Clinical Chemistry 53: 922-932, 2007. First published March 15, 2007; 10.1373/clinchem.2006.077784
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Supplemental Data
Right arrow All Versions of this Article:
clinchem.2006.077784v1
53/5/922    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (3)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Manley, S. E.
Right arrow Articles by Luzio, S. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Manley, S. E.
Right arrow Articles by Luzio, S. D.
Related Collections
Right arrow Endocrinology and Metabolism
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2007;53:922-932.)
© 2007 American Association for Clinical Chemistry, Inc.


Automation and Analytical Techniques

Comparison of 11 Human Insulin Assays: Implications for Clinical Investigation and Research

Susan E. Manley1,a, Irene M. Stratton3, Penelope M. Clark2 and Stephen D. Luzio4

1 Department of Clinical Biochemistry and 2 The Regional Endocrine Laboratory, University Hospital Birmingham NHS Foundation Trust, University of Birmingham, Birmingham, United Kingdom.
3 Oxford Centre for Diabetes, Endocrinology & Metabolism, University of Oxford, Oxford, United Kingdom.
4 Diabetes Research Unit, Cardiff University, Cardiff, United Kingdom.

aAddress correspondence to this author at: Clinical Biochemistry, University Hospital Birmingham NHS Foundation Trust, Selly Oak Hospital, Raddlebarn Road, Birmingham B29 6JD, United Kingdom. Fax 44-121-414-0078; e-mail susan.manley{at}uhb.nhs.uk.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Background: The American Diabetes Association task force on standardization of insulin assays in 1996 showed wide variation in assay bias. Newer assays are specific for insulin, with several now available on automated immunoassay analyzers.

Methods: In 2004, we compared 11 commercially available insulin assays by analyzing 150 serum samples (99 fasting/51 postprandial) from study participants with various degrees of glucose intolerance (exclusions being type 1 diabetes, insulin treatment, or presence of insulin antibodies). All assays were calibrated against International Reference Preparation 66/304. One assay was not specific for insulin and another was an RIA; 10 assays used enzyme/chemiluminescent labels. Bland–Altman difference plots were modified to use the mean insulin from all assays on the x-axis as a common comparator.

Results: As in the 1996 study, insulin values from the different assays varied by a factor of 2, with the nonspecific assay ranking in the middle of the distribution. Spearman rank correlation coefficients, for ranking samples vs the mean, were 0.983–0.997. Both offsets and concentration-dependent differences were seen in the modified difference plots. Imprecision (mean CV) for automated assays (3%) was not significantly different from manual assays (5%). Similar values were obtained when one automated assay was run in laboratories in both the UK and the US. Results of 1 assay showed lower insulin concentrations in heparinized plasma than in serum.

Conclusions: Assay performance must be considered before comparing insulin results. The 2-fold variation in insulin results may be related to specificity, manufacturers’ calibration procedures or conversion factors.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
In 1996, the American Diabetes Association (ADA)1 task force on standardization of insulin assays reported widely disparate results (1). Results for plasma and serum from the 17 assays studied, mostly RIAs, varied by a factor of 2. Use of the same insulin reference preparation did not improve comparability, and the same assay method run in 2 laboratories yielded different results. Only 2 of the assays studied in the ADA comparison, an RIA and ELISA, were available for inclusion in this study.

Because of this bias, the task force did not recommend the use of different insulin assays within a population or research study unless their guidelines for assay validation and comparison with other laboratories were followed. To address these issues, major clinical trials (2)(3)(4)(5)(6) have organized comparisons of insulin assays with patient samples or sample exchanges with other laboratories.

Immunometric assays specific for insulin (based on low cross-reactivity to proinsulin/s quoted by the manufacturers) are widely available, commercially, with many assays run on automated platforms (7)(8)(9)(10). These assays, however, may not recognize insulin analogs for treatment of diabetes (11)(12)(13)(14). No international reference method for insulin has as yet been established (15).

External quality assessment schemes still show large differences in insulin values, e.g., the March 2003 UK National External Quality Assessment Service (UK NEQAS) Guildford, which involved 43 assays and 17 methods, found that the trimmed mean insulin concentration in human serum samples for all laboratories was 87 pmol/L, range 40–145 pmol/L, and for serum with added human recombinant insulin was 274 pmol/L, range 160–475 pmol/L (Gwen Wark, personal communication, March 7, 2006).

Specificity, sensitivity, and working ranges are key issues for measurement of circulating insulin for the clinical investigation of hypoglycemia (7). Calculated indices such as those derived from Homeostasis Model Assessment (HOMA) for insulin resistance and ß-cell function, widely quoted in research, are dependent on the absolute value for insulin and on specificity in particular (16).

We compared commercially available insulin assays by assaying serum samples from study participants with varying degrees of glucose tolerance.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
In this study, performed in 2004, insulin concentrations in fasting and postprandial serum samples from study participants with varying degrees of glucose tolerance were measured by use of 11 insulin assays performed in 7 UK and 2 US laboratories. Reconstituted, lyophilized QC material was also circulated. Total proinsulin was measured in one center, insulin compared in serum and heparinized plasma in another, and the same assay compared in 2 laboratories in the UK and US.

Ethics approval was obtained from the South East Wales Local Research Ethics Committee and complies with the current revision of the Helsinki Declaration.

patients
Fasting or postprandial blood samples were collected from 138 participants, 18–75 years of age. Sample donors were either nondiabetic or had impaired fasting glucose, impaired glucose tolerance, or type 2 diabetes. Patients with type 1 diabetes or type 2 diabetes requiring insulin treatment or who were recorded as insulin antibody–positive were excluded. Anonymized data were collected, including age, ethnicity, height, weight, and whether the sample was taken fasting or postprandial. All patients gave written informed consent.

The 150 blood samples (fasting/nonfasting 99/51) were collected from 43 healthy persons (fasting/nonfasting 14/29), 20 patients diagnosed with impaired fasting glucose or impaired glucose tolerance (fasting/nonfasting 18/2), and 87 patients with type 2 diabetes (fasting/nonfasting 67/20). Both fasting and nonfasting samples were obtained from 12 study participants. The mean (SD) age and body mass index of the normoglycemic study participants were 43.8 (14.3) years and 28.5 (6.2) kg/m2. The corresponding values for those with impaired fasting glucose or impaired glucose tolerance were 62.0 (10.1) years and 31.0 (4.9) kg/m2, and for those with type 2 diabetes 60.0 (9.4) years and 31.9 (5.6) kg/m2.

collection of blood
Over a 12-week period in clinics at the Diabetes Research Unit, Cardiff University, UK, blood samples were obtained from study participants via a vacutainer or intravenous cannula with a 3-way tap and collected into a plain vacutainer (yellow, 3.5 mL, Greiner Vacuette Z Serum Sep Clot Activator, Greiner Bio-One) or lithium heparin vacutainer (green, 4 mL, Greiner Vacuette Lithium Heparin). Samples were placed on a roller immediately and serum left to clot for 30 min. Blood samples were spun at 2000g for 5 min at 4 °C. No samples with hemolysis or lipemia were collected. Immediately after centrifugation aliquots (0.5 mL) were pipetted into tubes (Sarstedt) labeled with laboratory and sample codes. The tubes were frozen immediately at –20 °C and transferred to a –70 °C freezer within 24 h.

qc materials
Trilevel, lyophilized QC material (Lyphochek Immunoassay Plus Control, Bio-Rad Laboratories) was reconstituted according to manufacturer instructions, divided into aliquots, and stored at –20 °C.

shipment and measurement of samples and qcs
Samples were shipped on dry ice by an overnight delivery service. On receipt, laboratories stored samples at –20 °C (or –70 °C if available) immediately and did not allow the samples to thaw.

insulin and proinsulin assays
Study organizers selected the most commonly used, commercially available insulin assays from those used by participants in the UK NEQAS Guildford Peptide Hormone schemes. Assay information obtained from the manufacturers is shown in Table 1 . Manual and automated assays for clinical investigation and research were included; some were approved for research use only. Assays from the US were included, one from the previous comparison (1) and another used for major epidemiological studies (17)(18). The Tosoh ST AIA assay was included twice, once from a US research laboratory and once from a routine UK hospital laboratory. All laboratories requested serum when given the option of serum or heparinized plasma. Both types of samples were analyzed with the Biosource-Invitrogen (previously Medgenix) assay. Participating centers were asked to use their laboratory’s standard operating procedures to analyze samples soon after receipt. Serum samples were measured in batches of 30 over a 5-day period. Reconstituted trilevel QC material provided by the study was measured with each batch of samples. Total proinsulin was measured in serum by an ELISA (Linco Total Proinsulin ELISA, Linco Research) at Linco Research, St Charles, MO, with interassay imprecision (CVs) of 2.9%–8.3% at concentrations of 6.52–80.6 pmol/L.


View this table:
[in this window]
[in a new window]

 
Table 1. Details of assays for human insulin provided by manufacturers and imprecision determined in study.

role of the sponsors
The funding organizations had no role in the design of the study, data collection, analysis, or interpretation, or preparation of the manuscript other than diagnostic companies confirming the information published in leaflets for Table 1Up , and did not approve or disapprove of, or delay publication of the work.

statistical analysis
Data were analyzed by use of the Statistical Analysis System (SAS Institute) (19). Because there is no consensus regarding statistical methods for comparison of this number of insulin assays without a reference method, we compared individual assays. We found no aberrant assays, so we then compared each of the 12 sets of assay results with the overall mean. The method of Carstensen (20) was not appropriate, as it requires replicate measurements within an assay. Difference plots (21)(22) were modified to use the mean of all results for the x-axis as a common comparator in the Figs. in this paper. Bland–Altman difference plots for pairs of assays (one assay compared with one other) are presented in the online Data Supplement (see the Data Supplement that accompanies the online version of this article athttp://www.clinchem.ort/content/vol53/issue5). Each data point, zero (nil difference) lines, mean differences, and 1.96 SD lines are shown. Ordinary least squares linear regression was used for calculation of intercepts and slopes in the modified difference plots; Spearman rank correlation coefficients were calculated for each assay vs the mean.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
insulin assays
Serum was assayed in 7 UK and 2 US laboratories between May and July 2004 with commercial insulin assays (Table 1Up ) described by the manufacturer as standardized against the first International Reference Preparation 66/304 (IRP 66/304). Eleven immunometric assays using various separation and detection systems were included, i.e., 4 immunoenzymometric assays, 2 immunochemiluminometric assays, 1 immunoelectrochemiluminometric assay, 3 ELISAs, and 1 RIA. Only the ELISA from Mercodia was not specific for insulin, showing 54% cross-reactivity with intact proinsulin (manufacturer’s data). Six assays were run on automated, high-throughput immunoassay analyzers used in routine clinical laboratories, and 6 were run manually. Sample volume varied from 20 to 100 µL, not including dead volume for sampling with automated analyzers or the volume for duplicate assay if recommended by the manufacturer. Sensitivity, as stated by the manufacturers, ranged from 0.7 to 12 pmol/L. Some manufacturers provided details on the experiment used to determine sensitivity; it was usually determined by replicate analyses of a matrix material free of insulin. The mean of the zero signal plus 2 SD was used for 6 assays and mean plus 2.5 SD for another, but was not defined for the others. The upper concentration limit above which dilution was recommended ranged from 600 to 6945 pmol/L. Conversion factors, from mIU/L to pmol/L, varied from 6 to 7.46; Tosoh updated its conversion factor from 6 to 7.17 in July 2006 and DPC from 6 to 7.217.

assay imprecision
Imprecision, the mean of CVs for trilevel QC analyzed on 5 days, varied from 1.4% for 1 automated assay to 6.7% for a manual ELISA (Table 1Up ). The mean CV for automated assays (Table 1Up ) was 3.0% (range 1.4%–5.6%) and 5.4% for manual methods (range 3.9%–6.7%), but this difference was not significant. Imprecision for the low QC varied from 1.1% to 12.1% (Table 2 ) and for the high QC from 1.1% to 7.3%. CVs were constant across the range for 2 automated assays (mean CVs 1.4% and 2.5%), but imprecision for the low QC in 2 laboratories using the same assay was 9.4% in one laboratory and 3.1% in the other.


View this table:
[in this window]
[in a new window]

 
Table 2. Insulin values for quality controls and serum samples, pmol/L.

Insulin values for the low and mid-range QCs varied by a factor of 3, with mean values ranging from 35.9 to 106.5 pmol/L for the low QC and 83.8 to 291.1 pmol/L for the mid-range QC. For the high QC, values varied by a factor of 4, from 167.4 pmol/L to 709.8 pmol/L. One assay (DPC Immulite 1000) consistently reported low values for QCs. The direction and magnitude of bias with reconstituted QCs was not necessarily reflected in the results for serum (Table 2Up ). The source of insulin (bovine or recombinant human) was not stated by the manufacturer of the QC materials.

insulin measurement in heparinized plasma and serum
Measured insulin concentrations were higher in serum, median (IQ range) 132 (90–205) pmol/L, minimum 38 pmol/L, and maximum 1387 pmol/L, than plasma, 121 (85–168) pmol/L, minimum 27 pmol/L, and maximum 1402 pmol/L, with differences proportional to the concentration (Fig. 1 ). The linear regression equations were:

Formula
or

Formula
Corresponding equations in mIU/L (conversion factor 7.46) were:

Formula
or

Formula


Figure 1
View larger version (18K):
[in this window]
[in a new window]

 
Figure 1. Comparison of insulin (pmol/L) in serum and lithium heparinized plasma samples, n = 150, using Biosource-Invitrogen assay only.

Scattergram (solid line)line of identity A = B. Difference plot (solid line) zero line, (dashed lines) mean difference, 2 SD lines.

insulin and proinsulin
Insulin results for serum from the different assays varied by a factor of 2 for median and maximum values (Table 2Up ). The median and maximum values for the nonspecific assay ranked in 7th place (starting with the lowest value). Minimum values for insulin in human samples varied by a factor of >20 (Table 2Up ), with values of <10 pmol/L obtained from 2 automated assays (Tosoh and Roche) and the nonspecific ELISA (Mercodia), and from 11.5 to 38 pmol/L for the others. Spearman correlation coefficients (r2) vs the mean ranged from 0.983 to 0.997 (Table 3 ). Similar results were obtained when serum was run on the same automated assay (Tosoh) in 2 laboratories (Table 2Up ). The results fell on or around the line of identity, with a mean difference of 1.92 pmol/L (95% CI –0.12 to 3.97), P = 0.0647, (Fig. 2 ). Total proinsulin ranged from 2 to 168 pmol/L with median (IQ range) 23 (12–43) pmol/L.


View this table:
[in this window]
[in a new window]

 
Table 3. Mean differences for insulin for each assay from the mean of all 12 assays for 150 serum samples, linear regression equations A differences {Delta}a from mean values vs mean insulin and B actual value ya vs mean insulin, and correlation coefficients.


Figure 2
View larger version (16K):
[in this window]
[in a new window]

 
Figure 2. Comparison of TOSOH ST AIA insulin results (pmol/L) from UK and US laboratories in serum samples, n = 150.

Scattergram (solid line) line of identity A = B. Difference plot (solid line) zero line, (dashed lines) mean difference, 2 SD lines.

difference plots
The modified difference plots (Fig. 3 ) and usual difference plots (see Figs. in the online Data Supplement) showed constant differences or differences that were proportional to the concentration. The difference plots for one assay vs another, available electronically, validate the use of the mean value for a common x-axis and will be of interest to laboratories running a particular assay/s.


Figure 3
Figure 3
Figure 3
View larger version (86K):
[in this window]
[in a new window]

 
Figure 3. Difference plots for human insulin in 150 serum samples measured by 12 assays in UK or US, on left 0–150 pmol/L and right 0–1500 pmol/L.

Difference plot (solid line) zero line, (dashed lines) mean difference, 2 SD lines.

The slope and intercept for differences vs the mean are presented in Table 3Up as linear regression equations A, and for insulin values vs mean (scattergrams not shown) in Table 3Up as B. Marked differences from the means were found for manual assays from Linco ELISA and Dako, with slopes of –0.32 and –0.20 (Table 3AUp ), with the Linco RIA giving higher results, slope 0.19. The Biosource-Invitrogen assay that did not report results <38 pmol/L showed high concentration-dependent differences, as did the Mercodia ELISA that was reported to cross-react with total proinsulin, whereas the other 2 ELISAs had negative concentration-dependent differences (Fig. 3Up ). The MLT assay showed an offset with an intercept of –17 pmol/L. Of the automated assays, Abbott AxSYM and Bayer Advia Centaur showed positive, concentration-dependent differences (N.B., conversion factors >6.0), the DPC Immulite 1000 showed negative differences, and the Roche Elecsys E170 and Tosoh ST AIA had results closest to the mean of the assays.


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Since the ADA report on insulin standardization in 1996 (1), specific insulin assays have been introduced that incorporate nonradioisotopic labels and are suitable for automated platforms. In this comparison of 11 insulin assays, values measured for serum varied by a factor of 2, as reported earlier (1). However, progress with an international reference method and preparation has been slow, although recently mass spectrometric methods for insulin have been described (23).

Specificity of assays, calibration procedures, specimen type, assay performance, and conversion factors may all contribute to interassay variation. In this study, total proinsulin concentrations were in accord with previously published reports of 10% (proinsulin:insulin) for samples from persons without diabetes and 20% for patients with type 2 diabetes (24), with no samples displaying inappropriate hyperproinsulinemia (25). All assays except 1 were reported by manufacturers to be specific for insulin, and 1 was a specific RIA (Linco) that was included in the previous ADA comparison. The Tosoh assay, run in the US and UK for this study, produced similar results in the different laboratories, in contrast to the previous ADA comparison. In US epidemiological studies this assay was chosen recently to replace the nonspecific Pharmacia RIA100, which had been withdrawn because antibody stocks were exhausted. The nonspecific assay (Mercodia Iso-Insulin) ranked in the middle of the distribution of the assays in this study. This manufacturer, Mercodia, also produces a specific insulin assay with no detectable cross-reactivity with proinsulin and des 31–32 and 32–33 split proinsulins.

All manufacturers reported that their assays were calibrated against the same insulin reference preparation (IRP 66/304) prepared from human pancreas with a stated number of units per ampule. Details of manufacturers’ calibration procedures and calibrants (26) were not requested by the authors. Mean (range) recovery of IRP 66/304 in 14 methods has been reported to be 89% (75% to 127%), (27). Some assays showed overrecovery (Abbott AxSYM 127%), some underrecovery (DPC Immulite 75%), and some had recoveries within 10% (Bayer Centaur 100%, Dako 91%, Mercodia ELISA 99%, MLT 98%, and Roche Elecsys 103%). It remains to be seen whether the use of a human recombinant insulin preparation, calibrated in molar and mass terms, will reduce the interassay variability.

All participating laboratories measured insulin in serum, although heparinized plasma was also available. For illustrative purposes, both sample types were compared in 1 assay, and in that case higher insulin values were obtained in serum. Because these differences may be assay specific, individual laboratories should confirm with the assay manufacturer the procedures recommended for specimen collection and compare results for plasma and serum as required (28). The differences observed could be clinically significant when insulin concentrations or derived mathematical functions are compared for patient care or research.

Five assays were run on automated analyzers, and no significant difference in imprecision was observed between automated and manual assays. Bias and imprecision for lyophilized QCs in the individual assays did not always mirror those seen in serum (Table 2Up ), an important point for external quality assessment schemes. The use of nonhuman insulin in QCs and matrix effects may explain these differences. Although the security of baseline or low concentrations is important (28), sensitivity, as reported by the manufacturers, was determined in a number of ways that were not comparable. Use of the minimum detectable concentration from precision profiles may yield more reliable estimates.

A conversion factor of 6 for recombinant human insulin has been reported (1)(29). In this study, however, the conversion factors used by the manufacturers were found to vary considerably, with the potential to influence insulin values by up to 15%. For 3 of 4 specific assays using conversion factors >6, reported mean values were in the top half of the distribution, and the other value was in the middle. Assays using a factor of 6 ranked mainly in the bottom half, except for the specific RIA (Linco) that gave the highest results. Interestingly, an ELISA from the same company, also using 6 as a factor, gave the lowest results.

Because of potential assay interference, our study has limitations that prohibit the extrapolation of results to patients with insulin antibodies. The results are also not applicable to patients treated with insulin analogs, because these cannot be detected with some specific insulin assays (11)(12)(13)(14). A limited number of batches of reagent and calibrators were used in this study. Because there is no reference method for insulin or consensus on standard statistical techniques for such multiassay analyses (30), the mean insulin value for each sample was used as a common comparator. This approach may not be appropriate for a smaller number of assays.

Various techniques have been used over the years for comparing measurements from different studies and for reducing assay variation. For major research studies, insulin has been measured in central laboratories (3). In some studies data have been divided into ranks, but this method may not be reliable because populations differ (31), and it is not helpful when considering individuals. Results for other analytes such as hemoglobin A1c have been aligned to those reported by major clinical trials while the international reference method was being evaluated (32). In the case of human chorionic gonadotropin, clearly defined calibration procedures for the international standard and careful selection of antibodies reduced assay variation (33). Intermethod differences of 2- to 5-fold in prostate-specific antigen measurement results (34) were eliminated by the introduction of standard calibrators and guidelines, but differences in the recovery of international standards on automated platforms still varied by –5% to +22%. These results may be related to data transfer between the primary standards and master calibrator, and between master calibrator and individual reagent set calibrators, a potential issue for all analytes.

International diabetes and clinical chemistry organizations should be aware of the need for comparability of insulin values for patient care, clinical research studies, and any resulting metaanalyses or clinical research networks (35). Clinicians and researchers should be alerted to the fact that serum insulin measurements from the various assays now available can differ by a factor of 2.


   Acknowledgments
 
Grant/funding support: Diabetes UK provided a small research grant. Diagnostic companies provided insulin kits and covered some despatch costs: Abbott (C. Thompson); Bayer now Siemens (A. Thompson); DakoCytomation (H. Masters, C. Smellie); DPC now Siemens (C. Field, J. Hayes); Linco US (R. Gingerich, J. Mistry, A. Tunnicliffe); Biosource-Invitrogen, previously Medgenix (M. Salmon); Mercodia (P. Atkinson); M.L.T. now Invitron (H. Jeffreys, S. Woodhead); Roche (A. D’Souza); Tosoh UK and US (S. Marivoet). Sponsor for centers’ meeting – Mr. M. Fineman, Amylin Pharmaceuticals, San Diego, CA.

Financial disclosures: Dr. P.M. Clark is a member of the Specialist Advisory Group for Immunoassay in UK NEQAS for which she receives no remuneration. Dr. S.E. Manley and Mrs. I.M. Stratton received grant support from the Novo Nordisk Research Foundation, for a study on measures of glycemic control and have subsequently received funding, reagents, and equipment for this study from Bayer and DPC (both now Siemens), Roche, and Tosoh.

Acknowledgements: We would like to thank Diabetes UK for their grant, Professor Doug Altman, University of Oxford, for advising on statistical issues, Dr. Gwen Wark, UK NEQAS Scheme Organiser, for EQA data supplied with permission from the EQA scheme, Mrs. Rachel Round for administration, and Professor Stephen Gough, Miss Janet Smith, Dr. Malcolm Nattrass from University Hospital Birmingham NHS Foundation Trust and Dr. Tim Barrett from Birmingham Children’s Hospital NHS Trust for their helpful comments and support.

Also the following in the UK: A. Powell, P. Griffiths, (Birmingham Children’s Hospital); R. Round; M. Palma, M. Peyton, J. Babrah (Clinical Biochemistry, University Hospital Birmingham); R. Williams, (Derriford Hospital, Plymouth); G. Dunseath, A. Clark (Diabetes Research Unit, Cardiff University); G. Wark (Royal Surrey County Hospital, Guildford); R. D’Souza, R. Elms, (St George’s Hospital, Tooting); S. Fell, B. Starling (Walsall Manor Hospital): in the US: H-M. Wiedmeyer, A. Tennill (Diabetes Diagnostic Laboratory, Columbia, University of Missouri Health Care); J. Mistry (Linco Research Inc, St Charles, Missouri).


   Footnotes
 
1 Nonstandard abbreviations: ADA, American Diabetic Association; UK NEQAS, United Kingdom National External Quality Assessment Service; HOMA, Homeostasis Model Assessment; IRP 66/304, International Reference Preparation 66/304.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Robbins DC, Andersen L, Bowsher R, Chance R, Dinesen B, Frank B, et al. Report of the American Diabetes Association’s task force on standardization of the insulin assay. Diabetes 1996;45:242-256.[Abstract]
  2. Cull CA, Manley SE, Stratton IM, Haw N, Ross IS, et al. Approach to maintaining comparability of biochemical data during long-term clinical trials. Clin Chem 1997;43:1913-1918.[Abstract/Free Full Text]
  3. . UK Prospective Diabetes Study Group. UK Prospective Diabetes Study (UKPDS) 33 Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes. Lancet 1998;352:837-853.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  4. . UK Prospective Diabetes Study Group. UK Prospective Diabetes Study (UKPDS) XI: Biochemical risk factors in type 2 diabetic patients at diagnosis compared with age-matched normal subjects. Diabetic Med 1994;11:534-544.[ISI][Medline] [Order article via Infotrieve]
  5. . UK Prospective Diabetes Study Group. UK Prospective Diabetes Study XII. Differences between Asian, Afro-Caribbean and white Caucasian type 2 diabetic patients at diagnosis of diabetes. Diabetic Med 1994;11:670-677.[ISI][Medline] [Order article via Infotrieve]
  6. Meigs JB, Haffner SM, Nathan DM, D’Agostino RB, Wilson PW. Sample exchange to compare insulin measurements between the San Antonio Heart Study and the Framingham Offspring Study. J Clin Epidemiol 2001;54:1031-1036.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  7. Clark PM. Assays for insulin, proinsulin(s) and C-peptide. Ann Clin Biochem 1999;36:541-564.[ISI][Medline] [Order article via Infotrieve]
  8. Sapin R. Insulin assays: previously known and new analytical features. Clin Lab 2003;49:113-121.[Medline] [Order article via Infotrieve]
  9. El Kenz H, Bergmann P. Evaluation of immunochemiluminometric assays for the measurement of insulin and C-peptide using the ADVIA Centaur. Clin Lab 2004;50:171-174.[Medline] [Order article via Infotrieve]
  10. Morris ER, Dinesen B, Burnett MA, Christopher P, Bassett PA, Manley SE. A new ELISA for specific insulin measurements in type II diabetic patients. Biochem Soc Transactions 1992;21:25S.
  11. Heald AH, Bhattacharya B, Cooper H, Ullah A, McCulloch A, Smellie S, et al. Most commercial insulin assays fail to detect recombinant insulin analogues. Ann Clin Biochem 2006;43:306-308.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  12. Owen WE, Roberts WL. Cross-reactivity of three recombinant insulin analogs with 5 commercial insulin immunoassays. Clin Chem 2004;50:257-259.[Free Full Text]
  13. Sapin R, Le Galudec V, Gasser F, Pinget M, Grucker D. Elecsys insulin assay: free insulin determination and the absence of cross-reactivity with insulin lispro. Clin Chem 2001;47:602-605.[Free Full Text]
  14. Moriyama M, Hayashi N, Ohyabu C, Mukai M, Kawano S, Kumagai S. Performance, evaluation and cross-reactivity from insulin analogs with the ARCHITECT insulin assay. Clin Chem 2006;52:1423-1426.[Abstract/Free Full Text]
  15. Stenman U-H. Immunoassay standardization: is it possible, who is responsible, who is capable?. Clin Chem 2001;47:815-820.[Free Full Text]
  16. Wallace TM, Matthews DR. The assessment of insulin resistance in man. Diabetic Med 2002;19:527-534.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  17. Bravata DM, Wells CK, Concato J, Kernan WN, Brass LM, Gulanski BI. Two measures of insulin sensitivity provided similar information in a US population. J Clin Epidemiol 2004;57:1214-1217.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  18. Yang EJ, Kerver JM, Park YK, Kayitsinga J, Allison DB, Song WO. Carbohydrate intake and biomarkers of glycemic control among US adults: the third National Health and Nutrition Examination Survey (NHANES III). Am J Clin Nutr 2003;77:1426-1433.[Abstract/Free Full Text]
  19. . SAS Institute: Statistical Analysis System 6th ed 1990:1042 SAS Institute Cary, NC. .
  20. Carstensen B. Comparing and predicting between several methods of measurement. Biostatistics 2004;5:399-413.[Abstract]
  21. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-310.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  22. Twomey PJ. How to use difference plots in quantitative method comparison studies. Ann Clin Biochem 2006;43:124-129.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  23. Van Uytfanghe K, Rodríguez-Cabaleiro D, Stöckl D, Thienpont LM. New liquid chromatography–electrospray ionisation–tandem mass spectrometry measurement procedure for quantitative analysis of human insulin in serum. Rapid Commun Mass Spectrom 2007;21:1-3.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  24. Nagi DK, Knowler WC, Mohamed-Ali V, Bennett PH, Yudkin JS. Intact proinsulin, des 31,32 proinsulin, and specific insulin concentrations among nondiabetic and diabetic subjects in populations at varying risk of type 2 diabetes. Diabetes Care 1998;21:127-133.[Abstract]
  25. Warren-Perry MG, Manley SE, Ostrega D, Polonsky K, Mussett S, Brown P, Turner RC. A novel point mutation in the insulin gene giving rise to hyperproinsulinemia. J Clin Endocrinol Metab 1997;82:1629-1631.[Abstract/Free Full Text]
  26. Niederhau CM, Reinauer H. Evaluation of the IMx insulin analyzer [Abstract]. Clin Chem 1996;42:S190.
  27. Wark G. UK NEQAS Guildford Peptide Hormones. Annual Review 2005. Available from UK NEQAS Guildford Peptide Hormones scheme, Clinical Laboratory, Royal Surrey County Hospital, Guildford, Surrey, GU2 7XX, UK.
  28. Clark PM, Wood PJ, Teale DJ, Williams R. Inconsistencies in the measurement of serum/plasma insulin [Abstract]. ACB FOCUS 2003;P85:A14.
  29. Volund A, Brange J, Drejer K, Jensen I, Markussen J, Ribel U, et al. In vitro and in vivo potency of insulin analogues designed for clinical use. Diabetic Med 1991;8:839-847.[ISI][Medline] [Order article via Infotrieve]
  30. Horsti J, Uppa H, Vilpo JA. Poor agreement among prothrombin time international normalized ratio methods: comparison of seven commercial reagents. Clin Chem 2005;51:483-485.[Free Full Text]
  31. . The DECODE Study Group. Are insulin resistance, impaired fasting glucose and impaired glucose tolerance all equally strongly related to age?. Diabetic Med 2005;22:1476-1481.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  32. Manley SE, Round RA, Smith JM. Calibration of HbA1c and its measurement in the presence of variant haemoglobins: report on questionnaire to manufacturers. Ann Clin Biochem 2006;43:135-145.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  33. Bristow A, Berger P, Bidart JM, Birker S, Norman R, Stenman UH, et al. Establishment, value assignment and characterization of new WHO reference reagents for five molecular forms of human chorionic gonadotropin. Clin Chem 2005;51:177-182.[Abstract/Free Full Text]
  34. Roddam AW, Rimmer J, Nickerson C, Ward AM, . on behalf of the NHS Prostate Cancer Risk Management Programme. Prostate-specific antigen: bias and molarity of commercial assays for PSA in use in England. Ann Clin Biochem 2006;43:35-48.[CrossRef][ISI][Medline] [Order article via Infotrieve]
  35. Heller S, Kinmonth AL. A clinical research network in diabetes for the UK. Diabetic Med 2004;21:1061-1063.[CrossRef][ISI][Medline] [Order article via Infotrieve]



The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
D. Rodriguez-Cabaleiro, K. Van Uytfanghe, V. Stove, T. Fiers, and L. M. Thienpont
Pilot Study for the Standardization of Insulin Immunoassays with Isotope Dilution Liquid Chromatography/Tandem Mass Spectrometry
Clin. Chem., August 1, 2007; 53(8): 1462 - 1469.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Supplemental Data
Right arrow All Versions of this Article:
clinchem.2006.077784v1
53/5/922    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (3)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Manley, S. E.
Right arrow Articles by Luzio, S. D.
Right arrow Search for Related Content