|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Automation and Analytical Techniques |
1 Ludwig-Maximilians-University, Medizinische Klinik Innenstadt, Munich, Germany.
2 Centre Hospitalier Universitaire Vaudois, Département de médecine interne Lausanne, Lausanne, Switzerland.
3 Albert-Ludwigs-University, Medical Department 2, Freiburg, Germany.
4 Ruprecht-Karls-University, Institute of Pharmacology, Heidelberg, Germany.
aAddress correspondence to this author at: Medizinische Klinik Innenstadt, Klinikum der LMU Mänchen, Ziemssenstr. 1, D-80336 Mänchen, Germany. Fax 49-89-5160-4428; E-mail martin.reincke{at}med.uni-muenchen.de.
| Abstract |
|---|
|
|
|---|
Methods: We tested 104 participants (27 with primary aldosteronism, 30 with essential hypertension, and 47 healthy controls) with the intravenous saline infusion test (2.0 L isotonic saline over 4 h), with repetitive sampling. In all blood samples, aldosterone concentration was measured by an in-house RIA after extraction and chromatography, by 2 commercially available RIAs without extraction (Aldosterone Maia, Adaltis; Active Aldosterone, Diagnostics Systems Laboratories) and by an automated CLIA (Advantage, Nichols Institute Diagnostics).
Results: Correlation coefficients for results of pairs of assays ranged from 0.74 to 0.98. Agreement between commercial assays and in-house RIA was best at the low to intermediate concentrations after saline infusion. Mean (SD) Adaltis and DSL RIA results were 2- to 3-times higher [healthy participants: 78 (25) ng/L and 56 (18) ng/L, respectively] than those obtained by Nichols CLIA [17 (8) ng/L] and in-house RIA [23 (18) ng/L]. Aldosterone concentrations measured by the Nichols CLIA were below the limit of detection (limit of the blank) in 27 of 47 healthy participants.
Conclusions: Aldosterone concentrations reported by the Adaltis and DSL nonextraction RIAs were consistently higher than those produced by the Nichols CLIA and the in-house RIA. The convenient Nichols CLIA showed better agreement with the in-house RIA, but the concentrations in healthy participants were frequently undetectable by this method. Uncritical application of cutoff values from the literature must be avoided.
| Introduction |
|---|
|
|
|---|
Bioassays and double isotope derivative assays are ponderous and complex methods, whereas RIAs for aldosterone(4) allow laboratorians to assay many samples and are frequently used. Only recently, an automated CLIA for measurement of aldosterone has been reported(5)(6).
Currently, most immunoassays use rabbit polyclonal antisera with various affinities and specificities(5). Furthermore, most of the assays in use do not involve chromatography or extraction steps(7)(8). Inadequate standardization, poor interlaboratory reproducibility, and the limited comparability of different immunoassays remain problematic and produce difficulties in defining cutoff values for PA(9)(10), essentially requiring each laboratory to establish its own reference intervals.
In this study, we compared a newly introduced, commercial, automated CLIA and 2 established commercial aldosterone RIAs with an in-house RIA after extraction and chromatography as comparison method to measure aldosterone concentrations during dynamic testing for PA.
| Materials and Methods |
|---|
|
|
|---|
We based the diagnosis of PA on the following biochemical criteria(11): repeatedly increased aldosterone/renin ratios (according to cutoff values of the local laboratory using the Adaltis RIA for measurement of aldosterone), increased urinary aldosterone excretion [>15 µg/day, or a previous pathological saline infusion test (serum aldosterone at 240 min >80 ng/L or 223 nmol/L)]. Differential diagnosis of PA was based on an algorithm (combining the results of computed tomography/magnetic resonance imaging and a posture test)(12) and/or selective vein catheterization. The latter was performed in 15 of 25 patients. We confirmed the diagnosis of PA in 25 patients. In 2 patients with suspected PA, diagnostic procedures had not been finished by the end of this study. Twelve of the 25 patients had bilateral adrenal hyperplasia, 4 had an aldosterone-producing adenoma and underwent surgery. For the remainder, subtype evaluation had not been completed at the end of this study.
With the exception of spironolactone, which was withdrawn
6 weeks before testing, patients took their regular antihypertensive medication while being studied. Sodium intake was unrestricted. None of the subjects had chronic renal failure (creatinine >132.6 µmol/L), a condition that previously had been associated with falsely increased results when direct methods had been used(7). Mean (SD) serum creatinine was 70.8 (8.9) µmol/L for healthy individuals, 70.8 (17.7) µmol/L for patients with essential hypertension, and 79.6 (26.5) µmol/L for patients with PA. Participants did not take contraceptives or other potentially cross-reactive medication.
The study protocol was approved by the ethics committee of the University of Freiburg, according to the requirements of the Declaration of Helsinki, and written informed consent was obtained from all participants.
aldosterone assay methods
In all blood samples, we determined aldosterone concentration by 4 different methods. As a comparison method, we used the specific in-house RIA (plasma samples) established at the Steroid Laboratory of the University of Heidelberg, using tritiated aldosterone ([1,2,4,6, 3H]aldosterone; Amersham Biosciences) and a rabbit antialdosterone antiserum, raised and characterized in the steroid laboratory, as described elsewhere(13). Before RIA, we performed recovery-corrected extraction and chromatographic purification, thereby efficiently removing cross-reacting serum components. In more detail, we carried out the chromatographic separation of aldosterone, as previously described(14), with minor modifications. Celite (Celite 545 AW; Sigma Aldrich) was used as an inert support for partition chromatography. The stationary phase consisted of 30% formamide in water; while the mobile phase consisted of a mixture of ethyl acetate in n-hexane with increasing polarity, eluting aldosterone with 50% ethyl acetate in n-hexane. The limit of blank [determined as mean (2SD) of blank] of the method was 1.5 pg per tube (7.5 ng/L). The recovery of a known amount of aldosterone determined repeatedly in quality control samples was 103.8 (8.2)%.
Furthermore, aldosterone concentration was determined by 3 commercially available assays according to the protocol given by the manufacturers. We used 2 RIAs without extraction in serum (Aldosterone Maia) and plasma samples (Active Aldosterone, DSL). Furthermore, serum aldosterone concentration was measured on a fully automated immunochemiluminescence analyzer (Aldosterone, Nichols Institute Diagnostics)(15). Additional characteristics of the assays are given in Table 1
.
|
statistical analysis
Results are displayed as mean (SD), except for Fig. 1
where, for reasons of clarity, SE bars are shown. Statistical analysis of the data was performed using Microcal Origin 6.0 and SAS 6.12. Pearson coefficient of correlation was calculated to describe the linear component of correlation. For the Nichols assay, values below the limit of detection (as given by the manufacturer, <15 ng/L) were set to 15 ng/L. This was the case in 57% of samples from healthy participants, 3% from patients with PA, and 46% from patients with essential hypertension. To compare the 3 commercial assays with the in-house RIA (reference method), we plotted the results of the assays as described by Bland and Altman(16), showing the difference (given as percent of the mean) against the mean of reference method and each commercial assay, respectively. Furthermore, to compare the commercial assays with the reference method, we performed Deming regression analysis with Analyze-It Clinical Laboratory (V 1.67, Analyze-It Ltd.).
|
| Results |
|---|
|
|
|---|
Overall correlation between assays ranged from 0.74 to 0.98 (Pearson coefficient of correlation) and increased during the course of saline infusion: Correlation between assays before saline infusion was from 0.74 to 0.95, whereas at the end of saline infusion, correlation was between 0.93 and 0.98 (Table 2
).
|
At lower concentrations, aldosterone measured by Adaltis RIA and DSL RIA showed a better correlation than each of these RIAs with Nichols CLIA. In the range of 0150 ng/L (0416 nmol/L), aldosterone (as determined by Adaltis or DSL) measurements by Nichols CLIA produced values below the limit of detection in a substantial part of the samples.
Despite good overall correlation, absolute aldosterone concentrations differed dramatically (Table 3
and Fig. 1
). Mean aldosterone concentrations measured by Adaltis RIA and DSL RIA were 2 to 3 times higher than those obtained by the reference method (in-house RIA). In contrast, mean values produced by Nichols CLIA better corresponded to those of the reference method (Table 3
). In the Deming regression analysis, using the in-house RIA as the reference method, we determined that a comparison with the Nichols CLIA revealed a slope of 1.2 (r = 0.91), whereas we found larger deviations for the Adaltis RIA (slope 1.5, r = 0.73) and the DSL RIA (slope 1.4, r = 0.85). These findings were confirmed when analyzing the data by means of the Bland-Altman plot (Fig. 2
). Whereas the results obtained by the Nichols CLIA are scattered symmetrically on both sides of the x-axis (difference = 0%), values obtained by the DSL and Adaltis RIA show a distribution clearly above the x-axis, with a mean deviation of 75% to 100% from the reference method. For all 3 commercial assays, absolute deviation from the reference method and the scatter around the mean deviation was most pronounced in the very low range of aldosterone concentrations [<80 ng/L (220 nmol/L)] (Fig. 2
).
|
|
| Discussion |
|---|
|
|
|---|
However, in our study comparing the Nichols CLIA with 2 commercial RIAs and, as reference method, an in-house aldosterone assay with extraction and chromatography, we found major differences in absolute mean values between these assays despite fairly high interassay correlations.
The discrepancies among assay results could be a consequence of different factors: First, the specificity of the antisera or antibodies varies between assays. As a consequence, reported aldosterone concentrations might be falsely high because of cross-reacting steroids or other interfering substances, especially if assays without extraction and chromatography are used. Although crossreactivity of the antibodies reported by the manufacturers does not show major differences (Table 1
), this does not exclude the influence of other potentially interfering substances. Second, variability could derive from the fact that aldosterone is measured in either plasma or serum. Although not explicitly investigated in our study, the choice of sample material seems to play a minor role. We found no better correlations between methods using the same or different sample material. Finally, and probably most important, calibration of the assays seems to be different, which, of course, has a strong impact on the results obtained. This might explain the systematically higher values for Adaltis and DSL RIAs when compared with the Nichols CLIA assay. An independent reference method based on mass spectrometry would be a way to further investigate this aspect.
Best correlation among the assays was found in the lower to middle range of values observed at the end of the saline infusion test, when Pearson coefficient of correlation ranged from 0.93 to 0.98. In contrast, at the upper and lower end of the concentration range discrepancies among assays increased. The increasing scatter of data points in the Bland-Altman plots for values below mean aldosterone concentrations of 80 ng/L clearly points to a significant lack of sensitivity of all commercial assays, making results in this concentration range less reliable.
Although an ideal assay for routine clinical application should be practicable, fast, and cost-effective, as well as accurate across the expected physiologic and pathologic ranges, these requirements are sometimes conflicting. It is important to keep in mind that a fast and simple assay providing wrong results may cause more harm than good. All three commercially available methods investigated in our study skipped the time-consuming and tedious steps of extraction and chromatographic separation of the samples. Major advantages of the recent Nichols CLIA assay are the automated system, time effectiveness, the option for simultaneous determination of renin concentration from the same specimen, and the lack of radioactive hazards. However, according to our results, the sensitivity of the Nichols CLIA assay might be an issue especially in healthy participants, because in this group a high percentage of samples fall below the limit of detection.
The inadequate correlation among methods has important clinical implications. Dynamic test results, depending on the assay used, might be interpreted as abnormal if cutoff values from the literature are applied. For the saline infusion test, for example, different cutoff values have been recommended: Ganguly(23) refers to 85 ng/L (240 nmol/L), whereas in a recent review, Mulatero et al.(10) recommend 50 ng/L (140 nmol/L). The definition of assay-specific cutoff levels and a comparison of the assays in terms of accuracy in detecting PA were beyond the scope of our study. This would certainly have required not only a larger population but also control of potentially interfering preanalytic factors, such as antihypertensive medication, before saline infusion test. Whatever cutoff is used, however, their application will result in a higher number of false positive tests if aldosterone is measured by Adaltis RIA or DSL RIA. In contrast, measuring aldosterone with Nichols CLIA, despite an overall acceptable correlation of values, would lead to a number of false-negative test results.
Many questions related to biochemical testing in diagnosis and differentiation of the subtypes of primary aldosteronism are, of course, related to difficulties in standardization of preanalytic conditions. Medication, salt intake, local protocol for performing dynamic tests, and many other potentially confounding factors vary among studies. However, our study emphasizes the need for harmonization and standardization of aldosterone assays as a prerequisite to compare results from different studies. The impact of the variability on the aldosterone/renin ratio, which is being used increasingly as a screening variable, can be expected to be in the same magnitude as the differences in absolute aldosterone values. This will enormously affect the result of the ratio, and thereby also the prevalence of positive test leads to a distinct study. Uncritical application of published cutoff values may lead to serious consequences for the patients affected. Standardized assays, calibrated against a mass spectrometric reference method, would improve patient management and would allow more confidence in comparing results obtained in different laboratories.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
M. Stowasser Update in Primary Aldosteronism J. Clin. Endocrinol. Metab., October 1, 2009; 94(10): 3623 - 3630. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Taylor, D. P. Cooper, R. D. Gordon, and M. Stowasser Measurement of Aldosterone in Human Plasma by Semiautomated HPLC-Tandem Mass Spectrometry Clin. Chem., June 1, 2009; 55(6): 1155 - 1162. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Funder, R. M. Carey, C. Fardella, C. E. Gomez-Sanchez, F. Mantero, M. Stowasser, W. F. Young Jr., and V. M. Montori Case Detection, Diagnosis, and Treatment of Patients with Primary Aldosteronism: An Endocrine Society Clinical Practice Guideline J. Clin. Endocrinol. Metab., September 1, 2008; 93(9): 3266 - 3281. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Cavalier, A. Carlisi, J.-P. Chapelle, and P. Delanaye Analytical Quality of Calcitonin Determination and Its Effect on the Adequacy of Screening for Medullary Carcinoma of the Thyroid Clin. Chem., May 1, 2008; 54(5): 929 - 930. [Full Text] [PDF] |
||||
![]() |
J Manolopoulou, M Bielohuby, S J Caton, C E Gomez-Sanchez, I Renner-Mueller, E Wolf, U D Lichtenauer, F Beuschlein, A Hoeflich, and M Bidlingmaier A highly sensitive immunofluorometric assay for the measurement of aldosterone in small sample volumes: validation in mouse serum J. Endocrinol., February 1, 2008; 196(2): 215 - 224. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Oberleithner, C. Riethmuller, H. Schillers, G. A. MacGregor, H. E. de Wardener, and M. Hausberg Plasma sodium stiffens vascular endothelium and reduces nitric oxide release PNAS, October 9, 2007; 104(41): 16281 - 16286. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Stowasser and R. D. Gordon Aldosterone Assays: An Urgent Need for Improvement Clin. Chem., September 1, 2006; 52(9): 1640 - 1642. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |