Cancer Diagnostics |
1 Department of Microbiology & Molecular Cell Biology, Virginia Prostate Center, Eastern Virginia Medical School, Norfolk, VA.
2 Fred Hutchison Cancer Center, Seattle, WA.
3 Center for Prostate Disease Research, Department of Surgery, Uniformed Services University of the Health Sciences, Rockville, MD.
4 University of Pittsburgh Cancer Institute, Hillman Cancer Center, Pittsburgh, PA.
5 Cancer Therapy and Research Center, Institute for Drug Development, San Antonio, TX.
6 Department of Pathology, Johns Hopkins Medical Institutes, Baltimore, MD.
7 Department of Pathology, University of Alabama at Birmingham, Birmingham, AL.
8 Cancer Biomarkers Research Group, Division of Cancer Prevention, National Cancer Institute, Bethesda, MD.
9 Department of Medicine, University of Texas Health Sciences Center, San Antonio, TX.
aAddress correspondence to this author at: Department of Microbiology and Molecular Cell Biology, Eastern Virginia Medical School, 700 W. Olney Rd., Norfolk, VA 23507. Fax 757-446-5766; e-mail semmesoj{at}evms.edu.
| Abstract |
|---|
|
|
|---|
Methods: Two sequential experimental phases were conducted to establish interlaboratory calibration and standardization of the surface-enhanced laser desorption (SELDI) instrumental and assay platform output. We first established whether the output from multiple calibrated Protein Biosystem II SELDI-ionization time-of-flight mass spectrometry (TOF-MS) instruments demonstrated acceptable interlaboratory reproducibility. This was determined by measuring mass accuracy, resolution, signal-to-noise ratio, and normalized intensity of three m/z "peaks" present in a standard pooled serum sample. We next evaluated the ability of the calibrated and standardized instrumentation to accurately differentiate between selected cases of prostate cancer and control by use of an algorithm developed from data derived from a single site 2 years earlier.
Results: When the described standard operating procedures were established at all laboratory sites, the across-laboratory measurements revealed a CV for mass accuracy of 0.1%, signal-to-noise ratio of
40%, and normalized intensity of 1536% for the three pooled serum peaks. This was comparable to the intralaboratory measurements of the same peaks. The instrument systems were then challenged with sera from a selected group of 14 cases and 14 controls. The classification agreement between each site and the established decision algorithm were examined by use of both raw peak intensity boosting and ranked peak intensity boosting. All six sites achieved perfect blinded classification for all samples when boosted alignment of raw intensities was used. Four of six sites achieved perfect blinded classification with ranked intensities, with one site passing the criteria of 26 of 28 correct and one site failing with 19 of 28 correct.
Conclusions: These results demonstrate that "between-laboratory" reproducibility of SELDI-TOF-MS serum profiling approaches that of "within-laboratory" reproducibility as determined by measuring discrete m/z peaks over time and across laboratories.
| Introduction |
|---|
|
|
|---|
Several laboratories have demonstrated the feasibility of using mass spectrometric proteomic pattern analysis for the diagnosis of several categories of tumors, including ovarian, breast, lung, pancreas, and prostate cancer (5)(6)(7)(8)(9)(10)(11)(12)(13)(14)(15)(16)(17)(18)(19)(20)(21)(22)(23)(24). In one of the earliest contributions, investigators from Eastern Virginia Medical School (EVMS)1 and the Fred Hutchison Cancer Research Center (FHCRC) demonstrated that surface-enhanced laser desorption/ionization time-of-flight mass spectrometry (SELDI-TOF-MS) protein profiles may have clinical utility as a diagnostic tool for the early detection of prostate cancer, including prostatic adenocarcinoma (PCa) (6)(8). Specifically, using just nine differentially expressed protein peaks, the assay achieved an overall PCa case/control correct classification rate >90%. Also of interest was the high specificity, 97%, which is significantly better than that of prostate-specific antigen (PSA) and would therefore be of obvious clinical use for PCa screening. This is particularly important given the recent evidence that current PSA screening practices do not detect the majority of prostate cancers, including those that are high grade (25).
Recently, there has been considerable controversy concerning the SELDI profiling approach (26); the most publicized of which have been several commentaries by Diamandis and the analysis of publicly available data posted by Petricoin and coworkers. The major concern raised by Diamandis (27)(28) has been whether the SELDI-based approaches are reproducible, whereas the concerns raised by Sorace and Zahn (29) and Baggerly et al. (30) deal with study design bias. There is also a great deal of confusion and controversy as to whether the use of high-throughput proteomic techniques such as SELDI can improve the early detection of prostate cancer, including PCa (6)(9)(15)(28)(31). In addition, clinical issues as to whether SELDI detects equally well both high-grade (Gleason score
7) and low-grade (Gleason score
6) cancers of the prostate remain to be determined. Similarly, a very important comparison will be the use of SELDI-TOF-MS in the early detection of PCa compared with the present implementation and future use of assays for PSA. All of these issues can and should be addressed in any appropriately designed validation study (32).
It is the goal of this collaborative projectthe Early Detection Research Network (EDRN)-Prostate-SELDI Investigational Collaboration (EPSIC)to use state-of-the-art protein profiling technology to develop and validate high-throughput screening methods for the early detection of PCa. A successful validation study of SELDI profiling is a necessary first step toward demonstration of clinical viability of the assay. In addition, we have included an initial first phase to determine reproducibility of the SELDI assay as a specific response to the question of platform reproducibility. This report describes the methods, results, and pitfalls encountered in the initial assay reproducibility assessment phase of our analysis.
| Materials and Methods |
|---|
|
|
|---|
|
interlaboratory calibration and standardization of instrument output
SELDI-TOF-MS is a modification of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. The principal difference is the addition of an affinity chromatographic step incorporating specific chemical modification of the desorption surface. Thus, sample complexity is reduced "on-chip" in a convenient, easily manipulatable format before laser desorption. The specific system used for these studies is the Protein Biosystem II model outfitted with the metal affinity (IMAC3-Cu) ProteinChip®. Each of the six laboratory sites was equipped with a Protein Biosystem II (Ciphergen Biosystems Inc.) and a Biomek® 2000 Workstation liquid-handling robot (Beckman Instruments), which included an eight-channel pipette and wash tool and a custom-configured mixer/shaker. The robot is adjusted to prepare the samples and spot them onto the Ciphergen ProteinChip arrays by use of the Ciphergen 12-array Bioprocessor® unit. Sinapinic acid matrix addition was also performed robotically. These detailed preparative protocols have been described elsewhere (6)(19), and the complete standard operating procedure is included in the Data Supplement that accompanies the online version of this article athttp://www.clinchem.org/content/vol51/issue1/ .
The synchronization and standardization process for phase IA of the study was performed as follows: Aliquots of pooled normal human sera prepared at EVMS were used as QC samples. This QC sample consists of pooled serum obtained from 360 healthy individuals (197 women and 163 men), giving almost 2 L of serum. Serum from each individual was collected by venipuncture into a 10-mL SST VacutainerTM Tube. Blood was allowed to clot at room temperature for 30 min, and the tubes were centrifuged at 1500g for 10 min. Each individual serum sample was then decanted and pooled into a 3-L beaker on ice. The pooled serum was separated into 0.4-mL aliquots and stored at 80 °C. Each aliquot of the QC pool has therefore undergone only one freeze-thaw cycle. Spectra obtained for these QC samples at EVMS were used as benchmarks for the participating sites, and a QC sample is run on one randomly chosen spot of every chip processed. Protocols and a checklist of required criteria were sent to each site, and technical staff from EVMS and Ciphergen Biosystems coordinated a visit to each institution to standardize the Protein Biosystem II instruments and prepare the robotics for serum processing. The QC serum samples were assayed, and the resulting data were used to evaluate the performance of both the Biomek 2000 robotic sample-processing station and the Protein Biosystem II SELDI-TOF-MS instrument at each site. Instrument calibration of the spectra was performed externally with the All-in-1 peptide calibrator (Ciphergen), which contains seven peptides in the mass range of 1084.257033.61 Da. Instrument calibration was performed weekly, and the most recent calibration equation was applied.
After routine instrument performance was tested, standardization of output was accomplished by adjusting the laser intensity, detector voltage, and detector sensitivity of each instrument so that three consistently present protein peaks in the QC serum (m/z 5910, 7773, and 9297 ± 0.2%) were displayed to specific criteria (see Table 1
). Additionally, the resolution values for all three peaks were required to be >400, with signal-to-noise ratios (S/N)
40 for m/z 5910 and
80 for m/z 7773 and 9297. When these criteria were met, each onsite researcher then performed an evaluation phase. This consisted of running two additional Bioprocessors (12 ProteinChips/each) of QC and sending the data to EVMS for examination and initial analysis. The same raw data were also sent to the FHCRC for final analysis.
During the course of the initial validation phase, the QC pool was assayed weekly. If any set of QC spectra did not meet the criteria specified, instrumental data collection settings were adjusted. If the QC spectra still did not meet the expected criteria, robot protocols were checked and adjusted. If this did not resolve the problem, the Protein Biosystem II instrumentation was then checked by the Ciphergen engineer.
samples and data collection
After phase 1A standardization of the SELDI-TOF-MS system at a site, each site received 14 PCa and 14 non-PCa sera from EVMS. All serum samples used in this study were collected and assayed under a protocol approved by the respective Institutional Review Boards for each laboratory site. Criteria used to select the 14 PCa and 14 non-PCa samples from 186 PCa patients and 219 age-matched healthy men included that (a) the spectra show no signs of protein degradation, (b) at least 1 mL of serum remain in the serum bank, and (c) samples were correctly classified by the current classifier. We did not use samples that were marginally classified because the objective of phase 1B was platform validation and not biological validation of the previously derived classifier. The operator laboratory sites were blinded to the mass locations of the three diagnostic peaks. Samples were coded and analyzed blindly by each of the six laboratories, including EVMS. Each of the six sites prepared these samples on IMAC3-Cu ProteinChips, using their Biomek robotic system, and analyzed them on the SELDI-TOF-MS system. No effort was made to ensure that the ProteinChips used were from the same lot, and multiple lots were used within and across laboratory sites. All sites sent their primary data and processed spectra to the coinvestigators at the FHCRC, the Data Management and Coordinating Center of the EDRN, which analyzed each sites ability to identify the peaks required for accurate analysis and classification of these samples.
data analysis
The classifier used in this study was constructed by use of boosting (boosting logistic regression and boosting decision tree). The details have been described previously (8)(33). The peak identification and alignment method used has also been described (34). Three diagnostic peaks were fixed for this classifier and used in phase 1B. The maximum intensity value in each of the three windows centered at the diagnostic peaks, with a half window width of ±0.2% of mass value, was used for each triplicate, and the median intensity value was selected. The median intensity value for each diagnostic peak was then used in the classifier to calculate a predication score. If the prediction score value was >0, the predictive probability for this sample being a cancer was greater than that for being a control, and the sample would be classified as a cancer; otherwise, it would be classified as a control. Scores of all 28 samples were calculated for each site by use of this fixed classifier, and the results were compared with the scores calculated on the original spectra generated 17 months earlier at the EVMS site.
| Results |
|---|
|
|
|---|
|
|
|
The entire phase IA, including data analysis, was completed in
6 months. Table 1
in the online Data Supplement shows the timelines for site visits, data collection, and analysis. Once the initial observations and evaluation of the QC spectra were completed, each validation laboratory was then directed to continue to the next phase of the study, phase IB.
frequent and accurate instrument calibration is essential for reproducibility
The values for both m/z and peak intensity are critical components to the construction of the SELDI-TOF-MS spectral profile. Because consistency for these measures would be critical to reproducible performance, we examined the impact of accurate mass designation and spectral intensity on cross-site agreements. For this experiment, we compared the spectral data from each site under two calibration scenarios. In the first scenario, the spectral output was correctly calibrated by optimized analysis of a 7-in-1 peptide calibrator; in the second, the spectral output from the 7-in-1 peptide calibrator was poorly resolved. The poor calibration occurred because two of the peptides peaks from the 7-in-1 peptide solution had observed mass values disparate from theoretical, leading to a poorer calibration. The right-hand panels of Fig. 2
show QC peaks from the poor calibration attempt. The left-hand panels show QC peaks based on the 7-in-1 peptide calibrator that demonstrated adequate calibration. When the poor calibration data were used, the mass designation values where dramatically shifted outside the acceptable range. The misclassification rate based on the poor calibration was 93%, whereas use of the adequate calibration led to a misclassification rate of 0%. If the calibration was performed with the 7-in-1 peptide calibrator but the two poorly resolved peptide peak measures were omitted from the calibration, the misclassification error was 0% as well (data not shown).
|
phase ib: classification of PCA case/control samples
The second step in the EPSIC phase I design was to determine whether, after instrument calibration and output standardization, the separate sites could achieve comparable correct classification rates when challenged with the same previously characterized 14 PCa and 14 control samples. The 14 PCa and 14 control sera were provided by EVMS and selected from the larger EVMS-2002 pool originally used to construct the classifier. These 28 samples were chosen based on being easiest to separate by the classifier. The direct assessment of the between-site agreement was to apply the classifier previously developed from EVMS-2002 data gathered 2 years previously to the current data developed at each site. Our pass criterion for this phase IB study was that at least two predesignated sites, EVMS and the University of Alabama at Birmingham (the EDRN-designated Biomarker Discovery Laboratory and the Biomarker Validation Laboratory, respectively) achieve correct classification of at least 26 of 28 samples. Thus, the challenge was to both reproduce the classification results from 2 years earlier when the same samples were analyzed and to demonstrate that the standardized instruments would match this output. The scores and classifications of the 28 samples for all six sites compared with the original spectra are shown in Fig. 3
. For good agreement, the values of the scores for all six sites should be near the values for the original scores and also achieve a high degree of correct classification. As shown in Fig. 3
, correct classification was achieved for data from all six sites. In particular, the majority of the values of scores for PCa spectra were near their corresponding original scores (Fig. 3
, filled circles). Interestingly, the classification of control samples, although 100% correct within the decision window, demonstrated more variation in their scores within each sample. Also of note, there appears to have been larger intersite variation for control samples; most notably with the scores from two sites, EVMS and Walter Reed Army Medical Center, Center for Prostate Disease Research, moving toward zero.
|
among-site agreement
We next examined the individual spectral output from each laboratory site for the fine scale agreement at each of the peak values used in the diagnostic algorithm. In Fig. 4
, median intensity across three replicates per sample was plotted for a random sample of PCa and controls from the EVMS-2002 study as representative of the training data set. In addition, Fig. 4
shows the median sample values from the EVMS-2002 study for the 14 PCa and 14 control samples selected for the validation study, together with the diagnostic spectra for the same 28 samples from the six sites. A few points are worth noting. Peak 1, the most dominant diagnostic peak, has the best agreement among the participating laboratories. When we binned the window values, the major features of the peaks, intensities, were captured. Peaks 2 and 3 are weak peaks, and agreement among the laboratories seemed more difficult to achieve for these peaks (see Figs. A and B in the online Data Supplement). Therefore, a classifier based on strong peaks seems more likely to be reproducible across multiple laboratories. All three peaks in all laboratories were shifted to the left when compared with the peaks in the same 28 samples and that in the training samples, indicating that the training sample may have bias in calibration.
|
We also used a ranked intensity approach to select for diagnostic m/z peaks from the larger EVMS-2002 dataset because this approach might be expected to be more robust than the raw intensity method. However, the ranked intensity method led to selection of a major discriminating m/z value that corresponded to a "shoulder" peak (see Fig. C in the online Data Supplement). This was in contrast to the separate peaks (dominant m/z value within the local spectral neighborhood) used in the raw intensity approach. When the ranked intensity approach was evaluated by use of the data for the 14 cases and 14 controls from each of the six sites, four of the sites achieved perfect classification, one of the sites passed with 26 of 28 correctly classified, and one site failed with 19 of 28 correctly classified (data not shown).
| Discussion |
|---|
|
|
|---|
The first objective evaluation of the SELDI-TOF-MS methodology was for the ability to standardize the output of six Ciphergen Protein Biosystem II instruments. Specifically, we are asking whether the platform and assay conditions can be transferred from an originating laboratory site to other secondary sites. This is an important test because it also requires reproducibility of established constant conditions over time. Thus, regardless of the actual identities of the proteins that give rise to the observed SELDI-TOF-MS peaks, the expression patterns will need to be shown to be reproducibly observed from site to site. To successfully accomplish this first test (phase IA), each site had to adopt strict standard operating procedures. We found it absolutely necessary for each site to monitor continuously, whenever data were gathered, the spectral output of each instrument and adjust it as necessary to achieve the exact display of several peaks consistently found in the QC serum. Indeed, it is obvious from our results that automated software/hardware improvement is needed in this regard to ensure reproducible optimization of instrumental operation. It is also clear that automation of all sample preparation steps is essential to achieving accurate reproducibility. Several instrument parts were also replaced to keep the instruments in optimal performance, and this list has also been included in the online Data Supplement.
Once the instruments at each of the six laboratory sites had been calibrated and the output standardized, they were ready to be challenged by PCa case and control samples. In this test, each instrument was required to assay case/control samples from the same source. These case/control test samples originated from the larger EVMS-2002 sample set that was used to generate the decision algorithm being challenged. Thus, a static decision algorithm and standardized assay approach was being evaluated. Two models were used to initially select the 14 cancers and 14 controls that were to be separated in phase 1B. One model used boosting applied to aligned peak raw intensities, and these results are presented in full. The other model, which we include for comparison, ranked the peak intensities within each sample from 1 to 100. (Note that there are more than 100 peaks to rank; therefore, peak intensities are aggregated into bins by use of the rank approach.) Boosting was then applied to the rank values. Cancer and control cases that were furthest apart according to both models were selected for phase 1B. In phase 1B, we evaluated the performance of both of these classifiers. The classifier built by boosting of the peak intensities had perfect classification for all six sites. The classifier built by boosting of the ranked intensities had perfect classification for four sites, passable classification at one site, and failed our criteria for classification at one site. It is interesting that the inherently more robust ranked intensity approach had more difficulty achieving perfect classification compared with the raw intensity value approach. We believe that this is attributable to the use of a shoulder peak m/z value in the final decision structure. This would suggest that complex peak shapes will be more difficult to accurately reproduce and may fail as robust marker events. This result underscores the necessity of defining peak vs "noise" events from the m/z spectral data as a required preliminary step.
Although the EPSIC passed the criteria set for phase 1B using either value selection approach, there are several other issues worth discussion. The first is the role of calibration. The process for calibrating the SELDI-TOF-MS instrument is explained in the manufacturers manual, but the importance of the process is not made very clear. The described approach involves the use of a peptide calibrator, which is run on the SELDI instrument, and the spectral output is used to calibrate the conversion of time of flight to mass/charge (m/z) values. We recommend that the peptide calibration equation is always first checked by its residual plot for goodness of fit and then compared with the spectral peaks of the pooled serum for mass designation. An inadequate calibration equation can lead to a significant shift of the m/z values for the peak maximum. If this had occurred during our analysis, the sites using this poor calibration would have failed to classify correctly. These poorly calibrated spectra are detected only by careful examination of the profiles of the QC spectral peak locations on the magnitudes of the observed m/z shifts of peaks. A cursory visual inspection of the overall spectra will likely miss this important problem and consequently lead to failure of the classification comparisons. We are currently developing analytical software approaches to incorporate external/internal reference peaks to prevent this from occurring.
The optimization/standardization of peak intensities is also important for successful classification. For all three diagnostic peaks, cases have lower intensities than controls, and the diagnostic peak in Fig. 4
is the most dominant diagnostic peak feature for this classifier. Compared with the other sites, spectra obtained at the EVMS site showed lower intensities for the control samples. This may explain the observation that the control scores from the EVMS laboratory site were more shifted toward zero than were the scores from the other sites. In addition, peak 1, the most dominant diagnostic peak, had the best agreement among the sites. When we binned the windows, the major features of the peaks, such as intensities, were captured because maximum values within the bin usually took the values near the peaks. Peaks 2 and 3 are relatively weak peaks, and agreement among the sites was more difficult to achieve for these peaks features (see the online Data Supplement). Therefore, a classifier based on strong peaks seems to have higher likelihood of being reproducible across multiple laboratories. All three peaks in the spectra from all of the sites were shifted to the left when compared with the same peaks in the original training samples, indicating that the training set may possess some bias in calibration.
The Protein Biosystem II instruments at several sites required repair or parts replacement over the course of phases AC (see Table 2
in the online Data Supplement). In addition, several sites required that the detector voltage be incrementally increased over time to produce QC spectra that consistently passed the specified criteria. This is a result of detector decay over time, which is known to occur in this instrumentation. We recommend that, in addition to continual monitoring of the QC spectra, laboratories should keep log books of usage to anticipate replacement and detector voltage drift events.
In summary, it may be helpful to reconsider the comments of Diamandis (27)(28). Diamandis raised concern that the "discriminating peaks are not consistent either within a group or among groups of individuals". This comment, which was derived from a "metaanalysis" of the published literature, has now been experimentally addressed in our study in which we show clearly that the same three diagnostic peaks, at least the first strong diagnostic peak, were identified at multiple sites and were effective at differentiating case/control samples at all sites. In addition, the likelihood of separate sites identifying the same peaks has been disputed (38)(39). We have not revealed the identities (i.e., the m/z values) of the diagnostic peaks used in this study because the ongoing studies (phase II) require the sites to remain blind to the identities of these peaks. Although concern for lack of "analytical sensitivity" and whether the diagnostic peaks are from "high-abundant molecules" was raised, these issues alone become more academic if the assay is clinically useful and reproducible. However, if, as we have described, the larger, more prominent peaks are the most reproducible, then indeed this process is limited at the level of effective analytical sensitivity. Specifically, improvements in the sensitivity that are reflected in stronger m/z peak S/N will improve the robustness of measurements; consequently, upfront sample depletion and concentration steps may be needed. Ultimately, what was experimentally called for was a "published report that similar data can be obtained by using different batches of SELDI chips, different technologists, or by using the same conditions at a later time". Our current EPSIC study provides this information. We have demonstrated that under the strict operating procedures that we have described, we are able to achieve across-laboratory reproducibility of SELDI-TOF-MS analysis. We would therefore suggest that the next important step in the evaluation of this assay approach is a demonstration of the robustness of clinical classification as is slated for our phase II design.
In conclusion, we show that, if the protocols described here are adhered to, SELDI-TOF-MS profiling can provide a reproducible diagnostic assay platform, providing that the measured events are separable m/z peak values. We are currently examining the robustness of the assay across patient sample sets. This phase II analysis will address the ability of the current combination of SELDI-TOF-MS assay and decision algorithm to perform as a robust diagnostic tool for the detection of prostate cancer.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
J. F. Timms, R. Cramer, S. Camuzeaux, A. Tiss, C. Smith, B. Burford, I. Nouretdinov, D. Devetyarov, A. Gentry-Maharaj, J. Ford, et al. Peptides Generated Ex Vivo from Serum Proteins by Tumor-Specific Exopeptidases Are Not Useful Biomarkers in Ovarian Cancer Clin. Chem., February 1, 2010; 56(2): 262 - 271. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. LEBRECHT, D. BOEHM, M. SCHMIDT, H. KOELBL, R. L. SCHWIRZ, and F. H. GRUS Diagnosis of Breast Cancer by Tear Proteomic Pattern Cancer Genomics Proteomics, May 1, 2009; 6(3): 177 - 182. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Lakka Klement, T.-T. Yip, F. Cassiola, L. Kikuchi, D. Cervi, V. Podust, J. E. Italiano, E. Wheatley, A. Abou-Slaybi, E. Bender, et al. Platelets actively sequester angiogenesis regulators Blood, March 19, 2009; 113(12): 2835 - 2842. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. LEBRECHT, D. BOEHM, M. SCHMIDT, H. KOELBL, and F. H. GRUS Surface-enhanced Laser Desorption/Ionisation Time-of-flight Mass Spectrometry to Detect Breast Cancer Markers in Tears and Serum Cancer Genomics Proteomics, March 1, 2009; 6(2): 75 - 83. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Sturgeon, B. R. Hoffman, D. W. Chan, S.-L. Ch'ng, E. Hammond, D. F. Hayes, L. A. Liotta, E. F. Petricoin, M. Schmitt, O. J. Semmes, et al. National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines for Use of Tumor Markers in Clinical Practice: Quality Requirements Clin. Chem., August 1, 2008; 54(8): e1 - e10. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. N. McGuire, J. Overgaard, and F. Pociot Mass spectrometry is only one piece of the puzzle in clinical proteomics Briefings in Functional Genomics, February 28, 2008; (2008) eln005v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. Liotta and E. F. Petricoin Putting the "Bio" back into Biomarkers: Orienting Proteomic Discovery toward Biology and away from the Measurement Platform Clin. Chem., January 1, 2008; 54(1): 3 - 5. [Full Text] [PDF] |
||||
![]() |
R. E. Banks Preanalytical Influences in Clinical Proteomic Studies: Raising Awareness of Fundamental Issues in Sample Banking Clin. Chem., January 1, 2008; 54(1): 6 - 7. [Full Text] [PDF] |
||||
![]() |
D. McLerran, W. E. Grizzle, Z. Feng, W. L. Bigbee, L. L. Banez, L. H. Cazares, D. W. Chan, J. Diaz, E. Izbicka, J. Kagan, et al. Analytical Validation of Serum Proteomic Profiling for Diagnosis of Prostate Cancer: Sources of Sample Bias Clin. Chem., January 1, 2008; 54(1): 44 - 52. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. McLerran, W. E. Grizzle, Z. Feng, I. M. Thompson, W. L. Bigbee, L. H. Cazares, D. W. Chan, J. Dahlgren, J. Diaz, J. Kagan, et al. SELDI-TOF MS Whole Serum Proteomic Profiling with IMAC Surface Does Not Reliably Detect Prostate Cancer Clin. Chem., January 1, 2008; 54(1): 53 - 60. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Pelikan, W. L. Bigbee, D. Malehorn, J. Lyons-Weiler, and M. Hauskrecht Intersession reproducibility of mass spectrometry profiles and its effect on accuracy of multivariate classification models Bioinformatics, November 15, 2007; 23(22): 3065 - 3072. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Seam, D. A. Gonzales, S. J. Kern, G. L. Hortin, G. T. Hoehn, and A. F. Suffredini Quality Control of Serum Albumin Depletion for Proteomic Analysis Clin. Chem., November 1, 2007; 53(11): 1915 - 1920. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Machaalani, M. Arlotto, K. A. Waters, E. Gozal, F. Berger, and M. Dematteis A Novel Method of Tissue Collection and Storage: Validation Using SELDI-TOF MS Analysis Clin. Chem., July 1, 2007; 53(7): 1387 - 1389. [Full Text] [PDF] |
||||
![]() |
J. Albrethsen Reproducibility in Protein Profiling by MALDI-TOF Mass Spectrometry Clin. Chem., May 1, 2007; 53(5): 852 - 858. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. M Thompson, M S. Lucia, and C. M Tangen Commentary: The ubiquity of prostate cancer: echoes of the past, implications for the present Int. J. Epidemiol., April 23, 2007; (2007) dym056v1. [Full Text] [PDF] |
||||
![]() |
M. Gross, I. Top, I. Laux, J. Katz, J. Curran, C. Tindell, and D. Agus {beta}-2-Microglobulin Is an Androgen-Regulated Secreted Protein Elevated in Serum of Patients with Advanced Prostate Cancer Clin. Cancer Res., April 1, 2007; 13(7): 1979 - 1986. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. C.S. Cho, T. T.C. Yip, R. K.C. Ngan, T.-T. Yip, V. N. Podust, C. Yip, H. H.Y. Yiu, V. Yip, W.-W. Cheng, V. W.S. Ma, et al. ProteinChip Array Profiling for Identification of Disease- and Chemotherapy-Associated Biomarkers of Nasopharyngeal Carcinoma Clin. Chem., February 1, 2007; 53(2): 241 - 250. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Aivado, D. Spentzos, U. Germing, G. Alterovitz, X.-Y. Meng, F. Grall, A. A. N. Giagounidis, G. Klement, U. Steidl, H. H. Otu, et al. From the cover: Serum proteome profiling detects myelodysplastic syndromes and identifies CXC chemokine ligands 4 and 7 as markers for advanced disease PNAS, January 23, 2007; 104(4): 1307 - 1312. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Mosley, F. W. K. Tam, R. J. Edwards, J. Crozier, C. D. Pusey, and L. Lightstone Urinary proteomic profiles distinguish between active and inactive lupus nephritis Rheumatology, December 1, 2006; 45(12): 1497 - 1504. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Nedelkov, U. A. Kiernan, E. E. Niederkofler, K. A. Tubbs, and R. W. Nelson Population Proteomics: The Concept, Attributes, and Potential for Cancer Biomarker Research Mol. Cell. Proteomics, October 1, 2006; 5(10): 1811 - 1818. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Drake, E. E. Schwegler, G. Malik, J. Diaz, T. Block, A. Mehta, and O. J. Semmes Lectin Capture Strategies Combined with Mass Spectrometry for the Discovery of Serum Glycoprotein Biomarkers Mol. Cell. Proteomics, October 1, 2006; 5(10): 1957 - 1967. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Cornett, J. A. Mobley, E. C. Dias, M. Andersson, C. L. Arteaga, M. E. Sanders, and R. M. Caprioli A Novel Histology-directed Strategy for MALDI-MS Tissue Profiling That Improves Throughput and Cellular Specificity in Human Breast Cancer Mol. Cell. Proteomics, October 1, 2006; 5(10): 1975 - 1983. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. L. Hortin The MALDI-TOF Mass Spectrometric View of the Plasma Proteome and Peptidome Clin. Chem., July 1, 2006; 52(7): 1223 - 1237. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. L. Hortin, S. A. Jortani, J. C. Ritchie Jr, R. Valdes Jr, and D. W. Chan Proteomics: A New Diagnostic Frontier Clin. Chem., July 1, 2006; 52(7): 1218 - 1222. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. P. Diamandis Serum Proteomic Profiling by Matrix-Assisted Laser Desorption-Ionization Time-of-Flight Mass Spectrometry for Cancer Diagnosis: Next Steps Cancer Res., June 1, 2006; 66(11): 5540 - 5541. [Full Text] [PDF] |
||||
![]() |
C. Agostini and M. Facco The promising future of proteomics in sarcoidosis. Am. J. Respir. Crit. Care Med., May 15, 2006; 173(10): 1053 - 1054. [Full Text] [PDF] |
||||
![]() |
E. Kriegova, C. Melle, V. Kolek, B. Hutyrova, F. Mrazek, A. Bleul, R. M. du Bois, F. von Eggeling, and M. Petrek Protein Profiles of Bronchoalveolar Lavage Fluid from Patients with Pulmonary Sarcoidosis Am. J. Respir. Crit. Care Med., May 15, 2006; 173(10): 1145 - 1154. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Jimeno and M. Hidalgo Molecular biomarkers: their increasing role in the diagnosis, characterization, and therapy guidance in pancreatic cancer. Mol. Cancer Ther., April 1, 2006; 5(4): 787 - 796. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Ernst, C. Melle, B. Schimmel, A. Bleul, and F. von Eggeling Proteohistography--Direct Analysis of Tissue with High Sensitivity and High Spatial Resolution Using ProteinChip Technology J. Histochem. Cytochem., January 1, 2006; 54(1): 13 - 17. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. Lin and W. A. Kibbe Irrational Exuberance in Clinical Proteomics Clin. Cancer Res., November 15, 2005; 11(22): 7963 - 7964. [Full Text] [PDF] |
||||
![]() |
H. W. Ressom, R. S. Varghese, M. Abdel-Hamid, S. A.-L. Eissa, D. Saha, L. Goldman, E. F. Petricoin, T. P. Conrads, T. D. Veenstra, C. A. Loffredo, et al. Analysis of mass spectral serum profiles for biomarker selection Bioinformatics, November 1, 2005; 21(21): 4039 - 4045. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. E. Banks, A. J. Stanley, D. A. Cairns, J. H. Barrett, P. Clarke, D. Thompson, and P. J. Selby Influences of Blood Sample Processing on Low-Molecular-Weight Proteome Identified by Surface-Enhanced Laser Desorption/Ionization Mass Spectrometry Clin. Chem., September 1, 2005; 51(9): 1637 - 1649. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Master Diagnostic Proteomics: Back to Basics? Clin. Chem., August 1, 2005; 51(8): 1333 - 1334. [Full Text] [PDF] |
||||
![]() |
A. Karsan, B. J. Eigl, S. Flibotte, K. Gelmon, P. Switzer, P. Hassell, D. Harrison, J. Law, M. Hayes, M. Stillwell, et al. Analytical and Preanalytical Biases in Serum Proteomic Pattern Analysis for Breast Cancer Diagnosis Clin. Chem., August 1, 2005; 51(8): 1525 - 1528. [Full Text] [PDF] |
||||
![]() |
N. Jeffries Algorithms for alignment of mass spectrometry proteomic data Bioinformatics, July 15, 2005; 21(14): 3066 - 3073. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Goggins Molecular Markers of Early Pancreatic Cancer J. Clin. Oncol., July 10, 2005; 23(20): 4524 - 4531. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Baumann, U. Ceglarek, G. M. Fiedler, J. Lembcke, A. Leichtle, and J. Thiery Standardized Approach to Proteome Profiling of Human Serum Based on Magnetic Bead Separation and Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry Clin. Chem., June 1, 2005; 51(6): 973 - 980. [Abstract] [Full Text] [PDF] |
||||
![]() |