Clinical Chemistry 48: 1096-1097, 2002;
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kazmierczak, S. C.
Right arrow Articles by Gourley, G. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kazmierczak, S. C.
Right arrow Articles by Gourley, G. R.
Related Collections
Right arrow General Clinical Chemistry
(Clinical Chemistry. 2002;48:1096-1097.)
© 2002 American Association for Clinical Chemistry, Inc.


Technical Briefs

Direct Spectrophotometric Method for Measurement of Bilirubin in Newborns: Comparison with HPLC and an Automated Diazo Method

Steven C. Kazmierczak1a, Alex F. Robertson2, Paul G. Catrou3, Kimberly P. Briley3, Bill L. Kreamer4 and Glenn R. Gourley4

1 Department of Pathology, Oregon Health and Sciences University, 3181 SW Sam Jackson Park Rd., Portland, OR 97201; Departments of
2 Pediatrics and
3 Pathology and Laboratory Medicine, Brody School of Medicine of East Carolina University, Greenville, NC 27858;
4 Department of Pediatrics and Waisman Center, University of Wisconsin, 1500 Highland Ave., Madison, WI 53705

aaddress correspondence to this author at: Department of Pathology, Oregon Health and Sciences University, Mailcode L471, 3181 SW Sam Jackson Park Rd., Portland, OR 97201; fax 503-494-8148, e-mail kazmierc{at}ohsu.edu

The measurement of serum total bilirubin is one of the most frequently performed tests in newborns (1)(2)(3). Direct spectrophotometry for the measurement of bilirubin in sera from newborns is a simple and rapid method that requires minimal sample for analysis (4). The direct spectrophotometric assay is based on the absorbance of bilirubin at 454 nm; by contrast, hemoglobin absorbs equally at both 454 and 528 nm. Subtraction of the absorbance at 528 nm from that at 454 nm eliminates the effect of hemolysis and yields a value that can be attributed primarily to bilirubin. Unfortunately, other pigments, such as carotenoids, also absorb at 454 nm, thus limiting this technique to neonates <2–3 weeks of age.

Direct spectrophotometric methods have been compared with diazo (Jendrassik–Gróf-based) bilirubin methods (5)(6)(7)(8)(9), but sample hemolysis can increase or decrease the results obtained by the latter procedure, depending on the concentrations of hemoglobin and bilirubin in the sample (10)(11). Most such studies either do not describe the effect of sample hemolysis on the agreement between methods or report poorer agreement between methods or increased scatter around the regression line (6)(8)(9).

HPLC for measurement of serum bilirubin concentrations is labor-intensive and not practical for routine use; however, it is not subject to interference from hemoglobin or lipemia. We used HPLC to investigate a direct spectrophotometric procedure for the measurement of bilirubin in the plasma of neonates and measured the hemoglobin concentrations in all samples to assess the impact of hemolysis.

This study was performed after approval by the Institutional Review Boards of both participating hospitals. Informed consent was obtained from mothers before collection of blood from newborns. Thirty-seven blood samples were obtained from neonates <4 days of age (mean age, 48 h; range, 12–86 h). The samples were collected into heparin-containing, amber-colored containers after heel puncture with a capillary puncture device (Tenderfoot®; International Technidyne Corp.). No specimens were collected from infants undergoing phototherapy. The samples were immediately sent to the hospital laboratory, where the plasma was separated and then divided into two aliquots. One aliquot was immediately assayed for bilirubin for clinical purposes by a diazo method in an AU640 analyzer (Olympus). The second aliquot was placed in an amber-colored container and frozen at -70 °C for subsequent analysis of hemoglobin and bilirubin by HPLC and bilirubin by a direct spectrophotometric method.

We measured bilirubin concentrations by HPLC according to a previously described method (12). Hemoglobin and bilirubin concentrations were measured simultaneously by HPLC, except that the diode-array wavelength of detection was 404 nm (4 nm bandwidth) with a reference of 464 nm (4 nm bandwidth). Hemoglobin solutions (Sigma-Aldrich) were used to construct an appropriate calibration curve. While the HPLC analysis was being performed, aliquots of these same samples were used to measure bilirubin concentrations by a direct spectrophotometric method with a bilirubinometer (LEICA UNISTAT®; Leica Inc.). The Leica instrument measures the bilirubin concentration in 20 µL of serum without dilution. Absorbance is measured at 460 and 550 nm and is used to calculate the total bilirubin concentration in the sample. A weekly calibration was performed with the use of an assayed glass calibration cuvette. Total imprecision (CV) according to the manufacturer is ~15% and 10% at bilirubin concentrations of 21 µmol/L (1.2 mg/dL) and 130 µmol/L (7.6 mg/dL), respectively, and the reportable range is 0.0–684.0 µmol/L (0.0–40.0 mg/dL). The time required for bilirubin analysis with this instrument is ~15 s.

The median bilirubin concentration measured in the 37 specimens by HPLC was 197 µmol/L (11.5 mg/dL) [range, 46.2–295.8 µmol/L (2.7–17.3 mg/dL)]. A regression analysis performed according to the Deming method (13) between the HPLC (x) and Olympus analyzer (y) results yielded a slope (SD) of 0.93 (0.07) and a y-intercept (SD) of 13.7 (13.7) µmol/L (r = 0.89). The Deming regression analysis for the HPLC (x) vs Leica bilirubinometer (y) results yielded a slope (SD) of 1.06 (0.03) and a y-intercept (SD) of -9.4 (5.3) µmol/L (r = 0.99). The scatter around the regression line was much less for the Leica bilirubinometer vs the HPLC results (Sy|x = 5.4 µmol/L) compared with the Olympus analyzer vs HPLC results (Sy|x = 16.3 µmol/L). The mean squares of the residuals were significantly different (F = 8.9; P <0.001).

Bland–Altman (14) difference plots (Fig. 1 ) revealed little dependence of the disagreement between methods on bilirubin concentration. However, the plots readily showed greater disagreement between bilirubin concentrations measured by HPLC and the Olympus analyzer compared with the HPLC vs Leica bilirubinometer methods. The maximum differences between the HPLC and the Olympus analyzer results varied from -42.8 µmol/L (-2.5 mg/dL) to 59.9 µmol/L (3.5 mg/dL), with a mean (SD) difference of -1.2 µmol/L (22.3 µmol/L). The maximum differences between the HPLC and the Leica bilirubinometer results varied from -24.0 µmol/L (-1.4 mg/dL) to 15.4 µmol/L (0.9 mg/dL), with a mean (SD) difference of -1.7 µmol/L (8.3 µmol/L).



View larger version (20K):
[in this window]
[in a new window]
 
Figure 1. Bland–Altman plots of bilirubin results obtained by HPLC vs Olympus analyzer (A) and HPLC vs Leica bilirubinometer (B).

The solid line represents the mean difference in measured bilirubin concentrations between methods, and the dashed lines are ± 2 SD.

Virtually all samples in this study showed evidence of hemolysis. The median hemoglobin concentration was 35.7 µmol/L [230 mg/dL; range, 8.7–197.6 µmol/L (56–1275 mg/dL)]. Because of the complex nature of the mechanism of hemoglobin interference in bilirubin measurements, there was no direct relationship between hemoglobin concentrations and the magnitude and direction of the interference bias. A plot of the hemoglobin concentration in each sample vs the magnitude of the interference bias, calculated as HPLC-measured bilirubin minus Olympus analyzer- or Leica analyzer-measured bilirubin concentrations, revealed a slope and y-intercept of 0.001 and -7.5 µmol/L, respectively, for the Olympus analyzer procedure and 0.000 and -2.1 µmol/L, respectively, for the Leica bilirubinometer procedure. Neither slope was statistically significantly different from zero (P >0.99).

In conclusion, the measurement of total bilirubin with a Leica direct spectrophotometric procedure provides better agreement with HPLC than does a diazo (Jendrassik–Gróf-based) procedure in an Olympus analyzer, and the Leica method provides the advantages of small instrument size, small sample volume, and rapid turnaround. However, it should be noted that none of the specimens in this study was visibly lipemic, and we did not evaluate the effects of lipemia on the Leica and Olympus methods. In addition, all samples were obtained from newborns <4 days of age.


References

  1. Harish R, Sharma DB. Transcutaneous bilirubinometry in neonates: evaluation of Minolta Air shields jaundicemeter. Indian Pediatr 1998;35:264-267.[Medline] [Order article via Infotrieve]
  2. Maisels MJ, Kring E. Transcutaneous bilirubinometry decreases the need for serum bilirubin measurements and saves money. Pediatrics 1997;99:599-601.[Free Full Text]
  3. Donzelli G, Pratesi S. Transcutaneous bilirubinometry in healthy preterm newborns. Clin Biochem 2000;33:505-508.[ISI][Medline] [Order article via Infotrieve]
  4. . Leica Inc.. Leica Unistat bilirubinometer instruction manual 1997:23 Leica Buffalo, NY. .
  5. Hajzer S. Comparison of direct spectrophotometric determinations of bilirubin with candidate reference method in sera of newborns. J Clin Chem Clin Biochem 1989;27:445-449.[ISI][Medline] [Order article via Infotrieve]
  6. Vink KLJ, Schuurman W, van Gansewinkel R. Direct spectrophotometry of bilirubin in serum of the newborn, with use of caffeine reagent. Clin Chem 1988;34:67-70.[Abstract/Free Full Text]
  7. Harrison SP, Barlow IM. Three direct spectrophotometric methods for determination of total bilirubin in neonatal and adult serum, adapted to the Technicon RA-1000 analyzer. Clin Chem 1989;35:1980-1986.[Abstract/Free Full Text]
  8. Rolinski B, Kuster H, Ugele B, Horn K. Total bilirubin measurement by photometry on a blood gas analyzer: potential for use in neonatal testing at the point of care. Clin Chem 2001;47:1845-1847.[Free Full Text]
  9. Harkness RA, Lawrence CR, Renshaw A, Barr I, Brown SS, Rinsler MG. Assessment of the performance and clinical utility of a ward-side room bilirubinometer. Ann Clin Biochem 1983;20:149-152.[ISI][Medline] [Order article via Infotrieve]
  10. Kroll MH, Ruddel M, Blank DW, Elin RJ. A model for assessing interference. Clin Chem 1987;33:1121-1123.[Abstract/Free Full Text]
  11. Kroll MH, Chesler R. Rationale for using multiple regression analysis for complex interferences. Eur J Clin Chem Clin Biochem 1992;30:415-424.[ISI][Medline] [Order article via Infotrieve]
  12. Bhutani VK, Gourley GR, Adler S, Kreamer B, Dalin C, Johnson LH. Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia. Pediatrics 2000;106:E17.
  13. Cornbleet PJ, Gochman N. Incorrect least-squares regression coefficients in method comparison analysis. Clin Chem 1979;25:432-438.[Abstract/Free Full Text]
  14. Bland JW, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurements. Lancet 1986;1:307-310.[ISI][Medline] [Order article via Infotrieve]



The following articles in journals at HighWire Press have cited this article:


Home page
Ann Clin BiochemHome page
J. M Kirk
Neonatal jaundice: a critical review of the role and practice of bilirubin analysis
Ann Clin Biochem, September 1, 2008; 45(5): 452 - 462.
[Abstract] [Full Text] [PDF]


Home page
Arch. Dis. Child. Fetal Neonatal Ed.Home page
D De Luca, E Zecca, M Corsello, E Tiberi, C Semeraro, and C Romagnoli
Attempt to improve transcutaneous bilirubinometry: a double-blind study of Medick BiliMed versus Respironics BiliCheck
Arch. Dis. Child. Fetal Neonatal Ed., March 1, 2008; 93(2): F135 - F139.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kazmierczak, S. C.
Right arrow Articles by Gourley, G. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kazmierczak, S. C.
Right arrow Articles by Gourley, G. R.
Related Collections
Right arrow General Clinical Chemistry


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS