|
|
||||||||
1
Department of Biotechnology, University of Turku, Tykistökatu 6A, 6th Floor, FIN-20520 Turku, Finland.
2
PerkinElmer Life Sciences/Wallac Oy, PO Box 10, FIN-20101 Turku, Finland.
aAuthor for correspondence. Fax 358-2-333-8050; e-mail piia.tarkkinen{at}utu.fi.b Current address: AboaTech Ltd, FIN-20520 Turku, Finland.
| Abstract |
|---|
|
|
|---|
Methods: A noncompetitive, kinetic CRP immunoassay was developed that uses individual, porous microparticles as the solid phase. The microparticles were covalently coated with a monoclonal capture antibody, and the monoclonal detection antibody was labeled with europium. The one-step binding reaction was stopped by washing after 2 min of incubation, and the fluorescence signal of individual particles was measured.
Results: The analytical detection limit (mean of zero calibrator + 3 SD) was 0.00016 mg/L CRP. Clinical samples were diluted 400-fold before assay to cover the CRP concentration range of 0.0641200 mg/L. The assay correlated well with the Dade Behring N High Sensitivity CRP assay (for 010 mg/L, r = 0.969, Sy|x = 0.68, n = 54; for 0350 mg/L, r = 0.969, Sy|x = 11.7, n = 100). The within- and between-run CVs based on calculated concentrations were, respectively, 916% and 14% at 0.11 mg/L, 4.512% and 8.2% at 4.2 mg/L, and 3.56.3% and 4.4% at 105 mg/L, with a CV <15% at 0.2 mg/L and above.
Conclusions: Use of the kinetic microparticle approach combined with time-resolved fluorometry allows ultrasensitive quantification of CRP in whole blood in 2 min with a linear assay range spanning more than four orders of magnitude.
| Introduction |
|---|
|
|
|---|
5 mg/L, measurement of concentrations relevant to heart-related risk prediction requires a detection limit at least 10 times lower (1)(12), thus giving rise to a need for a new generation of CRP assays. Quantitative measurement of CRP as a marker of inflammation is already among the most common tests performed in clinical laboratories (20), making it a strong candidate for point-of-care testing (POCT) as well. Consequently, the turnaround times of the assays have become increasingly important, and CRP assays suitable for POCT have been developed at a rapid pace (21)(22)(23). However, although the overall sensitivity of CRP assays has improved considerably because of the new applications described above, the present high-sensitivity assays facilitate CRP measurements only in clinical laboratory rather than in POC conditions. Correspondingly, the assays suitable for POCT generally use detection technologies that do not allow the detection of low analyte concentrations. Therefore, the present POC assays for CRP are well suited for the detection of, e.g., bacterial or viral infections or for follow-up of antiinflammatory therapy, but are inadequate for detecting CRP at the concentrations required for heart disease risk assessment.
In this study, we explored the possibility of using a kinetic one-step assay concept to develop a CRP assay capable of measuring a wide range of concentrations and suitable for POC situations. Porous microparticles served as a mobile, high-capacity solid phase to facilitate efficient collection of analyte molecules in a limited time. The high specific activities of lanthanide chelates (24)(25)(26) and the superior sensitivity and dynamic range of time-resolved fluorometry (27)(28)(29)(30) allowed sensitive detection of both very low and high analyte quantities in submicroliter volumes of whole blood with a single sample dilution. Because of the sensitive detection technology, results could be obtained long before the immunoassay reached equilibrium without compromising assay performance.
| Materials and Methods |
|---|
|
|
|---|
83% of the total pore volume.
covalent coating of the microparticles
The epoxy-surfaced microparticles were covalently coated with a monoclonal, CRP-specific capture antibody purchased from Medix Biochemica. The coating was performed batchwise, using
1 x 104 microparticles per reaction. For immobilization, 15 ng of antibody/particle was added in 24 nL of phosphate-buffered saline (8.0 g/L NaCl, 0.2 g/L KCl, 1.2 g/L Na2HPO4 · 7 H2O, and 0.2 g/L KH2PO4), pH 7.2. The reaction was incubated for 20 h at room temperature in a Rotamix mixer (Heto Lab Equipment). After coating, the particles were washed three times with 2 mmol/L Tris-HCl, pH 7.5, containing 0.1 g/L Tween 20. The coated particles were stored at 4 °C.
labeling of the detection antibody
A monoclonal CRP-specific antibody (purchased from HyTest Ltd.) was labeled with the intrinsically fluorescent Eu3+ chelate of 4-[2-(4-isothiocyanatophenyl)ethynyl]-2,6-bis{[N,N-bis(carboxymethyl)amino]methyl}pyridine (PerkinElmer Life Sciences/Wallac Oy). The reaction was carried out overnight at 4 °C with a 75-fold molar excess of the chelate in 50 mmol/L NaHCO3, pH 9.8. Excess label was removed by gel filtration on a Superdex 200 HR 10/30 column (Pharmacia Biotech) with Tris-saline-azide buffer (6.1 g/L Tris, 9.0 g/L NaCl, and 0.5 g/L NaN3), pH 7.5, as elution buffer. The fractions containing the antibody were pooled, and diethylenetriamine pentaacetic acid-treated bovine serum albumin (PerkinElmer Life Sciences/Wallac Oy) was added to a final concentration of 1 g/L. The labeling degree of the antibody was 5.5 Eu3+/IgG molecule.
crp calibrators and clinical samples
The CRP calibration material was acquired from HyTest Ltd. According to the manufacturer, the material had been produced from human pleural fluid or plasma, and its purity was >98%. Two hundred EDTA-whole blood samples collected for routine CRP measurements were obtained from the Turku University Central Hospital. The hematocrit (Hct) of each sample was determined by centrifugation (Biofuge Hemeo; Heraeus Instruments). From each of the 100 samples used in the correlation studies, a small volume (200 µL) of whole blood was set aside for the microparticle assay, and the remainder was centrifuged for 15 min at 2600g (Centrifuge 5804 R; Eppendorf) to separate the plasma. The whole blood samples were analyzed fresh, whereas the plasma samples were stored deep-frozen for 1 week at -70 °C before being assayed in the Dade Behring BN II nephelometer with the Dade Behring N High Sensitivity CRP assay. In the microparticle assays, DELFIA Assay Buffer (PerkinElmer Life Sciences/Wallac Oy) was used as the diluent for both the calibrators and clinical samples.
assay kinetics
All immunoassays were performed with individual microparticles coated with an anti-CRP capture antibody as solid phase. The kinetics of the different steps of the sandwich-type CRP immunoassay were determined using an assay volume of 330 nL/particle (3 particles/µL), the sample volume being one-third of the total volume. The assays were performed with CRP calibrator and detection antibody concentrations of 0.5 and 9.3 mg/L, respectively, and the concentrations were kept invariable for each assay condition. The binding reactions were performed in DELFIA Assay Buffer containing 3.5 g/L Tween 20. For practical reasons, the incubations were performed using a batch of 21 microparticles in an assay volume of 7 µL/tube. The reactions were incubated at room temperature in a vortex-type mixer (Scientific Industries) for different time periods (01150 min), after which the reaction was interrupted by washing twice with 200 µL of DELFIA Wash Solution (PerkinElmer Life Sciences/Wallac Oy). The particles were transferred and dried on a Cyclopore track-etched polycarbonate membrane (Whatman Polyfiltronics) to measure the signal of individual microparticles in a time-resolved microfluoro-meter (30). For each time point, four replica particles were measured.
In addition to the one-step kinetics (i.e., analyte and detection antibody added simultaneously in the reaction), the kinetics of the analyte and the detection antibody binding reactions were studied separately as well. These assays were performed essentially in the same way as above, but with the analyte and tracer incubated sequentially and the missing reagent being replaced by DELFIA Assay Buffer.
kinetic crp immunoassay
Effect of microparticle and detection antibody concentrations.
To examine the performance of a kinetic microparticle immunoassay, we studied the effect of microparticle concentration, using 1, 10, or 100 particles/µL (i.e., assay volume of 1000, 100, or 10 nL/particle, respectively). The total assay volumes were adjusted to 620 µL to facilitate manual handling of the reactions and to include at least 20 microparticles per reaction. Accordingly, the reactions were performed using the following three combinations: 20 particles/20 µL, 60 particles/6 µL, and 600 particles/6 µL. The detection antibody concentration (6.2 mg/L) and the ratio of sample volume to total volume (one-third) were kept constant for each assay condition. Experiments to assess the influence of different detection antibody concentrations were performed with 3 microparticles/µL (i.e., assay volume of 330 nL/particle). Three detection antibody concentrations, 6.2, 31.0, and 124.0 mg/L, were studied at variable analyte concentrations. All reactions were performed in a one-step format, i.e., by dispensing the detection antibody and the analyte simultaneously onto the particles. Otherwise, the same assay procedure as described under "Assay Kinetics" was used with an incubation time of 2 min.
Calibration curve and clinical samples.
To avoid variations in incubation times and temperatures that could distort the quantitative interpretation of the results, the clinical samples and the CRP calibrators used for generating the calibration curve were run together in same assay cycles. The 2-min assays were performed essentially in the same way as described above, with a total assay volume of 330 nL/particle and a detection antibody concentration of 6.2 mg/L. The analytical detection limit of the kinetic assay was calculated as the mean signal of the zero calibrator + 3 SD.
comparison method
The Dade Behring N High Sensitivity CRP assay was used as the comparison method in this study. The assay was performed in the Dade Behring BN II nephelometer with the manufacturers reagents as directed. According to the manufacturer, the assay can be used to measure CRP concentrations in the range of
0.181100 mg/L.
| Results |
|---|
|
|
|---|
20 min. The rapid kinetics can be explained by the 15-fold molar excess of the antibody compared with the amount of analyte used. The kinetics could be further improved by increasing the antibody excess, although the effect on background signal and assay sensitivity should also be taken into consideration. The analyte-binding step reached equilibrium at
240 min. The one-step kinetics, however, were unexpectedly slow, not reaching plateau even after 20 h of incubation.
|
|
As demonstrated previously, the number of microparticles per microliter is in linear proportion to the assay kinetics (32). The same effect was observed here as well: increasing the number of particles per microliter by a factor of 10 correspondingly increased assay kinetics 10-fold (data not shown). However, assuming that the correlation between the number of particles and the kinetics remained constant, generating a microparticle CRP assay that reached equilibrium in
2 min would require reducing the assay volume per particle by a factor of 600 (Table 1
). In other words, instead of 3 microparticles/µL, 1800 particles/µL would be required to reach the desired kinetics, an approach clearly not feasible. Because the aim of the study, however, was to develop a rapid assay for measuring CRP under POC situations, another alternative, a kinetic approach, was taken under investigation.
|
kinetic crp immunoassay
Effect of microparticle concentration.
In common endpoint immunoassays, the sensitivity of an assay is strongly dependent on the sample volume or, in other words, on the total amount of analyte per reaction. In this study, we investigated the effect of the total analyte quantity on another kind of assay approach, a kinetic microparticle application, where the one-step binding reaction was interrupted before reaching equilibrium. We studied the effect of sample amount by varying the assay volume per particle but keeping the ratio of sample to assay volume (i.e., analyte concentration) constant. The proportion of sample was invariably one-third of the total volume, and a similar 2-min incubation time was used for each assay. Fig. 3
shows the fluorescence intensities at different CRP concentrations in total assay volumes of 10, 100, and 1000 nL/microparticle (i.e., 600 particles/6 µL, 60 particles/6 µL, and 20 particles/20 µL, respectively).
|
Effect of detection antibody concentration on the high-dose hook phenomenon.
Because no major difference was found between the assay volumes studied (101000 nL/particle), we selected a 330-nL volume for further analysis because of the relative ease of handling and our earlier experience with this assay volume. Subsequently, we studied the dynamics of the assay and the extent of high-dose hook interference typical of quantitative one-step assays (33) in detail, using various amounts of the detection antibody. The dynamic range of the kinetic microparticle assay was more than four orders of magnitude (Fig. 4
) at each of the detection antibody concentrations studied (6.2, 31.0, and 124.0 mg/L). Furthermore, the signal obtained at a CRP concentration of 20 mg/L was, under each assay condition, at least equivalent to the signal achieved at a CRP concentration of 2 mg/L. Calculated from the most heavily bending curve (i.e., with a detection antibody concentration of 6.2 mg/L), CRP concentrations
12 mg/L had no effect on result interpretation at an assay range of 03 mg/L CRP. To cover the CRP concentration range of 0.0641200 mg/L, clinical samples were diluted 400-fold before analysis.
|
Calibration curve and clinical samples.
The calibration curve for the 2-min kinetic microparticle assay was established using an assay volume of 330 nL/particle. A typical curve based on four replicate particles is shown in Fig. 5
together with a within-assay precision profile. The analytical detection limit of the assay was 0.00016 mg/L CRP, calculated as the mean signal of the zero calibrator + 3 SD. In addition, the calibration curve was linear up to 3 mg/L CRP, i.e., more than four orders of magnitude.
|
The within- and between-run variations for EDTA-whole blood samples are shown in Table 2
. Because of the poor stability of whole blood samples, the assays were performed on 2 sequential days, two runs per day, with four replicate particles per assay. As could be predicted, the highest imprecision was seen at the lowest CRP concentrations. The functional detection limit of the assay, i.e., the lowest clinical CRP concentration that could be measured with a CV <15%, was estimated to be 0.2 mg/L.
|
We studied the correlation between the microparticle (y; EDTA whole blood) and the Dade Behring N High Sensitivity CRP (x; EDTA plasma) assays with 100 patient samples, using both Hct-corrected and uncorrected concentration values obtained from the microparticle assay (Fig. 6
). In addition to the total CRP concentration range, the hs-CRP range from 0 to 10 mg/L was analyzed separately. The slopes for the uncorrected concentration values (Fig. 6A
) were close to those expected based on plasma (and thus also analyte) displacement by red blood cells in the sample (for hs-CRP, y = 0.604x, Sy|x = 0.458; for total CRP, y = 0.571x, Sy|x =7.733), yielding lower concentrations measured in the microparticle assay compared with the Dade Behring assay. After Hct correction (Fig. 6B
), however, the slope was largely normalized (for hs-CRP, y = 0.933x, Sy|x =0.680; for total CRP, y = 0.852x, Sy|x =11.735). The intercepts for the Hct-corrected hs-CRP and total-CRP curves were 0.075 ± 0.137 (r = 0.969; n = 54) and 2.135 ± 1.396 mg/L (r = 0.969; n = 100), respectively (P <0.0001).
|
In the whole blood samples used in the correlation studies (n = 100), the Hct values ranged from 0.21 to 0.47. In addition, a mean Hct value was calculated from a total of 200 EDTA whole blood samples collected for routine CRP analysis at the local hospital (mean Hct, 0.357; range, 0.210.50). The CRP concentrations obtained with individual Hct or the mean Hct values for correction were compared with each other to determine the bias between the correction methods. Linear regression analysis (Fig. 7A
) and difference analysis (Fig. 7B
) both showed excellent agreement between the two methods for Hct correction, although some deviation of single patient samples was noted. For CRP concentrations <5 mg/L, the regression analysis yielded a slope of 1.053 ± 0.016 (Sy|x =0.135) and a y-intercept of -0.079 ± 0.033 mg/L (r = 0.996; n = 41). The corresponding values over the entire range of CRP concentrations (0250 mg/L) were as follows: slope, 1.040 ± 0.006 (Sy|x = 2.947); y-intercept, -0.221 ± 0.355 (r = 0.998; n = 100). The difference analysis revealed no bias at CRP concentrations of 05 mg/L and a bias of 1.05 mg/L over the entire CRP concentration range (0250 mg/L).
|
Because the risk of a future coronary event has been found to increase as a function of hs-CRP concentration (9)(10)(11)(12)(13)(14)(15)(16), we evaluated the effect of fixed Hct correction on cardiovascular risk assessment, using the quintile cutoff values of 0.7, 1.2, 2.0, and 3.9 mg/L suggested by Rifai and Ridker (34). In the population with CRP <5 mg/L (n = 41), one sample with a Hct of 0.28 was classified into a different quintile based on a slightly overestimated CRP concentration because the Hct correction based on the individually determined or predetermined values produced CRP concentrations of 3.7 and 3.9 mg/L, respectively (Fig. 7A
, inset).
| Discussion |
|---|
|
|
|---|
Another important advantage of using a stable lanthanide chelate as a label was that unprocessed whole blood could be used as the sample material in the one-step assay without disturbance from red blood cells or EDTA, which is a strong chelating agent and a commonly used anticoagulant in blood samples. Because of the extremely sensitive assay concept, the EDTA-whole blood samples used in the correlation studies were diluted 400-fold before being assayed to cover a CRP concentration range of 0.0641200 mg/L. The range could have easily been shifted by altering the dilution factor, but the wide dynamic range of the microparticle assay eliminated the need for multiple sample dilutions. In addition, a high dilution factor is very likely to have a positive contribution in reducing any blood-derived assay interference from the whole blood samples.
The effect of Hct correction in whole blood samples became evident in the correlation study between the microparticle and the Dade Behring N High Sensitivity assays. The lower slope obtained for the uncorrected concentration values was directly related to Hct and the displacing effect of red blood cells in the sample. Thus, after correction for Hct, the slope was normalized. The overall correlation between the microparticle and the Dade Behring assays was good, but with some deviation seen at higher CRP concentrations. The best slopes between the two assays were seen at concentrations <10 mg/L, and samples above this limit gave systematically lower concentrations in the microparticle assay compared with the Dade Behring assay. In the latter assay, CRP concentrations <10 mg/L are assayed at an initial sample dilution of 1:20, whereas samples with high concentrations are automatically diluted 1:100 (>10 mg/L) or 1:400 (>50 mg/L), which may have an effect on assay linearity. In the microparticle assay, all samples were assayed at a 1:400 dilution.
Only a minor bias was noted between concentrations corrected by use of an individually determined Hct value or a predetermined mean Hct value. As was also discussed by Pettersson et al. (35), typical variations in Hct values have fairly insignificant effects on assay outcomes, rendering individual Hct determinations unnecessary in the case of many test analytes. However, the use of a predetermined mean Hct value unavoidably leads to slightly overestimated concentrations when the Hct is low and, correspondingly, slightly underestimated concentrations when the Hct is high. As a result, incorrect classification of single patient samples with low or high Hct values is possible, especially near cutoff concentrations. In this study, however, only 1 patient sample (n = 41) was classified into a different quintile based on a slightly overestimated CRP concentration. For markedly increased concentrations, the bias becomes even more insignificant.
The effect of the high-dose hook phenomenon on the one-step kinetic microparticle assay was also insignificant. Considering the predilution of clinical samples, only a CRP concentration >4800 mg/L would affect result interpretation, and this concentration is far higher than the concentrations seen in clinical situations. In fact, CRP concentrations in the range of 100-1000 mg/L are rare and indicate a very serious clinical condition (36)(37)(38). An assay that also covers this range, however, facilitates sensitive follow-up of the effect of medication, e.g., in field circumstances, where training in instrument operation and parallel diagnostic methods may be limited. In conclusion, because the clinical CRP concentration range for evaluating the risk of future cardiac complications is
0.55 mg/L and concentrations >5 mg/L indicate a bacterial or viral infection, the microparticle assay developed can be used for any application involving measurement of CRP concentrations in clinical samples.
We have previously demonstrated the clinical functionality of a customary endpoint assay that uses similar microparticles coated with Fab fragments specific for prostate-specific antigen (PSA) (39). When we used the same assay volume (330 nL/particle) as in this study, the equilibrium of the one-step assay was reached in 240 min. Slower kinetics were observed for the one-step CRP assay, where reaching equilibrium took >1200 min. The difference in analyte-binding kinetics between the two assays was, however, smaller (160 min for PSA; 240 min for CRP). The kinetic differences are likely to originate directly from the much larger size of CRP (114 kDa) compared with PSA (33 kDa), which not only slows the diffusion rate in the microparticle pores, but also makes entering the pores more difficult because of steric hindrance. The extensive differences in the analyte and the one-step kinetics of the CRP assay can be explained by the increased molecular size as well: whereas in the first case the analyte and detection antibody diffuse separately in the reaction mixture and microparticle pores, in the latter case they form a liquid-phase complex before entering the pores. Because of the large molecular excess of the detection antibody and the pentameric molecular form of CRP, the complex may become notably large compared with single analyte and antibody molecules. In addition to the size-dependent effects mentioned above, this may also slow down and complicate the solid-phase binding reaction. If the current CRP assay were to be performed as an endpoint measurement, however, the slow kinetics could at least to some extent be improved by use of microparticles with a larger pore size or, alternatively, use of antibody fragments instead of intact antibodies for antigen capture and detection (39)(40). However, to reach a 2-min incubation time with the current assay design would require reducing the assay volumes to 3.0 and 0.6 nL/particle in the PSA and CRP assays, respectively. In addition to poor assay sensitivities, this would lead to impractical assay volumes or to the use of a notably higher number of microparticles per assay. Given that individual microparticles are measured, the customary equilibrium/endpoint assay is not feasible for POCT.
The kinetic assay format allowed us to generate a clinically functional, ultrasensitive CRP assay with an incubation time of only 2 min. Another key aspect of the assay developed was the independence of assay performance from microparticle concentration. Only a 1.5-fold change in the net signal was detected regardless of the 100-fold difference in the surface-to-volume ratio, obviously because of the low percentage of sample bound during the 2-min incubation time. In other words, because none of the assays was near equilibrium and the amount of bound analyte remained lower than the amount of unbound analyte throughout the 2-min incubation period, moderate changes in microparticle concentration or, more accurately, in the total assay volume per particle were irrelevant. Thus, the kinetic microparticle assay measured the initial rate of analyte association rather than the total amount of the antigen.
The revolution in assay and detection technologies has recently facilitated the development of genuinely portable instrumentation for extralaboratory testing, and commercial, portable POC CRP assays with extremely short turnaround times have already been developed. For example, the 2-min reflectometric NycoCard® CRP Whole Blood test (Axis Shield) (21) and the turbidimetric, 3-min QuikRead® CRP whole blood/serum/plasma assay (Orion Diagnostica) (23) represent pioneering quantitative test systems suitable for extralaboratory testing. In the flow-through NycoCard assay, the results are corrected with an assumed Hct value of 0.40, and the assay also facilitates visual interpretation of results. Both assay procedures, however, require manual dispensing of the assay-specific reagents and the sample. Because of the detection technologies used in these assays, the detection limits of the assays are high, 10 mg/L for the NycoCard and 8 mg/L for the QuikRead, and the measurement ranges cover less than two orders of magnitude. Thus, neither of these assays comes close to the sensitivity or the dynamic range seen in the 2-min kinetic microparticle approach. In fact, the difference in analytical detection limits between the microparticle and the above-mentioned assays is more than 30 000-fold.
Although many assay steps were performed manually in the present study, total automation of the assays can be accomplished. The fact that the kinetic assay concept described can easily bear changes in assay volumes or microparticle concentration gives much more flexibility in adjusting the assay conditions with an instrument platform, and it also enables more robust production of the assay itself. In addition, the batchwise procedure for coating the particles increases the ease and flexibility of production. Because it is substantially faster to perform than most of the current high-sensitivity immunoassays, the kinetic assay format should be well suited for POC purposes. Because assay systems for extralaboratory use require at least some extent of automation to reduce the risk of human error and to increase the accuracy of the results, discontinuing the reactions before equilibrium should not present a notable additional challenge for instrumentation. The assay-to-assay variation typical of manually performed kinetic assays could easily be avoided by exact device adjustment and programming of the incubation times and temperatures.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
R. Vikstedt, P. von Lode, T. Takala, K. Irjala, O. Peltola, K. Pettersson, and P. Suominen Rapid One-Step Immunofluorometric Assay for Measuring Soluble Transferrin Receptor in Whole Blood Clin. Chem., October 1, 2004; 50(10): 1831 - 1833. [Full Text] [PDF] |
||||
![]() |
P. von Lode, J. Rainaho, and K. Pettersson Quantitative, Wide-Range, 5-Minute Point-of-Care Immunoassay for Total Human Chorionic Gonadotropin in Whole Blood Clin. Chem., June 1, 2004; 50(6): 1026 - 1035. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |