Clinical Chemistry 51: 368-375, 2005. First published November 24, 2004; 10.1373/clinchem.2004.036665
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow All Versions of this Article:
clinchem.2004.036665v1
51/2/368    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via ISI Web of Science (6)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Du, H.
Right arrow Articles by Cheng, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Du, H.
Right arrow Articles by Cheng, J.
Related Collections
Right arrow General Clinical Chemistry
Right arrow Proteomics and Protein Markers
Right arrow Drug Monitoring and Toxicology
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2005;51:368-375.)
© 2005 American Association for Clinical Chemistry, Inc.


Drug Monitoring and Toxicology

Development of Miniaturized Competitive Immunoassays on a Protein Chip as a Screening Tool for Drugs

Hongwu Du1,5,6, Moutian Wu2, Weiping Yang3, Gu Yuan4, Yimin Sun5, Yuan Lu5, Shan Zhao5, Qingyun Du3, Jun Wang4, Sheng Yang2, Mangen Pan5, Ying Lu5, Shan Wang2 and Jing Cheng1,5,6,a

1 Department of Biological Sciences and Biotechnology and 2 Institute of Biomedicine, Tsinghua University, Beijing, Peoples Republic of China.
3 China Doping Control Center, Beijing, Peoples Republic of China.
4 Aviva Antibody Corporation, San Diego, CA.
5 Key Laboratory of Bioorganic Chemistry and Molecular Engineering of Ministry of Education, College of Chemistry, Peking University, Beijing, Peoples Republic of China.
6 National Engineering Research Center for Beijing Biochip Technology, Beijing, Peoples Republic of China.

aAddress correspondence to this author at: Department of Biological Sciences and Biotechnology, Tsinghua University, Beijing, 100084, Peoples Republic of China. Fax 86-10-62773059; e-mail jcheng{at}tsinghua.edu.cn.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Background: Doping in sports has become a serious problem. Gas chromatography–mass spectrometry (GC-MS) serves as an effective reference method, but it is limited by low throughput and is therefore not suitable for large-scale screening. Use of protein chips for high-throughput screening of all athletes for prohibited substances could become an important complementary tool to GC-MS.

Methods: We developed a protein chip based on an aldehyde-activated glass slide containing 10 physically isolated arrays. The chip was used to screen urine from 1347 athletes for prohibited substances and to screen a negative control group consisting of 200 females and 120 males. Urine samples from 66 individuals known to be abusers, provided by the China Doping Control Center (CDCC), and 129 standard prohibited substances were tested as positive controls.

Results: All 1347 urine samples screened by means of the protein chips were also subjected to reference analysis by GC-MS at the CDCC. There was no qualitative difference between the results obtained with the two methods. The correlation coefficient (r2) for the quantitative results obtained with the protein chip and GC-MS was 0.991.

Conclusions: The protein chip could be used to screen for a series of 16 prohibited drugs in urine samples. This system has the potential to become an effective screening method to test substances prohibited by the International Olympic Committee.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Cheating in sports through use of prohibited substances has become increasingly common, and drug-abusing athletes are endeavoring to be one step ahead of the testing authorities. To minimize doping in sports and to promote health, fairness, and equality for athletes worldwide, the World Anti-Doping Agency (WADA)1 has initiated a series of efforts to combat this problem. Despite impressive progress made in detecting doping agents, many believe that the current methods alone cannot be effective in preventing doping in sports (1). In fact, no reliable method is available to detect certain substances, such as human growth hormone and other substances with similar chemical structures (2)(3)(4)(5).

A change in attitude toward doping, with systematic screening for all athletes in major sporting events, may be an effective way to combat the problem. It should be an indispensable part of the Olympic Games (6)(7), and indeed, the number of athletes screened for doping in recent Olympic Games, such as the Sydney and Salt Lake City games, increased significantly (8)(9).

There are ~140 substances that have been used for doping (10). They range from small synthetic chemical molecules to peptides such as insulin and erythropoietin. The list of prohibited substances, issued by WADA, increases with pharmaceutical developments each year. Because of the vast differences in the physical and chemical characteristics of the prohibited substances, it remains a challenge to find a general method that can detect them all in a single assay.

A combination of gas chromatography (GC) and mass spectrometry (MS) is still the most reliable drug testing method and has been used since the 1972 Olympic Games (11)(12). This method can detect and differentiate minute amounts of analytes. However, its usage as a screening tool is limited because of the cost of the instrument and the lengthy procedures required for sample preparation. The use of this method is further restricted by the limited differential power of the molecular weights of analytes. As a result, the GC-MS method is reserved as a reference tool for low-molecular-weight drugs that may be abused by winners of sporting events (13).

Biochip technology is one of the fastest growing fields in the analytical sciences (14)(15). It is becoming a method of choice for simultaneous and high-throughput measurement of multiple analytes by use of nonporous and porous solid supports combined with fluorescence-based detection (16)(17). Biochip immunoassays promise to revolutionize both clinical diagnosis (18) and drug testing (19) in situations where there is a need to simultaneously detect multiple analytes. We have developed a novel protein chip system based on a competitive immunoassay design for the simultaneous detection of up to 16 prohibited substances in a single drop of urine. In this assay, anti-drug antibody competes for drug immobilized on the protein chip and drug present in the sample. Anti-drug antibody bound to the slide is then quantified by use of a Cy3-labeled second antibody (Fig. 1 ), and the entire process can be completed in 2 h. This method is highly sensitive and reproducible, and for routine applications one technician can process hundreds of samples per day.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 1. Schematic of the competitive immunoassay principle used to test for drugs on the protein chip.

Ab, antibody.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
samples from athletes
Urine samples (from 557 female and 790 male athletes) were collected between June 2002 and March 2003. All urine samples were stored at –20 °C. The negative control group consisted of urine samples provided by 200 female and 120 male athletes and had been confirmed as drug free by GC-MS. Sixty-six confirmed positive urine samples (21 female and 45 male; 30 positive for anabolic steroids, 24 positive for stimulants, and 12 positive for narcotics) were collected by the China Doping Control Center (CDCC) between November 1982 and December 2002.

antibodies and reagents
Twenty-three conjugated antigens and their corresponding antibodies were purchased from Fitzgerald Industries International, Inc. Six antibodies were produced by Aviva Antibody Corporation. Five drug–bovine serum albumin (BSA) conjugates were synthesized at the Key Laboratory of Bioorganic Chemistry and Molecular Engineering, Peking University, China. Cy3-labeled goat anti-mouse IgG (H+L; 1.0 mg/vial; product code PA43002; absorbance maximum, 550 nm; emission maximum, 570 nm) was purchased from Amersham Pharmacia Biotech. The CDCC provided 129 standard drugs, including 50 anabolic steroids, 68 stimulants, and 11 narcotics. All test substances were chosen from the 2004 Prohibited List published by the World Anti-Doping Agency (10). The antibodies used for the protein chip assays are summarized in Table 1 .


View this table:
[in this window]
[in a new window]
 
Table 1. Sources of antibodies used on the protein chip (n = 29).

preparation of protein chip substrates
Glass slides chemically modified with aldehyde groups were used as the substrate to covalently bind BSA-conjugated molecules at the designated locations. The slides were cleaned with 100 g/L chromic acid for 6 h, followed by rinsing with deionized water. Slides were then dipped into a 2 mol/L sodium hydroxide solution and then 4 mol/L hydrochloric acid, each for 30 min, followed by rinsing with deionized water and then drying under a stream of nitrogen. Cleaned slides were silanized for 8 h using 3-glycidoxypropyltrimethoxysilane in ethanol (40 mL/L). The glass surface was washed with toluene, acetone, and deionized water, after which the slides were dipped in 4 mol/L hydrochloric acid again for 30 min and then immersed into 50 mmol/L NaIO4 for 1 h to complete the preparation process. The contact angles of the aldehyde-activated slides were measured by use of a contact angle system (Model OCA; DataPhysics Instruments GmbH) for quality-control purposes. Slides were stored in a desiccated box at room temperature for a maximum of 3 months.

printing of protein chips
Ten 9 x 9 arrays of BSA-conjugated drugs were printed on each slide. For a peptide hormone, the peptide was printed directly. On each slide, one sample can be tested on one 9 x 9 array for a variety of analytes, and up to 10 samples can be analyzed in parallel on one protein chip. A contact printing robot (PixSys 5500; Cartesian Technologies) with a stealth microspotting pin (Model SMP3; TeleChem International) was used to print the protein chips on the aldehyde-activated slides. The concentration each the printed protein (drug-BSA) was 500 mg/L in 400 mL/L glycerol or Protein Printing Buffer (TeleChem International). The drug-BSA conjugate was reacted on the protein chip for 6 h in a humidified chamber. The slide was then stored at room temperature for up to 1 month.

immunoassay procedure
A competitive immunoassay design was used to test the 16 WADA-prohibited substances on the protein chips. A molded polyester frame was attached to the substrate to partition 10 arrays on the protein chip surface (Fig. 2A ). This protein chip consisted of 16 different drug-BSA conjugates and 11 positive or negative controls to form a 9 x 9 array. Each material was printed in triplicate (Fig. 2 , B and C). The protein chips were immersed in blocking solution [a 1:10 dilution of sheep serum in phosphate-buffered saline (PBS), pH 7.4] for 30 min at room temperature and then rinsed three times with PBS containing 0.5 mL/L Tween 20, pH 7.4 (PBS-Tween A). A mixture of the anti-drug mouse monoclonal antibodies and a urine sample containing the drug was then applied to the gridded reaction chamber formed by the polyester frame covering the surface of the protein chip. The protein chip was then maintained at 37 °C in a humidified chamber for 30 min. The protein chip was then rinsed three times with PBS-Tween A, and the secondary antibody (Cy3-labeled goat anti-mouse IgG) was applied to the protein chip and incubated at 37 °C for 30 min. The protein chip was then washed again and scanned for the presence of bound Cy3-labeled secondary antibody by use of a laser confocal scanner (GenePix 4000B; Axon Instruments) or a charge-coupled device-based scanner (EcoScan-100; CapitalBio Corporation). The analog fluorescent signal was converted to digital signal by data analysis software (GenePix Pro 4.0; Axon Instruments). The results obtained from the protein chip were later compared with those obtained by GC-MS at CDCC.



View larger version (58K):
[in this window]
[in a new window]
 
Figure 2. Typical appearance and layout of the arrays.

(A), schematic showing the glass slide and the polyester frame forming the reaction chambers above each array. (B) image of one of the 9 x 9 arrays. (C), layout of the array (each drug/substance is printed in triplicate).

PCP, phencyclidine; TCA, tricyclic antidepressants; hCG, human chorionic gonadotropin; LH, luteinizing hormone.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
prohibited substances detected on protein chips
An example of an image of the biochip obtained with a samples negative for amphetamine is shown in Fig. 3 . Examples for 14 additional drugs are provided in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol51/issue2/. Each drug was arrayed in triplicate on the aldehyde-activated protein chip (boxed areas in Fig. 3 ). As expected, the three test spots for each drug bound the anti-drug antibody and then the bound Cy3-labeled second antibody to give a fluorescent signal. In each case, all of the mouse IgG control spots (upper and low rows of nine spots and the two central groups of three spots) were positive, as would be expected from reaction of the immobilized mouse IgG control with the goat anti-mouse conjugate used in the assay.



View larger version (22K):
[in this window]
[in a new window]
 
Figure 3. Results of analysis for amphetamine on a protein chip.

The gray box indicates the expected binding of an antibody to its counterpart. The results for 14 additional drugs are provided in the online Data Supplement (available at http://www.clinchem.org/content/vol51/issue2/).

influence of different matrices on the fluorescence signal on protein chips
The potential effect on the fluorescent signal of different samples and solutions, such as urine, water, or PBS, was evaluated. We found that different solutions could dramatically affect the signals for certain tested substances. We have noticed a significant signal decrease for steroids when the solution was changed from PBS to urine. This may be attributable to the binding of some endogenous steroid interferents with the corresponding antibodies. Most of the exogenous drugs have a comparable ratio of PBS to blank urine, commonly <1.50, but we noticed that certain analytes, such as amphetamine, generated an exceptionally high ratio of ~2.50 (Table 2 ). We repeated the assay in a 96-well plate and found that all exogenous analytes containing amphetamine had a low PBS-to-blank urine ratio of 1.13–1.18. This suggests that some nondoping substances in human urine may have interfered with the interaction of amphetamine and its antibody.


View this table:
[in this window]
[in a new window]
 
Table 2. Fluorescent signals of substances analyzed in different test solutions.

detection limit and cutoff value
In principle, the fluorescent signal at the corresponding location is decreased when a tested substance is present in the sample because the drug in the sample competes with drug immobilized on the protein chip (BSA-drug) for the anti-drug antibody. Within the linear measurement range, the decrease in fluorescent signal was proportional to the amount of drug in the sample. This method can therefore be used for both qualitative and quantitative determination of the presence of substances in a sample. A calibration curve for amphetamine measured with use of the biochip is shown in Fig. 4 . Calibration curves for nine additional prohibited substances are provided in the online Data Supplement. The detection limit is defined as the lowest concentration of an analyte that can be detected by the protein chip. This concentration corresponds to a signal that is 3 SD lower than the mean of the negative control and ranged from 0.2 µg/L for morphine to 19 µg/L for methadone. The detection limit and the cutoff values (the 50% inhibitory concentration) for the 10 drugs are summarized in Table 3 .



View larger version (10K):
[in this window]
[in a new window]
 
Figure 4. Detection of amphetamine.

Calibration curves were produced by use of different concentrations (0–1024 µg/L) of each substance added to a drug-free urine. FLU., fluorescence. The results for nine additional prohibited substances are shown in the online Data Supplement (available at http://www.clinchem.org/content/vol51/issue2/).


View this table:
[in this window]
[in a new window]
 
Table 3. Detection limits and cutoff values for some representative drugs.

assay precision
In the precision studies, standard samples, including 200 blank urines confirmed by the doping-control-analysis method at CDCC, were repeatedly analyzed (n = 300) by five technicians using different batches of protein chips. The between-batch CV for all analytes was 16%, and the within-batch CV was 13% (Table 4 ).


View this table:
[in this window]
[in a new window]
 
Table 4. Assay imprecision for different drugs.

qualitative analysis
In a typical screening procedure, we collected urines from 141 Chinese gymnastic athletes and selected morphine as a representative example drug for qualitative analysis. Fig. 5 shows representative protein chip assay results for eight specimens, including a negative specimen (Fig. 5A ), five specimens positive for morphine on (Fig. 5 , B–F), a specimen positive for dihydrocodeine (Fig. 5G ), and a specimen positive for pethidine (Fig. 5H ). The absence of a fluorescence signal at the boxed location indicated a positive result. We also detected cross-reactivity with the anti-morphine antibodies for samples containing dihydrocodeine and pethidine. The mean (SD) morphine signal was 0.429 (0.12), and the critical value (used in significance testing, which is the value that a test statistic must exceed for the null hypothesis to be rejected) was 0.232 (P <0.05), which is equivalent to 1.7 µg/L morphine. The sample would be positive for morphine if the measured signal was below the critical value. All positive samples were confirmed by GC-MS at CDCC.



View larger version (71K):
[in this window]
[in a new window]
 
Figure 5. Representative protein chip assay results.

Boxed area in each panel indicates the triplicate test spots for morphine. Signal decrease is proportional to the abundance of morphine in the sample. (A), negative sample; (B–D), results for morphine (1, 10, and 100 µg/L, respectively) added to blank urine; (E), morphine-positive urine sample; (F), 100-fold-diluted morphine-positive sample; (G), cross-reaction with dihydrocodeine-positive sample; (H), cross-reaction with pethidine-positive sample.

quantitative analysis
In addition to serving as a qualitative screening tool for large numbers of samples, the protein chip can also be used for quantitative analysis. Table 5 shows the results of six replicate tests for samples from four methamphetamine drug abusers detected by both the protein chip and a GC-MS method. The correlation coefficient (r2) for the protein chip and GC-MS results was 0.991, indicating the comparability of the results obtained by these two methods for quantifying methamphetamine in urine.


View this table:
[in this window]
[in a new window]
 
Table 5. Comparison of methamphetamine concentrations obtained with the protein chip and GC-MS.


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Protein chips can be applied to functional studies of proteins, can facilitate systematic studies of the function of thousands of proteins, and can provide an integrated understanding of biological systems (20)(21)(22). The design of diagnostic protein chips has focused on clinical applications (23)(24). Ekins and Chu (25) suggested that quantitative immunoassays could be developed that use microspots of antibodies on nonporous solid supports, thereby enabling multianalyte assays, but it was only recently that such testing biochips have become a reality (26)(27)(28)(29). A sophisticated clinical profiling system, the Evidence developed by Randox Laboratories Ltd., is the premier example of the commercial multianalyte-analysis protein chip system intended for routine clinical diagnostic applications.

The protein chip technology should be applicable to a wide range of substances, including stimulants, narcotics, anabolic agents, and peptide hormones. In principle, nearly all prohibited drugs can be tested by this technology. The limit at the moment is the availability of the corresponding antibodies and antigen conjugates. In this study, in addition to the available commercial anti-drug antibodies, we also synthesized and produced several drug-BSA conjugates and the corresponding antibodies. More antibodies will need to be produced to detect more prohibited substances.

We have developed a protein chip suitable for use in miniaturized immunoassays for doping control. This protein chip can be used to simultaneously analyze many prohibited drugs on a single array as well as to detect other suspicious drugs. For example, a sample from a female athlete that had been confirmed methamphetamine positive by CDCC tested positive on the protein chip, as expected. However, it also was positive for morphine and human chorionic gonadotropin. Later, this sample was retested by CDCC: The positive morphine result was confirmed by GC-MS, and the positive result for human chorionic gonadotropin was confirmed by a pregnancy test.

One of the major factors influencing the performance of protein chips is the development of an appropriate substrate for immobilization of printed proteins (30)(31). In the present work, aldehyde-activated glass slides were found to be the most suitable substrate for protein immobilization. Experimentally, aldehyde-activated slides have some distinct advantages. They were more stable and demonstrated lower CVs for quantitative work. Additionally, proteins immobilized on aldehyde-activated slides were resistant to removal by PBS–100 mL/L Tween 20, acetonitrile, and trifluoroacetic acid during washing steps. This indicated that the aldehyde surface provides better immobilization. In addition, the molecules covalently attached to the aldehyde-activated slide surface retained their ability to interact with target proteins in solution.

Quality control is very important in a biochip immunoassay, and we have set up a series of measures to control the quality of the assay results. We first chose the glass slides from more than 10 available companies and then selected each slide for roughness, flatness, and thickness. We next measured the contact angle after chemical modification of the slides and chose only eligible slides (contact angle in range 60–70 degrees) as substrates for printing protein chips. We then examined the protein chips by Cy5 fluorescence scanning to confirm that the mean CV of the printing was <10%. Finally, if the mean CV of the signal with mouse IgG (control spots) was >20% within a protein chip, the assay was then automatically considered to have failed.

Although specific antibodies against dihydrocodeine and pethidine were not available for this study, samples positive for these two substances could also be detected because of the cross-reactivity of the morphine antibody used in the assay. Such cross-reactivity is well known in immunologic methods for opiates (32) and results from the chemical similarities of dihydrocodeine and pethidine to morphine. The presence of dihydrocodeine and pethidine in the samples was confirmed by ELISA.

Cross-reactivity may be a valuable asset for protein chips because it enables detection of more analytes with limited numbers of antibodies. Although the WADA list of prohibited substances extends to 140 different drugs, their chemical entities can be subgrouped, and some compounds within each group have similar core structures. Cautious screening of antibodies that recognize the core structure, optimization of the antigen immobilization ratio, and suitable data processing may allow the use of ~30–50 antibodies to detect all of the prohibited substances on the list because our protein chip serves the purpose of an initial screen. Under such conditions, all questionable positive samples could be identified and a list of potential substances obtained. The suspicious samples would be confirmed by GC-MS analysis.

When a specific analytical method is not available for a new drug, the cross-reactivity may be a useful means for detection. For example, tetrahydrogestrinone (THG) is a new steroid that caused alarm in the sporting world when the US Anti-Doping Agency first found it in samples collected at the 2003 US Outdoor Track and Field Championships. If an antibody against gestrinone (a steroid structurally related to THG) were available, it might be possible to identify all samples containing gestrinone, THG, or other structurally related substances.

The definition of a screening procedure by WADA is "an analytical test procedure whose purpose is to identify those samples which are suspicious with respect to containing a prohibited substance or metabolite or marker of a prohibited method and which require additional confirmatory testing". The criteria for accepting a screening result and allowing the testing of the sample to proceed must be scientifically valid and should include negative and positive controls in addition to the samples being tested. Screening procedures for threshold substances are not required to meet quantitative or uncertainty requirements (33). Our preliminary results indicate that protein chips could be used to simultaneously screen for traces of up to 16 popular drugs in urine. With further optimization, the current protein chip system may evolve into an acceptable screening method to test for all of the substances prohibited by WADA in the coming Olympic Games in Beijing in 2008.


   Acknowledgments
 
This work was funded by the Beijing Natural Science Foundation of China (No. H010210640121) and the National Hi-Tech Program of China (No. 2002AA2Z2011). We are grateful to Profs. Larry Kricka and Brian Caddy for critical comments and suggestions on the manuscript.


   Footnotes
 
1 Nonstandard abbreviations: WADA, World Anti-Doping Agency; GC, gas chromatography; MS, mass spectrometry; CDCC, China Doping Control Center; BSA, bovine serum albumin; PBS, phosphate-buffered saline; IOC, International Olympic Committee; and THG, tetrahydrogestrinone.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Laure P, Binsinger C, Lecerf T. General practitioners and doping in sport: attitudes and experience. Br J Sports Med 2003;37:335-338.[Abstract/Free Full Text]
  2. Skibeli V, Nissen-Lie G, Torjesen P. Sugar profiling proves that human serum erythropoietin differs from recombinant human erythropoietin. Blood 2001;98:3626-3634.[Abstract/Free Full Text]
  3. Catlin D. The insiders view to detection of drugs in sport. British Medical Association. Drugs in sport 2002:76-88 BMJ Books London. .
  4. Kniess A, Ziegler E, Kratzsch J, Thieme D, Muller RK. Potential parameters for the detection of hGH doping. Anal Bioanal Chem 2003;376:696-700.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  5. Pelissier-Alicot AL, Leonetti G. Growth hormone and doping. Ann Biol Clin (Paris) 2003;61:41-48.[Medline] [Order article via Infotrieve]
  6. Jenkins P. Doping in sport. Lancet 2002;360:99-100.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  7. . World Anti-Doping Agency. World anti-doping code 2003:1-8 World Anti-Doping Agency Montreal, Canada. .
  8. Mendoza J. The war on drugs in sport: a perspective from the front-line. Clin J Sport Med 2002;12:254-258.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  9. Gray P. Drug testing at the Sydney Olympics. Med J Aust 2001;174:203-204.[ISI][Medline] [Order article via Infotrieve]
  10. The 2004 prohibited list. Montreal, Canada: World Anti-Doping Agency. http://www.wada-ama.org/docs/web/standards_harmonization/code/list_standard_2004.pdf (accessed June 2004)..
  11. Forbes GB. The effect of anabolic steroids on lean body mass: the dose-response curve. Metabolism 1985;34:571-573.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  12. Wilson JD. Androgen abuse by athletes. Endocr Rev 1988;9:181-189.[ISI][Medline] [Order article via Infotrieve]
  13. Duntas LH, Parisis C. Doping in challenge to the endocrinologist. A reappraisal in view of the Olympic Games of 2004. Hormones 2003;2:35-42.
  14. Cheng J, Sheldon EL, Wu L, Uribe A, Gerrue LO, Carrino J, et al. Preparation and hybridization analysis of DNA/RNA from E. coli on microfabricated bioelectronic chips. Nat Biotechnol 1998;16:541-546.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  15. van Ingen C. Biology’s century: just the beginning for microarrays. Nat Genet 2002;32(Suppl):463.[CrossRef]
  16. Lander ES. Array of hope. Nat Genet 1999;21(Suppl):3-4.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  17. Stahler P. Microarray technology: an array of opportunities. Nature 2002;416:885-891.[CrossRef][Medline] [Order article via Infotrieve]
  18. Cahill DJ. Protein and antibody arrays and their medical applications. J Immunol Methods 2001;250:81-91.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  19. Birchard K. Past, present, and future of drug abuse at the Olympics. Lancet 2000;356:1008.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  20. MacBeath G, Schreiber SL. Printing proteins as microarrays for high-throughput function determination. Science 2000;289:1760-1763.[Abstract/Free Full Text]
  21. Zhu H, Bilgin M, Bangham R, Hall D, Casamayor A, Bertone P, et al. Global analysis of protein activities using proteome chips. Science 2001;293:2101-2105.[Abstract/Free Full Text]
  22. MacBeath G. Protein microarrays and proteomics. Nat Genet 2002;32(Suppl):526-532.
  23. Borrebaeck CA. Antibodies in diagnostics—from immunoassays to protein chips. Immunol Today 2000;382:279-282.
  24. Ng JH, Iiag LL. Biomedical applications of protein chips. J Cell Mol Med 2002;6:329-340.[ISI][Medline] [Order article via Infotrieve]
  25. Ekins RP, Chu FW. Multianalyte microspot immunoassay—microanalytical "compact disk" of the future. Clin Chem 1991;37:1955-1967.[Abstract/Free Full Text]
  26. Haab BB, Dunham MJ, Brown PO. Protein microarrays for highly parallel detection and quantitation of specific proteins and antibodies in complex solutions. Genome Biol 2001;2:1-13.
  27. Belov L, de la Vega O, dos Remedios CG, Mulligan SP, Christopherson RI. Immunophenotyping of leukemias using a cluster of differentiation antibody microarray. Cancer Res 2001;61:4483-4489.[Abstract/Free Full Text]
  28. Robinson WH, DiGennaro C, Hueber W, Haab BB, Kamachi M, Dean EJ, et al. Autoantigen microarrays for multiplex characterization of autoantibody responses. Nat Med 2002;8:295-301.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  29. Mezzasoma L, Bacarese-Hamilton T, Di Cristina M, Rossi R, Bistoni F, Crisanti A. Antigen microarrays for serodiagnosis of infectious diseases. Clin Chem 2002;48:121-130.[Abstract/Free Full Text]
  30. Lal SP, Christopherson RI, dos Remedios CG. Antibody arrays: an embryonic but rapidly growing technology. Drug Discov Today 2002;7(Suppl):143-149.[ISI][Medline] [Order article via Infotrieve]
  31. Angenendt P, Glokler J, Sobek J, Lehrach H, Cahill DJ. Next generation of protein microarray support materials: evaluation for protein and antibody microarray applications. J Chromatogr A 2003;1009:97-104.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  32. Wainer BH, Wung WE, Connors M, Rothberg RM. The specificity of antimorphine and antimeperidine antibodies and their reactivity with opioid peptides. J Pharmacol Exp Ther 1979;208:498-506.[Abstract/Free Full Text]
  33. International standard for laboratories. Published by World Anti-Doping Agency, Montreal, Canada. http://www.wada-ama.org/docs/web/standards_harmonization/code/lab_standard_3.pdf (accessed June 2004)..




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow All Versions of this Article:
clinchem.2004.036665v1
51/2/368    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via ISI Web of Science (6)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Du, H.
Right arrow Articles by Cheng, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Du, H.
Right arrow Articles by Cheng, J.
Related Collections
Right arrow General Clinical Chemistry
Right arrow Proteomics and Protein Markers
Right arrow Drug Monitoring and Toxicology
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS