Methods: We developed a microfluidic flow cell consisting of a network of chambers and channels molded into a polydimethylsiloxane substrate. The substrate was aligned and reversibly bound to the microarray printed on a standard glass slide to form a functional microfluidic unit. The microfluidic units were placed on an engraved, disc-shaped support fixed on a rotational device. Centrifugal forces drove the sample and buffers directly onto the microarray surface.
Results: This microfluidic system increased the hybridization signal by
10fold compared with a passive system that made use of 10 times more sample. By means of a 15min automated hybridization process, performed at room temperature, we demonstrated the discrimination of 4 clinically relevant Staphylococcus species that differ by as little as a single-nucleotide polymorphism. This process included hybridization, washing, rinsing, and drying steps and did not require any purification of target nucleic acids. This platform was sensitive enough to detect 10 PCR-amplified bacterial genomes.
Conclusion: This removable microfluidic system for performing microarray hybridization on glass slides is promising for molecular diagnostics and gene profiling.
| Introduction |
|---|
|
|
|---|
18 h for detection of nucleic acids and are thus too slow for the rapid diagnosis of infectious diseases. Microfluidics is an emerging technology allowing the movement of minimum volumes in microscopic channels and chambers that are microfabricated in silicon, hard plastic, or soft elastomer polydimethylsiloxane (PDMS)1
(13). Fluid propulsion and controlled valves must be designed to allow sequential displacement of liquids into the desired channels and chambers (14). Microfluidic systems for nucleic acid hybridization with micropumps (15), pneumatic pumps (16), and syringe pumps (6) have been developed. However, these systems are complex, expensive, and require special procedures for arraying bioprobes and for detecting hybridization signals. In this study, we built and tested a removable microfluidic structure allowing DNA hybridization on a glass slide microarray. After hybridization, microarrays can be analyzed externally with a standard scanner based on confocal microscopy for glass slide microarray analysis. We show that this microfluidic device allows the differentiation of single-nucleotide polymorphisms (SNPs) to identify Staphylococcus species in 15 min. | Materials and Methods |
|---|
|
|
|---|
fabrication of the elastomeric flow cells
The microfluidic structures were fabricated using the PDMS replicating techniques reported by Duffy et al. (17). A wide variation of PDMS structures can be molded by use of microfabricated SU-8 micromolds (17)(18)(19)(20)(21). Two types of photoresist (i.e., SU-8 25 and SU-8 100) available from Microchem Inc. were used. SU-8 25 was used for the microchannel structures, and SU-8 100 was used for the reagent chambers. In the first step, SU-8 25 was processed on a 15.24-cm reclaimed Si wafer (Addison Engineering) to obtain the structures for the microchannels 25 µm in depth and the alignment marks for the second SU-8 25 layer. Subsequently, a thick layer (250 µm) of SU-8 100 was spin-coated over the substrate on which the molds for the microchannels had been created. This thicker layer was used to define the mold for the much larger reagent reservoirs. Because crosslinked SU-8 photoresists have lower optical transparency than their unexposed surroundings, the alignment marks can be observed readily, even when they are completely covered with a thick layer of the unexposed photoresist. In the pattern design, we compensated for possible alignment errors between the two layers of photoresist. The channels and chambers overlapped 50 µm in the connection areas to avoid possible disconnections caused by misalignment. Six identical molds were fabricated simultaneously on the 15.24-cm Si wafer for faster replication.
PDMS was purchased from Dow Corning. For polymerization molding of the flow cell, the base (Sylguard 184 silicone elastomer) and the curing agent were thoroughly mixed in a weight proportion of 10:1. Because of the thickness of the structures, low-temperature curing (i.e., 65 °C) in a convection oven was preferred over high-temperature baking. High temperature (e.g., 150 °C) causes substantial thermal stress at the interface between the SU-8 patterns and the Si substrate that can actually crack the substrate and peel off the SU-8 structures. Leveling of the PDMS on the substrate is required to achieve a uniform thickness over the entire flow cell.
preparation of glass slides
The glass surface was functionalized based on the chemical reactions described by Joos et al. (22). All chemical reactions were carried out in polypropylene jars. Surfaces used were 25 x 75-mm microscope glass slides (VWR International). After sonication (Branson 1210 ultrasonic cleaner; Branson Ultrasonics Corporation) for 1 h in deionized water, the slides were sonicated for 1 h in 40 mL of 100 g/L NaOH, washed several times with deionized water, and dried under a stream of nitrogen. The slides were then sonicated in an aminopropyltrimethoxysilane solution (2 mL of water, 38 mL of methanol, and 2 mL of aminopropyltrimethoxysilane) for 1 h, washed with methanol, dried, and baked for 15 min at 110 °C. The amine-modified slides were activated by sonication overnight in 40 mL of 1,4-dioxane containing 0.32 g (2 mmol) of carbonyldiimidazole as a coupling agent, washed with dioxane and diethyl ether, and dried under a stream of nitrogen.
microarray production
Microarrays were fabricated based on the method previously reported by Schena et al. (23). Oligonucleotide probes at 10 µmol/L in phosphate-buffered saline (pH 7.4; Sigma-Aldrich) supplemented with 1 mmol/L EDTA were diluted 2-fold by the addition of Array-it Microspotting Solution Plus (Telechem International). Capture probes were spotted in duplicate with a Virtek SDDC-2 arrayer (Bio-Rad Laboratories) equiped with SMP2 pins (Telechem International). Each spot had a volume of 0.6 nL and a diameter of 6080 µm. Subsequently, the slides were dried overnight, washed by immersion in boiling 0.1% Igepal CA-630 for 5 min, rinsed in boiling ultrapure water for 5 min, and dried by centrifugation for 5 min under reduced pressure (SpeedVac plus from Thermo Savant). The slides were stored at room temperature in a dry and oxygen-free environment.
pcr amplification and amplicon labeling
Amplicons of 368 bp were generated either by standard PCR or by asymmetrical PCR amplification of purified staphylococcal genomic DNA using the Staphylococcus-specific primers described previously (24). Genomic DNAs were purified from strains S. aureus ATCC 43300, S. epidermidis ATCC 14990, S. haemolyticus ATCC 29970, and S. saprophyticus ATCC 35552, as described previously (24). Fluorescent Cy dyes were incorporated during asymmetric PCR amplification (25). Cy-labeled dUTP nucleotides (Amersham Biosciences) were used at a concentration of 0.02 mM in a 25-µL PCR mixture containing 0.02 mM dATP, 0.05 mM dCTP, 0.05 mM dGTP, 0.05 mM dTTP, 5 mM KCl, 1 mM Tris-HCl (pH 9), 0.1 mL/L Triton X-100, 2.5 mM MgCl2, 0.5 U of Taq DNA polymerase (Promega), 0.2 µM primer TstaG765, primer TstagG422 at 0.2 µM for standard PCR or at 0.005 µM for asymmetric PCR, and 1 x 10-4 to 1 ng of purified staphylococcal genomic DNA (equivalent of 10 to 1 x 105 genome copies per PCR reaction). Thermal cycling for PCR amplification (180 s at 94 °C followed by 40 cycles of 5 s at 95 °C, 30 s at 55 °C, and 30 s at 72 °C) was carried out on a PTC-200 DNA Engine thermocycler (MJ Research). For detection limit assays, PCR amplifications were performed using the equivalent of 1, 10, 100, 1000, and 10 000 genome copies purified from S. aureus strain ATCC 43300 to evaluate the minimum number of bacterial genome copies that can be detected by the microfluidic platform. Approximately 9 Cy-labeled nucleotides per amplicon were incorporated during PCR.
dna microarray hybridization and data acquisition
We mixed 5 µL of amplified PCR reaction mixture containing the Cy-labeled PCR amplicons with 15 µL of hybridization buffer (8x SSPE (OmniPur; EM Sciences), 0.4 g/L polyvinylpyrrolidone and 400 mL/L formamide). Passive hybridization was performed in self-sticking, 20 µL, 15 x 13-mm Hybri-well hybridization chambers (Sigma-Aldrich). Amplicons produced by standard PCR were denatured by heating at 95 °C for 5 min. Hybridization buffer containing the labeled sample was introduced into the chambers, and hybridization was conducted for 5 min at room temperature. After hybridization, the microarrays were washed with 2x SSPE containing 1 g/L sodium dodecyl sulfate for 5 min at room temperature and rinsed once with 2x SSPE for 5 min. The microarrays were then dried by centrifugation at 1350g for 3 min. The slides were then scanned using a ScanArray 4000XL (Packard Bioscience Biochip Technologies), and fluorescent signals were analyzed using its QuantArrayTM software.
For flow-through hybridization, a unit consisting of a glass slide and our homemade flow cell was used. This unit was placed on a prototype plastic disc support, and the whole platform was fixed on the hub of a motor controlled by a computer (Fig. 1C
). The labeled sample was prepared in the same way as for passive hybridization. Sample (2 µL) and 10 µL of washing and rinsing buffer were loaded on the microfluidic unit just before spinning the disc. The disc was spun at different speeds to sequentially burst the centrifugal valves and to allow the sample (12g), washing buffer (44g), and rinsing buffer (50g) to flow through the hybridization chamber. The disc was spun at high speed (100g) for 1 min to dry the slide and then scanned as described above for passive hybridization.
|
| Results |
|---|
|
|
|---|
In the flow-through system, the total amount of targets transported to the spot from the bulk solution was approximated by diffusion layer theory (29). Under laminar flow conditions, when a sample solution flows over the capture arrays, the target DNA in a layer close to the spot surface will be depleted during the hybridization reaction. Close to the surface of the chamber, the velocity of the fluid approaches zero and can be regarded as forming a diffusion boundary layer having a thickness inversely proportional to the cubic root of the stream velocity (29). Within this layer, depletion to some degree is expected despite the mass transport caused by convection (30). The "diffusion layer thickness" can be estimated from Eq. 1
. The equation is derived from a case of diffusion layer inside a circular tube (29) and has been adapted for laminar flow between two parallel plates of infinite width shown in Fig. 2
:
![]() | (1) |
|
Total accumulation of captured DNA molecules on the surface over a certain amount of time (which is proportional to the inverse of the maximum flow velocity when the sample volume is constant) is then obtained from Eq. 2
, which was adapted for our self-contained flow cells from the equation for flux in a circular tube (29):
![]() | (2) |
For passive hybridization, the accumulation of targets on the hybridization surface can be predicted by Eq. 3
(31):
![]() | (3) |
flow-through hybridization in 15 min
The flow-through hybridization unit combines a PDMS flow cell juxtaposed with a glass slide a few minutes before the experiment without performing any surface treatment or adhesion step. This microfluidic system allows robust control of valve opening to sequentially release the contents of different chambers. Indeed, Fig. 3
shows that the bursting range in rotation per min for each of the 3 centrifugal valves of our CD microfluidic system do not overlap.
|
Each staphylococcal amplicon at 10 nmol/L generated by asymmetric PCR was hybridized by both passive and flow-through hybridizations. For flow-through hybridization, loading of the reagents was performed immediately before spinning the disc platform to avoid reagent evaporation. A spin speed was selected to obtain a sample flow rate of 0.4 µL/min in the hybridization chamber, which corresponds to a hybridization time of 5 min considering that 2 µL of sample was loaded on the microfluidic unit. This hybridization time is identical to the time used in the passive hybridization experiments. After the hybridization step, the spin speed of the platform was increased to sequentially burst the centrifugation valves, releasing 10 µL of washing buffer and 10 µL of rinsing buffer, respectively, into the hybridization chamber. These two buffers flowed through the hybridization chamber with a mean flow rate of 2.2 µL/min in
9 min. Finally, there was a 1-min drying step (high spin speed). The total time for the entire hybridization process was therefore
15 min. Subsequently, the PDMS microfluidic flow cells were pealed off, and the hybridized microarrays were scanned. Plotting of the fluorescence intensity revealed that flow-through hybridization in a 140-nL chamber was more sensitive than passive hybridization in the 20-µL chamber. Five min of hybridization with 10 nmol/L of amplicon generated from S. aureus, S. epidermidis, S. haemolyticus, and S. saprophyticus showed ratios of 9.5, 13.5, 18.7, and 6.9, respectively, between flow-through and passive hybridizations (Fig. 4
). Hybridizations of amplicons generated by standard PCR amplification from the equivalent of 1, 10, 100, 1000, or 10 000 genome copies were performed with the flow-through hybridization system. We found that the equivalent of as little as 10 genome copies of starting material was sufficient to produce an unambiguous hybridization signal (Fig. 5
).
|
|
differentiation of 4 clinically important STAPHYLOCOCCUS species
Staphylococcus-specific PCR primers targeting the tuf gene were used to amplify a 368-bp fragment from S. aureus, S. epidermidis, S. haemolyticus, and S. saprophyticus purified genomic DNAs. Species-specific capture probes targeting these 4 staphylococcal species were arrayed on glass slides and hybridized with the 4 different staphylococcal amplicons. The results demonstrated that it was possible to detect and differentiate the 4 different staphylococcal tuf amplicons (Fig. 6
). The S. epidermidis-specific oligonucleotide probe was designed in a specific area of its genome that differs from the S. aureus sequence by only a a SNP. A nucleotide analog was added at a strategic location in the S. epidermidis probe to make it more discriminative for the S. aureus amplicon. With this strategy, hybridization of S. aureus amplicons gave a fluorescence signal
6 times stronger with the S. aureus probe compared with the S. epidermidis probe (Fig. 6
). The other oligonucleotide capture probes had at least 3 nucleotide mismatches with the nonhomologous Staphylococcus amplicons, and no substantial cross-hybridization was observed. In addition, a mixture of the 4 Staphylococcus amplicons (each at 1 nmol/L), hybridized on the same microarray by our microfluidic system, showed fluorescence hybridization signals with intensities similar to those shown in Fig. 6
for each of the 4 species-specific capture probes.
|
| Discussion |
|---|
|
|
|---|
This study reports the merging of standard microarray glass slide technology with a simple, low-cost microfluidic device. We demonstrated that nanoliter volumes of liquid can be moved precisely into channels and chambers on the glass slide surface created with a microfluidic elastomeric flow cell juxtaposed above the slide. This custom microarray hybridization microfluidic platform is easy to use, automated, and rapid. It uses standard glass slides compatible with commercial arrayers and scanners found in most academic departments. In this removable microfluidic system, the hybridization chamber is composed of a microfluidic network engrafted on a disposable, low-cost elastomeric material. This elastomeric material reversibly sticks without any adhesive or chemical reaction to the glass slide, forming together the microfluidic unit. Placed on a plastic CD-like support, the microfluidic units are spun at different speeds to control fluid movements. To simplify hybridization experiments using this device, buffer compositions and capture probe sequences are optimized to be compatible with room temperature hybridizations to avoid the need for a heating device. Furthermore, this microfluidic system allows a drastic reduction in the volume of reagents needed for microarray hybridizations and does not require a PCR amplicon purification step, which may be time-consuming.
Among various pumping methods attempted in research of flow-through DNA chips, one is electroosmotic pumping (EOP) (35)(36)(37)(38)(39)(40). However, DNA has a high negative electrophoretic mobility because of the large number of negative charges carried by its phosphate groups at most pH values. Consequently, EOP of DNA requires a high electroosmotic mobility buffer to overcome the negative electrophoretic mobility of the DNA molecules. Furthermore, DNA hybridization buffers often contain high concentrations of salt that reduce the electroosmosis effect and make the EOP a less effective approach (16). Pumping with mechanical pressure presents some advantages over the EOP approach as it is insensitive to pH, to the charge of moving molecules, and to salt concentration. However, high back pressures can be generated because of the high flow resistance caused by the small dimensions of the microchannels. Consequently, leakage is often a problem if the unit is not sealed. With the CD platform, the reagents are delivered by centrifugal force generated over the entire length of the liquid element. Therefore, local high pressure is avoided, and as a result, interface sealing is readily accomplished. In fact, no leakage was observed with this approach. In the present study, PDMS was selected to make the microchambers and channels. PDMS is a low-cost material that can be easily prototyped and can make reversible and watertight seals with glass slides. Each microfluidic unit, composed of the molded PDMS juxtaposed on a glass slide, was placed in a custom-made plastic disc support (Fig. 1
). Centripetal force was used to move the liquid into the microfluidic chambers and channels as described previously by us (41)(42). The rapidity of the hybridization reactions prevents reagent evaporation problems that may be encountered in slower standard microarray hybridization methods.
In a passive hybridization system, a hybridization event requiring collision between a capture probe and the analyte relies solely on diffusion. In such systems, sensitivity is increased by means of long hybridization periods (43)(44). One advantage of flow-through hybridization is that, because of the shallow architecture of the microchannels, the probability of collision between the probe and the analyte is increased by the much shorter diffusion distance. This allows fluid movement of nanoliter volumes over the capture probes, thereby accelerating the hybridization kinetics (44)(45)(46). Using a microfluidic system, Chung et al. (47) showed a 6-fold increase in hybridization efficiency with the flow-through hybridization performed in a 33-µL chamber compared with the passive hybridization performed in a chamber of the same volume. In our work, we increased even further the kinetics of hybridization with a much smaller hybridization chamber (140 nL) combined with flow-through hybridization. Flow-through hybridization in such a small hybridization chamber allowed a substantial reduction in the reagent and sample volumes (1/10) while reducing hybridization time to 5 min. This rapid hybridization by means of our microfluidic system increased the kinetics of hybridization by an average of 10.5-fold compared with passive hybridization using 368-bp PCR amplicons as nucleic acid targets. By contrast, 30 min was required for hybridization in the flow-through hybridization system described by Chung et al. (47). Furthermore, there is no need to purify the target PCR amplicons before hybridization, thereby reducing the overall time of the assay.
To be useful for the diagnosis of infectious diseases in clinical laboratories, a molecular test should be highly sensitive, specific, and ideally, rapid and inexpensive. Our system showed a detection limit of 500 amol of amplified target. This detection limit is comparable to those obtained with microfluidic devices that are more complex to manufacture (mainly irreversibly bounded plastic devices actuated with active valves) and requiring much longer hybridization times (6)(48). One system using chemiluminescence has a detection limit of 250 amol, but requires a 3-h hybridization period (49), too long for practical use in clinical diagnostics (50). Here we report a microfluidic system that allows detection of PCR amplicons generated from the amplification of 10 bacterial genome copies, which is at least 1000 times more sensitive than results obtained by other groups using microarray hybridization on microfluidic devices (15). The addition of an amplicon concentration step, as well as the use of a microfluidic device for faster PCR thermocycling, should allow detection of even lower copy numbers.
In terms of specificity, this simple system differentiated 4 different Staphylococcus species with a post-PCR hybridization protocol of only 15 min. An artificial mismatch created by the addition of an inosine was incorporated in the S. epidermidis capture probe to further destabilize binding events with the S. aureus amplicons differing by only SNP. The nucleotide analog was located 4 nucleotides upstream from the SNP present in the S. aureus genomic DNA sequence. Our actual device design can accommodate up to 150 spots in the 140-nL hybridization chamber. This should be sufficient for some practical applications in the diagnostic field, but future improvements in spot density and hybridization chamber geometry will substantialtly increase the number of spots in the microfluidic hybridization chamber.
In conclusion, this microfluidic microarray system could become a tool of choice for the rapid identification of nucleic acid sequences in fields such as molecular diagnostics, detection of bioterrorism agents, food quality survey, and forensic analysis. This prototype paves the way toward the development of micro-total analysis system (µ-TAS) on a CD support, where sample preparation, PCR amplification, and microarray hybridization and detection could be accomplished with a single portable device for point-of-care diagnostic applications.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
K. Boissinot, A. Huletsky, R. Peytavi, S. Turcotte, V. Veillette, M. Boissinot, F. J. Picard, E. A. Martel, and M. G. Bergeron Rapid exonuclease digestion of pcr-amplified targets for improved microarray hybridization. Clin. Chem., November 1, 2007; 53(11): 2020 - 2023. [Full Text] [PDF] |
||||
![]() |
D. J. Carter and R. B. Cary Lateral flow microarrays: a novel platform for rapid nucleic acid detection based on miniaturized lateral flow chromatography Nucleic Acids Res., May 11, 2007; 35(10): e74 - e74. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |