window]
 
Table 1. Patient demographics.

data analysis
Stability of busulfan after 3 months storage was analyzed with non-least-squares linear regression (GraphPad PRISM, Ver. 2.0; GraphPad Software Inc.). Pharmacokinetic data were calculated according to the standard noncompartmental model with the TopFit 2.1 program (18).


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
electrospray ionization conditions
Different mobile phases were tested to optimize the sensitivity of the electrospray ionization. There was only a minor influence of the organic solvent used (500 mL/L methanol vs 500 mL/L acetonitrile). In the presence of acetic acid, the spectra showed a base peak at m/z 269.2 [M + Na+] with additional peaks at m/z 247.2 [M + H+], m/z 264.2 [M + NH4+], and m/z 285 [M + K+] of variable intensity. When we used 10 mmol/L ammonium acetate–acetonitrile (50:50 by volume) as the solvent, the base peak was [M + NH4+] with other peaks {[M + Na+] and [M + K+]} of only minor intensity. However, the overall sensitivity was insufficient. Decreasing the pH of the buffer solution by the addition of 10 mL/L acetic acid markedly improved signal height to optimize the sensitivity.

chromatographic separation
Chromatograms resulting from the extractions of a blank plasma sample, a quality-control sample at the limit of quantification (5 µg/L), and plasma samples from one patient are shown in Fig. 1 . Under the LC-MS conditions described, the retention times of busulfan and [2H8]busulfan were 5.8 and 5.7 min, respectively. Because of the highly selective detection method, there were no interfering peaks present in 10 blank samples investigated in the time window of 4–6.5 min when the analytes were eluted from the column. The gradient program was used to ensure that all contaminants were washed from the column within a run time of <10 min, ensuring a prolonged life of the analytical column. No deterioration of the column was observed in >350 injections of the extracted plasma samples.



View larger version (20K):
[in this window]
[in a new window]
 
Figure 1. Representative chromatograms from extractions of a blank plasma (A); 5 µg/L busulfan and 400 µg/L [2H8]busulfan (B); and a patient plasma sample (C).

assay validation
Calibration curves were linear over the concentration range used for busulfan, with correlation coefficients of 0.9992–0.9999 (n = 4). The mean (± SD) y-intercept was 0.000975 (±0.00692), and the mean (± SD) slope was 0.00093 (±0.000007) over 4 independent assay days.

The accuracy and precision values are shown in Table 2 . The limit of detection and limit of quantification were 2 and 5 µg/L, respectively. Recoveries of 10, 100, and 2000 µg/L busulfan after extraction were >87% (Table 3 ).


View this table:
[in this window]
[in a new window]
 
Table 2. Intra- and Interday accuracy and precision of busulfan samples.


View this table:
[in this window]
[in a new window]
 
Table 3. Recovery of busulfan from extracted plasma samples.1

Busulfan is known to react with nucleophilic groups, such as amines, in a substitution reaction. This reaction may occur during storage; therefore, the stability of busulfan in plasma samples under different conditions was investigated. After 3 months of storage at -20 °C, the busulfan concentration differed by a mean of 2.9% (range, -7.7% to 12.6%) from that originally quantified (concentration range, 9–1895 µg/L), indicating long-term stability. Furthermore, because there is a possibility that samples may be left accidentally at room temperature or in a typical refrigerator for some time during acquisition or processing, stability at 4 °C and at room temperature was investigated. The degradation of 105 µg/L busulfan in plasma under these conditions is shown in Fig. 2 . Although busulfan is stable for up to 24 h at 4 °C, the busulfan concentration decreased to ~75% within 24 h at room temperature. After 6 days at room temperature, only 20% of the initial concentration of busulfan remained. Extracted samples redissolved in mobile phase were stable for 12 h at room temperature and for up to 1 month at -20 °C.



View larger version (14K):
[in this window]
[in a new window]
 
Figure 2. Stability of busulfan at room temperature (•), at 4 °C ({blacksquare}), and at -20 °C ({diamondsuit}) over a period of 144 h.

clinical samples
Steady-state plasma concentrations (CSSs) of busulfan at and after the administration of the eighth busulfan dose are shown in Fig. 3 . The mean minimum busulfan plasma concentration before the eighth dose in the four patients was 497 µg/L (range, 405–603 µg/L). The mean area under the plasma concentration–time curve [(AUC); 0–26 h] for busulfan was 1839 µmol/L · min. Plasma concentrations of busulfan displayed linear pharmacokinetic behavior with a mean elimination half-life of 3.6 h. Complete pharmacokinetic data of all patients are given in Table 4 .



View larger version (13K):
[in this window]
[in a new window]
 
Figure 3. CSS of busulfan (µg/L) after administration of the final (eighth) dose in four patients. Data are means; error bars indicate SD.


View this table:
[in this window]
[in a new window]
 
Table 4. Pharmacokinetic data from patients derived from the eighth dose of intravenous busulfan (Busulfex).


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
In the last 10 years, pharmacokinetically guided administration of chemotherapeutic agents has become of increasing interest because the assessment of pharmacokinetic parameters such as AUC, drug CSS, and drug exposure time above a certain plasma concentration may lead to less toxicity and higher efficacy compared with administration on the basis of body surface area only. For example, AUC-guided administration has been investigated extensively for carboplatin, leading to a more precise prediction of the effect of carboplatin in terms of dose-limiting toxicity mainly as thrombocytopenia (19)(20)(21).

Busulfan has been used since the 1950s, and currently it is one of the most frequently used chemotherapeutic agents in high-dose regimes for patients undergoing HSCT (1). The majority of clinical studies investigating pharmacokinetics of busulfan have demonstrated a pharmacodynamic relationship between busulfan CSS and outcome in patients with busulfan–cyclophosphamid preparative regimens in HSCT, as well as busulfan-related toxicity (for example, venoocclusive disease or intestinal pneumonia) (9). Thus, several investigators have recommended TDM of busulfan to maximize the likelihood of engraftment and reduce busulfan-related toxicity (9)(22).

This report describes a new, highly sensitive and specific LC-MS assay for the quantification of busulfan in small volumes of patient plasma. This method requires only 200 µL of human plasma for accurate determination of busulfan concentrations because of a low limit of quantification (5 µg/L), which is 10-fold lower than the majority of previous reports (11)(12)(14)(15). Furthermore, this volume is more appropriate when attempting TDM of plasma busulfan concentrations in children when conducting AUC-guided administration.

A further advantage of this method is that it does not require a derivatization step in contrast to the majority of methods described previously (10)(11)(12)(13)(14)(15)(16)(23). In addition, a simple liquid–liquid extraction was sufficient to produce chromatography free from interference of endogenous compounds in plasma. Although most previous methods have also used liquid–liquid extraction, the present method used diethyl ether as the solvent in comparison with diethyl ether-dichloromethane (10), ethyl-acetate (11)(12)(13)(16), toluene (14), or dichloromethane (23). Of all of these solvents, diethyl ether is the quickest to evaporate under N2 at a lower temperature and, compared with chlorinated solvents, has a lower environmental hazard potential. Taken together, the use of diethyl ether for the extraction of busulfan from plasma and the absence of a derivatization step markedly reduce assay time. Zuccaro et al. (17) described a LC-MS method using particle beam evaporation together with electron-impact ionization. In addition to a comparatively low sensitivity with a limit of quantification of 100 µg/L, this kind of LC-MS system is too temperamental for routine laboratory use.

Apart from assay duration, accuracy and precision as well as extraction recovery must also be considered when assessing the quality of any particular method. Intra- and interassay validation data demonstrate that even at the limit of quantification of the present assay, accuracy and precision are retained (see Tables 2Up and 3Up ). Recovery of the internal standard, [2H8]busulfan, was not assessed because of the high degree of similarity with busulfan; however, recoveries of busulfan were 87–92% over a concentration range of 10–2000 µg/L. Previous methods have reported recoveries of 90–102% (10)(12)(13)(14)(15)(16). Therefore, it can be concluded that the method of extraction in the present assay, although quick and simple, is just as effective as previous methods used. Unless blood samples are left at room temperature or in a typical refrigerator for more than a few hours or up to 1 day, respectively, the described method is well suited for the rapid and accurate TDM of busulfan during conditioning before HSCT. Additionally, the new formulations of intravenous busulfan, such as Busulfex (dissolved in dimethylacetamide) and a liposomal form, require clinical trials that include an evaluation of pharmacokinetic parameters, especially in children. Thus, the present method may also be helpful for clinical research.


   Acknowledgments
 
This research was supported by the Robert Bosch Foundation (Stuttgart, Germany) and Federal Ministry for Education and Research (Berlin, Germany) Grant 01 GG 9846.


   Footnotes
 
1 Nonstandard abbreviations: HSCT, hematopoietic stem cell transplantation; TDM, therapeutic drug monitoring; LC-MS, liquid chromatography–mass spectrometry; CSS, steady-state plasma concentration; and AUC, area under the concentration–time curve.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Hassan M. The role of busulfan in bone marrow transplantation. Med Oncol 1999;16:166-176.[ISI][Medline] [Order article via Infotrieve]
  2. Hassan M, Ljungman P, Bolme P, Ringden O, Syruckova Z, Bekassy A, et al. Busulfan bioavailability. Blood 1994;84:2144-2250.[Abstract/Free Full Text]
  3. Grochow LB. Busulfan disposition: the role of therapeutic monitoring in bone marrow transplantation induction regimens. Semin Oncol 1993;20:18-25.[ISI][Medline] [Order article via Infotrieve]
  4. Vassal G, Koscielny S, Challine D, Valteau-Couanet D, Boland I, Deroussent A, et al. Busulfan disposition and hepatic veno-occlusive disease in children undergoing bone marrow transplantation. Cancer Chemother Pharmacol 1996;37:247-253.[ISI][Medline] [Order article via Infotrieve]
  5. Jones RJ, Lee KS, Beschorner WE, Vogel VG, Grochow LB, Braine HG, et al. Venoocclusive disease of the liver following bone marrow transplantation. Transplantation 1987;44:778-783.[ISI][Medline] [Order article via Infotrieve]
  6. McDonald GB, Hinds MS, Fisher LD, Schoch HG, Wolford JL, Banaji M, et al. Veno-occlusive disease of the liver and multiorgan failure after bone marrow transplantation: a cohort study of 355 patients. Ann Intern Med 1993;118:255-267.[Abstract/Free Full Text]
  7. Schuler U, Schroer S, Kuhnle A, Blanz J, Mewes K, Kumbier I, et al. Busulfan pharmacokinetics in bone marrow transplant patients: is drug monitoring warranted?. Bone Marrow Transplant 1994;14:759-765.[ISI][Medline] [Order article via Infotrieve]
  8. Vassal G. Pharmacologically-guided dose adjustment of busulfan in high-dose chemotherapy regimens: rationale and pitfalls [Review]. Anticancer Res 1994;14:2363-2370.[ISI][Medline] [Order article via Infotrieve]
  9. McCune JS, Gibbs JP, Slattery JT. Plasma concentration monitoring of busulfan: does it improve clinical outcome?. Clin Pharmacokinet 2000;39:155-165.[ISI][Medline] [Order article via Infotrieve]
  10. Funakoshi K-I, Yamashita K, Chao W-F, Yamaguchi M, Yashiki T. High-performance liquid chromatographic determination of busulfan in human serum with on-line derivatization, column switching and ultraviolet absorbance detection. J Chromatogr B Biomed Appl 1994;660:200-204.[ISI][Medline] [Order article via Infotrieve]
  11. Heggie JR, Wu M, Burns RB, Ng CS, Fung HC, Knight G, et al. Validation of a high-performance liquid chromatography assay method for pharmacokinetic evaluation of busulfan. J Chromatogr B Biomed Appl 1997;692:437-444.
  12. Rifai N, Sakamoto M, Lafi M, Guinan E. Measurement of plasma busulfan concentration by high-performance liquid chromatography with ultraviolet detection. Ther Drug Monit 1997;19:169-174.[ISI][Medline] [Order article via Infotrieve]
  13. Quernin M-H, Poonkuzhali B, Montes C, Krishnamoorthy R, Dennison D, Srivastava A, et al. Quantification of busulfan in plasma by gas chromatography-mass spectrometry following derivatization with tetrafluorothiophenol. J Chromatogr B Biomed Appl 1998;709:47-56.
  14. Quernin M-H, Poonkuzhali B, Medard Y, Dennison D, Srivastava A, Krishnamoorthy R, et al. High-performance liquid chromatographic method for quantification of busulfan in plasma after derivatization by tetrefluorothiophenol. J Chromatogr B Biomed Appl 1999;721:147-152.
  15. Bleyzac N, Barou P, Aulagner G. Rapid and sensitive high-performance liquid chromatographic method for busulfan assay in plasma. J Chromatogr B Biomed Appl 2000;742:427-432.
  16. Lai W-K, Pang C-P, Law L-K, Wong R, Li C-K, Yuen PM-P. Routine analysis of plasma busulfan by gas chromatography-mass fragmentography. Clin Chem 1998;44:2506-2510.[Abstract/Free Full Text]
  17. Pichini S, Altieri I, Bacosi A, Di Carlo S, Zuccaro P, Iannetti P, et al. High-performance liquid chromatographic-mass spectrometric assay of busulfan in serum and cerebrospinal fluid. J Chromatogr 1992;581:143-146.[ISI][Medline] [Order article via Infotrieve]
  18. Heinzel G, Woloszczak R, Thomann P. TopFit version 2.0: pharmacokinetic and pharmacodynamic data analysis for the PC 1993:2.5-2.66 Gusta Fischer New York. .
  19. van den Bongard HJ, Mathot RA, Beijnen JH, Schellens JH. Pharmacokinetically guided administration of chemotherapeutic agents. Clin Pharmacokinet 2000;39:345-367.[ISI][Medline] [Order article via Infotrieve]
  20. Calvert AH. Dose optimisation of carboplatin in adults. Anticancer Res 1994;14:2273-2278.[ISI][Medline] [Order article via Infotrieve]
  21. Ghazal-Aswad S, Tilby MJ, Lind M, Baily N, Sinha DP, Calvert AH, et al. Pharmacokinetically guided dose escalation of carboplatin in epithelial ovarian cancer: effect on drug-plasma AUC and peripheral blood drug-DNA adduct levels. Ann Oncol 1999;10:329-334.[Abstract/Free Full Text]
  22. Slattery JT, Risler LJ. Therapeutic monitoring of busulfan in hematopoietic stem cell transplantation. Ther Drug Monit 1998;20:543-549.[ISI][Medline] [Order article via Infotrieve]
  23. Peris J-E, Latorre J-A, Castel V, Verdeguer A, Esteve S, Torres-Molina F. Determination of busulfan in human plasma using high-performance liquid chromatography with pre-column derivatization and fluorescence detection. J Chromatogr B Biomed Appl 1999;730:33-40.



The following articles in journals at HighWire Press have cited this article:


Home page
Drug Metab. Dispos.Home page
A. J. L. Cooper, I. R. Younis, Z. V. Niatsetskaya, B. F. Krasnikov, J. T. Pinto, W. P. Petros, and P. S. Callery
Metabolism of the Cysteine S-Conjugate of Busulfan Involves a {beta}-Lyase Reaction
Drug Metab. Dispos., August 1, 2008; 36(8): 1546 - 1552.
[Abstract] [Full Text] [PDF]


Home page
JCOHome page
G. Hempel, D. Oechtering, C. Lanvers-Kaminsky, T. Klingebiel, J. Vormoor, B. Gruhn, and J. Boos
Cytotoxicity of Dimethylacetamide and Pharmacokinetics in Children Receiving Intravenous Busulfan
J. Clin. Oncol., May 1, 2007; 25(13): 1772 - 1778.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (10)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mürdter, T. E.
Right arrow Articles by Schwab, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mürdter, T. E.
Right arrow Articles by Schwab, M.
Related Collections
Right arrow Drug Monitoring and Toxicology
Right arrow Automation and Analytical Techniques