Clinical Chemistry 51: 1070-1071, 2005; 10.1373/clinchem.2005.048215
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bakker, A. J.
Right arrow Articles by Zijlstra, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bakker, A. J.
Right arrow Articles by Zijlstra, A.
Related Collections
Right arrow Clinical Immunology
Right arrow Proteomics and Protein Markers
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2005;51:1070-1071.)
© 2005 American Association for Clinical Chemistry, Inc.


Letters to the Editor

(Micro)Albuminuria: Antigen Excess Detection in the Roche Modular Analyzer

Andries J. Bakkera, Ageeth Bierma-Ram, Halbe Keidel, Haye Syperda and Appie Zijlstra

Department of Clinical Chemistry, Klinisch Chemisch Laboratorium, Leeuwarden, The Netherlands

aAddress correspondence to this author at: Department of Clinical Chemistry, Klinisch Chemisch Laboratorium, PO Box 850, 8901 BR Leeuwarden, The Netherlands. Fax 31-582882227; e-mail a.j.bakker{at}kcl.znb.nl.


To the Editor:

Measurement of urinary albumin is performed both for detection of microalbuminuria (1) and for establishing the selectivity of urinary protein excretion when renal function deteriorates (2); it thus requires assays that can measure albumin concentrations from a few milligrams per liter to several grams per liter. Falsely low concentrations are encountered regularly in immunoturbidimetric assays as a result of antigen excess (3). Procedures to detect antigen excess include additional testing with qualitative measurement of urinary albumin by reagent strip (dipstick) (4)(5)(6), quantitative measurement of urinary total protein (7), or albumin analysis with multiple dilutions of the sample (8). Another procedure uses an excessive amount of antibody, which can lengthen the calibration curve considerably (9). All of these procedures are costly. Mathematical procedures based on the kinetic differences between samples with and without antigen excess have been proposed (9)(10), but they falsely suggest antigen excess in samples with low antigen concentrations. An efficient procedure involves adding additional albumin after the measurement is completed and interpreting the additional change in turbidity (9). Here we report our experience with such a procedure on the MODULAR® ANALYTICS P-type analyzer (Roche GmbH).

For measurement of urinary albumin [modified for Modular from Ref. (11)], we incubated a sample with 45 g/L PEG-6000 in phosphate-buffered saline (reagent 1, consisting of 10 mmol/L phosphate, 140 mmol/L NaCl, 1 mL/L Triton X-100, pH 7.4) for 1.5 min. We then added an antibody against human serum albumin (reagent 2, consisting of Dako Q328 diluted 10-fold in saline) and monitored the development of turbidity at 18-s intervals for the next 3 min to establish the albumin concentration. We then added more albumin (reagent 3, consisting of an equivalent amount of 200 mg/L albumin prepared by dilution of a human control serum in saline) and monitored the change in turbidity for an additional 1.5 min; this served as the so-called "prozone check". When the absorbance of the analyzer-generated prozone is less than the lower cutoff limit (0.200), a flag is issued, which initiates an automatic rerun of the 10-fold diluted sample. The detailed settings for the MODULAR ANALYTICS P-type analyzer are given in Table 1 in the Data Supplement that accompanies the online version of this letter at http://www.clinchem.org/content/vol51/issue6/. The calibrators (2–600 mg/L) consisted of purified human albumin (prod. no. A3782; Sigma) dissolved in phosphate-buffered saline for which the concentrations had been established as described previously (11).

The slope of the calibration curve (sensitivity) was lower at higher concentrations, and the absorbance decreased at concentrations above ~1200 mg/L (Fig. 1A ). The additional absorbance change for the prozone check after addition of excess antigen, the so-called prozone value, increased slightly with increasing albumin concentrations to ~250–300 mg/L (Fig. 1A ). This increasing prozone value could be explained by the fact that the absorbance change produced by the endogenous albumin in the sample may not be complete at 3 min after addition of reagent 2 and therefore contributes to the prozone turbidity. At higher albumin concentrations, the prozone value decreased, and when the curve reached concentrations ≥900–1200 mg/L, the prozone value was reduced to almost 0. Although this system worked with dilutions of a human serum, we also tested the procedure with urine samples from diabetic patients and from patients with renal failure. During the first 2 weeks of the investigation of the prozone measurement, the prozone value calculated by the Modular analyzer was recorded for all samples. Because the frequency of samples with microalbuminuric results was considerably lower than the frequency of samples with normoalbuminuric results, the prozone value was recorded for samples with microalbuminuric results for an additional 2 months and for samples with macroalbuminuric results for another 2 months. In total, 167 samples with normoalbuminuric results, 168 samples with microalbuminuric results, and 140 samples with macroalbuminuric results were assayed. The results for the prozone values, including those obtained in the rerun, are displayed in Fig. 1B .



View larger version (26K):
[in this window]
[in a new window]
 
Figure 1. Absorbance and prozone numbers for diluted control serum (A), and prozone numbers for urine samples (B).

(A), shown is the absorbance ({diamondsuit}) for various dilutions of a control serum measured in the assay for urinary albumin as well as the prozone number obtained for each dilution ({blacktriangleup}). (B), prozone number ({blacktriangleup}) obtained for each albumin concentration in all urine samples.

This procedure with the 0.200 cutoff value for the prozone check effectively detected all samples in which antigen excess caused falsely low microalbuminuric or normal values. The same kind of procedure could be useful for other assays in which antigen excess might cause falsely low values. In our experience, falsely low values attributable to antigen excess occur in daily practice in a limited number of samples, which makes an efficient procedure for antigen excess detection very important.

In conclusion, an effective and efficient procedure for the detection of antigen excess could be included in the assay for urinary albumin on the Roche Modular analyzer and could eliminate the need for additional tests to prevent reporting of falsely low results from antigen excess.


Footnotes

Editor’s note: The manufacturer has promised a reply, but none has been received at press time.


References

  1. Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parriott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002;48:436-472.[Abstract/Free Full Text]
  2. Newman DJ, Price CP. Renal function and nitrogen metabolites. Burtis CA Ashwood ER eds. Tietz textbook of clinical chemistry, 3rd ed 1999:1204-1270 WB Saunders Philadelphia. .
  3. Deaton CD, Maxwell KW, Smith RS. Laser nephelometry. Ritchie RF eds. Automated immunoanalysis 1978:375-407 Marcel Dekker New York. .
  4. Rifai N, Gubar K, Silverman LM. Immunoturbidimetry: an attractive technique for the determination of urinary albumin and transferrin. Clin Biochem 1987;20:179-181.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  5. Krämer BK, Jesse U, Ress KM, Schmülling R-M, Risler T. Enzyme-linked immunosorbent assay for urinary albumin at low concentrations. Clin Chem 1987;33:609-611.[Abstract/Free Full Text]
  6. Paloheimo L, Pajari-Backas M, Pitkänen E. Evaluation of an immunoturbidimetric microalbuminuria assay. J Clin Chem Clin Biochem 1987;25:889-892.[ISI][Medline] [Order article via Infotrieve]
  7. Jury DR, Mikkelsen DJ, Dunn PJ. Prozone effect and the immunoturbidimetric measurement of albumin in urine. Clin Chem 1990;36:1518-1519.[Free Full Text]
  8. Landgraf-Leurs MMC, Modi E, Horn K, Landgraf R. Immunoturbidimetric assay for the determination of microalbuminuria using the Hitachi analyser. J Clin Chem Clin Biochem 1987;25:683-687.[ISI][Medline] [Order article via Infotrieve]
  9. Anderson RJ, Sternberg JC. A rate nephelometer for immunoprecipitin measurement of specific proteins. Ritchie RF eds. Automated immunoanalysis 1978:409-469 Marcel Dekker New York. .
  10. Deveril I. Kinetic measurements of the immunoprecipitin reaction using the centrifugal analyser. Price CP Spencer K eds. Centrifugal analysers in clinical chemistry 1980:109-124 Praeger Publ. New York. .
  11. Bakker AJ. Immunoturbidimetry of urinary albumin: prevention of adsorption of albumin: influence of other urinary constituents. Clin Chem 1988;34:82-86.[Abstract/Free Full Text]




This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bakker, A. J.
Right arrow Articles by Zijlstra, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bakker, A. J.
Right arrow Articles by Zijlstra, A.
Related Collections
Right arrow Clinical Immunology
Right arrow Proteomics and Protein Markers
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS