To the Editor:
Measurement of urinary albumin is performed both for detection of microalbuminuria (1) and for establishing the selectivity of urinary protein excretion when renal function deteriorates (2); it thus requires assays that can measure albumin concentrations from a few milligrams per liter to several grams per liter. Falsely low concentrations are encountered regularly in immunoturbidimetric assays as a result of antigen excess (3). Procedures to detect antigen excess include additional testing with qualitative measurement of urinary albumin by reagent strip (dipstick) (4)(5)(6), quantitative measurement of urinary total protein (7), or albumin analysis with multiple dilutions of the sample (8). Another procedure uses an excessive amount of antibody, which can lengthen the calibration curve considerably (9). All of these procedures are costly. Mathematical procedures based on the kinetic differences between samples with and without antigen excess have been proposed (9)(10), but they falsely suggest antigen excess in samples with low antigen concentrations. An efficient procedure involves adding additional albumin after the measurement is completed and interpreting the additional change in turbidity (9). Here we report our experience with such a procedure on the MODULAR® ANALYTICS P-type analyzer (Roche GmbH).
For measurement of urinary albumin [modified for Modular from Ref. (11)], we incubated a sample with 45 g/L PEG-6000 in phosphate-buffered saline (reagent 1, consisting of 10 mmol/L phosphate, 140 mmol/L NaCl, 1 mL/L Triton X-100, pH 7.4) for 1.5 min. We then added an antibody against human serum albumin (reagent 2, consisting of Dako Q328 diluted 10-fold in saline) and monitored the development of turbidity at 18-s intervals for the next 3 min to establish the albumin concentration. We then added more albumin (reagent 3, consisting of an equivalent amount of 200 mg/L albumin prepared by dilution of a human control serum in saline) and monitored the change in turbidity for an additional 1.5 min; this served as the so-called "prozone check". When the absorbance of the analyzer-generated prozone is less than the lower cutoff limit (0.200), a flag is issued, which initiates an automatic rerun of the 10-fold diluted sample. The detailed settings for the MODULAR ANALYTICS P-type analyzer are given in Table 1 in the Data Supplement that accompanies the online version of this letter at http://www.clinchem.org/content/vol51/issue6/. The calibrators (2600 mg/L) consisted of purified human albumin (prod. no. A3782; Sigma) dissolved in phosphate-buffered saline for which the concentrations had been established as described previously (11).
The slope of the calibration curve (sensitivity) was lower at higher concentrations, and the absorbance decreased at concentrations above
1200 mg/L (Fig. 1A
). The additional absorbance change for the prozone check after addition of excess antigen, the so-called prozone value, increased slightly with increasing albumin concentrations to
250300 mg/L (Fig. 1A
). This increasing prozone value could be explained by the fact that the absorbance change produced by the endogenous albumin in the sample may not be complete at 3 min after addition of reagent 2 and therefore contributes to the prozone turbidity. At higher albumin concentrations, the prozone value decreased, and when the curve reached concentrations
9001200 mg/L, the prozone value was reduced to almost 0. Although this system worked with dilutions of a human serum, we also tested the procedure with urine samples from diabetic patients and from patients with renal failure. During the first 2 weeks of the investigation of the prozone measurement, the prozone value calculated by the Modular analyzer was recorded for all samples. Because the frequency of samples with microalbuminuric results was considerably lower than the frequency of samples with normoalbuminuric results, the prozone value was recorded for samples with microalbuminuric results for an additional 2 months and for samples with macroalbuminuric results for another 2 months. In total, 167 samples with normoalbuminuric results, 168 samples with microalbuminuric results, and 140 samples with macroalbuminuric results were assayed. The results for the prozone values, including those obtained in the rerun, are displayed in Fig. 1B
.
|
This procedure with the 0.200 cutoff value for the prozone check effectively detected all samples in which antigen excess caused falsely low microalbuminuric or normal values. The same kind of procedure could be useful for other assays in which antigen excess might cause falsely low values. In our experience, falsely low values attributable to antigen excess occur in daily practice in a limited number of samples, which makes an efficient procedure for antigen excess detection very important.
In conclusion, an effective and efficient procedure for the detection of antigen excess could be included in the assay for urinary albumin on the Roche Modular analyzer and could eliminate the need for additional tests to prevent reporting of falsely low results from antigen excess.