|
|
||||||||
Automation and Analytical Techniques |
1 Laboratoire de Biochimie A et de Neurobiologie and
2 Laboratoire dImmunologie et dHistocompatibilité, Hôpital St-Louis AP-HP, 1 avenue Claude Vellefaux, 75010 Paris, France.
3 Laboratoire de Biochimie Spécialisée, Hotel Dieu, 9 quai Moncousu, 44093 Nantes Cedex 1, France.
aAuthor for correspondence. Fax 33-1-4249-9247; e-mail thierry.le-bricon{at}sls.ap-hop-paris.fr.
| Abstract |
|---|
|
|
|---|
Methods: With the Capillarys ß1-ß2+® reagent set, proteins were separated at 7 kV for 4 min in 15.5 cm x 25 µm fused-silica capillaries (n = 8) at 35.5 °C in a pH 10 buffer with online detection at 200 nm. Serum samples with different electrophoretic patterns (n = 265) or potential interference (n = 69) were analyzed and compared with agarose gel electrophoresis (AGE; Hydrasys®-Hyrys®, Hydragel protein(e) 15/30® reagent set; Sebia).
Results: CVs were <3.5% for albumin, <11% for
1-globulin, <4.1% for
2-globulin, <7.4% for ß-globulin, and <5.8% for
-globulin (3 control levels); measured throughput was 60 samples/h. In patients without paraprotein (n = 116), the median differences between CE and AGE were -5.4 g/L for albumin, 4.0 g/L for
1-globulin, 0.7 g/L for
2-globulin, 0.6 g/L for ß-globulin (P <0.001 for all fractions), and -0.1 g/L for
-globulin (not significant). More samples had at least one
-migrating peak detected by CE (n = 135 vs 130; paraprotein detection limit,
0.50.7 g/L), but fewer were quantified (n = 84 vs 91) because of
- to ß-migration shifts. There was a 1.2 g/L median difference between CE and AGE for
-migrating paraprotein quantification (n = 69; P <0.001). Several ultraviolet-absorbing substances (lipid emulsion, hemoglobin) or molecules (contrast agent, gelatin-based plasma substitute) induced CE artifacts.
Conclusions:The Capillarys instrument is a reliable CE system for serum protein analysis, combining advantages of full automation (ease of use, bar-code identification, computer-assisted correction of
1-globulins) with high analytical performances and throughput.
| Introduction |
|---|
|
|
|---|
During the last decade, capillary electrophoresis (CE) has emerged as a powerful analytical tool (3)(4). High voltage applied in a thin (internal diameter <100 µm) fused-silica tube allows fast and efficient separation of ionized molecules or complexes (5). Protein analysis is usually performed in free solution with use of highly basic buffers to minimize interactions with capillary walls. Direct quantification is obtained in the low ultraviolet (UV) range (
200 nm via the peptide bonds) at the cathodic end of the capillary (6)(7). The first automated multicapillary instrument designed for routine serum protein analysis by CE, the Paragon CZE2000® (Beckman), was commercialized in 1994. Its advantages include high precision and resolution, slightly improved sensitivity for paraprotein detection (vs AGE), use of small amounts of sample and reagents, full automation, and high throughput (8)(9)(10)(11)(12)(13).
The aim of this study was to evaluate the performance of a recently launched automated multicapillary CE instrument (Capillarys®; Sebia) (14) for serum protein analysis. We compared the results with those obtained with our standard semiautomated AGE technique (Hydrasys®-Hyrys®; Sebia) (13).
| Materials and Methods |
|---|
|
|
|---|
1 nL). With the Capillarys ß1-ß2+® reagent set (Sebia), separation is obtained by applying a voltage of 7 kV for 4 min in eight fused-silica capillaries (total/effective length: 17.5/15.5 cm; i.d., 25 µm) in a pH 10 buffer at 35.5 °C controlled by Peltier effect. Online detection is at 200 nm in an optical cell (i.d., 100 µm) located at the cathodic end of the capillary and connected to the detector by eight optic fibers. In healthy individuals, six main protein fractions are detected in the following order:
-globulins, ß2-globulins, ßl-globulins,
2-globulins,
l-globulins, and albumin (Fig. 1
|
AGE of serum proteins was performed on a semiautomated Hydrasys system (Sebia) using the Hydragel protein(e) 15/30® reagent set (13). Briefly, after manual sample distribution (10 µL), separation is obtained in an 8 g/L agarose gel with the following automated steps: application for 30 s; separation in Tris-barbital buffer, pH 9.2, at a constant 20 W at 20 °C until 33 V-h have accumulated (
7 min), drying for 10 min at 65 °C, staining with amidoschwarz (4 g/L in an acidic solution), and destaining (0.5 g/L citric acid solution). Total analysis time is 45 min for 30 samples. Densitometric scanning of gels is performed at 570 nm (Hyrys densitometer; Sebia).
Quantitative results for each fraction of the CE or AGE electropherogram are expressed in g/L by multiplying the percentage of total area under the curve by total serum protein concentration.
analytical evaluation
Capillarys throughput was determined by analyzing 13 x 8 samples (all with bar codes) in a row. Carryover was assessed by alternating serum samples with distilled water (CDM Lavoisier; n = 30). The detection limit and precision studies were conducted according to the guidelines of the French Society of Clinical Biology (15), as follows: the detection limit was defined as 3 SD above the mean of 30 consecutive analyses of a blank (here distilled water), and for imprecision, CVs (within and between runs) were calculated based on 21 runs of at least two control specimens. A normal serum (both quantitatively and qualitatively), one with a low albumin concentration (26.9 g/L), and one containing a well-defined
-migrating paraprotein (27.7 g/L) were selected. Assay linearity was evaluated by gradually mixing 60 g/L solutions (in 9.0 g/L NaCl solution; Aguettant) of human albumin (Sigma) and
-globulins (Sigma) to obtain concentrations of 60, 50, 40, 30, 20, 10, and 5 g/L for each fraction (at a constant total protein concentration of 60 g/L). Sensitivity for paraprotein detection was checked by mixing samples containing either a cathodic
- (5.7 g/L) or anodic
-monoclonal (6.3 g/L) peak with a normal serum (15.0 g/L polyclonal
-globulins) down to the lowest peak concentration of 0.1 g/L; this was considered as the highest dilution giving a still discernible band.
We evaluated the effect of storage conditions, using a normal serum specimen kept at room temperature or 4 °C for 1, 8, 24, or 48 h and a lyophilized control (Sebia) stored for 1 week at 4 or 1 month at -20 °C.
specimens and interference study
A total of 432 blood samples obtained from hospitalized patients were grouped as follows. Group A (n = 265) contained carefully selected sera covering a wide range of electrophoretic patterns: within reference values (n = 25); single well-defined monoclonal immunoglobulin (n = 64; up to 93 g/L) or oligoclonal profiles/weak bands (n = 66); polyclonal hyper-
-globulinemia (1585 g/L; n = 30), hypo-
-globulinemia (0.56.0 g/L; n = 30), hypoalbuminemia (1030 g/L; n = 25), hyper-
l-globulinemia (>3 g/L) or hyper-
2-globulinemia (>9 g/L; n = 25). Group B (n = 98) contained unselected specimens from our daily routine that were analyzed blindly by AGE and CE. Group C (n = 69) contained samples with potential interference, including hemolysis (n = 19), lactescence (triglycerides >10 mmol/L; n = 7), icterus (total bilirubin >50 µmol/L; n = 20), increased C-reactive protein (CRP; >100 mg/L; n = 9), fibrinogen (sodium heparinate plasma; n = 5), and cryoglobulins (n = 9).
Blood was collected on a 7-mL tube without anticoagulant (Becton Dickinson) and centrifuged for 15 min at 1500g at 20 °C. Total serum protein was measured and AGE performed within 8 h before storage at -20 °C (groups A and C) or was analyzed immediately by CE (group B). Cryoglobulin samples were run on the Capillarys after 24 h of storage at 4 or 37 °C.
We evaluated the effects of UV-absorbing substances on the CE electropherogram by adding to 90 µL of a normal serum, 10 µL of a 70 g/L hemoglobin solution prepared from hemolysates (16), a sodium and meglumine ioxitalamate contrast agent (Telebrix®; Guerbet), or a lipid emulsion for parenteral nutrition (Ivelip® 10%; Baxter). We also evaluated the effect of a gelatin-based plasma substitute (Plasmion®; Roger Bellon) by mixing 50 µL of the solution with 50 µL of normal serum.
This study was conducted in accordance with the requirements of the Helsinki Declaration of 1975 as revised in 1996.
other biochemical markers
Total serum protein was determined by the Biuret reaction on a Modular® system (Roche). Bilirubin, triglyceride, and CRP concentrations were determined by standard biochemical methods using Roche reagent sets (Modular); hemoglobin was measured on a BN-100® nephelometer (Dade-Behring).
In case of discordant results between CE and AGE for group A patients, we performed serum immunofixation (IFE) using the Hydragel 4 IF® reagent set (Hydrasys; Sebia) with antibodies against IgG, IgA, IgM, and
and
light chains.
statistics
Statistical analysis was performed with Sigmastat®, Ver. 2.0 (Jandel Scientific). Results are expressed as the median and range or the mean and SD according to the population distribution (KolmogorovSmirnov test). Linear regression was used for analytical evaluation of CE linearity, and Deming regression analysis (17) was used for comparison of CE and AGE. Statistical significance was set at P <0.05 for all tests.
| Results |
|---|
|
|
|---|
-globulins (r = 0.98), with the detection limit (absorbance of 0.010) corresponding to
3.5 g/L of a mixture containing 5 parts albumin and 1 part
-globulin. The detection limit was 0.5 g/L for the anodic
-migrating paraprotein and 0.7 g/L for the more cathodic
-migrating paraprotein. Storage conditions had no significant impact (both quantitative and qualitative) on the Capillarys electropherogram (data not shown).
|
method comparison study
The results of the method comparison for protein fraction quantification in the subgroup of group A patients without paraprotein (n = 116) are presented in Table 2
. Differences between methods were proportional to the protein fraction concentration for albumin (r = 0.52) and
1-globulins (r = 0.34; linear regression, P <0.001); we observed no significant difference between methods for
-globulins (Table 2
). In the paraprotein subgroup, the following results were observed for
-migrating paraprotein quantification (n = 69): mean (SD) slope, 0.96 (0.04); mean (SD) intercept, 1.4 (0.2); r = 0.99; median difference, 1.2 g/L by CE (P <0.001). A BlandAltman plot of individual differences is provided in Fig. 2
.
|
|
The total number of samples with at least one detectable abnormal peak in the
2 to the
zone was 135 by CE and 130 by AGE (Fig. 3
). There was a 100% agreement in the subgroup of patients with a single well-defined paraprotein by AGE (n = 64). In others subgroups, additional peaks detected by CE corresponded to small or faint bands identified as monoclonal by IFE in 70% of cases. Fewer peaks could be quantified by CE than AGE because 15 of them (all quantified by AGE) were shifted from the
zone (by AGE) to the ß2 zone (identified as IgG in 58%, IgM in 34%, IgA in 8%;
in 58%, and
in 42%). Five patients with hypo-
-globulinemia by AGE had a small detectable peak by CE; only the three quantitatively most important peaks (0.7, 0.9, and 2.1 g/L) were detected and identified by IFE as oligoclonal, IgG
, and free
or IgD
or IgE
, respectively. In group B, results for
-globulins (CE/AGE; n = 98) were 62 and 71 normal, respectively; 17 and 15 with a single well-defined quantified peak; and 19 and 12, respectively, with a small peak or an oligoclonal profile.
|
in vivo (group c) and in vitro interferences
Hemolyzed serum samples had increased ß1-globulins and/or
2-globulins. Patients with a CRP <240 mg/L (n = 4) had no evident modification of the
zone of the electropherogram; a faint band in the anodic region of the
-globulins (close to fibrinogen) was observed in the two specimens with concentrations >300 mg/L. Icteric samples had no apparent modification of the CE electropherogram, whereas lipemic samples had increased
1-globulins with, in some cases, an additional peak close to albumin. After cryoglobulin storage at 4 °C, no peak was detected in the
zone of the Capillarys electropherogram in seven of nine samples. At 37 °C, a peak was detected in 100% of cases that corresponded to the measured cryoprecipitate (Table 3
). Free hemoglobin, Telebrix, and Ivelip were responsible for the appearance of additional peaks in the ß1 zone (hemoglobin), at the junction of the
1-/
2-globulin zones (Telebrix), or in the
1-globulin zone (Ivelip). Addition of Plasmion increased the
-globulin fraction in a polyclonal-like way, a phenomenon confirmed in vivo in a patient treated with a similar product (Gélofusine®; B Braun Médical). Figures showing the interference of selected compounds (lipids, cryoglobulins, contrast agent, plasma substitute, and fibrinogen) are available as a Data Supplement accompanying the online version of this article athttp://www.clinchem.org/content/vol49/issue11/.
|
| Discussion |
|---|
|
|
|---|
10% for sampling, dilution, and injection) to electrophoretic separation and quantification of protein fractions (CV, 1.27.1%), except for the flat, often large and heterogeneous
1-globulin zone (CV >10% with irreproducible integration). Similar results have been published with the Paragon CZE2000 (8)(9) and the earlier 4.41 version of the Capillarys (14). Despite good correlation with the Hydrasys-Hyrys for albumin and
-globulins (r = 0.99), there was an
200% positive difference vs AGE for
1-globulins and a 15% negative difference for albumin. Reports in the literature indicate similar results for the Paragon CZE2000 (
100% difference for
1-globulins, but lower difference for albumin) (8)(9) and the Capillarys (14), highlighting the need for an adjustment of the protein fraction reference interval (9)(18). Comparisons with immunonephelometry and colorimetry suggest that quantification of albumin and
1-globulin in the low-UV range is more accurate than protein staining (8)(9), especially for
1-acid glycoprotein because of its high sialic acid content. Overall, the throughput for the Capillarys (
60 samples/h) was better than for AGE (
30 samples/h with the Hydrasys-Hyrys) (13) and the Paragon CZE2000 (
40 samples/h) (8).
The detection limits for paraproteins in the Capillarys were in the 0.5 g/L range, similar to the detection limits reported for the Paragon CZE2000 (8) and for the semiautomated Hydrasys-Hysis AGE system (13). Importantly, there were no well-defined single paraproteins missed by CE, as reported for a few cases with the earlier 4.41 buffer and software version of the Capillarys (14) or the Paragon CZE2000 (19)(20). In oligoclonal, hypo- and hyper-
-globulinemic subgroups, the Capillarys displayed higher sensitivity than AGE for the detection of weak monoclonal bands (70% of positivity by IFE), as has been observed with the Paragon CZE2000 (20). In 30% of discordant results with AGE (difficult and/or complex profiles of the
-globulins), however, abnormalities detected by CE were not confirmed by IFE, the gold standard method. They might correspond to other anodic
-proteins, such as small amounts of fibrinogen, CRP, or degraded products of the C3 complement (14). Because degradation of the C3 fraction of the complement is a temperature-dependent phenomenon (14), serum storage at 4 °C for 48 h or at -20 °C for more prolonged periods is recommended. Concerning paraprotein quantification, there was no gain in the number of peaks quantified by CE because of
- to ß2-migration shifts observed in 11.5% of paraproteins (mostly IgM or IgG). Importantly, however, the median difference with AGE was limited to 1.2 g/L with only five outliers (maximum difference, -9.0 g/L, or -17%), an improvement over previously published CE methods (9)(19).
Hemolysis, icterus, and turbidity (and associations of these) represent the most frequently encountered serum interferences in clinical chemistry. The electrophoretic mobility of hemoglobin in the Capillarys was similar to that reported with the Hydrasys-Hyrys (21), the Paragon CZE2000 (8)(9), and the earlier, 4.41 version of the Capillarys (14):
2-globulins (haptoglobin-hemoglobin complex) and ß-globulins for free hemoglobin. Icterus had no effect on the CE electropherogram (up to 300 µmol/L total bilirubin), but an increase in the baseline or, sometimes, a well-defined VLDL peak appeared between albumin and other
1-globulins proteins in hyperlipemic samples. An optional computer-assisted correction is available in the present 4.51 version, which deletes the area under the curve between albumin and
1-globulins. Use of this correction improved the automated integration of
1-globulins (
38% improvement of the analytical precision) and the correlation between CE and AGE [r = 0.90; slope = 1.29 (0.08); intercept = 1.2 (0.2)] and reduced the difference with AGE to +1.8 g/L (Fig. 4
). After adjustment of reference values for
1-globulins, this correction could allow an easier screening of
1-antitrypsin deficiencies by CE (22). A computer-supported algorithm has also been developed by Beckman for the Paragon CZE2000 to recognize deficiency variants of
1-antitrypsin (23). Fibrinogen (heparinized samples or incompletely clotted serum) migrates in the anodic part of the
-globulins, close to C3. The high sensitivity of CE might also lead to the detection of endogenous proteins present in low concentrations (undetected by AGE), as observed in two specimens with a CRP >300 mg/L (faint band in the anodic
-globulin zone). The clinical incidence of this phenomenon might be limited because, in our experience, only
0.5% of hospitalized patients have a serum CRP >300 mg/L, and it rarely exceeds 500 mg/L. Interestingly, controlling the temperature of the capillary at 35.5 °C also prevented cryoglobulins from precipitating during electrophoretic migration, thus allowing their detection and accurate quantification, an advantage over AGE.
|
Overcoming interference by exogenous UV-absorbing substances is a new challenge in CE, not present in standard gel-based methods with indirect quantification by dye binding. As observed previously (14)(24), an iodinated radioopaque agent (Telebrix) was detected in the form of a
1-/
2-monoclonal-like peak, leading to a large converse error for all other fractions. During intravenous infusion of lipid emulsions (parenteral nutrition), the Capillarys algorithm should efficiently correct the
1-globulin zone. We also established that gelatin-based plasma substitutes falsely increase
-globulins in a polyclonal-like way. Visual inspection of serum specimens and knowledge of patient treatments are, therefore, of primary importance to avoid gross quantitative errors and misleading interpretations. Centrifugation (high speed or ultracentrifugation) for lipids or dialysis for small molecules might also help reduce interference. Considering the short half-lives of these drugs (e.g., 4 h for Gélofusine), performing CE of serum proteins at a distance from drug administration might be the safest alternative.
In conclusion, the Capillarys is an automated multicapillary zone electrophoresis instrument designed for human serum protein analysis. This high-throughput analytical system can save a substantial amount of technical time compared with AGE methods. Although not currently available (but available on the Paragon CZE2000), the most interesting perspective of the Capillarys is the immunologic identification of monoclonal proteins (25)(26)(27). Analysis of low-protein biological fluids (e.g., cerebrospinal fluid and urine) remains difficult (28)(29) because of the relative low sensitivity of UV detection and the presence of UV-absorbing interfering substances (especially in urine). More specialized analyses, such as for hemoglobin A1C (30) or transferrin sialoforms (31), have been successfully developed on research monocapillary electrophoresis systems, but are not currently available on multicapillary electrophoresis systems.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
/
quantitation. Clin Chem 1988;34:2196-2204.
1-Antitrypsin deficiencies masked by a clinical capillary electrophoresis system (CZE 2000). Clin Biochem 2000;33:79-80.[CrossRef][ISI][Medline]
[Order article via Infotrieve]
The following articles in journals at HighWire Press have cited this article:
![]() |
X. Bossuyt and G. Marien Detection of Monoclonal Proteins by Capillary Zone Electrophoresis: Comparison of 2 Multichannel Automated Systems Clin. Chem., January 1, 2007; 53(1): 152 - 153. [Full Text] [PDF] |
||||
![]() |
P. Vermeersch, G. Marien, and X. Bossuyt A Case of Pseudoparaproteinemia on Capillary Zone Electrophoresis Caused by Geloplasma Clin. Chem., December 1, 2006; 52(12): 2309 - 2311. [Full Text] [PDF] |
||||
![]() |
P. Luraschi, I. Infusino, I. Zorzoli, G. Merlini, C. Fundaro and, and C. Franzini Heavy Chain Disease Can Be Detected by Capillary Zone Electrophoresis Clin. Chem., January 1, 2005; 51(1): 247 - 249. [Full Text] [PDF] |
||||
![]() |
K. Gijbels, J. De Coster, and X. Bossuyt Interference by Gelatin-Based Plasma Substitutes in Capillary Zone Electrophoresis Clin. Chem., August 1, 2004; 50(8): 1473 - 1475. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |