Clinical Chemistry 51: 1830-1835, 2005; 10.1373/clinchem.2005.053595
Right arrow Help viewing high resolution images
Right arrow Return to article
Click here to download a slide of this image in PowerPoint format.
(Downloading may take up to 30 seconds.
If the slide opens in your browser, select File -> Save As to save it.)

Click on image to view larger version.



Figure 1. Bland–Altman plots comparing the two hCG methods (A) and the two FSH methods (B).

Results are shown as the ratio of the Roche Elecsys to the Bayer Centaur method (for hCG) or ratio of the Vitros ECi to the Bayer Centaur method (for FSH) vs the average hormone concentration. The dashed lines indicate the limits of acceptability defined as 2 SD of the mean ratio for hCG [mean (SD), 0.98 (0.30)] and FSH [1.05 (0.38)].