| Abstract |
|---|
|
|
|---|
Methods: We developed and evaluated three immunoassays (U-MidOC, U-LongOC, and U-TotalOC) for the detection of various molecular forms of urine OC (U-OC). We evaluated the association of U-OC with other markers of bone turnover and with bone mass in 1044 elderly women and studied seasonal and circadian variation of U-OC.
Results: U-OC correlated with other bone turnover markers [Spearman correlation (r), 0.300.57; P <0.0001], demonstrating the association between U-OC and skeletal metabolism. There was also a significant association between bone metabolism assessed by U-OC quartiles and bone mass assessed by total body bone mineral content (P <0.0001). The seasonal effects appeared to be rather small, but we observed a significant circadian rhythm similar to the one reported for serum OC with high values in the morning and low values in the afternoon.
Conclusions: The three immunoassays had unique specificities toward different naturally occurring U-OC fragments. U-OC concentrations measured with any of these assays correlated with bone turnover rates assessed by conventional serum markers of bone metabolism. The measurement of OC in urine samples could be used as an index of bone turnover in monitoring bone metabolism.
| Introduction |
|---|
|
|
|---|
5800) produced by osteoblastic cells during bone formation (4)(5)(6). A fraction of OC enters the blood, where it can be detected, and circulating OC has been widely measured to assess bone turnover (7)(8). In addition to the newly synthesized OC derived from osteoblasts, the circulating OC pool probably also includes molecules derived from the resorption process when OC embedded in the bone matrix is released (9)(10). The main route of circulating OC catabolism is renal filtration and degradation (11), and immunoreactive OC is present in urine (12). Urine OC (U-OC) is a heterogeneous pool of various OC fragments (13)(14)(15) that reflect potentially diverse degradation cascades and that consist mainly of the middle portion of the molecule truncated at both the amino and carboxy termini. Intact, unfragmented OC has not been found in urine.
The introduction of serum OC (S-OC) assays into routine clinical practice has been limited by the presence of multiple forms of OC in the circulation and differences among assays in the ability to detect these forms (16). It is unclear whether OC fragments are generated by proteolysis of intact OC in blood, are formed during biosynthesis, or whether some fragments are derived directly from bone resorption. OC molecules of resorptive origin would be of clinical interest, but the circulating resorptive forms may be masked in serum samples by de novo-synthesized intact OC, which is inevitably recognized by most immunoassays in addition to the shorter fragments. Circulating intact OC and large fragments are degraded in circulation or peripheral organs (11), whereas smaller OC fragments might be more resistant to degradation and accumulate in urine. Urine could therefore potentially offer a better source for OC fragments of resorptive origin. In addition, urine samples can be collected easily, and the problems related to the instability of S-OC after sampling (17) are likely to be less severe for U-OC, which is presumably an end-product of fragmentation.
U-OC is attractive as a new candidate marker of bone turnover despite such limitations as the gradual age-related renal impairment and the tendency to show higher preanalytical variability than serum assays (18). The latter could be related to the use of creatinine for normalizing urinary marker values, which contributes a second source of variability (19). Urine as a sample material is, however, intriguing because it traditionally has been used for measurement of resorption markers and not markers of osteoblastic origin.
At the moment, there are no commercial methods available for the measurement of U-OC. Most assays are probably unable to detect the small fragments present in urine, and the concentrations in urine are low. RIAs for U-OC have used a polyclonal guinea pig antibody against OC purified from human bone (20) or a polyclonal chicken antibody against a synthetic midmolecule peptide sequence (residues 1236) (14), and two-site immunofluorometric assays (13) have used murine monoclonal antibodies (Mabs) for various epitopes of human OC (21) in two different two-site combinations. Limited data are available on the analytical performance of U-OC assays and the possible confounding factors that might have an influence on the measured U-OC, such as biological variation and stability of immunoreactivity after sampling.
In this report we describe the development and characterization of three immunoassays for the measurement of OC fragments in human urine and provide data on preanalytical variability.
| Materials and Methods |
|---|
|
|
|---|
-carboxyglutamic acid (Gla)-containing forms of OC. The antibodies were biotinylated with a 50-fold molar excess of biotinisothiocyanate (bio-Mabs) and labeled with a 200-fold molar excess of europium(III) chelate (Eu-Mabs) as described previously (21). A synthetic peptide of human OC (hOC) amino acids 149 (Gla at positions 17, 21, and 24; Advanced Chemtech) was biotinylated with a 30-fold molar excess of biotinisothiocyanate (bio-hOC) as described previously (15). Bio- and Eu-Mabs and bio-hOC were used in three combinations to generate two sandwich assays, bio-6F9/Eu-3H8 (U-MidOC) and bio-2H9/Eu-6F9 (U-LongOC), and one competitive assay, bio-hOC/Eu-3H8 (U-TotalOC). A synthetic peptide of hOC amino acids 143 (Gla at positions 17, 21, and 24; Advanced Chemtech) was used as a calibrator. The calibrator was diluted in 50 mmol/L Tris-HCl, 150 mmol/L NaCl, 5 g/L NaN3 (pH 7.75) buffer containing 10 g/L bovine serum albumin and stored in aliquots at 70 °C.
immunoassay procedures for urinary oc
For the U-MidOC and U-LongOC assays, 400 ng of bio-6F9 (U-MidOC) or bio-2H9 (U-LongOC) in 50 µL of Delfia Assay Buffer was added to the wells of streptavidin-coated microtiter plates, incubated with continuous shaking at room temperature (22 °C) for 30 min, and washed twice with Delfia Wash Solution. The samples or calibrators (10 µL) were added, followed by 100 ng of 3H8 (U-MidOC) or Eu-2H9 (U-LongOC) in 50 µL of Assay Buffer containing 5 mmol/L EDTA. After 2 h of shaking at room temperature (22 °C), the plates were washed six times, and 200 µL of Delfia Enhancement Solution was added. The plates were shaken for 30 min before time-resolved fluorescence measurement in a Victor2 Multilabel Counter. Streptavidin-coated plates were purchased from Innotrac Diagnostics, and Delfia Assay Buffer, Delfia Wash Solution, Delfia Enhancement Solution, and the Victor2 Counter were from Perkin-Elmer Life Sciences/Wallac.
For the U-TotalOC assay, a competitive assay design was used. Streptavidin-coated microtiter plates were prewashed once and coated with bio-hOC (1.7 ng/well) in 50 µL of Assay Buffer for 30 min with continuous shaking. The plates were washed four times, and samples or calibrators (30 µL) were added together with 2 ng of Eu-3H8 in 100 µL of Assay Buffer. The plates were incubated for 1 h, after which the wells were washed four times and 200 µL of Delfia Enhancement Solution was added. After incubation for 30 min with shaking, the time-resolved fluorescence was measured.
Calibrators and samples were analyzed in duplicate in all assays.
assay evaluation
We determined assay imprecision with control samples prepared by diluting a urine sample (obtained as the first morning void from an 8-year-old boy) in zero calibrator in two concentrations, low and high (respective OC concentrations, 9.1 and 44.7 µg/L in the U-MidOC assay, 0.08 and 0.55 µg/L in the U-LongOC assay, and 125 and 380 µg/L in the U-TotalOC assay). Within-run imprecision was defined as the variation for 12 replicate determinations and was also evaluated as the mean CV of all the measurements (duplicates) performed for the cohort of elderly women (n = 1019; details below). We determined total imprecision by calculating the variation for control samples from 15 separate assays (25 for the competitive assay). The detection limits (limits of the blank) were determined by use of calibrator free of analyte (zero calibrator) and defined as the concentration corresponding to a signal 2 SD above (two-site assays) or below (competitive assay) the mean of the zero calibrator measured in 12 replicates.
We evaluated dilution linearity by diluting three urine samples into zero calibrator in two-, four-, and eightfold dilutions and calculated the linearity for each dilution as a percentage of the value obtained from the diluted sample compared with the value obtained from undiluted sample. We evaluated the recovery of added analyte by adding a known amount of calibrator peptide to three individual urine samples containing a known amount of OC and calculating the result as the percentage of the added analyte recovered. The exact amounts of added analyte were 79 and 19 µg/L for U-MidOC assay, 10.5 and 2.4 µg/L for the U-LongOC assay, and 219 and 75 µg/L for the U-TotalOC assay.
We determined the ability of each assay to recognize different molecular forms of OC present in human urine by use of OC fragments isolated from a urine sample. The sample was collected as the first morning void from a 11-year-old boy and stored at 70 °C. The total urine OC pool was isolated and fractionated as described previously (15). Briefly, the urine sample was extracted with a C18 solid-phase column followed by immunoaffinity chromatography involving Mabs 6F9, 2H9, and 3H8 (all in one column). Different molecular forms were separated by fractionation on a C4 HPLC column, and collected fractions were analyzed for OC with all three assays. To determine the molecular masses and sequences of isolated fragments, we analyzed fractions containing OC by matrix-assisted laser-desorption/ionization time-of-flight mass spectrometry (MALDI-MS) and N-terminal sequencing as described previously (15).
in vitro stability and circadian and seasonal variations
Fresh first morning void urine was collected from three healthy females and stored at 4 °C until delivered to the laboratory within 2 h after collection. Samples were stored in aliquots either at 4 °C or room temperature (22 °C) for 0 h, 2 h, 4 h, 8 h, 1 day, 3 days, or 5 days. After incubation, samples were frozen and stored at 70 °C until assayed simultaneously.
To study circadian variation, we had nine healthy young females (median age, 24.5 years; interquartile range, 2426 years) following their typical daily routine collect all urine voids during 1 day at 0600 (first morning void), 0900, 1200, 1500, 1800, 2100, and 2400, and the following morning at 0600. First morning voids were first stored at 4 °C and frozen at 70 °C within 2 h. The samples collected during the daytime were immediately stored at 70 °C, and samples collected in the evening were temporarily stored at 20 °C and then at 70 °C the following day.
We evaluated seasonal variations using samples from the cohort of elderly women described below. Urine samples were collected throughout all seasons and stored at 80 °C. The number of samples collected each month from January to December was 96, 118, 111, 127, 83, 53, 32, 50, 88, 104, 112, and 69, respectively.
cohort of elderly women
The elderly women were members of the Malmö OPRA study. For that study, women 75 years of age were randomly selected from the population files of the city of Malmö, Sweden, during 19951999. A total of 1604 women were sent invitations to participate the week after their 75th birthday. Of those, 152 women did not participate because of illness, 376 declined to enter the study, and 32 women were not reached despite several attempts, leaving a total of 1044 participants. No exclusion criteria were used. Informed consent was obtained from all participants, and all parts of the study were approved by the local ethics committee (22). Urine samples were obtained as the first morning void and stored at 80 °C. Nonfasting serum samples were obtained between 0800 and 1300 and stored at 80 °C. Urine was available for 1019 women and serum for 1024 women. Urine samples were analyzed for three U-OC fragments and urine creatinine and serum samples for two markers of bone formation and two markers of bone resorption (details of the measurements are given in the following sections). Bone mineral contents (BMCs) and areal bone mineral densities (aBMDs) of the hip, lumbar spine, and total body were assessed together with serum and urine sampling; results from the hip (femoral neck) were obtained for 951 women, from the lumbar spine for 974 women, and from the total body for 931 women.
other measurements
Serum bone-specific alkaline phosphatase (S-BoneALP) was measured by the Metra BAP immunoassay (Quidel Corporation). Serum intact OC and N-mid-OC (S-TotalOC) was measured by an in-house immunoassay (23). Briefly, S-TotalOC is a one-step sandwich assay based on Mabs bio-2H9 and Eu-6F9 and synthetic OC peptide 149 as a calibrator with within- and between-run CVs <5% and <8%, respectively. Serum C-terminal cross-linked telopeptides of type I collagen (S-CTX) was measured by the Elecsys ß-CrossLaps immunoassay (Roche Diagnostics). Serum tartrate-resistant acid phosphatase 5b (S-TRACP5b) was measured by a solid-phase, immunofixed, enzyme activity assay (within-run CV, 1.8%; between-run CV, 2.2%) as described previously (24). Urine creatinine was measured in accordance with the Jaffe reaction. All measurements of each individual assay were made within the same laboratory. BMC and aBMD of the femoral neck, lumbar spine (L2L4), and total body were assessed by a dual-energy x-ray absorptiometry technique (Lunar® DPX-L).
statistical analysis
All results from U-OC measurements were corrected for urine creatinine before analyses and are presented as the median (interquartile range) unless otherwise stated. The ShapiroWilk test for gaussian distribution and the Levene test for homogeneity of variance were applied for all markers, and because of nongaussian distributions, statistical calculations were done only after logarithmic transformation or with a nonparametric test if logarithmic transformation did not give a gaussian distribution. Comparisons between the groups were made with one-way ANOVA with Tukey or Dunnett post hoc adjustment in the case of gaussian distribution and with the nonparametric Wilcoxon test followed by Bonferroni adjustment when appropriate if distributions were not gaussian. Nonparametric Spearman correlation was used for correlation analysis. The Statistical Analysis System Enterprise Guide 2 program (SAS 8.2; SAS Institute) was used in all analyses, and P <0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
The assays varied in their ability to recognize different naturally occurring molecular forms of U-OC. Prepubertal urine contained two major pools of immunoreactive OC fragments eluting at distinct times during HPLC fractionation (peak I and a twin peak II+III; Fig. 2A
). According to MALDI analysis, the first pool (peak I) consisted of OC fragments starting predominantly from residue Asp14. The monoisotopic mass of the most predominant ion produced by peak I was 2126.93 ([M+H]+; Fig. 2B
), which corresponds to the calculated monoisotopic molecular mass of the human OC fragment containing residues 1431 (theoretical mass, 2126.89 for [M+H]+). In contrast, the double peak II+III contained fragments with more extended amino termini starting from residue Gly7 or Pro9. The predominant ion produced by peak II was 2808.37 ([M+H]+; Fig. 2B
), which corresponds to the calculated molecular mass of the human OC fragment containing residues 731 (theoretical mass, 2808.24 [M+H]+). Theoretical molecular masses were calculated without Gla modification because
-carboxylation is destroyed in MALDI-MS and is not included in the molecular masses of observed ions (25). The identities of the fragments were confirmed by N-terminal sequencing. The U-TotalOC assay detected U-OC fragments present in all three peaks, reflecting the total concentration of immunoreactive OC in urine. The two-site assay U-MidOC recognized only OC fragments present in peaks II and III and failed to detect the more truncated fragments present in peak I. The two-site assay U-LongOC detected only the longest fragments present in peak III, but not as effectively as U-MidOC.
|
biological and preanalytical variability
We evaluated seasonal variation by comparing results obtained from different individuals sampled in different months throughout the year. The number of individuals in each month was different, being smallest for July (n = 32) and largest for April (n = 127). There was no statistically significant trend for U-OC values with regard to sampling month when U-OC was measured with either the U-LongOC (P = 0.25, nonparametric Wilcoxon test) or U-TotalOC (P = 0.93; Fig. 3B
) assay. For the U-MidOC values, there was a minor trend (P = 0.02) for greater values in autumn compared with the rest of year (Fig. 3A
). In contrast, a circadian rhythm was observed for U-OC with all three assays (Fig. 3
, C and D). U-OC values were highest during the night and early morning and decreased in the afternoon. This was followed by an increase later in the evening and night.
|
The results of stability studies are shown in Table 2
. The immunoreactivity detected by the two-site assays U-LongOC and U-MidOC did not differ from the initial value when urine samples were stored at either 4 °C or 22 °C for up to 5 days (P = 0.190.99, one-way ANOVA with Dunnett test). Concentrations measured by the U-TotalOC assay appeared to be lower after incubation at either 4 or 22 °C compared with the initial values. The decrease in immunoreactivity was more pronounced at 4 °C and for the longest timepoints, i.e., 3 and 5 days (P = 0.0001 and <0.0001, respectively, one-way ANOVA with Dunnett test). In addition, results obtained already after 4 h (P = 0.02) and 1 day (P = 0.04) were different from the initial value. We also observed a similar trend in samples incubated at 22 °C, but it was less pronounced (presumably because of higher variation), and the difference compared with the initial value did not reach statistical significance until after 5 days (P = 0.01).
|
association with bone mass
U-OC concentrations measured with the U-MidOC, U-LongOC, or U-TotalOC assays had a statistically significant negative correlation with BMC measured at the total body, spine, or hip (P <0.0001 for all associations). Women were further classified into four groups according to U-OC values, 1 being the lowest quartile of U-OC and 4 the highest, and the total body BMC in each quartile group is shown in Fig. 4A
. There was a significant negative association between BMC and U-OC quartiles (P <0.0001 for all assays, Wilcoxon nonparametric test), and BMC values were lower in the high turnover quartiles. There was also a significant negative association between T-scores for femoral neck BMD and U-OC (Fig. 4B
). In addition, the percentage of individuals with a T-score below 2.5 was greater (
3942%) in quartile 4 than in quartile 1 (
2226%; P <0.001). The results were similar whether U-OC was measured with the U-MidOC, U-LongOC, or U-TotalOC assay.
|
correlation with bone turnover rate
U-OC correlated significantly (P <0.0001) with bone turnover rate assessed by the serum markers of bone turnover used in this study (Table 3
). The outcome of correlation analysis with the U-MidOC and U-LongOC assays was nearly identical. U-MidOC and U-LongOC values were strongly correlated with the bone resorption marker S-CTX, and the r value for U-OC and S-CTX was 0.51 (95% confidence interval, 0.460.56) for both U-OC assays. U-MidOC and U-LongOC results also were strongly correlated with S-TRACP5b [r = 0.44 (0.390.49) and 0.41 (0.410.46), respectively]. Interestingly, the correlation of U-MidOC and U-LongOC with the bone formation marker S-BoneALP appeared to be lower [r = 0.38 (0.320.43) and 0.36 (0.300.41), respectively] than the correlation with resorption markers. Correlation coefficients for U-TotalOC and bone markers were slightly smaller. The r value for U-TotalOC and S-CTX was 0.38 (0.330.43), and that for U-TotalOC and S-TRACP5b was also 0.38 (0.330.43). The correlation coefficient for U-TotalOC and S-BoneALP was only marginally smaller, 0.30 (0.240.36).
|
We also evaluated the correlation between S-OC and U-OC. U-MidOC and U-LongOC values were strongly correlated with S-totalOC [r = 0.57 (0.520.61) and 0.55 (0.510.59), respectively]. The correlation between U-TotalOC and S-totalOC [r = 0.30 (0.230.35)] was lower and equal to the correlation for U-TotalOC and S-BoneALP [r = 0.30 (0.240.36)].
In addition to the large cohort of elderly women, U-OC was also evaluated in a smaller cohort of pre- and postmenopausal women (n = 91), and U-OC detected with any of these assays was associated with bone turnover rate and was able to distinguish between postmenopausal women receiving or not receiving hormone replacement therapy (See the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol51/issue3/).
| Discussion |
|---|
|
|
|---|
Immunoreactive U-OC can be classified into at least two main categories according to the size of fragments, the longer fragments starting from residues Gly7 or Pro9, and the shorter ones having residue Asp14 or Leu16 at the amino terminus (15). We developed assays that detected longer and shorter fragments (U-TotalOC) or longer fragments exclusively (U-MidOC and U-LongOC), and U-OC measured with all three assays was able to provide insight into bone metabolism in the cohort of elderly women. In this cohort, U-OC correlated significantly with other markers of bone turnover, namely, S-CTX, S-TRACP5b, S-BoneALP, and S-TotalOC, emphasizing the connection between U-OC and bone turnover rate. However, although we observed a significant correlation between U-OC and conventional serum markers of bone turnover, the correlation between different markers generally appeared to be rather modest [Spearman correlation (r), 0.300.57], which should also be taken into account when indexes to assess the balance between resorption and formation are considered (1)(26). The distinct significant association between total body BMC and bone metabolism assessed by U-OC quartiles further demonstrates the connection between U-OC and skeletal metabolism.
In addition to the association with BMC, we also evaluated the association of U-OC with aBMD assessed at the femoral neck. Femoral neck BMD as assessed by dual-energy x-ray absorptiometry is currently the most widely used index to assess bone and the key diagnostic tool in osteoporosis (27)(28). It should be recognized, however, that the assessment of regional BMD in a particular skeletal site reflects only events taking place at that site and thus only in a fraction of the skeleton; it also provides information only in two dimensions. Bone turnover markers reflect the total body skeletal metabolism; thus, it may not be physiologically correct to relate biochemical markers such as U-OC to BMD. A topic for further studies would be the combination of short-term monitoring of U-OC, or turnover in general, with long-term follow-up of changes in BMC or BMD to understand the correlation between bone markers and bone mass.
The performance of all three U-OC assays was highly similar in the tested groups, and we found no clear evidence for qualitative differences with regard to U-OC fragments in different clinical samples. In particular, the discriminatory power in the quartile analysis and the correlation with bone turnover markers were nearly identical for U-MidOC and U-LongOC (which actually detects a small subset of fragments detected by U-MidOC). The U-OC assays were simple to perform, and the detection limits and dynamic ranges were adequate for measurement of urine samples, with the exception of a few too dilute urine samples, which were excluded from the evaluations. The variations in the competitive U-TotalOC assay were greater than for the two-site assays, and the total variation for U-TotalOC was particularly high (>20%). This was observed primarily for the control samples, and the calibration curve for the U-TotalOC assay constructed with synthetic OC was highly reproducible with overlapping calibration curves and within-assay CVs <10%. The high total variation may partly be attributable to the control samples, which were prepared by diluting urine containing a very high concentration of U-OC and collected from a prepubertal boy. The molecular fragments generated in this setting and their relative proportions may be different from those of postmenopausal women and older individuals, and a premenopausal undiluted adult urine sample or urine obtained from, e.g., a postmenopausal woman shortly after ovariectomy would probably serve as a more appropriate model for control samples in the future. By reducing the analytical variability by careful validation, the value of U-OC in clinical settings could become even more pronounced.
Measurements were performed with urine samples collected as the first morning void, but a second morning void and 24-h urine collections were used as well and found to be appropriate sample material (data not shown). The concentrations were highest and most easily measurable in the first morning voids, which can be recommended as a sample material for further studies. Furthermore, the immunoreactivity in urine samples seemed to tolerate short-term storage at 4 °C or even room temperature before freezing at 20 or 70 °C. This in vitro stability of OC in urine facilitates the collection and storage of urine samples, and in this aspect U-OC appears to be superior to S-OC. However, U-OC may exhibit some prestorage variability resulting from sample handling immediately after the collection. Despite the carefully controlled collection and storage of urine samples in the stability study, we were unable to address this aspect in the current study in detail, and this remains to be evaluated.
Circulating OC concentrations demonstrate a circadian rhythm characterized by a decrease during the morning to a low around noon. This is followed by a gradual increase that peaks after midnight. The difference between the highest and lowest values ranges from 10% to 20% depending on the assay (29)(30). Pronounced diurnal variation has also been observed for resorption markers such as urinary collagen type I cross-linked N-telopeptides (NTx) (31) and CTX (32). Resorption markers reach their maximum and minimum at around 0500 and 1400, respectively, and the magnitude of variation appears to be higher for collagen degradation products than, for example, for S-OC (32). Taylor et al. (12) found a diurnal variation of U-OC after the one reported for S-OC, i.e., high concentrations in the morning and low concentrations in the afternoon. Our results with all three U-OC assays were consistent with this rhythm.
In addition to circadian variability, minor seasonal fluctuations in S-OC, with concentrations decreasing from January to July and then increasing to a peak in winter, have been reported (30)(33). More recent studies, however, have found no indication of an important wintertime increase and only a borderline summertime increase, if any may occur (34). The effect of sampling season on urine OC values has not been investigated previously. We studied the seasonal variability in the cross-sectional study of elderly women and did not detect any major seasonal variation in U-OC concentrations. There was a small but statistically significant tendency for higher values in late autumn and winter when U-OC was measured with the U-MidOC assay, which might be related to the summertime increase in sunlight and vitamin D. The results obtained with the two other assays displayed no significant variability regarding the month of sampling, which may be related to the lower sensitivities of these assays to detect minor changes. Nevertheless, the seasonal effects appeared to be rather small in comparison with other sources of measurement noise and not of sufficient magnitude to confound the use of U-OC measurements.
The clinical properties of the U-OC assays appear to be distinct from those of S-OC assays. The correlations between the U-OC assays and the well-known resorption marker S-CTX were significantly higher, without overlapping confidence limits, than the correlation with the bone formation marker S-BoneALP. The correlations between the U-OC assays and another resorption marker, S-TRACP5b, also tended to be higher than those between the U-OC assays and S-BoneALP, but the confidence limits were slightly overlapping. We also recently showed that U-MidOC and U-LongOC assay results were predictive for fractures in a large prospective study of elderly women (22). Briefly, when women in the highest quartile of bone turnover markers were compared with all other women, the odds ratios for sustaining a clinical vertebral fracture were 2.71 (95% confidence interval, 1.504.89) for the U-MidOC and 2.75 (1.524.96) for the U-LongOC assays. Together with the resorption markers S-CTX and S-TRACP5b, these two U-OC assays were the only bone turnover markers that were statistically significant, and the odds ratios for the U-OC assay results were comparable to the odds ratios for S-CTX [1.94 (1.053.58)] and S-TRACP5b [2.28 (1.264.15)]. Interestingly, the four S-OC assays evaluated were not predictive for fractures. Differences between U-OC and S-OC have also been published by others. Gundberg et al. (35) observed discordance between the measurements of S-OC and urinary excretion of its breakdown product (Gla) in patients with Paget disease, which is characterized by high bone turnover with an increased number and appositional rate of osteoblasts seen in the affected sites with no net bone loss (36). In such patients, urinary excretion of Gla was normal, but S-OC concentrations were increased threefold (35). More recently, Åkesson et al. (37) monitored patients after tibia osteotomy (n = 14) and found that U-OC values had kinetics similar to those of resorption markers such as urinary deoxypyridinoline and distinct from the kinetics of S-OC. Different kinetics were also observed for S-OC and U-OC when alendronate was administered to osteoporotic postmenopausal women (n = 19), and the changes in U-OC, but not in S-OC, were correlated with changes in resorption markers (serum or urinary collagen NTx and CTX) (14). We have observed similar differences in the kinetics of U-OC and S-OC in the follow-up of postmenopausal women (n = 76) after alendronate treatment in a placebo-controlled alendronate intervention trial (38).
Osteoclastic bone resorption in vitro releases immunoreactive OC, both intact and fragmented molecules, from bone matrix (10)(39). Because bone hydroxyapatite is able to absorb only intact OC molecules but not OC fragments (40)(41), circulating intact OC, whether released from biosynthesis in osteoblasts or derived from resorption, may traffic back to the skeleton. Circulating intact OC, which is not adsorbed to hydroxyapatite, and the largest OC fragments are likely to be degraded by proteases in the circulation or peripheral organs (11)(42). Small OC fragments are more resistant to degradation and probably accumulate in the urine as midmolecule fragments because the middle portion of the molecule is more resistant to proteolysis as a result of structural properties of the molecule (15)(43)(44). Because of the rapid clearance from the circulation to bone or by proteolytic activity, S-OC probably reflects rather acute changes in bone metabolism, whereas OC fragments accumulating in the urine serve more as an index of basal bone turnover rate, most likely predominantly resorption.
In summary, we have developed three immunoassays for the detection of OC fragments in urine. The assays have unique specificities toward different naturally occurring U-OC fragments. U-OC concentrations followed a circadian rhythm similar to the one reported for S-OC, but we observed no major seasonal variations. High concentrations of OC in urine were associated with low bone mass, and U-OC detected with any of these assays correlated with bone turnover rate assessed by conventional serum markers of bone metabolism. Furthermore, there is accumulating evidence that U-OC and S-OC assays give different information concerning bone turnover in clinical settings. The measurement of U-OC thus provides an additional means to monitor bone metabolism and may have applications in diagnostics related to disorders of bone metabolism.
| Acknowledgments |
|---|
| Footnotes |
|---|
-carboxyglutamic acid; MALDI-MS, matrix-assisted laser desorption/ionization time-of-flight mass spectrometry; BMC, bone mineral content; aBMD, areal bone mineral density; S-BoneALP, serum bone-specific alkaline phosphatase; CTX and NTx, C- and N-terminal cross-linked telopeptides of type I collagen; and TRACP5b, tartrate-resistant acid phosphatase isoenzyme 5b. | References |
|---|
|
|
|---|
-carboxyglutamic acid and serum osteocalcin as bone markers: studies in osteoporosis and Pagets disease. J Clin Endocrinol Metab 1983;57:1221-1225.
-carboxyglutamic acid-containing protein from bone. Proc Natl Acad Sci U S A 1976;73:1447-1451.The following articles in journals at HighWire Press have cited this article:
![]() |
K. K. Ivaska, J. Lenora, P. Gerdhem, K. Akesson, H. K. Vaananen, and K. J. Obrant Serial Assessment of Serum Bone Metabolism Markers Identifies Women with the Highest Rate of Bone Loss and Osteoporosis Risk J. Clin. Endocrinol. Metab., July 1, 2008; 93(7): 2622 - 2632. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Kraenzlin Biochemical Markers of Bone Turnover and Osteoporosis Management IBMS BoneKEy, July 1, 2007; 4(7): 191 - 203. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. K. Ivaska, K. Pettersson, A. Nenonen, K. Uusi-Rasi, A. Heinonen, P. Kannus, and H. K. Vaananen Urinary Osteocalcin Is a Useful Marker for Monitoring the Effect of Alendronate Therapy Clin. Chem., December 1, 2005; 51(12): 2362 - 2365. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |