| Abstract |
|---|
|
|
|---|
Methods: To design assays for immunodetection of proBNP, NT-proBNP, and BNP, we used a panel of BNP- and NT-proBNPspecific monoclonal antibodies (MAbs). All MAbs were tested in 2-site combinations in time-resolved fluoroimmunoassays with recombinant or synthetic antigens and plasma from heart failure (HF) patients. ProBNP and related molecules were assayed in HF plasma samples and plasma extracts by means of gel filtration fast protein liquid chromatography (FPLC) before and after protein fractionation on Sep-Pak C18 cartridges.
Results: The limits of detection for BNP, proBNP, and NT-proBNP assays were 0.4, 3, and 10 ng/L, respectively. Gel filtration-FPLC studies revealed 1 peak of NT-proBNP (
25 kDa), 1 peak of proBNP (
37 kDa), and 2 peaks of BNP immunoreactivity, a major peak (
37 kDa) for proBNP, and a minor peak (
4 kDa) for BNP. In patient plasma, the molar concentration of NT-proBNP was almost 10 times that of proBNP. The mean proBNP:BNP ratio in patient plasma was 6.3, ranging from 1.8 to 10.8.
Conclusions: ProBNP is the major BNP-immunoreactive form in human blood. The proBNP:BNP ratio in plasma samples is dependent on the methods used for sample handling and for the measurement of the peptides.
| Introduction |
|---|
|
|
|---|
2535 kDa) (1). Similar studies have detected BNP immunoreactivity in 2 fractions, 1 with a molecular mass characteristic of BNP-32 peptide (
3 kDa) and the other with high-molecular-mass, most likely the BNP precursor (proBNP) molecule (
3040 kDa) (2)(3) with posttranslational modifications (glycosylation) (4). No conclusive evidence exists, however, to elucidate the ratios of the immunoreactive forms of BNP and proBNP in the blood. We sought to better define the molecular forms of proBNP and its derivatives in blood samples of HF patients by use of a GF-fast protein liquid chromatography (FPLC) separation technique. | Materials and Methods |
|---|
|
|
|---|
preparation of monoclonal antibodies
The hybridoma cell lines producing monoclonal antibodies (MAbs) were obtained by use of a standard procedure (5). To obtain NT-proBNP-specific antibodies, mice were immunized with synthetic peptides corresponding to sequences 124, 1327, 2845, 4660, and 6176 of NT-proBNP molecules conjugated to the carrier protein. For BNP MAb production, both synthetic peptides corresponding to sequences 110, 1122, 1723, and 2632 of BNP and the whole BNP molecule conjugated to the carrier protein were used as immunogens. Humane treatment of mice was in accordance with the Committee on Care and Use of Laboratory Animals.
sandwich immunofluorescent assay
To perform the sandwich immunofluorescent assay (IFA), we used biotinylated MAbs as capture antibodies and MAbs labeled with stable europium (III) chelate of 2,2',2'',2''';-[[4-[(4-isothiocyanatophenyl)ethynyl]pyridine- 2,6-diyl]bis(methylenenitrilo)]tetrakis(acetic acid) as detection antibodies. The protocols of antibody labeling with biotin-isothiocyanate and europium chelate have been described by Katrukha et al. (6). Mixtures of equal quantities (200 ng per well) of biotinylated and Eu3+-conjugated antibodies in 50 µL of buffer A (0.05 mol/L Tris-HCl, pH 7.7, 9 g/L NaCl, 0.1 mL/L Tween 40, 5 g/L bovine serum albumin, and 0.5 g/L NaN3) were incubated in streptavidin-coated plates (Wallac-Perkin-Elmer) with 50 µL of tested sample or the calibrator for 30 min at room temperature with gentle shaking. After washing with buffer B (0.01 mol/L Tris-HCl, pH 7.8, 0.15 mol/L NaCl, 0.25 mL/L Tween 20, and 0.5 g/L NaN3) we added 200 µL per well of enhancement solution (1.75 mol/L NaSCN, 1 mol/L NaCl, 50 mL/L glycerol, 200 mL/L 1-propanol, 0.005 mol/L Na2CO3, 0.05 mol/L glycine-NaOH, pH 10.0). The plate was incubated at room temperature for 3 min with gentle shaking, and fluorescence was measured on a Victor 1420 multilabel counter (Wallac-Perkin-Elmer). Total time of the assay was 40 min.
beckman access bnp assay
The Access BNP assay uses a capture MAb specific to the ring structure, with part of the C-terminal fragment of BNP and detection MAb specific to the N-terminal part of the BNP molecule, and enables the detection of both BNP and proBNP (7).
patients and blood samples
Diagnosis of HF was based on patient symptoms of dyspnea, orthopnea, lung rales, or leg edema and confirmed by echocardiography studies (GE Vivid 7 ultrasound imaging system) and x-ray examination. The preliminary diagnosis was made by a cardiologist and further confirmed by an HF expert. Blood samples were collected from patients with left ventricular ejection fractions (measured by modified Simpsons method) <30% [mean (SD) 23.4%, (5.2%)] and left-ventricular end-systolic volume >90 mL. Venous blood was collected into EDTA-containing Vacuette tubes (Greiner Bio-One) and centrifuged at 3000g (15 min, +4 °C). Serum samples were obtained from blood collected in plastic tubes, incubated for 30 min at room temperature, and centrifuged at 5000g (30 min, +20 °C). Plasma and serum samples were stored at 70 °C before use. For MAb testing, pooled serum (39 patients with severe HF) or pooled plasma (10 HF patients) was used as a source of endogenous antigens. As a negative (non-HF) control we used pooled serum or plasma from 10 healthy donors. Pooled plasma of healthy donors was also used for dilution of recombinant NT-proBNP, proBNP, and synthetic BNP before applying on GF column and for assay validation. All studies using human blood samples were in accordance with the current revision of the Helsinki Declaration.
protein extraction from plasma samples
BNP and proBNP were extracted from plasma samples of HF patients by use of Sep-Pak C18 cartridges (Millipore-Waters) according to the method described by Shimizu et al. (2). The eluates were dried for 23 h with a reduced pressure evaporator Mini DNA (Heto Lab Equipment).
gf-fplc
Individual patient plasma samples or Sep-Pak C18 extracts, reconstituted in 0.1 mol/L sodium phosphate buffer, pH 7.4, containing 0.3 mol/L NaCl and 0.005 mol/L EDTA, were centrifuged at 10 000g for 10 min, and supernatants (150 µL) were applied onto Superdex 75 10/300 GL GF column (GE Healthcare) equilibrated with the same buffer. Proteins were eluted at a flow rate of 0.8 mL/min. The column was calibrated with a set of standard proteins (GE Healthcare): albumin (67 kDa), ovalbumin (43 kDa), chymotrypsinogen (25 kDa), ribonuclease A (13.7 kDa), and aprotinin (6.52 kDa). Recombinant NT-proBNP, proBNP, and synthetic BNP were reconstituted in pooled plasma from healthy donors before being loaded onto the Superdex 75 column.
| Results |
|---|
|
|
|---|
The antibody combination 50E1263224C51122 manifested the highest detection limit in 1-step sandwich-IFA with both synthetic and endogenous antigens and was used to study molecular forms of proBNP and its derivatives in plasma samples of HF patients.
development of an immunoassay for quantification of nt-probnp in human blood
We selected 84 MAbs for their ability to recognize recombinant NT-proBNP in ELISA. Among those, 14 MAbs were specific to region 124, 24 MAbs to region 1327, 19 MAbs to region 2845, 13 MAbs to region 4660, and 15 MAbs to region 6176. Precise epitope mapping of newly generated antibodies was performed by use of a library of synthetic peptides containing overlapping sequences. All MAbs were tested in pairs (capture and detection) in sandwich-IFA with recombinant NT-proBNP or proBNP and with pooled serum or plasma from HF patients as a source of endogenous antigens. All MAbs with remote epitopes being used in 2-site combinations were able to recognize recombinant NT-proBNP and proBNP with a low detection limit (10100 ng/L). Only a few MAb combinations reacted with the antigen from serum and plasma, however. Thus, none of the 13 MAbs that were specific to region 4660 and recognized recombinant NT-proBNP were able to recognize endogenous antigen when used in 2-site combinations with any other antibody.
MAbs pairs with 1 antibody specific to region 1327 and 1 antibody specific to region 6176 recognized endogenous antigen with high sensitivity. When MAbs specific to the N-terminal region or to region 2845 were tested in pairs with antibodies recognizing sequence 6176, the ratio of signals from the endogenous to recombinant peptide was significantly lower than in the assays using MAbs specific to epitopes 1327. Only MAb 29D12, which is specific to peptide 512, elicited high signals upon interaction with either serum or plasma samples being tested in pairs with MAbs specific to peptide 6776. The assay 15C4637113G121320 was selected and used in further studies because it exhibited the highest sensitivities to both recombinant and native antigens.
development of an immunoassay for quantification of probnp in human blood
To develop an assay specific for proBNP, we tested all NT-proBNP- and BNP-specific MAbs in a sandwich-IFA with recombinant proBNP and pooled serum or plasma from HF patients used as antigens. Similarly to their reaction to NT-proBNP in described above studies, MAbs reacted differently with recombinant proBNP and the antigen found in serum or plasma from HF patients. The majority of tested MAb combinations interacted effectively with recombinant proBNP, but far fewer pairs recognized endogenous antigen. Similarly to the previous experiments with NT-proBNP, MAbs against region 2860 of proBNP (able to recognize recombinant proBNP) interacted only weakly with the antigen in patient serum or plasma being tested in all possible 2-site combinations with other antibodies. Antibody pairs including MAbs specific to region 1327 of NT-proBNP and MAbs specific to regions 1122 and 2632 of BNP elicited the highest signals with endogenous antigen. Among antibodies specific to the N-terminal fragment of NT-proBNP, only 29D12 (peptide 512) elicited a relatively high signal when tested in 2-site combinations with endogenous proBNP. In contrast to the NT-proBNP assay, antibodies specific to region 6176 of NT-proBNP in combination with antibodies against region 1122 or 2632 of BNP manifested weak interactions or no interaction with the endogenous antigen but were able to interact with recombinant proBNP.
Among all the antibody combinations tested, the antibody pair 50E116F3 manifested the highest sensitivity to recombinant and endogenous proBNP. The 50E1 antibody is directed against site 2632 of BNP, and MAb 16F3 is specific to peptide 1320 of NT-proBNP.
ifa validation
Typical calibration curves for determination of BNP, proBNP, and NTproBNP and serial dilutions of human plasma samples are shown in Fig. 1
, A B, and C, respectively. The detection limits of the assays were 0.4 ng/L for BNP, 10 ng/L for NT-proBNP, and 3 ng/L for proBNP. The detection limit was defined as a concentration, (measured 20 times in a single run) producing a signal 2 SD above the mean for a calibrator that is free of analyte. The BNP assay was linear in the range of 1200 000 ng/L (r2 = 0.9985). The NT-proBNP assay was linear in the range of 15100 000 ng/L (r2 = 0.9972), and the proBNP assay was linear in the range of 10180 000 ng/L (r2 = 0.9976). Within-assay imprecision (CV) was assessed by measuring 20 replicates of 3 different concentrations of calibrator in pooled normal human plasma and was always <6% (Table 1
). Recoveries from pooled patient or individual patient plasma samples, which were enriched with 3 concentrations of analyte, are also shown in Table 1
. The specificity of antibodies used in the assays was confirmed by Western blotting analysis (data not shown). All antibodies detected recombinant proBNP. MAbs 50E1 and 24C5 also detected synthetic BNP, and 15C4, 16F3, and 13G12 also detected recombinant NT-proBNP. Cross-reactivity of atrial natriuretic peptide in the BNP assay was <0.5%.
|
|
gf-fplc study of plasma samples from hf patients
Patient diagnoses and peptide concentrations in plasma samples are presented in Table 2
. Plasma samples from 7 HF patients were subjected to GF to characterize the molecular forms of proBNP-derived products and to establish their ratios. BNP, proBNP, and NT-proBNP were quantified in fractions with the help of immunoassays 50E124C5, 50E116F3, and 15C413G12, respectively (described above). No preconcentration of specimens was necessary. Mean recoveries in GF studies determined by sandwich-IFA were 68%, 44%, and 59% for BNP, proBNP, and NT-proBNP, respectively.
|
Fractions analysis by the NT-proBNP assay revealed 1 peak of immunoreactivity with a maximum in fraction 13 (
25 kDa; Fig. 2A
). The position of the NT-proBNP activity maximum did not coincide with those established for recombinant NT-proBNP (fraction 15;
15.3 kDa) and recombinant proBNP (fraction 14;
20.6 kDa). ProBNP assay measurements of immunological activities in plasma fractions revealed a maximal concentration in fraction 12 (
37 kDa; Fig. 2B
). A broad concentration peak with maxima in 1213 fractions (
2537 kDa) was observed in the plasma fractions of patient 1.
|
The analysis of the fractions by the BNP assay revealed the presence of 1 major and 1 smaller peak (Fig. 2C
). A minor peak was detected at
46 kDa (fractions 18 and 19; almost coincident with BNP standard) and 1 major peak was detected at
2537 kDa [fraction 12 (6 patients); fractions 12 and 13 (1 patient)]. In all plasma samples, the position of the major BNP peak coincided completely with the position of the immunoreactivity peak measured by proBNP assay (Fig. 2B
).
gf-fplc study of plasma extracts from sep-pak c18 column
To establish the effects of preliminary extraction on the proBNP:BNP ratio, plasma samples from 7 patients described in the aforementioned experiments were applied onto Sep-Pak C18 cartridges. The mean recovery, determined by sandwich-IFA, for this procedure was 30% for BNP and
7.5% for proBNP. The eluates were dried in evaporator, dissolved in a phosphate-salt buffer, and loaded onto a Superdex 75 column. Fractions were analyzed with our experimental BNP and proBNP assays. BNP immunodetection revealed the presence of 2 peaks of immunological activity (Fig. 2D
) in the same fractions as in the experiments with the untreated samples. However, the ratios of the peaks of immunological activities of proBNP and BNP were significantly different from those of nonextracted samples. The values of proBNP:BNP ratio in nonextracted plasma samples and in plasma samples from the same patients measured after extraction on Sep-Pak C18 cartridges are shown in Table 3
. The content of proBNP (1st peak) and BNP (2nd peak) were determined and compared. Ratios of immunologically active proBNP and BNP in plasma samples before extraction varied between 1.8 and 10.8 (mean 6.3), whereas in samples subjected to preliminary extraction, ratios were significantly lower, ranging from 0.3 to 3 (mean 1.5).
|
To demonstrate that the results of the above studies were not dependent on the specificity of our in-house BNP assay, we used a commercial BNP assay (Beckman Access) together with our in-house assay to analyze a representative plasma sample from patient 7 before and after preliminary peptide extraction on Sep-Pak C18 cartridge. The immunological activity profiles of both the commercial and in-house assays (Fig. 3
) have identical patterns and differ solely in the absolute values, with higher concentrations measured by the Beckman immunoassay. Similar to the results of fraction testing obtained with our in-house assay, the ratio of proBNP:BNP measured by the Beckman assay was significantly higher in plasma before preliminary extraction (Fig. 3A
) than in the extracted samples (Fig. 3B
).
|
| Discussion |
|---|
|
|
|---|
37 kDa. The analysis of the fractions with the BNP-specific assay was able to detect both BNP and proBNP, revealing the presence of 2 peaks. The position of the major peak coincided exactly with the peak of proBNP, whereas the 2nd (minor) peak was localized in the area of 46 kDa and was considered to be related to BNP. The ratios of immunoreactive BNP forms (proBNP:BNP) in plasma of HF patients varied from patient to patient, but proBNP content exceeded that of BNP in all cases (by 1.8- to 10.8-fold). On the basis of these findings, we concluded that proBNP was the major antigen contributing to immunological activity measured by both the Beckman commercial assay and our experimental BNP assay in HF patients. To develop reliable and highly sensitive assays for precise and specific detection of proBNP and proBNP-derived peptides, we generated large panels of MAbs specific to NT-proBNP and BNP, with epitopes covering the whole sequences of both molecules. As a source of endogenous protein for pair testing we used not only synthetic/recombinant antigens but also serum and plasma samples from HF patients. This approach helped us resolve 2 issues. First, we identified the epitopes that are changed or modified in the endogenous antigens compared with synthetic/recombinant molecules. Second, we identified MAb combinations that recognized endogenous molecules with the same sensitivity and efficiency as their "artificial" counterparts. We demonstrated that MAbs specific to the central part (region 2860) of the NT-proBNP molecule are almost unable to recognize the antigen in human blood, a phenomenon that may be attributable to posttranslational modification(s) of endogenous NT-proBNP. We have since learned that the central portion of the NT-proBNP molecule is glycosylated, and polysaccharide residues prevent antibodies from interaction with the endogenous antigen. Schellenberger et al. (4) reported that in CHO cells proBNP is expressed as a glycoprotein. Several sites of glycosylation located in the central part of the molecule were documented. It was also demonstrated that proBNP in patient blood is glycosylated. Because antibodies specific to the central, glycosylated part of the NT-proBNP molecule are unable to recognize endogenous antigen, MAbs specific to other regions not affected by glycosylation should be used in NT-proBNP and proBNP assays.
For many years chromatography was the prevailing method of analysis of proBNP, NT-proBNP, and BNP circulating in human blood. Because of the relatively low sensitivity of the immunoassays used to quantify antigens in the samples after chromatography, concentration of the sample before applying it onto column was necessary to compensate for the dilution during chromatography (2)(3)(8).
Previous studies revealed that plasma samples from HF patients contain significant amounts of proBNP. In some samples proBNP concentration was even greater than that of BNP, but usually reported proBNP:BNP ratios were not >2 (2)(3). We observed significantly higher differences between proBNP and BNP concentrations than those previously reported. In our studies the proBNP:BNP ratio varied from 1.8 to 10.8, with a mean of 6.3. The major experimental difference between our study and previous studies was that we skipped the step of plasma extraction/concentration before applying the sample onto the GF column. The high sensitivity of our research assays allowed us to quantify analytes even in the diluted samples after GF-FPLC, leading us to assume that preliminary extraction of the proteins might change proBNP:BNP ratios. To test this hypothesis, we compared the concentrations of both analytes in plasma samples before and after extraction on Sep-Pak C18 cartridges performed according to Shimizu et al. (2). We demonstrated that proBNP:BNP ratios in the samples before extraction and after extraction on Sep-Pak C18 differed significantly. In nonextracted samples, the mean ratio was
6.3, whereas in the extracted samples, molar concentrations of both peptides were almost equal (mean 1.5). The reason for the apparent increase in the magnitude of the BNP peak (relative to proBNP) should be sought in the incomplete elution of proBNP (a significantly bigger molecule than BNP) from Sep-Pak C18 cartridges. We found that extraction changes the peptide ratio significantly, decreasing apparent proBNP content in the sample. Thus we feel that studies of BNP and proBNP content in patient plasma that are performed without preliminary peptide extraction are more relevant, and the results of our study better reflect the real concentrations and/or ratios of analytes.
To verify that differences in proBNP:BNP ratios were not biased by the specificity of our in-house BNP assay, we analyzed 1 of the plasma samples by both a commercial BNP assay (Beckman Access) and our in-house BNP assay (Fig. 3
). For both assays the results of BNP immunoreactivity measurements were very similar. When the sample was not subjected to the preliminary extraction, both assays showed 2 activity peaks: the major peak corresponded to proBNP, and the minor peak almost corresponded to BNP. The proBNP:BNP ratios were 8.6 (Beckman) and 10.8 (in-house). After preliminary extraction, the ratio changed: the peak heights of proBNP and BNP immunoreactivities became practically identical (proBNP:BNP ratio 0.9 Beckman, 1.09 in-house). It should be also noted that our data are in good agreement with recently published results by Hawkridge et al. (9), who observed no detectable free BNP in HF patient plasma. Quantitative differences of our data vs findings of Hawkridge et al. (9) could be explained by methodological differences.
In conclusion, our data suggest that proBNP is the predominant form displaying BNP immunoreactivity in patients with HF. We postulate that direct proBNP measurements by assays, using one antibody specific to the BNP and another to the NT-proBNP part of the molecule, may be of clinical and commercial value. Additional studies of large cohorts of HF patients will be required to obtain experimental evidence for these statements.
| Acknowledgments |
|---|
Financial disclosures: K.R.S., N.N.T., V.L.F., and A.G.K. are employees of HyTest Ltd., which manufactures and markets reagents for BNP testing. F.S.A. has received research funding from the majority of in vitro diagnostic companies that market natriuretic peptide assays. In addition, he has consulted for Abbott Diagnostics, Ortho-Clinical Diagnostics, Biosite, and Senera. No potential conflict of interest relevant to this article was reported.
Financial disclosures: None declared.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
N. N. Tamm, K. R. Seferian, A. G. Semenov, K. S. Mukharyamova, E. V. Koshkina, M. I. Krasnoselsky, A. B. Postnikov, D. V. Serebryanaya, F. S. Apple, M. M. Murakami, et al. Novel Immunoassay for Quantification of Brain Natriuretic Peptide and Its Precursor in Human Blood Clin. Chem., September 1, 2008; 54(9): 1511 - 1518. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Vickery, M. C. Webb, C. P. Price, R. I. John, N. A. Abbas, and E. J. Lamb Prognostic value of cardiac biomarkers for death in a non-dialysis chronic kidney disease population Nephrol. Dial. Transplant., June 18, 2008; (2008) gfn341v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Hammerer-Lercher, B. Halfinger, B. Sarg, J. Mair, B. Puschendorf, A. Griesmacher, N. A. Guzman, and H. H. Lindner Analysis of Circulating Forms of proBNP and NT-proBNP in Patients with Severe Heart Failure Clin. Chem., May 1, 2008; 54(5): 858 - 865. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. R. Seferian, N. N. Tamm, A. G. Semenov, A. A. Tolstaya, E. V. Koshkina, M. I. Krasnoselsky, A. B. Postnikov, D. V. Serebryanaya, F. S. Apple, M. M. Murakami, et al. Immunodetection of Glycosylated NT-proBNP Circulating in Human Blood Clin. Chem., May 1, 2008; 54(5): 866 - 873. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Szinnai, N. G. Morgenthaler, K. Berneis, J. Struck, B. Muller, U. Keller, and M. Christ-Crain Changes in Plasma Copeptin, the C-Terminal Portion of Arginine Vasopressin during Water Deprivation and Excess in Healthy Subjects J. Clin. Endocrinol. Metab., October 1, 2007; 92(10): 3973 - 3978. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S. Apple, A. H.B. Wu, A. S. Jaffe, M. Panteghini, R. H. Christenson, NACB COMMITTEE MEMBERS, R. H. Christenson, F. S. Apple, C. P. Cannon, G. Francis, et al. National Academy of Clinical Biochemistry and IFCC Committee for Standardization of Markers of Cardiac Damage Laboratory Medicine Practice Guidelines: Analytical Issues for Biomarkers of Heart Failure Circulation, July 31, 2007; 116(5): e95 - e98. [Full Text] [PDF] |
||||
![]() |
R. L. Gundry and J. E. Van Eyk Unraveling the Complexity of Circulating Forms of Brain Natriuretic Peptide Clin. Chem., July 1, 2007; 53(7): 1181 - 1182. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||