| Abstract |
|---|
|
|
|---|
Approach: This is a narrative review based on method evaluations conducted in the laboratories of the authors as well as on relevant publications, especially comparative evaluation studies, from the literature. Publications considered here have been collected by the authors over the past 30 years of involvement as methods for HDL-C made the transition from their early use in lipid research laboratories to clinical laboratories and the recent emergence of homogeneous assays.
Content: The presentation includes descriptions of methodologies, including homogeneous, precipitation, electrophoresis, and ultracentrifugation assays. Reference methods and recommended approaches for assessing accuracy are described. Accuracy and imprecision are summarized in the context of the National Cholesterol Education Program (NCEP) standards for analytical performance. The effects of interfering substances and preanalytical sources of variation are presented.
Summary: Homogeneous assays have been shown to be reasonably well suited for use in routine clinical laboratories, generally meeting the NCEP criteria for precision, accuracy, and total error. However, discrepant results compared with the reference methods have been observed with some of the assays, and the sources of discrepancies are not well characterized. Some homogeneous reagents have not been thoroughly evaluated. At least three of the reagents have experienced successive adjustments in formulation; hence, the reagents may not yet be fully optimized. For these reasons, the homogeneous assays cannot be confidently recommended for use in long-term clinical trials and other research applications without thorough validation.
| Introduction |
|---|
|
|
|---|
This review, drawing on two previous publications including the same authors (3)(4), critically examines the revolutionary new generation of homogeneous assays within the historical context of the evolution of HDL-C methods. Biochemical and physical characteristics of HDL that facilitate separations are discussed. Consensus opinions regarding the clinical utility of HDL-C are presented together with guidelines for use in characterizing and treating dyslipidemias. In addition, the characteristics of the various assay systems are summarized in relation to nationally established analytical performance criteria, and demonstration of traceability is summarized in relation to accepted targets for accuracy.
| HDL |
|---|
|
|
|---|
|
Classically, HDL particles are defined in terms of hydrated density as the fraction with a density between 1.063 and 1.21 kg/L as obtained by preparative ultracentrifugation (10). Because there is little lipid-associated material with a density >1.21 kg/L, HDL is often measured as the cholesterol with a density >1.063 kg/L. Traditionally, the fraction corresponding to HDL as separated by electrophoresis is designated the
-lipoproteins. However, in common practice this and the fraction separated by chemical precipitation or chromatography are also referred to as HDL, although these fractions may not be exactly equivalent. Thus, HDLs are defined in terms of the analytical procedure used to isolate them and include a family of similar particles that vary in size and composition. By contrast, VLDL particles include a substantially greater proportion of lipids, especially TGs, with a density <1.006 kg/L. LDL particles, which carry most of the circulating cholesterol, range in density from 1.006 to 1.063 kg/L. These classes correspond to pre-ß- and ß-lipoproteins, respectively, when separated by electrophoresis.
metabolism
HDLs are secreted from the liver or intestine as disk-shaped nascent particles that consist mainly of phospholipids and apo AI (11). In the circulation, additional phospholipids, cholesterol, and the minor apolipoproteins are transferred from TG-rich particles to these small, dense HDL particles, designated HDL3, converting them to larger and less dense spherical particles. Cholesteryl esters are produced from free cholesterol on the particle surface by the action of lecithin cholesterol acyltransferase in the presence of its cofactor apo AI and move to the core, contributing to enlargement of the HDL particle. These cholesteryl ester-rich particles, referred to as HDL2, contain twice as many cholesterol molecules per unit of apolipoproteins compared with HDL3. In vitro, HDL2 can be converted back to HDL3 in the presence of hepatic lipoprotein lipase (12).
The HDL cholesteryl esters are cleared through the liver by one of three mechanisms: (a) selective removal from HDL by the hepatic HDL receptors; (b) transfer from HDL to apo B-containing lipoproteins via the mediation of cholesterol ester transfer protein, then uptake by the liver through receptors for these TG-rich lipoproteins; or (c) removal with apo E-rich HDL, which is recognized by the hepatic remnant receptors (11). These processes constitute the reverse cholesterol transport mechanism by which cellular and lipoprotein cholesterol is delivered back to the liver for reuse or disposal, a process that is considered to contribute in part to the cardioprotective role of HDL. Another presumed protective role of HDL is in resisting oxidation of LDL, a process that makes lipoproteins more atherogenic (13).
clinical relevance
Many epidemiologic investigations have demonstrated the strong and independent inverse association of HDL, measured in terms of either its cholesterol or apo AI content, to risk of CAD [reviewed in (14)]. The risk of CAD increases 23% for every 10 mg/L decrease in HDL-C (15); thus, higher HDL-C concentrations are considered protective. Many factors, some modifiable, are known to affect HDL-C values (16), including age, gender, genetic factors, cigarette smoking, physical exercise, dietary intake, obesity, and certain drugs (Table 2
).
|
Population distributions for HDL-C have been established in various studies with means of
450 mg/L in men and
550 mg/L in women. The Lipid Research Clinics Program Prevalence Study (17), a major North American study funded by the NIH, used the heparin/MnCl2 precipitation procedure with MnCl2 at 0.046 mol/L in EDTA-plasma specimens. Today, a higher concentration of MnCl2 (0.092 mol/L) would be recommended for EDTA plasma, giving slightly lower HDL-C values. The most recent large population data are from the US National Health and Nutrition Examination Survey III (18), which used a dextran sulfate (50 000 Da) with the MgCl2 method.
| Consensus Guidelines and Medical Decision Points for Diagnosis and Treatment |
|---|
|
|
|---|
600 mg/L considered "protective". Subsequent consensus statements in 1997 from the American Heart Association for primary prevention (21) and from the American Heart Association and American College of Cardiologists for secondary prevention (22) reinforced the enhanced role of HDL-C. In 2001, the NCEP increased the high-risk medical decision point to <400 mg/L (23). The European Atherosclerosis Society and other international consensus panels have also issued guidelines (24)(25). Dissemination of these guidelines has enhanced the use of HDL-C in the diagnosis and management of patients at increased risk for coronary heart disease and led to increasing demand for measurements by clinical laboratories. Appropriate classification of patients based on the cutpoints mandates the use of accurate methods for HDL-C measurement.
Treatment decisions are primarily based on LDL-C because this atherogenic lipoprotein has well-established interventions that have been shown in randomized clinical trials to be efficacious (18)(20). Some patients will have total cholesterol and/or LDL-C values below the high-risk medical decision points but be at increased risk because of low HDL-C (<350 mg/L), in many cases but not always associated with increased TGs. The efficacy of treating low HDL-C values is not as well established but still considered worthwhile. Fortunately, many of the therapies appropriate for high LDL-C and TGs have a beneficial effect on HDL-C, e.g., increased physical exercise, weight loss, and smoking cessation (Table 2
). More recently, some of the cholesterol-lowering drugs, specifically niacin, gemfibrozil, statins, and resins (in decreasing order of effectiveness), have been recognized for increasing HDL-C concentrations and thereby further contributing to overall risk reduction. Thus, measurement of HDL-C not only improves the identification of patients at increased risk for CAD, but also is useful in monitoring efficacy of treatment, especially with therapies likely to affect HDL-C. Some have advocated the use of ratios, either total cholesterol or LDL-C to HDL-C, as a better indicator of overall risk. However, the ratio is considered to have no physiologic significance, and NCEP guidelines focus on the individual values (18).
| Guidelines for Analytical Performance |
|---|
|
|
|---|
|
In developing guidelines for analytical performance, the expert panel considered the requisite analytical performance needed for reliable medical decisions in relation to biological variation and the capabilities of routine laboratories [for a more detailed explanation of the goal setting process, see Refs. (26)(27)]. The current primary goal is that total error (bias + 1.96CV) be within 13% of the true value. The total error term, combining the contributions of imprecision (random error) and inaccuracy or bias (systematic error), represents the maximum tolerable error in measurement of a single specimen to 95% tolerance limits, i.e., in 19 of every 20 measurements. The underlying specifications for imprecision and bias consistent with the total error goal are informative. Imprecision, in proportional units or CV, should be
4%. At low HDL-C values, this proportional target becomes difficult to achieve, e.g., at 250 mg/L the proportional goal would be 10 mg/L; therefore, the precision guideline shifts to an absolute or standard deviation target of 17 mg/L for HDL-C values <420 mg/L. The bias should be <5% from the true value, determined by the accepted RM. Thus, in this example the total error target of 13% is equal to 5% plus 1.96 times 4%.
The imprecision of a particular method can be assessed by replicate analysis. Results are most representative with actual patient specimens or pools prepared from serum freshly collected and frozen. Commercial control materials, because of changes in their analyte and matrix properties during the manufacturing process, may not truly depict performance on patient specimens. For reasons of practicality, such pools are generally lyophilized for long-term stability and often artificially supplemented with concentrated lipoproteins, sometimes of animal source, which can substantially alter their analyte and matrix characteristics (29). Such artificial materials often give not only distorted indications of imprecision, but also unreliable estimations of bias. The term "commutable" refers to a material that is similar to patient specimens in measurement characteristics, often not achieved in even the best commercial control materials. Currently the only approach considered reliable for establishing the accuracy of an HDL-C assay is a direct comparison study on actual patient specimens with a RM. The CDC, in collaboration with the National Heart, Lung and Blood Institute, has established programs (30) for standardizing the lipid and lipoprotein analytes to defined RMs for total cholesterol, LDL- and HDL-C, and TGs. The CDC coordinates a network of reference laboratories, the Cholesterol Reference Method Laboratory Network (CRMLN), which offers direct comparison protocols using fresh patient specimens for reliable transfer of accuracy. The HDL RM and the CRMLN program will be described in detail after a general description of the analytical techniques used.
| Evolution of Methods for Measurement of HDL-C |
|---|
|
|
|---|
| Ultracentrifugation |
|---|
|
|
|---|
Ultracentrifugation as a separation technique is not only tedious and time-consuming, but the labile lipoproteins can be substantially altered by the high salt concentrations and centrifugal forces used. Furthermore, there is a plethora of different types of equipment, making conditions extremely difficult to reproduce from one laboratory to another, and separations are highly dependent on the skills of the technician. Achieving complete and reproducible recovery is difficult, even for experienced technologists, and fractions may be cross-contaminated. In addition, the fractions are heterogeneous and contain other functional lipoproteins. For example, the HDL fraction with a density between 1.063 and 1.210 kg/L may contain considerable amounts of the atherogenic apo B-containing lipoprotein(a). In summary, although ultracentrifugation has been highly useful in research and as a comparison method for validation of other methods, this approach is not considered practical for routine analytical measurements.
| Electrophoresis |
|---|
|
|
|---|
(HDL), pre-ß (VLDL), and ß (LDL). Lipoproteins can be separated using a variety of electrophoretic media, such as paper, agarose gel, cellulose acetate membrane, and polyacrylamide gel, each used with various buffers (31)(32). The common lipophilic stains are Oil Red O, Fat Red 7B, and Sudan Black. Electrophoresis was long considered primarily useful for qualitative analysis because the lipophilic dyes are not specific for a particular class of lipid and because, with the varying composition of the lipoproteins, the staining may not be proportional to cholesterol content. A technique using precipitation of the bands with phosphotungstic acid after agarose gel electrophoresis was reported to allow more reliable quantification of lipoprotein cholesterol but gives discrepant results in samples with atypical lipoproteins (33). Subsequent use of specific enzyme reagents, e.g., cholesterol esterase and oxidase with a peroxidase indicator, to stain electrophoretic plates for cholesterol improved the quantification (34).
Even more promising is a recent modification that uses the enzyme cholesterol dehydrogenase rather than cholesterol oxidase, with the dye nitroblue tetrazolium chloride, which is insoluble and stable after reduction. Electrophoretic systems using this latter (35)(36) or another color system (37) have been reported to give acceptably accurate quantification of HDL-C and the other lipoproteins. A distinct advantage of the electrophoretic methods is the visual presentation that facilitates observation of atypical lipoproteins. Nevertheless, both ultracentrifugation and electrophoresis have disadvantages for use in the clinical laboratory, especially when the workload is high. The more practical chemical precipitation methods that had been used in the research laboratories were generally adopted as the method of choice when measurement of HDL-C moved into routine clinical laboratories.
| Other Methods |
|---|
|
|
|---|
| Chemical Precipitation (First-Generation) Methods for Separation of HDL |
|---|
|
|
|---|
|
A major problem with the precipitation methods has been interference from increased TGs, which make aggregated lipoproteins lighter, preclude the sedimentation of the insoluble aggregates, and yield turbid supernates (48). The resulting cloudy supernatant is contaminated with lipoproteins other than HDL, leading to an overassessment of HDL-C. The extent of TG interference and the cutoff above which results are no longer reliable vary widely depending on the particular precipitation reagent and concentration. Turbid supernates can sometimes be cleared by centrifugation at higher g forces (49) or by filtration (50). Alternatively, specimens can be diluted (40), which decreases the background density of the solution, or the lightest TG-containing particles can be removed by ultracentrifugation before precipitation (51).
Each of the precipitation reagents evolved with modifications reported to change the selectivity or performance. For example, the heparin-MnCl2 method (Table 4
) was first described with 1.0 mol/L MnCl2 (51). Subsequent studies (49) concluded that this concentration was appropriate for serum specimens but that a higher concentration, 2.0 mol/L, improved specificity in EDTA-plasma specimens, presumably to compensate for chelation of some of the divalent cation by EDTA and to avoid incomplete precipitation of apo B-containing lipoproteins. The higher concentration also decreased the extent of interference from high TGs. The original dextran sulfate-MgCl2 precipitation method (Table 4
), with a 500 000-Da material (52), was reported to give low HDL-C values, which was remedied by substituting a dextran sulfate of 50 000 Da (46). The phosphotungstic acid procedures (Table 4
) have been described with and without MgCl2, adjusted to different pH values, and at several concentrations (42)(53). A common version of this method (subsequently designated as PTA) has been reported to give lower values for HDL-C compared with heparin and dextran sulfate methods (54), an important observation in light of published evaluations of the third-generation methods. Similarly, a variety of polyethylene glycol molecular weights, concentrations, and pH conditions have been reported (47)(55). The accepted RM accuracy targets for HDL-C, which will be described in more detail subsequently, use heparin-MnCl2 (30) and dextran sulfate (50 000 Da)-MgCl2 (56) to achieve precipitation.
For more than a decade, quantification of HDL-C in routine laboratories was performed almost exclusively by one of the first-generation precipitation techniques. Although the separation step was performed manually, the required equipment was readily available and cholesterol in the supernate could be easily measured by various automated analyzers. Most laboratories relied on commercial reagents; early methods often used phosphotungstic acid (53), which is inexpensive and readily available. However, phosphotungstic acid reagents were found to be somewhat unstable with extended storage, and separations were reportedly less rugged, being sensitive to separation conditions and technique (54), which led to gradual replacement by the dextran sulfate reagent (Fig. 1
) (46). The precipitation methods were capable of achieving precision consistent with the NCEP targets (Table 3
), as demonstrated by their performance in lipid specialty laboratories (54), but in routine laboratories often did not (57), in part because of the need for manual pipetting and other manipulations.
|
With improvements in automated analyzers, specimen volume requirements progressively decreased and the large volume required for manual pipetting in the precipitation step was increasingly excessive in relation to blood volumes collected for other laboratory tests. The additional manipulations required to clear specimens with high TGs interfered with laboratory workflow. As the HDL-C measurement became increasingly common in the 1990s, pressures to reduce operating costs in the laboratory gradually moved other methods toward full automation, and the labor-intensive manual steps became increasingly untenable. Some higher-volume laboratories attempted to streamline the precipitation methods with automated pipetting stations, but in most cases the pretreatment step continued to be performed manually, making HDL-C one of the more tedious common procedures in the clinical laboratory. A result was a second generation of more streamlined methods.
| Second-Generation Methods |
|---|
|
|
|---|
A method developed for small and low-volume laboratories streamlined the separation with a plastic device (Spin-Pro) containing premeasured precipitation reagent (60). The user simply poured the specimen in the top of the plastic tube and placed the tube in the centrifuge. During centrifugation, the device accurately measured the sample, which was mixed with the precipitation reagent, and the resulting precipitate sedimented. After centrifugation, the supernate rose into the central well, and the device could be placed directly in the sample tray of a chemistry analyzer. Highly sophisticated compact analyzers, developed for physicians offices and other point-of-care sites, integrated HDL separation and quantification from plasma or whole blood applied to cassettes or reagent-impregnated strips, eliminating the need for separate pretreatment steps (61)(62).
| Accuracy Targets |
|---|
|
|
|---|
There is no definitive method for HDL-C; indeed, considering the complexity of the lipoprotein particles, it may be impossible to unambiguously define the fraction of interest and develop such a method. Cholesterol is a readily defined molecule, providing a basis for a definitive method. By contrast, lipoproteins are complex, heterogeneous, and polydisperse particles with overlapping properties that have not been fully characterized, and the various separation methods can obtain different populations of particles with varying associations with CAD risk. There is no compelling evidence that the particular fraction obtained by the CDC RM is a better indicator of CAD risk than the fraction obtained by any other method. Thus, the CDC method is accepted as the RM based on consensus in consideration of its use as the accuracy target for population studies from which decision points were taken and not because the method is based on an unambiguous definition and thoroughly validated and specific quantification of the most protective fraction of lipoprotein particles.
In fact, the current emphasis on the RM as the accuracy target, although necessary to achieve standardization of results, might actually lead to rejection of even more specific methods, which might actually better separate a risk-negative population of lipoprotein particles. Developers of new methods should be aware of this potential and consider that a new method, perhaps not in agreement with the RM because it fractionates a different population of lipoprotein particles, might need to be validated by demonstrating better prediction of risk in population studies, e.g., in cross-sectional retrospective studies. Accumulation of compelling evidence supporting better prediction by a different fractionation might necessitate redefining the RM.
designated comparison method
Few laboratories have an ultracentrifuge and the skills required to perform the CDC RM for HDL-C. In addition, ultracentrifugation is expensive and requires a large sample volume, typically 5.0 mL, which complicates the process of obtaining specimens for comparison. Therefore, the CRMLN laboratories developed and validated a modified dextran sulfate (50 000 Da) procedure as the Designated Comparison Method (DCM) to provide results approximately equivalent to those of the CDC RM but without the need for ultracentrifugation (56). The MgCl2 precipitation reagent concentration was decreased slightly from that used in the AACC Selected Method to make the HDL-C values slightly higher and more consistent with those of the CDC RM. Comparison studies with the less costly DCM, closely standardized to the CDC RM, are available through the CRMLN laboratories (30).
crmln
Because of matrix and analyte changes in processed control materials, judgments of accuracy based on such materials may be compromised. The approach considered most reliable for assuring accuracy is a direct comparison using actual representative patient specimens with a method of established accuracy (65). The CRMLN, coordinated by the CDC and consisting now of five experienced US laboratories and several international partner laboratories in Canada, Europe, Japan, and South America, all closely standardized to the CDC RM, offers a protocol, based on NCCLS guidelines (66), whereby diagnostic manufacturers can assure accuracy by completing a comparison study with the DCM (67). Agreement within specified limits (5% for HDL-C), as well as within other limits for total and LDL-cholesterol and expected in the near future for TG, qualifies the method for certification. On the basis of comparison results, appropriate adjustment of calibrator setpoints can be made to bring performance into agreement with the accuracy target. Diagnostics manufacturers as well as distributors and instrument partners are encouraged to certify at least every 2 years and to ensure that every production lot is calibrated to maintain accuracy consistent with the accuracy targets, which can be accomplished by ongoing participation in the CDC/CRMLN program. The website of the AACC (www.aacc.org/standards/cdc/cholesterolinfo.stm) provides details of the comparison studies, contact information for CDC/CRMLN, and a listing of the commercial methods that have qualified for certification. Clinical laboratories can promote standardization by encouraging suppliers of their reagent sets to participate in the CRMLN comparison process. In addition, clinical laboratories can also make arrangements with a CRMLN laboratory to perform their own direct comparisons, which are especially useful in the case of new or modified methods. Modest user fees support the CRMLN program.
Third-Generation Homogeneous Assays
The provision of a reliable standardization program was a factor in improving accuracy in HDL-C measurements, but imprecision remained a problem with the conventional precipitation methods. A major breakthrough was reported in 1994 (68), with the first of a series of so-called "homogeneous" methods capable of full automation and considered third generation (Table 4
). The term homogeneous is based on a previous convention in clinical chemistry used to describe an immunoassay that does not require separation of free from bound label, i.e., an assay that can be fully automated in a single reaction vessel. Some have described the new assays with the term "direct", which implies that the analyte is measured directly. Because all existing methods, including those requiring pretreatment, measure HDL-C directly, the term homogeneous is considered more descriptive.
The new-generation homogeneous assays do not require off-line pretreatment and separation, eliminating the manual pipetting, mixing, and centrifugation steps. This first homogeneous assay for HDL-C (Fig. 2A
), described by International Reagents Corporation (Kobe, Japan), was highly innovative but limited in application with four successive reagent additions. The first reagent contained polyethylene glycol, which caused aggregation of the apo B-containing chylomicrons, VLDL, and LDL. The second reagent protected or blocked the aggregated lipoproteins with antibodies to apo B and apo C. The third reagent included the cholesterol reaction enzymes (cholesterol esterase, cholesterol oxidase, and peroxidase), which reacted only with the unprotected HDL-C. A fourth and final reagent stopped the enzymatic color reaction and cleared the reaction mixture with guanidine salts, which solubilized the aggregates. HDL-C was quantified based on the final reaction absorbance, monitored at 600 and 700 nm. This breakthrough method, designated subsequently as IRC, introduced the possibility of full automation and set the stage for subsequent, two-reagent homogeneous methods. Although not suitable for many analyzers because of the four reagent additions, the IRC method is still distributed in Japan.
|
In 1995, Kyowa Medex Co. (Tokyo, Japan) (69) introduced two clever innovations (Fig. 2B
): (a) sulfated
-cyclodextrins together with Mg2+ were found to selectively block but not precipitate chylomicrons and VLDL, providing selectivity without the need for a clearing reagent; and (b) the specificities of the enzymes cholesterol esterase and cholesterol oxidase toward HDL-C were enhanced by covalently linking polyethylene glycol molecules to the enzymes, excluding cholesterol in the larger LDL particles. Polyethylene glycol 6000 Da in size was found to optimize the specificity at concentrations much lower than those used previously to precipitate lipoproteins, leading the developers to speculate that the modified enzymes were able to distinguish the lipoprotein classes based on their size and/or charge. These innovations facilitated a fully automated homogeneous assay with only two successive reagent additions. The commercial version of this reagent (designated subsequently as PEG) included in the first reagent
-cyclodextrin sulfate together with dextran sulfate, Mg2+, and detergents with buffering to pH 7. The second reagent contained the modified enzymes and substrates. This two-reagent assay, now also distributed through Roche Diagnostics (Indianapolis IN; formerly Boehringer Mannheim Corporation), offered a practical homogeneous assay suitable for many of the common chemistry analyzers. The original commercial reagent sets included the second enzyme-containing reagent in lyophilized form, necessitating reconstitution. A second version (70) introduced in mid-1998 provides both reagents in liquid form. A third version with decreased Mg2+ concentration, apparently to reduce carryover effects, is currently in evaluation by one of the authors (M.A.N.). (Users should recognize that this and other homogeneous reagents have been successively modified; different versions with various measurement properties may be supplied by various distributors and instrument partners.)
Subsequently, other innovative homogeneous reagents were developed, also in Japan. Daiichi Pure Chemicals Company (Tokyo, Japan) offered a homogeneous assay (Fig. 2C
) using a synthetic polymer together with a polyanion to block the non-HDL lipoproteins (59)(60)(71). A detergent then exposed only cholesterol in HDL to the enzymes, giving specificity for HDL-C. Commercial reagent sets, also distributed by Genzyme Corporation (Cambridge, MA) and other distributors and instrument partners, include two reagent additions, the first with the polyanion and polymer blocking agents and the second with detergent, enzymes, and substrates. The first commercial version required reconstitution of the lyophilized enzyme reagent. In mid-1998, a second version was introduced with both reagents as liquids (72)(73). A subsequent evaluation (74) indicated that the fully liquid formulation was substantially modified (Fig. 2
) to improve specificity and decrease potential interference. A third version of this reagent without Mg2+ was described in poster presentations from Daiichi Pure Chemicals (Tokyo, Japan) at the 2000 AACC Annual Meeting (75)(76).
A fourth homogeneous method (59)(77)(78), involving immunoinhibition (Fig. 2D
), was introduced by Wako Pure Chemicals Industry (Osaka, Japan). Antibody to human apo B in the first reagent reacts with the apo B-containing lipoproteins, chylomicrons, VLDL, and LDL, blocking their reaction with the enzymes added in a second reagent and giving specificity for HDL-C. The original reagent set required reconstitution of the enzyme reagent; however, both reagents in the second version introduced in mid-1998 are liquids.
A fifth homogeneous method was introduced by Denka Seiken Co. (Niigata, Japan) (78)(79) and was also distributed through Polymedco Inc. (Cortlandt Manor, NY) and by Randox Laboratories Limited (Crumlin, UK). This method (Fig. 2E
) reportedly uses a selective reagent with cholesterol esterase and oxidase to generate peroxidase from the cholesterol in lipoproteins other than HDL. The initial non-HDL-derived peroxidase is scavenged by the enzyme catalase. A second reagent includes an inhibitor of catalase and a surfactant to release specifically HDL-C, which develops color through the usual peroxidase sequence.
These commercial homogeneous methods better conform to the workflow patterns in the modern clinical laboratory; sample handling can be as simple as placing a bar-coded specimen tube on the automated analyzer. Elimination of manual steps can provide cost savings; higher reagent costs are generally offset by decreased labor costs. Additionally, these newer methods improve precision through more consistent pipetting of smaller specimen volumes as well as precise temperature control and reaction timing, which facilitate achieving the NCEP analytical performance goals. Clinical laboratories are rapidly adopting the homogeneous methods (Fig. 1
), although questions regarding specificity, especially on specimens with unusual lipoprotein composition, have been raised. Performance characteristics of each of the common commercially available homogeneous assays as well as some of the potential problems are described in detail subsequently.
polyethylene glycol-modified enzyme method (peg)
The PEG assay has been extensively evaluated in a variety of studies, including CRMLN certification (www.aacc.org/standards/cdc/cholesterolinfo.stm), comparisons with the RM and DCM, and multicenter studies (Table 5
).
|
Imprecision.
Most of the published evaluation studies report total CVs <3.1% (Table 6
).
|
Dynamic range.
The dynamics range spans from
30 mg/L up to at least 1500 mg/L.
Specificity.
Okamoto et al. (80), using gel chromatography to separate lipoproteins by size, demonstrated that the PEG assay includes the apo E-containing HDL particles with the rest of HDL-C, in contrast to a PTA method that precipitated this population, consequently yielding HDL-C results
10% lower. Because the apo E-containing HDL particles are also included by the CDC RM, the homogeneous assay was considered more accurate. Okazaki et al. (81) supported these findings by demonstrating close agreement of the PEG method with a HPLC procedure: PEG = 1.026HPLC - 7.5 mg/L (r = 0.998). Another study similarly observed correlation with a conventional PTA precipitation procedure in normo- (r = 0.987) and hypertriglyceridemic (r = 0.953) samples; but supporting the previous studies, that y-intercept (
570 mg/L) was consistent with a negative bias by PTA (82). apo E-containing particles are known to account for
10% of the total HDL-C, consistent with the bias observed in these studies and also consistent with a previously observed negative bias of the PTA method in comparison with dextran sulfate and heparin precipitation methods (9)(83).
Accuracy.
The homogeneous PEG HDL-C assay was compared with the CDC RM (PEG = 1.068RM - 17 mg/L; r = 0.993; mean bias, 2.2%) and the DCM (PEG = 1.037DCM + 4 mg/L; r = 0.996; mean bias, 4.5%) in specimens with TGs <2000 mg/L (84). In a multicenter study performed before the commercial release of the reagent by six laboratories in Europe, comparisons with a PTA precipitation method gave results with correlation coefficients of 0.960.99, but slopes that ranged from 0.95 to 1.08. In this study, which used a calibrator with the distributors preliminary HDL-C setpoint, the PEG assay was negatively biased on average by 3% compared with the CDC RM. The distributor adjusted the calibrator by the suggested 3% to make results consistent with the RM (84)(85), an adjustment expected to bring total error to
7.5%, within the 13% NCEP target. An earlier evaluation of this assay using a calibrator from the manufacturer (Kyowa Medex) gave a slope of 0.87 with a negligible y-intercept that produced negative total errors of 1521% (58) compared with another CDC standardized method.
The first commercial reagents introduced to the market did not reproduce the encouraging results of the multicenter study with the subsequent calibrator adjustment. Correlation coefficients were acceptable (r = 0.987), but the slope was 12% high with a y-intercept of only 10 mg/L (59). At the clinical decision cutpoints of 350 and 600 mg/L, total errors of 16.2% and 14.5%, respectively, were observed. However, subsequent calibrator lots yielded results as expected from the reference standardization (PEG = 0.994RM + 21 mg/L; r = 0.997) (72)(86). Other investigators observed similar agreement (77). These studies illustrate the challenges in appropriately assigning target values to calibrators. The concentration of the calibrator for the PEG assay now appears to be appropriately assigned, which is supported by a US proficiency survey (87). Roche, a distributor of the PEG method, now has each new reagent lot routinely validated for accuracy by the CRMLN laboratory in Rotterdam. Users of all HDL methods should be aware of the lot-to-lot calibration issue and are advised to confirm the accuracy of each new lot by overlap studies and/or by direct comparison studies with one of the CRMLN laboratories.
The second, fully liquid PEG reagent version, introduced in 1998, was evaluated in another multicenter study and gave similar performance characteristics (88). The two generations were correlated in seven different laboratories (r = 0.989) with a mean deviation of only 2 mg/L. A comparison with the DCM (PEG = 0.968DCM + 24.9 mg/L; r = 0.996) produced total errors at the decision points (350 and 600 mg/L) of 7.3% and 3.4%, respectively (84).
Interferences.
The PEG assay is free of TG interference up to
10 000 mg/L regardless of whether the TGs are in chylomicrons or VLDL. With the fully liquid reagent, hypertriglyceridemic samples should be diluted with saline (9 g/L NaCl) to bring TG concentrations below 18 000 mg/L (88). LDL-C concentrations of 3000000 mg/L did not cause a significant bias, whereas LDL-C concentrations >6000 mg/L produced falsely increased HDL-C values (88). Hemoglobin up to 10 g/L did not interfere, but at higher concentrations a slight negative bias was observed. Bilirubin concentrations >100 mg/L yielded discrepant results when the homogeneous PEG assay was compared with precipitation procedures, but the studies did not make clear which assay was most affected. However, interferences from hemoglobin and bilirubin seem much less pronounced in this homogeneous assay compared with conventional precipitation procedures. Increased free fatty acids did not interfere, whereas paraproteinemias gave discrepant results in some samples compared with the PTA procedure. Heparin as anticoagulant does not interfere, but results with EDTA plasma should be multiplied by 1.06 to compensate for dilution caused by osmotic shifting of water from the red cells (26).
Patients with type III hyperlipoproteinemia, characterized by cholesterol enrichment of the VLDL because of impaired clearance of remnant lipoproteins, were observed to give erroneous values with the PEG assay, one falsely high and one low (89)(90), apparently because of remnant particles reacting atypically with the PEG assay. In contrast, the PEG assay overestimated HDL-C by 72% in hyperlipidemic (C57BL/6) mice with accumulation of cholesterol-rich remnant particles (91), made so by a high-fat, high-cholesterol diet. In apo E-deficient mice, which also accumulate remnant lipoproteins, the overestimation was 228%. The discordance seemed to be related to the degree of hyperlipidemia because the apo E-deficient mice had extremely high VLDL-cholesterol (VLDL-C) and LDL-C concentrations. By contrast, mice with normal cholesterol and TG concentrations demonstrated good agreement between the PEG assay and PTA precipitation. On the basis of these studies, users would be prudent to validate the PEG and other homogeneous assays before use in studies of animals and humans expected to have unusual lipoprotein characteristics such as increased remnants.
A recent evaluation of this method in patients with diabetes inferred but did not explicitly state minimal interference (92). Whether liver disease interferes with the PEG assay, as has been observed with other homogeneous assays, is not currently known.
synthetic polymer/detergent (spd) method
The earliest version of this reagent, which included dextran sul