Clinical Chemistry 53: 1503-1510, 2007. First published June 7, 2007; 10.1373/clinchem.2006.078485
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow 078485.Supplemental Data
Right arrow All Versions of this Article:
clinchem.2006.078485v1
53/8/1503    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Xu, Y.
Right arrow Articles by Li, Q.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Xu, Y.
Right arrow Articles by Li, Q.
Related Collections
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2007;53:1503-1510.)
© 2007 American Association for Clinical Chemistry, Inc.


Automation and Analytical Techniques

Multiple Fluorescent Labeling of Silica Nanoparticles with Lanthanide Chelates for Highly Sensitive Time-Resolved Immunofluorometric Assays

Ye Xu1 and Qingge Li1,2,a

1 Molecular Diagnostics Laboratory, Department of Biomedical Sciences, and the Key Laboratory of the Ministration of Education for Cell Biology and Tumor Cell Engineering, School of Life Sciences, Xiamen University, Xiamen, Fujian 361005, China.
2 The Key Laboratory of Chemical Biology of Fujian, Xiamen University, Xiamen, Fujian 361005, China.

aAddress correspondence to this author at: Department of Biomedical Sciences, School of Life Sciences, Xiamen University, Xiamen, Fujian 361005, China. Fax 86-592-2187363; e-mail qgli{at}xmu.edu.cn.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussions
References
 
Background: Time-resolved immunofluorometric assays (TrIFA) using lanthanide-labeled nanoparticles have greatly increased the sensitivity of immunoassays. Current labeling strategies, however, use either physical doping of lanthanide chelates into preformed nanoparticles or covalent linking of lanthanide chelates to precursors used for making nanoparticles; both these strategies have drawbacks.

Methods: Luminescent Eu(III) and Tb(III) chelates were covalently coated on the surface of preformed silica nanoparticles to which detection antibodies or bridging proteins for antibody binding were conjugated. We used the resulting conjugates in TrIFA for detection of hepatitis B surface antigen (HBsAg) and hepatitis B e antigen (HBeAg), both individually and simultaneously. We compared the results of the newly established method with results of an ELISA for serum samples. Positive samples identified by TrIFA but not by ELISA were confirmed by additional assays, including real-time PCR detection of viral DNA.

Results: The prepared nanoparticle conjugates were homogeneous in size, at ~55 (5) nm in diameter [mean (SD)], were stable for long-time storage (>2 years), and contained more chelates [6.86 x 105 for Eu(III), 4.73 x 104 for Tb(III)] per nanoparticle than particles made as previously reported. The TrIFA established for HBsAg had a comparable or lower detection limit (0.0092 µg/L) than existing nanoparticle-based TrIFA or ELISA. The TrIFA for HBeAg had a much lower detection limit [10.0 National Centre Unit (NCU)/L] than ELISA and detected HBeAg in 5 samples missed by the ELISA method. Simultaneous TrIFA for both HBsAg and HBeAg was achieved with detection limits (0.033 µg/L for HBsAg and 27.0 NCU/L for HBeAg) close to those of the individual assays.

Conclusions: Covalent surface labeling of silica nanoparticles with lanthanide chelates provides good fluorescent labels that can be used in TrIFA for highly sensitive and robust detection of clinical targets.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussions
References
 
Time-resolved fluorometry (TRF)1 using lanthanide chelates has been widely used in ultrasensitive quantitative immunoassays (1)(2)(3)(4)(5). The unique advantage of TRF is its high sensitivity compared with other routinely used immunoassays, such as ELISA. Currently, the most widely used TRF-based immunoassay is the dissociation-enhanced, lanthanide-induced fluorescence immunoassay (6), in which nonfluorescent Eu(III) chelate is used as label and the amount of Eu(III) is measured through a dissociation and fluorescence enhancement step. Another reported system is the enzymatically amplified time-resolved fluorescence immunoassay (7), in which the alkaline phosphatase used as the label is detected through a fluorogenic reaction with Tb(III). These methods both use an indirect labeling strategy that needs a signal development step before fluorescence detection.

The signal development step of lanthanide-based immunoassays involves extra reaction procedures that are time-consuming and vulnerable to ambient lanthanide contamination. If a fluorescent lanthanide chelate can be used directly as a label, then the signal development step becomes unnecessary and fluorescence can be measured from the solid phase immediately after the immune reactions. Direct labeling has not found wide applications in TRF-based immunoassays, however, mainly because a lanthanide chelate must be highly fluorescent, thermodynamically stable, water soluble, and covalently reactive to serve as a direct label. Such properties are hard to find in any single chelate.

While the development of new chelates is still ongoing, efforts to explore new features of existing fluorescent chelates are generating some intriguing results. An interesting feature found is that the Eu(III) chelate shows fluorescence-enhancing effects in a multiple labeling system (8)(9)(10). In such a system, a carrier molecule such as streptavidin (SA) (8) or a polymeric compound (11)(12) carrying hundreds of europium chelates is used to label antibody for highly sensitive TRF immunoassays. A breakthrough to further boost the multiple-labeling strategy is the development of nanoparticles containing thousands of fluorescent lanthanide chelates for TRF immunoassays (13)(14)(15)(16)(17)(18)(19)(20)(21). These nanoparticles are conjugated to SA or antibody to create a detection reagent, which allows direct measurement of fluorescence from the nanoparticles. Polystyrene nanoparticles containing fluorescent Eu(III) chelates have been used for ultrasensitive detection of prostate-specific antigen (13)(20). Recently, Eu(III), Tb(III), Sm(III), and Dy(III) chelate-dyed latex nanoparticles were also synthesized and used in sandwich-type immunoassays (16). At about the same time, silica nanoparticles containing Eu(III) or Tb(III) chelates were prepared and evaluated for TRF immunoassay (22)(23)(24)(25). Silica nanoparticles containing lanthanide chelates are prepared either by physical doping into the preformed silica nanoparticles or prelinking to precursors for on-site silica nanoparticle formation.

In the present study, we covalently linked lanthanide chelates onto the surface of preformed silica nanoparticles. We then conjugated these particles to detection antibodies and used the resulting conjugates in time-resolved immunofluorometric immunoassay (TrIFA) of hepatitis B surface antigen (HBsAg) and hepatitis B e antigen (HBeAg), both individually and simultaneously, with detection limits being 20- to 30-fold lower than those of ELISA. Evaluation of a large number of clinical samples with the current method detected all positive sera identified by ELISA and 5 HBeAg-positive results that were missed by ELISA. These data indicate that our new labeling approach is suitable for highly sensitive TrIFA in clinical settings.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussions
References
 
preparation of lanthanide-coated nanoparticles
The scheme for preparation of Eu(III)- and Tb(III)-coated nanoparticles is described in Fig. 1A . Preparation details are provided in the Materials and Methods of the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol53/issue8. Briefly, we synthesized bare silica nanoparticles using reverse microemulsion (26)(27). We introduced amino groups to the surfaces of these particles by treatment with (3-aminopropyl)trimethoxylsilane (APTMS) (28). We used the aminated nanoparticles for coating Eu(III) ligand 4,4-bis(1,1,2,2,3,3-heptafluoro-4,6-hexanedion-6-yl)chlorosulfo-o-terphenyl (BHHCT) or Tb(III) ligand N,N,N1, N1-[2,6-bis(3-aminomethyl-1-pyrazolyl)-phenylpyridine] tetrakis (BPTA), followed by mixing with EuCl3 or TbCl3 solutions.


Figure 1
View larger version (11K):
[in this window]
[in a new window]

 
Figure 1. Schematic illustration of preparation of lanthanide chelate–coated silica nanoparticles and their site-specific conjugation with antibody or bridging protein.

(A), bare silica nanoparticles were first aminated using APTMS. Eu(III) ligand BHHCT was directly introduced through reaction between chlorosulfo group of BHHCT and amino groups on nanoparticles. After loading with EuCl3, the nanoparticles were recovered by centrifugation. The nanoparticles obtained were treated sequentially with TEOS, APTMS, BHHCT, and EuCl3 as above another 4 times. Tb(III) ligand BPTA was introduced through EDC to link the carboxylic acids groups of BPTA and amino groups on nanoparticles. After loading with TbCl3, the nanoparticles were recovered by centrifugation and ready for use. (B), the antibody was first oxidized by NaOI4 to generate aldehyde groups and then was conjugated to Eu(III)-BHHCT–coated nanoparticles through amino groups on the surface. (C), Tb(III)-BPTA–coated nanoparticles were first linked with oxidized Dextran 500 through Schiff reaction and then SA was immobilized onto the surface of nanoparticles.

characterization of fluorescent nanoparticles
The number of nanoparticles in 1 mL suspension was calculated according to the density of the silica particle (1.96 g/cm3), average particle diameter, and particle concentration, as reported (13)(23). The apparent number of chelates per nanoparticle was calculated by comparing the signal of a known number of nanoparticles in washing buffer B (10 mmol/L Tris-HCl, pH 9.1, containing 0.5 g/L Tween 20) with a free chelate calibrator in the same buffer. Spectra analysis and lifetime measurements of the nanoparticles are described in the online Data Supplement.

nanoparticle labeling
We conjugated anti-HBsAg IgG to Eu(III)-coated silica nanoparticles using a site-specific method (29) sketched in Fig. 1BUp . Conjugation of SA to Tb(III)-BPTA–coated nanoparticles was bridged with oxidized Dextran 500, sketched in Fig. 1CUp . Detailed protocols are provided in the online Data Supplement.

nonspecific binding of antibody-conjugated lanthanide nanoparticles
We carried out nonspecific binding studies with particle concentrations ranging from 4.1 x 10–12 mol/L to 1.3 x 10–10 mol/L, which covers the concentration (6.5 x 10–11 mol/L) used for calibration curve studies. Detailed protocols are described in the online Data Supplement.

immunoassays
After monoclonal antibody B20 (5 µg/mL in 20 mmol/L Tris-HCl buffer, pH 7.4) was physically coated on the microtitration wells (50 µL/well), we added 50 µL HBsAg standard solutions or serum samples to each well. We incubated the plate at 37 °C for 45 min followed by 3 washes with washing buffer A [10 mmol/L PBS (10 mM sodium phosphate, pH 7.4, containing 0.15 M NaCl), pH 7.4, containing 0.5 g/L Tween 20]. Then we added 50 µL Eu(III)-BHHCT–coated nanoparticles conjugated with monoclonal anti-HBsAg antibody S04 [diluted 2000-fold with dilution buffer (10 mmol/L Tris-HCl, pH 7.8, containing 2 g/L bovine serum albumin, 1 g/L NaN3, and 9 g/L NaCl)] to each well. The plate was incubated at 37 °C for 90 min, washed 5 times with washing buffer B, and subjected to TRF measurement. TrIFA of HBeAg was carried out similarly, except that we used biotinylated anti-HBeAg antibody and SA-conjugated, Tb(III)-BPTA–coated silica nanoparticles. Detailed procedures are provided in the online Data Supplement.

We carried out simultaneous assay of HBsAg and HBeAg with Eu(III)-BHHCT– and Tb(III)-BPTA–coated nanoparticles as follows. Both anti-HBsAg monoclonal antibody B20 and anti-HBeAg monoclonal antibody KE (each diluted to 5 µg/mL with 20 mmol/L Tris-HCl buffer, pH 7.4) were simultaneously coated on the microtitration wells (50 µL/well) by physical adsorption. Then we added 50 µL HBsAg and HBeAg standard solutions or serum samples to each well. The plate was incubated at 37 °C for 45 min and washed 3 times with washing buffer A. We added biotinylated anti-HBeAg (50 µL) antibody to each well and incubated the plate at 37 °C for 45 min. After washing the plate 3 times with washing buffer A, we added 50 µL of the Eu(III)-BHHCT–conjugated anti-HBsAg and Tb(III)-BPTA–conjugated SA to each well. The plate was incubated at 37 °C for 90 min, washed 5 times with washing buffer B, and subjected to TRF measurement.

For comparison studies, we comparatively analyzed all serum samples with ELISA reagent sets. For samples showing inconsistent results between TrIFA and ELISA, we performed real-time PCR according to the manufacturer’s manual to verify the existence of viral DNA. We carried out analytical recovery of TrIFA by adding standard HBsAg or HBeAg to 3 human serum samples (1 negative sample and 2 positive samples) and detecting the concentrations of HBsAg or HBeAg. We also subjected the reference panels for HBsAg and HBeAg assays to TrIFA for validation study.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussions
References
 
preparation and characterization of lanthanide chelate–coated silica nanoparticles
Bare silica nanoparticles had spherical form and uniform size [55 ( (5)) nm in diameter, mean (SD)] (Fig. 2, A and B ). After APTMS treatment, we confirmed the existence of amino groups on the surfaces by observing salicylaldehyde-mediated yellow color. When Eu(III)-BHHCT was coated to silica nanoparticles, aggregation occurred immediately after mixing BHHCT with the nanoparticles in ethanol. The aggregated nanoparticles were dispersed by ultrasonic treatment. We used a multiple-round coating strategy to increase overall chelate loading. With Eu(III), maximal loading was achieved after 5 rounds of coating, probably owing to gradual saturation of all BHHCT linkage sites on the nanoparticle (some of these sites might have been blocked in the first 4 rounds of reaction because of aggregation). Coating of Tb(III)/BPTA did not result in any aggregation, and extra rounds of coating did not increase the overall fluorescence intensity.


Figure 2
View larger version (99K):
[in this window]
[in a new window]

 
Figure 2. TEM images.

(A), bare silica nanoparticles (25 000x magnification), (B), silica nanoparticle with amino groups (195 000x magnification), (C), Eu(III)-BHHCT–coated nanoparticles (195 000x magnification), and (D), Tb(III)-BPTA–coated nanoparticles (195 000x magnification).

The Eu(III)-BHHCT– and Tb(III)/BPTA–coated nanoparticles showed no change in size and shape compared with the bare silica nanoparticles, but both had clusters of black spots on the particle surface (Fig. 2, C and DUp ). The clustered distribution of Eu(III)-BHHCT on the silica nanoparticles functionalized with APTMS was similar to that of gold nanoparticles when adsorbed on the same functionalized surfaces (30). The fluorescence spectra of coated lanthanide chelates showed no difference from those of free chelates. The lifetimes of free and coated Eu(III)-BHHCT were 0.346 and 0.398 ms, respectively, whereas the lifetimes of the free and coated Tb(III)-BPTA were 2.645 and 2.224 ms. In contrast, the lifetime of Eu(III)-BHHCT encapsulated inside silica nanoparticles prepared by different methods was no longer than 0.300 ms (24), and the lifetime of Tb(III)-BPTA chelates decreased significantly [from 2.68 ms (free chelates) to 1.52 ms (encapsulated) (23)]. As lifetime change is an indicator of fluorescence quenching, the longer lifetime of lanthanide chelates indicated less fluorescence quenching with the surface-coated nanoparticles. Eu(III)-BHHCT even had increased lifetime after coating, suggesting that fluorescence enhancement might have occurred. Such an observation is consistent with multiple labeling of protein carriers (8)(9)(10).

The apparent number of Eu(III)-BHHCT per nanoparticle was calculated at 6.86 x 105, which is higher than that observed with commercial polystyrene nanoparticles of even larger size [3.1 x 104 chelates per nanoparticle of 107 nm in diameter (13)]. The apparent number of Tb(III)-BPTA per nanoparticle was 4.73 x 104, which is also significantly higher than that of physically encapsulated chelates in silica nanoparticles of the similar size [1500 chelates per silica nanoparticle of 42 nm in diameter (23)].

conjugation of antibody to lanthanide-coated silica nanoparticles
The existence of amino groups after lanthanide coating, which was detected by salicylaldehyde-mediated yellow color, allowed us to use a site-specific labeling strategy to conjugate antibody to Eu(III)-BHHCT–coated nanoparticles (Fig. 1BUp ). To achieve the highest sensitivity and specificity, we screened for the best combination of capture and detection antibody pair among 9 anti-HBsAg antibodies including 4 polyclonal antibodies and 5 monoclonal antibodies. The optimal antibody pair with HBsAg was monoclonal antibody S04 as the detection antibody and monoclonal B20 as the capture antibody. The labeling ratio was also optimized with this antibody pair, and we found that 0.8 mg antibody and 4 mg (0.051 nmol) nanoparticles gave the highest signal-to-noise ratio. Unlike Eu(III)-BHHCT–coated nanoparticles, direct conjugation of antibody to Tb(III)-BPTA–coated nanoparticles failed to give any meaningful results. 1-Ethyl-3-(3-dimethylaminopropyl)carbodiimide hydrochloride (EDC) chemistry used for BPTA conjugation might be responsible for this failure, because excess EDC might block part of the amino groups on the silica nanoparticles, making them less accessible to antibody conjugation. By using oxidized Dextran 500 as a neutral and hydrophilic arm, we successfully linked SA to the nanoparticles, which allowed subsequent conjugation of biotinylated anti-HBeAg antibody to nanoparticles (Fig. 1CUp ). The conjugates prepared were fairly stable. Conjugate stock solutions (10 mg/mL; i.e., 1.3 x 10–7 mol/L) could be kept at 4 °C for 2 years without obvious loss of detection sensitivity.

nonspecific binding of nanoparticle conjugates
Overall nonspecific binding of Eu(III)-BHHCT–coated nanoparticles was <0.002% (Table 1 ). The nonspecific binding of antibody-conjugated and nonconjugated Eu(III) nanoparticles showed no obvious difference between anti-HBsAg antibody–coated or SA-coated wells; however, the nonspecific binding on SA-coated wells was always lower than that of antibody-coated wells. When additional nanoparticles were added, SA-coated wells showed a slight increase in nonspecific binding, whereas antibody-coated wells showed a negligible increase. The nonspecific binding of Tb(III) nanoparticles onto antibody-coated wells was also <0.002%. The nonspecific binding of antibody conjugates prepared in the present work was comparable with chelate-doped nanospheres (21), though slightly higher than those delicately modified nanospheres with PEG-tethered chains (31).


View this table:
[in this window]
[in a new window]

 
Table 1. Nonspecific binding of Eu(III)-BHHCT–coated nanoparticles to anti-HBsAg–coated microtitration wells.

TRIFA of HBSAG
The dynamic range of the nanoparticle-based TrIFA of HBsAg was 0.05–111.1 µg/L (Fig. 3A ). The detection limit, defined as the concentration corresponding to a signal that is 3 SD above the zero calibrator, was 0.0092 µg/L. The CV along the entire concentration range was between 1.76% and 8.72% (n = 8). The analytical recovery was in the range of 99.1%–106.5%, CV 2.31%–4.40% (Table 2 ). In contrast, the linear range of ELISA was 0.2–5.0 µg/L, with a detection limit of 0.2 µg/L. When 338 human serum samples were analyzed with both TrIFA and ELISA, the cutoff value was defined as 2.1 times the mean signal of the negative control (Fig. 3B ). Based on the cutoff value, the 2 methods gave completely consistent results for all serum samples. Experiments with the reference panel of HBsAg also generated values as expected (data not shown).


Figure 3
View larger version (17K):
[in this window]
[in a new window]

 
Figure 3. TrIFA of HBsAg using Eu(III)-BHHCT–coated nanoparticles.

(A), calibration curves (solid symbols) and precision profiles (open symbols) of HBsAg assay using Eu(III)-BHHCT–coated nanoparticles. The CVs are expressed as variations of concentration of 8 replicas. (B), analysis of 338 human serum samples using TrIFA (left) and ELISA (right) for HBsAg in 4 experiments. The cutoff value was calculated as 2.1 times the mean signal of the negative control. The cutoff values for each experiment are as follows: a = 3955 counts, b = 3446 counts, c = 2920 counts, d = 2828 counts, e = 0.105 AU.


View this table:
[in this window]
[in a new window]

 
Table 2. Analytical recovery.

TRIFA of HBEAG
The linear range of the nanoparticle-based TrIFA of HBeAg was 15.0 to 1.11 x 104 NCU/L (Fig. 4A ). The detection limit was 10.0 NCU/L. The CVs along the whole concentration range were within 3.99% to 12.11% (n = 8). The analytical recovery was in the range of 95.1% to 105.0%, with CVs of 2.39% of 9.09% (Table 2Up ). In contrast, the linear range of ELISA was narrower (3.0 x 102 to 9.6 x 103 NCU/L), and the detection limit was higher (3.0 x 102 NCU/L). When 340 human serum samples were analyzed with both TrIFA and ELISA using 2.1 times the signal of the negative control as the cutoff value, 6 ELISA-negative samples gave positive signals by TrIFA (Fig. 4B ). To clarify this difference, these samples were subjected to ELISA tests for HBsAg; HBs, HBe, and HB core antibodies; and real-time PCR assay for hepatitis B virus DNA. Five of 6 samples were confirmed as truly hepatitis B virus-positive samples (see Table S1 in the online Data Supplement). One sample seemed to be false positive, although the result needs further investigation. These results demonstrated that TrIFA had higher sensitivity than ELISA in HBeAg detection. TrIFA of the reference panel of HBeAg also generated expected results (data not shown).


Figure 4
View larger version (16K):
[in this window]
[in a new window]

 
Figure 4. TrIFA of HBeAg using Tb(III)-BPTA–coated nanoparticles.

(A), calibration curves (solid symbols) and precision profiles (open symbols) for HBeAg assay using Tb(III)-BPTA–coated nanoparticles. The CVs are expressed as variations of concentration of 8 replicas. (B), analysis of 340 human serum samples using TrIFA (left) and ELISA (right) for HBeAg in 4 experiments. The cutoff value was calculated as 2.1 times the mean signal of the negative control. The cutoff values for each experiment are as follows: a = 20 810 counts, b = 26 354 counts, c = 25 417 counts, d = 22 510 counts, e = 0.105 AU.

simultaneous TRIFA of HBSAG and HBEAG
Before simultaneous TrIFA of HBsAg and HBeAg, we first investigated the potential interference between signals of Tb(III) and Eu(III) in simultaneous detection. Under the detection conditions for Tb(III), Eu(III) gave negligible interference signal. However, under the detection conditions for Eu(III), Tb(III) generated interference signal. We established a correction method (see Correction Methods and Fig. S1 in the online Data Supplement) and used it to build up universal calibration curves for simultaneous TrIFA under different conditions (see Fig. S2 in the online Data Supplement). We then studied the interference between the 2 antibody pairs and noticed that the capture antibody EB3 used in the single assay of HBeAg had strong nonspecific binding to the detection antibody S04 for the HBsAg assay, resulting in high background signals in the HBsAg detection. Thus we changed EB3 to KE in the simultaneous assay after reselection of antibody pairs. Although the substitution of KE for EB3 caused a slight decrease in the assay sensitivity of HBeAg, the interference between the 2 antibody pairs was eliminated. According to the calibration curves of the simultaneous TrIFA of HBsAg and HBeAg (see Fig. S2 in the online Data Supplement), the linear range was 0.11 to 65.41 µg/L for HBsAg and 33.0 to 8.1 x 103 NCU/L for HBeAg; the detection limit was 0.033 µg/L for HBsAg and 27.0 NCU/L for HBeAg. The CVs (n = 8) along the whole concentration range were <10% for both assays.

When 307 human serum samples (including the 6 samples that showed inconsistent results between individual TrIFA and ELISA in the HBeAg assay) were analyzed with simultaneous TrIFA as well as ELISA using the same cutoff value definition as above, concordant results were obtained for all the samples in HBsAg (see Table S2 in the online Data Supplement). The simultaneous assay results for HBeAg showed that, among those 6 differential samples, only 2 (no. 174 and 314) were detected as HBeAg positive in the simultaneous assay. The results of other samples were concordant with ELISA. These results showed that the sensitivity of the simultaneous assay was slightly lower than the individual assay (at least in the case of HBeAg), though still higher than ELISA.


   Discussions
Top
Abstract
Introduction
Materials and Methods
Results
Discussions
References
 
Nanoparticle-based TrIFA using lanthanide chelate labels provides an ultrasensitive tool for clinical detection. Taking advantage of both multiple labeling and versatile surface modification of silica nanoparticles, we have directly coated Eu(III) and Tb(III) chelates on the surface of nanoparticles. The particles prepared are successfully used in detection of HBsAg and HBeAg, with a detection limit that is lower than (or equivalent to) other nanoparticle-based TrIFAs and is ~20- to 30-fold lower than ELISA. Moreover, we successfully detected both antigens simultaneously in a single assay. Thus, the present labeling strategy may help lanthanide chelate–based fluorescent TrIFA gain more clinical applications.

Our covalent surface coating approach has many advantages over existing labeling strategies. First, compared with multiple labeling strategy using proteins (8) or polymers (12), our coating process is more straightforward and controllable. Silica nanoparticles are chemically inert, and the coating reaction can be carried out in either organic or water solutions depending on the solubility of the chelate used. Separation of finished nanoparticles from free chelates or unconjugated proteins can be simply conducted by centrifugation. Second, the surface coating approach obviates many drawbacks inherent in the encapsulation strategy currently used for loading lanthanide chelates to latex or silica nanoparticles. For example, covalent coating eliminates the potential leakage problem often encountered in physical doping/dying. So far, physical dying/doping is nearly exclusively used with latex nanoparticles and frequently applied in silica nanoparticle labeling. It has been shown in silica nanoparticles that chelate leakage compromises its performance as label in TrIFA (23)(24)(25). Although the leakage problem has not been well explored with latex particles, its potential harm to immunoassay cannot be ignored (32). Third, covalent surface labeling of silica nanoparticles allows flexibly adjusting/maximizing the number of chelates coated on each particle without jeopardizing the uniformity of the nanoparticles. The spongelike, porous nature of silica nanoparticles provides much more surface area than solid nanoparticles do, which allows increased numbers of chelates to be loaded on every silica nanoparticle. In contrast, in either physical doping (23)(24)(25) or covalent binding-copolymerization (22)(33), the amount of lanthanide chelates has to be delicately controlled to avoid irregular (or even no) nanoparticle formation, limiting the number of chelates that can be loaded on each nanoparticle. That may explain why our surface coating strategy allows more chelates per nanoparticle than the encapsulation methods (16)(22). Fourth, fluorescence quenching was not observed with our surface-coated chelates. In contrast, fluorescence quenching is readily observed with encapsulated chelates (23)(24). Lack of quenching further makes the effective chelate number on surface-coated nanoparticles higher than inside-encapsulated nanoparticles. Finally, despite that organic fluorophores encapsulated inside silica nanoparticles are more photostable than free fluorophores (27), the surface-coated lanthanide chelates are also very stable for routine use. Such an observation may be because lanthanide chelates themselves are more photostable than organic fluorophores (23)(25) or because in TRF detection, the pulsed excitation light resource causes less photobleaching than constant irradiation in conventional fluorometric detection.

Despite many advantages shown by lanthanide chelate–coated silica nanoparticles, whether they can achieve high sensitivity in immunoassays also depends on many other factors. For example, the binding reagents used, the experimental conditions, and nonspecific binding are all crucial to the sensitivity of the immunoassay, in addition to the delectability of the label (9)(34). On screening antibody pairs, optimizing antibody conjugation format, evaluating nonspecific binding of the conjugated antibodies, and interference correction, ultrahigh sensitivity for both individual and simultaneous detection of HBsAg and HBeAg was realized. Although these factors may mean increased labor intensity for development of a TrIFA assay, it is worthwhile to consider them to achieve the best sensitivity. The low nonspecific binding of the antibody conjugates further confirms the advantages of the hydrophilicity of silica nanoparticles. Through these simple optimizations, our particles give a detection limit for HBsAg of 0.0092 µg/L, which is slightly lower than latex nanoparticles (0.028 µg/L) (17) and silica nanoparticles (0.023 µg/L) (22). The high sensitivity is especially meaningful for HBeAg, because it is an indicator of hepatitis B virus replication and can remain undetected by classic ELISA in certain samples.

We also established direct, simultaneous detection of 2 analytes using Eu(III) and Tb(III) chelate–coated silica nanoparticles. By using a novel correction algorithm, nearly the same sensitivity was achieved in the simultaneous assays as in the individual assays. Although simultaneous detection using Eu(III) and Tb(III) chelates has been previously achieved by stepwise detections (35)(36), to our knowledge, this is the first time that simultaneous detection was achieved in a single assay. Other simultaneous detection formats have been reported utilizing Eu(III) and Sm(III) chelates (37)(38). However, because Sm(III) chelate has rather low fluorescence yield and relatively short fluorescence lifetime, the Eu(III) and Sm(III) dual-label TrIFA has much lower sensitivity than the individual Eu(III) label assay. Thus, our dual-labeling strategy may allow more dual analysis to be performed simultaneously.

In conclusion, we developed a multiple labeling strategy for coating fluorescent lanthanide chelates on the surface of silica nanoparticles for TrIFA. This strategy was successfully demonstrated in both individual and simultaneous detection of HBsAg and HBeAg, with the detection limit lower than existing labeling approaches. Validation with a large number of clinical samples indicates that our labeling strategy may serve as a solid step toward the clinical use of nanoparticle TRF-based immunoassays.


   Acknowledgments
 
Grant funding/support: This work was partially supported by the Xiamen Scientific Development Program (no. 3502Z20055008).

Financial disclosures: None declared.

Acknowledgments: We are grateful to Xilin Zhao, Yongyou Zhang, and Jinping Cheng for critical reading of the manuscript and Jieli Zhang for technical assistance. We are also indebted to Dr. H. Matsumoto and Dr. Jingli Yuan for kindly providing BHHCT and BPTA.


   Footnotes
 
1 Nonstandard abbreviations: TRF, time-resolved fluorometry; SA, streptavidin; TrIFA, time-resolved immunofluorometric immunoassay; HBsAg, hepatitis B surface antigen; HBeAg, hepatitis B e antigen; APTMS, (3-aminopropyl)trimethoxylsilane; BHHCT, 4,4-bis(1,1,2,2,3,3-heptafluoro-4,6-hexanedion-6-yl)chlorosulfo-o-terphenyl; BPTA, N,N,N1, N1-[2,6-bis(3-aminomethyl-1-pyrazolyl)-phenylpyridine] tetrakis; EDC, 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide hydrochloride; NCU, National Centre Unit.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussions
References
 

  1. Hemmila I. Fluoroimmunoassays and immunofluorometric assays. Clin Chem 1985;31:359-370.[Abstract/Free Full Text]
  2. Soini E, Lovgren T. Time-resolved fluorescence of lanthanide probes and applications in biotechnology. Anal Chem 1987;18:105-154.
  3. Diamandis EP, Christopoulos TK. Europium chelate labels in time-resolved fluorescence immunoassays and DNA hybridization assays. Anal Chem 1990;62:1149A-1157A.[Medline] [Order article via Infotrieve]
  4. Hemmila I, Laitala V. Progress in lanthanides as luminescent probes. J Fluorescence 2005;15:529-542.[CrossRef]
  5. Yuan J, Wang G. Lanthanide complex-based fluorescence label for time-resolved fluorescence bioassay. J Fluorescence 2005;15:559-568.[CrossRef]
  6. Hemmila I, Dakuba S, Mukkala V-M, Siitari H, Lovgren T. Europium as a label in time-resolved immunofluorometric assays. Anal Biochem 1984;137:335-343.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  7. Christopoulos TK, Diamandis EP. Enzymatically amplified time-resolved fluorescence immunoassay with terbium chelates. Anal Chem 1992;64:342-346.[Medline] [Order article via Infotrieve]
  8. Morton RC, Diamandis EP. Streptavidin-based macromolecular complex labeled with a europium chelator suitable for time-resolved fluorescence immunoassay applications. Anal Chem 1990;62:1841-1845.
  9. Diamandis EP. Signal amplification in time-resolved fluorometry. Clin Chem 2001;47:380-381.[Free Full Text]
  10. Diamandis EP. Multiple labeling and time-resolvable fluorophores. Clin Chem 1991;37:1486-1491.[Abstract/Free Full Text]
  11. Scorilas A, Diamandis EP. Polyvinylamine-streptavidin complexes labeled with a europium chelator: a universal detection reagent for solid-phase time resolved fluorometric applications. Clin Biochem 2000;33:345-350.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  12. Scorilas A, Bjartell A, Lilja H, Moller C, Diamandis EP. Streptavidin-polyvinylamine conjugates labeled with a europium chelate: applications in immunoassay, immunohistochemistry, and microarrays. Clin Chem 2000;46:1450-1455.[Abstract/Free Full Text]
  13. Harma H, Soukka T, Lovgren T. Europium nanoparticles and time-resolved fluorescence for ultrasensitive detection of prostate-specific antigen. Clin Chem 2001;47:561-568.[Abstract/Free Full Text]
  14. Huhtinen P, Soukka T, Lovgren T, Harma H. Immunoassay of total prostate-specific antigen using europium(III) nanoparticle labels and streptavidin-biotin technology. J Immunol Methods 2004;294:111-122.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  15. Huhtinen P, Pelkkikangas AM, Jaakohuhta S, Lovgren T, Harma H. Quantitative, rapid europium(III) nanoparticle-label-based all-in-one dry-reagent immunoassay for thyroid-stimulating hormone. Clin Chem 2004;50:1935-1936.[Free Full Text]
  16. Huhtinen P, Kivela M, Kuronen O, Hagren V, Takalo H, Tenhu H, et al. Synthesis, characterization, and application of Eu(III), Tb(III), Sm(III), and Dy(III) lanthanide chelate nanoparticle labels. Anal Chem 2005;77:2643-2648.[Medline] [Order article via Infotrieve]
  17. Valanne A, Huopalahti S, Vainionpaa R, Lovgren T, Harma H. Rapid and sensitive HBsAg immunoassay based on fluorescent nanoparticle labels and time-resolved detection. J Virol Methods 2005;129:83-90.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  18. Valanne A, Huopalahti S, Soukka T, Vainionpaa R, Lovgren T, Harma H. A sensitive adenovirus immunoassay as a model for using nanoparticle label technology in virus diagnostics. J Clin Virol 2005;33:217-223.[Web of Science][Medline] [Order article via Infotrieve]
  19. Soukka T, Antonen K, Harma H, Pelkkikangas AM, Huhtinen P, Lovgren T. Highly sensitive immunoassay of free prostate-specific antigen in serum using europium(III) nanoparticle label technology. Clin Chim Acta 2003;328:45-58.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  20. Soukka T, Harma H, Paukkunen J, Lovgren T. Utilization of kinetically enhanced monovalent binding affinity by immunoassays based on multivalent nanoparticle-antibody bioconjugates. Anal Chem 2001;73:2254-2260.[Medline] [Order article via Infotrieve]
  21. Soukka T, Paukkunen J, Harma H, Lonnberg S, Lindroos H, Lovgren T. Supersensitive time-resolved immunofluorometric assay of free prostate-specific antigen with nanoparticle label technology. Clin Chem 2001;47:1269-1278.[Abstract/Free Full Text]
  22. Tan MQ, Wang GL, Hai XD, Ye ZQ, Yuan JL. Development of functionalized fluorescent europium nanoparticles for biolabeling and time-resolved fluorometric applications. J Mater Chem 2004;14:2896-2901.[CrossRef]
  23. Ye Z, Tan M, Wang G, Yuan J. Preparation, characterization, and time-resolved fluorometric application of silica-coated terbium(III) fluorescent nanoparticles. Anal Chem 2004;76:513-518.[Medline] [Order article via Infotrieve]
  24. Ye ZQ, Tan MQ, Wang GL, Yuan JL. Novel fluorescent europium chelate-doped silica nanoparticles: preparation, characterization and time-resolved fluorometric application. J Mater Chem 2004;14:851-856.[CrossRef]
  25. Tan MQ, Ye ZQ, Wang GL, Yuan JL. Preparation and time-resolved fluorometric application of luminescent europium nanoparticles. Chem Mater 2004;16:2494-2498.[CrossRef][Web of Science]
  26. Bagwe RP, Yang CY, Hilliard LR, Tan WH. Optimization of dye-doped silica nanoparticles prepared using a reverse microemulsion method. Langmuir 2004;20:8336-8342.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  27. Santra S, Wang KM, Tapec R, Tan WH. Development of novel dye-doped silica nanoparticles for biomarker application. J Biomed Optics 2001;6:160-166.[CrossRef]
  28. Waddell TG, Leyden DE, DeBello MT. The nature of organosilane to silica-surface bonding. J Am Chem Soc 1981;103:5303-5307.[CrossRef][Web of Science]
  29. Qian WP, Xu B, Wu L, Wang CX, Yao DF, Yu F, et al. Controlled site-directed assembly of antibodies by their oligosaccharide moieties onto APTES derivatized surfaces. J Colloid Interface Sci 1999;214:16-19.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  30. Westcott SL, Oldenburg SJ, Lee TR, Halas NJ. Formation and adsorption of clusters of gold nanoparticles onto functionalized silica nanoparticle surfaces. Langmuir 1998;14:5396-5401.[CrossRef][Web of Science]
  31. Matsuya T, Tashiro S, Hoshino N, Shibata N, Nagasaki Y, Kataoka K. A core-shell-type fluorescent nanosphere possessing reactive poly(ethylene glycol) tethered chains on the surface for zeptomole detection of protein in time-resolved fluorometric immunoassay. Anal Chem 2003;75:6124-6132.[Medline] [Order article via Infotrieve]
  32. Hakala H, Mukkala V-M, Sutela T, Hovinen J. Synthesis and properties of nanospheres copolymerised with luminescent europium(III) chelates. Org Biomol Chem 2006;4:1383-1386.[CrossRef][Web of Science]
  33. Hai X, Tan M, Wang G, Ye Z, Yuan J, Matsumoto K. Preparation and a time-resolved fluoroimmunoassay application of new europium fluorescent nanoparticles. Anal Sci 2004;20:245-246.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  34. Jackson TM, Ekins RP. Theoretical limitations on immunoassay sensitivity. Current practice and potential advantages of fluorescent Eu3+ chelates as non-radioisotopic tracers. J Immunol Methods 1986;87:13-20.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  35. Yuan J, Wang G, Majima K, Matsumoto K. Synthesis of a terbium fluorescent chelate and its application to time-resolved fluoroimmunoassay. Anal Chem 2001;73:1869-1876.[Medline] [Order article via Infotrieve]
  36. Eriksson S, Vehniainen M, Jansen T, Meretoja V, Saviranta P, Pettersson K, et al. Dual-label time-resolved immunofluorometric assay of free and total prostate-specific antigen based on recombinant Fab fragments. Clin Chem 2000;46:658-666.[Abstract/Free Full Text]
  37. Leinonen J, Lovgren T, Vornanen T, Stenman UH. Double-label time-resolved immunofluorometric assay of prostate-specific antigen and of its complex with alpha 1-antichymotrypsin. Clin Chem 1993;39:2098-2103.[Abstract]
  38. Matsumoto K, Yuan JG, Wang GL, Kimura H. Simultaneous determination of alpha-fetoprotein and carcinoembryonic antigen in human serum by time-resolved fluoroimmunoassay. Anal Biochem 1999;276:81-87.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]



The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
X. Xia, Y. Xu, X. Zhao, and Q. Li
Lateral Flow Immunoassay Using Europium Chelate-Loaded Silica Nanoparticles as Labels
Clin. Chem., January 1, 2009; 55(1): 179 - 182.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow 078485.Supplemental Data
Right arrow All Versions of this Article:
clinchem.2006.078485v1
53/8/1503    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Xu, Y.
Right arrow Articles by Li, Q.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Xu, Y.
Right arrow Articles by Li, Q.
Related Collections
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS