Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 51: 472-474, 2005; 10.1373/clinchem.2004.043802
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Singh, R. J.
Right arrow Articles by Binder, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Singh, R. J.
Right arrow Articles by Binder, S.
Related Collections
Right arrow Endocrinology and Metabolism
(Clinical Chemistry. 2005;51:472-474.)
© 2005 American Association for Clinical Chemistry, Inc.


Letters to the Editor

Precisely Wrong? Urinary Fractionated Metanephrines and Peer-Based Laboratory Proficiency Testing

Ravinder J. Singh1,a, Stefan K. Grebe1,2, Bingfang Yue3, Alan L. Rockwood3, John C. Cramer4, Zoltan Gombos5 and Graeme Eisenhofer6

Departments of,1 Laboratory Medicine and Pathology, and 2 Medicine, Mayo Clinic, Rochester, MN
3 ARUP Institute for, Clinical and Experimental Pathology, Salt Lake City, UT
4 Laboratory of Bioorganic Chemistry, National Institute of Diabetes, and Digestive and Kidney Diseases, NIH, Bethesda, MD
5 Department of Pathology, and Laboratory Medicine, Berkshire Health System, Pittsfield, MA
6 Clinical Neurocardiology Section, National Institute of, Neurological Disorders and Stroke, NIH, Bethesda, MD

aAddress correspondence to this author at: Department of Laboratory Medicine and Pathology, Mayo Clinic, 200 1st St. SW, Rochester, MN 55905. Fax 507-284-9758; e-mail Singh.Ravinder{at}mayo.edu.


To the Editor:

Measurements of urinary fractionated metanephrines (normetanephrine and metanephrine), the O-methylated metabolites of norepinephrine and epinephrine, provide a sensitive test for diagnosis of pheochromocytoma. We present evidence suggesting inaccurate metanephrine and normetanephrine calibration by US laboratories.

Metanephrines are present in urine mainly as sulfate- and glucuronide-conjugated metabolites produced from free metanephrines by the actions of conjugating enzymes (1). An acid hydrolysis step is usually performed to liberate the free metanephrines from the conjugated metabolites. This step minimizes requirements for high analytical sensitivity, simplifying subsequent measurement.

Urinary fractionated metanephrines are usually measured by HPLC with electrochemical detection (HPLC-EC). Gas chromatography with mass spectrometry (GC-MS) and liquid chromatography with tandem mass spectrometry (LC-MS/MS) are more recent alternatives offering high sample throughput and improved analytical specificity (2)(3).

Clinical laboratories in the US participate in proficiency testing programs, primarily aimed at determining agreement in diagnostic test results among laboratories. Identical survey samples are distributed to participating laboratories, with results stratified according to testing methodologies. Laboratories that report results deviating by more than a certain amount from the mean of their peer groups fail the proficiency test.

Until the spring of 2004, the only source of commercially available urine metanephrine calibrators in the US was Bio-Rad. As with the survey samples used for proficiency testing, these calibrators are supplied as lyophilized urine, which must be reconstituted and taken through the same sample-processing steps as patient samples. One laboratory not using these commercial calibrators (Mayo Clinic Endocrine Laboratory) reported significantly higher results for urinary fractionated metanephrines than all other participating laboratories during three successive proficiency surveys in 2003/2004. Mayo laboratory calibrators are prepared in house from normetanephrine and metanephrine supplied by Sigma-Aldrich Chemical Co. The Mayo laboratory suspected that differences between their in-house calibrator and the commercial calibrator used by other laboratories might have been responsible for their successive failures to pass proficiency testing. Comparisons of the two calibrators by LC-MS/MS indicated 33% higher values for normetanephrine and 40% higher values for metanephrine in the commercial calibrator than in the in-house calibrator (Table 1 ). Values in the commercial calibrator were only slightly higher with than without the acid hydrolysis step, indicating that most of the metanephrines were in the free rather than the conjugated form.


View this table:
[in this window]
[in a new window]
 
Table 1. Differences between assigned and measured values for Bio-Rad calibrators.1

To resolve the discrepancy, two independent laboratories (one at the NIH and the other at ARUP) were asked by those overseeing the proficiency program to investigate the problem. The calibrators used by the NIH laboratory (also prepared in house from normetanephrine and metanephrine supplied by Sigma-Aldrich) showed close agreement by HPLC-EC with those supplied by the Mayo laboratory. Values for the commercial calibrator were 36% higher than assigned values for normetanephrine and 50% higher than assigned values for metanephrine (Table 1Up ). Again, values were only slightly higher with an acid hydrolysis step, confirming that most of the metanephrines in the commercial calibrator were in the free form. The ARUP laboratory routinely used the Bio-Rad calibrator for GC-MS measurements of urinary fractionated metanephrines. When compared with freshly prepared calibration solutions from Sigma-Aldrich, the concentrations of metanephrines in the Bio-Rad calibrator were also calculated to be significantly higher than assigned values (Table 1Up ). Possible degradation of free metanephrines during the acid hydrolysis step was determined to be insignificant (<3%).

The three laboratories used different methods of analysis (HPLC-EC, LS-MS/MS, and GC-MS), indicating that the observed differences were unrelated to methodology or instrumentation. However, all three laboratories used the same source for their in-house-prepared calibrators. Thus, an impure starting material from Sigma-Aldrich could explain the discrepancy in assigned and measured values for the calibrators from Bio-Rad. To examine this possibility, samples of the Sigma-supplied normetanephrine and metanephrine were tested by elemental analysis. The results (Table in the Data Supplement that accompanies the online version of this letter at http://www.clinchem.org/content/vol51/issue2/) confirmed that the purity of the Sigma-supplied material was >98%, as stipulated by the manufacturer. This analysis and the highly consistent findings from the three independent laboratories indicate that the values assigned to the Bio-Rad calibrator are indeed 24–33% lower than their real values.

If most of the US laboratories participating in the proficiency testing program have been using the Bio-Rad calibrator, then this would explain the differences in results of proficiency testing between the Mayo laboratory and the other laboratories participating in the program. It is important to note that this program is peer-based, not accuracy-based. Although peer-based programs provide an adequate system for assessing agreement among members of a peer group, they do not guarantee accuracy of results.

Finally, Bio-Rad has informed one of us (A.L.R.) that their calibrator is for the Bio-Rad HPLC assay and cannot be assumed to work with other methods. If this indeed is the official position of Bio-Rad, then those laboratories using their calibrator, but not their assay reagents, should consider changing to an accurate calibrator prepared in house or provided by an alternative commercial source.


References

  1. Eisenhofer G. Free or total metanephrines for diagnosis of pheochromocytoma: what is the difference?. Clin Chem 2001;47:988-989.[Free Full Text]
  2. Crockett DK, Frank EL, Roberts WL. Rapid analysis of metanephrine and normetanephrine in urine by gas chromatography-mass spectrometry. Clin Chem 2002;48:332-337.[Abstract/Free Full Text]
  3. Taylor RL, Singh RJ. Validation of liquid chromatography-tandem mass spectrometry method for analysis of urinary conjugated metanephrine and normetanephrine for screening of pheochromocytoma. Clin Chem 2002;48:533-539.[Abstract/Free Full Text]

A representative for Bio-Rad responds:

Steven Binder

R1 Bio-Rad Laboratories, Clinical HPLC Group, 4000 Alfred Nobel Dr., Hercules, CA 94547

aE-mail steve_binder{at}bio-rad.com


To the Editor:

The Bio-Rad Metanephrines by HPLC method was first introduced in 1984. This method uses an extensive sample clean-up for urine, based on two ion-exchange columns and several pH changes. The results were compared with a published gas chromatography–mass spectrometry method (1) and showed good agreement. Consistency of calibration has been maintained over time by side-by-side comparison of lots, as well as by the method of standard additions.

There is currently no certified reference material (CRM) available for use in this test. One of the primary requirements for a CRM is the use of a matrix similar to the patient specimen (2). We have observed that the calibrator developed for mass spectrometric methods may be a working solution of the analytes prepared in dilute acid or methanol (3). This type of calibrator cannot be tested in the Bio-Rad HPLC method because the analytical recovery, in the absence of appropriate concentrations of salts, proteins, and buffers, does not resemble the analytical recovery observed in human urine. Similarly, the Bio-Rad calibrator, which is prepared with human urine and appropriate preservatives, may not demonstrate the expected behavior when other methods are used, and such usage is not consistent with current concepts of metrology.


References

  1. Canfell C, Binder SR, Khayam-Bashi H. Quantitation of urinary normetanephrine and metanephrine by reversed-phase extraction and mass-fragmentographic analysis. Clin Chem 1982;28:25-28.[Abstract/Free Full Text]
  2. Müller MM. Implementation of reference systems in laboratory medicine. Clin Chem 2000;46:1907-1909.[Free Full Text]
  3. Taylor RL, Singh RJ. Validation of liquid chromatography-tandem mass spectrometry method for analysis of urinary conjugated metanephrine and normetanephrine for screening of pheochromocytoma. Clin Chem 2002;48:533-539.



The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
E. Grouzmann, B. Mathian, and T. Buclin
Calibration of Fractionated Metanephrines in Urine: Still an Issue?
Clin. Chem., October 1, 2008; 54(10): 1738 - 1739.
[Full Text] [PDF]


Home page
Clin. Chem.Home page
R. J. Singh and G. Eisenhofer
High-Throughput, Automated, and Accurate Biochemical Screening for Pheochromocytoma: Are We There Yet?
Clin. Chem., September 1, 2007; 53(9): 1565 - 1567.
[Full Text] [PDF]


This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Singh, R. J.
Right arrow Articles by Binder, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Singh, R. J.
Right arrow Articles by Binder, S.
Related Collections
Right arrow Endocrinology and Metabolism


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS