Methods: Immunoassays were performed on serum supplemented with Silwet L-720 on the IMMULITETM 2500 and AxSYMTM analyzers. Direct effects of the surfactant on the chemiluminescent detection step of immunoassays and on antibody immobilization on the solid phase were examined.
Results: Increasing the final surfactant concentration from 0 to 400 mg/L in serum significantly increased (
51%) the apparent total triiodothyronine (TT3) concentrations measured on the IMMULITE 2500 but not the AxSYM analyzer. Several other competitive, but not noncompetitive, assays were also significantly affected by the surfactant on the IMMULITE 2500 analyzer. The effect was independent of serum components, and the surfactant had no direct effect on chemiluminescence reactions. The capture antibody, however, was displaced from the solid phase by incubation with solutions containing surfactant under conditions similar to the IMMULITE TT3 assay.
Conclusions: The Silwet L-720 surfactant, which is used to coat the inner surfaces of tubes, appears to account for previously reported immunoassay interference by BD Vacutainer SST blood collection tubes. One of the mechanisms for the interference is the desorption of antibodies from the solid phase by the surfactant. The results identify an important factor in the selection of suitable blood collection tube surfactants and provide an approach for solving similar tube-assay interference problems in the future.
| Introduction |
|---|
|
|
|---|
In the present study, we performed a series of experiments to determine the specific mechanism underlying the positive interference with serum collected in BD blood collection tubes for TT3 measured on the IMMULITE 2500 analyzer. Silwet L-720 was positively identified as the source of assay interference from the BD collection tubes, and the mechanism appears to be a result of the desorption of capture antibodies from the polystyrene (PS) solid-support beads by the surfactant.
| Materials and Methods |
|---|
|
|
|---|
Blood samples were drawn after written informed consent was obtained from apparently healthy volunteers (age range, 3060 years; 10 males and 10 females) by trained technologists using a butterfly connected to a evacuated tube holder. The glass tubes were filled to capacity and inverted 5 times after blood draw to promote clotting. Serum was obtained after clotting for 30 min at room temperature, followed by centrifugation at 2000g for 5 min, which is a deviation from the manufacturers recommendations of <1300g for 10 min. All specimens were processed within 2 h of collection. Serum was transferred with disposable plastic Pasteur pipettes into a 500-mL plastic container and kept covered at room temperature until testing within 3 h, or was stored at 4 °C between testing intervals for up to 3 days. All experiments involving the surfactant were done by adding the surfactant to a serum pool.
laboratory analysis
Surfactant solutions.
The polydimethylsiloxane surfactant Silwet L-720 (GE Silicone/OSI Specialties) was diluted in methanol to concentrations of 0, 1, 2, 4, 8, 12, 16, and 20 g/L. The Silwet L-720 concentrations used in this study are similar to the polydimethylsiloxane surfactant concentrations described by Anraku and Shoji (2) in the original evacuated blood collection tube patent.
Immunoassays.
Serum pool TT3 concentrations from glass tubes, with different concentrations of surfactant, were measured in duplicate in random order on an IMMULITE 2500 analyzer (Diagnostic Products Corporation) and an AxSYMTM analyzer (Abbott Laboratories). Three different reagent lots and one calibrator lot were used for the IMMULITE 2500 analyzer during the study, and two reagent lots and one calibrator lot were used for the AxSYM analyzer. Both the IMMULITE 2500 and AxSYM TT3 assays gave satisfactory internal and external quality-control results during the study and showed similar intra- and interassay precision, as described previously (8).
The serum pool with different concentrations of surfactant was also analyzed on the IMMULITE 2500 analyzer for thyroid-stimulating hormone [thyrotropin (TSH)], and sex hormone-binding globulin (SHBG) by an immunometric assay, with labeled detection antibody and capture antibodies immobilized on PS beads. Total thyroxine (TT4), cortisol, progesterone, and thyroxine-binding globulin (TBG) were also measured on the IMMULITE 2500 analyzer but by competitive immunoassays using limiting amounts of immobilized antibodies and labeled hormones. The serum pool samples were analyzed in duplicate in random order. All of the noncompetitive and competitive immunoassays described above were from the same reagent and calibrator lots.
testing of various surfactant concentrations on measurements with immunoassays
To test the effects of various surfactant concentrations on TT3, progesterone, cortisol, TT4, TSH, SHBG, and TBG results, we added 40 µL of surfactant solutions (0, 1, 2, 4, 8, 12, 16, and 20 g/L) to nonevacuated plain plastic test tubes with no additives (BD Labware) and left them overnight at room temperature to evaporate the methanol (n = 6 per surfactant concentration). We then added 2 mL of a serum pool from glass blood collection tubes to each test tube and vortex-mixed the tubes to give final surfactant concentrations of 0, 20, 40, 80, 160, 240, 320, and 400 mg/L, which span the concentration of surfactant used in the patent description of evacuated blood collection tubes (2). The serum pool samples were assayed in duplicate for TT3 on both the IMMULITE 2500 and AxSYM analyzers, whereas progesterone, cortisol, TT4, TSH, SHBG, and TBG were assayed in singleton on the IMMULITE 2500 analyzer. Various concentrations of surfactant, as described above, were also added to phosphate-buffered saline (PBS) to determine the effect of surfactant concentrations, in the absence of serum components, on the chemiluminescent signal generation with the IMMULITE 2500 TT3 assay. It is important to note that the addition of surfactant to the tubes does not necessarily reflect the silicone coating procedure used by BD.
study of surfactant on t3-alkaline phosphatase conjugate activity
To study potential T3-alkaline phosphatase (ALP) activity changes with increasing surfactant concentrations, 010 µL of a 2 g/L surfactant solutions were added to each plain plastic test tube and left overnight at room temperature to evaporate the methanol (n = 6 for each surfactant concentration). After the tubes had dried, 50 µL of T3-ALP conjugate from the reagent wedge (lot L2T3A2211) of an IMMULITE 2500 TT3 assay reagent set was added to each test tube, vortex-mixed, and incubated at 37 °C for 30 min with intermittent agitation. The final surfactant concentration ranged from 0 to 400 mg/L. ALP activity was measured by the addition of 200 µL of a chemiluminescent substrate, a phosphate ester of adamantyl dioxetane from the IMMULITE 2500 TT3 assay (18)(19). The sustained emission of light from the hydrolysis of the chemiluminescent substrate was photomultiplied and quantified for 8 s at 477 nm with continuous tracing on a LUMATTM LB 9507 luminometer (EG and G Berthold). Changes in relative light units were measured and compared with that exhibited by the blank tube containing no surfactant after subtraction from background emissions.
effect of pretreatment of ps beads with surfactant on serum tt3 measurements
To determine whether surfactant directly interferes with the PS beads and/or the murine anti-T3 antibodies on the PS beads, we added 40 µL of surfactant solutions (0, 2, 8, and 20 g/L) to each plain plastic test tube and left it overnight at room temperature to evaporate the methanol. One PS bead coated with anti-T3 antibodies (lot L2T31221) from an IMMULITE TT3 assay reagent set and 2 mL of multidiluent 1 (lot L2M1Z0109) were added to each test tube, vortex-mixed, and incubated at room temperature for 30 min (n = 6 per surfactant concentration). The final concentrations of surfactant incubated with the PS beads were 0, 40, 160, and 400 mg/L. The beads were separated from the liquid phase by decantation, air-dried for 30 min, and transferred to a clean plastic test tube. These beads were then carefully added to an empty TT3 bead pack and used to assay a serum pool from glass tubes for TT3 in duplicate on the IMMULITE 2500 analyzer. For a negative control, a sham experiment was also performed in which the PS beads underwent the same conditions as described above but with no pretreatment with surfactant or methanol solutions (n = 6). The serum pool was also analyzed as usual for TT3 in duplicate on the IMMULITE 2500 analyzer (n = 6).
sodium dodecyl sulfatepolyacrylamide gel electrophoresis and immunoblotting of surfactant solutions after incubation with ps beads
To determine whether increasing concentration of surfactant desorbs anti-T3 antibodies from the surface of the PS bead, we suspended the PS beads in 300 µL of PBS containing various concentrations of surfactant and incubated the solutions for 30 min at 37 °C to mimic the IMMULITE TT3 assay conditions. Three different bead solutions were pooled for each surfactant concentration. PS beads incubated with 300 µL of a 10 g/L sodium dodecyl sulfate (SDS) solution in PBS for 5 min at 95 °C were used as positive control for the Western blot analysis. Before loading on gels, we concentrated the samples by collecting the retentate from a MicronTM 10 concentrator (Millipore). We then mixed 15-µL aliquots of the controls, PBS-surfactant solutions, and molecular markers (6250 kDa; NOVEXTM) with 5 µL of 4x NuPAGETM lithium dodecyl sulfate sample buffer and 1 µL of 2-ß-mercaptoethanol and heated the mixture at 70 °C for 5 min. These aliquots were then loaded on a NuPAGE precast Tris-glycine gradient gel (4%20% acrylamide; Invitrogen) and electrophoresed in NuPAGE MOPS buffer at 90 V for 1.5 h. Proteins in the gel were electrophoretically transferred to a nitrocellulose membrane in a transfer buffer (12 mmol/L Tris and 96 mmol/L glycine in 200 mL/L methanol) in a NOVEX XCell II transfer apparatus. The transfer was conducted at 25 V for 2.5 h. The efficiency of the protein transfer was confirmed by staining with GelCode Blue reagent (Pierce Biotechnology). The membrane was then blocked with 50 g/L nonfat dried milk in PBS containing 1 mL/L Tween 20 for 1 h at room temperature. After the membrane was washed 3 times in PBS-Tween 20 (5 min each wash at room temperature), the antibodies desorbed from the PS beads by surfactant were detected by an enhanced chemiluminescent Western blot detection system, with a 1:250 dilution of sheep anti-mouse IgG antibodies conjugated with horseradish peroxidase and luminol as a substrate (Amersham Biosciences). The membrane was enclosed in a piece of Saran Wrap and exposed to Kodak X-OMATTM AR x-ray film in an x-ray film cassette. Detection of positive signals on the x-ray film occurred in 7 min. The exposed x-ray film was developed in a Kodak RPX-OMAT automated photo image detection system.
statistical analyses
The means of duplicates from each specimen for TT3 results were used for statistical analysis. All other serum analytes from each specimen were analyzed in singletons. The results are reported as mean (SD). Results for all measured analytes obtained were compared by the 2-tailed Student t-test or ANOVA. Statistical significance was defined as P <0.05. Statistical analyses were performed with StatViewTM software (Ver. 5.0; SAS Institute).
| Results |
|---|
|
|
|---|
|
effect of surfactant on other immunoassays
To determine whether other assays on the IMMULITE 2500 analyzer were also affected by the surfactant, we selected assays that were either unaffected or affected by BD blood collection tubes from our previous study (8). As shown in panels A through D of Fig. 2
, the competitive assays for progesterone (P = 0.04), cortisol (P = 0.0003), TT4 (P <0.0001), and TBG (P = 0.0002), when measured on the IMMULITE 2500 analyzer, were affected by increasing concentration of surfactant. In contrast, as shown in panels E and F of Fig. 2
, the noncompetitive assays for SHBG and TSH were not affected by the surfactant (P >0.05).
|
effect of surfactant on chemiluminescent signal
To rule out a possible interaction of the surfactant with a component of the serum as the source of the IMMULITE TT3 assay interference, we mixed various concentrations of surfactant with PBS. As shown in Fig. 3
, increasing the concentration of surfactant from 0 to 400 mg/L in PBS produced a significant decrease in the binding of the tracer, and therefore, the chemiluminescent signal (relative light units) decreased from 3.1 x 107 at 0 mg/L (maximum binding of T3-ALP conjugate to PS bead) to 2.5 x 107 at 400 mg/L Silwet L-720 surfactant in the IMMULITE TT3 assay (P <0.0001). Thus, the interference with the surfactant happens in the absence of serum and appears to be attributable to a direct effect on the assay.
|
effect of surfactant on t3-alp conjugate activity
To determine whether the observed decrease in the chemiluminescent signal in the presence of the surfactant was attributable to alterations in signal generation, we incubated increasing concentrations of Silwet L-720 with T3-ALP conjugate for 30 min at 37 °C and measured the chemiluminescent signal intensity on a luminometer. As illustrated in Fig. 4
, increasing surfactant concentrations did not significantly alter the kinetics or magnitude of chemiluminescent responses from hydrolysis of adamantyl 1,2-dioxetane phosphate by ALP (P = 0.89). Thus, these findings demonstrate that the surfactant does not directly affect the activity of the T3-ALP conjugate or the generation of the chemiluminescent signal.
|
effect of pretreatment of ps beads with surfactant on serum tt3 measurements
We next examined the effect of Silwet L-720 on the PS solid support used in the IMMULITE 2500 analyzer. Pretreatment of PS beads with increasing concentrations of Silwet L-720 surfactant for 30 min at room temperature markedly increased the apparent TT3 concentration in a serum pool from 2.65 nmol/L at 0 mg/L to a peak of 3.79 nmol/L at 160 mg/L and 3.36 nmol/L at 400 mg/L surfactant (Fig. 5
). By unpaired t-test, the apparent mean serum TT3 concentrations from PS beads incubated with 40, 160, and 400 mg/L Silwet L-720 surfactant were all significantly higher than that of a serum pool from glass tubes (Fig. 5
). The results obtained with untreated beads (sham) and 0 mg/L Silwet L-720 surfactant were not significantly different from the results obtained with the serum pool (P >0.05; Fig. 5
). Hence, pretreatment of PS beads with Silwet L-720 surfactant falsely increased the TT3 values compared with nonpretreated beads, which suggests that the surfactant may be interfering with the assay either by binding to the PS beads and/or interfering with the anti-T3 capture antibody.
|
effect of surfactant on desorption of capture antibody from ps solid-support beads
To determine whether Silwet L-720 surfactant desorbs anti-T3 antibodies from PS beads, we performed SDS polyacrylamide gel electrophoresis and immunoblotting on solutions that were incubated at 37 °C for 30 min with PS beads and different concentrations of surfactant. As expected, a 50-kDa mouse IgG heavy chain band was detected when the beads were treated with 10 g/L SDS (Fig. 6
, lane 1) but not in the empty control lane (Fig. 6
, lane 2). No bands were evident when the PS beads were incubated with the PBS negative control solution (Fig. 6
, lane 3). Bands corresponding to the heavy chain of mouse IgG were, however, detected after treatment of the beads with 20 (lane 4), 40 (lane 5), or 240 (lane 6) mg/L Silwet L-720 surfactant, respectively. These results indicate that Silwet L-720 surfactant can desorb capture antibodies from the surface of the PS beads supplied with the IMMULITE 2500 TT3 assay, which would be predicted to lead to a falsely increased result for competitive-type assays.
|
| Discussion |
|---|
|
|
|---|
ALP, the reporter enzyme used in the IMMULITE 2500 analyzer, is a metalloenzyme that requires for activity the binding of zinc to the active site as well as other divalent cations, such as magnesium (25). In the case of IMMULITE 2500, a chemiluminescent substrate is used for ALP. It is well known that various components of serum or urine, even in minute amounts, can interfere significantly with photon emission in chemiluminescent reactions (26). Results from this study (Fig. 4
) reveal, however, that the surfactant had no effect on the ALP reporter enzyme, which was not completely unexpected because our previous study had shown that only a small number of IMMULITE 2000 assays (progesterone, TT3, cortisol, TBG, TT4, insulin) gave a positive bias ranging from 11.9% to 38.6% with BD SST compared with glass blood collection tubes (8). In addition, ALP is frequently used in many other immunoassay systems, such as the AxSYM, which was not affected by the interference with the BD blood collection tubes (8).
The adsorption of proteins on PS surfaces has been studied extensively (27). The findings from these studies have revealed that the PS surface characteristics such as hydrophobicity, surface charge, and coadsorption of or exchange with surfactants, copolymers, and other proteins are important factors for the strength of the noncovalent binding of proteins. These factors are also well known to be particularly important to the stability and specificity of antibodies in immunoassays (28). It is also evident that the amount adsorbed, the orientation, and the conformation of antibodies on the PS beads is very important to the performance of immunoassays (28). It is conceivable that the surfactant may also affect the assay by one of the other potential mechanisms, such as by denaturation of the anti-T3 antibodies on the surface of the PS bead (29). Previous studies have shown that polydimethylsiloxanes, with their lipophilic qualities, can attract and denature or induce conformational changes in proteins (30)(31)(32)(33). Thus, any component in the serum that interferes with the interaction of antibodies with the surface of the PS bead may affect the performance of immunoassays. Because of the competitive format for the TT3 assay, any of the above-hypothesized mechanisms assay will lead to a decrease in T3-ALP conjugate binding and chemiluminescent signal and, subsequently, falsely increase the measured serum TT3 concentration.
As shown in Fig. 6
, increasing the concentration of surfactant was found to desorb mouse anti-T3 IgG on the surface of the PS beads. Surfactants are known to adsorb strongly to hydrophobic surfaces such as the PS beads, and they potentially can displace larger, less hydrophobic molecules, such as antibodies (29)(34)(35). In addition to Silwet L-720, we also tested the detergents Tween 20 and SDS on the TT3 assay on the IMMULITE 2500 analyzer and found no effect with Tween 20 but a marked increase in TT3 with SDS. (data not shown). This is consistent with the common use of Tween 20 as a mild surfactant in the wash solutions of ELISAs and with the dissociation of the capture antibody observed with SDS in Fig. 6
. Several competitive and noncompetitive immunoassays that were either unaffected or affected by collection of serum into BD tubes from our previous study (8) were examined for the effect of the Silwet L-720 surfactant (Fig. 2
). The immunoassays that showed a significant positive bias with BD blood collection tubes (progesterone, cortisol, TT4, and TBG) in our previous study demonstrated a significant increase with increasing concentrations of surfactant added to a serum pool from glass tubes (Fig. 2
, A, B, C, and D, respectively). The immunoassays that were unaffected by the BD blood collection tubes (SHBG and TSH) in our previous study did not show a significant increase with increasing concentrations of surfactant added to serum pools (Fig. 2, E and F
, respectively). The reason for the difference in the response to the surfactant by the different assays is most likely related to their different formats. Noncompetitive assays use excess capture and detector antibodies; therefore, they would be expected to be less affected by the partial desorption from the solid support or possible denaturation of antibodies. The reason for the various degrees of interference at the same surfactant concentration in the competitive assays may be attributable to differences in the strength of noncovalent binding of the antibodies to the different types of solid phase used for each type of assay. In addition, the sample volumes differ considerably among assays, which would change the final concentration of surfactant for the different immunoassays.
In conclusion, the present study shows a significant effect of the Silwet L-720 surfactant in BD blood collection tubes on the IMMULITE 2500 TT3 and other immunoassays. The surfactant present in the serum promoted the desorption of anti-T3 antibodies from the surface of the PS beads, thus lowering the chemiluminescent signal and falsely increasing the TT3 concentration. Although they are potentially an important source of preanalytical errors, detection and prevention of interferences from blood collection tube additives is a challenging problem for most clinical laboratories. In addition to performing a careful comparison study in which they change tubes, clinical laboratorians should have a heightened awareness of this problem and should be vigilant for any possible changes in assay performance with new lots of tubes and/or any other types of tube changes. Results from this study also reveal that the identification of suitable tube surfactants should involve assessment of the ability of the surfactant to cause desorption and/or possible denaturation of antibodies from the various solid supports used for immunoassays. Improving the affinity of antibodies to solid supports, or perhaps covalent attachment of antibodies, could also minimize this type of problem in the future and may make immunoassays more robust to other types of interferences.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
G. Dimeski and A. Carter Magnesium contamination from terumo blood collection tubes. Clin. Chem., August 1, 2006; 52(8): 1612 - 1614. [Full Text] [PDF] |
||||
![]() |
A. K. Stankovic and G. Parmar Assay interferences from blood collection tubes: a cautionary note. Clin. Chem., August 1, 2006; 52(8): 1627 - 1628. [Full Text] [PDF] |
||||
![]() |
S. Wang, V. Ho, A. Roquemore-Goins, and F. A. Smith Effects of Blood Collection Tubes, including Pediatric Devices, on 16 Common Immunoassays. Clin. Chem., May 1, 2006; 52(5): 892 - 893. [Full Text] [PDF] |
||||
![]() |
L. J. Kricka, J. Y. Park, M. B. Senior, and R. Fontanilla Processing Controls in Blood Collection Tubes Reveals Interference Clin. Chem., December 1, 2005; 51(12): 2422 - 2423. [Full Text] [PDF] |
||||
![]() |
L. J. Kricka and J. Y. Park Additive-Aggravated Assays: An Authoritative Answer Clin. Chem., October 1, 2005; 51(10): 1767 - 1767. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |