Clinical Chemistry 52: 1609-1610, 2006; 10.1373/clinchem.2006.071357
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Develter, M.
Right arrow Articles by Bossuyt, X.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Develter, M.
Right arrow Articles by Bossuyt, X.
Related Collections
Right arrow General Clinical Chemistry
Right arrow Clinical Immunology
(Clinical Chemistry. 2006;52:1609-1610.)
© 2006 American Association for Clinical Chemistry, Inc.


Letters to the Editor

Can Heparin Plasma Be Used instead of Serum for Nephelometric Analysis of Serum Proteins?

Mieke Develter1, Norbert Blanckaert1, Arnost Komárek2 and Xavier Bossuyt1,a

1 Laboratory Medicine, Immunology, University Hospital Leuven, Leuven, Belgium
2 Department of Biostatistics, University Leuven, Leuven, Belgium

aAddress correspondence to this author at: Department of Laboratory Medicine, Immunology, University Hospital Leuven, Herestraat 49, B-3000 Leuven. Fax 32-13-347042; e-mail xavier.bossuyt{at}uz.kuleuven.ac.be.


To the Editor:

Nephelometry is widely used for quantification of serum proteins. The main manufacturers of nephelometers, Beckman-Coulter and Dade-Behring, recommend serum as a sample for most tests. For a limited number of tests, e.g., lipoprotein(a), they state that serum and heparin plasma can be used. Serum should be separated from contact with cells as soon as possible (1), but blood should be allowed to sit long enough for complete clot formation because inadequate clot formation with residual fibrin strands can clog the sampling pipette. Use of plasma circumvents these clotting-related problems. We therefore investigated whether lithium-heparin plasma can be used for nephelometric analysis of a whole array of serum proteins.

Serum and plasma samples were collected simultaneously. The serum-clot contact time was 4 h, whereas plasma was separated from cells within 2 h (1). Analysis was performed with an Immage instrument (Beckman-Coulter) for the following 22 analytes: albumin, prealbumin (PAB), {alpha}1-acid glycoprotein, {alpha}1-antitrypsin, {alpha}2-macroglobulin (AMG), ß2-microglobulin, factor B, C1 inhibitor, C1q, C3, C4, C5, ceruloplasmin (CERU), IgA, IgG, IgG1, IgG2, IgG3, IgG4, IgM, and {kappa} and {lambda} light chains. All reagents were from Beckman-Coulter except for the reagents for IgG subclasses, which were from Sanguin, and the reagents for C1q, C5, C1 inhibitor, and ß2-microglobulin, which were supplied by DakoCytomation. Samples were analyzed daily for albumin, PAB, {alpha}1-antitrypsin, C3, C4, C5, CERU, IgA, IgG, and IgM. The other assays were performed weekly with preanalytical storage of samples at 4 °C for a maximum of 6 days. The samples included normal, low, and/or increased values.

For the sample comparison, Pearson correlation (Analyze-It for Microsoft Excel, Ver. 1.62), Passing–Bablok, and Bland–Altman analyses were performed (R software, Ver. 2.2.0) (2). For the evaluation of clinical significance, the reference-change value (RCV) was calculated (3) with Z being the 0.975 percentile of the gaussian distribution, and CVa and CVi being the analytical and intraindividual CVs of the test, respectively:

Formula

The multiplication by {surd}2 is because 2 values are being compared. The intraindividual CV was obtained from Westgard QC (4), and the analytical CV was estimated by a 4-month cumulative CV of quality-control results at an analyte concentration close to the sample value.

In total, 508 analyses were performed on 170 serum and 170 plasma samples from 163 patients (84 males and 79 females). The Pearson correlation revealed excellent correlation, except for IgG2, which had an outlier. Omitting the outlier gave better correlation (correlation coefficient = 0.99; 95% confidence interval, 0.98–1.0; Table 1 ). Passing–Bablok and Bland–Altman analyses revealed no significant differences (P <0.05) between plasma and serum for 15 and 12 of the 22 tests, respectively. Eleven and 3 tests (C4, CERU, and {lambda} light chain) were significantly different in either or both statistical analyses, respectively (Table 1 ).


View this table:
[in this window]
[in a new window]
 
Table 1. Tests, number of analyses/test, reference ranges, concentration ranges, correlation, and Passing–Bablok and Bland–Altman analyses (serum and heparin plasma samples).1

We next evaluated whether the statistically significant differences observed for these 14 tests were also clinically relevant by the RCV technique. This analysis revealed no clinically relevant differences between plasma and serum values, except for 1 AMG determination (of 24).

Passing–Bablok analysis revealed (a) a constant bias for AMG, (b) a proportional bias for IgA, and (c) a nonlinear relationship between serum and heparin plasma for PAB and IgG4. Visual examination of the regression line for PAB and IgG4 showed only slight nonlinearity. We further evaluated the nonlinearity for these 2 tests by use of the cusum test, as described in detail by Passing and Bablok (5). Briefly, a score is assigned to each data point: 0 for data points on the regression line and sqrt(L/l) and –sqrt(l/L) for measurements above and under the regression line, respectively (L denotes the number of measurements under the line and l those above). The data points are sorted along the fitted line, and cumulative sums of sorted scores are computed. If any of these cusums exceeds the threshold for a given significance level, nonlinearity can be concluded. For both PAB and IgG4, the threshold was exceeded at the 5% significance (P <0.05), but not at the 1% significance level (P >0.01).

In conclusion, our results suggest that for nephelometric quantification of specific proteins, serum and lithium-heparin plasma samples can be used interchangeably. It should be mentioned that plasma samples may form a cryoprecipitate when stored. This can be resolved with additional centrifugation (6).


References

  1. . Clinical and Laboratory Standards Institute. Procedures for the handling and processing of blood specimens, approved guideline—third edition. CLSI document H18–A3, Vol. 24, No. 38 2004 CLSI Wayne, PA. .
  2. Verzani J. Using R for Introductory Statistics 2004:433pp Chapman & Hall/CRC Press Boca Raton, FL. .
  3. Harris EK, Brown SS. Temporal changes in the concentration of serum constituents in healthy men. Distribution of within-person variances and their relevance to the interpretation of differences between successive measurements. Ann Clin Biochem 1979;16:169-176.[ISI][Medline] [Order article via Infotrieve]
  4. Westgard Quality Corporation. Desirable specifications for total error, imprecision, and bias, derived from biologic variation. http://www.westgard.com/biodatabase1.htm (accessed March 2006)..
  5. Passing H, Bablok W. A new biometrical procedure for testing the equality of measurements from two different analytical methods. J Clin Chem Clin Biochem 1983;21:709-720.[ISI][Medline] [Order article via Infotrieve]
  6. Heinrich RT. A heparin-precipitable fraction of human plasma. II. Occurrence and significance of the fraction in normal individuals and in various disease states. J Clin Invest 1957;36:605-616.[ISI][Medline] [Order article via Infotrieve]




This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Develter, M.
Right arrow Articles by Bossuyt, X.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Develter, M.
Right arrow Articles by Bossuyt, X.
Related Collections
Right arrow General Clinical Chemistry
Right arrow Clinical Immunology


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS