Clinical Chemistry 52: 1568-1574, 2006. First published June 8, 2006; 10.1373/clinchem.2006.069039
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow 069039.Supplemental Data
Right arrow All Versions of this Article:
clinchem.2006.069039v1
52/8/1568    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (8)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kort, S. A.R.
Right arrow Articles by Blankenstein, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kort, S. A.R.
Right arrow Articles by Blankenstein, M. A.
Related Collections
Right arrow Cancer Diagnostics (since 2002)
Right arrow Proteomics and Protein Markers
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2006;52:1568-1574.)
© 2006 American Association for Clinical Chemistry, Inc.


Automation and Analytical Techniques

Comparison of 6 Automated Assays for Total and Free Prostate-Specific Antigen with Special Reference to Their Reactivity toward the WHO 96/670 Reference Preparation

Sheila A.R. Kort1, Frans Martens1, Hilde Vanpoucke2, Hans L. van Duijnhoven3 and Marinus A. Blankenstein1,a

1 Endocrine Laboratory, Department of Clinical Chemistry, VU University Medical Center, Amsterdam, The Netherlands.
2 Laboratory of Clinical Chemistry, H. Hartziekenhuis, Roeselare-Menen vzw, Belgium.
3 Algemeen Klinisch Laboratorium, Elkerliek Ziekenhuis, Helmond, The Netherlands.

aAddress correspondence to this author at: Department of Clinical Chemistry, VU University Medical Center, De Boelelaan 1117, 1081 HV Amsterdam, The Netherlands. Fax 31-20-4443895; e-mail MA.Blankenstein{at}vumc.nl.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Background: Prostate-specific antigen (PSA) assays have historically produced different results. Our aim was to investigate the comparability of assay results of selected commercially available assay methods designed to measure total, free, or complexed PSA (tPSA, fPSA, and cPSA).

Methods: We measured tPSA, fPSA, and cPSA in 70 samples and in the WHO PSA 96/670 reference preparation with 6 assays (Beckman-Coulter Access, Abbott ARCHITECT and AxSYM, Bayer Advia Centaur, DPC IMMULITE 2000, and Roche Modular Analytics E170). We also calculated the fPSA/tPSA ratio.

Results: The mean deviations from the expected tPSA and fPSA values for the WHO 96/670 reference preparation were 0.37 (range, 0.01–1.32) and 0.19 (range, 0.05–0.49) µg/L, respectively. When plotted against the expected WHO 96/670 reference preparation value, regression slopes varied from 0.99 to 1.22 and r2 from 0.9996 to 1.000. When total PSA was measured in mixtures of sera with high and low tPSA concentrations, the mean (SD) slope of regression of different assays against an in-house method was 1.04 (0.09). In these specimens, the fPSA/tPSA ratio was 0.11–0.14 with different methods. The tPSA and fPSA values in patient samples measured in different assays and plotted against ARCHITECT gave regression slopes from 0.88 to 0.97. The results of the studied assays for tPSA in serum samples agreed within 15%, from each other, and all results for the WHO 96/670 reference preparation were within 6.8% (confidence interval, 1.7%–15.2%) of the expected value. The results for fPSA were more diverse.

Conclusions: Differences among PSA assays appear to have decreased since introduction of the WHO 96/670 reference preparation, but further efforts are needed to harmonize fPSA assays.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Prostate cancer (PCa) 1 remains an important public health problem (1)(2). Prostate-specific antigen (PSA) testing is used for both early detection and monitoring of PCa (3).

Assays for total PSA (tPSA) are based on the immunochemical detection of 2 major PSA forms: free, or uncomplexed, PSA (fPSA) and complexed, the dominant form of PSA, bound to {alpha}1-antichymotrypsin (PSA-ACT). A third form, PSA bound to {alpha}2-macroglobulin, is not detected by currently available assays (4). Other PSA complexes are considered to be of less significance because of their very low concentrations in blood (5).

Because benign prostate hyperplasia and PCa can both increase tPSA, the ratio fPSA/t PSA is often used (6). Patients with higher fPSA/tPSA ratios are less likely to have PCa than patients with lower ratios (7)(8). Detection of fPSA is based on antibodies specific for at least one PSA epitope masked in the complexed form. Detection of the complexed PSA form is based on at least one epitope specific for the PSA-ACT complex.

The equimolarity, or lack thereof, of tPSA assays has been a matter of debate (9)(10). An assay for tPSA is called equimolar when it generates the same signal for the same concentrations of fPSA and PSA-ACT complex and thus is not affected by changes in proportions of fPSA and PSA-ACT complex. A lack of equimolarity hampers interchangeability of results of different assays (11). The use of 2 monoclonal antibodies free of steric hindrance and the use of optimal incubation times are ingredients for an assay with equimolar detection.

Traditionally, different assays for PSA have produced different results on the same patient sample. The reasons for this include the differences in epitope recognition by various assays and the use of different reference preparations. To circumvent the latter problem, an international reference preparation has been devised in which 90% of the PSA is complexed to ACT and the remaining 10% is fPSA (12)(13). This proportion is similar to that in the circulation in patients with PCa. Manufacturers of several PSA assays have used this WHO 96/670 reference preparation for calibration, but intermethod differences persist (14). The purpose of the present study was to assess the responsiveness to the WHO preparation of various assays for fPSA and tPSA and to compare the performance of these automated assays on clinical specimens that contain various PSA concentrations.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
We performed experiments with the WHO 96/670 reference preparation (First International Standard), with patient serum samples, and with serum pools.

The PSA 90:10 WHO 96/670 reference preparation, which contains 500 µg/L tPSA after reconstitution in an fPSA/tPSA ratio of 0.10, was a gift of Abbott Laboratories (Abbott Park, IL). Reconstitution was done at 1 location, and for all solutions we used the recommended 2 mL of 10 g/L bovine serum albumin in 20 mmol/L phosphate-buffered saline (pH 7.4) to create a solution containing 500 µg/L PSA. This stock solution was diluted with 18 mL of buffer to create a 50 µg/L solution. We diluted the 50 µg/L stock solution with buffer to generate 6 concentrations: 20, 10, 4, 2.5, 0.5, and 0.2 µg/L. Each solution, and a portion of buffer, was stored in single-use volumes and frozen at –20 °C.

We obtained serum samples by selecting 70 anonymized surplus samples <6 months old with PSA concentrations (range) of <0.1 to 50 µg/L as measured by one method (ARCHITECT Total PSA; Abbott). From these samples, multiple 500-µL aliquots were made and stored at –20 °C. Finally, we prepared independent serum pools with expected values, as measured by our in-house method (ARCHITECT) for tPSA, by mixing sera with high and low tPSA concentrations in various proportions. Pools and aliquots were analyzed within 3 weeks.

test procedure
Assays were performed according to the instructions of their manufacturers. Only 1 reagent lot and 1 calibration were used. The serum aliquots were kept frozen and thawed only once and analyzed within 4 h of thawing. The methods used were the Access (Beckman-Coulter), ARCHITECT (Abbott), AxSYM (Abbott), Advia Centaur (Bayer), IMMULITE 2000 (DPC), and Modular Analytics E170 (Roche). Our in-house method, ARCHITECT, was arbitrarily used to provide a concentration value for comparison of methods. Of the 6 selected methods, 5 measured tPSA and fPSA and 1 measured tPSA and cPSA. For the latter assay, fPSA was calculated as the difference between tPSA and cPSA. Assay characteristics are summarized in Table 1 , and the performance characteristics of the assays as run in the participating laboratories are shown in Table 2 .


View this table:
[in this window]
[in a new window]
 
Table 1. Characteristics of tPSA, fPSA, and cPSA assays.


View this table:
[in this window]
[in a new window]
 
Table 2. Imprecision data of different PSA assays for tPSA and fPSA.

statistical evaluation
We analyzed results of measurements of the WHO 96/670 reference preparation in buffer by linear regression and Pearson correlation and used Passing and Bablok regression analysis for the serum aliquots to determine slope and intercept. P values ≤0.05 were considered to reflect statistical significance.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
measured TPSA and FPSA in the who 96/670 reference preparation
Comparisons of results obtained with the WHO 96/670 reference preparation of each of the several other methods and the arbitrarily selected comparison method (Architect) are shown in Fig. 1 and Table 3 .


Figure 1
View larger version (17K):
[in this window]
[in a new window]
 
Figure 1. Comparison of tPSA (A) and fPSA (B) in the WHO 96/670 reference preparation, measured by 6 different assays.

tPSA and fPSA were measured in buffered solutions of the reference preparation by 6 (tPSA) or 5 (fPSA) methods, and the results were plotted against the expected values. Slopes, intercepts, and correlation coefficients of the different lines are presented in Table 3Up . Acc, Access; Arch, ARCHITECT; Axs, AxSYM; Cen, Centaur; Immu, IMMULITE.


View this table:
[in this window]
[in a new window]
 
Table 3. Slopes, intercepts, and correlation coefficients for comparison of different assays for tPSA and fPSA in the WHO 96/670 reference preparation.1

measured TPSA and FPSA in serum pools
The values for serum pools measured by various methods plotted against the arbitrarily assigned PSA values by the in-house method are shown in Table 4 . Slopes of the regression lines varied between 0.97 and 1.19, and the intercepts ranged from 0.23 to 0.12.


View this table:
[in this window]
[in a new window]
 
Table 4. Characteristics of the regression lines representing the relationships between tPSA and fPSA results obtained with different assays in serum pools with arbitrarily assigned PSA values.1

patient serum samples
The relationships between the assays for tPSA and fPSA in patient serum samples are shown in Fig. 2 . Seventy patient specimens were tested, those with tPSA <24 µg/L and fPSA <20 µg/L are shown in Fig. 2 . When the results were plotted against those obtained with the ARCHITECT assay, the slopes varied between 0.88 and 0.97, whereas all r2 values were >0.983. Results (Bland–Altman plots) obtained by various assays showed substantial differences; i.e., 0.5 to 1.0 µg/L for tPSA and 0.12 to 0.40 µg/L for fPSA. The data for tPSA are illustrated in Fig. S1 and the data for fPSA in Fig. S2 of the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol52/issue8/ .


Figure 2
View larger version (11K):
[in this window]
[in a new window]
 
Figure 2. tPSA (A) and fPSA (B) in patient sera measured by 6 different assays.

Seventy anonymized surplus samples with tPSA concentrations of <0.1 to 50 µg/L as measured by one method were analyzed for tPSA (5 methods) and fPSA (4 methods). For clarity, the tPSA scale was limited to 24 µg/L. Results >24 µg/L did not deviate from the calculated line. Cen, Centaur; Imm(u), IMMULITE.

Passing and Bablok regression characteristics for fPSA/tPSA ratios in patient serum samples, comparing the ARCHITECT with the other 5 assays, are shown in Table 5 . The values obtained with the original calibrator were recalculated against the WHO 96/670 reference preparation.


View this table:
[in this window]
[in a new window]
 
Table 5. Characteristics for fPSA/tPSA ratio in patient serum samples between the ARCHITECT and other assays: Passing and Bablok regression.

Passing and Bablok regression of the recalculated values against those obtained with the original calibration showed >20% deviation from the ideal slope for the Access assay, whereas for all other methods the deviation was <7%.

FPSA/TPSA ratio in serum and buffer
The values for the fPSA/tPSA ratio measured in the WHO 96/670 reference preparation and the serum pools with arbitrarily assigned values are presented in Table 6 .


View this table:
[in this window]
[in a new window]
 
Table 6. fPSA/tPSA ratios for different assays measured in WHO 96/670 reference preparation buffered solution and in independent patient serum pools with assigned expected tPSA values.


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
This study demonstrates that overall the results of the tested assays agreed well and that all gave PSA results close to the WHO 96/670 reference preparation value. Lot-to-lot differences were not accounted for, however, as all measurements were made with a single lot of each reagent.

Results for tPSA appears to be less diverse than those for fPSA (Figs. 1Up and 2Up ). This obviously has consequences for the intermethod differences in the fPSA/tPSA ratio.

The differences between assays were larger when the WHO reference preparation was analyzed than when the assays were used for clinical samples. This may indicate that the assays are fine-tuned to yield comparable results on actual specimens. In addition it demonstrates that standardizing against an international standard does not necessarily mean that the results will agree for specimens. This is illustrated by the 2 assays from the same manufacturer, which not only are calibrated against the WHO preparation but also use the same monoclonal antibodies. They differ with respect to detection system (chemiluminescence for the ARCHITECT and fluorescence for the AxSYM), microparticle sizes, and incubation times. The differences between Figs. 1Up and 2Up may also be related, in part, to matrix differences (buffer vs serum) (15). Obviously, correct measurement of patient samples is preferred over correct measurement of a standardized buffer solution.

As shown in Table 6Up , in the tPSA range from 2.5 to 10 µg/L in the WHO 96/670 reference preparation buffered solution, the fPSA/tPSA ratios for the AxSYM, E170, and IMMULITE 2000 remained within 10% of the expected ratio for the 90:10 WHO reference preparation. Agreement between results for the fPSA/tPSA ratio appears to be dependent on the matrix, however. This confirms that results obtained for fPSA or tPSA from different assays or manufacturers are still not interchangeable.

Because the detection of tPSA in serum is less prone to differences among assays, the effects on the fPSA/tPSA ratio are caused mainly by variation in fPSA. This once again emphasizes the principle and need for assay equimolarity (11)(14)(16). It also demonstrates that the effect of specimen matrix should be taken into account when developing assays.

The Advia Centaur fPSA/tPSA ratio appeared to be higher in buffer than in serum. This may be attributable to either a higher (calculated) fPSA or a lower tPSA. Because the results in Fig. 1AUp and Fig. 2AUp show that Advia Centaur tPSA values are comparable to those of other assays, it appears that not all cPSA is detected.

Higher fPSA/tPSA ratios were obtained with the Access. According to the Beckman Access package insert, the Access calibrators for tPSA and fPSA are composed of human PSA with no reference to the WHO 96/670 reference preparation. Further information from Beckman stated that the original Hybritech Tandem R calibration was based on an internal reference preparation of purified human PSA, which yielded 16% higher PSA results than the WHO 96/670 calibration. This may also explain the difference from an ideal slope in the Passing and Bablok regression when the original values were plotted against the results obtained by recalculation to the WHO reference preparation.

The fPSA/tPSA results obtained in serum with the IMMULITE 2000 were lower than those obtained with other methods. In serum this method gave the highest tPSA results (Fig. 2AUp ) and the lowest fPSA results (Fig. 2BUp ). Because this is in contrast to the results obtained with the WHO reference preparation in buffer, this method also appears to show a matrix effect. Assay design may also play a role because the IMMULITE total PSA has a polyclonal-monoclonal design and the free PSA is a monoclonal-polyclonal assay.

Both the ARCHITECT and AxSYM yielded higher values for 2 samples. After we diluted the samples, the differences remained, which led us to conclude that the Abbott assays probably bind to a different site on PSA, which may not be recognized equally well by other antibodies. Alternatively, these specimens may contain aberrant PSA forms (17).

In conclusion, the differences among PSA assays appear to have decreased since introduction of the WHO 96/670 reference preparation, but the results obtained with different assays are not yet interchangeable (18)(19). Taking into account the clinical use of the fPSA/tPSA ratio, however, it is imperative that further efforts are undertaken to optimize assays to yield more comparable results for fPSA. Future steps toward PSA harmonization may include, for example, development of a reference method. Until then, monitoring of patients should be done with one method only. When a new assay must be used during monitoring, the results of the 2 assays may need to be compared on the same sample from each patient.


   Acknowledgments
 
We thank Abbott Diagnostics for providing us with the WHO PSA 96/670 reference preparation and reagents with no further obligations.


   Footnotes
 
1 Nonstandard abbreviations: PCa, prostate cancer; tPSA, fPSA, and cPSA, total, free (uncomplexed), and complexed prostate-specific antigen, respectively; and ACT, {alpha}1-antichymotrypsin.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Auvinen A, Rietbergen JB, Denis LJ, Schröder FH, Prorok PC. Prospective evaluation plan for randomised trials of prostate cancer screening. The International Prostate Cancer Screening Trial Evaluation Group. J Med Screen 1996;3:97-104.[Medline] [Order article via Infotrieve]
  2. Andriole GL, Levin DL, Crawford ED, Gelmann EP, Pinsky PF, Chia D, et al. Prostate Cancer Screening in the Prostate, Lung, Colorectal and Ovarian (PLCO) Cancer Screening Trial: findings from the initial screening round of a randomized trial. J Natl Cancer Inst 2005;97:433-438.[Abstract/Free Full Text]
  3. Armbruster DA. Prostate-specific antigen: biochemistry, analytical methods, and clinical application. Clin Chem 1993;39:181-195.[Abstract]
  4. Patel D, White PA, Milford Ward A. A comparison of six commercial assays for total and free prostate specific antigen (PSA): the predictive value of the ratio of free to total PSA. BJU Int 2000;85:686-689.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  5. Ward AM, Catto JW, Hamdy FC. Prostate specific antigen: biology, biochemistry and available commercial assays. Ann Clin Biochem 2001;38:633-651.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  6. Gion M, Mione R, Barioli P, Barichello M, Zattoni F, Prayer-Galetti T, et al. Percent free prostate-specific antigen in assessing the probability of prostate cancer under optimal analytical conditions. Clin Chem 1998;44:2462-2470.[Abstract/Free Full Text]
  7. Gann PH, Ma J, Catalona WJ, Stampfer MJ. Strategies combining total and percent free prostate specific antigen for detecting prostate cancer: a prospective evaluation. J Urol 2002;167:2427-2434.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  8. Tanguay S, Begin LR, Elhilali MM, Behlouli H, Karakiewicz PI, Aprikian AG. Comparative evaluation of total PSA, free/total PSA, and complexed PSA in prostate cancer detection. Urology 2002;59:261-265.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  9. Tewari PC, Williams JS. Analytical characteristics of seminal fluid PSA differ from those of serum PSA. Clin Chem 1998;44:191-193.[Free Full Text]
  10. Blase AB, Sokoloff RL, Smith KM. Five PSA methods compared by assaying samples with defined PSA ratios. Clin Chem 1997;43:843-845.[Free Full Text]
  11. Stephan C, Klaas M, Muller C, Schnorr D, Loening SA, Jung K. Interchangeability of measurements of total and free prostate-specific antigen in serum with 5 frequently used assay combinations: an update. Clin Chem 2006;52:59-64.[Abstract/Free Full Text]
  12. Rafferty B, Rigsby P, Rose M, Stamey T, Gaines Das R. Reference reagents for prostate-specific antigen (PSA): establishment of the first international standards for free PSA and PSA (90:10). Clin Chem 2000;46:1310-1317.[Abstract/Free Full Text]
  13. Stamey TA, Chen Z, Prestigiacomo AF. Reference material for PSA: the IFCC standardization study. International Federation of Clinical Chemistry. Clin Biochem 1998;31:475-481.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  14. Roddam AW, Rimmer J, Nickerson C, Ward AM. Prostate-specific antigen: bias and molarity of commercial assays for PSA in use in England. Ann Clin Biochem 2006;43:35-48.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  15. Ishibashi M. Standardization of prostate-specific antigen (PSA) assays: can interchangeability of PSA measurements be improved?. Clin Chem 2006;52:1-2.[Free Full Text]
  16. Semjonow A, De Angelis G, Oberpenning F, Schmid HP, Brandt B, Hertle L. The clinical impact of different assays for prostate specific antigen. BJU Int 2000;86:590-597.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  17. Van Duijnhoven HL, Péquériaux NC, Van Zon JP, Blankenstein MA. Large discrepancy between prostate-specific antigen results from different assays during longitudinal follow-up of a prostate cancer patient. Clin Chem 1996;42:637-641.[Abstract/Free Full Text]
  18. Blijenberg BG, Yurdakul G, Van Zelst BD, Bangma CH, Wildhagen MF, Schroder FH. Discordant performance of assays for free and total prostate-specific antigen in relation to the early detection of prostate cancer. BJU Int 2001;88:545-550.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  19. Chan DW, Sokoll LJ. WHO first international standards for prostate-specific antigen: the beginning of the end for assay discrepancies?. Clin Chem 2000;46:1291-1292.[Free Full Text]



The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
C. M. Sturgeon, M. J. Duffy, U.-H. Stenman, H. Lilja, N. Brunner, D. W. Chan, R. Babaian, R. C. Bast Jr., B. Dowell, F. J. Esteva, et al.
National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines for Use of Tumor Markers in Testicular, Prostate, Colorectal, Breast, and Ovarian Cancers
Clin. Chem., December 1, 2008; 54(12): e11 - e79.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
F. H. Jansen, M. Roobol, C. H. Bangma, and R. H. N. van Schaik
Clinical Impact of New Prostate-Specific Antigen WHO Standardization on Biopsy Rates and Cancer Detection
Clin. Chem., December 1, 2008; 54(12): 1999 - 2006.
[Abstract] [Full Text] [PDF]


Home page
Ann Clin BiochemHome page
D. B. P. Eekers, A. Laschet, M. de Groot, E. Roelofs, A. Kester, K. Delaere, P. Lambin, F. van Gils, M. Nap, and J. ten Kate
Why determine only the total prostate-specific antigen, if the free-to-total ratio contains the information?
Ann Clin Biochem, May 1, 2008; 45(3): 270 - 274.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow 069039.Supplemental Data
Right arrow All Versions of this Article:
clinchem.2006.069039v1
52/8/1568    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (8)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kort, S. A.R.
Right arrow Articles by Blankenstein, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kort, S. A.R.
Right arrow Articles by Blankenstein, M. A.
Related Collections
Right arrow Cancer Diagnostics (since 2002)
Right arrow Proteomics and Protein Markers
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS