| Abstract |
|---|
|
|
|---|
Methods: Herpes simplex virus (HSV) was chosen as a model to compare the unlabeled probe method to an in-house reference assay using dual-labeled, minor groove binding probes. A saturating dsDNA dye (LCGreen® Plus) was used for real-time PCR. HSV-1, HSV-2, and an internal control were differentiated by PCR amplicon and unlabeled probe melting analysis after PCR.
Results: The unlabeled probe technique displayed 98% concordance with the reference assay for the detection of HSV from a variety of archived clinical samples (n = 182). HSV typing using unlabeled probes was 99% concordant (n = 104) to sequenced clinical samples and allowed for the detection of sequence polymorphisms in the amplicon and under the probe.
Conclusions: Unlabeled probes and amplicon melting can be used to detect and genotype as few as 10 copies of target per reaction, restricted only by stochastic limitations. The use of unlabeled probes provides an attractive alternative to conventional fluorescence-labeled, probe-based assays for genotyping and detection of HSV and might be useful for other low-copy targets where typing is informative.
| Introduction |
|---|
|
|
|---|
This report describes the limits of copy number detection, typing accuracy, and melting reproducibility with the use of unlabeled probes. Herpes simplex virus (HSV) was chosen to develop an unlabeled probe assay based on the availability of quantified DNA targets for HSV-1 and HSV-2, characterized clinical samples, the necessity of low-level detection (13)(14)(15)(16), and an established in-house reference assay (17).
| Materials and Methods |
|---|
|
|
|---|
HSV genomic DNA standards purified from HSV-1 (1 x 104 DNA copies/µL, Macintyre strain, ABI) and HSV-2 (1 x 104 DNA copies/µL, ABI) were used as quantified control standards. We subcloned 180-bp glycoprotein D amplicons from the ABI quantified HSV-1 and HSV-2 DNA into pCR® II-TOPO® (Invitrogen) following the manufacturers protocols. These clones, designated pHSV-1 and pHSV-2, confirmed the sequence of the glycoprotein D gene of the ABI-quantified HSV-1 and HSV-2 and determined the range of detection for the unlabeled probe method.
hsv reference assay
For comparison, we used an in-house reference assay, an Epoch minor groove binding probe-based assay (Nanogen) (17). The minor groove binding probe was 100% complementary to a region of the glycoprotein D gene in HSV-1 and -2. Samples were run on an ABI 7900 real-time PCR instrument as described by Stevenson et al. (17). Because the reference assay does not discriminate between HSV-1 and -2, only the presence or absence of HSV was tested.
primers and probe
Oligonucleotides, manufactured by Integrated DNA Technologies, were desalted and resuspended in 1x TE (10 mmol/L Tris, pH 8.0, 1 mmol/L EDTA). Primers for HSV were numbered from the ATG start codon of glycoprotein D gene. Primers used for HSV asymmetric real-time PCR were 458F and 603R. The unlabeled oligonucleotide probe (HSVP) was designed with 100% complementarity to HSV-2 and modified with a 3' amino C6 modification to prevent DNA polymerase extension (18). We used HSVFS and HSVRS to amplify HSV from clinical samples for sequencing and HSVnest for sequencing to confirm HSV type. We designed primers ICF and ICR to amplify the IC. Primer and probe sequences are summarized in Table 1
.
|
real-time pcrs and melting analysis
PCR amplification was performed in a LightCycler® (Roche Diagnostics). Real-time PCR and unlabeled probe melting analysis reactions contained 1x Roche FastStart DNA Hybridization Mix (includes dNTPs with dUTP and 1 mmol/L MgCl2), 0.1 µmol/L forward primer 458F, 0.5 µmol/L reverse primer 603R, 0.5 µmol/L probe HSVP, 0.15 µmol/L forward primer ICF, 0.15 µmol/L reverse primer ICR, 0.8x LCGreen Plus, an additional 2 mmol/L MgCl2, 0.25 units uracil–DNA glycosylase, and 1 µL DNA, in a total volume of 10 µL. Cycling and postamplification melting analysis were performed using the following conditions at a programmed transition rate of 20 °C/s: [55 °C(5:00) + 95 °C(10:00) + (95 °C(0:01) + 60 °C(0:01) + 72 °C(0:10)) x 55 cycles + 95 °C(0:00) + 40 °C(0:20) + 50 °C
95 °C at 0.5 °C/s + 40 °C(0:00)]. We used an arbitrary crossing threshold (CT) cutoff of 50 cycles to determine whether a sample was positive. After LightCycler amplification and melting analysis, samples were placed at 4 °C for
10 min before high-resolution melting analysis using an HR-1TM instrument (Idaho Technology). The HR-1 uses the same capillaries as the LightCycler and is capable of continuous, high-resolution melting and fluorescence data acquisition (19). For HR-1 melting analysis, samples were removed from 4 °C and melted from 60 °C to 95 °C at 0.3 °C/s. We analyzed HR-1 melting data by use of Idaho Technology High Resolution Melting Analysis software (version 1.5f).
detection and typing of hsv by unlabeled probe and amplicon melting
We blinded 182 previously assayed clinical HSV samples and tested them using unlabeled probes. Samples were amplified and typed by low-resolution melting in a LightCycler. All samples were subsequently melted in the HR-1 to confirm type. To assess genotypes for HSV-1 or -2, unlabeled probe and amplicon melting were required. A subset of indeterminate samples was reamplified if we observed ambiguous HSV typing by melting analysis. The 2nd amplification of indeterminate samples was performed as previously described, but without IC primers, and melted on the HR-1 to verify HSV type.
sequencing
We reamplified 129 positive and negative HSV samples, as determined by amplicon and unlabeled probe melting analysis, using primers HSVFS and HSVRS for sequencing analysis. Reaction conditions consisted of 1x CoralLoad [Qiagen, proprietary, contains Tris · Cl, KCl, (NH4)2SO4, 1.5 mmol/L MgCl2, orange dye, red dye; pH 8.7 (20 °C)], an additional 2 mmol/L MgCl2, 0.3 µmol/L forward primer HSVFS, 0.3 µmol/L reverse primer HSVRS, 0.5 mmol/L dNTPs, 1 units Qiagen Taq DNA polymerase, and 2 µL of the extracted clinical sample in a final volume of 40 µL. Cycling was performed in an ABI 9700 using the following conditions: 94 °C(3:00) + [94 °C(0:10) + 65 °C(0:30) + 72 °C(0:45)] x 45 cycles. We performed postamplification analysis by use of a 2% agarose gel to observe products. Excess primers and nucleotides were removed using a Qiagen PCR Cleanup set following manufacturers protocol. All positive amplification products were submitted to the ARUP sequencing facility for dye terminator sequencing with primer HSVnest. Some HSV-positive samples could not be adequately amplified with primers HSVFS and HSVRS and were sequenced from the 458F and 603R recovered amplicon.
dynamic range of hsv detection and typing
Plasmids pHSV-1 and pHSV-2 were quantified by A260/280 measurements to estimate copy number. We compared dilution series of each plasmid to ABI-quantified standards diluted to 1000 copies of HSV-1 or HSV-2 per reaction. We adjusted plasmid concentrations to the ABI-quantified source at 1000 copies per reaction, so the CTs were within 0.5 of the quantified control. To determine the dynamic range, we performed a serial dilution of both plasmids at 1:10 dilutions from 1 x 109 to 1 copy of target per reaction.
| Results |
|---|
|
|
|---|
|
|
dynamic range
Universal primers 458F and 603R amplified HSV-1 and 2 with equal efficiency based on ABI-quantified controls and cloned plasmids pHSV-1 and pHSV-2 (Fig. 2A
). The range of detection for amplicon and unlabeled probe analysis was <10 copies to 1 x 109 copies of HSV per reaction (data not shown). A representative dilution experiment for low copy number detection is shown in Fig. 2B
. The unlabeled probe assay could reproducibly detect and type 10 copies of HSV per reaction. Lower copy numbers of target will not always amplify HSV but will exhibit IC amplification. The example in Fig. 2B
shows HSV amplification from the same dilution series, at an expected concentration of 1 copy per reaction for pHSV-1 and -2. One set exhibits amplification; the 2nd set did not amplify HSV but displays IC amplification at the expected CT. High-resolution melting analysis of HSV samples with positive amplification from Fig. 2B
for 10 and 1 copy are shown in Fig. 2C
. Samples with an estimated 10 copies of pHSV-1 or pHSV-2 exhibit the signature melting patterns expected for both amplicon and unlabeled probe melting analysis. HSV-positive samples of an estimated 1 copy show amplicon melting profiles but have weak or absent probe melting signals. The unlabeled probe method can reproducibly detect and type 10 or more copies of HSV per reaction. Amplicon melting was sufficient for the detection of HSV for some samples with <10 copies of target per reaction but was not accurate for typing because of the amplicon melting temperature shifts sometimes observed when very low copies of target were amplified (Fig. 2
, B and C).
|
detection of hsv in clinical samples
We blinded 110 archived clinical samples and tested them for the presence of HSV; 87 samples were from cerebrospinal fluid and 23 were from genital (6), serum/plasma (6), oral (3), and miscellaneous (8) sources. The unlabeled probe method detected 91 of 94 HSV-positive reference samples, 16 of 16 HSV-negative samples, and 3 false negatives. The 3 undetected positive samples had CT values that would calculate to <10 copies of HSV per reaction based on the reference assay. Given that some samples were >2 years old, and HSV could not be detected, it was not clear whether the samples degraded over time or the unlabeled probe technique simply did not detect HSV. The 3 negative samples were additionally processed with sequencing primers HSVFS and HSVRS, as well as with equimolar primers 458F and 603R without the unlabeled probe. In each case, the reactions failed to produce an amplicon, indicating that the samples had degraded (data not shown).
To further support the sensitivity of the unlabeled probe method, we performed an additional study. A 2nd set of 72 HSV reference samples <3 months old was assayed using the unlabeled probe method. Thirty-one of the clinical samples were from cerebrospinal fluid, and 41 were from genital (11), serum/plasma (10), oral (5), and miscellaneous (15) sources. Examination of this 2nd set showed 100% concordance with the reference assay. The unlabeled probe method detected 45 of 45 HSV-positive reference samples, 27 of 27 HSV-negative samples, and no false negatives. Of the 45 positive clinical samples, 4 had CT values indicating <10 copies of HSV per reaction, suggesting a sensitivity limited only by stochastic factors, similar to the reference assay.
typing of hsv in clinical samples
We sequenced 103 positive HSV clinical samples to verify HSV type. We included 36 HSV-1, 68 HSV-2, and 4 indeterminates based on glycoprotein D sequencing. Indeterminate samples were positive for HSV but, because of insufficient sample quantity, could not be accurately typed by unlabeled probes or sequenced. Representative melting profiles for 15 clinical samples melted on the LightCycler and on HR-1 are shown in Fig. 3, A and B
. These examples were chosen to highlight the capability of the unlabeled probe technique to detect sequence variations in the amplicon and under the unlabeled probe. Two samples, 1 HSV-1 and 1 HSV-2, displayed lower-temperature unlabeled probe melting than control HSV samples, whereas amplicon melting profiles were less affected. Sequence analysis confirmed that both samples had single-base sequence variations under the probe, a G
T at position 533 for HSV-1 and a C
T at position 551 for HSV-2. Of the 103 sequence-verified samples, 18 had at least 1 sequence variation in the amplicon, and 9 of these had a sequence variation under the probe. Of the 18 samples with sequence variations in the amplicon, only 1 sample had an amplicon melting profile that was altered by >0.5 °C (data not shown). Of the 9 samples with a sequence variation under the probe, all displayed aberrant probe melting profiles compared with controls. However, all samples were still accurately typed based on high-resolution probe melting and amplicon melting.
|
typing of samples without ic
Of the 103 HSV samples typed using unlabeled probes, 18 samples initially could not be accurately typed because of aberrant melting profiles or weak probe signals. The exclusion of the IC primers improved HSV amplification and detection and, for some samples, allowed for accurate typing by unlabeled probe melting. Nine indeterminate samples had aberrant unlabeled probe melting that was reproducible without the IC primers. These samples possessed sequence variations under the unlabeled probe as previously described. Of the remaining 9 indeterminate samples, 2 were HSV negative (CT >50), 2 were HSV-2, and 5 displayed amplicon melting but no unlabeled probe melting transition, presumably because of insufficient ssDNA amplification. These 5 were considered HSV+ (see Table 3
), and only 1 could be sequenced.
|
| Discussion |
|---|
|
|
|---|
Equimolar concentrations of primers 458F and 603R were examined to determine whether amplicon melting alone could detect and accurately type HSV. Equimolar primers could detect HSV with 100% concordance but were not consistently accurate for typing when sequence variations were present or at lower levels of target DNA. Very low levels of target amplicon may exhibit variable melting temperatures, outside the expected variation for a given amplicon (Fig. 2B
). Variability of amplicon melting temperatures at low copy numbers is not entirely understood and may be amplicon specific. Multiplexed HSV-1– or -2–specific primers were examined to determine whether type-specific amplification could accurately type HSV by amplicon melting alone. In our experience, 3' mismatches were not sufficient to specifically amplify HSV-1 or -2 by real-time PCR, and the issue was not pursued (data not shown). However, experimentation with different-length amplicons or a different locus may indicate that amplicon melting alone is sufficient for HSV typing using universal primers.
Inconclusive typing of HSV can occur when sequence polymorphisms are present under a fluorescence-labeled probe (22). The use of a dsDNA binding dye and unlabeled probe allows for a confirmatory amplicon melt, which permits HSV typing when sequence polymorphisms are present as long as reaction target copy numbers are >10. Below 10 copies per reaction, stochastic effects limit assay reproducibility, a problem common to many assays that can only be solved by increasing reaction volumes or by concentration of sample before amplification.
The range of HSV detection and typing using the unlabeled probe method was 10 to 1 x 109 copies per reaction. Ten copies of HSV per reaction may be detected, but not reproducibly typed, and concentrations of >1 x 109 copies per reaction were not tested. The universal 458F and 603R amplified HSV-1 and -2 with equal efficiency throughout the dynamic range, based on CT.
HR-1 melting analysis improves melting signal to noise and decreases baseline fluctuations compared with the LightCycler (19). For samples in which target copy number is >100 per reaction, HR-1 melting is advantageous to accurately type the target. When the target copy number is <10 copies per reaction, or when samples display aberrant melting profiles, a 2nd amplification reaction without IC primers may be performed to improve the HSV amplicon and unlabeled probe melting signal. This reflex reaction was performed 10% of the time to verify type or to confirm that the sample contained HSV. Aberrant melting curves were observed approximately 5% of the time, most often in positive samples. In comparison, the reference assay, which initially processes samples as single reactions, would have a repeat rate of 74% based on the number of positive clinical samples processed, since all positive samples with a CT >30 cycles are reanalyzed to confirm an HSV-positive sample (17).
The unlabeled probe method displayed 100% detection concordance when samples were not degraded. Typing positive samples was successful in 103 of 104 samples. The missed sample, sequence verified to be HSV-1, was detected by amplicon melting but could not be accurately typed owing to a lack of unlabeled probe signal. Another clinical sample reproducibly showed aberrant melting profiles that did not correlate to HSV-1 or -2. Gel analysis of the amplicon showed an unknown band at 50 bp. A no-DNA control is regularly included and produces no primer dimers based on melting and agarose gel analysis, suggesting the band was not primer dimer. Further analysis of the band was not pursued, and the sample was considered HSV negative by both the unlabeled probe and reference assay.
Unlabeled probes can detect low copy numbers of HSV from clinical samples with sensitivity similar to that of fluorescence-labeled, probe-based assays. The unlabeled probe method detected HSV from a variety of clinical samples with an overall 98% concordance to the reference assay and correctly typed 99% of HSV-1– or -2–positive samples, even when sequence variations were present. Typing is most accurate with high-resolution melting, although approximately 81% of HSV samples were genotyped by LightCycler melting analysis alone. Unlabeled probes are an attractive alternative to conventional fluorescence-labeled, probe-based assays. Genotyping, low-level detection of target, and ICs to monitor extraction, amplification, and detection can all be achieved with unlabeled probes.
| Acknowledgments |
|---|
Financial disclosures: Aspects of high-resolution melting analysis are licensed from the University of Utah to Idaho Technology. Dr. Carl T. Wittwer holds equity interest in Idaho Technology.
Acknowledgments: We thank Diana Mohl, Arthur Turlak, and Tyson Georgi for clinical samples and documentation; Jeff Stevenson and Westin Hymas for reagents and technical support; Michael Liew and Rebecca Margraf for technical advice; and Maria Erali for help with the manuscript. Aspects of high-resolution melting analysis are licensed from the University of Utah to Idaho Technology.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
K. De Leeneer, I. Coene, B. Poppe, A. De Paepe, and K. Claes Genotyping of Frequent BRCA1/2 SNPs with Unlabeled Probes: A Supplement to HRMCA Mutation Scanning, Allowing the Strong Reduction of Sequencing Burden J. Mol. Diagn., September 1, 2009; 11(5): 415 - 419. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Lyon and C. T. Wittwer LightCycler Technology in Molecular Diagnostics J. Mol. Diagn., March 1, 2009; 11(2): 93 - 101. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Zhou, R. J. Errigo, H. Lu, M. A. Poritz, M. T. Seipp, and C. T. Wittwer Snapback Primer Genotyping with Saturating DNA Dye and Melting Analysis Clin. Chem., October 1, 2008; 54(10): 1648 - 1656. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |