Clinical Chemistry 53: 152-153, 2007; 10.1373/clinchem.2006.074229
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via ISI Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Related Collections
Right arrow Clinical Immunology
(Clinical Chemistry. 2007;53:152-153.)
© 2007 American Association for Clinical Chemistry, Inc.


Letters to the Editor

Detection of Monoclonal Proteins by Capillary Zone Electrophoresis: Comparison of 2 Multichannel Automated Systems

Xavier Bossuyta and Godelieve Mariën

Laboratory Medicine, University Hospitals Leuven, Katholieke Universiteit Leuven, Leuven, Belgium

aAddress correspondence to this author at: Laboratory Medicine, Immunology, University Hospitals Leuven, Herestraat 49, B-3000 Leuven, Belgium. Fax 32-16-34-70-42; e-mail xavier.bossuyt{at}uz.kuleuven.be.


To the Editor:

Capillary zone electrophoresis (CZE) is an alternative method for separation of serum proteins (1). Two dedicated and automated multichannel instruments are available, the Paragon 2000 (Beckman-Coulter) and the Capillarys (Sebia). The sensitivity and specificity of Paragon 2000 for detection of monoclonal proteins have been reported to be 95%(2)(3)(4) and 78%(4), respectively. The low reported specificity reflects the frequent occurrence of slight abnormalities in the electropherogram at the anodal part of the {gamma}-globulin fraction (fibrinogen region)(4).

Only 2 studies (5)(6) evaluated the Capillarys for detection of monoclonal proteins and reported a high sensitivity, and prospective studies in a routine clinical setting are lacking.

We studied 597 consecutive samples submitted to the laboratory to test for the presence of a monoclonal protein or to reevaluate a known monoclonal protein. All samples were analyzed by the Capillarys (software version 5.2.1) and Paragon 2000 (software version 1.6.02) systems and by immunofixation, which was used as the reference standard. Immunofixation was performed on the Hydrasys Automate (Sebia) according to the manufacturer’s instructions, with the use of Hydragel 4 immunofixation gels.

Immunofixation revealed the presence of a distinct monoclonal protein in 246 of the 597 samples. The distribution of the types of monoclonal proteins is given in Table 1 . The sensitivity of Capillarys for detecting a monoclonal protein was 90% [95% confidence interval (CI), 85%–93%]. This sensitivity was comparable to that of the Paragon, which was 91.5% (95% CI, 87%–95%; P >0.5 by {chi}2).


View this table:
[in this window]
[in a new window]

 
Table 1. CZE analysis of 246 consecutive samples positive by immunofixation.

In 267 samples, immunofixation revealed no abnormalities. The specificity of the Capillarys was calculated to be 87% (95% CI, 83%–91%), higher than the specificity of the Paragon, which was 58% (95% CI, 52%–66%; P <0.0001 by {chi}2). The lower specificity of the Paragon 2000 was attributable to the previously described (4) disturbed morphology at the fibrinogen position. If such disturbed morphology was not considered abnormal, then the specificity of the Paragon was 80% (95% CI, 75%–88%;P ≤ 0.03 for comparison with Capillarys by {chi}2).

In 84 samples, immunofixation was difficult to interpret (presence of M-protein not excluded, M-protein very faint or not identifiable, differentiation between monoclonal and polyclonal pattern questionable). The results for the Paragon 2000 and Capillarys for these samples are available from the authors.

Quantification of the monoclonal protein by delimitation was possible in 87 samples. There was a good correlation between the Capillarys and Paragon (r = 0.99; Pearson). Bland-Altman analysis revealed an intercept of 0.81 (95% CI, 0.558–1.096) and a slope of 0.96 (95% CI, 0.95–0.98).

As shown above, each system may fail to detect monoclonal proteins. For example, the Paragon 2000 failed to detect a monoclonal protein in a sample in which immunofixation revealed an IgM{kappa} monoclonal protein. This monoclonal protein was clearly detected by the Capillarys. Conversely, the Capillarys failed to detect a significant IgM{kappa} monoclonal protein and a IgG{kappa} monoclonal protein. In both samples the CZE electropherograms showed a minor disturbance in morphology of the {gamma}-region but not the obvious monoclonal peak that was observed with agarose gel electrophoresis. A remarkable finding on the Capillarys analysis, observed in both samples, was an unusually long separation between the albumin fraction and the {alpha}1-globulin fraction.


Acknowledgments

We thank Sebia for providing the reagents to perform this study.


References

  1. Bossuyt X. Separation of serum proteins by automated capillary zone electrophoresis [Review]. Clin Chem Lab Med 2003;41:762-772.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  2. Katzmann JA, Clark R, Sanders E, Landers JP, Kyle RA. Prospective study of serum protein capillary electrophoresis immunotyping of monoclonal proteins by immuno-subtraction. Am J Clin Path 1998;110:503-509.[ISI][Medline] [Order article via Infotrieve]
  3. Bossuyt X, Marien G. False-negative results in detection of monoclonal proteins by capillary zone electrophoresis: a prospective study. Clin Chem 2001;47:1477-1479.[Free Full Text]
  4. Mariën G, Vranken G, Demuylder M, Blanckaert N, Bossuyt X. Clinical capillary zone electrophoresis of serum proteins: balancing high sensitivity and high specificity. Clin Chem 2003;49:1419-1420.[Free Full Text]
  5. Bossuyt X, Lissoir B, Mariën G, Maisin D, Vunckx J, Blanckaert N, et al. Automated serum protein electrophoresis by Capillarys®. Clin Chem Lab Med 2003;41:704-710.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  6. Gay-Bellile C, Bengoufa D, Houze P, Le Carrer D, Benlakehal M, Bousquet B, et al. Automated multicapillary electrophoresis for analysis of human serum proteins. Clin Chem 2003;49:1909-1915.[Abstract/Free Full Text]




This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via ISI Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Related Collections
Right arrow Clinical Immunology


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS