| Abstract |
|---|
|
|
|---|
Methods: After protein precipitation (500 µL methanol/zinc sulfate added to 500 µL plasma), samples were injected into the HPLC system and extracted online. The extracts were then back-flushed onto the analytical column and detected with an atmospheric pressure chemical ionization-triple quadrupole mass spectrometer monitoring the deprotonated molecular ions [M-H] of 15-F2t-IsoP (m/z = 353
193) and the internal standard 15-F2t-IsoP-d4 (m/z = 357
197).
Results: In human urine, the assay was linear from 0.025 to 80 µg/L and in human plasma from 0.0025 to 80 µg/L (r2>0.99). Interday accuracy and precision for concentrations above the lower limit of quantification were <10%. Concentrations of 15-F2t-IsoP in urine of 16 healthy individuals ranged from 55348 ng/g creatinine. In 16 plasma samples from healthy individuals, free 15-F2t-IsoP was detectable in all samples and concentrations were 325 ng/L.
Conclusions: Our assay meets all predefined method performance criteria, allows for analysis of >80 samples/day, and has sufficient sensitivity for quantifying 15-F2t-IsoP concentrations in plasma and urine from healthy individuals. It is, thus, suitable for clinical routine monitoring and the analysis of samples from larger clinical trials.
| Introduction |
|---|
|
|
|---|
(PGF2
), are considered a reliable index of in vivo oxidative stress (1). Unlike the enzymatically formed prostaglandins, F2-IsoPs are predominately formed by free-radicalcatalyzed peroxidation of arachidonic acid in situ in the phospholipid domain of cell membranes (2). After cleavage, presumably by phospholipases, F2-IsoPs circulate in free form in the plasma and are finally excreted in the urine (2)(3).
The mechanism of formation of F2-IsoPs includes several steps. Initial arachidonoyl radicals undergo endocyclization and are reduced to 4 F-ring regioisomers. Each regioisomer can form 8 racemic diastereomers leading to 64 theoretically possible different F2-IsoPs (4). In addition, endoperoxide intermediates can rearrange in vivo to form other types of isoprostanes (5)(6)(7) or highly reactive
-ketoaldehydes (8). Other classes of IsoPs can be formed in vivo by free radical catalyzed peroxidation of fatty acids such as eicosapentanoic or docosahexanoic acid (9)(10)(11).
In vitro several IsoPs including 15-F2t-IsoP (8-iso-PGF2
/iPF2
-III; for nomenclatures see Supplemental Data Table 1), the most extensively studied F2-IsoP isomer, possess biological activity such as nonspecific vasoconstriction (12)(13)(14)(15), bronchoconstriction (16)(17), and modulation of platelet function (13)(18)(19). It remains unresolved whether the effects observed in vitro at high concentrations are also relevant in vivo at physiological concentrations, which are considerably lower.
Since F2-IsoPs were first characterized in humans (20), increasing evidence has indicated that they provide a specific and sensitive assessment of lipid peroxidation (LPO) in vivo. Although F2-IsoPs are not a major product of LPO, their characteristics favor F2-IsoPs as specific markers of LPO. F2-IsoPs have been measured with gas chromatography/mass spectrometry (GC/MS) (20)(21)(22), GC-tandem MS (MS/MS) (23), liquid chromatography MS (LC/MS) (3)(24)(25)(26)(27), and immunoassays.
GC-MS assays require extensive sample preparation, including solid-phase extraction (SPE), thin layer chromatography (TLC), and derivatization reactions to protect the polar groups (28). Commercial immunoassays require extensive sample purification (SPE/TLC) and are susceptible to cross-reactivity because F2-IsoPs and their metabolites share a 1,3-syn-hydroxycyclopentane ring, which is the major determinant of antigenicity (28). Comparison of F2-IsoP measurement in urine by immunoassay and GC-MS has revealed considerable inconsistencies (29) attributable in part to the fact that the peak observed in many GC-MS assays after standard SPE and TLC sample purification is comprised of more than 1 F2-IsoP isomer (28).
All LC-MS methods published to date require SPE for sample purification (3)(24)(25)(26)(27). We report here the development and validation of an HPLC-MS/MS assay for the quantification of 15-F2t-IsoP with automated online extraction. Manual sample preparation is limited to a simple protein precipitation step.
| Materials and Methods |
|---|
|
|
|---|
/iPF2
-III), 15-F2t-IsoP-d4 (3,3,4,4-d4-15-F2t-IsoP, IS,
98% d4), 15(R)-F2t-IsoP, 9ß,11
-15-F2t-IsoP, prostaglandin F2
(PGF2
), 15(R)-PGF2
, 5-trans-PGF2
, 9
,11ß-PGF2, 9ß,11
-PGF2, and 5-trans-9ß,11
-PGF2 (Supplemental Data Fig. 1) were purchased from Cayman Chemical.
calibrators and quality control samples
Stock solutions (1 g/L) of all compounds were prepared in methanol and stored in polypropylene screw-top tubes at 80 °C. Working solutions for quality control and calibration were prepared by dilution of the stock solutions with methanol and were also stored in polypropylene tubes. Plasma and urine samples used for assay development and validation were obtained from healthy volunteers and pooled human plasma from the local blood bank. Blood samples were collected into EDTA tubes containing no other additives. Samples from healthy volunteers and patients were collected during various clinical trials. All protocols were approved by the local institutional review boards at both institutions. All study participants provided written informed consents and studies were conducted in full compliance with the principals of good clinical practice as set forth in the International Conference on Harmonization Harmonized Tripartite Guidelines (version April 1996), the US Code of Federal Regulations (21 CFR 50, 54, 56, and 312) and the principles stated in the Declaration of Helsinki (version 11, October 2000). The use of blood bank samples for assay validation and quality control was institutional review board exempt (Colorado Multiinstitutional Review Board).
sample preparation and protein precipitation
The protein precipitation solution (methanol/0.2 mol/L ZnSO4, 7:3, v/v) contained the internal standard (IS) 15-F2t-IsoP-d4 at 4 µg/L; 500 µL protein precipitation solution was added to an equal volume of plasma or urine. After vortex-mixing for 1 min and centrifugation (13 000g, 10 min, 4 °C), the supernatant was transferred into an HPLC vial and placed in the autosampler at 4 °C.
Creatinine concentrations in urine were measured with a Beckman Synchron LX® system using the Jaffe rate method.
automated online extraction and hplc conditions
The HPLC system consisted of 2 G1312A binary pumps, 2 G1322A vacuum degassers, a G1329A/G1330A thermostated autosampler and a G1316A thermostated column compartment (all Agilent 1100 series) with an integrated 6-port Rheodyne column switching valve, as shown in Supplemental Data Fig. 2. Sciex API4000 or API5000 triple quadrupole mass spectrometers were used as the detector. The HPLC system and the mass spectrometers were controlled by the Analyst software (Applied Biosystems).
We injected 500 µL of the samples onto a 4.6 x 12.5mm Eclipse XDB-C8 5-µm extraction column (Agilent Technologies) with a mobile phase of 30% methanol and 70% 0.1%-formic acid, flow rate 5 mL/min. After 1 min, the switching valve was activated and the analytes back-flushed from the extraction column onto a Phenomenex Synergi Hydro-RP 80Å, 3.0 x 250mm, 4-µm column filled with polar endcapped-C18 material. We used methanol and 0.1% formic acid, flow rate 0.6 mL/min, with the following gradient: 01 min 63% methanol; 1.110 min 63%98%. The analytical column was kept at 98% for 1 min and then reequilibrated to the starting conditions. After 9 min the column-switching valve was switched back into the extraction position and the extraction column reequilibrated to the starting conditions; the total run time between injections of 13 min. Both columns were maintained at 60 °C.
ms/ms analysis
The HPLC system was interfaced with the mass spectrometer with an atmospheric pressure chemical ionization (APCI) source. Nitrogen (purity: 99.999%) was used as collision-activated dissociation gas. The mass spectrometer was run in the negative multiple reaction monitoring (MRM) mode. The declustering potential was set to 70 V, the entrance potential (EP) to 5V, the interface to 400 °C, and the collision energy to 36 eV. The first quadrupole (Q1) was set to select the deprotonated molecular ions [M-H] of 15-F2t-IsoP (m/z= 353) and 15-F2t-IsoP-d4 (IS, m/z= 357), and the 3rd quadrupole (Q3) to select the characteristic product ions of 15-F2t-IsoP (m/z 193) and 15-F2t-IsoP-d4 (IS, m/z 197). Peak area ratios obtained from MRM mode of the mass transitions for 15-F2t-IsoP (m/z 353
193) and 15-F2t-IsoP-d4 (IS, m/z 357
197) were used for quantification.
calibration and quality control samples
Calibration and QC samples were prepared by enriching plasma or urine with 15-F2t-IsoP. To account for endogenous 15-F2t-IsoP, the ratio of endogenous 15-F2t-IsoP peak area divided by the IS peak area of unenriched matrix was subtracted from area ratios of enriched samples (corrected analyte area/IS area ratio). The calibration curves (0, 0.0005, 0.001, 0.0025, 0.005, 0.01, 0.025, 0.04, 0.1, 1, 10, 20, 40, 80 µg/L, n = 6 per concentration) were constructed with nonweighted linear regression. Concentrations in plasma were calculated as nanograms per liter, in urine as nanograms per gram creatinine. QC samples were prepared in human urine or plasma at concentrations of 0.04 µg/L, 0.1 µg/L, 20 µg/L, and 40 µg/L 15-F2t-IsoP.
matrix effects/ion suppression
To detect ionization efficiency changes attributable to coeluting matrix substances, we tested urine and plasma from 10 different healthy volunteers and aliquots of 2 different pooled human plasma samples. After protein precipitation, samples were extracted online and back-flushed onto the analytical column as described above. We infused 15-F2t-IsoP (10 mg/L dissolved in H2O/methanol, 7:3, v/v) postcolumn via a tee at 10 µL/min using a syringe pump. The extent of ion suppression was established by monitoring the intensity of the ion currents in MRM mode (m/z = 353
m/z = 193) at the retention times of analyte and IS (30).
validation procedures
The assay was completely validated according to the FDA Center for Drug Evaluation and Research guidelines for bioanalytical method validation (31).
Acceptance criteria.
The performance of the assay was considered acceptable if the precision (CV%) at each concentration was
15% for intraday and interday variability, the expected result was within 15% of the assigned concentration for both intra- and interday variability, and the calibration curve showed a correlation coefficient r2 of 0.99 or better.
Lower limit of quantification.
The lower limit of quantification (LLOQ) was determined as the lowest concentration of the calibration curve consistently yielding results within 20% of the nominal concentration and a precision
20%.
Precision and recovery.
The intraday precision, interday precision, and recovery were determined by analysis of QC samples containing 0.04 µg/L, 0.1 µg/L, 20 µg/L, and 40 µg/L 15-F2t-IsoP (n = 6/concentration). Samples were extracted and analyzed on 3 different days (n = 6/ concentration and day).
Extraction Recoveries.
The extraction recoveries were determined by comparing the signals for 15-F2t-IsoP obtained after extraction of QC samples (n = 6) with the signals of extracted matrix enriched with the respective concentrations of 15-F2t-IsoP after the extraction procedure.
Stability studies.
We tested stability in urine and plasma during 3 freeze-thaw cycles at 0.04, 0.1, 20, and 40 µg/L (n = 3). Samples were kept frozen at 80 °C and thawed at room temperature. Within-batch stability in fresh plasma was tested at 0.1 µg/L. Samples were kept at 80 °C, 20 °C, +4 °C, or at room temperature. After 4, 8, 12, 24, 48, and 168 h samples were extracted, analyzed, and compared with freshly prepared samples.
Dilution integrity.
Dilution integrity was established with freshly prepared urine and plasma samples enriched with 40 µg/L 15-F2t-IsoP. Dilutions (1:1, 1:10, 1:100, and 1:400, n = 6) were made with fresh unenriched urine or plasma. Deviations from the nominal concentrations after dilution were calculated.
Carryover effect.
Carryover was assessed by analyzing urine or plasma samples enriched with 15-F2t-IsoP at the upper limit of quantification (80 µg/L, n = 3) followed by blank methanol samples.
Cross validation.
Random spot urine samples obtained from 30 healthy volunteers were divided into aliquots, shipped on dry ice, and stored at 80 °C until analysis at the University of Colorado Health Sciences Center and at the University of Pennsylvania Medical Center. Investigators at both laboratories were blinded to the concentrations in the samples. Samples were analyzed in duplicate and quantified with an abbreviated standard curve (0, 0.001, 0.0025, 0.005, 0.01, 0.025, 0.04, 0.1, 0.25, 0.5, and 1 ng/L, n = 3/batch). Correlations were analyzed by linear regression (SPSS 14.0).
| Results |
|---|
|
|
|---|
m/z = 193 was selected for quantification. The most abundant product ion detected for 15-F2t-IsoP-d4 was at m/z = 197. Based on this the transition m/z = 357
m/z = 197 was selected for quantification of the IS.
|
No ion suppression was detected for most of the urine and for all the tested plasma samples (Figs. 2A
and 2B
). However, 2 of 10 urine samples showed a loss of signal intensity at the retention times of 15-F2t-IsoP and IS by a factor of
3 (Fig. 2C
).
|
15-F2t-IsoP eluted with a mean (SD) retention time of 8.30 (0.12) min and the IS with a retention time of 8.27 (0.13) min.
In urine, the LLOQ was 0.025 µg/L (with API5000) and linearity was up to 80 µg/L (0.231x + 0.011, r2 = 0.9971). With the API4000 detector, the assay was linear from 0.04 (LLOQ) to 80 µg/L (y = 0.204x0.098, r2 = 0.9995, Fig. 3A
). In plasma, the LLOQ was 0.0025 µg/L (with API5000) and linearity up to 80 µg/L (y = 0.976x 0.002, r2 = 0.9912). With the API4000, the linear range was 0.01 (LLOQ) to 80 µg/L (y = 1.180x 0.075, r2 = 0.9996, Fig. 3B
). At the tested concentrations of 0.04, 0.1, 20, and 40 µg/L, intraday recoveries in urine and plasma were 93.8%106.4% and intraday precisions
6.8% (Table 1A
). Interday recoveries were 91.1%106.2%, and interday precisions
8% (Table 1B
).
|
|
The mean (SD) absolute recovery of 15-F2t-IsoP after protein precipitation of urine and plasma was 96.7% (8.1%) and 100.2% (9.5%), respectively.
15-F2t-IsoP was stable in urine and plasma for at least 3 freeze-thaw cycles (Figs. 4A
and 6B). Samples were considered stable if they contained analyte concentrations not statistically different from the corresponding baseline samples as tested by ANOVA in combinations with Dunnetts post hoc test. In addition, the mean concentrations had to be within 15% of baseline. In fresh human plasma, the concentration of 15-F2t-IsoP was stable for at least 168 h when stored at +4 °C or below. At room temperature, the concentration of 15-F2t-IsoP was stable for at least 48 h; after 168 h the measured concentration increased to 189.2% (26.6%) of the nominal concentration, mainly because of to autooxidation of arachidonic acid resulting in in vitro formation of 15-F2t-IsoP. To confirm stability of endogenous 15-F2t-IsoP in nonenriched samples from healthy individuals, we tested 6 plasma and 6 urine samples, stored for 24 h at 4 °C or room temperature, and during 3 freeze-thaw cycles. The 15-F2t-IsoP concentrations in plasma relative to immediately analyzed controls were (all n = 6): after 24 h at 4 °C, 95.0% (1.9%); after 24 h at room temperature, 89.4% (8.0%); after 1 freeze-thaw cycle, 95.3% (4.6%); after 2 freeze-thaw cycles, 96.8% (4.0%); and after 3 freeze-thaw cycles, 106.2% (4.0%). The 15-F2t-IsoP concentrations in urine relative to immediately analyzed controls were (all n = 6): after 24 h at 4 °C, 102.3% (3.4%); after 24 h at room temperature, 102.6% (3.7%); after 1 freeze-thaw cycle, 95.9% (2.3%); after 2 freeze-thaw cycles, 97.0% (4.5%); and after 3 freeze-thaw cycles, (117.7%) (2.1%). The results showed that except after 3 freeze-thaw cycles in urine, 15-F2t-IsoP was stable in native plasma and urine samples at the low concentrations found in healthy persons.
|
Recovery was not affected by dilution of urine or plasma samples. The mean recoveries for dilutions of 1:1, 1:10, 1:100, and 1:400 of samples enriched with 40 µg/L in urine and plasma were between 100.7%99.9% (urine) and 100.2%104.1% (plasma).
Carryover was not observed after analysis of samples with an 15-F2t-IsoP concentration of 80 µg/L Analysis of urine and plasma samples from healthy subjects showed the endogenous 15-F2t-IsoP peak clearly separated from adjacent peaks (Figs. 5A
and 5B
). Neighboring peaks were identified by enriching urine or plasma from healthy persons with commercially available isomers of 15-F2t-IsoP and PGF2
. All of the tested isomers showed the ion transition m/z = 353
193.
|
With the exception of the 2 9ß-isomers of 15-F2t-IsoP and PGF2
that do not occur naturally (9ß,11
-15-F2t-IsoP and 9ß,11
-PGF2), all isomers were chromatographically separated from 15-F2t-IsoP (Figs. 5C
and 5D
).
Cross-validation of urine samples from 30 healthy volunteers showed good correlation of results obtained by the 2 laboratories (r2 = 0.95, P <0.001, Supplemental Data Fig. 3).
Concentrations of 15-F2t-IsoP in urine of healthy persons (n = 16) ranged from 55 to 348 ng/g creatinine. In plasma from the same healthy persons, free 15-F2t-IsoP was detectable in all the samples (Fig. 5B
), with concentrations ranging between 325 ng/L (n = 16).
| Discussion |
|---|
|
|
|---|
Increased use of MS detection has led to a revival of the use of column-switching techniques for automated online sample preparation, used for many years for HPLC analyses (33). To date, SPE has been used for sample preparation in all published LC-MS methods for the analysis of 15-F2t-IsoP (3)(23)(24)(25)(26)(27). Only one method employed automated SPE, but the extraction was carried out offline with an automated SPE workstation (27).
Matrix effects should be assessed during development and validation of an analytical method (34). Among the atmospheric pressure ionization techniques, electrospray ionization (ESI) has been reported to be particularly susceptible to matrix effects (35)(36). When we evaluated ESI during method development, all tested urine samples caused severe ion suppression, resulting in a drop of signal intensity by an order of magnitude around the retention of our analytes of interest (Fig. 2D
). After the ionization mode was switched to APCI, ion suppression was not detected in any of the plasma samples and was detected in only 2 of 10 tested urine samples. These findings are in line with observations by King et al. (35), who reported that plasma samples prepared by acetonitrile protein precipitation caused much more severe ion suppression in ESI than APCI. Interestingly, for 15-F2t-IsoP, ion suppression was observed for urine but not plasma, independent of the ionization mode. The fact that matrix effects can alter ionization efficiency in both ESI and APCI highlights the importance of the use of a stable isotope-labeled analyte as IS (37).
All previous LC-MS methods for 15-F2t-IsoP used negative ion mode ESI, but none were tested for matrix effects (3)(24)(25)(26)(27). Even if a stable isotope-labeled analyte is used as the IS (37), the loss of signal intensity will negatively affect the LLOQ. For compounds such as 15-F2t-IsoP, which normally occur at very low concentrations, and urine samples that may be subject to varying degrees of dilution, the LLOQ may not be low enough to cover the lower concentration range of the analyte.
Method development was mainly carried out using an API4000 mass spectrometer. During our study, an API5000 mass spectrometer became available. Compared with the API4000, the API5000 mass spectrometer improved the LLOQ by a factor of 1.6 in urine and 4.0 in plasma. This improvement is critical for plasma, for which the concentrations in healthy persons were 325 ng/L. At these concentrations the API4000 mass spectrometer was not sufficiently sensitive.
For IsoPs, there are 4 regioisomers, each theoretically comprised of 16 isomers (4). The regioisomers differ by the length of the alkyl side chains and/or by the position of the 3rd hydroxyl group on the side chain and can, therefore, be distinguished by monitoring typical daughter ions generated in MRM mode. Within a given group of regioisomers (e.g., 15-F2-IsoP), however, all the isomers show the same fragmentation pattern and thus must be separated chromatographically to allow for specific quantification. Only 3 of the 16 theoretically possible isomers in the 15-F2-IsoP group are commercially available; 15-F2t-IsoP, 15(R)-F2t-IsoP, and 9ß, 11
-15-F2t-IsoP were tested together with PGF2
and 5 PGF2 isomers for chromatographic separation (Fig. 5
). The only 2 compounds that coeluted with 15-F2t-IsoP were the 2 9ß-isomers of 15-F2t-IsoP and PGF2
. 9ß, 11
-compounds cannot occur naturally (5)(6)(7)(38)(39). However, theoretical considerations suggest that half of the 15-F2t-IsoP formed will be 9ß, 11ß and half will be 9
, 11
.
It must be noted that our study was not designed to determine reference intervals. Our intention was to demonstrate that the 15-F2t-IsoP concentrations found in urine and plasma of healthy persons were within the range of the reliable response of our assay.
It is not surprising, however, that in this situation in which multiple isomers must be separated chromatographically for reliable quantification, the normal values for 15-F2-IsoP in biological samples reported in the literature differ widely. Reported mean (SD) normal values measured by LC-MS/MS in urine of healthy volunteers range from 250 (220) ng/g creatinine (27) to 1110 (450) ng/g creatinine (3). The method by Liang et al. (27), which reported the lowest normal values, also achieved the best chromatographic separation of 4 isomers [15(R)-F2t-IsoP, 15-F2t-IsoP, 15(R)-PGF2
, PGF2
] and one unknown peak. The above-mentioned isomers were separated in the same sequence by our method; however, it allowed for additional separation of 9
, 11ß-PGF2, which eluted between the 15-F2t-IsoP and 15(R)-PGF2
peaks. The naturally occurring PGF2 isomer can be identified in urine and plasma (Fig. 5, C and D
, peak 3). Separation of 9
, 11ß-PGF2 is crucial because it elutes shortly after the 15-F2t-IsoP peak, but is not an isoprostane. The superior chromatographic separation achieved by our method most likely explains why our measured values of 15-F2t-IsoP in urine of healthy volunteers are lower than the values reported in the literature for comparable analytical methodologies. The selection of the analytical column during method development proved to be critical to achieve the desired chromatographic separation.
| Acknowledgments |
|---|
| Footnotes |
|---|
, prostaglandin F2
; LPO, lipid peroxidation; SPE, solid-phase extraction; TLC, layer chromatography; IS, internal standard; LLOQ, Lower limit of quantification; ESI, electrospray ionization; and APCI, atmospheric pressure chemical ionization. | References |
|---|
|
|
|---|
, in the rat: evidence for interaction with thromboxane-A2 receptors. J Clin Invest 1992;90:136-141.[ISI][Medline]
[Order article via Infotrieve]
, a potent agonist of the vascular thromboxane endoperoxide receptor, is a platelet thromboxane endoperoxide receptor antagonist. Prostaglandins 1992;44:155-163.[CrossRef][ISI][Medline]
[Order article via Infotrieve]
on retinal vessels. Am J Physiol-Reg I 1998;43:R1406-R1416.
)-III and iPE(2)-III are mediated via the thromboxane A(2) receptor in vivo. Circulation 2000;101:2833-2840.
in isolated perfused rat lung. J Appl Physiol 1993;74:460-465.
, in rabbit lung in situ. Am J Physiol 1992;263:H660-H663.[ISI][Medline]
[Order article via Infotrieve]
is not mediated by thromboxane receptor isoforms. J Biol Chem 1996;271:14916-14924.
increases platelet adhesion and reduces the antiadhesive and antiaggregatory effects of NO. Arterioscl Throm Vas 1998;18:1248-1256.
) levels are elevated in individuals with non-insulin-dependent diabetes mellitus. Febs Letters 1995;368:225-229.[CrossRef][ISI][Medline]
[Order article via Infotrieve]
in human plasma and urine by liquid chromatography-electrospray ionization mass spectrometry. J Chromatogr B 2000;746:17-24.[CrossRef]
using liquid chromatography-tandem mass spectrometry: increased excretion in diabetics. Prostaglandins Other Lipid Mediat 2000;62:173-181.[CrossRef][ISI][Medline]
[Order article via Infotrieve]
and 2,3-dinor-8-isoprostaglandin-F-2
in human urine using liquid chromatography-tandem mass spectrometry. Free Radical Bio Med 2003;34:409-418.[CrossRef]
-I: An index of lipid peroxidation in humans. Proc Natl Acad Sci U S A 1998;95:3449-3454.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||