Various methods have been used to measure protein in urine and cerebrospinal fluid (CSF). Urinary protein can be determined by turbidimetric assays using trichloroacetic acid (1), sulfosalicylic acid (2), or benzethonium chloride (BTC) (3)(4); the biuret assay (5)(6); or protein dye-binding assays using Coomassie Brilliant Blue (CBB) (7)(8)(9)(10)(11)(12)(13), Pyrogallol Red (PYR)-molybdate (14)(15), or Ponceau S (16). The BTC, CBB, and PYR methods can be automated. Shephard and Whiting (17) used benzalkonium chloride (BC) as a precipitating agent for the determination of protein in urine and CSF by nephelometry. In this study we developed an automated turbidimetric method using BC and compared it with conventional automated methods.
Albumin fraction V (bovine), disodium molybdate · 2 H2O, NaOH, sodium azide, sodium benzoate, potassium oxalate, and succinic acid were obtained from Merck; BTC, Coomassie Brilliant Blue G 250, EDTA, and PYR were from Sigma; BC was from Serva; and phosphoric acid was from Riedel de Haen. A commercially available PYR reagent (M-TP) and protein calibrators were obtained from Beckman Coulter. A globulin solution containing mostly IgG and IgM and no albumin was obtained from the Refik Saydam Center of Hygiene, Turkey.
We developed a new method that uses the same principle as the method of Iwata and Nishikaze (3) but with BC instead of BTC. BC is a detergent and causes protein denaturation in an alkaline medium. The NaOH/EDTA reagent (reagent 1) contained 33 mmol/L EDTA and 0.5 mol/L NaOH. The BC solution (5 g/L; reagent 2) was prepared from a 50 g/L aqueous stock solution of BC. Calibrators containing 50, 100, 200, 500, and 1000 mg/L protein were prepared from a Beckman serum calibrator (Multical). In a manual procedure, we mixed 2.0 mL of NaOH/EDTA with 200 µL of sample and immediately added 0.5 mL of BC reagent. After a 10-min incubation, we measured the absorbance at 450 nm. We then adapted our method to a Synchron® LX 20 Pro analyzer, using the following conditions: reagent 1 volume, 200 µL; reagent 2 volume, 50 µL; sample volume, 20 µL; primary wavelength, 470 nm; secondary wavelength, 700 nm; blank reading (sample blank), 4872; reaction time, 10 min; reaction reading, 700748; calibrators, 50, 100, 200, 500, and 1000 mg/L.
We compared the performance characteristics of our method with those of the Bradford CBB assay (7), the PYR assay (14), and the BTC assay (3). We adapted the PYR and BTC methods to the same analyzer and used the Beckman Coulter M-TP assay, which is also based on the PYR-molybdate method, for comparison. It has previously been demonstrated that the PYR-molybdate method is analytically and clinically comparable to the reference biuret method (15). We measured the total protein concentrations in 100 urine and 88 CSF samples simultaneously by each of the methods. Urine samples were assayed on the day of collection after centrifugation, and CSF samples were stored at 20 °C before analysis. Deming regression analysis, the Pearson correlation coefficient between the methods, and all other statistical calculations were performed with the Microsoft Excel Analyze-it program. The study was approved by the Local Ethics Committee of our hospital.
Samples containing 100-5000 mg/L protein were prepared by adding known amounts of protein calibrators to a urine pool. The PYR assay was linear up to 3000 mg/L; the BC, BTC, and M-TP assays were linear up to 1500 mg/L; and the CBB assay was linear up to 1000 mg/L. Serial dilutions of a sample containing 200 mg/L protein were prepared (range, 6.25200 mg/L), and a protein concentration that produced an absorbance +3 SD above that for a zero calibrator (n = 21 replicates) was accepted as the detection limit of the method. The detection limits were 50 mg/L for the CBB, PYR, and M-TP methods and 25 mg/L for the BTC and BC methods.
The results of our imprecision studies (n = 21) and for recovery of protein added to urine and CSF are shown in Table 1
. Albumin and globulin solutions at the same concentration were prepared, and absorbance ratios were calculated (Table 1
). We measured the total protein concentrations of 100 urine and 88 CSF samples in duplicate by the five methods simultaneously and averaged the duplicates for each method. Deming regression and Pearson correlation analyses were performed. We also adapted the nephelometric BC method described by Shephard and Whiting (17) to the Beckman Coulter Immage nephelometer and determined its performance characteristics separately. The nephelometric BC method was linear at protein concentrations of 501000 mg/L. Within-run CVs were 5.7%, 5.1%, and 3.6% for samples with mean protein concentrations of 127, 497, and 844 mg/L, respectively. Between-run CVs for the same samples were 12%, 5.7%, and 5.9%, respectively.
|
Methods for determining total protein in urine and CSF samples have different advantages and disadvantages. Biuret-gel-filtration or biuret-precipitation methods are accurate, but they are time-consuming and impractical for routine use. Dye-binding methods using CBB and PYR are rapid, simple, and suitable for automation (15)(18)(19) but tend to give different responses to different types of proteins (9)(10)(11)(12)(13)(18). The BTC assay can also be automated (4). In our study, the PYR method was linear up to 3000 mg/L and seemed to be more advantageous than the other methods, but albumin-globulin response studies showed that globulins produced 50% lower responses than did albumin; adding sodium dodecyl sulfate to the reagent decreased this difference to 20%, but this modification also decreased the upper limit linearity of the method to 1500 mg/L. The M-TP, BTC, and BC methods had also 20% lower responses to globulins, whereas the CBB method had a 30% difference, although sodium dodecyl sulfate was added as suggested (20). In the imprecision studies, the BC method had the lowest within- and between-run CVs. In between-run imprecision studies, the BC method had a significantly lower CV than the BTC method, and this was thought to be related to reagent stability. This finding is particularly important because the BTC and BC methods are based on the same principle. The BC method is also more advantageous than the BTC method in a manual procedure because the BC method has a 10-min incubation time at room temperature, whereas the BTC method requires incubation for 4060 min (3)(21) at room temperature.
For urinary total protein, the 150200 mg/L concentration range is important because this range is used for some clinical decisions. In this concentration range, the BC assay measured 100% of the expected protein concentration. Marshall and Williams (22) compared the Sigma PYR and CBB assays in their study and obtained the equation: CBB = 0.685PYR + 17 mg/L. They proposed using urinary protein as the calibrator and found that it increased the agreement between the methods (slope, 0.972; y-intercept, 16 mg/L). We also found a 25% difference between the CBB and M-TP methods in urine samples but good agreement in CSF samples. Our results showed that, for urine samples, the results obtained with the CBB method do not agree with those obtained with the M-TP, BC, BTC, or PYR methods. The comparison studies showed that BC correlates better with the M-TP method than do the BTC and CBB methods for urine and CSF samples, based on the slope, intercept, and correlation (r) values.
BC is a widely used disinfectant and is much less expensive than the other reagents, especially the dyes CBB and PYR. It also inhibits bacterial contamination, and this can be a reason for the better reagent stability. Shephard and Whiting (17) compared four protein precipitantsBC, BTC, trichloroacetic acid, and sulfosalicylic acid for nephelometryand found that BC had good precision and agreed most closely with the trichloroacetic acidPonceau S dye-binding method. In our hands, the nephelometric method of Shephard and Whiting was linear 501000 mg/L protein, similar to the measuring range (781250 mg/L) reported by those authors (17). In imprecision studies, the precision and linearity of our method were better than those of the nephelometric method. Nephelometers are also expensive and are not used in many in small- and medium-sized clinical laboratories.
In conclusion, we present a new turbidimetric method that correlates better with the M-TP method than do the BTC and CBB methods; is more precise than the M-TP, CBB, PYR, and BTC methods; has better precision and recovery in the critical decision concentration range; has satisfactory albumin-globulin response and analytical range; uses less expensive and accessible chemicals; and is suitable for automation.