Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 51: 1054-1057, 2005; 10.1373/clinchem.2005.048199
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (9)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Halldórsdóttir, A. M.
Right arrow Articles by Eby, C. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Halldórsdóttir, A. M.
Right arrow Articles by Eby, C. S.
Related Collections
Right arrow Lipids, Lipoproteins, and Cardiovascular Risk Factors
(Clinical Chemistry. 2005;51:1054-1057.)
© 2005 American Association for Clinical Chemistry, Inc.


Technical Briefs

Soluble CD40 Ligand Measurement Inaccuracies Attributable to Specimen Type, Processing Time, and ELISA Method

Anna Margrét Halldórsdóttir1, Joshua Stoker2, Rhonda Porche-Sorbet1 and Charles S. Eby1,a

Divisions of1 Laboratory Medicine and 2 Cardiology, Washington University School of Medicine, St. Louis, MO;

aaddress correspondence to this author at: Washington University School of Medicine, St. Louis, MO 63110; fax 314-362-1461, e-mail eby{at}pathbox.wustl.edu

CD40 ligand (CD40L) is a member of the tumor necrosis factor superfamily and is produced in a variety of cells, including platelets. The soluble form (sCD40L) is a mediator of both inflammatory and hemostasis processes and has been implicated in the pathogenesis of atherosclerosis. Clinical studies have revealed increased sCD40L in patients with unstable angina (1) and identified an association between increased sCD40L and future risk for death or nonfatal myocardial infarction (2)(3). While prospectively measuring sCD40L in a cohort of persons at risk for cardiovascular complications, we identified both preanalytical and analytical sources of error. This report documents the effects of specimen type (serum and plasma), processing (time and temperature), and commercial reagent selection on sCD40L ELISA results. These findings raise concerns about the accuracy of sCD40L results reported in recent clinical studies.

After obtaining informed consent, we enrolled 147 patients older than 60 years referred for diagnostic cardiac catheterization in an Institutional Review Board-approved study to evaluate the value of clinical, echocardiographic, and biomarker variables for prediction of future cardiovascular complications. When combined with clinical predictors, B-type natriuretic peptide and C-reactive protein, but not sCD40L, were independent predictors of death or cardiovascular hospitalization at 6 months (data not shown). The unexpectedly poor correlation between undiluted plasma sCD40L results and clinical outcomes in this study motivated us to perform the following investigations.

Venous blood from the 147 study participants [mean (SD) age, 70.8 (6.9) years] was collected into plastic tubes containing tripotassium EDTA (BD Vacutainer; Becton Dickinson) before catheterization and placed on ice for 1–4 h before processing. Blood from 10 control individuals [mean (SD) age, 38.7 (8.4) years] was collected into both EDTA and plain glass tubes (Becton Dickinson) and maintained at room temperature for 30 min before processing. Study and control samples were centrifuged twice: first at 2790g for 5 min to separate cells from the plasma/serum and then at 16 000g for 3 min to remove any residual platelets. Supernatants were aliquoted and stored at –70 °C.

We assessed the effects of time and temperature on measured sCD40L concentrations by collecting whole blood from a single healthy individual into a syringe and immediately aliquoting it into EDTA-containing and plain glass tubes. For every time point analyzed, plasma and serum tubes were kept on cells at room temperature, and in 1 experiment they were also kept on ice. Samples were then centrifuged and stored as described above. Selected samples from the time–temperature experiment were either filtered through a 0.2 µm syringe filter or were ultracentrifuged at 200 000g for 4 h at 4 °C to remove potential remaining platelet microparticles before repeat sCD40L testing.

sCD40L concentrations were measured with an sCD40L assay (Quantikine®; R&D Systems). According to the package insert, the R&D ELISA is suitable for measuring sCD40L in serum and plasma and is linear within the analytical range of the assay (0.0625–4 µg/L). The stated lower limit of detection is 0.0042 µg/L. Reference intervals for serum (0.675–38.373 µg/L; mean, 8.273 µg/L; n = 44) and platelet-poor EDTA plasma (0.106–11.831 µg/L; mean, 2.987 µg/L; n = 16) were provided by the manufacturer. The manufacturer’s protocols were followed.

We retested selected samples with 2 sCD40L assays (BMS 239 and BMS 293) from Bender MedSystems. The BMS 239 is suitable only for testing serum, whereas the BMS 293 is a high-sensitivity assay designed both for plasma and serum (package inserts). The lower limit of detection for the BMS 239 is 0.095 µg/L, and the lower limit of detection for the BMS 293 is 0.005 µg/L. The manufacturer’s protocols were followed.

When establishing a central 95% interval for sCD40L with 10 control plasma samples diluted 1:5, per the R&D package insert, we found that all results were below the lowest point on the calibration curve (0.0625 µg/L). Following discussions with the manufacturer’s technical consultants, we tested control and patient plasma samples undiluted. The distribution of sCD40L concentrations in undiluted plasma for the 147 patients is shown in Fig. 1A . When the 2 control and 8 patient plasma samples with sCD40L concentrations exceeding the upper limit of the calibration curve were diluted 1:2 and 1:5 in calibration diluent, the results were not linear (data not shown). To determine whether the nonlinear dilution response was a systematic analytical problem, we added recombinant sCD40L to serum and plasma samples with sCD40L concentrations <0.2 µg/L to produce a predicted concentration of 2 µg/L. Serial 2-fold dilutions of these samples in the calibration diluent also produced nonlinear results (data not shown).



View larger version (20K):
[in this window]
[in a new window]
 
Figure 1. sCD40L results depend on specimen type/processing and ELISA method.

(A), distribution of undiluted plasma sCD40L concentrations in 147 patients undergoing cardiac catheterization. Each data point represents 1 patient. Specimens were tested undiluted. Dotted lines represent the dynamic range of the assay. (B and C), comparison of 2 commercial sCD40L ELISA tests, using plasma and serum specimens from the time-temperature experiments. Dashed lines represent the lines of unity. Results have been corrected for dilution. (B), comparison of the R&D assay with the Bender BMS 239 for serum samples. The results of the regression analysis were as follows: y = 0.96x + 0.42 µg/L. (C), comparison of the R&D ELISA with the Bender BMS 293 for plasma samples. The results of the regression analysis were as follows: y = 0.96x + 6.08 µg/L. (D), effect of sample storage conditions on measured sCD40L concentrations. Each data point represents the mean of 2 experiments, the error bars represent SD. Serum ({circ}) and plasma (•) samples drawn from a healthy donor were stored on cells at room temperature for different lengths of time before processing. The graph shows the change in measured sCD40L concentrations with time. Results have been corrected for dilution.

Interestingly, when we compared the values for undiluted serum and plasma samples from the control group, the mean sCD40L concentration in undiluted serum (1.33 µg/L) was 6-fold higher than in undiluted plasma (0.24 µg/L). There was a weak correlation (r = 0.223) between serum and plasma concentrations for undiluted samples.

To evaluate the correlation between sCD40L concentrations measured by ELISAs from different manufacturers, we compared results obtained for selected serum and plasma samples by the R&D ELISA and the Bender BMS MedSystems ELISAs. The R&D and Bender BMS 239 assays showed good correlation for serum samples (Fig. 1BUp ), but for plasma, the correlation was poor between the new Bender high-sensitivity BMS 293 assay and the R&D ELISA (Fig. 1CUp ).

Finally, we investigated the impact of the processing variables time and temperature on sCD40L determinations. When serum from a healthy donor was stored on cells at room temperature, there was a 6- to 7-fold increase in sCD40L concentrations after 180 min (Fig. 1DUp ). sCD40L concentrations in similarly treated plasma samples did not increase, and most values were below the analytical range. When either serum or plasma was stored on ice, no increase in sCD40L concentration was observed over 180 min (data not shown).

To examine whether the time-dependent increase in serum sCD40L concentrations was attributable to release of platelet microparticles producing membrane sCD40L, we either filtered or ultracentrifuged specimens before repeat testing. No difference was observed (data not shown).

Accurate measurement of an analyte is essential for its clinical diagnostic utility. Despite reports showing an association between increased sCD40L and cardiovascular complications (2)(3), there is poor agreement among studies regarding sCD40L ranges for controls or cases with similar cardiovascular risk factors (1)(2)(3)(4)(5)(6)(7). In addition, the methods sections in some reports fail to specify whether serum or plasma was tested or whether plasma was tested with an ELISA designed for testing serum, and they provide few sample-processing details (2)(3)(6)(8).

After reviewing the literature and the manufacturers’ product specifications, we decided to measure plasma sCD40L with the R&D ELISA. We were disappointed to discover that the only sCD40L ELISA suitable for plasma testing at that time lacked the sensitivity to measure sCD40L in 100% of controls when plasma was diluted 1:5, according to the manufacturer’s recommendations.

When we tested undiluted control and patient plasma samples, 10 of 157 (6%) gave exceedingly high results, and serial dilutions of these specimens produced nonlinear results. After we shared these findings with the manufacturer, the R&D sCD40L ELISA was briefly withdrawn from the market while changes were made in the assay diluent to address the presence of heterophilic antibodies in some samples. However, no changes were made to increase the sensitivity of the assay.

Most clinical studies have measured sCD40L in serum with either the R&D or Bender 293 ELISAs, reducing the problem of analytical insensitivity. Thom et al. (9) reported that mean measured sCD40L concentrations were 9-fold higher in serum than in plasma when assayed with a Bender sCD40L ELISA, which is consistent with our results. In addition, we have shown that the agreement between the R&D and Bender 239 ELISA methods for measuring sCD40L in serum was excellent (Fig. 1BUp ).

However, we have also shown that the serum sCD40L concentration increases significantly with time in samples stored at room temperature (Fig. 1DUp ). This is in agreement with previously published data (9)(10) and represents the combination of in vivo sCD40L, which is likely to be the physiologically relevant component, and ex vivo-released sCD40L.

Platelets are activated during the process of clot retraction, and sCD40L shedding from the platelet surface probably accounts for the progressive increase in serum concentrations. Shed sCD40L could be bound to platelet microparticles (7). However, in our experiments, filtration and ultracentrifugation did not lead to a decrease in serum concentrations of sCD40L, suggesting that ex vivo-released sCD40L is not bound to intact membrane. It may therefore be impossible to distinguish between in vivo and ex vivo release of sCD40L.

The measurement of sCD40L concentrations in human blood with the R&D ELISA is therefore problematic for the following reasons: the assay lacks sensitivity for measuring sCD40L concentrations in diluted plasma samples; testing of serum is problematic because of ex vivo release of sCD40L; there is poor correlation between plasma and serum samples; and the linearity of measurements obtained with the reformulated assay reagents has not been evaluated.

Recently, Bender MedSystems began selling a high-sensitivity sCD40L ELISA (Bender 293) suitable for plasma and serum testing. A preliminary evaluation confirmed that it is more sensitive than the R&D sCD40L ELISA test for plasma, but no further studies have been performed.

In summary, investigators should carefully consider the choice of specimen type, specimen-handling procedures, and properties of the commercial ELISA tests when measuring sCD40L concentrations in blood because each of these variables can critically affect measured sCD40L concentrations. The optimum strategy would be to measure sCD40L in platelet-free plasma by a sensitive analytical method.


Acknowledgments

We thank Bender MedSystems (Vienna, Austria) for supplying the high-sensitivity sCD40L ELISA.


References

  1. Aukrust P, Müller F, Ueland T, Berget T, Aaser E, Brunsvig A, et al. Enhanced levels of soluble and membrane-bound CD40 ligand in patients with unstable angina. Circulation 1999;100:614-620.[Abstract/Free Full Text]
  2. Heeschen C, Dimmeler S, Hamm C, van den Brand M, Boersma E, Zeiher A, et al. Soluble CD40 ligand in acute coronary syndromes. N Engl J Med 2003;348:1104-1111.[Abstract/Free Full Text]
  3. Varo N, de Lemos J, Libby P, Morrow D, Murphy S, Nuzzo R, et al. Soluble CD40L. Risk prediction after acute coronary syndromes. Circulation 2003;108:1049-1052.[Abstract/Free Full Text]
  4. Schönbeck U, Varo N, Libby P, Buring J, Ridker P. Soluble CD40L and cardiovascular risk in women. Circulation 2001;104:2266-2268.[Abstract/Free Full Text]
  5. Viallard J, Solanilla A, Gauthier B, Contin C, Déchanet J, Grosset C, et al. Increased soluble and platelet-associated CD40 ligand in essential thrombocythemia and reactive thrombocytosis. Blood 2002;99:2612-2614.[Abstract/Free Full Text]
  6. Aggarwahl A, Schneider D, Terrien E, Sobel B, Dauerman H. Increased coronary arterial release of interleukin-1 receptor antagonist and soluble CD40 ligand indicative of inflammation associated with culprit coronary plaques. Am J Cardiol 2004;93:6-9.[ISI][Medline] [Order article via Infotrieve]
  7. Jinchuan Y, Zonggui W, Jinming C, Li L, Xiantao K. Upregulation of CD40-CD40 ligand system in patients with diabetes mellitus. Clin Chim Acta 2003;339:85-90.
  8. Yan J, Zhu J, Gao L, Wu Z, Kong X, Zong R, et al. The effect of elevated serum soluble CD40 ligand on the prognostic value in patients with acute coronary syndromes. Clin Chim Acta 2004;343:155-159.[CrossRef][Medline] [Order article via Infotrieve]
  9. Thom J, Gilmore G, Yi Q, Hankey J, Eikelboom J. Measurement of soluble P-selectin and soluble CD40 ligand in serum and plasma. J Thromb Haemost 2004;2:2067-2069.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  10. Nannizzi-Alaimo L, Rubenstein M, Alves V, Leong G, Phillips D, Gold H. Cardiopulmonary bypass induces release of soluble CD40 ligand. Circulation 2002;105:2849-2854.[Abstract/Free Full Text]



The following articles in journals at HighWire Press have cited this article:


Home page
Eur Heart JHome page
D. A. Morrow, M. S. Sabatine, M.-L. Brennan, J. A. de Lemos, S. A. Murphy, C. T. Ruff, N. Rifai, C. P. Cannon, and S. L. Hazen
Concurrent evaluation of novel cardiac biomarkers in acute coronary syndrome: myeloperoxidase and soluble CD40 ligand and the risk of recurrent ischaemic events in TACTICS-TIMI 18
Eur. Heart J., May 1, 2008; 29(9): 1096 - 1102.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
B. T. Ivandic, E. Spanuth, D. Haase, H.-G. Lestin, and H. A. Katus
Increased Plasma Concentrations of Soluble CD40 Ligand in Acute Coronary Syndrome Depend on in Vitro Platelet Activation
Clin. Chem., July 1, 2007; 53(7): 1231 - 1234.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
F. S. Apple, L. A. Pearce, A. Chung, R. Ler, and M. M. Murakami
Multiple Biomarker Use for Detection of Adverse Events in Patients Presenting with Symptoms Suggestive of Acute Coronary Syndrome
Clin. Chem., May 1, 2007; 53(5): 874 - 881.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
D. A. Morrow and J. A. de Lemos
Benchmarks for the Assessment of Novel Cardiovascular Biomarkers
Circulation, February 27, 2007; 115(8): 949 - 952.
[Full Text] [PDF]


Home page
CirculationHome page
R. S. Vasan
Biomarkers of Cardiovascular Disease: Molecular Basis and Practical Considerations
Circulation, May 16, 2006; 113(19): 2335 - 2362.
[Full Text] [PDF]


Home page
Clin. Chem.Home page
M. Weber, B. Rabenau, M. Stanisch, A. Elsaesser, V. Mitrovic, C. Heeschen, and C. Hamm
Influence of Sample Type and Storage Conditions on Soluble CD40 Ligand Assessment
Clin. Chem., May 1, 2006; 52(5): 888 - 891.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (9)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Halldórsdóttir, A. M.
Right arrow Articles by Eby, C. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Halldórsdóttir, A. M.
Right arrow Articles by Eby, C. S.
Related Collections
Right arrow Lipids, Lipoproteins, and Cardiovascular Risk Factors


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS