Clinical Chemistry 50: 1068-1071, 2004; 10.1373/clinchem.2004.031575
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (19)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kristensen, G. B.B.
Right arrow Articles by Sandberg, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kristensen, G. B.B.
Right arrow Articles by Sandberg, S.
Related Collections
Right arrow Evidence Based Laboratory Medicine and Test Utilization
(Clinical Chemistry. 2004;50:1068-1071.)
© 2004 American Association for Clinical Chemistry, Inc.


Technical Briefs

Standardized Evaluation of Instruments for Self-Monitoring of Blood Glucose by Patients and a Technologist

Gunn B.B. Kristensen1,a, Kari Nerhus1, Geir Thue1 and Sverre Sandberg1,2

1 NOKLUS, Norwegian Center for Quality Improvement of Primary Care Laboratories, Division of General Practice, Department of Public Health and Primary Health Care, University of Bergen, Ulriksdal 8c, N-5009, Norway;2 Laboratory of Clinical Biochemistry, Haukeland University Hospital, Bergen, Norway

aauthor for correspondence: fax 47-55586710, e-mail gunn.kristensen{at}isf.uib.no

Self-monitoring of blood glucose (SMBG) is recommended to improve metabolic control for patients with diabetes because tight glycemic control can decrease microvascular complications in individuals with type 1 and type 2 diabetes (1)(2). The worldwide market for SMBG is $2.7 billion per year, with annual growth estimated to be 10–12% (3). The performance of these instruments is therefore an important issue.

In most studies evaluating instruments for SMBG, the patient is not involved, although user errors account for a large portion of the total error in SMBG (4). In International Organization for Standardization document ISO/FDIS 15197, it is therefore recommended that a user performance evaluation should be performed in addition to testing by a medical laboratory technologist (MLT) (5). We have developed a procedure based on these recommendations in which instruments are tested simultaneously by an experienced technologist and a group of patients under the same conditions, using blood from the same patients. The procedure was evaluated with two instruments: Glucometer Dex from Bayer (1999) and GlucoMen Glyco from Menarini (2002).

Norwegian patients with diabetes (type 1 and type 2) participated. For each meter, 100 patients were randomly divided into two groups, as shown in Supplement 1 of the Data Supplement that accompanies the online version of this Technical Brief at http://www.clinchem.org/content/vol50/issue6/. Before the evaluation, one group was given specific training by a MLT (1–2 h) on how to use the new meter, in addition to standard written instructions, whereas the other group received the meter by mail with the same standard instructions. Three lots of strips were used in the test. After receiving their new meter, the user manual, and 50 strips, the patients familiarized themselves with the meter for 2 weeks. In the period between 2 and 4 weeks, the patients performed five glucose measurements in duplicate on themselves on 5 different days. After 4 weeks, the patients met individually with the MLT for a consultation. During the consultation, the patients first performed two measurements on themselves with no instructions from the MLT. Within 5 min, the MLT took a sample from a different finger. With this blood sample, glucose measurements were performed twice on two different meters and twice with the reference method. Because the meters had no underfill detection, the MLT visually checked the strips and marked all measurements with strips not completely filled with blood. The MLT also observed whether the instrument was coded correctly. Finally, the patients completed two questionnaires, one about the user manual and one about the user-friendliness of the device. The study protocol was approved by the Norwegian Regional Committee for Medical Research Ethics.

The comparison method for the Glucometer Dex was a glucose dehydrogenase method using hemolyzed blood with reagents from Roche Diagnostics (Unimate 7) run on a Cobas Fara centrifugal analyzer (Roche Diagnostics). The comparison method for GlucoMen Glyco was a method that measures glucose in plasma on an Advia 1650 with reagents from Bayer (Glucose Hexokinase Method II; prod. no. B01-597-01). Both methods were verified by SRM 965 from the NIST and by control solutions verified by the isotope dilution–gas chromatography/mass spectrometry reference method.

Imprecision (SD and CV) was calculated by use of paired measurements, based on the formula:

where d is the difference between measurements, and n is the number of duplicate samples. The criterion promoted by Burnett (6) was used for detection of outliers.

According to ISO/FDIS 15197 (5) and the NCCLS (7), the number of measurements deviating more than ±20% (results >=4.2 mmol/L) and >0.83 mmol/L (results <4.2 mmol/L) should be <5%. None of the meters met these requirements when used by patients, but the GlucoMen Glyco met the requirements when used by the MLT (Fig. 1 and Table 1 ). However, several other, usually stricter quality specifications have been suggested (8)(9)(10)(11), e.g., the American Diabetes Association suggests a total error of ±10% (9).



View larger version (25K):
[in this window]
[in a new window]
 
Figure 1. Difference between Glucometer Dex or GlucoMen Glyco results (first measurement) and reference method results (mean of two measurements) plotted against the reference method results.

(A), patients (n = 93 and 90 for Glucometer Dex and GlucoMen Glyco respectively); (B), MLT (n = 95 and 92 for Glucometer Dex and GlucoMen Glyco, respectively). Measurements with strips filled incompletely with blood are indicated with closed symbols. Open symbols indicate measurements with different lots of test strips. Lines represent limits suggested from American Diabetes Association (inner lines) (9) and in ISO/FDIS 15197 (outer lines) (5).


View this table:
[in this window]
[in a new window]
 
Table 1. Analytical quality.

As can be seen in Table 1Up , the measurements performed by the patients had significantly poorer precision than measurements performed by the MLT (P <0.05). On the basis of patient-derived quality specifications and simulation studies, it has been suggested that imprecision should be <5% (9)(10). In a study where the analytical quality of five SMBG systems was investigated, the CV varied from 5% to 11% when patients used the instruments compared with 2.3–5.9% when a MLT used them (12), which is in line with our results.

One important issue of incorporating a user test in the evaluation of SMBG devices is to discover user errors that will not be detected by the MLT. This was found for the GlucoMen Glyco, but not for the Glucometer Dex. For the GlucoMen Glyco, 6% of the measurements performed by the patients were done with too little blood, whereas the MLT had no such measurements. In addition, 16% of the patients had not coded the instrument correctly (Fig. 1Up and Table 1Up ). If measurements related to these errors were excluded from the calculations, the analytical quality of the meter would also be acceptable in the hands of the patients. Regarding the Glucometer Dex, there was a problem with the strip that led to poor uptake of blood. As many as 24% of the measurements performed by the patients with the Glucometer Dex at the consultation were done with too little blood. The MLT, as well as the patients who received training, had significantly fewer incompletely filled strips at the consultation than the group of untrained patients (15%, 18%, and 31%, respectively; Fig. 1Up and Table 1Up ). This problem thus would have been detected by the MLT, but was more obvious when the patients used this meter.

To evaluate the lot-to-lot variation, we included only the measurements performed by the MLT. For the Glucometer Dex, the analytical quality of one lot was significantly poorer than the quality of the two others, even when we excluded the results for incompletely filled strips. For the GlucoMen Glyco there was no difference in analytical quality among the three lots, but all had a negative bias compared with the reference method, as shown in Supplement 2 in the online Data Supplement. Lot-to-lot variation may be a considerable problem with SMBG devices and could be a major factor in loss of analytical quality (12). From 2004 onward, the quality of all lots on the Norwegian market will be examined.

Educational efforts might influence the performance of SMBG (13)(14)(15). For the Glucometer Dex, the precision obtained both at the consultation and at home was better for the patient group trained on meter use by the MLT compared with the group that received only written instructions [CV, 12% and 28% for the trained patients and the untrained patients, respectively, at the consultation (P <0.05) vs 5.6% for the trained patients and 8.8% for the untrained patients at home (P <0.05)]. For the GlucoMen Glyco, the precision obtained at home was better for the trained group than for the nontrained group (CV, 5.2% vs 7.6%; P <0.05). However, 12 patients in the nontrained group compared with 2 patients in the trained group had coded the GlucoMen Glyco incorrectly.

User errors that were assessed in the evaluations were highlighted in the questionnaires (Supplements 3 and 4 in the online Data Supplement). Regarding the Glucometer Dex, ~25% of the patients commented on the problem of poor uptake of blood. In the case of the GlucoMen Glyco, 13% answered that it was difficult to apply blood to the test strip, and 17% found it difficult to code the instrument. Manufacturing of SMBG instruments according to patients’ wishes may lead to improvements in acceptability, compliance, and glucose control (16).

Each evaluation lasted ~5 months. One month was used to prepare the work, 2.5 months were needed to complete the practical work, and 1 month was needed for result evaluation. The costs were estimated to be approximately NOK 150 000 (US $20 000) for each evaluation.

It is essential that important shortcomings of SMBG devices are disclosed before the instruments are made commercially available. A procedure for evaluating new instruments and strips should therefore be standardized, including both a user part and a part that deals with analytical quality in the hands of experienced technologists. The evaluation should not be too costly to perform. We believe that our procedure fulfills these demands, and the Norwegian Health Authorities have decided that all SMBG instruments marketed in Norway should be examined by a procedure similar to the one described in this study. In addition, all lots of strips on the market will be tested in a special survey because they cannot be included in the procedure for practical reasons.


Acknowledgments

The Glucometer Dex and GlucoMen Glyco instruments and strips used and tested by the MLT and the patients in the study were kindly supplied by Bayer Diagnostics (Tarrytown, NY) and Menarini (Firenzi, Italy), respectively. The National Office for Social Insurance in Norway provided financial support for the study. The study is part of the Global Campaign of Diabetes Mellitus launched by the IFCC.


References

  1. . DCCT Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977-986.[Abstract/Free Full Text]
  2. . UKPDS Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes. Lancet 1998;352:837-853.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  3. Thayer A. Deciphering diseases. Chem Eng News 1999;30:19-28.
  4. . Center for Devices and Radiological Health (CDRH). Human factors in self-monitoring of blood glucose: Task IV report April 1990;Vol. 1 US Food and Drug Administration Rockville, MD. .
  5. International Organization for Standardization. In vitro diagnostic test systems—requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO/TC 212/SC. Final Draft International Standard ISO/FDIS 15197. Geneva, Switzerland: ISO, 2003..
  6. Burnett RW. Accurate estimation of standard deviations for quantitative methods used in clinical chemistry. Clin Chem 1975;21:1935-1938.[Abstract]
  7. . National Committee for Clinical Laboratory Standards. Point-of-care blood-glucose testing in acute and chronic care facilities; approved guideline, 2nd ed. NCCLS document C30–A2 2002 NCCLS Wayne, PA. .
  8. . American Diabetes Association. Consensus statement on self-monitoring of blood glucose. Diabetes Care 1987;10:95-99.[Web of Science][Medline] [Order article via Infotrieve]
  9. . American Diabetes Association. Consensus statement on self-monitoring of blood glucose. Diabetes Care 1996;19:S62-S66.
  10. Boyd JC, Bruns DE. Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose. Clin Chem 2001;47:209-214.[Abstract/Free Full Text]
  11. Skeie S, Thue G, Sandberg S. Patient-derived quality specifications for instruments used in self-monitoring of blood glucose. Clin Chem 2001;47:67-73.[Abstract/Free Full Text]
  12. Skeie S, Thue G, Nerhus K, Sandberg S. Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician. Clin Chem 2002;48:994-1003.[Abstract/Free Full Text]
  13. Proposed strategies for reducing user error in capillary blood glucose monitoring. The National Steering Committee for Quality Assurance in Capillary Blood Glucose Monitoring. Diabetes Care 1993;16:493-498.[Abstract]
  14. One J. Standardization and recommendations for the appropriate use of automated devices for self-monitoring blood glucose. Rinsho Byori 2001;49:1190-1198.[Medline] [Order article via Infotrieve]
  15. Barcelo A, Robles S, White F, Jadue L, Vega J. [An intervention to improve diabetes control in Chile]. Rev Panam Salud Publica 2001;10:328-333.[Medline] [Order article via Infotrieve]
  16. Halimi S, Charpentier G, Grimaldi A, Grenier JL, Baut F, Germain B, et al. Effect on compliance, acceptability of blood glucose self-monitoring and HbA(1c) of a self-monitoring system developed according to patient’s wishes. The ACCORD study. Diabetes Metab 2001;27:681-687.[Web of Science][Medline] [Order article via Infotrieve]



The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
J. Mahoney and J. Ellison
Assessing the Quality of Glucose Monitor Studies: A Critical Evaluation of Published Reports
Clin. Chem., June 1, 2007; 53(6): 1122 - 1128.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
G. B.B. Kristensen, K. Nerhus, G. Thue, and S. Sandberg
Results and Feasibility of an External Quality Assessment Scheme for Self-Monitoring of Blood Glucose
Clin. Chem., July 1, 2006; 52(7): 1311 - 1317.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
G. B.B. Kristensen, N. G. Christensen, G. Thue, and S. Sandberg
Between-Lot Variation in External Quality Assessment of Glucose: Clinical Importance and Effect on Participant Performance Evaluation
Clin. Chem., September 1, 2005; 51(9): 1632 - 1636.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
W. E. Winter
A Rosetta Stone for Insulin Treatment: Self-Monitoring of Blood Glucose
Clin. Chem., June 1, 2004; 50(6): 985 - 987.
[Full Text] [PDF]


This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (19)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kristensen, G. B.B.
Right arrow Articles by Sandberg, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kristensen, G. B.B.
Right arrow Articles by Sandberg, S.
Related Collections
Right arrow Evidence Based Laboratory Medicine and Test Utilization


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS