Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 51: 2103-2109, 2005. First published August 25, 2005; 10.1373/clinchem.2005.053835
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow A correction has been published
Right arrow All Versions of this Article:
clinchem.2005.053835v1
51/11/2103    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (17)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Windelberg, A.
Right arrow Articles by Ueland, P. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Windelberg, A.
Right arrow Articles by Ueland, P. M.
Related Collections
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2005;51:2103-2109.)
© 2005 American Association for Clinical Chemistry, Inc.


Automation and Analytical Techniques

Automated Assay for the Determination of Methylmalonic Acid, Total Homocysteine, and Related Amino Acids in Human Serum or Plasma by Means of Methylchloroformate Derivatization and Gas Chromatography–Mass Spectrometry

Amrei Windelberga, Ove Årseth, Gry Kvalheim and Per Magne Ueland

1 LOCUS for Homocysteine and Related Vitamins, and Section for Pharmacology, Institute of Medicine, University of Bergen, Bergen, Norway.

aAddress correspondence to this author at: LOCUS for Homocysteine and Related Vitamins, Section for Pharmacology, Institute of Medicine, University of Bergen, Armauer Hansens Hus, N-5021 Bergen, Norway. Fax 47-55-97-4605; e-mail amrei.windelberg{at}farm.uib.no.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Background: The combined measurement of methylmalonic acid (MMA) and total homocysteine (tHcy) in serum or plasma is useful in diagnosing and distinguishing between cobalamin and folate deficiencies. We developed and validated an isotope-dilution gas chromatography–mass spectrometry (GC-MS) method with automated sample workup for the determination of MMA, tHcy, and the related amino acids Met, total cysteine (tCys), Ser, and Gly in serum or plasma.

Methods: Serum or plasma samples (100 µL) were treated with a reductant (dithioerythritol), deproteinized with ethanol, and derivatized and extracted in a single step by the addition of methylchloroformate and toluene. All liquid handling was performed in 96-well (1 mL) microtiter plates by a robotic workstation. The N(S)-methoxycarbonyl ethyl ester derivatives were analyzed by GC-MS in the selected-ion monitoring mode.

Results: Detection limits (signal-to-noise ratio, 5:1) were between 0.03 µmol/L (MMA) and 10 µmol/L (Ser, tCys). The assay was linear to 100 µmol/L for MMA and tHcy and to 1000 µmol/L for Met, tCys, Ser, and Gly. The within-day CVs ranged from 0.7% to 3.6% (n = 20), and the between-day CVs from 2.1% to 8.1% (n = 20). The recovery was between 79% and 99% for the different analytes.

Conclusion: This assay combines a simple and automated sample preparation with selective and sensitive GC-MS analysis and is well suited for the combined measurement of MMA, tHcy, and the related amino acids.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Methylmalonic acid (MMA)1 and total homocysteine (tHcy) in serum or plasma increase in cobalamin-deficient individuals and are clinically useful indicators of cobalamin status. tHcy is also increased in folate deficiency and is used as an indicator of this deficiency (1). tHcy is thus a less specific indicator of vitamin B12 function than MMA, also because the concentration of tHcy is influenced by diverse genetic and lifestyle factors and disease states(2). In addition, an increased tHcy concentration increases the risk of cardiovascular disease(3).

Numerous tHcy assays have been published, most of which are based on liquid or gas chromatography (4)(5)(6). More recently, homogeneous immunologic tHcy assays adapted to commercial platforms have gained widespread use(7)(8). The determination of MMA in serum or plasma has been accomplished by gas chromatography–mass spectrometry (GC-MS)(9)(10)(11)(12)(13), capillary electrophoresis with laser-induced fluorescence detection(14), and liquid chromatography–tandem mass spectrometry(15)(16).

Simultaneous determination of MMA and tHcy in the same serum or plasma specimen is practical because of the complementarity of these 2 analytes in the diagnosis of cobalamin and folate deficiencies. The inclusion of other amino acids related to Hcy metabolism, such as Met, total cysteine (tCys), Ser, and Gly, in such an assay may give additional information about metabolic abnormalities influencing the concentration of tHcy. Such multiplexing could be achieved by GC after alkylchloroformate derivatization, which allows simultaneous esterification of carboxylic groups and acylation of amino and thiol groups, creating N(S)-alkoxycarbonyl alkyl ester derivatives. This strategy was proposed by Husek (17)(18), who introduced alkylchloroformates as general derivatizing reagents in GC.

Methods based on chloroformate derivatization for the determination of tHcy (19)(20)(21)(22)(23)(24)(25)(26) or MMA(27) have been published. Most of them use ethylchloroformate(20)(22)(23)(25)(26)(27), but other chloroformates have been used as well(19)(21)(24). The advantages include fast and simple sample preparation in an aqueous medium at room temperature. The performance of the plasma tHcy methods based on chloroformate derivatization has been published previously(19)(20)(21)(22)(23)(24)(25)(26), whereas no such method has been optimized for the quantification of MMA in serum or plasma.

Here we describe an automated isotope-dilution GC-MS assay for the simultaneous determination of MMA, tHcy, and the related amino acids Met, tCys, Ser, and Gly based on derivatization with methylchloroformate (MCF). All liquid handling was performed in 96-well microtiter plates to enable high-throughput sample preparation.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
materials
L-Cysteine (Cys), L-Ser, D,L-dithioerythritol, pyridine, and toluene were obtained from Fluka. Gly, L-homocystine, and L-Met were from Sigma Chemical Co. MMA and the derivatization reagent, MCF, were purchased from Aldrich. D,L-d3-Ser and D,L-d2-Cys were obtained from Medical Isotope; d2-Gly, D,L-d8-homocystine, and d3-MMA were from Cambridge Isotope Laboratories; and D,L-d4-Met was from C/D/N Isotopes. All other chemicals and solvents were of the highest purity available. Human serum for assay calibration and validation experiments was obtained by pooling surplus sera from routine testing. For precision studies, the pooled serum was treated with mercaptoethanol (20 mmol/L) and then dialyzed 4 times against 10 volumes of phosphate-buffered saline (137 mmol/L NaCl, 2.7 mmol/L KCl, 2.9 mmol/L KH2PO4, 14 mmol/L Na2HPO4) containing 4 mmol/L EDTA.

sample collection
EDTA-plasma was obtained by collecting blood into Vacutainer Tubes (Becton Dickinson); the final EDTA concentration in the samples was 4 mmol/L. The EDTA-blood samples were centrifuged within 60 min. Serum was obtained by collecting blood into Vacutainer Tubes with no additive. Blood was allowed to clot at room temperature for 30 min before isolation of the serum fraction. Plasma and serum were stored at –20 °C until use.

sample preparation
For routine measurements, sample preparation and all liquid handling were performed in 96-well (1 mL) microtiter plates by a robotic workstation (ATplus 2; Hamilton). Serum or plasma samples (100 µL) were transferred from sample tubes into the microtiter plates, where they were mixed with 25 µL of D,L-dithioerythritol (200 mmol/L) containing d8-homocystine and d2-Cys (30 µmol/L and 1 mmol/L, respectively) and incubated at room temperature for 20 min to reduce the disulfide bonds of Hcy and Cys. Samples were then deproteinized by adding 450 µL of ethanol containing the remaining deuterated internal standards (2.6 µmol/L d3-MMA, 5.6 µmol/L d4-Met, 22 µmol/L d2-Gly, and 44 µmol/L d3-Ser). Aliquots (380 µL) of the supernatant obtained after centrifugation (3 min at 5800g) were transferred into an empty microtiter plate and mixed with 300 mL/L of water and 50 µL of pyridine; 250 µL of 200 mL/L in toluene was then added. Mixing was achieved by repeated pipetting. After incubation at room temperature for 6 min to obtain phase separation, 500 µL of the aqueous phase was replaced by water (450 µL), and the samples were mixed again. Most of the lower aqueous phase was removed before the plates were manually sealed and transferred to the sample tray of the autoinjector. An aliquot (1.5 µL) of the toluene layer was used for GC-MS analysis.

gc-ms
A Thermo Finnigan trace GC ultra system coupled to a Fisons MD800 mass spectrometer was used in the electron ionization mode. The analytes were separated on a CP Sil 24-CB low-bleed/MS capillary column from Varian [15 m x 0.25 mm (i.d.); film thickness, 0.25 µm]. Samples were injected in the splitless mode, and the oven temperature program was as follows. The initial temperature of 75 °C was increased at a rate of 45 °C/min to 85 °C, which was maintained for 1 min, and then increased at 30 °C/min to 125 °C and further at 120 °C/min to 290 °C. This temperature was maintained for 2 min. Helium was used as carrier gas and delivered at a flow rate that was increased from 1.1 to 2.2 mL/min during the run. The interface temperature was 250 °C, the source temperature was 200 °C, and the electron energy was 70 eV. The analytes were quantified in serum or plasma by measuring the area ratios of analyte vs deuterated internal standard and comparing these ratios with the area ratios obtained from external standards with known analyte concentrations.

assay calibration
The assay was calibrated by replicate measurements of an external standard, which consisted of pooled serum with known concentrations of all analytes, and an assay blank of phosphate-buffered saline. Additional serum and plasma with known metabolite concentrations were used as quality controls. The concentrations of the calibrators were determined either by established methods [tHcy by HPLC (28) and MMA by capillary electrophoresis(14)] or by an amino acid standard solution (containing 2.5 mmol/L Gly, 2.5 mmol/L Ser, 2.5 mmol/L Met, and 2.5 mmol/L Cys in 0.1 mol/L hydrochloric acid) obtained from Sigma.

linearity and limit of detection
The linear range and the limit of detection (LOD) of the assay were determined by adding MMA, Hcy, Gly, Ser, Met, and Cys at concentrations from 0.003 to 1000 µmol/L to phosphate-buffered saline. The peak-area ratios of the analytes to their deuterated forms were plotted against their concentrations. The LOD was defined as a signal-to-noise ratio of 5:1.

recovery
Pooled serum was divided into 3 portions, and 2 concentrations (medium and high; Table 1 ) of all analytes were added to 2 portions. At each analyte concentration, 10 replicates were analyzed in 1 run. The recovery (percentage) was calculated as:


View this table:
[in this window]
[in a new window]
 
Table 1. Analytical recovery of the assay.1

precision
Dialyzed serum was divided into 3 portions that were supplemented with low (0.05 µmol/L MMA, 3.3 µmol/L Hcy, 10 µmol/L Met, 40 µmol/L Ser, 60 µmol/L Gly, and 60 µmol/L Cys), medium (0.15 µmol/L MMA, 10 µmol/L Hcy, 30 µmol/L Met, 120 µmol/L Ser, 180 µmol/L Gly, and 180 µmol/L Cys), or high (0.45 µmol/L MMA, 30 µmol/L Hcy, 90 µmol/L Met, 360 µmol/L Ser, 540 µmol/L Gly, and 540 µmol/L Cys) concentrations of all analytes. Within-day precision was determined by assaying 20 replicates of each concentration on 1 day. Between-day precision was determined by assaying the same samples on 20 different days over a period of 6 weeks.

method comparison
Surplus plasmas from 50 routine determinations of tHcy by an immunologic assay (ADVIA Centaur®; Bayer) were reassayed by this GC-MS method. The tHcy concentrations ranged from 5 to 35 µmol/L in these samples.

The method described here was also validated as part of an external quality assessment program for MMA and tHcy determination (29). During a period of 16 months, 16 serum samples covering the concentration range 0.17–1.20 µmol/L MMA and 16 EDTA-plasma samples covering the concentration range 4.9–64.5 µmol/L tHcy were analyzed, and the individual results were compared with the mean concentrations obtained by all participating laboratories.

metabolites in plasma and serum from healthy individuals
Plasma and serum samples were obtained from 120 healthy blood donors (mean age, 43.3 years; range, 20–65 years). One half of the donors, 22 men and 38 women, were fasting; the other half, 42 men and 18 women, had eaten a light breakfast 2–3 h before blood sampling.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
gc-ms
The analytes showed retention times between 2.4 and 4.4 min and eluted in the following order: MMA (2.4 min), Gly (3.1 min), Ser (3.9 min), Met (4.0 min), tCys (4.2 min), and tHcy (4.4 min). The total run time was 6 min. The mass spectra of the MCF derivatives of all 6 analytes and their deuterated internal standards were obtained in the full-scan acquisition mode. For selected-ion monitoring analysis, the recorded ion pairs (labeled/unlabeled) were m/z 174/177 for MMA/d3-MMA, 161/163 for Gly/d2-Gly, 173/179 for Ser/d3-Ser, 235/239 for Met/d4-Met, 206/208 for Cys/d2-Cys, and 233/237 for Hcy/d4-Hcy. These ions were chosen on the basis of ion abundance, detection limit, and the absence of interfering material in biological matrices.

The high chromatographic resolution and separation from matrix components (Fig. 1 ) suggested no interference. Succinic acid had a retention time of 2.8 min and was clearly separated from MMA. We avoided peaks that interfere with the determination of Ser by monitoring different fragments for Ser (m/z 173) and its deuterated internal standard (m/z 179). Matrix effects were assessed by comparing the analytical recoveries of medium (0.15 µmol/L MMA, 10 µmol/L Hcy, 30 µmol/L Met, 120 µmol/L Ser, 180 µmol/L Gly, and 180 µmol/L Cys) analyte concentrations in phosphate-buffered saline, dialyzed serum, serum, and plasma. Recoveries were ≥80% for all analytes and all matrices, and the variation of recovery in the different matrices was between 6% and 15% for the different analytes.



View larger version (53K):
[in this window]
[in a new window]
 
Figure 1. GC-MS chromatogram for derivatized MMA, tHcy, and related amino acids and their internal standards in typical human plasma.

The selected ion traces for the analytes are shown in blue; the traces for the deuterated internal standards are shown in red. Different retention windows were used for each analyte and internal standard pair, and all peaks were normalized to the highest peak within each retention window.

To obtain adequate sensitivity, we divided the data acquisition into different retention windows so that only 2 ions were monitored simultaneously. A chromatogram of a plasma sample is shown in Fig. 1Up .

linearity and limit of detection
The LOD (signal-to-noise ratio, 5:1) were 0.03 µmol/L for MMA, 0.1 µmol/L for tHcy, 3 µmol/L for Gly, 1 µmol/L for Met, 10 µmol/L for Ser, and 10 µmol/L for tCys. Least-squares linear regression analysis of the peak-area ratios vs analyte concentration indicated that the assay was linear from the LOD to 100 µmol/L for MMA and tHcy and to 1000 µmol/L for Ser, Gly, Met, and tCys. The equations for the regression lines were as follows: y = 0.083x + 0.008 (r2 = 0.9999) for MMA; y = 0.060x + 0.005 for tHcy (r2 = 0.9999); y = 0.009x + 0.091 (r2 = 0.9989) for Gly; y = 0.002x + 0.012 (r2 = 0.9991) for Ser; y = 0.034x + 0.317 (r2 = 0.9987) for Met; and y = 0.003x + 0.052 for tCys (r2 = 0.9985; Fig. 2 ).



View larger version (22K):
[in this window]
[in a new window]
 
Figure 2. Linear dynamic ranges for derivatized MMA, tHcy, Ser, Gly, Met, and tCys.

The axes were log-transformed to cover the broad concentration ranges in one figure. The regression lines are calculated by relative least-squares linear regression. They are slightly curved because of the logarithmic scaling.

recovery and precision
Recoveries for all analytes were between 79% and 99% (Table 1Up ). The results of the precision studies are shown in Table 2 . Within-day CVs ranged from 0.7% to 3.6%, and between-day CVs ranged from 2.1% to 8.1%. At medium and high concentrations, the within-day CVs were <2.0% for all analytes, and the between-day CVs were <3.5% for all analytes except for tCys (6%).


View this table:
[in this window]
[in a new window]
 
Table 2. Precision of the assay.1

method comparison
Comparison of tHcy determined in 50 plasma samples with this GC-MS method and an immunologic assay (ADVIA Centaur®) showed good correlation. The equation for the regression line was as follows: y = 1.056x – 0.960 (r2 = 0.990; Sy|x = 0.650). Method comparison as part of the external quality assessment program gave a correlation of y = 0.918x + 0.013 (r2 = 0.998; Sy|x = 0.013) for MMA and y = 0.979x + 0.009 (r2 = 1.000; Sy|x = 0.370) for tHcy.

sample throughput and ruggedness
With automated sample preparation, 96 samples can be prepared in 1.5 h, and 288 samples can be prepared in 4 h. A single GC-MS instrument equipped with an autosampler can assay 130 samples in 24 h. The injection liner was changed every 1000 samples, and the column life exceeded 2000 injections. The ion sources of the mass spectrometers were cleaned every 2000 samples to maintain the sensitivity of the assay.

metabolites in plasma and serum from healthy individuals
The concentrations of MMA, tHcy, Gly, Ser, Met, and tCys were measured in both serum and EDTA plasma from fasting and nonfasting individuals. Median concentrations and 25th to 75th percentiles are shown in Table 3 . Median plasma concentrations in fasting individuals were 0.16 µmol/L for MMA, 9.2 µmol/L for tHcy, 230 µmol/L for Gly, 111 µmol/L for Ser, 26.5 µmol/L for Met, and 261 µmol/L for tCys. The serum concentrations of Ser and Gly were 21%–28% higher than those of plasma. For all other analytes, the serum samples contained only slightly higher amounts of metabolites (2%–7%) than the plasma samples. The difference between fasting and nonfasting metabolite concentrations was <10%. The measured concentrations were similar to those reported by others for MMA (30), tHcy(4), Gly(31), Ser(31), Met(30), and tCys(30) in healthy individuals.


View this table:
[in this window]
[in a new window]
 
Table 3. Concentrations of the analytes in plasma and serum from healthy donors.


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
This GC-MS method offers the combined measurement of MMA, tHcy, and the related amino acids Gly, Ser, Met, and tCys in plasma or serum. All sample preparation is adapted to 96-well microtiter plates, and the liquid handling is performed by a robotic workstation. The fast and simple derivatization with MCF combined with short retention times of the analytes ensures high sample throughput. The GC-MS analysis is characterized by selectivity, sufficient sensitivity, and good precision, with CVs often <3%.

In this assay, we derivatized the samples with MCF in an aqueous medium containing ethanol and measured the analytes as N(S)-methoxycarbonyl ethyl esters. Derivatization with ethylchloroformate in the presence of ethanol gave the same MMA derivative (m/z 174), whereas different derivatives were obtained for Hcy and the other amino acids. This is in agreement with the observation that the alkyl groups in the carbamate and thiocarbonate moieties of the derivatives are derived from the chloroformate, whereas the alkyl group in the ester moiety is derived from the alcohol. A suggested reaction mechanism for the ester formation is an alcohol exchange reaction between the alcohol and the mixed anhydride formed by a reaction between the alkylchloroformate and the carboxyl group (32).

Most assays for tHcy based on chloroformate derivatization use chloroformate and an alcohol with the same alkyl group, with ethylchloroformate and ethanol being a common combination (20)(23)(25)(26). We used the combination of MCF and ethanol because it gave superior assay performance in terms of retention times and resolution of the analytes. Ethanol was also used as a protein-precipitating agent in our method to avoid the addition of strong acids such as trichloroacetic acid, which may consume chloroformate and form pyridine salts, causing rapid deterioration of the column(33).

Column life exceeded 2000 injections under the conditions that we used, which makes column stability another advantage of our method. With another assay based on ethylchloroformate derivatization and chloroform extraction, we observed a column life of ~600 injections. We have not systematically investigated the reasons for the extended column life, which could be caused by the different derivatizing agent, the extraction solvent (toluene vs chloroform), or the temperature programming.

Chloroformate derivatization is usually followed by liquid–liquid extraction of the derivatives into chloroform (20)(21)(23)(24) or, recently, chloroform/isooctane(26); however, with the robotic pipetting device, the chloroform and aqueous phases were not mixed well enough in the microtiter plate to enable extraction of the derivatives into the organic solvent. This problem was solved by the use of toluene as the extraction solvent. We added toluene together with MCF, thereby combining the derivatization and extraction steps.

Two features of the design of this GC-MS method caused some concern. First, to liberate Hcy and Cys from their disulfides, we used D,L-dithioerythritol as reductant, which may consume chloroformate and cause assay interference (33); however, all selected-ion monitoring traces used to quantify the analytes were similar in the absence and presence of D,L-dithioerythritol. Furthermore, the use of deuterated internal standards in this method corrects for variable derivatization, which may occur if the amount of MCF becomes limiting. Second, for simplicity, our GC-MS method did not include a step for the removal of plasma lipids, which in other assays has been accomplished by hexane extraction(21)(34) or cation-exchange chromatography or solid-phase extraction(20)(23)(24)(35). Injection of plasma lipids may cause rapid column deterioration(33). In the present method, we increased the oven temperature (to 290 °C) at the end of each run to remove potentially adhesive material.

Despite the complementarity of tHcy and MMA determinations in the diagnosis of cobalamin and folate deficiencies, these 2 metabolites are usually measured by separate methods (10)(16)(30)(36)(37). Our method is optimized for the combined measurement of these 2 markers. Met, tCys, Ser, and Gly were also included in the assay because these amino acids are related to Hcy metabolism(38). Hcy is an intermediate in sulfur amino acid metabolism, and it is remethylated to Met by the folate- and cobalamin-dependent methionine synthase reaction or metabolized to Cys along the trans-sulfuration pathway(38). The major source of the folate-linked 1-carbon units needed for Hcy remethylation is the conversion of Ser to Gly by serine hydroxymethyltransferase. Ser also acts as a substrate in the trans-sulfuration pathway(39).

In conclusion, we have developed and validated a fully automated assay with high-throughput sample preparation for the combined determination of MMA, tHcy, and related amino acids in serum or plasma. Other attractive features of this method are low sample requirement (100 µL), simple derivatization in an aqueous medium at room temperature, selective and accurate quantification of the analytes with isotope dilution and selected-ion monitoring, and the adaptation of all steps of the method to a 96-well microtiter format. The method can easily be adapted to measure other carboxylic acids and amino acids. The present work may motivate further development of chloroformates as general derivatization agents.


   Acknowledgments
 
This work was funded by The Foundation to Promote Research into Functional Vitamin B12 Deficiency. We thank Pål Holm for supplying the serum and plasma samples from the fasting and nonfasting individuals.


   Footnotes
 
1 Nonstandard abbreviations: MMA, methylmalonic acid; tHcy, total homocysteine; GC-MS, gas chromatography–mass spectrometry; tCys, total cysteine; MCF, methylchloroformate; and LOD, limit of detection.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Allen RH, Stabler SP, Savage DG, Lindenbaum J. Diagnosis of cobalamin deficiency I: usefulness of serum methylmalonic acid and total homocysteine concentrations. Am J Hematol 1990;34:90-98.[ISI][Medline] [Order article via Infotrieve]
  2. Ueland PM, Refsum H, Schneede J. Determinants of plasma homocysteine. Robinson K eds. Homocysteine and vascular disease 2000:59-84 Kluwer Academic Publishers London. .
  3. Ueland PM, Refsum H, Beresford SAA, Vollset SE. The controversy over homocysteine and cardiovascular risk. Am J Clin Nutr 2000;72:324-332.[Abstract/Free Full Text]
  4. Ueland PM, Refsum H, Stabler SP, Malinow MR, Andersson A, Allen RH. Total homocysteine in plasma or serum: methods and clinical applications. Clin Chem 1993;39:1764-1779.[Abstract]
  5. Ubbink JB. Assay methods for the measurement of total homocyst(e)ine in plasma. Semin Thromb Hemost 2000;26:233-241.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  6. Rasmussen K, Moller J. Total homocysteine measurement in clinical practice. Ann Clin Biochem 2000;37:627-648.
  7. Nexo E, Engbaek F, Ueland PM, Westby C, O’Gorman P, Johnston C, et al. Evaluation of novel assays in clinical chemistry: quantification of plasma total homocysteine. Clin Chem 2000;46:1150-1156.[Abstract/Free Full Text]
  8. Ducros V, Demuth K, Sauvant MP, Quillard M, Causse E, Candito M, et al. Methods for homocysteine analysis and biological relevance of the results. J Chromatogr B Analyt Technol Biomed Life Sci 2002;781:207-226.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  9. Marcell PD, Stabler SP, Podell ER, Allen RH. Quantitation of methylmalonic acid and other dicarboxylic acids in normal serum and urine using capillary gas chromatography-mass spectrometry. Anal Biochem 1985;150:58-66.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  10. Rasmussen K. Solid-phase sample extraction for rapid determination of methylmalonic acid in serum and urine by a stable-isotope-dilution method. Clin Chem 1989;35:260-264.[Abstract/Free Full Text]
  11. Rifai N, Hagen T, Bradley L, Sakamoto M. Determination of serum physiological concentration of methylmalonic acid by gas chromatography-mass spectrometry with selected ion monitoring. Ann Clin Biochem 1998;35:633-636.
  12. Kushnir MM, Komaromy-Hiller G. Optimization and performance of a rapid gas chromatography-mass spectrometry analysis for methylmalonic acid determination in serum and plasma. J Chromatogr B Biomed Sci Appl 2000;741:231-241.[CrossRef][Medline] [Order article via Infotrieve]
  13. Yazdanpanah M, Chan PC, Evrovski J, Romaschin A, Cole DE. An improved assay for plasma methylmalonic acid using chemical ionization gas chromatography mass spectrometry. Clin Biochem 2003;36:617-620.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  14. Schneede J, Ueland PM. Application of capillary electrophoresis with laser-induced fluorescence detection for routine determination of methylmalonic acid in human serum. Anal Chem 1995;67:812-819.[Medline] [Order article via Infotrieve]
  15. Kushnir MM, Komaromy-Hiller G, Shushan B, Urry FM, Roberts WL. Analysis of dicarboxylic acids by tandem mass spectrometry. High-throughput quantitative measurement of methylmalonic acid in serum, plasma, and urine. Clin Chem 2001;47:1993-2002.[Abstract/Free Full Text]
  16. Magera MJ, Helgeson JK, Matern D, Rinaldo P. Methylmalonic acid measured in plasma and urine by stable-isotope dilution and electrospray tandem mass spectrometry. Clin Chem 2000;46:1804-1810.[Abstract/Free Full Text]
  17. Husek P. Amino acid derivatization and analysis in 5 minutes. FEBS Lett 1991;280:354-356.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  18. Husek P. Chloroformates in gas chromatography as general purpose derivatizing agents. J Chromatogr B Biomed Sci Appl 1998;717:57-91.[CrossRef][Medline] [Order article via Infotrieve]
  19. Kataoka H, Takagi K, Makita M. Determination of total plasma homocysteine and related aminothiols by gas chromatography with flame photometric detection. J Chromatogr B Biomed Appl 1995;664:421-425.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  20. Pietzsch J, Julius U, Hanefeld M. Rapid determination of total homocysteine in human plasma by using N(O,S)-ethoxycarbonyl ethyl ester derivatives and gas chromatography mass spectrometry. Clin Chem 1997;43:2001-2004.[Free Full Text]
  21. Sass JO, Endres W. Quantitation of total homocysteine in human plasma by derivatization to its N(O,S)-propoxycarbonyl propyl ester and gas chromatography-mass spectrometry analysis. J Chromatogr A 1997;776:342-347.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  22. Myung S-W, Kim M, Min H-K, Yoo E-A, Kim K-R. Determination of homocysteine and its related compounds by solid-phase microextraction-gas chromatography-mass spectrometry. J Chromatogr B Biomed Sci Appl 1999;727:1-8.[CrossRef][Medline] [Order article via Infotrieve]
  23. Sigit JI, Hages M, Brensing KA, Frotscher U, Pietrzik K, von Bergmann K, Lutjohann D. Total plasma homocysteine and related amino acids in end-stage renal disease (ESRD) patients measured by gas chromatography-mass spectrometry. Comparison with the Abbott IMx homocysteine assay and the HPLC method. Clin Chem Lab Med 2001;39:681-690.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  24. Shinohara Y, Hasegawa H, Tagoku K, Hashimoto T. Simultaneous determination of methionine and total homocysteine in human plasma by gas chromatography-mass spectrometry. J Chromatogr B Biomed Sci Appl 2001;758:283-288.[CrossRef][Medline] [Order article via Infotrieve]
  25. Haddad R, Mendes MA, Hoehr NF, Eberlin MN. Amino acid quantitation in aqueous matrices via trap and release membrane introduction mass spectrometry: homocysteine in human plasma. Analyst 2001;126:1212-1215.[CrossRef][Medline] [Order article via Infotrieve]
  26. Husek P, Matucha P, Vrankova A, Simek P. Simple plasma work-up for a fast chromatographic analysis of homocysteine, cysteine, methionine and aromatic amino acids. J Chromatogr B Analyt Technol Biomed Life Sci 2003;789:311-322.[ISI][Medline] [Order article via Infotrieve]
  27. Husek P, Liebich HM. Organic acid profiling by direct treatment of deproteinized plasma with ethyl chloroformate. J Chromatogr B Biomed Appl 1994;656:37-43.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  28. Fiskerstrand T, Refsum H, Kvalheim G, Ueland PM. Homocysteine and other thiols in plasma and urine: automated determination and sample stability. Clin Chem 1993;39:263-271.[Abstract]
  29. Moller J, Rasmussen K, Christensen L. External quality assessment of methylmalonic acid and total homocysteine. Clin Chem 1999;45:1536-1542.[Abstract/Free Full Text]
  30. Allen RH, Stabler SP, Savage DG, Lindenbaum J. Metabolic abnormalities in cobalamin (vitamin B12) and folate deficiency. FASEB J 1993;7:1344-1353.[Abstract]
  31. Altamura C, Maes M, Dai J, Meltzer HY. Plasma concentrations of excitatory amino acids, serine, glycine, taurine and histidine in major depression. Eur Neuropsychopharmacol 1995;5(Suppl 1):71-75.
  32. Wang J, Huang ZH, Gage DA, Watson JT. Analysis of amino acids by gas chromatography-flame ionization detection and gas chromatography mass spectrometry: simultaneous derivatization of functional groups by an aqueous-phase chloroformate-mediated reaction. J Chromatogr A 1994;663:71-78.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  33. Husek P. Determination of homocysteine by gas chromatography-mass spectrometry following treatment with chloroformates: a comment. J Chromatogr B Biomed Sci Appl 2000;740:289-290.[CrossRef][Medline] [Order article via Infotrieve]
  34. Husek P. Simultaneous profile analysis of plasma amino and organic acids by capillary gas-chromatography. J Chromatogr B Biomed Appl 1995;669:352-357.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  35. Husek P, Simek P. Advances in amino acid analysis. LC GC 2001;19:986-999.
  36. Moller J, Rasmussen K. Homocysteine in plasma: stabilization of blood samples with fluoride. Clin Chem 1995;41:758-759.[Free Full Text]
  37. Magera MJ, Lacey JM, Casetta B, Rinaldo P. Method for the determination of total homocysteine in plasma and urine by stable isotope dilution and electrospray tandem mass spectrometry. Clin Chem 1999;45:1517-1522.[Abstract/Free Full Text]
  38. Finkelstein JD. Methionine metabolism in mammals. J Nutr Biochem 1990;1:228-237.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  39. Wagner C. Biochemical role of folate in cellular metabolism. Bailey LB eds. Folate in health and disease 1995:23-42 Marcel Dekker New York. .



The following articles in journals at HighWire Press have cited this article:


Home page
JAMAHome page
M. Ebbing, O. Bleie, P. M. Ueland, J. E. Nordrehaug, D. W. Nilsen, S. E. Vollset, H. Refsum, E. K. Ringdal Pedersen, and O. Nygard
Mortality and Cardiovascular Events in Patients Treated With Homocysteine-Lowering B Vitamins After Coronary Angiography: A Randomized Controlled Trial
JAMA, August 20, 2008; 300(7): 795 - 804.
[Abstract] [Full Text] [PDF]


Home page
PediatricsHome page
A.-L. Bjorke-Monsen, I. Torsvik, H. Saetran, T. Markestad, and P. M. Ueland
Common Metabolic Profile in Infants Indicating Impaired Cobalamin Status Responds to Cobalamin Supplementation
Pediatrics, July 1, 2008; 122(1): 83 - 91.
[Abstract] [Full Text] [PDF]


Home page
Cancer Epidemiol. Biomarkers Prev.Home page
S. E. Vollset, J. Igland, M. Jenab, A. Fredriksen, K. Meyer, S. Eussen, H. K. Gjessing, P. M. Ueland, G. Pera, N. Sala, et al.
The Association of Gastric Cancer Risk with Plasma Folate, Cobalamin, and Methylenetetrahydrofolate Reductase Polymorphisms in the European Prospective Investigation into Cancer and Nutrition
Cancer Epidemiol. Biomarkers Prev., November 1, 2007; 16(11): 2416 - 2424.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
S. Sen, A. Hinderliter, P. K. Sen, J. Simmons, J. Beck, S. Offenbacher, E. M. Ohman, and S. M. Oppenheimer
Aortic Arch Atheroma Progression and Recurrent Vascular Events in Patients With Stroke or Transient Ischemic Attack
Circulation, August 21, 2007; 116(8): 928 - 935.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Clin. Nutr.Home page
O. Midttun, S. Hustad, J. Schneede, S. E Vollset, and P. M Ueland
Plasma vitamin B-6 forms and their relation to transsulfuration metabolites in a large, population-based study
Am. J. Clinical Nutrition, July 1, 2007; 86(1): 131 - 138.
[Abstract] [Full Text] [PDF]


Home page
JAMAHome page
B. F. Cole, J. A. Baron, R. S. Sandler, R. W. Haile, D. J. Ahnen, R. S. Bresalier, G. McKeown-Eyssen, R. W. Summers, R. I. Rothstein, C. A. Burke, et al.
Folic Acid for the Prevention of Colorectal Adenomas: A Randomized Clinical Trial
JAMA, June 6, 2007; 297(21): 2351 - 2359.
[Abstract] [Full Text] [PDF]


Home page
J. Clin. Endocrinol. Metab.Home page
P. I. Holm, S. Hustad, P. M. Ueland, S. E. Vollset, T. Grotmol, and J. Schneede
Modulation of the Homocysteine-Betaine Relationship by Methylenetetrahydrofolate Reductase 677 C->T Genotypes and B-Vitamin Status in a Large-Scale Epidemiological Study
J. Clin. Endocrinol. Metab., April 1, 2007; 92(4): 1535 - 1541.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Clin. Nutr.Home page
S. J Eussen, L. C de Groot, L. W Joosten, R. J Bloo, R. Clarke, P. M Ueland, J. Schneede, H. J Blom, W. H Hoefnagels, and W. A van Staveren
Effect of oral vitamin B-12 with or without folic acid on cognitive function in older people with mild vitamin B-12 deficiency: a randomized, placebo-controlled trial.
Am. J. Clinical Nutrition, August 1, 2006; 84(2): 361 - 370.
[Abstract] [Full Text] [PDF]


Home page
NEJMHome page
K. H. Bonaa, I. Njolstad, P. M. Ueland, H. Schirmer, A. Tverdal, T. Steigen, H. Wang, J. E. Nordrehaug, E. Arnesen, K. Rasmussen, et al.
Homocysteine Lowering and Cardiovascular Events after Acute Myocardial Infarction
N. Engl. J. Med., April 13, 2006; 354(15): 1578 - 1588.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow A correction has been published
Right arrow All Versions of this Article:
clinchem.2005.053835v1
51/11/2103    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (17)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Windelberg, A.
Right arrow Articles by Ueland, P. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Windelberg, A.
Right arrow Articles by Ueland, P. M.
Related Collections
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS