Technical Briefs |
1 ARUP Institute for Clinical and Experimental Pathology, ARUP Laboratories Inc, 500 Chipeta Way, Salt Lake City, UT 84108
2 Department of Pathology, University of Utah Health Sciences Center, Salt Lake City, UT 84108
aauthor for correspondence: fax 801-584-5207, e-mail juenkejm{at}aruplab.com
Gabapentin is a novel anticonvulsant drug that was introduced in the early 1990s and later approved (1995) for use in the US as an adjunctive treatment of partial seizures with or without secondary generalization in persons >3 years of age. Although structurally similar to
-aminobutyric acid (GABA), gabapentin does not interact with GABA receptors, nor is it converted to GABA or a GABA agonist (1). Gabapentin is widely studied therapeutically. Its initial and approved use as adjunctive epileptic therapy has been broadened, with many additional indications. These include treatment for neuropathic pain after spinal cord injury (2)(3)(4), posttraumatic stress disorder (5), poststroke pain syndrome (6)(7)(8), alcohol withdrawal(9), migraine therapy (10), hot flashes associated with prostate cancer treatment (11), and postoperative pain after cancer surgery (12)(13).
The general mechanism by which gabapentin exerts its anticonvulsant action is unknown. It is not appreciably metabolized in the liver, nor does it induce liver enzymes. It circulates relatively unbound in serum, with a protein bound fraction of
3%. It has a volume of distribution of
58 L. Because gabapentin does not bind to protein, it can be removed by hemodialysis if medically necessary. Gabapentin is renally eliminated with an elimination half-life of
6 h and clearance proportional to creatinine clearance. Impaired renal function substantially decreases the clearance of gabapentin (14). Gabapentin exhibits saturable absorption, making it a nonlinear drug and kinetically less predictable. A doseresponse pattern is apparent for plasma gabapentin concentrations and for clinical effects within the dosage range 600-1800 mg/day. Seizure control has not been seen with trough plasma concentrations <2 mg/L. A majority of patients at suggested doses fall within a 210 mg/L range. The major side effects of the drug include somnolence, dizziness, ataxia, fatigue, and nystagmus. No serious or irreversible effects occur after overdose. Therefore, monitoring of trough plasma gabapentin concentrations is most useful to establish compliance. In treating chronic pain and addictions, higher trough concentrations of 1530 mg/L are maintained. Administration of gabapentin does not influence the pharmacokinetics of conventional anticonvulsant drugs, nor are the pharmacokinetics of gabapentin modified by the presence of other anticonvulsant drugs (1).
Gabapentin measurements have traditionally involved a derivatization step. Underivatized it exhibits essentially no ultraviolet absorbance. Several types of analysis have been performed, including spectrofluorimetry (15)(16)(17) and HPLC with a variety of detection methods, including mass spectrometry and capillary electrophoresis, but most frequently, fluorescent derivatization (18)(19)(20)(21)(22)(23). Gas chromatography (GC) has also been used with detection by mass spectrometry (MS) and flame ionization (24)(25). Here we describe a rapid and cost-effective isocratic liquid chromatography method that uses commonly available ultraviolet detection for gabapentin quantification without the need for solid-phase extraction.
The assay system included an automated HP1090 liquid chromatograph with autosampler, diode array detector, and a Perkin-Elmer Nelson 1022TM Integrator for result recording. The detector wavelength was set at 335 nm, the absorbance maximum for the 2,4,6-trinitrobenzene sulfonic acid (TNBSA; Pierce Scientific) derivative of gabapentin. The analytes were separated on a 50 x 3.0 mm (i.d.) LunaTM 5-µm phenyl-hexyl column (Phenomenex) with a 4 x 3.0 mm (i.d.) phenyl guard cartridge (Phenomenex) as precolumn. The flow rate was 1.0 mL/min with an approximate backpressure of 5800 kPa at ambient temperature. The mobile phase consisted of acetonitriletype I waterconcentrated glacial acetic acid (500:500:1 by volume). HPLC-grade acetonitrile and acetic acid were purchased from Fisher Scientific. The mobile phase was filtered and degassed before use through a 0.45 µm nylon membrane under reduced pressure.
The internal standard (IS), 3-amino-2-naphtholic acid, was purchased from Aldrich, and gabapentin stock was obtained from Pfizer. The primary stock solutions (100 g/L) of gabapentin and IS were prepared in type I water and methanol, respectively.
Samples were prepared by transferring 0.5 mL of patient sample, positive control, negative control (blank), and calibrators into respective microcentrifuge tubes; 50 µL of IS solution was then added to each tube, followed by 0.7 mL of acetonitrile. The tubes were capped, vortex-mixed for
30 s, and centrifuged for 5 min at
7287g. The supernatant was transferred to a clean, screw-top tube, in which it was completely dried down under desiccated air at 5060 °C. To each tube, we added 0.5 mL of 0.1 mol/L sodium bicarbonate, pH 8.5 (Sigma), and 0.2 mL of freshly prepared derivatization solution. The derivatization solution contained TNBSA prepared daily in 0.1 mol/L sodium bicarbonate (400 µL TNBSA/mL of bicarbonate). The tubes were then tightly capped and incubated at 90 ± 10 °C for 20 min. After incubation, the tubes were centrifuged for 2 min at 1121g. The caps were then removed from the tubes, and the derivatization was stopped by the addition of 0.25 mL of 100 g/L sodium dodecyl sulfate (Sigma) and 0.2 mL of 1 mol/L HCl (Fisher). The solution was transferred to a glass autosampler vial and capped tightly; 4060 µL was then injected on the column by the autosampler. The ratio of the peak height of the analyte divided by the peak height of the IS was used to quantify the analytes of interest from the calibration curve.
We evaluated the linearity of the method by analyzing in-house C18-filtered blood bank plasma to which gabapentin was added at the following concentrations: 1, 2, 5, 10, 20, 30, 40, 50, and 60 mg/L. This experiment was performed three times over several weeks. Each calibrator was then analyzed in duplicate, and concentrations were calculated from a four-point calibration curve constructed with the 1, 10, 30, and 60 mg/L calibrators. On the basis of a criterion of an accuracy of 85115% of true value, the assay was linear to 60 mg/L. The mean CV was 4.8%.
Analytical performance at 1 mg/L included drug recoveries of 94101% and a CV of 3.5%. This concentration was determined as the lowest concentration tested that produced a peak twice as great as the baseline noise of a plasma blank. As a result, 1 mg/L was selected as the lower limit of quantification and as the low calibrator for the assay.
Shown in Fig. 1
are chromatograms obtained from a blank plasma (panel A), a representative patient (panel B), and the 10 mg/L calibrator (panel C). The peaks of interest, the IS at 2.473 min and gabapentin at 2.800 min, were followed by four constant derivatization peaks that eluted in the next 3.5 min for a total run time of 6.3 min. After further study (data not shown), the first peak was identified as GABA. The other peaks remain undetermined, but studies have concluded that these peaks are not attributable to the use of sodium dodecyl sulfate. A small side post peak was also seen with the IS when heparinized plasma was used, but it was found to be inconsequential in correlation. The peak was not present in EDTA plasma.
|
To check for specimen interference, we processed 25 deidentified patient samples. Two drugs that potentially will interfere at 335 nm are baclofen (Lioresal®) and tiagabine (Gabatril®), both of which coelute with gabapentin. Baclofen and tiagabine concentrations of 100 and 300 µg/L, respectively, produced a false gabapentin concentration of
2.6 mg/L (data not shown). Other drugs assayed at therapeutic concentrations that had no interference included carbamazepine and its epoxide and hydroxy metabolites, oxcarbazepine and it monohydroxylated metabolite, zonisamide, levetiracetam, phenytoin and its metabolites, felbamate, lamotrigine, clonazepam, phenobarbital, primidone, acetaminophen, salicylate, ibuprofen, amitriptyline, nortriptyline, desipramine, doxepin and nordoxepin, imipramine, valproic acid, topiramate, mephenytoin and Nirvanol, amiodarone and desethylamiodarone, methsuximide and normethsuximide, ethotoin, clozapine, and sertraline.
Carryover was studied by injecting a high calibrator (60 mg/L) followed by blank plasma over several runs, over several days. No carryover was observed for the 60 mg/L calibrator.
We compared 30 deidentified patient samples testing positive for gabapentin with another reference laboratory GC-MS assay. The linear regression equation for correlation, where y is the HPLC method, was: y = 1.05x + 0.84 mg/L. The 5% slope correlated to no significant difference in patient results and was deemed acceptable.
In conclusion, this report describes a robust assay for the measurement of gabapentin by HPLC with ultraviolet detection that uses TNBSA as a chromogenic derivatization agent. The mechanism of binding to primary amines allows for selectivity toward the analyte of choice. Many coadministered medications either do not have a primary amine site or structurally inhibit binding of the chromophore at the site. The product is stable for at least 48 h at room temperature, whereas derivatives from other methods used are stable for 412 h. Use of this chromophore in conjunction with a phenyl-hexyl column in place of the common C18 column enhances selectivity for gabapentin. The phenyl-hexyl column uses hexyl alkyl groups instead of traditional propyl chains, giving it added stability in an acidic mobile phase and high selectivity for aromatic, amine, and polar compounds. Column lifetime generally supercedes 2000 injections, with the precolumn being replaced every other day or
300 injections and the column washed with acetonitrile during daily instrument maintenance. The extraction process avoids the use of solid-phase extraction, which substantially decreases set-up time. Cost savings are seen in the savings from solid-phase extraction columns, less expensive derivatization reagents, and technologist time compared with our previous GC-MS assay (24). Each step in the extraction causes a change in color, easily identifying each step in the procedure, and allows for quick extraction troubleshooting. Use of the HPLC method reduced our sample reporting time by one-half from the previous GC-MS assay. Although this chromatographic assay is limited by the two extremely rare interferences, it has been a reliable alternative to current methods that could easily be placed in production in a small laboratory.