General Clinical Chemistry |
1 Analytical Chemistry Division and 2 Statistical Engineering Division, National Institute of Standards and Technology, Gaithersburg, MD.
aAddress correspondence to this author at: National Institute of Standards and Technology, 100 Bureau Dr., Stop 8392, Gaithersburg, MD 20899-8392. Fax 301-977-0685; e-mail nathan.dodder{at}nist.gov.
| Abstract |
|---|
|
|
|---|
Methods: The SRM was produced from 2 serum pools with different creatinine concentrations. The concentrations were certified using a higher-order isotope-dilution GC-MS method and an isotope-dilution LC-MS method. The LC-MS method is a potential higher-order reference measurement procedure.
Results: The GC-MS mean (CV) concentrations were 67.0 (0.9%) µmol/L for serum pool 1 and 346.1 (0.45%) µmol/L for serum pool 2. The LC-MS results were 66.1 (0.2%) µmol/L and 346.3 (0.2%) µmol/L, respectively. For serum pool 1, there was a 1.4% difference between the mean GC-MS and LC-MS measurements, and a 0.10% difference for serum pool 2. The results from the 2 methods were combined to give the certified concentrations and expanded uncertainties.
Conclusions: The certified concentration (expanded uncertainty) of SRM 967 was 66.5 (1.8) µmol/L for serum pool 1 (a value close to the diagnostically important concentration of 88.4 µmol/L) and 346.2 (7.4) µmol/L for serum pool 2 (a concentration corresponding to that expected in a patient with chronic kidney disease).
| Introduction |
|---|
|
|
|---|
The NKDEP recommended that a serum reference material with a creatinine concentration of 88.4 µmol/L (1.00 mg/dL) be developed (3). This value corresponds to a glomerular filtration rate (GFR) of approximately 60 mL · min–1 · 1.73 m–2. CKD is defined as a GFR of <60 mL · min–1 · 1.73 m–2 for
3 months (3). The GFR describes the ability of the kidneys to filter waste products from the blood and is estimated based on the patients serum creatinine concentration, age, sex, and race. Therefore, 88.4 µmol/L is near the critical concentration that would determine a positive or negative diagnosis. Also, compared with higher creatinine concentrations, errors associated with the calibration or measurement precision of creatinine at this relatively low concentration will have a greater impact on the error of the estimated GFR (3).
To meet the recommendations set by the NKDEP, NIST has developed standard reference material (SRM) 967, Creatinine in Frozen Human Serum (4). This SRM consists of 2 serum pools with target creatinine concentrations of approximately 88.4 µmol/L (the diagnostically critical concentration) and 354 µmol/L (4 mg/dL, a concentration corresponding to that in a patient with CKD). The creatinine concentrations were certified using 2 independent methods. The isotope-dilution gas chromatography–mass spectrometry (GC-MS) method (5) is considered to be a higher-order reference measurement procedure by the NCCLS (6) and the Joint Committee on Traceability in Laboratory Medicine (JCTLM) (7). The isotope-dilution LC-MS method is similar to a procedure developed at the Laboratory of the Government Chemist that was approved by the JCTLM as a higher-order reference measurement procedure (7)(8). The LC-MS method has fewer sample preparation steps and is a higher-throughput method than the GC-MS method. The commutability of SRM 967 was then verified in a study organized by the NKDEP.
SRM 967 has a frozen serum matrix. Frozen serum more closely matches the native state of clinical samples than a lyophilized matrix. Reference materials supplied by other metrology institutes are either lyophilized (9) or do not have the target creatinine concentrations described above (10).
| Materials and Methods |
|---|
|
|
|---|
preparation of srm 967
The pools of human serum used for SRM 967 were prepared by Solomon Park Research Laboratories.
2
Blood was collected from healthy, postmenopausal, adult women following CLSI guidelines (13). The resulting serum master pool of approximately 3 L was split into 2 pools. Pool 1 was not enriched with additional creatinine. Pool 2 was enriched with an appropriate amount of reagent-grade creatinine to bring the concentration up to approximately 354 µmol/L. Both pools were passed through filters with a 0.2-µm pore size. No preservatives were added. One-milliliter aliquots of the pools were placed in 3-mL amber glass vials and capped with Teflon stoppers and aluminum seals. The vials were frozen at –80 °C until analysis.
preparation of the calibration standards
Certification of the creatinine concentrations in SRM 967 was performed at NIST. We made calibration solutions that contained known unlabeled:labeled creatinine mass ratios and internal standard solutions that contained known masses of labeled creatinine. The internal standard solutions were added to the samples at the beginning of the sample preparation; the mass of the added labeled creatinine was approximately equal to the mass of unlabeled creatinine in the sample. To achieve this 1:1 ratio, we performed a preliminary quantification in which a wider range of mass ratios was used. Once the approximate creatinine concentration was measured, the quantity of internal standard necessary for a 1:1 ratio was calculated. After sample processing, we ran the calibration standards and samples in the same set on the mass spectrometer. The unlabeled:labeled creatinine peak area ratios in the samples were converted to mass ratios using data from the calibration standard runs and either a bracketing method or linear regression method, as described below. The mass ratios were then solved for the mass of the unlabeled creatinine, and the concentration of unlabeled creatinine in each sample was calculated.
For each analytical method, we gravimetrically prepared an independent stock solution of labeled creatinine to make the calibration standards and sample internal standard solutions. We weighed approximately 1.25 mg labeled creatinine (either creatinine-13C2 or creatinine-d3) into a 50-mL volumetric flask, added 50 mL water, and calculated the concentration of the solution (approximately 0.025 mg/g). This solution was split into 3 sets, each set containing a different mass of labeled creatinine. One set was used as internal standards for the samples from serum pool 1, the 2nd set as internal standards for the samples from serum pool 2, and the 3rd set to prepare the calibration standards. The internal standard aliquots were stored at –20 °C.
The accuracy of the quantification was limited by the accuracy of the mass of unlabeled creatinine in the calibration standards. To test for bias in the GC-MS calibration standards, 2 independent solutions of unlabeled creatinine were gravimetrically prepared as follows. We weighed approximately 20 mg solid creatinine SRM 914a (14) into a volumetric flask and added 20 mL water. We transferred 2.5 mL of the solution to a 100-mL volumetric flask and added water to a final volume of 100 mL. We then calculated the concentration (approximately 0.025 mg/g) of the stock solution.
We made calibration standards from both unlabeled stock solutions such that the unlabeled:labeled creatinine mass ratios of the 2 sets of calibration standards were offset from each other by 0.1 units and ranged from 0.8 to 1.2. To test for bias, we ran the calibration standards as a single set. The linearity of the resulting calibration curve confirmed that the 2 independently prepared calibration standard sets were not biased. The calibration standards for the GC-MS method were derivatized as described below and reconstituted in hexane to the same concentration as the GC-MS samples, approximately 10 mg/L, and stored at –20 °C until analysis. The calibration standards for the LC-MS method were made in a similar manner, except we prepared an independent solution of unlabeled creatinine for each of the 3 sample sets. We diluted the LC-MS calibration standards with 10 mmol/L ammonium acetate to the same concentration as the LC-MS samples, approximately 1.6 mg/L, and stored them at –20 °C until analysis.
quality control samples
SRM 909b Human Serum, which had been previously certified for creatinine concentrations (15)(16), was analyzed along with SRM 967 to confirm the accuracy of the analysis. SRM 909b consists of lyophilized serum (2 pools), and each vial was reconstituted with 10.00 mL water and allowed to equilibrate for 1.5 h before being prepared along with SRM 967.
quantification by gc-ms
Two aliquots from each SRM 967 vial and 1 aliquot from each SRM 909b vial were measured. The aliquots were added gravimetrically to the creatinine-13C2 internal standard solutions and equilibrated overnight at 5 °C. Ion-exchange chromatography was necessary to separate creatine from creatinine before derivatization, because these 2 compounds will form the same derivative. Amberlite IRC-50 ion-exchange resin (Chemical Dynamics) was washed in water and soaked in 1.0 mol/L HCl for 3.5 h with occasional agitation. The resin was then rinsed with water and stored in excess 0.1 mol/L HCl until use. The resin was slurry packed into 20 cm by 10-mm columns using water. The volume of resin in each column was 5 mL. The resin was washed with 150 mL water, then the samples were added to the columns. We eluted the creatine with 75 mL water; this fraction was discarded. We eluted the creatinine with 75 mL of 1.0 mol/L ammonium hydroxide. We used a separate set of samples to measure the densities using the Lang–Levy pipette method (17).
The derivatization reaction required the samples to be completely free of water. To remove the water and ammonium hydroxide, the samples were freeze dried and then reconstituted in 100% ethanol. We removed the solid residue by passing the samples through polyvinylidene fluoride syringe filters (25-mm diameter, 0.45-µm pore size). The samples were evaporated to near dryness under vacuum, and the creatinine was converted into a derivative according to the described procedure (5). The samples were then solvent exchanged into hexane. This brought the final concentration of the samples to approximately 10 mg/L. The samples were stored at –20 °C until analysis.
We performed the measurements by use of an Agilent 5972 GC-MS. The injection volume was 1 µL. The GC column was 30-m long, with a 0.25-mm internal diameter, and a 0.25-µm thick DB-5ms stationary phase (Agilent). The GC oven temperature program was 130 °C for 2 min, 12 °C/min to 250 °C, 250 °C for 0.5 min. The mass spectrometer was operated in the electron impact ionization mode with selected ion monitoring of the [M-73]+ ions at m/z 150 and 152 for the unlabeled and labeled forms, respectively.
We ran 3 separately prepared sets of samples on the GC-MS. Each set consisted of 10 samples: 4 samples of SRM 967 pool 1, 4 samples of SRM 967 pool 2, 1 sample of SRM 909b pool 1, and 1 sample of SRM 909b pool 2. Calibration was by bracketing, i.e., each sample was measured in duplicate, in between duplicate measurements of the 2 calibration standards with unlabeled:labeled peak area ratios just below and above that of the sample. We calculated the mass ratios by linear interpolation between the bracketing standards for each sample. We then repeated the measurements on a 2nd day, with the order of the standards reversed. The results of the 2-day measurements were averaged to arrive at the mass ratios from which the creatinine concentrations in the samples were calculated.
quantification by lc-ms
Two aliquots from each SRM 967 vial and 1 aliquot from each SRM 909b vial were measured. Aliquots of SRM 967 and SRM 909b were added gravimetrically to the creatinine-d3 internal standard solutions and equilibrated overnight at 5 °C. The proteins were precipitated by adding 3 volumes of ice-cold ethanol to each cold tube and vortex-mixing. After standing for 5 min, the samples were centrifuged at 900g for 20 min. The supernatant, containing the creatinine, was removed and concentrated to dryness using a N2 stream. Each sample was reconstituted in 1 mL water and filtered through a polyvinylidene fluoride syringe filter (13-mm diameter, 0.2-µm pore size). We diluted the samples to a concentration of approximately 1.6 mg/L with 10 mmol/L ammonium acetate and stored them at –20 °C until analysis.
We performed the measurements by use of an Agilent 1100 series LC-MS. The injection volume was 4 µL, corresponding to approximately 6 ng creatinine. The liquid chromatography column was a 15-cm-long, 2.0 mm internal diameter, 3 µm particle diameter, LUNA C18 (2) (Phenomenex). The gradient mobile phase program was 10 mmol/L ammonium acetate for 7 min, ramped to 20% 10 mmol/L ammonium acetate and 80% acetonitrile by 7.1 min, and held for 13 min. The flow rate was 0.2 mL/min. The column temperature was 23 °C. The mass spectrometer was operated using positive mode electrospray ionization and selective ion monitoring of the (M+H)+ ions at m/z 114 and 117 for creatinine and creatinine-d3, respectively. The ionization source parameters were drying gas temperature 350 °C, N2 gas flow 12 L/min, nebulizer pressure 170 kPa (25 psi), capillary 1500 V, and fragmentor 120 V.
We ran 3 separately prepared sets of samples on the LC-MS. Each set consisted of 14 samples: 6 samples of SRM 967 pool 1, 6 samples of SRM 967 pool 2, 1 sample of SRM 909b pool 1, and 1 sample of SRM 909b pool 2. Each set was run as follows: the 5 calibration standards were run 1st; followed by the samples; then the samples were measured again in the reverse order; and last, the 5 calibration standards were run in reverse order. We calculated a composite linear regression, using a slope-intercept model, from the peak areas of the calibration standards. We used the linear calibration to calculate the mass ratios using the average of the duplicate sample peak area measurements, which we then used to calculate the creatinine concentration in the samples.
| Results |
|---|
|
|
|---|
|
|
Among 12 GC-MS measurements, the largest was 69.2 µmol/L. This point was identified as an outlier by Grubb and Dixon tests and excluded from the statistical analysis and calculation of the certified concentrations. The GC-MS mean concentrations (SD) for SRM 967 were 67.0 (0.6) µmol/L for pool 1 and 346.1 (1.6) µmol/L for pool 2. The LC-MS results were 66.1 (0.2) µmol/L for pool 1 and 346.3 (0.8) µmol/L for pool 2. For pool 1, there was a 1.4% difference between the mean GC-MS and LC-MS measurements, and a 0.10% difference for pool 2. There was no evidence of inhomogeneity within or among the vials, or of a concentration trend corresponding to the vial filling order. The results of the control measurements are listed in Table 2
and were within 1.0% of the certified values for SRM 909b.
|
| Discussion |
|---|
|
|
|---|
calculation of certified concentrations and expanded uncertainties
The results from the 2 methods were combined using a Bayesian approach (18)(19)(20). This approach, intended for certifying data from a small number of analytical methods, assumes that both the means and the variances of the methods could be different. The means were combined by Eq. 1
, where
is the combined mean and c1 and c2 are the means from the 2 methods. The combined mean is the certified concentration.
![]() | (1) |
![]() | (2) |
, for each serum pool, to account for undetected interferences and the uncertainty in the purity of the reference standard SRM 914a (other sources of type B uncertainty were considered negligible). The type A and B uncertainties were combined using Eq. 3
![]() | (3) |
± U) that is expected to encompass a large fraction (approximately 95%) of the measurements obtained by subsequent analyses. Table 3
|
commutability validation
A commutability validation study was organized by the NKDEP. Commutability refers to the ability of the SRM to give similar results to real patient samples when analyzed by different analytical methods. The experimental design followed a protocol recommended by the CLSI (22). Briefly, creatinine was measured in SRM 967 and individual patient serum samples using routine laboratory methods and the NIST LC-MS method described above. SRM 967 was found to be commutable with 15 methods from 7 in vitro diagnostic equipment manufacturers (23).
| Acknowledgments |
|---|
Financial disclosures: None declared.
Acknowledgments: We thank Karen Phinney, Mary Satterfield, Katherine Sharpless, and Stephen Wise, all from NIST, for reviewing the manuscript.
| Footnotes |
|---|
2 Certain commercial instruments and materials are identified in this report to adequately specify the experimental procedures. Such identification does not imply endorsement by the National Institute of Standards and Technology, nor does it imply that the instruments and materials identified are the best available for the purpose. ![]()