Proteomics and Protein Markers |
1 Department of Medical Sciences, Cardiology, and 2
Department of Medical Sciences, Clinical Chemistry, Uppsala University, Uppsala, Sweden.
3 Uppsala Clinical Research Centre, Uppsala, Sweden.
aAddress correspondence to this author at: Department of Clinical Chemistry and Pharmacology, University Hospital, SE-751 85 Uppsala, Sweden. Fax 46-186113703; e-mail per.venge{at}akademiska.se.
| Abstract |
|---|
|
|
|---|
Methods: Cardiac troponin concentrations were measured in a random sample of patients with unstable coronary artery disease included in the GUSTO IV trial (n = 696) by the AccuTnI (Beckman-Coulter Diagnostics), Architect cTnI (Abbott Diagnostics), Immulite 2500 cTnI (Diagnostics Products Corporation), and Elecsys 2010 cTnT (Roche Diagnostics) assays and related to the 1-year mortality. The primary cutoff concentrations were based on the 99th percentile upper reference limits and an imprecision (CV)
10%.
Results: The sensitivities of the AccuTnI and Architect cTnI assays in identifying patients who died within 1 year were equal and were significantly higher (P <0.05) than those of the Immulite 2500 cTnI and the Elecsys cTnT assays. The concordance between the AccuTnI and Architect cTnI assays was 97%, but concordances between the Architect cTnI and the Elecsys cTnT assays were 89%92% with more at-risk patients (P <0.01 to P <0.001) identified by the Architect cTnI assay.
Conclusions: The Architect cTnI assay has clinical performance similar to that of the AccuTnI, probably as a result of the inclusion of a monoclonal antibody against troponin I epitope 4149 in the assay.
| Introduction |
|---|
|
|
|---|
10% more patients with an adverse outcome than the AxSYM assay (Abbott, Abbott Park, IL) and the Liaison assay (Diasorin, formerly Byk-Sangtec). In addition, compared with the troponin T assay of Roche, the AccuTnI assay was superior. The choice of antibodies in the cTnI assays was different. The AccuTnI assay is based on a pair of monoclonal antibodies lying next to each other and directed against epitopes in the heart-specific and the stable region of the molecule close to the NH2 terminus (epitopes 2440 and 4149), whereas the 2 other assays incorporate one antibody directed toward the heart-specific region close to the NH2 terminus (i.e., in the region of epitopes 20-39) and the other antibody directed against epitopes in the stable part closer to the COOH terminus (epitopes 8791 and 80110, respectively). With the accumulating knowledge of the presence in blood of many different forms of troponins and interfering factors in the form of autoantibodies (12)(13)(14), we postulated that the assay configuration with respect to epitope specificity of the antibodies determines the clinical performance, possibly because different forms and molecular complexes of troponins are identified by the assays. It appears that epitope 4149 is critical. Recently, Abbott Diagnostics redesigned their troponin I assay and included in their Architect assay a monoclonal antibody directed toward this epitope. The aim of this study was therefore to determine whether, as a result of these modifications, the Architect assay has achieved clinical performance similar to that of the Beckman Coulter assay. For this purpose, using the Abbott Architect assay and the Beckman Coulter AccuTnI assay, we measured the troponin concentrations in samples from a random group of patients with unstable angina and nonST-elevation myocardial infarction taken from the Global Utilization of Strategies to Open Occluded Arteries IV (GUSTO IV) trial and related the concentrations to the 1-year mortality. For comparison, we also measured the troponin I concentrations with the Immulite 2500 assay from Diagnostics Products Corporation (DPC, Los Angeles, CA), which has a antibody configuration similar to that of the old AxSYM and Liaison assays, as well as the troponin T concentrations with the Elecsys 2010 assay from Roche (Mannheim, Germany). | Patients and Methods |
|---|
|
|
|---|
3 times the upper reference limit (URL) as reported previously in detail (15). At 12 months, all-cause mortality data were collected for all patients. From the GUSTO IV trial cohort, 700 samples were randomly selected for the analyses shown in this report. Four samples contained insufficient serum for the analyses. Blood anticoagulated with EDTA was obtained from a direct vein puncture at randomization, which occurred at a median of 8.5 h from symptom onset in the present subpopulation. Plasma was prepared within 30 min of collection by centrifugation at 2000g at room temperature for 20 min.
One reference population consisted of 442 apparently healthy persons participating in the Sweden, Women and Men and Ischemic Heart Disease (SWISCH) casecontrol study on risk factors for coronary artery disease in elderly men and women during 20012002 (16). The Architect assay was used to measure cTnI in EDTA-anticoagulated plasma from 424 of these persons (plasma samples from 18 participants were missing). Written consent was obtained from all participants, and the protocol was approved by the local ethics committee. The second reference population consisted of 747 apparently healthy persons: 435 men [mean (SD) age, 42 (9) years] and 312 women [44 (10) years]. These blood samples had been collected as part of a health screening program of employees at Pharmacia Diagnostics, Uppsala, Sweden. Before use in this study, the samples were deidentified. cTnI in the serum samples of this reference population was measured with the DPC assay. For the preparation of serum, blood was obtained in Vacutainer Tubes (Becton Dickinson) and allow to clot at room temperature for 6090 min, after which the serum was obtained by centrifugation for 10 min at 2000g and thereafter stored at 70 °C until used.
Troponin measurements were made on freshly frozen samples (70 °C) that had been thawed twice at most. As shown previously with the AccuTnI assay, troponin concentrations are stable in plasma after several freezethaw cycles (17). All assays were performed with reagents supplied by the respective companies and performed according to their instructions.
According to the manufacturers, the minimum detectable concentrations of cTnI were 0.01 µg/L for the AccuTnI,
0.01 µg/L for the Architect, and 0.10 µg/L for the Immulite 2500. Total imprecision (as CVs) was 4.1%8% (range, 0.0511 µg/L) for the AccuTnI, 3%5.8% (range, 0.113.7 µg/L) for the Architect cTnI, and 6.2%21.7% (range, 0.1240.9 µg/L) for the Immulite 2500 cTnI assays. The minimum cTnT concentration detectable by the Elecsys assay, according to the manufacturer, was 0.01 µg/L, and total imprecision was 4%9% (range, 0.430 µg/L).
Mortality at 12 months was used as the clinical endpoint in the clinical performance evaluations of the assays.
statistics
The Fisher exact test was used to compare the clinical performances of the assays. Differences in sensitivities were calculated by differences between 2 proportions. Calculations were performed by the statistical software Statistica for Windows (Ver. 7.0; Statistica) and Statistical Package for Social Sciences (SPSS 12.0) software (SPSS Inc.).
| Results |
|---|
|
|
|---|
10% was 0.03 µg/L; for the Architect cTnI assay, it was 0.032 µg/L; and for the cTnT assay, it was 0.03 µg/L. With respect to the Immulite 2500 assay, we defined the imprecision profile of the assay from the CV obtained with duplicate measurements of samples from a cohort of patients from the GUSTO IV trial (n = 812). The lowest concentration at which the CV was
10% was 0.33 µg/L. The 99th percentile URLs of seemingly healthy reference populations used in this report were partly adopted from the manufacturer and partly from our own results. For the AccuTnI and Architect cTnI assays, the 99th percentile URLs for troponin I were established by means of samples obtained from the SWISCH cohort (n = 424) and were 0.04 and 0.038 µg/L, respectively. For the cTnI assay on Immulite 2500 and for the cTnT assay, the 99th percentile URLs were established in another cohort of seemingly healthy persons (n = 747) and were 0.22 and <0.01 µg/L, respectively. Two different cutoff concentrations were used for the AccuTnI and Architect cTnI assays and were based on the fact that among the 424 persons, 3 had concentrations measured with the Architect assay that were regarded as obvious outliers. The cutoff concentrations based on the 421 persons were 0.027 and 0.032 µg/L, respectively.
The correlation between the Architect cTnI and AccuTnI assays was very good (r = 0.97; y = 1.007x + 0.161 µg/L; Sy|x = 1.90 µg/L; n = 802). The overall agreement between the 2 assays is illustrated in Fig. 1
by means of a BlandAltman plot and shows an overall difference between the methods of 5.8% with the AccuTnI giving the higher concentrations.
|
The sensitivities for all 4 assays in identifying persons who died within the 1-year follow-up period are given in Fig. 2
. As shown, the sensitivities of the Architect cTnI and AccuTnI assays were identical at similar cutoff concentrations, whereas the sensitivities of the cTnT and the Immulite 2500 cTnI assays were lower than the sensitivities of the Architect cTnI and AccuTnI assays (P = 0.05) when the 10% CV cutoffs were used. In addition, the sensitivities based on the 99th percentile cutoffs of the cTnT assay and the Immulite 2500 cTnI assay were lower than the sensitivity of the AccuTnI assay when the 99th percentile URL for persons <60 years of age was adopted (P <0.05).
|
We calculated the concordances between the assays for the Architect cTnI, AccuTnI, and Elecsys cTnT assays, based on either the 10% CV cutoff concentrations or 99th percentile URL for the reference populations. When we used the 10% CV or the 99th percentile URL cutoffs, we found an overall concordance of 97%. As shown in Table 1
, with results based on the truncated 99th percentiles, the clinical performances of the AccuTnI and Architect cTnI assays were very similar, with 1.8% and 1.3% of the study population identified only by either the Architect cTnI assay or the AccuTnI assay, of which 1 death had occurred among those individuals identified by the Architect cTnI assay only. The results based on the untruncated 99th percentile URL gave similar results with a 96% concordance. For comparison, we also estimated the concordance based on the lower cutoff limits based on the 99th percentile URL for the persons <60 years of age in the SWISCH cohort as analyzed by AccuTnI (0.02 µg/L) and the 99th percentile URL as reported by the manufacturer of the Architect cTnI assay (0.017 µg/L). The concordance based on these lower cutoff limits was also 97% with similar clinical performances. Also shown in Table 1
is a comparison between the Architect cTnI and the Elecsys cTnT assays; the overall concordance was 92% when the 99th percentile URLs were used as cutpoints. Significantly more individuals were identified by the Architect cTnI assay only than by the Elecsys cTnT assay only (P <0.01, Fisher exact test), with 5 deaths among those identified by the cTnI assay alone but no deaths among those identified by the cTnT assay alone. Thus, 8% of all deaths occurred in the cohort identified only by the Architect cTnI assay. When we used the 10% CV cutoff concentrations, the differences in clinical performances between the cTnI and cTnT assays were even more striking (P <0.001) with an overall concordance of 89%. Comparison of the AccuTnI and the Elecsys cTnT assays gave results identical to those obtained in the Architect cTnI and Elecsys cTnT comparison and were almost identical to previously published results based on the Fast Revascularisation during InStability in CAD trial (FRISC II) cohort (10)(11).
|
| Discussion |
|---|
|
|
|---|
The reason for the improvement in clinical performance after inclusion of antibodies against epitope 4149 in cTnI assays is not entirely clear. From an analytical point of view, the Liaison and Immulite 2500 cTnI assays are very sensitive, well-performing assays and are in that respect comparable to the AccuTnI and Architect cTnI assays; however, their clinical performance is not as high (11). Thus, the question is: What do these different assays measure? Indeed, they seem to identify different molecules at the lower end of the concentration range; we found no correlations between the 2 assay principles, as illustrated by the comparison between the AccuTnI and Liaison cTnI, when results from the reference populations were compared. It is also well documented that cTnI is present in many different forms in the circulation, some of which may be the consequence of truncation of the molecule as a result of proteolytic breakdown and some of which may be the result of complex formations between cardiac troponins I, T, and C in various molecular constellations (12)(18). An interesting observation is the presence of circulating autoantibodies against cTnI in a substantial number of study participants and the possible interference of these antibodies with the assays (13)(14) because the autoantibodies seem to be directed preferentially to the more C-terminal regions of the molecule and not to the N-terminal region including epitope 4149. As shown previously, the concordance between the old AxSYM assay and the Elecsys cTnT assay was very good,
95%, and with very similar clinical performances for the 2 assays (10). This shows that the 2 assays probably identify the same patients but miss some patients identified by the AccuTnI and Architect cTnI assays. This could indicate that the old AxSYM assay and the Elecsys cTnT assay identify the same molecular complexes of troponins I and T and also suggests that additional unique molecular forms of cTnI are identified by the AccuTnI and Architect cTnI assays.
The study cohort in this report was the GUSTO IV trial patients, who had been recruited as part of a clinical trial of the antithrombotic drug abciximab (ReoProTM) and were included if they had symptoms and signs of acute coronary syndrome without electrocardiographic indications of ST-segment elevations (19). In our previous report we showed that troponin T concentrations >0.01 µg/L strongly predicted an acute myocardial infarction within the next 30 days (6), whereas troponin T as a risk predictor of death within 1 year was weaker but significant. The present data obtained with improved cTnI assays indicate that the predictive power of cardiac troponins probably was underestimated by the troponin T results. From a clinical point of view, the differences in performance of the different assays should be of considerable interest because the mortality was also high in the cohorts identified with increased concentrations only by the AccuTnI or the Architect cTnI assays. Thus, 1-year mortality was similarly high in patients with Architect TnI concentrations exceeding the 99th percentile cutoff of 0.027 µg/L irrespective of undetectable or detectable troponin T concentrations (11% vs 10%; Table 1
). The clinical relevance of identifying patients with very low but increased concentrations of cardiac troponins is also illustrated by the fact that early intervention may reduce the mortality by
30% in those with increased cTnI concentrations as measured with the AccuTnI assay (10). In our previous study on the FRISC II patient cohort, the average time from symptoms to blood sampling was
40 h (5) and was substantially longer than the 8.5 h in the GUSTO IV trial. In spite of these time differences, we obtained almost identical predictive results in the 2 studies of cardiac troponin measurements, which might indicate that the unique molecular forms of cTnI identified by the assays remained in the circulation for a substantial time period after the acute event.
One limitation of the present study is the exact definition of cutoff concentrations for the fair comparison of different assays. Two criteria for cutoffs were used in this study, and regardless of which principle was adopted, similar results were obtained. The cutoff concentrations at which different assays have a CV
10% are based on different criteria: in some reports, they are based on results from 1 center, whereas in others they are based on mean concentrations obtained in multicenter studies (17)(20)(21). Thus, comparisons of assays based on concentrations from such different approaches are obviously quite uncertain. In our study, we therefore preferred to use cutoffs based on the 99th percentile URLs of well-defined reference populations. In this regard, the SWISCH cohort is unique because it was collected randomly in a community but matched with respect to sex and age to the FRISC II cohort of patients. All participants were invited to have a thorough medical examination by a cardiologist, and only those found healthy at this examination were included in the reference cohort. The inclusion criteria for the GUSTO IV trial were similar to those for FRISC II, which means that the SWISCH cohort is well matched to the GUSTO IV cohort as well. Compared with other studies, the 99th percentile of the SWISCH cohort was somewhat higher, which likely is attributable to differences resulting from inclusion of older persons because, as we show in the present study with the Architect cTnI and also in a previous report with the AccuTnI assay, cTnI concentrations tend to increase with age (11). The uncertainty in defining true 99th percentile URLs based on 424 individuals is obvious because the cutoff concentration is equal to the individual having the fifth highest concentration. If 1 or 2 more seemingly healthy individuals with somewhat increased concentrations are included in the reference population, this may have profound consequences on the 99th percentile limit, which is one major reason to eliminate possible outliers from the cohort. We based our evaluations on several cutoff limits ranging from 0.017 µg/L to 0.038 µg/L for the Architect cTnI assay and from 0.02 to 0.04 µg/L for the AccuTnI assay, and the clinical performance results were very similar irrespective of cutoff limit. It should, however, be emphasized that the aim of the present study was not to define the optimal clinical cutoffs for the different assays, but to illustrate the fact that changes in antibody configurations have impacts on the capacity of the assay to identify patients at risk.
In conclusion, we show that the clinical performance of troponin assays varies. This is true between assays that measure troponin I or T, but also among various cTnI assays. For the measurement of cTnI, inclusion in the assays of at least 2 monoclonal antibodies against epitopes in the N-terminal part of the heart-specific region of the molecule seemed critical to obtaining optimum performance. For as yet unidentified reasons, such cTnI assays also identify more patients at risk than the currently available cTnT assay.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
T. Tsutamoto, C. Kawahara, M. Yamaji, K. Nishiyama, M. Fujii, T. Yamamoto, and M. Horie Relationship between renal function and serum cardiac troponin T in patients with chronic heart failure Eur J Heart Fail, July 1, 2009; 11(7): 653 - 658. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Venge, S. James, L. Jansson, and B. Lindahl Clinical Performance of Two Highly Sensitive Cardiac Troponin I Assays Clin. Chem., January 1, 2009; 55(1): 109 - 116. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.M. Cobbaert, C.W. Weykamp, E.C.H.J. Michielsen, H. Baadenhuijsen, and M.P. van Dieijen-Visser Time-Dependent Instability of Cardiac Troponins in Human Plasma Spiked with NIST Reference Material 2921 Clin. Chem., December 1, 2008; 54(12): 2078 - 2079. [Full Text] [PDF] |
||||
![]() |
F. S. Apple, M. M. Murakami, R. Ler, D. Walker, M. York, and for the HESI Technical Committee of Biomarkers Wor Analytical Characteristics of Commercial Cardiac Troponin I and T Immunoassays in Serum from Rats, Dogs, and Monkeys with Induced Acute Myocardial Injury Clin. Chem., December 1, 2008; 54(12): 1982 - 1989. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Ilva, J. Lassus, K. Siirila-Waris, J. Melin, K. Peuhkurinen, K. Pulkki, M. S. Nieminen, H. Mustonen, P. Porela, and V.-P. Harjola Clinical significance of cardiac troponins I and T in acute heart failure Eur J Heart Fail, August 1, 2008; 10(8): 772 - 779. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R Tate, W. Ferguson, R. Bais, K. Kostner, T. Marwick, and A. Carter The determination of the 99th centile level for troponin assays in an Australian reference population Ann Clin Biochem, May 1, 2008; 45(3): 275 - 288. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E.F. Melanson, M. J. Tanasijevic, and P. Jarolim Cardiac Troponin Assays: A View From the Clinical Chemistry Laboratory Circulation, October 30, 2007; 116(18): e501 - e504. [Full Text] [PDF] |
||||
![]() |
P. A. Kavsak, A. M. Newman, V. Lustig, A. R. MacRae, G. E. Palomaki, D. T. Ko, J. V. Tu, and A. S. Jaffe Long-Term Health Outcomes Associated with Detectable Troponin I Concentrations Clin. Chem., February 1, 2007; 53(2): 220 - 227. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||