Technical Briefs |
1 Department of Laboratory Medicine, Childrens Hospital, Harvard Medical School, Boston MA;2 Department of Mathematics, Massachusetts Institute of Technology, Cambridge, MA;
aaddress correspondence to this author at: Department of Laboratory Medicine, Childrens Hospital, Harvard Medical School, 300 Longwood Ave., Boston MA 02155; fax 617-730-0383, e-mail mark.kellogg{at}childrens.harvard.edu
Tobacco use, primarily through cigarette smoking, continues to be a primary cause of preventable death worldwide. It has been well documented that exposure to tobacco smoke, which is very common, causes chronic lung and heart disease. Data from the 19881991 National Health and Nutrition Examination Survey data found that 87.9% of nonsmokers had detectable concentrations of serum cotinine (1). Because of the health risks associated with tobacco exposure, analysis of biomarkers of tobacco exposure has increased. Cotinine is the preferred serum biomarker for tobacco exposure (2)(3)(4)(5).
Nicotine, a natural product in tobacco, is rapidly absorbed through the lungs into the pulmonary venous circulation and then to the heart and other body organs. It has a biological half-life of only 1 h and is rapidly excreted in the urine. As such, its use to measure tobacco exposure is limited. Nicotine is metabolized primarily via oxidation of the
-carbon to cotinine and N-oxidation of the pyrrolidine ring (6)(7). Cotinine accounts for
90% of nicotine metabolites in serum and has a half-life of 1040 h (8)(9). This relatively longer half-life makes it suitable for assessing exposure to cigarette smoke.
Validated methods for cotinine analysis in passive smoke assessment generally require large sample volumes, which are unsuitable for pediatric populations. Available methods that use smaller sample volumes and/or less complex extractions do not provide adequate assay sensitivity, thus precluding their use in assessing passive cigarette smoke exposure (10). We describe the development of a sensitive single-step extraction and rapid method for serum cotinine based on ion spray tandem mass spectrometry (MS/MS).
Cotinine was purchased from Sigma Chemical Co., D,L-cotinine-methyl-d3 from Cambridge Isotope Laboratory, and HPLC-grade methanol, optima-grade ammonium hydroxide, and methylene chloride from Fisher Scientific. Mass spectrometry was conducted with a PE Sciex API 3000 triple quadrupole MS with a TurboIon Spray interface and Analyst software (Applied Biosystems). The liquid chromatography (LC) system consisted of a LC-10AT VP liquid chromatograph with a SIL-10AD autoinjector and a SCL-10A VP system controller (Shimadzu Scientific Institute, Inc.).
Stock solutions of cotinine and internal standard were each prepared to obtain a concentration of 100 mg/L in methanol. These solutions were found to be stable for at least 6 months when stored at 4 °C. The internal standard was prepared by dilution (1:2000) of the stock solution with 10 mL/L ammonia in methanol (1:99 by volume) to make a solution of 50 µg/L. This solution was stored in a flammable cabinet freezer and was stable up to 30 days. The stock calibrator solution was diluted (1:200) in methanol to obtain a working solution of 500 µg/L. This solution was then serially diluted to obtain the final calibrator concentrations of 200, 50, 20, 5, and 2 µg/L. Controls were prepared by adding stock solutions containing 500, 200, and 50 mg/L cotinine to pooled sera collected from smokers and nonsmokers at a volume ratio of 1:19. These solutions were then further diluted (1:100) with pooled serum, and the final concentrations of 268, 124, and 33 µg/L were determined by standard addition. Calibrator solutions and control aliquots were stored at 70 °C and were found to be stable up to 6 months. Monthly preparations of new solutions, which were run in parallel with existing solutions, were used to assess stability. Deviations >5% were considered significant. The mobile phase consisted of 100% methanol. Cotinine-free serum for use in calibration was created by pooling sera from individuals with no known passive smoke exposure for at least 5 days and found to have serum cotinine <2 µg/L as measured by an independent laboratory. New serum pools were tested against existing calibrators to verify that cotinine concentrations were less than the method detection limit of 0.6 µg/L.
A working diluent/internal standard solution was prepared by combining 800 µL of internal standard, 5.2 mL of methanol, 16 µL of ammonium hydroxide, and 6 mL of distilled water. We mixed 50 µL of sample, control, or calibrator with 200 µL of the working diluent/internal standard and 400 µL of methylene chloride. We then added an additional 50 µL of methanol to control and patient samples, and 50 µL of cotinine-free serum to the calibrator. All samples were vortex-mixed for 10 s and centrifuged at 16 000g for 1 min. We transferred 300 µL of the lower methylene chloride layer to a clean glass vial. A 30-µL injection was made into the tandem mass spectrometer, equipped turbo spray ion interface, at a flow rate of 0.2 mL/min; the total analysis time was 3 min. The tandem mass spectrometer was used in the positive-ion mode. Nitrogen was used as the drying gas at a flow rate of 8 L/min and for collision-activated dissociation. The ion source temperature was 450 °C, the declustering potential was 41 V, and the collision energy was 33 eV. Cotinine was detected in the multiple-reaction-monitoring mode at a m/z 177/80 and 177/98 transitions. Analyst software, Ver. 1.2 (Applied Biosystems/MDS SCIEX), was used for system control, data acquisition, and data processing.
The assay was linear to a cotinine concentration of 500 µg/L. The lower limit of detection was 0.54 µg/L, based on +2 SD from the mean of 20 replicates of drug-free sera (0.008 ± 0.271 µg/L) and 20 replicates of the 2 µg/L calibrator (2.02 ± 0.19 µg/L). Data were evaluated with the EP Evaluator Online statistics program (David G. Rhoads Associates, Inc.). Functional sensitivity (interassay CV <20%) of the method is 1.4 µg/L. Between-run imprecision (CV) at cotinine concentrations of 2, 5, 20, 33, 50, 124, and 268 µg/L was 11%, 9.5%, 6.1%, 7.2%, 4.0%, 5.6%, and 2.6% (n = 20), respectively. Recovery in serum samples to which cotinine had been added at concentrations of 2300 µg/L ranged from 92% to 103%. To further assess the accuracy of the method, we assayed 40 samples previously analyzed by LC-MS/MS at Mayo Clinic Laboratories (Rochester, MN and Wilmington, MA) (11). The results are summarized in Fig. 1
.
|
To assess interference, we assayed combinations of 51 different drugs and ascorbic acid. We mixed 20 µL of the drug cocktail with 50 µL of serum with a known cotinine concentration of 20 µg/L. Final concentrations of the tested substances were
1 mg/L. Samples were then extracted and tested as described above. None of the substances produced an interference >5% compared with the serum diluted with drug cocktail diluent. The list of drugs tested appears in Supplement 1, which accompanies the online version of this Technical Brief at http://www.clinchem.org/content/vol50/issue11/.
Ion suppression was assessed as recommended by Annesley (12). Briefly, the response of calibrators injected into the mobile phase was compared with the same amount of cotinine added to unextracted and extracted samples. To assure that slight differences in matrix were not causing suppression, we used 20 different cotinine-free serum samples. We observed a small decrease in signal, but because the internal standard and cotinine signals coincided, the ion suppression present was corrected when concentrations were calculated. Additionally, cotinine (10 µg/L) was infused at a constant rate into the mass spectrometer, and extracts from 20 different cotinine-free serum samples were injected. Signal decreased by <10% when serum samples were injected.
The total time to complete this assay is substantially shorter than for other methods that require extensive extractions before assay (11)(13). The simple extraction into methylene chloride and a 3-min assay time combine to allow 100 samples to be analyzed in <8 h. Additionally, the 50-µL sample size makes this an ideal assay for use in large research studies or epidemiologic surveys where multiple analytes and limited sample volumes are common.
The described assay is simple, sensitive, precise, and requires only a small serum sample volume. Without the need for a HPLC column, analysis time is rapid (3 min), and is sensitivity suitable for the qualitative detection of recent passive nicotine exposure. We conclude that this method is ideally suited for analysis of cotinine in passive and active smokers for both clinical and research purposes.