Clinical Chemistry 53: 98-103, 2007.
First published November 16, 2006; 10.1373/clinchem.2006.075184
(Clinical Chemistry. 2007;53:98-103.)
© 2007 American Association for Clinical Chemistry, Inc.
|
Automation and Analytical Techniques |
Molecular BeaconBased Temperature Control and Automated Analyses for Improved Resolution of Melting Temperature Analysis Using SYBR I Green Chemistry
Christoffer Nellåker1,a,
Ulf Wållgren2 and
Håkan Karlsson1
1 Department of Neuroscience, Karolinska Institutet, Stockholm, Sweden.
2 FLIR Systems AB, Danderyd, Sweden.
aAddress correspondence to this author at: Retzius väg 8, 171 77 Stockholm, Sweden. Fax: +468325325; e-mail christoffer.nellaker{at}ki.se.
 |
Abstract
|
|---|
Background: Melting temperature analysis of products amplified with SYBR I Green chemistry is a cheap and effective method for identification of sequence differences. When used in conventional quantitative real-time PCR instruments (qPCR), this method is limited by temperature variations over the heating block and low numbers of fluorescence measurements during the dissociation step, which hamper the ability of most instruments to report accurate and precise melting temperatures.
Methods: We designed a molecular beaconbased temperature indicator probe (Tm-probe) to control for variations in temperatures over the heating block of the instrument. In addition, we wrote an automated curve-fit analysis algorithm of dissociation data to use multiple data points with a gaussian curve fit to extrapolate precise melting temperatures.
Results: Use of the Tm-probe in conjunction with the analysis algorithm and multiple dissociations improved SDs of melting temperatures over a 96-well plate from 0.19 to 0.06 °C
Conclusions: Melting temperature analyses with SYBR I Green chemistry on conventional qPCR instruments can be improved by the use of a Tm-probe in conjunction with curve-fit analysis of data. Resolution improvement up to 3-fold is possible and allows additional melting temperatures to be identified.
 |
Introduction
|
|---|
Quantitative real-time PCR (qPCR) analyses are commonly used and are rapid methods for the analysis of nucleic acid levels within both clinical and basic sciences. This method uses fluorescent dyes to indicate levels of amplified PCR products as the reactions proceed (i.e., in real time). Sequence specificity of amplicon detection in qPCRs varies from the most specific Minor GrooveBinding-probes, to Taqman-probes (1), to the indiscriminate SYBR I Green detection system, which merely fluoresces with increased intensity when bound to double-stranded DNA. For economic reasons, however, SYBR I Green chemistry is one of the more widely used detection systems in qPCR. To check for erroneous products, a dissociation step must be run after the amplification has completed. During this step, the instrument gradually heats the amplified reactions and measures the decrease in fluorescence signal as the 2 strands of the products dissociate. The temperature at which the rate of signal decline is maximal (i.e., the peak of the negative derivative of the fluorescence measurements) is defined as the melting temperature (Tm) and is related to the base-pair composition of the product. Therefore in addition to detecting erroneous products, Tm analyses can indicate sequence variations between amplicons. Tm analyses have been recently employed for strain identification in clinical and veterinary virology (2)(3), typing of bacterial strains (4), identification of expression patterns of highly homologous genomic elements (5), and genotyping of HLA variants (6). Furthermore, this approach has previously been used to detect translocations in cancers (7) and to scan for single-nucleotide polymorphisms (8).
High-resolution Tm analyses were not the original purpose of most qPCR instruments. Temperature variations over the heating block and low numbers of fluorescence measurements during the dissociation step hamper the ability of most instruments to report accurate and precise Tms (9). To improve Tm analyses without acquiring a specialized instrument for the purpose (such as the HR-1TM from Idaho Technology), 2 major issues must be resolved, the temperature variations inherent to the heating block must be normalized and more precise Tms be calculated from low-resolution temperature data.
Tm analysis programs that apply a curve-fit method to the data points obtained during melting have been developed for other systems. We present a program that was adapted for the ABI Prism® 7000 with SDS v1.2.3, but the principles of this program can be adapted for other systems. In addition, we present the application of a Tm-probe used to control for temperature variations, similar to that applied for a microfluidic platform by Dodge et al. (10).
 |
Materials and Methods
|
|---|
qpcr and dissociation curve
We used an ABI Prism 7000 SDS (Applied Biosystems) with a Precision Plate Holder (Applied Biosystems), white Thermo-Fast® 96 detection plates (ABgene), and the version 1.2.3 SDS software package (Applied Biosystems). The Platinum SYBR Green qPCR SuperMix UDG (Invitrogen) was used in each of the 25-µL reactions. Unless stated otherwise, reactions were performed with 250 nmol/L of the forward (TCA GGT CAA CAA TAG GAT GAC AAC A) and reverse (CAA TGA GGG TCT ACA CTG GGA ACT) primers and 133 nmol/L of the Tm probe (FAM-TTT TTT T(-TAMRA)TC GGC CGC TCC CCC CCC CCC CCA GCG GCC GA-BHQ2). The passive reference dye ROX was not included. Unless stated otherwise, 3100 copies of a pcDNA2.1 (Invitrogen) TA cloning plasmid containing a HERV-W gag fragment from chromosome 3q26 were used as an amplification template (1) (see Fig. 5
). Signal-to-noise ratios were calculated as the signal fluorescence during the last cycles divided by the noise fluorescence during the first cycles (11). Reaction efficiencies of the assay in the presence or absence of the Tm-probe were calculated as previously described (1).

View larger version (9K):
[in this window]
[in a new window]
|
Figure 5. Demonstration of improved resolution in detection of 4 different amplification targets.
(A), target sequences of plasmid templates. Dashes indicate nucleotide identical to the alignment sequence. Open circles indicate gaps in sequences introduced to optimize the alignment. (B), Tm variations of amplicons from the 4 different templates detected with and without Tm-probe temperature correction. (C), the same data divided into the respective template sequences used, indicate GcTm calculated Tms, + indicate GcTm-calculated Tms corrected with Tm-probe.
|
|
tm probe design
To allow detection in an ABI Prism 7000 simultaneously as SYBR I Green fluorescence, a molecular beacon (12)(13) was designed to have a stem structure with a Tm higher than those observed for the target transcript amplicons (85 °C), for which the SYBR signal is minimal [Web page for tm analysis and generation of Fig. 2
folding http://www.bioinfo.rpi.edu/applications/mfold/dna/form1.cgi (14)(15)(16)]. During denaturation, the fluorescence of molecular beacons increases rather than decreases upon melting, allowing the derivative curve of dissociation data to be easily distinguished from that of any SYBR products. To obtain absorption and emission wavelengths appropriate for the instrument, we used a wavelength-shifting molecular beacon design (17). The molecular beacon was triple-labeled with 6-carboxyfluorescein (FAM) in the 5' end, 6-carboxytetramethylrhodamine (TAMRA) attached to the 6th thymidine from FAM, and Black Hole Quencher (BHQ)-2 in the 3' end (Fig. 1
). In the hybridized configuration the FAM in the Tm-probe absorbs the 485-nm excitation provided by the instrument. The high-energy state FAM undergoes fluorescence resonance energy transfer (FRET) to TAMRA, which in turn donates its energy through FRET to the BHQ-2. At temperatures >85 °C the molecular beacon undergoes a conformational change, and the TAMRA will no longer transfer any energy to BHQ-2 because it is no longer in close enough proximity and will thus fluoresce at 580 nm. The Tm probe was purchased from MedProbes (Eurogentec).

View larger version (10K):
[in this window]
[in a new window]
|
Figure 2. Derivative dissociation curve data from SYBR I Green detection (top) and TAMRA detection (bottom).
The raw data is displayed as a dashed line and the gaussian curve fit with GcTm is shown as an solid line. The labels n1, n2, and n3 show the data ranges used for curve fitting. Tm indicates the point on the gaussian curve determined to correspond to the Tm by GcTm.
|
|

View larger version (7K):
[in this window]
[in a new window]
|
Figure 1. Tm-probe design.
Structure of the molecular beacon used for the Tm-probe, showing the hairpin conformation and the positions of the dyes FAM, TAMRA, and BHQ-2.
|
|
tm analysis
We used the MATLABTM (The MathWorks) version 7.0.1.24704 with The Optimization Toolbox to write an automated analysis algorithm for data from Sequence Detection Software version 1.2.3 used in conjunction with an ABI Prism 7000. The program, gaussian curve fit analysis of TM (GcTm), was designed to determine Tms of amplicons by fitting gaussian curves to derivative data from dissociation analyses (Fig. 2
). The peak of the negative derivative data is automatically selected by taking the values differing from the mean derivative over all temperatures by at least 1.2 SDs. We also designed the program to use the Tm of the Tm-probe to normalize temperatures of amplified products reported from the instrument in each well. The Tm normalization calculation took the Tms, determined by GcTm, of the amplicon minus that of the corresponding Tm-probe plus the average of all the Tm-probes used in that experiment. The program is available for download at http://www.neuro.ki.se/kristensson/tmanalysis.html.
 |
Results
|
|---|
To evaluate the Tm analysis methods we used a plasmid template in 25-µL reactions in all wells of a 96-well plate. The Tms reported by the SDS software over the 96-well plate show a range of 0.6 °C in reported temperatures (SD 0.19 °C). Each Tm was reported as 1 of 3 discreet temperatures (Fig. 3
). Analysis with the GcTm algorithm of the dissociation data gave Tms with higher resolutions than those reported by the SDS software, based on a gaussian curve fitted to multiple data points (on average 5.25, Fig. 3
), but the systematic variations in reported temperatures remained (range 0.66 °C). To control for the temperature variations over the heating block, a molecular beaconbased Tm-probe was designed (Fig. 1
). Gaussian curves fitted to derivative dissociation curve data for the SYBR and TAMRA detectors are shown in Fig. 2
. The negative derivative dissociation data for the Tm-probe with the TAMRA detector displays an artifactual peak with changes in SYBR fluorescence, because in our assay SYBR has a broad emission spectrum and fluoresces an order of magnitude more than TAMRA. In the interest of cost-effectiveness, we used the lowest concentration of Tm-probe giving reliable Tms with low variation. The Tms for the Tm-probe calculated with GcTm showed variations over the heating block similar to those of the amplicons (Fig. 3
). Correcting the Tms of the amplicons by subtracting those of the Tm-probe (and adding the average Tm of all the Tm-probes) gave a profile over the 96-well plate with no apparent systematic variations with regard to well position (Fig. 3
). Variations in reported Tms had an SD of 0.18 °C for the SYBR alone and an SD of 0.12 °C when corrected with the Tm-probe (Fig. 4
). The Tm-probe showed a higher variation in Tms within wells than the amplicons. In addition, although the TAMRA fluorescence displayed spatial variation similar to that of SYBR I Green, the TAMRA fluorescence fluctuated more. Therefore the Tm-probe showed high accuracy but lower precision. To reduce variations, we set the ABI Prism 7000 to run 2 additional dissociation curve analyses immediately after the primary run. These dissociations were analyzed individually with GcTm and the resulting Tms averaged. The mean of 3 dissociations resulted in an SD of 0.17 °C for the SYBR alone and an SD of 0.06 °C when corrected with the Tm-probe (Fig. 4
). Additional dissociations beyond the 3 described did not further decrease the SDs of the corrected Tms. To illustrate the improved precision and accuracy, we analyzed 4 different plasmids with variations in the target sequences (Fig. 5A
). The addition of the Tm-probe allowed the identification of 3 distinct groups of Tms, indicating at least 3 unique sequences that were indistinguishable without correction of amplicon Tms (Fig. 5B
). Uncorrected and corrected Tms for each of the 4 plasmid templates are shown in Fig. 5C
to demonstrate that sequence variations do not necessarily give Tm differences. Although the Tm-probe described in the present work had no effect on reaction efficiency (data not shown), the fluorescence emitted by FAM, which is not transferred by FRET to TAMRA, increased the background and thereby reduced the signal-to-noise ratio from 9 to 4.5.

View larger version (33K):
[in this window]
[in a new window]
|
Figure 3. Distribution of reported temperatures over a 96-well plate in an ABI Prism 7000 for 1 template sequence.
Top, one example of the Tms reported by the SDS software. Upper middle, indicates the amplicon Tms reported by SYBR and calculated by the Tm-analysis program. Lower middle, Tm-probe Tms as calculated by GcTm Tm-analysis program. Bottom, the normalized Tms of the amplicons, (i.e., calculated Tms corrected for temperature variations with the Tm-probe data). The lower 3 plots represent data averaged for 3 dissociation curves.
|
|

View larger version (11K):
[in this window]
[in a new window]
|
Figure 4. Reported Tms for 1 template in a 96-well plate.
represents amplicon Tms for the SYBR dye alone calculated with the Tm-analysis program. + indicates amplicon Tms corrected with the Tm-probe. Single dissociation indicates data from a single dissociation curve, analyzed with GcTm. Triple dissociation indicates that the data is the mean of 3 repeated dissociation analyses on the same 96-well plate and analyzed with GcTm.
|
|
 |
Discussion
|
|---|
According to the manufacturers specifications of the ABI Prism 7000 instrument, the temperature uniformity over a 96-well block is ±0.5 °C, a feature that limits the obtainable resolution when Tms of amplicons are examined. Using the Tm of a plasmid template as a reporter, we found that higher Tms were reported consistently in the middle of the 96-well reaction plate and that Tms were reported as discreet temperatures (Fig. 3
). We wrote an analytical algorithm to extrapolate accurate Tms from multiple data points. Using this algorithm in conjunction with a molecular beaconbased Tm-probe, we improved resolution 3-fold.
Some caveats should be considered with the use of the proposed Tm analyses. The reason for the observed variations in Tm-probe melting temperatures between wells is not known. However, the tests performed have shown a greater variation in wells flanked by empty wells or in wells at the corners of the plate, possibly a sign of evaporation effects and salt-concentration changes affecting the Tm-probe to a larger extent than the amplicons.
Furthermore, for Tm analyses of multiple sequences in the same reactions, SYBR I Green chemistry may not be optimal and other dyes (such as LCGreen) have been suggested to be more appropriate (18). Although the use of LCGreen can improve the determination of amplicon Tms, it will not eliminate the need for 3 dissociation analyses to be performed, because the primary source of variation stems from the Tm-probe measurements rather than those of the amplicons. SYBR I Green chemistry is adequate for comparison of Tms of amplicons between wells. It should be noted that differing sequences can share the same Tm, as is the case with the plasmids corresponding to the HERV-W gag on chromosomes 5p13 and 12p13.
In conclusion, we report methodological improvements for Tm analyses on an ABI Prism 7000 instrument. By use of the addition of a Tm-probe in combination with a curve-fit method (GcTm) and repeated measurements to determine Tms of amplicons, we have improved Tm precision from ±0.36 °C to ±0.12 °C (with a 95% confidence interval).
 |
Acknowledgments
|
|---|
This study was generously supported by the Stanley Medical Research Institute and the Swedish Research Council (project no. K2006-21X-20047-01-3). The authors declare that they have no competing interests. We thank Cecilia Eckervig from MedProbes whose expertise aided the design of the Tm-probe. We also thank Dr. Yuanrong Yao in our laboratory for providing the plasmids used as templates.
 |
Footnotes
|
|---|
1 Nonstandard abbreviations: qPCR, quantitative real-time PCR; Tm, melting temperature; FAM, 6-carboxyfluorescein; TAMRA, 6-carboxytetramethylrhodamine; BHQ, Black Hole Quencher-2; FRET, fluorescence resonance energy transfer; GcTm, Gaussian curve fit analysis of Tm. 
 |
References
|
|---|
- Yao Y, Nellaker C, Karlsson H. Evaluation of minor groove binding probe and Taqman probe PCR assays: influence of mismatches and template complexity on quantification. Mol Cell Probes 2006;20:311-316.[ISI][Medline]
[Order article via Infotrieve]
- Waku-Kouomou D, Alla A, Blanquier B, Jeantet D, Caidi H, Rguig A, et al. Genotyping measles virus by real-time amplification refractory mutation system PCR represents a rapid approach for measles outbreak investigations. J Clin Microbiol 2006;44:487-494.[Abstract/Free Full Text]
- Pham HM, Konnai S, Usui T, Chang KS, Murata S, Mase M, et al. Rapid detection and differentiation of Newcastle disease virus by real-time PCR with melting-curve analysis. Arch Virol 2005;150:2429-2438.[CrossRef][ISI][Medline]
[Order article via Infotrieve]
- Harasawa R, Mizusawa H, Fujii M, Yamamoto J, Mukai H, Uemori T, et al. Rapid detection and differentiation of the major mycoplasma contaminants in cell cultures using real-time PCR with SYBR Green I and melting curve analysis. Microbiol Immunol 2005;49:859-863.[ISI][Medline]
[Order article via Infotrieve]
- Nellaker C, Yao Y, Jones-Brando L, Mallet F, Yolken RH, Karlsson H. Transactivation of elements in the human endogenous retrovirus W family by viral infection. Retrovirology 2006;3:44.[CrossRef][Medline]
[Order article via Infotrieve]
- Graziano C, Giorgi M, Malentacchi C, Mattiuz PL, Porfirio B. Sequence diversity within the HA-1 gene as detected by melting temperature assay without oligonucleotide probes. BMC Med Genet 2005;6:36.[CrossRef][Medline]
[Order article via Infotrieve]
- Bohling SD, Wittwer CT, King TC, Elenitoba-Johnson KS. Fluorescence melting curve analysis for the detection of the bcl-1/JH translocation in mantle cell lymphoma. Lab Invest 1999;79:337-345.[ISI][Medline]
[Order article via Infotrieve]
- Germer S, Higuchi R. Single-tube genotyping without oligonucleotide probes. Genome Res 1999;9:72-78.[Abstract/Free Full Text]
- Herrmann MG, Durtschi JD, Bromley LK, Wittwer CT, Voelkerding KV. Amplicon DNA melting analysis for mutation scanning and genotyping: cross-platform comparison of instruments and dyes. Clin Chem 2006;52:494-503.[Abstract/Free Full Text]
- Dodge A, Turcatti G, Lawrence I, de Rooij NF, Verpoorte E. A microfluidic platform using molecular beacon-based temperature calibration for thermal dehybridization of surface-bound DNA. Anal Chem 2004;76:1778-1787.[Medline]
[Order article via Infotrieve]
- Kutyavin IV, Afonina IA, Mills A, Gorn VV, Lukhtanov EA, Belousov ES, et al. 3'-minor groove binder-DNA probes increase sequence specificity at PCR extension temperatures. Nucleic Acids Res 2000;28:655-661.[Abstract/Free Full Text]
- Bonnet G, Tyagi S, Libchaber A, Kramer FR. Thermodynamic basis of the enhanced specificity of structured DNA probes. Proc Natl Acad Sci U S A 1999;96:6171-6176.[Abstract/Free Full Text]
- Tyagi S, Kramer FR. Molecular beacons: probes that fluoresce upon hybridization. Nat Biotechnol 1996;14:303-308.[CrossRef][ISI][Medline]
[Order article via Infotrieve]
- Zuker M. Mfold web server for nucleic acid folding and hybridization prediction. Nucleic Acids Res 2003;31:3406-3415.[Abstract/Free Full Text]
- SantaLucia J, Jr. A unified view of polymer, dumbbell, and oligonucleotide DNA nearest-neighbor thermodynamics. Proc Natl Acad Sci U S A 1998;95:1460-1465.[Abstract/Free Full Text]
- Peyret N. Prediction of nucleic acid hybridization: parameters and algorithms [PhD dissertation] 2000 Wayne State University Detroit, MI. .
- Tyagi S, Marras SA, Kramer FR. Wavelength-shifting molecular beacons. Nat Biotechnol 2000;18:1191-1196.[CrossRef][ISI][Medline]
[Order article via Infotrieve]
- Wittwer CT, Reed GH, Gundry CN, Vandersteen JG, Pryor RJ. High-resolution genotyping by amplicon melting analysis using LCGreen. Clin Chem 2003;49:853-860.[Abstract/Free Full Text]
The following articles in journals at HighWire Press have cited this article:

|
 |

|
 |
 
C. N. Gundry, S. F. Dobrowolski, Y. R. Martin, T. C. Robbins, L. M. Nay, N. Boyd, T. Coyne, M. D. Wall, C. T. Wittwer, and D. H.-F. Teng
Base-pair neutral homozygotes can be discriminated by calibrated high-resolution melting of small amplicons
Nucleic Acids Res.,
June 1, 2008;
36(10):
3401 - 3408.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. G. Herrmann, J. D. Durtschi, C. T. Wittwer, and K. V. Voelkerding
Expanded Instrument Comparison of Amplicon DNA Melting Analysis for Mutation Scanning and Genotyping
Clin. Chem.,
August 1, 2007;
53(8):
1544 - 1548.
[Abstract]
[Full Text]
[PDF]
|
 |
|