Methods: We examined potential assay interference for three types of tubes: plastic Greiner Bio-OneTM VacuetteTM; glass Becton Dickinson (BD) VacutainerTM; and plastic BD Vacutainer SSTTM tubes. Serum samples from apparently healthy volunteers (age range, 3060 years; 15 males and 34 females) were collected in different tube types and analyzed in 17 immunoassays (n = 49), 30 clinical chemistry tests (n = 20), and 33 immunology assays (n = 15). Tube effects were also examined by adding pooled serum to different tube types.
Results: TT3 values, when measured by the IMMULITETM 2000 but not the AxSYMTM analyzer, were significantly higher (P <0.0001) for SST (2.81 nmol/L) than either glass (2.15 nmol/L) or Vacuette (2.24 nmol/L) tubes. The effect was large enough to substantially shift the distribution of patient values, increasing the percentage of values above the reference interval from 11.3% to 35.8%. The degree of interference from SST tubes on TT3 differed among various tube lots and could be attributed to a tube additive shared by other plastic tubes. Results from several other tests statistically differed among tube types, but differences were not considered to be clinically significant.
Conclusions: Assay interferences from blood collection tubes represent challenges to clinical laboratories because they are not detected by the usual quality-control or proficiency testing programs. Laboratories can, however, address this problem by monitoring distribution of patients results.
| Introduction |
|---|
|
|
|---|
Blood collection tubes, however, represent much more complex devices than is commonly appreciated by laboratory workers. There are multiple components (Table 1
) that may influence clot formation, cause interaction with the tube and stopper surface, and either shed materials into the samples or adsorb components from the sample. Previous studies have suggested that serum separator tubes may have small analytical effects on many assays, but apart from the problem with therapeutic drugs, none of these effects was considered to be clinically significant (1)(3)(5)(6)(7)(8)(9)(10)(11)(12)(13). Mass spectrometry recently has shown that many types of tubes shed polymeric materials into samples and that these may interfere with some types of analyses, such as protein profiling (14). In the present study, evidence of a substantial effect of blood collection tubes on serum total triiodothyronine (TT3) results prompted an investigation of the effects of collection tubes on TT3 and other laboratory assays.
|
| Materials and Methods |
|---|
|
|
|---|
Blood samples were drawn after written informed consent from apparently healthy volunteers (age range, 3060 years; 15 males and 34 females) by trained technologists using a butterfly connected to a vacuum tube holder. Blood samples were collected into glass, Vacuette, and SST collection tubes in a randomized drawing order, and the tubes were filled to capacity. The blood collection tubes were inverted eight times after blood draw to ensure proper mixing of the blood with clot activators. Serum was obtained after clotting for 30 min at room temperature followed by centrifugation at 2000g for 5 min. All samples were processed within 2 h of blood collection. Serum drawn in Vacuette and SST tubes remained on the separator gel, whereas serum drawn in glass tubes was transferred into 13 x 75 mm test tubes. These samples were kept capped at room temperature until testing within 3 h or were stored between testing intervals at 4 °C for up to 2 days. As an alternative approach to test for possible tube interference from plastic tube additives, pooled serum collected from glass collection tubes was pipetted into each of the three tube types and placed on a roller mixer for 30 min before analysis.
laboratory analysis
Immunoassay analytes.
Serum TT3 concentrations in samples from 49 apparently healthy volunteers were measured in random order after collection in three different tube types on an IMMULITE 2000 analyzer (Diagnostic Products Corporation). Serum samples were analyzed in the same run in duplicate. Three reagent lots and one calibrator lot were used for the IMMULITE 2000 analyzer during the study. The reportable range of the IMMULITE 2000 TT3 assay is 0.629.24 nmol/L. The intra- (n = 21) and interassay (n = 21) CVs for the IMMULITE TT3 assay were, respectively, 6.7% and 8.1% at 1.36 nmol/L, 9.4% and 15% at 3.19 nmol/L, and 16% and 16% at 5.56 nmol/L. Serum samples from healthy volunteers (n = 23) were also assayed for TT3 concentrations by an AxSYM automated fluorescent analyzer (Abbott Laboratories). The reportable range of the AxSYM TT3 assay is 0.4612.32 nmol/L. The intra- (n = 80) and interassay (n = 80) CVs for the AxSYM TT3 assay were, respectively, 3.8% and 8.7% at 0.97 nmol/L, 0.7% and 5.5% at 2.40 nmol/L, and 1.2% and 4.2% at 5.82 nmol/L. All assays examined, including TT3, gave satisfactory internal and external quality-control results during the study.
The same serum samples were also analyzed on an IMMULITE 2000, according to the manufacturers instructions, for thyroid-stimulating hormone, sex hormone-binding globulin, growth hormone, ferritin, insulin, and ß2-microglobulin by immunometric principles with labeled detection antibodies and binding antibodies immobilized to polystyrene beads. Total and free thyroxine, folate, vitamin B12, cortisol, C-peptide, thyroid-binding globulin, dehydroepiandrostendione sulfate, and testosterone were measured by competitive immunoassays using limited immobilized antibodies and labeled hormones. Serum samples from each volunteer were analyzed singly in the same batch.
Routine chemistry analytes.
Routine chemistry analytes were measured in the sera collected in SST and Vacuette tubes from apparently healthy volunteers (n = 20) with a Synchron LXTM 20 Clinical System analyzer (Beckman Coulter). The serum samples were analyzed singly in random order and in the same analytical run. The methodology and range of assay imprecision obtained with the three quality controls (low, normal, and high concentration) for each analyte examined are shown in Table 1
of the Data Supplement that accompanies the online version of this article athttp://www.clinchem.org/content/vol51/issue2/ . Serum osmolality was determined with a FiskeTM 2400 osmometer (Fiske Associates).
Immunology tests.
Serum samples collected in SST and Vacuette tubes from 15 volunteers were tested for plasma proteins, including apolipoproteins A-I and B, albumin,
1-antitrypsin, ceruloplasmin, C-reactive protein, high-sensitivity C-reactive protein, rheumatoid factor, haptoglobin, C3 and C4, IgG, IgM, IgA, and IgE on an IMMAGETM nephelometer (Beckman-Coulter). Anti-cytomegalovirus IgM and IgG, anti-rubeola IgG, anti-thyroglobulin, anti-thyroid peroxidase, anti-cardiolipin IgG and IgM, anti-cyclic citrullinated peptide, and anti-Varicella zoster virus antibodies were assayed with the LabotechTM ELISA test system (Biochem ImmunoSystems). Anti-nuclear, anti-double stranded-DNA, anti-extractable nuclear antigen, and anti-Jo-1 antibodies were assayed by ELISA on the PhDTM System (Helix Diagnostics). Anti-EpsteinBarr virus-viral capsid antigen IgG antibodies were determined by a slide immunostaining method using horseradish peroxidase-conjugated anti-IgG and peroxide for detection. Serum samples from each volunteer were analyzed in random order in the same run in singleton. The methodologies and assay imprecision for the quality-control materials are shown in Table 2
of the online Data Supplement. All immunology assays were from the same reagent and calibrator lots.
|
studies of interference from tube components
To determine whether serum TT3 measured on the IMMULITE 2000 analyzer is affected by the clot activator, internal tube coating, or the lubricant from the rubber stopper, we used a dry gauze sponge to remove the clot activator, internal coating, and rubber stopper lubricant from each tube type, including plastic red- and tiger-top Vacutainer tubes. The separator gels were not disturbed during the process. A 2-mL aliquot of a serum pool from glass tubes was then added to the unaltered tubes and the cleaned glass, Vacuette, SST, and red- and tiger-top tubes (n = 5). Serum from each tube type was mixed on a roller mixer for 30 min at room temperature before analysis to assure contact of the serum with the entire tube surface. To examine the effect of serum contact with rubber stoppers, serum samples from each tube type were exposed to the respective rubber stoppers by inversion of the collection tube for 30 min, and the results were compared with serum samples from the same tube type that were not inverted (n = 5). Serum TT3 concentrations in these studies were analyzed in the same analytical run in duplicate.
statistical analysis
The means of the duplicate results for each TT3 sample were used for statistical analysis. All other serum analytes from each sample were analyzed in singleton. The results are reported as the mean (SD). Results for all measured analytes obtained from the three different collection tubes were compared by the two-tailed Student t-test or Wilcoxon signed-rank test for paired samples and by repeated-measures ANOVA. For TT3, PassingBablok regression analysis was used to compare the results among collection tube types; 95% confidence intervals for the slope and intercept were calculated from the standard error of regression and used to determine concordance with the target values of 1.00 and 0.00, respectively (15). Spearman rank correlation coefficients and BlandAltman plots were used to assess the difference between TT3 results obtained with different types of collection tubes (16). All P values were adjusted by use of Bonferroni correction for the multiple comparisons inherent in the pairwise testing procedures. Statistical analyses were performed with StatViewTM (Ver. 5.0; SAS Institute Inc.) and Analyze-ItTM for Microsoft Excel (Ver. 1.71; Analyze-It Software) software.
To determine whether statistically significant differences in analyte concentrations among tube types were clinically relevant, we used the significant change limit method, as described by Boyanton and Blick (17). Briefly, the mean for each analyte in the glass or Vacuette tube represented the initial value. The usual SD used for calculating the significant change limit was based on the mean SD of the quality-control data for the previous 6 months for each respective analyte. The quality-control material whose target mean most clearly matched the initial value from either glass or Vacuette tubes for each analyte was used to determine the usual SD. The significant change limit was calculated for each analyte by determining the range (±2.8 usual SD) from the mean of glass or Vacuette tubes. Serum analyte concentrations from the different collection tubes that exceeded their respective significant change limits were considered to be clinically significant.
| Results |
|---|
|
|
|---|
33% (Fig. 1A
|
In July 2002, there was a downward shift in serum TT3 results, which was apparently observed by other users of the IMMULITE 2000 TT3 assay as well (18) (Fig. 1B
, arrow a). In response, the manufacturer reformulated and recalibrated the assay beginning with lot no. 144 of the L2KT3 reagent set (18), and when the new assay was implemented by our laboratory in December 2002, we observed an upward shift of the serum TT3 results (Fig. 1B
, arrow b). No change in the reference interval of the assay was recommended by the manufacturer during this time period. However, an unexplained gradual upward shift in the serum TT3 results appeared, starting around January 2004. A reference interval study was performed on apparently healthy volunteers, and the reference interval was changed from 1.262.76 nmol/L to 1.543.31 nmol/L. The laboratory then began an investigation of the source of the shift in assay values. After identification of assay interference from blood collection tubes in July 2004, the serum TT3 results shifted back to the previous range after the laboratory switched from plastic SST tubes to Vacuette plastic collection tubes (Fig. 1B
, arrow c).
blood collection tube comparison of serum tt3
To assess the possible contribution of blood collection tubes to problems with the TT3 assay, serum samples were collected in three tube types: glass tubes (a red-top glass Vacutainer, no-additive blood tube; 10.25 x 64 mm); SST tubes (a gold-top Vacutainer plastic tube with gel separator; 13 x 75 mm); and Greiner Bio-One Vacuette tubes (a red and yellow-top plastic tube with gel separator; 13 x 75 mm). The means of the TT3 concentrations measured in 49 serum samples from each type of blood collection tube measured on the IMMULITE 2000 analyzer are shown in Table 2
. The results of PassingBablok regression analysis of serum TT3 concentrations among tube types are shown in Fig. 2
, A, C, and E. The serum TT3 concentrations obtained with the SST tubes showed a significant positive proportional difference compared with glass and Vacuette tubes (Fig. 2
, A and E, respectively). Serum TT3 concentrations from glass and Vacuette tubes did not demonstrate significant differences (Fig. 2C
).
|
For the BlandAltman plots of the 49 serum samples (Fig. 2
, B, D, and F), the mean (SD) differences 0.66 (0.061) nmol/L for SST vs glass tubes, 0.094 (0.028) nmol/L for Vacuette vs glass tubes, and 0.57 (0.049) nmol/L for SST vs Vacuette tubes, respectively.
effect of tube lot-to-lot variation on serum tt3 results
We investigated a defective lot of SST tubes as a possible source of the increase in TT3 concentrations by comparing the predominant lot of SST tubes in use at the time (lot no. 3308587) with a new lot of SST tubes (lot no. 4034512). We also tested multiple lots of the other tube types. We observed a significant lot-to-lot variation for serum TT3 concentrations with the SST tubes (lot nos. 3308587 vs 4034512; 2.97 and 2.59 nmol/L, respectively; n = 10; P = 0.003) but not with the glass collection tubes (lot nos. 3253846 vs 3029306; 2.32 and 2.34 nmol/L, respectively; n = 10; P = 0.80) or Vacuette (lot nos. B120307 vs B04040L; 2.33 and 2.34 nmol/L, respectively; n = 10; P = 0.22). Although the new lot of SST tubes appeared to yield lower TT3 results than the old lot, TT3 results from serum collected in the new lot of tubes were still higher than those with the other two tube types (P = 0.03).
tube types and different immunoassay methods
To determine whether tube types had similar effects on different TT3 methods, we measured serum TT3 in samples collected with the different collection tubes on both the IMMULITE 2000 and AxSYM analyzers (n = 23). We observed no significant differences in serum TT3 concentrations among different tube types on the AxSYM analyzer (P = 0.98). The serum TT3 concentrations in glass, Vacuette, and SST tubes measured on the AxSYM were 1.24, 1.23, and 1.25 nmol/L, respectively. In contrast, we found significant differences in serum TT3 concentrations from the different tube types on the IMMULITE 2000 analyzer (P = 0.02). The serum TT3 concentration in SST tubes measured on the IMMULITE was significantly higher (2.20 nmol/L) than in glass (1.68 nmol/L) or Vacuette (1.70 nmol/L) tubes. We found no significant difference between glass and Vacuette tubes, as measured on the IMMULITE 2000 analyzer.
dilution studies
We evaluated the effect of dilution on the SST tube interference in the TT3 assay by performing 1:1 to 1:3 dilution studies using human serum albumin (40 g/L) in phosphate-buffered saline as the diluent. The regression equations for observed (y) vs expected (x) serum TT3 values on the diluted samples were y =0.94x + 10.0 (r = 0.99) for glass, y = 1.05x 10.0 (r = 0.99) for Vacuette, and y = 0.98x + 0.5 (r = 0.99) for SST tubes, demonstrating that the interferent diluted linearly from 1:1 to 1:3 with an albumin solution. Sample dilution, therefore, was not a potential solution to the interference problem.
Addition of saline (8.5 g/L NaCl) to SST tubes for up to 48 h, however, did not extract any interfering substance, as determined by use of the saline incubated in the tubes as a diluent for the TT3 assay.
effects of tube components on tt3 assays
Incubation of serum originally collected in glass tubes and then transferred to the SST tubes revealed the presence of an interfering substance in the SST tubes. A serum pool obtained from glass collection tubes with a TT3 concentration of 2.05 nmol/L on repeat analysis after incubation in SST tubes for 30 min had a concentration of 3.25 nmol/L when measured on the IMMULITE 2000 analyzer (n = 4; P = 0.003). Analysis of the same serum samples on the AxSYM analyzer revealed no significant change in TT3 concentrations (glass tube, 1.25 nmol/L; SST tube, 1.19 nmol/L; P = 0.46). We observed no significant difference in the TT3 concentration when the serum pool obtained from glass tubes was compared with the same serum pool from glass tubes that was incubated in Vacuette tubes and measured on the IMMULITE 2000 (glass tube, 2.05 nmol/L; Vacuette tube, 2.07 nmol/L; P = 0.49) or the AxSYM (glass tube, 1.25 nmol/L; Vacuette tube, 1.29 nmol/L; P = 0.80) analyzer.
To determine whether TT3 measured on the IMMULITE 2000 analyzer was affected by the clot activator, internal tube coating, or lubricant on the rubber stopper, we transferred serum samples from glass tubes to tubes that were cleansed with a gauze sponge to remove coatings on the tube and rubber stopper, as described in the Materials and Methods. The results indicated that the TT3 concentrations were significantly higher in unaltered compared with cleaned tubes for SST (2.48 vs 1.99 nmol/L, respectively; n = 5, P = 0.01) but not glass (1.48 vs 1.57 nmol/L, respectively; n = 5; P = 0.43) or Vacuette tubes (1.45 vs 1.51 nmol/L, respectively; n = 5; P = 0.30). Although not statistically significant, measured TT3 concentrations were higher in unaltered compared with cleaned tubes for red-top (2.22 vs 1.97 nmol/L, respectively; n = 5; P = 0.13) and tiger-top (2.48 vs 2.13 nmol/L, respectively; n = 5; P = 0.14) Vacutainer collection tubes. These findings support the hypothesis that the coating in the collection tubes and the lubricant on the rubber stopper are the interferent from SST tubes.
To further establish whether the increased TT3 concentrations found in the SST tubes were attributable to contact with its rubber stopper or the materials applied to the stopper, we exposed serum samples from each tube type to the respective rubber stoppers by inversion of collection tubes and compared the results with serum samples from the same tube type that were not inverted (n = 5). TT3 concentrations were higher in samples exposed to the rubber stoppers compared with those that were not exposed to rubber stoppers for SST (2.74 vs 2.36 nmol/L, respectively; P = 0.03) but not the glass (1.98 vs 1.97 nmol/L, respectively; P = 0.96) and Vacuette collection tubes (1.96 vs 2.00 nmol/L, respectively; P = 0.45).
Recently, SST tubes with a new separator gel formulation, Vacutainer SST II (plastic tube; 13 x 100 mm; cat. no. 367955; lot no. 3293573), have been developed and marketed by BD. We examined whether TT3 concentrations obtained from SST and SST II tubes differed (n = 10). We found significant differences between the SST and SST II tubes (2.98 and 2.77 nmol/L, respectively; P = 0.02), but both the SST and SST II tubes gave higher TT3 results than glass and Vacuette tubes (2.34 and 2.33 nmol/L, respectively; P = 0.03). Therefore, the positive interference observed with the SST and SST II tubes on the IMMULITE 2000 TT3 assay is independent of the type of separator gel.
effect of tube types on other assays
In view of the effect of collection tubes on TT3 results, we tested other serum assays in the laboratory with samples collected in the three tube types. Serum samples from 49 apparently healthy volunteers were tested on the IMMULITE 2000 analyzer for the assays shown in Table 2
. We observed statistical differences (P <0.003) between tube types for 10 of the 17 serum analytes examined. The difference between glass and SST tubes was >10% for progesterone, TT3, cortisol, thyroid-binding globulin, total thyroxine, and insulin, whereas only thyroid-binding globulin differed by >10% for glass vs Vacuette tubes (Table 2
). Apart from TT3, none of the other test differences would be regarded as clinically important (Table 2
) based on the calculated significant change limit for each test (17). Similarly, we tested the effect of collecting serum in SST and Vacuette tubes on general chemistry analytes and immunology tests (Tables 1
and 2
, respectively, in the online Data Supplement). The difference between the two types of tubes (meanSST meanVacuette) was statistically significant (P <0.002) for 6 of 30 measured general chemistry analytes, but the magnitude of these differences was small (range, 3.15% to 5.18%) and not considered to be clinically relevant (17). For all of the immunology analytes examined, we observed no significant differences in results between the Vacuette and SST collection tubes (meanSST meanVacuette; Table 2
in the online Data Supplement).
| Discussion |
|---|
|
|
|---|
The substantial lot-to-lot variation of SST tube effects on TT3 suggests that the positive bias between TT3 results cannot be adequately compensated by an adjustment in the reference interval. We also observed higher measured serum TT3 concentrations in samples from plastic red-top (2.19 nmol/L) and tiger-top Vacutainer tubes (2.53 nmol/L) compared with glass tubes (1.80 nmol/L) and Vacuette (1.83 nmol/L) tubes (n = 25 for each tube type). Overall, these results suggest that the SST tubes and other plastic serum blood collection tubes from this manufacturer are not suitable for performing the TT3 assays on the IMMULITE 2000 analyzer, although the tubes appear to be suitable for TT3 analysis on the AxSYM analyzer. The SST tubes, however, appeared to function acceptably for all other laboratory tests that were examined (Table 2
and Tables 1
and 2
in the online Data Supplement). The effect of the blood collection tubes on any assay, however, should be considered in the context of the total allowable error; thus, the overall diagnostic utility of some of the other assays that showed a blood collection tube bias may still be adversely affected by the problem with the BD collection tubes.
The identity of the interfering substance in the SST tube is not known, but as shown in Table 1
, there are only a few possible candidates. The main difference among the glass, Vacuette, and SST tubes is the interior coating. A silicone polymer was used to coat the interior of the SST but not the glass or Vacuette tubes (Table 1
). Silicone is used to fix the silica powder (clot activator) to the plastic tubes and as a stopper lubricant (20). The silicone coating in the tubes is also used to reduce adherence of erythrocytes to the walls of the collection tubes (20). Previous reports have shown that silicone-coated collection tubes can interfere with ion-specific electrode determinations of both ionized magnesium (21)(22)(23)(24) and lithium (20), causing falsely increased concentrations. In addition, the water-soluble silicone polymer coating the interior of serum separator tubes has been shown to interfere negatively with avidin-biotin binding in an IRMA for thyrotropin, prolactin, and human chorionic gonadotropin (25). Interestingly, the higher TT3 results obtained in the red- and tiger-top compared with glass collection tubes may also be attributable to the silicone polymer because both of these tubes have silicone as an internal tube coating and stopper lubricant (26). It is important to note, however, that Vacuette tubes also contain silicone as a lubricant on the rubber stopper but not as a coating on the interior of the tube [Table 1
; personal communication with Greiner Bio-One technical services; Ref. (27)]. Hence, the presumably smaller amount of silicone present on the rubber stopper of Vacuette tubes may explain the small mean difference in measured TT3 in these tubes compared with the glass tubes, which have glycerin as the stopper lubricant with no internal tube coating [Tables 1
and 2
and Fig. 2D
; Ref. (26)].
In our study, removing the clot activator, internal tube coating, and lubricant from the rubber stopper with a gauze sponge significantly decreased the measured serum TT3 concentration in SST and other BD plastic serum tubes, thus supporting the hypothesis that an interferent, such as silicone, coating the tubes and as a stopper lubricant falsely increases the serum TT3 concentration measured by the IMMULITE TT3 assay. However, tube additives other than silicone used to coat the interiors of SST tubes (i.e., polyvinylpyrrolidone, polyethylene oxide, and polyvinyl alcohol) may also be responsible for the falsely increased serum TT3 concentrations in SST compared with glass or Vacuette tubes when measured on the IMMULITE 2000 analyzer (28).
The falsely increased serum TT3 concentrations obtained from the SST and other plastic BD serum tubes measured on the IMMULITE 2000 analyzer may also be dependent not only on the presence of silicone, but also on the quantity in the collection tubes. This point is supported by the observation that the measured TT3 in serum exposed to SST tubes with rubber stoppers was significantly higher than TT3 values for serum exposed to SST tubes without rubber stoppers. It is also conceivable that not all collection tubes were coated during production with a homogeneous layer of the silicone coating in the tubes and/or on the rubber stoppers. Hence, this possible variation in quantity of silicone in the collection tubes may explain the differences in measured serum TT3 concentrations among the tube types and among the different SST tube lots (Table 2
and Fig. 2
). The mechanism of interference of tube components with the IMMULITE 2000 TT3 assay is not clear. Further studies are warranted to elucidate the exact mechanism by which tube components interfere with the TT3 assay.
It is noteworthy that participation in internal and external quality assurance programs would not reveal the type of preanalytical problem described in this study. Proficiency samples received by the clinical laboratory for evaluation are contained in the same type of sealed vials or tubes; therefore, any effect of variations in collection tube components on the analyte test result(s) would have been excluded from these proficiency samples. This is of particular importance for clinical laboratories that receive serum or plasma samples collected in different types of tubes from different manufacturers. Although it was done only retrospectively, in this case the routine monitoring of moving averages based on patient data may be potentially useful for identifying future tube-related problems (Fig. 1B
). Moving average procedures are often criticized for (a) being time-consuming and insensitive compared with common quality-control rules (29)(30); (b) being complex, thus requiring sophisticated computer programs to smooth the patient data and to produce meaningful graphic summaries (29)(30); and (c) being unduly sensitive to outlying groups of patient results (29)(30). Despite these limitations, the moving average data would have alerted, in this case, the laboratory to an increase in TT3 concentrations in patient samples that conventional internal and external quality-control material missed.
Other than increased vigilance when inspecting laboratory results and improving the feedback between the clinical laboratory and clinicians, there is not much that clinical laboratories can do to readily detect blood collection tube problems. It is impractical for clinical laboratories to repeat a tube evaluation study with each new lot of tubes, but laboratories should consider comparing results from separate lots of the same tube type to detect any lot-to-lot variations when first evaluating a new tube. Similarly, it is impractical for tube manufacturers to test their tubes on all the various assay platforms, but they should ensure during the manufacturing process consistency in the amount and quality of any tube additives. Diagnostics companies could also help in the identification of future tube problems by providing detailed information not only on the tube type but the commercial source of the tubes that they use when determining the reference interval for any new assay. Any reference interval study done by diagnostics companies on previously developed assays with older tubes that are no longer widely used, such as glass tubes, should ideally be repeated with tube types that are currently used by their customers. Finally, all stakeholders in this issue should be more vigilant about the effect of blood collection tubes on laboratory assays and work together to prevent and minimize the problem.
Since submission of this manuscript, BD has sent out a technical bulletin (VS7313; www.bd.com/vacutainer/techbulletins/) describing the SST tube interference on various instrument platforms other than the Diagnostics Products Corporation [ADVIATM CENTAUR and ACS 180TM (Bayer Healthcare Diagnostics Division) and ACCESSTM, ACCESS 2, UNICELTM DXI800, SYNCHRON LX I 725 (Beckman Coulter Inc)] and on a variety of different assays (TT3, TT4, folate, vitamin B12, follicle-stimulating hormone, hepatitis B surface antigen, cancer antigen 27.29, and cortisol). BD has recommended alternative blood collection tubes that can be used in the interim until the SST tube interference problem is resolved.
| Acknowledgments |
|---|
NOTE ADDED IN PROOF. In response to immunoassay problems with the BD Vacutainer SST, SST II, and Microtainer tubes, BD has recently reformulated the tubes to reduce the amount of surfactant in the tubes to eliminate the assay interference (www.bd.com/vacutainer/techbulletins/). We have evaluated serum samples collected in the adjusted BD Vacutainer SST (13 x 100 mm and 13 x 75 mm) and the adjusted tiger-top (16 x 100 mm) tubes and have compared the results to BD glass (10.25 x 64 mm) and Greiner Vacuette (13 x 75 mm) blood collection tubes on the IMMULITE 2000 analyzer for the immunoassay analytes shown in Table 2
(n = 10 for each tube type in duplicate). No clinically significant differences (see Table 2
for significant change limit criteria) among the tube types were observed (data not shown); therefore, it appears that the new BD tubes have been successfully adjusted to reduce the assay interference and now yield results that are similar to glass and Vacuette tubes for the assays tested.
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
A. C. Madeo, A. Manichaikul, J. C. Reynolds, N. J. Sarlis, S. P. Pryor, T. H. Shawker, and A. J. Griffith Evaluation of the Thyroid in Patients With Hearing Loss and Enlarged Vestibular Aqueducts Arch Otolaryngol Head Neck Surg, July 1, 2009; 135(7): 670 - 676. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. E.P.B. Ballieux, N. I. Weijl, H. Gelderblom, J. van Pelt, and S. Osanto False-Positive Serum Human Chorionic Gonadotropin (hCG) in a Male Patient with a Malignant Germ Cell Tumor of the Testis: A Case Report and Review of the Literature Oncologist, November 1, 2008; 13(11): 1149 - 1154. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Stankovic and G. Parmar Assay interferences from blood collection tubes: a cautionary note. Clin. Chem., August 1, 2006; 52(8): 1627 - 1628. [Full Text] [PDF] |
||||
![]() |
S. Wang, V. Ho, A. Roquemore-Goins, and F. A. Smith Effects of Blood Collection Tubes, including Pediatric Devices, on 16 Common Immunoassays. Clin. Chem., May 1, 2006; 52(5): 892 - 893. [Full Text] [PDF] |
||||
![]() |
L. J. Kricka, J. Y. Park, M. B. Senior, and R. Fontanilla Processing Controls in Blood Collection Tubes Reveals Interference Clin. Chem., December 1, 2005; 51(12): 2422 - 2423. [Full Text] [PDF] |
||||
![]() |
L. J. Kricka and J. Y. Park Additive-Aggravated Assays: An Authoritative Answer Clin. Chem., October 1, 2005; 51(10): 1767 - 1767. [Full Text] [PDF] |
||||
![]() |
R. A.R. Bowen, Y. Chan, M. E. Ruddel, G. L. Hortin, G. Csako, S. J. Demosky Jr, and A. T. Remaley Immunoassay Interference by a Commonly Used Blood Collection Tube Additive, the Organosilicone Surfactant Silwet L-720 Clin. Chem., October 1, 2005; 51(10): 1874 - 1882. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. T. Holmes, A. Levin, B. Forer, and F. Rosenberg Preanalytical Influences on DPC IMMULITE 2000 Intact PTH Assays of Plasma and Serum from Dialysis Patients Clin. Chem., May 1, 2005; 51(5): 915 - 917. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |