|
|
||||||||
Technical Briefs |
Department of Laboratory Medicine, St. Josephs Hospital, 50 Charlton Ave. East, Hamilton, Ontario, L8N 4A6 Canada;
aauthor for correspondence: fax 905-521-6185, e-mail dgauthie{at}stjosham.on.ca
Gabapentin (Neurontin®), one of the relatively new antiepileptic drugs, has a novel mechanism of action that is not fully understood. Gabapentin has been shown to increase
-aminobutyric acid concentrations in the brain of epilepsy patients (1)(2). There is an approximate linear relationship between dose and plasma concentrations. A therapeutic range of 220 mg/L has been commonly accepted (2), although a higher minimum concentration of 6 mg/L has also been suggested (1).
Various analytical methods have been used for the determination of gabapentin in plasma, with liquid chromatography being the most commonly used approach in clinical laboratories. Gabapentin does not have a natural chromophore; therefore, it must be derivatized to be detected spectrophotometrically or fluorometrically unless detected by mass spectrometry (3). Gabapentin has been derivatized with 2,4,6-trinitrobenzenesulfonic acid for photometric detection (4)(5). Several approaches have been described for the preparation of fluorescent derivatives. Garcia et al. (6) derivatized gabapentin with fluorescamine for determination by capillary electrophoresis. However, the preparation of o-phthalaldehyde (OPA) derivatives of gabapentin has been the most popular and practical approach for liquid chromatographic determinations (7)(8)(9)(10)(11)(12).
In all of these procedures, plasma proteins are precipitated by different reagents and the resulting supernatant is treated with the derivatizing reagent. As a result, all endogenous amino compounds produce fluorescent derivatives, which produce very large solvent peaks, decreasing the sensitivity and increasing the length of the chromatographic run times. These procedures have another limitation in that the OPA derivatives must be injected almost immediately after preparation because of their instability.
We evaluated an alternative liquid chromatographic procedure to reduce extraneous peaks and improve the stability of OPA derivatives for the determination of gabapentin in plasma.
A 200-µL aliquot of a plasma sample was mixed with 200 µL of a 5 mg/L solution of the internal standard [1-(aminomethyl)cycloheptane acetic acid)] in a 15 g/L sodium borate (borax; Sigma-Aldrich) solution and applied to a 1-mL Bond Elut C18 extraction column (cat. no. 12102001; Varian) that had been activated by washing with one column volume of methanol and one column volume of water. High-purity Omnisolv brand solvents (methanol and acetonitrile) were obtained from Merck. Deionized water was distilled in a glass still. The sample was pushed through the column at a very slow rate (
90 s) by use of a 1-mL plastic syringe (cat. no. 309626; Becton-Dickinson) and an adapter (cat. no. 12131001; Varian) to fit the C18 extraction column. The extraction column was then drained by suction and washed with 200 µL of a 15 g/L sodium borate solution, and the compounds of interest were eluted with 250 µL of methanol.
The eluate was mixed with 200 µL of a 1 g/L methanolic solution of OPA containing 1 mL/L mercaptoethanol. A 3-µL aliquot of the mixture was injected on a Ultrasphere octyl silica column [15 cm x 4.6 mm (i.d.); 5-µm particle size; Beckman] protected with a C8 guard column. A mobile phase containing methanol (200 mL), acetonitrile (400 mL), water (500 mL), 700 g/L perchloric acid (0.5 mL; JT Baker), and a 160 g/L methanolic solution of tetramethylammonium hydroxide (0.5 mL; Sigma-Aldrich) was pumped through the column at a flow rate of 1.6 mL/min. The instrument used was a Varian integrated "PROSTAR" HPLC system consisting of a pump, autosampler, fluorescence detector, and data management system. The peaks were detected by use of an excitation wavelength of 350 nm and an emission wavelength of 450 nm. The drug and the internal standard showed fluorescence response without any delay.
The linearity of the method was determined by serial dilution (eight concentrations) of a 40 mg/L gabapentin calibrator in drug-free plasma: y = 0.1956x - 0.003 (r = 0.998). The lower limit of quantification was 0.3 mg/L, which is well below the therapeutic range (220 mg/L). The within-run CVs were 3.0% (n = 10) for both the low (mean = 0.5 mg/L) and the high (mean = 20.7 mg/L) gabapentin concentrations. Between-run CVs (n = 10) were 6.7% (mean = 0.6 mg/L) for the low concentration and 3.7% (mean = 20.6 mg/L) for the high concentration.
Quality-control sera (Bio-Rad TDM levels 1 and 3) containing acetaminophen, amikacin, amitriptyline, caffeine, carbamazepine, chloramphenicol, desipramine, digoxin, disopyramide, ethosuximide, gentamicin, imipramine, lidocaine, lithium, methotrexate, N-acetylprocainamide, netilimicin, nortriptyline, phenobarbital, phenytoin, primodone, procainamide, propranolol, quinidine, salicylate, theophylline, tobramycin, valproic acid, and vancomycin did not show any interfering peaks when processed. Hemolyzed or moderately lipemic samples did not affect the recovery of the drug or internal standard.
The chromatogram of drug-free plasma (Fig. 1A
) showed only a few early-eluting peaks. There were no peaks eluting after the internal standard. Gabapentin is a polar amino acid and cannot be isolated by liquidliquid extraction. For gas chromatographic procedures, gabapentin has been isolated with cation-exchange (13) or mixed-mode solid-phase extraction (14)(15). Cation-exchange extraction of gabapentin does not offer any advantage because all the endogenous amino compounds form cations and are coextracted. Furthermore, cation-exchange columns must be eluted with ammoniac methanol to recover gabapentin, and the ammonia must be completely removed by evaporation before derivatization with OPA. Kushnir et al. (15) reported that gabapentin is not significantly retained by commonly used 1-mL C18 or CN extraction columns. On the other hand, Lensmeyer et al. (4) isolated gabapentin in 9298% yields with a Bond Elut C18 extraction column. These authors passed the sample through a C18 column at a very slow rate. We also found that both gabapentin and the internal standard are recovered from plasma in a 9098% yield after extraction with 1-mL C18 extraction columns. The recovery of gabapentin is drastically reduced when the sample passes through the C18 sorbent at a relatively fast rate. However, endogenous amino compounds present in plasma are more polar than gabapentin and are not retained by C18 sorbent even when the sample is passed at a slow rate.
|
Different authors have used different compositions of OPA derivatizing agent, a mixture of a methanolic solution of OPA, a thiol, and borate buffer, but the fluorescent derivatives are produced in similar yields. The presence of water in the reagent makes the OPA derivative unstable, and the stability can be improved by decreasing the water content (16). In our described procedure, the derivatizing reagent does not contain borate buffer. The methanolic eluate obtained from the C18 extraction column contains traces of borate buffer, which is adequate to initiate the reaction. The resulting OPA derivatives are stable for at least 4 h at room temperature and for at least 24 h when stored at 48 °C.
Gabapentin OPA derivatives have been detected at excitation wavelengths of 230250 or 330350 nm with a common emission wavelength of 430450 nm. Excitation at 230250 nm provides
2.5 times the sensitivity of that observed at 330350 nm with the same attenuation, slit width, and emission wavelength. However, the baseline is unstable at excitation wavelengths of 230250 nm, and the detector must be set at a higher attenuation to decrease the noise. The procedure presented here is extremely sensitive even at an excitation wavelength of 350 nm.
-Aminocyclohexanepropionic acid (cat. no. 43,805-7; Aldrich) behaved similar to gabapentin and its analog for solid-phase extraction. It produced a fluorescent OPA derivative that eluted close to the gabapentin analog commonly used as the internal standard. However, the fluorescence response of this compound was much weaker than that of gabapentin, and it had to be used at a concentration of 50 mg/L. This did not affect the linearity of the procedure.
Acknowledgments
We thank Dr. A. Fraser of Victoria General Hospital, Halifax, and Pfizer Canada for supplying us with samples of gabapentin and 1-(aminomethyl)cycloheptane acetic acid.
References
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |