| Abstract |
|---|
|
|
|---|
Methods: The CD shuttled cholesterol reversibly between reduced and oxidized forms in the presence of thio-NAD and NADH. The production rate of thio-NADH correlated with the cholesterol concentration and was measured by the absorbance at 404/500 nm. This CD method was combined with an immunoaffinity separation procedure with specific monoclonal antibodies to apolipoprotein (apo) A1 and apo B-100 and used for RLP-C assay. Results were compared with a RLP-C method that uses cholesterol oxidase, peroxidase, and chromogenic substrate.
Results: The CD method could detect 0.10 x 10-3 mmol/L cholesterol and was at least 5 times more sensitive than the conventional enzymatic method. Within- and between-day imprecision (as CVs) of the RLP-C assay with the CD method was <4%. Regression analysis of RLP-C assays with the new (y) and conventional (x) cholesterol methods yielded: y = 1.02x - 0.008 mmol/L (Sy|x = 0.0065 mmol/L; r = 0.997; n = 297).
Conclusions: Serum RLP-C can be measured by the CD method. The CD method may be useful for other assays that require sensitive cholesterol measurements in biological materials.
| Introduction |
|---|
|
|
|---|
We have developed an enzymatic cycling method (6), using cholesterol dehydrogenase (CD; EC no. not certified) (7), that can detect 0.10 x 10-3 mmol/L cholesterol with one-point calibration. Using the CD method, we determined RLP-C concentrations in sera from healthy controls.
| Materials and Methods |
|---|
|
|
|---|
procedures
Principle of the CD method.
CEH and CD were used in the CD method (Fig. 1
). The cholesterol esters were hydrolyzed to free cholesterol and fatty acids by CEH. Free cholesterol changed to the oxidized form reversibly, and this cycling reaction was repeated several times in the presence of thio-NAD and NADH. Thio-NADH (reduced form) produced by the CD method was measured by the change in absorbance per minute at 404/500 nm as a measure of the cholesterol concentration.
|
Reagents for the CD method and application to an automated analyzer.
The first reagent contained, per liter, 8000 U of CEH, 1500 U of CD, 0.8 mmol thio-NADH, 2.5 g of sodium cholate, 1 mL of Triton X-100, and 10 mmol HEPES buffer (12). The second reagent contained 2.4 mmol/L NADH in 0.3 mol/L diethanolamine buffer (pH 10).
The assay was performed on a TBA-20R automatic analyzer (Toshiba) 2optimized for a rate assay measuring absorbance at 404/500 nm for 5 min at 37 °C with 10 µL of sample, 270 µL of the first reagent, and 90 µL of the second reagent.
Procedure for RLP-C.
Lipoproteins other than RLP in serum were precipitated with an immunoaffinity separation procedure using monoclonal antibodies to apolipoprotein A1 and B-100 (3)(4); the cholesterol in the supernatant was measured by the CD method as RLP-C. The calibration solution (International Reagents Corp.) contained 0.789 mmol/L (30.5 mg/dL) cholesterol.
The conventional RLP-C assay using the peroxidase (PO; donor:hydrogen peroxide oxidoreductase; EC 1.11.1.7) method (RLP-Cholesterol JIMRO II; Japan Immunoresearch Laboratory Co. Ltd.) was used as the comparison method (4)(5). Five calibration points were used to construct the calibration curve in the PO method.
samples
Fresh serum samples were randomly collected without anticoagulant as part of a mass examination at Kochi prefecture in cooperation with Osaka Medical Center for Cancer and Cardiovascular Diseases. This study was approved by the Committee for Ethical Standards of Osaka Medical Center for Cancer and Cardiovascular Disease. The 297 individuals studied in this research had no history of cardiovascular disease and were apparently healthy.
statistical analysis
Linear regression, reference intervals, and cutoff values were calculated by the least-squares method, the nonparametric method, and the 75th percentile of the population in this study group, respectively.
| Results |
|---|
|
|
|---|
|
The concentrations of thio-NAD and NADH in the CD method at 37 °C were assessed using a 0.026 mmol/L cholesterol solution. The reaction was increased by addition of the coenzymes (Fig. 2, B and C
). To achieve the necessary sensitivity and the absorbance range needed for measurement of cholesterol in RLP-C (02.587 mmol/L as RLP-C), we chose 0.6 mmol/L for both thio-NAD and NADH.
The optimal CD activity at 37 °C was assessed using a 0.026 mmol/L cholesterol solution. The sensitivity was increased in proportion to the amount of CD (Fig. 2D
). To achieve the necessary sensitivity and the absorbance range for the RLP-C assay, we used a CD activity of 1.13 kU/L.
The optimal CEH activity at 37 °C was assessed using three human sera with total cholesterol concentrations of 3.5, 5.9, and 10.8 mmol/L, respectively, which were diluted 1:401 (1 µL of serum in 400 µL of saline) in saline. The amount of CEH for the hydrolysis of cholesterol esters was fixed at 8 kU/L. This CEH activity hydrolyzed cholesterol esters of high-cholesterol sera (Fig. 2E
).
detection limit of cd method for cholesterol
The detection limit of the CD method, evaluated with a sequentially diluted cholesterol solution (1.29 mmol/L), was 0.10 x 10-3 mmol/L (mean + 3 SD of zero calibrator; n = 10).
performance of the new rlp-c assay
We measured 0, 0.098, 0.197, 0.336, and 0.789 mmol/L cholesterol solutions determined as RLP-C by the CD and conventional (PO) method (Fig. 3
). The CD method had a linear calibration curve (n = 5) that passed through the origin, with fivefold higher absorbance (sensitivity) than that of the PO method. The PO method had a nonlinear calibration curve (n = 5) that did not pass through the origin, and the increase in absorbance was disproportional with the concentration of cholesterol.
|
The linearity of the RLP-C assay with the CD method was evaluated by sequential dilution of a high RLP-C control serum with saline. The curve was linear up to 2.587 mmol/L RLP-C in this method.
Within- and between-day precision was determined by measuring three control sera containing 0.098, 0.341, 1.058 mmol/L cholesterol. Between-day testing was carried out on 10 points over a 2-week period. The assay was calibrated each testing day. The RLP-C assay with the CD method showed good within- and between-day precision [CVs, 0.573.1% (n = 20) and 1.03.8%, respectively; Table 1
]. On the other hand, the RLP-C assay with the conventional (PO) method had within- and between-day CVs of 1.86.0% and 2.07.5%, respectively.
|
The RLP-C assay with the CD method was not affected by hemoglobin (up to 5.0 g/L), bilirubin (up to 0.35 mmol/L), conjugated bilirubin (up to 0.28 mmol/L), or ascorbic acid (up to 2.84 mmol/L; data not shown).
serum rlp-c concentrations in apparently healthy people
The observed RLP-C concentrations in this study group (n = 297) with the CD method were 0.0800.437 mmol/L, whereas the results obtained by the conventional (PO) RLP-C assay were 0.0850.419 mmol/L. The 75th percentile was 0.217 mmol/L with the CD method and 0.220 mmol/L by the conventional (PO) assay.
The correlation by linear regression of RLP-C assay with the CD method (y) and conventional RLP-C assay (x) was: y = 1.02x - 0.008 mmol/L (Fig. 4
). The mean (SD) of x was 0.188 (0.088) mmol/L, and the mean (SD) of y was 0.184 (0.090) mmol/L. The mean difference between methods was 0.004 mmol/L, and the SD of residuals (Sy|x) was 0.0065 mmol/L. Samples with <0.13 mmol/L RLP-C were 4060% higher by the new method than by the conventional RLP-C assay.
|
| Discussion |
|---|
|
|
|---|
-4-cholesten-3-one) and reduced substrate (cholesterol) was estimated to be 10/min from the molar absorptivity. This CD method showed good linearity even at very low cholesterol concentrations, and the calibration curve passed through the origin of the coordinated axes. The detection limit of the CD method (10-4 mmol/L) was approximately one-tenth that of a PO method reported, for example, by Leary et al. (5) as 0.0013 mmol/L (0.08 mmol/L as serum RLP-C). The change in absorbance (sensitivity) of the CD method was at least 5 times higher than that of the PO method. If the sample volume for the cholesterol assay is increased from 10 µL, further improvement of sensitivity appears possible. The CD method meets the needs for an RLP-C assay. Serum RLP-C >0.194 mmol/L is a cardiovascular risk factor (4)(5). When serum samples with RLP-C of 0.194 mmol/L were diluted 61-fold with the immunoaffinity gel solution for the separation of RLP, the final cholesterol concentration in the supernatants was <0.003 mmol/L (0.194 divided by 61). The RLP-C assay with the CD method required only one calibrator, had suitable precision and linearity, and was not affected by the potential interferences tested.
In conclusion, the CD method is well suited for accurate RLP-C assays and is expected to provide an easy method for the measurement of very low concentrations of cholesterol. The CD method as a highly sensitive cholesterol assay method also appears suitable for measurement of very low cholesterol in other biological materials.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
H. R. Superko Advanced Lipoprotein Testing and Subfractionation Are Clinically Useful Circulation, May 5, 2009; 119(17): 2383 - 2395. [Full Text] [PDF] |
||||
![]() |
X.-M. Liang, H. Otani, Q. Zhou, Y. Tone, R. Fujii, M. Mune, S. Yukawa, and T. Akizawa Renal protective effects of pitavastatin on spontaneously hypercholesterolaemic Imai Rats Nephrol. Dial. Transplant., August 1, 2007; 22(8): 2156 - 2164. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Usui, H. Kakuuchi, M. Okamoto, Y. Mizukami, and M. Okazaki Differential Reactivity of Two Homogeneous LDL-Cholesterol Methods to LDL and VLDL Subfractions, as Demonstrated by Ultracentrifugation and HPLC Clin. Chem., November 1, 2002; 48(11): 1946 - 1954. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |