Evidence-Based Laboratory Medicine and Test Utilization |
1 Hennepin County Medical Center, Minneapolis, MN.
2 Medical College of Virginia, Richmond, VA.
3 Duke University Medical Center, Durham, NC.
4 University of California at San Francisco, San Francisco, CA.
5 University of Maryland School of Medicine, Baltimore, MD.
NACB COMMITTEE MEMBERS,
Baltimore, MD
Minneapolis, MN
Boston, MA
Cleveland, OH
Richmond, VA
Boston, MA
Durham, NC
Aarhus, Denmark
Nashville, TN
Cleveland, OH
San Francisco, CA
IFCC COMMITTEE ON STANDARDIZATION OF MARKERS OF CARDIAC DAMAGE (C-SMCD) MEMBERS,
Minneapolis, MN
Baltimore, MD
Rochester, MN
Innsbruck, Austria
Barcelona, Spain
Brecia, Italy
Milan, Italy
Brisbane, Australia
San Francisco, CA
aAddress correspondence to this author at: Director, Rapid Response Laboratories, University of Maryland School of Medicine, 22 S. Greene St., Baltimore, MD 21201. Fax 410-328-5880; e-mail rchristenson{at}umm.edu.
| Introduction |
|---|
|
|
|---|
| I. Overview of Analytical Issues for Acute Coronary Syndrome (ACS) Biomarkers |
|---|
|
|
|---|
| II. Analytical Biomarker Issues |
|---|
|
|
|---|
10% at the 99th percentile reference limit. Before introduction into clinical practice, cardiac biomarker assays must be characterized with respect to potential interferences, including rheumatoid factors, human anti-mouse antibodies, and heterophile antibodies. Preanalytical and analytical assay characteristics should include biomarker stability (over time and across temperature ranges) for each acceptable specimen type used in clinical practice and identification of antibody/epitope recognition sites for each biomarker. Analytical and preanalytical specifications developed by professional groups such as the IFCC should be followed (Level of Evidence: C).
a. cTn specifications
First, in the context of cTn, the epitopes recognized by the antibodies must be delineated. Epitopes located on the stable part of the cTnI molecule should be a priority. Specific relative responses need to be described for the following cTnI forms: free cTnI, the I-C binary complex, the T-I-C ternary complex, and oxidized, reduced, and phosphorylated isoforms of the 3 cTnI forms. The effects of different anticoagulants on binding of cTnI also need to be addressed. Second, the source of material used to calibrate cTn assays, specifically for cTnI, should be reported. A cTnI standardization subcommittee of the AACC in collaboration with the NIST has developed a primary reference material (SRM #2921) (3). Although this material demonstrated commutability with only 50% of current cTn assays, it will be of use in harmonizing cTnI concentrations across different assays (4)(5). At present, it appears that the only way to achieve complete standardization for cTnI would be for all manufacturers to agree on using the same antibody pairs for all commercial assays as well as a common reference material for calibration (6)(7). The IFCC C-SMCD is currently exploring the development of a serum-based secondary reference material. For cTnT, as there is only one assay manufacturer, harmonizing between assay generations has been consistent. Third, manufacturers need to use methods advocated by the CLSI to characterize detection limit, functional sensitivity, and total imprecision (8)(9). Key characteristics for cTn assays include determination of the distribution of values in a healthy reference population, the statistical determination of the 99th percentile cutoff for the reference population, and determination of the concentration corresponding to the 10% CV (total imprecision). Preanalytical factors that should be described include effect of storage time and temperature, effect of glass vs plastic tubes and gel separator tubes, and the influence of anticoagulants for plasma and whole-blood measurements. As more assay systems are devised for point-of-care (POC) testing, identical analytical criteria must apply to both central laboratory methodologies and POC testing systems. When measuring cTn by different methodologies within the same institution, assay results should be harmonized or a strategy implemented to avoid interpretative confusion by clinicians.
b. cardiac biomarker turnaround
Clinicians and laboratorians continue to support a goal for turnaround times (TATs) <60 min for cardiac biomarkers, but the largest study published to date has demonstrated that TAT expectations are not being met in a large proportion of hospitals (10). A College of American Pathologists Q-probe survey study of 7020 cTn and 4368 CK-MB determinations in 159 predominantly North American hospitals demonstrated that the median and 90th percentile TATs for troponin were 74.5 and 129 min, and for CK-MB, 82 and 131 min. In fact, fewer than 25% of hospitals were able to meet the <60-min TAT, defined as order-to-report time. A separate subanalysis of just POC testing systems was not reported. Recently published data have shown that implementation of POC cTn testing can decrease TATs to <30 min in cardiology critical-care and short-stay units (11). These data highlight the continued need for laboratory services and healthcare providers to work together to develop better processes to meet a <60-min TAT as requested by physicians.
c. biomarkers no longer recommended for use in the evaluation of acs
Use of aspartate aminotransaminase, total lactate dehydrogenase, and lactate dehydrogenase isoenzymes are not recommended for evaluation of cardiac injury and detection of myocardial infarction (MI). The use of total CK or CK-MB activity is an acceptable alternative for evaluating cardiac injury in institutions where cTn or CK-MB mass assays are not available or feasible. Total CK can also assist in improving myocardial tissue specificity when the ratio of CK-MB to total CK is greater than previously established reference intervals. This concept is emphasized in a statement from the American Heart Association (AHA) Council on Epidemiology and Prevention regarding case definitions for acute coronary heart disease in epidemiology and clinical research studies (12). The following recommendations were made to allow for a more accurate interpretation of recent trends in ACS during implementation of cTn assays and use of the European Society of Cardiology (ESC)/American College of Cardiology (ACC) consensus MI definition (13)(14) predicated on cTn: (a) simultaneous use of traditional biomarkers with cTn to determine the performance of new biomarkers; and (b) use of adjustment factors in databases and retrospective studies seeking to determine incidence and trends of MI before and after cTnderived studies.
d. determining biomarker decision cutoff characteristics for acs
The 99th percentile of a reference decision-limit (medical decision cutoff) for cTn assays should be determined in each local laboratory by internal studies using the specific assay that is used in clinical practice or validating a reference interval that is based on findings in the literature (13)(16). Desirable imprecision (expressed as %CV) of each cTn assay (and CK-MB mass assay) has been defined as
10% CV at the 99th percentile reference limit (13)(16). Unfortunately, the majority of laboratories have neither the resources to perform adequately powered 99th percentile reference studies nor the ability to carry out CLSI protocols to establish total imprecision criteria for the cTn assay that they plan to use in practice (17). Therefore, clinical laboratories must rely on the peer-reviewed published literature to assist in establishing both local reference limits and imprecision characteristics. Caution must be taken when comparing the findings reported in the manufacturers package inserts, which have been cleared by the FDA, with the findings reported in journals because of differences in total sample size, distributions by sex and ethnicity, age ranges, and statistical methods used to calculate the 99th percentile reported.
To date, very few in vitro diagnostic companies have published 99th percentile limits in their package inserts. There is no established guideline set by the FDA or other regulatory agencies to mandate a consistent evaluation of the 99th percentile reference limit for cTns. The largest and most diverse reported reference value study to date shows plasma (heparin) 99th percentile reference limits for 8 cTn assays (7 cTnI and 1 cTnT) and 7 CK-MB mass assays (18). This study was performed in 696 healthy adults (ages 1884 years) stratified by sex and ethnicity. There was a 13-fold difference between the lowest vs the highest measured cTnI 99th percentile limit. The lack of cTn assay standardization (there is no primary reference material that is commutable with all commercial methods, as noted earlier) and the differences in antibody epitope recognition between assays (different assays use different antibodies, as noted earlier) give rise to substantially discrepant concentrations. What is generally recognized, though, is that as long as one understands the characteristics of an individual assay and does not attempt to compare absolute concentrations between different assays, clinical interpretation should be acceptable for all assays.
For CK-MB (as has been recognized for years for total CK), all assays demonstrate a significant 2- to 3-fold higher 99th percentile limit for men vs women (18). Further, CK-MB can demonstrate up to 2- to 3-fold higher concentrations for African Americans vs Caucasiansdifferences attributed to between-race physiological differences in muscle mass. These data led to the class I recommendation that clinical laboratories should establish different CK-MB reference limits based on sex. Labs should also consider doing so for ethnic groups.
For cTn, expert consensus has emerged in support of the 99th percentile as the reference cutoff, in spite of whether the total imprecision of the assay is
10%. This has been supported by a recent study that has demonstrated that misclassification of patients who are ruled out using cTn assays with variable imprecision (10%25%) at the 99th percentile does not lead to significant patient misclassification over serial cTn orders (19). Further, whereas the literature has been enriched with studies appropriately addressing the total imprecision of cTn assays, as to what the lowest concentration will be to attain a 10% CV, the manufacturers package inserts often publish imprecision data primarily based on within-run or between-day precision. Again, there is no consistent regulatory specification regarding precision data that should be reported in the manufacturers package inserts. To better address day-to-day clinical laboratory practice, early findings from an IFCC C-SMCD study demonstrated that the total imprecision for 13 commercial assays [based on a 20-day CLSI protocol (20)] was unable to experimentally achieve a 10% CV at their 99th percentile limit. Improved 2nd-generation assays, however, have recently demonstrated 10% CVs at the 99th percentile (21). The ultimate goal will be to have all cTn assays attain a 10% CV at the 99th percentile reference limit to reduce any potential of false-positive analytic results attributable to imprecision in the low concentration range.
For clinical trials, to avoid the confusion of multiple centers using multiple assays, several approaches are recommended for cTn testing (15)(16). First, analyze all samples from trial centers in a core, central lab with a precise, well-defined assay. Second, provide all trial centers with the same well-defined, FDA-cleared assay. Third, uniformly define each centers assays by using the 99th percentile concentration (assay-dependent), thus reducing reliance on local laboratory criteria for cTn decision cutoffs. Fourth, use a multiple (2- to 3-fold) of the 99th percentile. Fifth, if trials decide to use cutoff values defined in earlier studies, the degree of imprecision at these concentrations should be reported.
e. european society of cardiology/american college of cardiology recommendations
An ESC/ACC consensus document along with the AHA/ACC guidelines for differentiating AMI and unstable angina codified the role of cTn by advocating that the diagnosis of AMI be based on increases of cTnI or T (preferred) or CK-MB mass above the 99th percentile cutoff in the appropriate clinical situation (14)(22). The guidelines recognized the reality that neither the clinical presentation nor the electrocardiogram had adequate clinical sensitivity and specificity. The guidelines do not suggest that all increases of these biomarkers should elicit a diagnosis of MI or high-risk profile, only those associated with the appropriate clinical, electrocardiogram, imaging, or pathological findings. When cTn increases are not due to acute ischemia, the clinician is obligated to search for another etiology for the elevation (6)(23). Updated guidelines addressing the revised universal definition of MI cosponsored by the Joint ESC-ACC-AHA-World Heart Federation (WHF) Task Force For The Redefinition of MI will soon be published. This document will support and coincide with the recommendations proposed in the current joint NACB IFCC document.
| Acknowledgments |
|---|
Financial Disclosures: The National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines Committee for Utilization of Biomarkers in Acute Coronary Syndromes and Heart Failure reports all reported relationships within the 2 years previous to this publication that may be relevant to this guidelines document. A document of those relationships may be found in the online Data Supplement at http://www.clinchem.org/content/vol53/issue4 .
| Footnotes |
|---|
The materials in this publication represent the opinions of the authors and committee members, and do not necessarily represent the official position of the National Academy of Clinical Biochemistry (NACB) or the International Federation of Clinical Chemistry (IFCC). The National Academy of Clinical Biochemistry is the academy of the American Association for Clinical Chemistry.
This article has been copublished in the April 3, 2007 online issue of Circulation.
© 2007 American Association for Clinical Chemistry and the American Heart Association, Inc.
1 Nonstandard abbreviations: NACB, National Academy of Clinical Biochemistry; ACS, acute coronary syndrome; cTn, cardiac troponin; FDA, US Food and Drug Administration; AMI, acute myocardial infarction; C-SMCD, Committee on Standardization of Markers of Cardiac Damage; CK-MB, creatine kinase MB; cTnI, cardiac troponin I; cTnT, cardiac troponin T; CLSI, Clinical Laboratory Standards Institute; POC, point-of-care; TAT, turnaround time; MI, myocardial infarction; AHA, American Heart Association; ESC, European Society of Cardiology; ACC, American College of Cardiology; and WHF, World Heart Federation. ![]()
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
J. R Tate, W. Ferguson, R. Bais, K. Kostner, T. Marwick, and A. Carter The determination of the 99th centile level for troponin assays in an Australian reference population Ann Clin Biochem, May 1, 2008; 45(3): 275 - 288. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Thygesen, J. S. Alpert, A. S. Jaffe, H. D. White, and On behalf of the Joint ESC/ACCF/AHA/WHF Task Force The universal definition of myocardial infarction: some issues and concerns: reply Eur. Heart J., May 1, 2008; 29(9): 1209 - 1210. [Full Text] [PDF] |
||||
![]() |
F. S. Apple, S. W. Smith, L. A. Pearce, R. Ler, and M. M. Murakami Use of the Centaur TnI-Ultra Assay for Detection of Myocardial Infarction and Adverse Events in Patients Presenting With Symptoms Suggestive of Acute Coronary Syndrome Clin. Chem., April 1, 2008; 54(4): 723 - 728. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Collinson, K. Pulkki, J. Suvisaari, J. Ravkilde, A. Stavljenic-Rukavina, A. Hammerer-Lercher, H. Baum, M. P. van Dieijen-Visser, and P. H. Laitinen How Well Do Laboratories Follow Guidelines on Cardiac Markers? The Cardiac Marker Guideline Uptake in Europe Study Clin. Chem., February 1, 2008; 54(2): 448 - 449. [Full Text] [PDF] |
||||
![]() |
S. E.F. Melanson, M. J. Tanasijevic, and P. Jarolim Cardiac Troponin Assays: A View From the Clinical Chemistry Laboratory Circulation, October 30, 2007; 116(18): e501 - e504. [Full Text] [PDF] |
||||
![]() |
F. S. Apple and M. M. Murakami Serum and Plasma Cardiac Troponin I 99th Percentile Reference Values for 3 2nd-Generation Assays Clin. Chem., August 1, 2007; 53(8): 1558 - 1560. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||