|
|
||||||||
Letters to the Editor |
1 Centre Hospitalier Universitaire Clermont-Ferrand Department of Immunology Hotel-Dieu Hospital Clermont-Ferrand, France
2 Université Clermont1 Faculty of Medicine-Pharmacy Clermont-Ferrand, France
3 Centre Hospitalier Universitaire Clermont-Ferrand Department of Rheumatology G Montpied Hospital Clermont-Ferrand, France
aAddress correspondence to this author at: Department of Immunology, Hotel-Dieu, Blvd. Leon Malfreyt, F-63058 Clermont-Ferrand, France. Fax 33-4-73-750-637; e-mail bevrard{at}chu-clermontferrand.fr.
To the Editor:
Measurement of immunoglobulin
and
free light chains (FLC) has improved the diagnostic evaluation and monitoring of monoclonal gammopathies (1), particularly light chain multiple myeloma(2), nonsecretory multiple myeloma, plasmacytoma, systemic amyloid light chain (AL) amyloidosis, heavy chain myeloma(3), and monoclonal gammopathy of undetermined significance(4). FLC measurement methods, however, are affected by analytical problems inherent to nephelometric techniques.
A 76-year-old woman was admitted for evaluation of bone pain and hypercalcemia. The pains were located in the left inguinal fold. Examination showed no lymphadenopathy or hepatosplenomegaly. X-rays of the skeleton showed 2 voluminous gaps in the pubic branches. Laboratory examinations showed hypercalcemia of 2.91 mmol/L and renal insufficiency. Complete blood cell count showed a macrocytic anemia with a hemoglobin concentration of 97 g/L, rouleaux formation, and circulating plasmacytes. Bone marrow contained 85% dystrophic plasmacytes. Serum total protein was 68 g/L. Protein electrophoresis showed hypogammaglobulinemia of 2.2 g/L [reference interval (RI) 7–11 g/L] with a moderate increase of the
-2 fraction (12.2 g/L, RI, 6–9 g/L). Immunofixation electrophoresis of blood and urine showed a major band of monoclonal
FLC migrating in the
-2 region. Serum FLC measurement showed a major increase in the
FLC (Table 1
). We arrived at a diagnosis of light chain
multiple myeloma.
|
In our laboratory, serum
and
FLC measurements are carried out with a nephelometric technique on the BNprospec automat (Dade Behring) using the Freelite reagents (The Binding Site) according to the manufacturers recommendations. A 1st FLC measurement is recommended on a sample diluted at 1:100. If the result is lower than the low end of the assay range, the machine automatically starts the assay again by diluting at 1:20 and then at 1:5. Likewise, if the concentration in FLC is higher than the high value of the range, the analyzer automatically starts serial dilutions and continues to a dilution of 1:32 000.
Initially, the
chains were <0.294 mg/L [RI, 3.3–19.4 mg/L (5)] and
chains were <0.405 mg/L [RI, 5.7–26.3 mg/L(5)]. In view of this unusual result (double negativity), a 1:32 000 dilution was made manually and the
-chain concentrations were then >60 100 mg/L. Measurement at a dilution of 1:2000 was then carried out for the
chains, and the result was <0.41 mg/L (Table 1
). Serum immunofixation confirmed the result, with a strong
monoclonal band. Thirty-five days later the
chains were 112 mg/L and
chains <0.405 mg/L. Given the patients history, however, a measurement at a dilution of 1:32 000 was made manually for
FLC, and the concentration obtained was 12 600 mg/L (Table 1
).
In this patient, we repeatedly observed that very high FLC concentrations yielded paradoxically low results. This situation is consistent with antigen excess. The "limitations" section of the product insert from the company does alert the user to the possibility of this phenomenon, but to our knowledge this report is the 1st to demonstrate that the problem of antigen excess in this assay is not merely theoretical but can actually occur clinically. The main risk is that a value within the RI, or one that is substantially underestimated, will be reported instead of the true value. To eradicate antigen excess, any suspect sample must be tested again at a higher dilution. Serum FLC measurements should always be interpreted together with other laboratory test results and clinical findings. This protocol requires a laboratory capable of performing serum protein electrophoresis, immunofixation, serum FLC measurement, and measurements of IgG, -A, and -M. Dissociation of these techniques among various laboratories can cause errors in interpretation.
In conclusion, we have illustrated that antigen excess can cause falsely low serum FLC results with nephelometric techniques. Clinically suspicious results should be repeated after sample dilution.
Acknowledgments
Grant/funding support: None declared.
Financial disclosures: None declared.
Footnotes
1 These authors contributed equally to this work. ![]()
References
The following articles in journals at HighWire Press have cited this article:
![]() |
A. Legg, J. A. R. Hobbs, G. P. Mead, A. R. Bradwell, S. Davern, and A. Solomon Monoclonal vs Polyclonal Free Light Chain Assays Am J Clin Pathol, June 1, 2009; 131(6): 901 - 902. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |