Endocrinology and Metabolism |
1 Institut de Recerca, 2 Serveis dEndocrinologia, and 3 Bioquímica, Hospital de la Santa Creu i Sant Pau, Barcelona, Spain. 4 Hospital de Sabadell, Institut Universitari Parc Taulí, Sabadell, Spain. 5 Universitat Autònoma de Barcelona, Barcelona, Spain.
aAddress correspondence to this author at: Hospital de la Santa Creu i Sant Pau, Servei de Bioquímica, Sant Antoni Maria Claret 167, 08025 Barcelona, Spain. Fax 34-93-2919196; e-mail fblancova{at}hsp.santpau.es.
| Abstract |
|---|
|
|
|---|
Methods: We compared two commercially available immunoassays, an ELISA and a time-resolved immunofluorometric assay (TR-IFMA), with RBA.
Results: We found excellent agreement between the RBA and ELISA for measurement of GAD65 autoantibodies (GADAs); they showed comparable analytical precision in the cutoff range and achieved similar diagnostic specificity. The ELISA identified more GADA-positive individuals among patients with new-onset type 1 diabetes than did the RBA [89% (95% confidence interval, 7895%), vs 71% (5882%); P <0.03]. For IA-2 autoantibodies (IA-2As), only the TR-IFMA achieved analytical performance and diagnostic accuracy comparable to that of the RBA. These results with the GADA ELISA and IA-2A TR-IFMA were consistent with those obtained blindly in the Diabetes Antibody Standardization Program 2003. The performance of the GADA TR-IFMA and IA-2A ELISA was unsatisfactory, and these tests were not subjected to clinical evaluation.
Conclusions: The GADA ELISA and IA-2A TR-IFMA behave comparably with RBA and are thus suitable for use in the clinical laboratory.
| Introduction |
|---|
|
|
|---|
The standardized radiobinding assay (RBA) detects GADAs and IA-2As in
7080% and 4070%, respectively, of patients with new-onset DM1, whereas the frequency of detection in the general population is <1% (1)(6). RBAs, although cumbersome and requiring use of 35S, reportedly provide more precise results and higher diagnostic sensitivity and specificity than ELISAs (2)(3)(4).
We compared the analytical and diagnostic accuracy of two commercially available immunoassays, an ELISA and a time-resolved immunofluorometric assay (TR-IFMA), with RBA for measuring GADA and IA-2A concentrations.
| Materials and Methods |
|---|
|
|
|---|
The Diabetes Antibody Standardization Program distributed in a blind fashion serum sample aliquots from 100 controls and 50 patients with new-onset DM1 to test the sera with any assay currently in use for the detection of diabetes and to determine which samples corresponded to patients with new-onset DM1.
assays
The assays chosen for this study were the RBA (9), TR-IFMA (DELFIA®; Perkin-Elmer; cat. no. 4011-0010 for GADAs and cat. no. 4012-0010 for IA-2As) (2)(3), and ELISA (RSR Ltd.; cat. no. RSR-GDE/96 for GADAs and cat. no. RSR-IAE/96 for IA-2As) (10).
Samples for analytical evaluation were assayed in triplicate in the three assays performed on different days, and accuracy and precision were assessed according to NCCLS recommendations (11). Sera from patients or control individuals used for diagnostic sensitivity and specificity evaluation were analyzed in duplicate. All analyses were performed by two well-trained laboratorians who did not have information regarding previous analytical results or clinical classification. Autoantibody concentration results are reported in WHO international units (kIU/L) and were calculated in relation to the National Institute for Biological Standards and Control (NIBSC) 97/550 reference standard, which was used to calibrate all assays. Diagnostic sensitivity was calculated as the percentage of sera from patients reported as positive based on the laboratory cutoffs for each assay, whereas diagnostic specificity was calculated as the percentage of control sera (group IV) reported as negative based on the same threshold.
| Results |
|---|
|
|
|---|
analytical evaluation
We measured recovery of diluted NIBSC 97/550 standard (added concentration) mixed with an in-house negative control (endogenous concentration; see Table 1
in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol50/issue8/). Recoveries at GADA concentrations near the diagnostic cutoff were 8899% for the RBA, 106% for the ELISA, and 89% for the TR-IFMA. Recoveries at IA-2A concentrations near the cutoff range were 159162% in the RBA, 4864% in the TR-IFMA, and 156221% in the ELISA. Imprecision values (CV) for serial dilutions of the NIBSC 97/550 standard are listed in Table 2 of the online Data Supplement. Within-run imprecision (CVWR) for GADA values near the cutoff was
20% for the RBA, ELISA, and TR-IFMA. Total imprecision (CVT) was also
20% for the RBA and ELISA but was almost 50% for the TR-IFMA. The CVWR for IA-2A near the cutoff values was
14% for the RBA, 9% for the ELISA, and 310% for the TR-IFMA. In contrast, the CVT was
20% for the RBA and ELISA and >50% for the TR-IFMA. The detection limits (calculated as the mean signal + 3 SD of a negative serum) for GADAs were 15.6 kIU/L in the RBA, 3.2 kIU/L in the ELISA, and 22.4 kIU/L in the TR-IFMA, and for IA-2As were 4.0 kIU/L in the RBA, 32.1 kIU/L in the ELISA, and 20.6 kIU/L in the TR-IFMA.
|
Given our previous experience of poor data replication of plasma samples from controls and patients, we excluded the TR-IFMA GADA and ELISA IA-2A methods from the following clinical evaluation although, in general, we found no major failures during the analytical evaluation and cutoff limit determinations.
clinical evaluation
The ELISA and RBA GADA showed diagnostic specificities of 99% (95% CI, 93100%) and 96 (8999)%, respectively. The highest diagnostic sensitivity for GADA in patients with new-onset DM1 was obtained with the ELISA (89%; 95% CI, 7895%), whereas the RBA had a sensitivity of 71 (5882)% in the same group of patients (P <0.03). GADA values in the four groups analyzed are shown in Fig. 1A
.
|
The IA-2A TR-IFMA had a diagnostic specificity similar to that of the RBA [99 (93110)% and 97 (9199)%, respectively]. The diagnostic sensitivity of the TR-IFMA and RBA in patients with new-onset DM1 was 37 (2651)% and 50 (3763)%, respectively (difference not significant). IA-2A TR-IFMA positivity in the group with nonclassified diabetes mellitus was not statistically different with respect to RBA (24% and 20%, respectively). No patients clinically classified as having DM2 were positive for IA-2As in either the RBA or TR-IFMA. IA-2A concentrations obtained with the RBA and TR-IFMA in the four groups studied are shown in Fig. 1B
.
Results obtained with each assay for patients with new-onset DM1 were subjected to ROC curve analysis (see Fig. 1
in the online Data Supplement). Areas under the ROC curves showed a significant difference (P <0.001) between GADA assays: 0.84 in the RBA (95% CI, 0.770.90) and 0.99 in the ELISA (95% CI, 0.951.00). IA-2A areas under ROC curves also showed a significant difference (P <0.001) between assays: 0.54 in the RBA (95% CI, 0.450.63) and 0.80 in the TR-IFMA (95% CI, 0.720.86).
PassingBablok regression analysis for GADA yielded the following equation: ELISA = 0.6 RBA 3.9 kIU/L (95% CI for intercept, 4.5 to 3.0 kIU/L; 95% CI for slope, 0.50.7), which shows a high correlation between methods (r = 0.94; P <0.001). A scatter plot of all samples analyzed for GADAs by RBA and ELISA is shown in Fig. 2A
. Seventeen of 141 patient samples yielded discordant results: all except one (belonging to the DM2 group) tested positive with the ELISA and negative with the RBA. Ten of these 17 samples were from patients with new-onset DM1, 4 were from patients with nonclassified diabetes mellitus, and 3 were from the DM2 patient group (potential LADA patients). One control sample was slightly positive with both assays, but two additional samples among the control sera were positive in the RBA.
|
Regarding IA-2As, PassingBablok regression analysis yielded the following equation: TR-IFMA = 2.9 RBA 1.0 kIU/L (95% CI for intercept, 6.0 to 2.2 kIU/L; 95% CI for slope, 1.94.2; r = 0.83; P <0.0001). A scatter plot of all samples analyzed for IA-2A is shown in Fig. 2B
. Twelve patient samples produced no concordant results: 9 were from patients with new-onset DM1, whereas the others were from patients with nonclassified diabetes mellitus. All but one, which was from a patient with new-onset DM1, tested positive with the RBA and negative with the TR-IFMA. One control sample was positive in the TR-IFMA, whereas two other control samples were positive in the RBA.
The diagnostic sensitivities and specificities obtained in the Diabetes Antibody Standardization Program 2003 for the GADA RBA and ELISA and the IA-2A RBA and TR-IFMA are listed in Table 1
.
| Discussion |
|---|
|
|
|---|
The CVWR for IA-2A concentrations near the cutoff was better for the TR-IFMA than for the RBA, although the CVT behaved in an opposite manner at the same threshold. Our imprecision values were similar to those reported previously (3). In PassingBablok regression analyses, the IA-2A RBA and TR-IFMA methods showed a proportional difference, which is not important for diagnosis of autoimmune diabetes given that, near the cutoff range, the assays correlated very well. However, this difference could yield a different quantitative risk assessment in the general population or in relatives of patients with DM1 because the IA-2A concentration has been suggested as being related to risk for DM1 (6).
The diagnostic sensitivities of IA-2A measurements obtained with the TR-IFMA and RBA in our patients with new-onset DM1 and in the Diabetes Antibody Standardization Program 2003 for patients with new-onset DM1 were consistent with the previously reported prevalence of IA-2A (4070%) (1), although lower than the prevalences reported by others using the same methods (3). None of the assays detected positivity in patients with DM2. It is well known that IA-2A prevalence is lower than that of GADAs in patients with LADA (usually <4%) (1). Thus, the number of patients with DM2 studied was too low to yield positive results.
The results obtained in the present study demonstrate that the ELISA for GADA and the TR-IFMA for IA-2A detection represent good alternatives to the corresponding RBAs. The ELISA and TR-IFMA are simple, easy to automate, and constitute highly reproducible methods suitable for routine use in the clinical laboratory. Moreover, both assays are based on reagents that are stable over long periods and avoid the use of harmful radioactive material. One disadvantage of the ELISA and TR-IFMA is the higher sample volume required compared with the RBA, particularly for autoantibody determination in newborns.