| Abstract |
|---|
|
|
|---|
Methods: We used real-time PCR with fluorescence resonance energy transfer technology to simultaneously detect and differentiate VAR, MPV, CPV/VAC, and CML. An internal control generated by cloning and mutating the PCR target gene facilitated monitoring of PCR inhibition in each individual test reaction.
Results: Strain differentiation results showed little interassay variability (CV, 0.40.6%), and the test was 100-fold more sensitive than virus culture on Vero cells. Low copy numbers of DNA could be detected with
95% probability (235849 genome copies/mL of plasma).
Conclusions: The real-time PCR assay can detect and differentiate human pathogenic orthopox viruses. The use of an internal control qualifies the assay for high sample throughput, as is likely to be needed in situations of suspected acts of biological terrorism, e.g., use of VAR.
| Introduction |
|---|
|
|
|---|
Smallpox was eradicated in 1979 but regained attention recently because of the potential use of its causative virus, VAR, as a bioterrorism agent. The virus is highly contagious, environmentally stable, and has fatality rates of 2050% (2). Human monkeypox is a similar but distinct disease that until recently occurred only in Central/West Africa. Person-to-person transmission of its causative agent, MPV, is less efficient, and the fatality rate is estimated to be 110% (3). By mid-2003, MPV had been introduced in North America as the result of importation of rodents (4). VAC was used worldwide as a vaccine to eradicate smallpox. Adverse events such as vaccinia necrosum or postvaccination encephalitis are rare, except in immunosuppressed persons (5).
CPV is endemic in Europe and Asia. Infection in humans is self-limiting and localized. Severe courses have been described in immunosuppressed patients (6).
CML can cause systemic and economically important disease in camels. It is still poorly understood, and infection in humans is rare.
In light of the serious public health concern, reliable tools for the detection and differentiation of orthopox viruses are needed. Once an orthopox virus is detected, it is important that the other human pathogenic species be reliably differentiated from VAR. Orthopox viruses are traditionally detected by electron microscopy, which does not allow differentiation among relevant species. Differentiation can be achieved by analyzing the morphology of pocks on infected chicken egg chorioallantoic membranes or the ceiling growth temperature of isolates. Serologic differentiation is also possible, as well as DNA endonuclease cleavage (7). All of these procedures are lengthy and require experienced handling in highly specialized laboratories. PCR allows primarily detection and simultaneous genotyping of orthopox viruses (8). However, a major drawback of conventional PCR assays is the laborious and time-consuming post-PCR processing. This problem can be overcome by real-time PCR technology, which allows detection of PCR products during the reaction by use of fluorescent DNA probes. With fluorescence resonance energy transfer (FRET) probes (9)(10), real-time PCR is also capable of differentiating strains (11). Recently, real-time assays for orthopox viruses have been described that target the hemagglutinin (HA) gene and differentiate VAR from the other orthopox viruses (12)(13).
In this study we present an assay that uses FRET real-time technology for the detection, differentiation, and quantification of VAR, MPV, CPV/VAC, and CML in a single PCR run. The assay contains an internal inhibition control to detect the inhibitory effect that some clinical samples have on PCR, making it suitable for high-throughput application when large numbers of samples have to be tested.
| Material and Methods |
|---|
|
|
|---|
proficiency study sample panel
A panel of 15 lyophilized plasma samples to which gamma-irradiated, heat-inactivated orthopox viruses (MPV, CPV, VAC, CML, and ectromelia virus) had been added was provided by the European Network for the Diagnosis of Imported Viral Diseases (ENIVD; Berlin, Germany). Viral concentration in the samples after resuspension in water ranged from 5 x 107 to 5 x 101 genome copies/mL.
plaque assay
MPV was grown for 3 days on subconfluent Vero cells overlaid with 10 g/L methylcellulose in medium solution in 24-well cell culture dishes. The overlay was removed, and cells were stained with methylene blue to visualize plaques.
nucleic acid extraction
For DNA extraction, we used a QIAamp Viral RNA Mini Kit (specified by the manufacturer, Qiagen, as also for use on DNA viruses) according to the manufacturers instructions.
plasmid genotype standards
We ligated a 275-bp insert of the HA gene of MPV, CML, CPV, and VAC into plasmid vectors and cloned it in Escherichia coli by means of a pCR 2.1-TOPO TA cloning reagent set (Invitrogen). Plasmids were purified with a Nucleo Spin Plasmid reagent set (Macherey Nagel) and sequenced with BigDye terminator cycle sequencing chemistry (Applied Biosystems) on an automatic ABI 377 DNA sequencer (Applied Biosystems). Plasmids were quantified spectrophotometrically. To obtain a genotype standard for VAR, we introduced the required point mutation into the MPV standard plasmid by means of a QuickChange site-directed mutagenesis reagent set (Stratagene). The recommendations of the manufacturer were followed.
internal control plasmid construct
The plasmid genotype standard for VAC was used to construct an internal control plasmid. Gene splicing by overlap extension was carried out with primers IcS (5'-atcgttcgttgagcgattagcagtattttgataattataccaaggataaaatatc-3') and IcAs (5'-ctgctaatcgctcaacgaacgataaagaataatggaattgggctccttataccaag-3') to introduce an alternative probe-binding site at nucleotide positions 178221. The resulting constructs of the correct length were cloned by means of a pCR 2.1-TOPO TA cloning reagent set (Invitrogen) and processed as described above.
oligonucleotide design
Sequence alignments were generated with default settings using the BLAST algorithm as provided by the National Center for Biotechnology Information (www.ncbi.nlm.nih.gov). GenBank accession number X65516 (VAR) served as the query sequence. Primers were chosen to facilitate amplification of all sequences in the alignment without base mismatches. Dual FRET probes were chosen within the 275-bp amplicon to allow differentiation of VAR, MPV, CML, and VAC/CPV based on sequence polymorphisms at the site of hybridization of the downstream probe (Fig. 1
). The probe was homologous for VAC/CPV and contained one mismatch for MPV, two for VAR, and three for CML.
|
conventional pcr
A PCR protocol for amplification of the gene coding for the 32-kDa adsorption protein via open reading frame D8L of all orthopox viruses was adapted previously and optimized in our laboratory. The reaction volume of 25 µL contained 1x PCR buffer II, 1.25 U of AmpliTaq Gold DNA polymerase (Applied Biosystems), 3 mM MgCl2, 200 µM each of the deoxynucleotide triphosphates, 400 nM primer D8L forward [5'-atgccgcaacaactatctcct-3' (14)], and 200 nM primer D8L reverse [5'-ctagttttgtttttctcgcgaa-3' (14)]. Thermal cycling included 95 °C for 15 min, followed by 45 cycles of 95 °C for 20 s, 56 °C for 10 s, and 2 °C for 60 s. Products were visualized on agarose gels.
real-time pcr
Each 30-µL reaction contained 5 µL of DNA, 3.3 mM MgCl2, 1x Platinum Taq polymerase reaction buffer (Invitrogen), 40 ng/µL bovine serum albumin (Sigma), 200 µM each of the deoxynucleotide triphosphates, 1 µM primer OPS3 (5'-tacttttgttactaatatcattag-3'), 1 µM primer OPAs4 (5'-agcagtcaatgatttaattgt-3'), 0.03 µM probe OPP1 (5'-gcttggtataaggagcccaattcXattattctt-3', where X is a ribose phosphate spacer nucleotide), 0.2 µM each of probes OPP2-640 (5'-tagctgctaaaagcgacgtct-3') and OPP2-705 (5'-atcgttcgttgagcgattagcag-3'), and 1 U of Platinum Taq DNA polymerase (Invitrogen). Probe OPP1 was 3'-labeled with fluorescein. Probes OPP2-640 and -705 were 3'-phosphorylated and 5'-labeled with the LCred640 and LCred705 dyes (Roche), respectively. Sealed LightCycler glass capillaries were used for amplification. Cycling conditions in a Roche LightCycler (Ver. 3.5) were as follows: 94 °C for 6 min and 40 cycles of 94 °C for 5 s, 51 °C for 10 s, and 72 °C for 15 s. Fluorescence was measured once per cycle at the end of the 51 °C segment.
A color compensation file was generated for all fluorescence measurements to compensate for the spectral overlap of the emission spectra between channels 2 and 3 of the LightCycler, as recommended by the manufacturer. The virus amplification signal was detected on fluorescence channel F2/Back-F1, and the internal control was detected on channel F3/Back-F1. Subsequent melting curve analysis involved an initial denaturation at 95 °C for 10 s, followed by an increase from 40 °C to 90 °C at a rate of 0.1 °C/s with continuous measurement of fluorescence. Melting points were calculated by the LightCycler operation software in a three-step procedure; the measured fluorescence signal was first plotted against the temperature to obtain a melting curve. The negative first derivative was then calculated to yield a maximum at the point of the strongest slope, termed the melting peak. Finally the maximum of a curve fit to the melting peak was recorded as the melting point.
statistical analysis
We examined different input concentrations of orthopox virus plasmid DNA to calculate the predicted proportion of positive results in replicate tests, using probit analysis as a nonlinear regression model. The Statgraphics plus (Ver. 5.0) software package was used for probit analysis.
| Results |
|---|
|
|
|---|
8 °C difference from CML and MPV).
|
|
Because our assay was intended for diagnostic use in a high-throughput setting, an internal control reaction was implemented that allowed in-tube detection of PCR inhibition without additional controls. A plasmid was constructed that contained the same primer binding sites as the virus but was detected by a heterologous downstream FRET probe (OPP2-705). The binding site for OPP2-705 had been introduced into the VAR plasmid by gene splicing and overlap extension, thereby removing the VAR probe binding site. After the concentration of the probe had been adjusted, 1 copy of the internal control plasmid could be amplified occasionally, and 20 copies were always detected without affecting the efficiency of amplification of the virus (data not shown).
To determine the limit of detection of our PCR compared with virus culture, we prepared serial 10-fold dilutions of a quantified MPV stock solution ranging from 15 000 to 0.015 plaque-forming units/mL in negative human plasma and subjected it to the plaque assay and real-time PCR. Dilution endpoints of 15 and 0.15 plaque-forming units/mL, respectively, were detectable. This corresponds to a ratio of DNA to infectious virus particles of
100:1. To determine the limit of detection in a statistically precise way, we added plasmid genotype standards in five different concentrations to human plasma, and plasma containing each concentration was amplified in five replicate reactions (5 x 5 = 25 reactions/genotype). Before extraction, the internal control plasmid was added to the lysis buffer at a concentration corresponding to 20 copies/PCR reaction. The observed proportions of positive results for each concentration for each genotype were subjected to probit regression analysis (Fig. 4
). The following virus genome-equivalents (geq)/mL of plasma were calculated to be detectable at a
95% chance: VAR, 849 geq/mL; MPV, 779 geq/mL; VAC, 843 geq/mL; and CML, 235 geq/mL. These detection limits correspond to 2.749.88 copies/PCR vial.
|
To mimic a real diagnostic situation, the novel assay was used in a blinded manner in a quality assurance study of the ENIVD. The sample panel contained different orthopox viruses added to human plasma. Virus concentrations ranged from 5 x 107 to 5 x 101 geq/mL, as indicated in data sheets provided by ENIVD. The real-time PCR detected virus in 14 of 15 virus-containing samples, whereas a conventional PCR assay gave only 11 of 15 correctly positive results. Results of strain differentiation in all samples were correct. The one sample that was not detected contained 3.6 x 106 copies/mL of ectromelia virus, a nonpathogenic orthopox virus for which our test had not been designed.
The specificity of our assay was confirmed by amplification of genomic DNA or RNA from 21 viral, bacterial, and fungal pathogens (Table 1
) that might be present in clinical samples submitted for orthopox virus testing (skin, blood, and organs). No amplification was observed in any of these materials.
|
| Discussion |
|---|
|
|
|---|
The HA gene chosen as the target sequence is highly specific for orthopox viruses: other pox viruses do not produce hemagglutinin (1). The suitability of this target gene has already been described (8)(12)(13). As in a previously published test intended only to separate VAR from other orthopox viruses (12), we chose FRET probes for detection of PCR products. Our assay differentiated VAR from other orthopox viruses by a particularly distinct FRET probe melting point, making it less susceptible to interference from sample impurities. This is also reflected in the small interassay CV for the melting points. As an improvement, our test also identifies the other human pathogenic orthopox viruses. The relevance of this feature has been illustrated by the recent introduction of a new orthopox virus (MPV) into North America (4).
The availability of plasmid genotype standards for all human pathogenic orthopox species makes it particularly easy to transfer our method to other laboratories. Moreover, in light of the new WHO safety regulations restricting work with variola virus DNA
500 bp in length (15), the availability of an MPV-derived, mutagenized plasmid provides a practical and compatible positive control for VAR. We have shown in this and an earlier study (11) that such cloned DNA fragments have the same properties as genomic DNA in FRET melting point analysis.
Once established, the sensitivity of a new PCR assay must be validated to qualify it for diagnostic use. Classic virologic quantification methods, e.g., plaque assays, can be used as reference methods, and we have shown that our assay is 100-fold more sensitive than virus culture. However, the disproportion between infectious particles and DNA copies in virus stock solutions may be problematic, making such results poorly comparable among laboratories. As an alternative approach we used quantified plasmid DNA to determine the analytical sensitivity in a statistically precise way, using probit analysis (18)(19). Human plasma was chosen as a mimic for clinical samples because patients with smallpox are viremic during the preeruptive stage of the disease, a window of opportunity to prevent the spread of the disease by identification of infected patients. Use of plasma also allows more precise definition of the experimental target concentrations than do solid samples, e.g., skin crusts. The sensitivities achieved by our assay are comparable to those of commercial US Food and Drug Administration-cleared PCR assays (20)(21). The good performance of our test in the international ENIVD quality-assurance study on PCR for orthopox viruses further underlines that our test is technically optimized.
In testing of patient samples, PCR inhibition may occur in a frequency of 0.342.1% of tests (20)(21). The use of inhibition controls is critical to prevent false-negative results (22). Use of an internal control that is copurified with the test samples and amplified with the same primers as the virus has been shown to be particularly efficient (17)(18)(23). It reduces test costs and laboratory effort by requiring no additional reactions. This in turn qualifies the test for high sample throughput, which might occur in case of a suspected event of biological terrorism.
In conclusion, the described assay can provide support to physicians who suspect on clinical grounds an infection with pathogenic orthopox viruses. Its high analytical sensitivity and inhibition control could qualify the test for exclusion diagnostics. The availability of plasmid standards and validation data make it an easily transferable method. However, we must emphasize that a positive laboratory diagnosis of an orthopox viral infection requires further confirmation by classic techniques such as electron microscopy, serology, or viral culture, given the enormous consequences for the public health sector that some of these diseases entail.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
S. Kramme, L. Van An, N. D. Khoa, L. Van Trin, E. Tannich, J. Rybniker, B. Fleischer, C. Drosten, and M. Panning Orientia tsutsugamushi Bacteremia and Cytokine Levels in Vietnamese Scrub Typhus Patients J. Clin. Microbiol., March 1, 2009; 47(3): 586 - 589. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kurth, J. Achenbach, L. Miller, I. M. Mackay, G. Pauli, and A. Nitsche Orthopoxvirus Detection in Environmental Specimens during Suspected Bioterror Attacks: Inhibitory Influences of Common Household Products Appl. Envir. Microbiol., January 1, 2008; 74(1): 32 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. G. Fedele, A. Negredo, F. Molero, M. P. Sanchez-Seco, and A. Tenorio Use of Internally Controlled Real-Time Genome Amplification for Detection of Variola Virus and Other Orthopoxviruses Infecting Humans J. Clin. Microbiol., December 1, 2006; 44(12): 4464 - 4470. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Savona and P. J. Danaher Vaccinia DNA in Blood After Smallpox Vaccination--Reply JAMA, September 20, 2006; 296(11): 1351 - 1352. [Full Text] [PDF] |
||||
![]() |
M. R. Savona, W. P. Dela Cruz, M. S. Jones, J. A. Thornton, D. Xia, T. L. Hadfield, and P. J. Danaher Detection of vaccinia DNA in the blood following smallpox vaccination. JAMA, April 26, 2006; 295(16): 1898 - 1900. [Full Text] [PDF] |
||||
![]() |
M. Niedrig, H. Meyer, M. Panning, and C. Drosten Follow-Up on Diagnostic Proficiency of Laboratories Equipped To Perform Orthopoxvirus Detection and Quantification by PCR: the Second International External Quality Assurance Study J. Clin. Microbiol., April 1, 2006; 44(4): 1283 - 1287. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Dreier, M. Stormer, and K. Kleesiek Use of Bacteriophage MS2 as an Internal Control in Viral Reverse Transcription-PCR Assays J. Clin. Microbiol., September 1, 2005; 43(9): 4551 - 4557. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. P. Fedorko, J. C. Preuss, G. A. Fahle, L. Li, S. H. Fischer, P. Hohman, and J. I. Cohen Comparison of Methods for Detection of Vaccinia Virus in Patient Specimens J. Clin. Microbiol., September 1, 2005; 43(9): 4602 - 4606. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |