Clinical Chemistry 48: 2236-2241, 2002;
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (11)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Penders, J.
Right arrow Articles by Delanghe, J. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Penders, J.
Right arrow Articles by Delanghe, J. R.
Related Collections
Right arrow General Clinical Chemistry
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2002;48:2236-2241.)
© 2002 American Association for Clinical Chemistry, Inc.

Quantitative Evaluation of Urinalysis Test Strips

Joris Penders1, Tom Fiers1 and Joris R. Delanghea1

1 Department of Clinical Chemistry University Hospital Ghent, De Pintelaan 185, B-9000 Ghent, Belgium.

aAuthor for correspondence. Fax 32-9-240-4985; e-mail joris.delanghe{at}rug.ac.be.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Background: Urine test strip results are generally reported in categories (i.e., ordinal scaled), but automated strip readers are now available that can report quantitative data. We investigated the possible use of these meters to complement flow cytometry of urine and compared reflectance readings with quantitative determinations of urinary glucose and microalbumin.

Methods: We compared URISYS 2400 (Roche) quantitative reflectance data with data from the UF-100 (Sysmex) and biochemical data for 436 nonpathologic and pathologic urine samples.

Results: Reproducibility of the reflectance signal was good for high- and low-concentration urine pools for protein (0.8% and 0.9% and 1.5% and 2.2% within and between runs, respectively), leukocyte esterase (1.1% and 1.0%; 5.1% and 1.2%), hemoglobin (1.7% and 1.1%; 8.9% and 1.1%) and glucose (2.1% and 0.5%; 6.5% and 2.3%). Fair agreement was obtained between UF-100 and test strip reflectance data for erythrocytes and hemoglobin (r = -0.680) and leukocytes and leukocyte esterase (r = -0.688). Higher correlations were observed for biochemical and test strip data comparing protein and albumin (r = -0.825) and glucose data (r = -0.851). The lower limits of detection for erythrocytes and leukocytes were 8 x 106/L and 19 x 106/L, respectively. The protein test (n = 220) detected 86% (95% confidence interval, 78–92%) of samples with <30 mg/L albumin with a specificity of 84% (95% confidence interval, 76–91%).

Conclusions: In urine test strip analysis, quantitative hemoglobin and leukocyte esterase reflectance data are complementary with flow cytometric results and glucose and albumin results.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Test strip analysis plays an important role in urinalysis as such, and the value of test strip urinalysis as a screening method has been thoroughly demonstrated (1). The reproducibility of (semi)automated readings is at least as good as visual readings (2), but most authors find the analytical, clinical, and labor cost-saving advantages of (semi)automated vs visual reading to be obvious (3).

Recently, the URISYS 2400 automated urine test strip analyzer (Roche Diagnostics) was introduced. This instrument offers the possibility to obtain reflectance readings. Test results, therefore, no longer need to be expressed in an ordinal scale. Access to the instrument’s raw data theoretically allows a higher analytical sensitivity for several analytes. Because microalbuminuria is generally regarded as an excellent marker for assessing early renal damage in common conditions such as diabetes and hypertension (4)(5)(6)(7) and as an early predictor of preeclampsia during pregnancy (8), the availability of highly sensitive test strip readers opens interesting perspectives for assessing this phenomenon.

Flow cytometry has been introduced for urinalysis (9)(10) to obtain quantitative data on urinary particles. The imprecision of urinary flow cytometry is far less than that of conventional urinary microscopy, but in some cases urinary flow cytometry reports erroneous results because of analytical interferences (e.g., calcium carbonate crystals falsely increase erythrocyte counts; confusion between yeast cells and erythrocytes). Langlois et al. (11) reported disagreement in erythrocyte counts between the UF-100 and the hemoglobin test strip reaction in 6.5% of cases. Of course, the frequency of erroneous results depends on the proportion of pathologic samples and on the preanalytical handling of samples. Combining diagnostic information provided by urinary flow cytometry and more quantitative test strip analysis therefore offers a theoretical basis for the development of diagnostic expert systems (11)(12).

In this study, we wanted to investigate the performance of quantitative urinary test strip analysis. In particular, we wanted to compare the reflectance readings of the protein test field with immunochemical microalbumin determinations. We also wanted to explore the possibilities of combining the two novel methods, particularly in the analysis of erythrocytes, leukocytes, and glucose, which are of major clinical importance.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
patients and samples
We studied 436 freshly collected urine samples submitted to our routine laboratory for diagnostic urinalysis. All samples were completely processed within 2–4 h after arrival. Test strip urinalysis was carried out before flow cytometry analysis (Sysmex UF-100; TOA Medical Electronics), using URISYS strips on a URISYS 2400 analyzer (Roche Diagnostics). Combur 10-Test M strips on a Miditron automated reflectance photometer (Roche) (13)(14), used in our routine laboratory, were used in parallel as a control. The strips include reagent pads for ordinal scale reporting of relative density, pH, leukocyte esterase, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and hemoglobin/myoglobin.

Day-to-day imprecision was assessed with control material: Liquichek Urinalysis Control Levels 1 and 2 (Bio-Rad). This is stable for 30 days when stored tightly capped at 2–8 °C.

urinary flow cytometry
The Sysmex UF-100 is a urinary flow cytometer-based walkaway instrument that performs automated microscopic analysis. It has been extensively evaluated for urinalysis (9)(10)(11)(15) as well as for cerebrospinal fluid analysis (16). The principle is based on argon laser flow cytometry. The UF-100 measures the conductivity and categorizes the particles on the basis of their shape, size, volume, and staining characteristics. The results are displayed in scattergrams, histograms, and as counts per microliter as well as counts per high-power field. The UF-100 automatically detects and counts red blood cells (RBCs), white blood cells (WBCs), bacteria, yeast cells, crystals, epithelial cells, small round cells, sperm cells, and casts. Particles that cannot be classified in one of the former categories are counted as "other cells".

urisys 2400
Urine test strip analysis was performed with the automated URISYS 2400. These test strips are the same as used in the Miditron but are supplied in a cassette holding 400 test strips for leukocyte esterase, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, hemoglobin/myoglobin, and pH.

The intensity of the reaction color of the test pad is detected by measuring the percentage of light reflected from the surface of the test pad. The higher the analyte, the higher the color intensity and, thus, the lower the reflectance. The reflectance value, expressed as a percentage within a range from 100% (white) to 0% (black), is therefore inversely related to the concentration of the analyte in the sample. Specific gravity (refractometry based) and clarity are measured in a flow cell, and color is rated with a specific algorithm against the blank pad on the test strip. Data are expressed in an ordinal scale (as "normal", "negative", "positive", or as nominal concentrations) on the reports, but (quantitative) reflectance data can be downloaded to floppy disks.

biochemical investigations
Albumin in urine was measured immunonephelometrically on 220 randomly selected samples with use of commercially available Behring antibodies on a Behring Nephelometer II analyzer (Dade Behring) standardized against the widely accepted WHO/College of American Pathologists Certified Reference Material 470. Glucose was measured by a hexokinase method standardized against Standard Reference Material 917a and Standard Reference Material 965 (n = 85), and total protein by a pyrogallol red method (17) (n = 129) with Standard Reference Material 917a as a standard and commercially available reagents for both measurement procedures (Roche) on a Modular P system (Roche).

statistics
P values <0.05 were considered significant. Agreement between automated flow cytometry and test strip data was evaluated by Spearman rank analysis. Multiple regression analysis was used to investigate a model relating leukocyte esterase and hemoglobin field reflectance. The lower limit of detection (18) was calculated as the mean value - 3 SD for a blank sample. Diagnostic accuracy was assessed by ROC analysis using commercially available statistical software (MedCalc®).


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
dilution and pH
Alkalinization of urine gives rise to erroneous results in particle counting. When monitoring our samples, we found a median pH of 6.5 (95% interval, 5–8). Hence, extremely alkaline urine samples were not present in the samples we investigated. Dilution was also monitored with a range of specific gravity of 1.005–1.033 and a median result of 1.017.

reproducibility
The within- and between-run CVs for protein, leukocyte esterase, hemoglobin, and glucose are summarized in Table 1 .


View this table:
[in this window]
[in a new window]
 
Table 1. Reproducibility of protein, leukocyte esterase, hemoglobin, and glucose on the URISYS 2400 analyzer.

comparison of protein reflectance results, albumin, and proteinuria
We found a close correlation between the log-transformed albumin results (x) and the test strip reflectance results (y): y (%) = 68.2 - 9.26[logx (mg/L)]; Spearman r = -0.825; P <0.001; Fig. 1 . Two discrepancies (0.9%) were found in which the test strips on both automated strip readers overestimated the urinary albumin concentration by >10-fold.



View larger version (16K):
[in this window]
[in a new window]
 
Figure 1. Correlation between protein (albumin) results obtained by quantitative test strip analysis (y) and albumin (x; Behring BN II Nephelometer; n = 220).

Protein field reflectance (y; %) = 68.2 - 9.26 log(albumin; mg/L); Spearman r = -0.825; P <0.001). The two outer solid lines represent the 95% prediction interval around the regression line. The dashed lines represent the lower limit of detection (- - -) with the 2 SD limits (- - - -).

Similarly, protein reflectance data and total protein measurements correlated well (Spearman r = -0.921; P <0.001), which is expected as long as the major protein is albumin. The lower limit of detection was 25 mg/L (CV = 2.2%) (18). On the basis of ROC analysis, diagnostic sensitivity was 86% (95% confidence interval, 78–92%) and specificity was 84% (95% confidence interval, 76–91%) for a reflectance value of 55.6% when the nephelometric assay was used as the comparison method at a cutoff of 30 mg/L (n = 220; 113 positive and 107 negative cases, respectively).

comparison of flow cytometric rbc and hemoglobin reflectance results
Agreement was fair between the flow cytometric RBC data (x) and the URISYS 2400 hemoglobin reflectance measurements (y) for counts above the upper reference limit (25 x 106 RBC/L on UF-100). The following regression equation was obtained: y (%) = 78.2 -19.4 logx (106 cells/L); Spearman r = -0.680; P <0.001; Fig. 2 . A lower limit of detection of 8 x 106 cells/L (CV = 1.2%) was calculated.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 2. Correlation between RBC counts (x; flow cytometry) and test strip hemoglobin concentration (y) for RBC counts >25 x 106 cells/L (n = 96).

Hemoglobin reflectance (y; %) = 78.2 - 19.4 log(RBC count; 106 cells/L); (Spearman r = -0.680; P <0.001. The two outer solid lines represent the 95% prediction interval around the regression line. The dashed lines represent the lower limit of detection (- - -) with the 2 SD limits (- - - -).

Because RBCs tend to lyse in urine, we investigated the linearity of the hemoglobin test strip pad. The UF-100 flow cytometer automatically determines the conductivity, so we performed multiple regression analysis on the study population, calculated in a regression model the hemoglobin reflectance vs RBCs and conductivity, and found that the t-value (as expected) for RBCs was -8.442 (P <0.0001). In contrast, conductivity showed a t-value of only 1.477 (P = 0.141). Additional dilution experiments were carried out. The ratio of hemoglobin reflectance to RBC count was constant when osmolality was >190 mosmol/L.

comparison of flow cytometric wbc and leukocyte esterase reflectance results
The correlation between WBC data (x) and leukocyte esterase measurements (y) is shown in Fig. 3 . When urinary WBC counts were higher than the upper reference limit (25 x 106 cells/L), the following regression equation was obtained: y (%) = 83.7 -15.4 logx (106 cells/L); Spearman r = -0.688; P <0.001. For WBCs, we found a lower limit of detection of 19 x 106 cells/L (CV = 1.2%).



View larger version (17K):
[in this window]
[in a new window]
 
Figure 3. Regression equation for urinary WBC counts above the upper reference limit (25 x 106 cells/L; n = 132).

Leukocyte esterase reflectance (y; %) = 83.7 - 15.4 log(WBC count; 106 cells/L); Spearman r = -0.688; P <0.001. The two outer solid lines represent the 95% prediction interval around the regression line. The dashed lines represent the lower limit of detection (- - -) with the 2 SD limits (- - - -).

Multiple regression analysis on the study population showed no effect of conductivity on test strip leukocyte esterase.

comparison of the glucose measurements
We also found good agreement between the hexokinase-based glucose method (x) and glucose test strip reflectance reading (y): y (%) = 53.9 - 19.3 logx (mmol/L); Spearman r = -0.851; P <0.001. Fig. 4 depicts the correlation between both measurement methods and the lower limit of detection of 0.17 mmol/L (CV = 1.3%) with the 2 SD limits.



View larger version (15K):
[in this window]
[in a new window]
 
Figure 4. Correlation between glucose measurements (glucose concentration different from 0) by the hexokinase-based method and glucose test strip reading (n = 85).

Glucose reflectance (y; %) = 53.9 - 19.3 log(glucose; mmol/L); Spearman r = -0.851; P <0.001). The two outer solid lines represent the 95% prediction interval around the regression line. The dashed lines represent the lower limit of detection (- - -) with the 2 SD limits (- - - -).


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
The change from urine microscopy to urinary flow cytometry has been accompanied by a significant decrease in imprecision (15). Because urinalysis test strips are often used for checking urinary flow cytometry data (11), there is a need for a more quantitative evaluation of urinalysis test strips. In this study, we compared the URISYS 2400 automated strip reader with the Sysmex UF-100 flow cytometer to evaluate the possible value of quantitative test strip data in a urinary expert system.

The turnaround time was short enough (2–4 h) not to affect the readings: time intervals can influence results of automated test strip analysis, especially leukocyte and erythrocyte ratings (19).

The detection limit of the URISYS 2400 protein assay was 25 mg/L if restricted to albumin alone and total protein was not considered. Because microalbuminuria is defined as excretion of 30–300 mg of albumin/24 h (20–200 µg/min, or 30–300 µg/mg of creatinine) (20), the protein field result allows identification of microalbuminuria cases in contrast to the classic reading and reporting of the strips, which can detect only albumin concentrations 150–200 mg/L or higher (21)(22). This implies that the test has the potential to offer a screening for microalbuminuria without increased cost, hence coming closer to the "urine test strip of the future" (23). This could improve level 1 urinary screening (12): it offers an improved, fast, reliable method that is easy to handle and usable in primary-care laboratories. It could be of special interest in patients with undiagnosed diabetes or hypertension in whom microalbuminuria is regarded as an excellent marker for assessing early renal damage (4)(5)(6)(7), particularly because many patients with non-insulin-dependent diabetes mellitus are asymptomatic and their diabetic state remains undiagnosed for years (24). It must be noted, however, that the sensitivity (86%) may be inadequate for patients with known diabetes in whom physicians do not want to miss microalbuminuria in their annual testing. Moreover, a 16% false-positive rate may be unacceptable for screening.

Because microalbuminuria is characterized by large intraindividual variability, an ordinal-scale answer might be misleading. It therefore is advantageous to have a better estimation of the true concentration.

As is the case for test strips in general, the method is based on the so-called "indicator error" principle, in which proton exchange between the indicator on the strip and the proteins in the solution produces a color change from yellow to green-blue. Of all the diagnostically relevant urinary markers, only albumin and transferrin accept protons well, so the potential error of missing Bence Jones proteinuria remains when test strip screening strategies are used in proteinuria (21), although many cases presenting with Bence Jones or tubular proteinuria show microalbuminuria (23).

In recent studies, the correlation (r) between UFC WBCs and WBCs by counting chamber was 0.93–0.98, and that of UFC RBCs and RBCs by counting chamber was 0.83–0.89 (10)(25)(26). In our study correlating test strip data with urinary flow results, r was -0.69 between the flow cytometric WBC count and the leukocyte esterase reaction, although the presence of esterase inhibitors in urine and severe proteinuria might negatively affect test results for leukocyte esterase (1)(27). No effect of conductivity on the leukocyte esterase field was noted. Test strip measurements had reasonable lower limits of detection for WBC (19 x 106/L).

In this study, the urinary hemoglobin concentration did not agree well with the RBC count obtained by flow cytometry. The hemoglobin measurement is based on the peroxidase principle. It is known that reducing substances (e.g., ascorbic acid) may lower the signal, whereas oxidizing substances may have a positive effect on measured hemoglobin concentration. Various low- and high-molecular mass inhibitors have been found in urine (28)(29). The presence of haptoglobin in urine enhances the peroxidase activity of hemoglobin (29)(30). Bacterial peroxidases can also contribute to total peroxidase activity in urine (23). On the other hand, the quantitative evaluation of test strips may help to eliminate analytical errors in RBC counting attributable to the presence of yeast cells or large amounts of calcium carbonate crystals (11)(15). As shown by our multiple regression model and dilution experiments, the hemoglobin field reflectance test is not influenced by dilution effects. As is the case for WBCs, the lower limit of detection for RBCs is acceptable (8 x 106/L).

The correlation coefficient for the glucose signal with the routine hexokinase-based method was -0.851. The presence of ascorbate oxidase on the glucose test field prevents interference by ascorbic acid.

Because of the procedure for applying the urine to the test strips on the URISYS 2400, improper dipping is no longer a problem, nor is confusion about sample identification or urine contamination caused by dipping the strip in the tube, which potentially leads to interferences with chromatographic methods (31).

In conclusion, quantitative urine test strip analysis provides reliable data on WBCs, RBCs, glucose, and albumin. This offers several possibilities: (a) The sensitivity for albumin may allow affordable screening for microalbuminuria, particularly in patients with undiagnosed renal damage. To fully explore the possibilities of the albumin test pad in first-line diagnosis, a formal study should be performed. (b) In addition, hemoglobin and leukocyte esterase reflectance data are useful for verifying flow cytometric data on RBCs and WBCs. This leads to improved elimination of occasional errors in the WBC and RBC counting channels of the flow cytometer (11).


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Bonnardeaux A, Somerville P, Kaye M. A study on the reliability of dipstick urinalysis. Clin Nephrol 1994;41:167-172.[Web of Science][Medline] [Order article via Infotrieve]
  2. Winkens RAG, Leffers P, Degenaar CP, Houben AW. The reproducibility of urinalysis using multiple reagent test strips. Eur J Clin Chem Clin Biochem 1991;29:813-818.[Web of Science][Medline] [Order article via Infotrieve]
  3. Peele J, Gadsden R, Crews R. Semi-automated vs. visual reading of urinalysis dipsticks. Clin Chem 1977;23:2242-2246.[Abstract/Free Full Text]
  4. Deferrari G, Repetto M, Calvi C, Ciabattoni M, Rossi C, Robaudo C. Diabetic nephropathy: from micro- to macroalbuminuria. Nephrol Dial Transplant 1998;13(Suppl 8):11-15.[Abstract/Free Full Text]
  5. Rodico J. Does antihypertensive therapy protect the kidney in essential hypertension?. J Hypertension Suppl 1996;14:S69-S75.[Medline] [Order article via Infotrieve]
  6. Viberti GC, Jarrett RJ, Wiseman MJ. Predicting diabetic nephropathy [Letter]. N Engl J Med 1984;311:1256-1257.[Web of Science][Medline] [Order article via Infotrieve]
  7. Mogensen C. Microalbuminuria predicts clinical proteinuria and early mortality in maturity-onset diabetes. N Engl J Med 1984;310:356-360.[Abstract]
  8. Nissell H, Kublickas M, Lunnell NO, Petterson E. Renal function in gravidas with chronic hypertension with and without superimposed preeclampsia. Hypertens Pregnancy 1996;15:127-134.
  9. Ben-Ezra J, Bork L, McPherson RA. Evaluation of the Sysmex UF-100 automated urinalysis analyzer. Clin Chem 1998;44:92-95.[Abstract/Free Full Text]
  10. Kouri T, Kähkönen U, Malminiemi K, Vuento R, Rowan M. Evaluation of Sysmex UF-100 urine flow cytometer vs chamber counting of supravitally stained specimens and conventional bacterial cultures. Am J Clin Pathol 1999;112:25-35.[Web of Science][Medline] [Order article via Infotrieve]
  11. Langlois M, Delanghe J, Steyaert S, Everaert K, De Buyzere M. Automated flow cytometry compared with an automated dipstick reader for urinalysis. Clin Chem 1999;45:118-122.[Abstract/Free Full Text]
  12. Kouri T, Fogazzi G, Gant V, Hallander H, Hofmann W, Guder WG. ECLM. European urinalysis guidelines. Scand J Clin Lab Invest 2000;60:1-96.[Web of Science][Medline] [Order article via Infotrieve]
  13. Whelan PV, McGill C, Ng K. Evaluation of the BMC Chemstrip urine analyzer and Chemstrip 10 UA test strip with emphasis on comparison with analytical, chemical, physical, and microscopic methods [Abstract]. Clin Chem 1993;39:1218.
  14. Gambke B, Kouri T, Kutter D, Nagel D, Schmid R, Scholer A. Miditron® multicenter evaluation. Klin Lab 1994;40:262-268.
  15. Delanghe JR, Kouri TT, Huber AR, Hannemann-Pohl K, Guder WG, Lun A, et al. The role of automated urine particle flow cytometry in clinical practice. Clin Chim Acta 2000;301:1-18.[Web of Science][Medline] [Order article via Infotrieve]
  16. Van Acker J, Delanghe J, Langlois M, Taes Y, De Buyzere M, Verstraete G. Automated flow cytometric analysis of cerebrospinal fluid. Clin Chem 2001;47:556-560.[Abstract/Free Full Text]
  17. Orsonneau JL, Douet P, Massoubre C, Lustenberger P, Bernard S. An improved pyrogallol red-molybdate method for determining total urinary protein. Clin Chem 1989;35:2233-2236.[Abstract/Free Full Text]
  18. Westgard JO. Method validation-the detection limit experiment. http://www.westgard.com/lesson29.htm (Accessed May 2002)..
  19. Froom P, Bieganiec B, Ehrenrich Z, Barak M. Stability of common analytes in urine refrigerated for 24 h before automated analysis by test strips. Clin Chem 2000;46:1384-1386.[Abstract/Free Full Text]
  20. Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002;48:436-472.[Abstract/Free Full Text]
  21. Boege F. Urinary proteins. Thomas L eds. Clinical laboratory diagnostics 1998:385-386 TH-Books Verlagsgesellschaft Frankfurt, Germany. .
  22. Pugia MJ, Lott JA, Kajima J, Saambe T, Sasaki M, Kuromoto K, et al. Screening school children for albuminuria, proteinuria and occult blood with dipsticks. Clin Chem Lab Med 1999;37:149-157.[Web of Science][Medline] [Order article via Infotrieve]
  23. Kutter D. The urine test strip of the future. Clin Chim Acta 2000;297:297-304.[Web of Science][Medline] [Order article via Infotrieve]
  24. Bennett PH, Haffner S, Kasiske BL, Keane WF, Mogensen CE, Parving HH, et al. Screening and management of microalbuminuria in patients with diabetes mellitus: recommendations to the Scientific Advisory Board of the National Kidney Foundation from an ad hoc committee of the Council on Diabetes Mellitus of the National Kidney Foundation. Am J Kidney Dis 1995;25:107-112.[Web of Science][Medline] [Order article via Infotrieve]
  25. Fenili D, Pirovano B. The automation of sediment analysis using a new urine flow cytometer (UF-100). Clin Chem Lab Med 1998;36:909-917.[Web of Science][Medline] [Order article via Infotrieve]
  26. Regeniter A, Haenni V, Risch L, Köchli HP, Colombo JP, Frei R. Urine analysis performed by flow cytometry: reference range determination and comparison to morphological findings, dipstick chemistry and bacterial culture results – a multicenter study. Clin Nephrol 2001;55:384-392.[Web of Science][Medline] [Order article via Infotrieve]
  27. Modde H, Rodriguez S. Urin Mikrobiologie. Colombo JP eds. Klinisch-Chemische urindiagnostik 1994:310 Labolife Verlagsgemeinschaft Rotkreuz, Switzerland. .
  28. Mattenheimer H, Adams EC. Quantitative determination of hemoglobin in urine. 1. The inhibitory effect of urine on the peroxidase-like activity of hemoglobin and on horseradish peroxidase. Z Klin Chem Klin Biochem 1967;5:48-54.[Web of Science][Medline] [Order article via Infotrieve]
  29. Ottomano C. Tecnologia ed automazione dell’ esame urine quantitativo. Riv Med Lab-JLM 2002;3:60-70.
  30. Mattenheimer H, Adams EC. The peroxidase-like activity of the hemoglobin-haptoglobin complex. Z Klin Chem Klin Biochem 1968;6:69-78.[Web of Science][Medline] [Order article via Infotrieve]
  31. Van Acker J, Verstraete A, Van Hamme M, Delanghe J. Falsely increased urinary caffeine attributable to contamination by urine test strips. Clin Chem 1999;45:1315-1317.[Free Full Text]



The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
B. Lumbreras-Lacarra, J. M. Ramos-Rincon, and I. Hernandez-Aguado
Methodology in Diagnostic Laboratory Test Research in Clinical Chemistry and Clinical Chemistry and Laboratory Medicine
Clin. Chem., March 1, 2004; 50(3): 530 - 536.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (11)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Penders, J.
Right arrow Articles by Delanghe, J. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Penders, J.
Right arrow Articles by Delanghe, J. R.
Related Collections
Right arrow General Clinical Chemistry
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS