|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Automation and Analytical Techniques |
1 Department of Laboratory Medicine, Min-Sheng General Hospital, Taoyuan, Taiwan; 2 Department of Core Technical Research, Bionime Corporation, Taichung, Taiwan; 3 IMCARMED GmbH, Saalfeld, Germany.
aAddress correspondence to this author at: Department of Core Technical Research, Bionime Corporation, Taichung, Taiwan. Fax +886-4-24952568; e-mail brown.hsu{at}bionime.com.
| Abstract |
|---|
|
|
|---|
Methods: During the sensor-fabrication process, barrel-plated gold electrodes were inserted into an injection-molded base. The electrode directly touched the electrical contact of a meter. We analyzed technical measurements for this biosensor, including tests of the measurement range, within-run imprecision, and between-meter imprecision. In clinical trials, experienced technicians tested 3 alternative sites (fingertip, palm, and arm). The results were simultaneously compared with plasma values obtained with the hexokinase method on the Olympus AU640 instrument. Analytical results were evaluated according to International Standards Organization 15197 (ISO 15197:2003) criteria and by Clarke error grid analysis (EGA), and CVs were calculated to evaluate within-run imprecision.
Results: The glucose measurement range was 0.6– 33.3 mmol/L (y = 0.96x + 0.07 mmol/L; r2 = 0.9977). The CVs in the within-run imprecision test were 1.7%–3.5%, and the overall CV was 2.1%, indicating good reproducibility of results. The Student t-tests of mean values from 5 meters revealed statistically insignificant differences (P > 0.05). In clinical trials, the agreement of the Rightest GM310 meter results with those of a laboratory method complied with ISO 15197:2003 criteria. In the EGA, 100% of the values were within the acceptable zones (A + B), and the proportion of values within zone A exceeded 95%.
Conclusions: The Bionime Rightest GM310 meter applied a simplified process for biosensor fabrication and displayed acceptable performance for monitoring glucose concentrations at alternative test sites.
| Introduction |
|---|
|
|
|---|
Pretreating SPCEs can enhance the electrochemical properties of carbon electrodes, including electrochemical response, response time, cyclical voltammetric behavior, and mechanical stability(4)(12). Therefore, such processes as electroplating of copper(13) and sputtering of palladium and gold(11)(14) have frequently been used to modify electrodes and thereby accelerate electron-transfer kinetics and reduce the potential required. Fabrication processes applied to most SPCE-based devices are extremely complex, however, and thus a simple process is required. This study investigated a novel technique for fabricating amperometric glucose biosensors that involved inserting barrel-plated gold electrodes into hollow holes in the injection-molded base, followed by immobilization of chemical reagents on a U-shaped furrow. This technique has the potential for producing manufacturing costs as low as those for the SPCE technique. Furthermore, gold particles generate large and specific areas that can immobilize most glucose oxidase molecules on the biosensing interface, and the excellent conductivity of gold accelerates electron transfer(14)(15)(16). This novel electrode-manufacturing technique for amperometric biosensors has been patented in the US(17).
Although commercially available glucose meters have been developed for alternative-site testing, the stability of measurements has been an ongoing topic of study in arm tests(18)(19). The aim of this study was to characterize and evaluate the clinical performance of the Bionime Rightest GM310 meter for alternative-site tests (e.g., fingertips, palms, and arms) with novel test strips fabricated by means of a one-step electrode-manufacturing process.
| Materials and Methods |
|---|
|
|
|---|
blood glucose meters
This study used Rightest SMBG system GM310 (Bionime) and a new test strip. The meter used amperometric electrochemistry to measure capillary blood glucose concentrations. The meter required a 0.6-µL blood sample and reported a venous plasma–equivalent glucose concentration in 5 s. The test strip was 40 mm long and 9 mm wide. The test strip consisted of an injection-molded base, enzyme-modified gold-plated electrodes, chemical reagent, and a cover film (Fig. 1
). The test strips were from the same batch (lot no. 1178062).
|
test procedures
The analytical measurement range was evaluated by assaying venous blood samples spiked with various glucose concentrations (range, 0.6–33.3 mmol/L) in heparin-containing tubes. Four replicates of each sample were tested with the meters. We centrifuged the samples, tested for the plasma glucose concentration in duplicate, and compared the results with those obtained with a glucose oxidase procedure on the YSI 2300 STAT Blood Glucose & Lactate Analyzer (Yellow Springs Instruments). Mean glucose values were calculated, and the meter results were plotted against the results obtained with the comparison method. We used 6 glucose concentrations (1.11–16.66 mmol/L) in venous blood samples with one meter to estimate within-run imprecision. The test procedure was designed on the basis of International Standards Organization (ISO) 15197:2003 criteria(20). In addition, we assessed between-meter imprecision with 5 meters; all meters measured the same venous blood sample (glucose concentration approximately 5 mmol/L) simultaneously in 10 replicates with the same batch of test strips. The imprecision in the mean values among the 5 meters was analyzed statistically with the paired Student t-test.
Supplementary Fig. 1 (see the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol54/issue10) describes the clinical trial procedures. Trained technicians and nurses performed measurements according to the user manual with the meter at room temperature. Manufacturer-recommended methods were used to obtain blood samples from the fingertip, palm, and arm. We used the hexokinase method on an Olympus AU640 instrument (Olympus America) as a comparison method. The clinical trial was funded by Bionime Technology, Taiwan.
statistical and clinical analyses
The clinical acceptability of the results obtained with the SMBG meter was evaluated by comparing them with those obtained with the Olympus AU640 method. We adopted 3 procedures for this analysis: (a) linear regression analysis, (b) Clarke error grid analysis (EGA), and (c) bias plots of accuracy criteria in accordance with ISO 15197:2003.
The EGA criteria specify that 95% of measurement results must fall in zone A and that 100% must fall in zones A plus B. The 5 zones in the EGA are based on their clinical acceptability. Measurements within zones A and B are clinically acceptable. Values in zone C lead to overcorrection of true glucose values. Values in zone D indicate the potential for dangerous treatments. Values in zone E may result in erroneous treatment. The EGA graph plots comparison-method values on the x axis and glucose values obtained with the meter on the y axis.
The primary objectives of the international standards (ISO 15197) are to establish requirements that result in acceptable performance. Performance criteria for blood glucose–monitoring systems were established on the basis of the accuracy (precision and trueness) required for individual glucose results. The accuracy determined by the degree of agreement between values obtained with blood glucose meters and laboratory plasma values was assessed further according to the criteria in ISO 15197:2003. The minimum acceptable performance for glucose-monitoring meters has been stipulated by ISO 15197:2003(20): 95% of the individual glucose results must be within ±0.83 mmol/L of the comparison-method results at glucose concentrations <4.20 mmol/L and within ±20% of the comparison-method results at glucose concentrations
4.20 mmol/L.
| Results |
|---|
|
|
|---|
|
The imprecision results are shown in Table 1
. Within-run CVs were 1.7%–3.5%, and the overall CV was 2.1%, excluding the SD value of 0.08 mmol/L for extra-low glucose concentrations (Table 1
). For between-meter tests, differences in mean values were not statistically significant (P > 0.05, paired t-test). The maximum mean difference among the 5 meters was 0.08 mmol/L or a 1.6% variation (Table 1
).
|
clinical evaluation of rightest blood glucose–monitoring system gm310
The capillary blood glucose concentration was measured at the fingertip, the palm, and the arm. Within 5 min after values were obtained with the Rightest GM310, we drew a venous sample from the same individual, centrifuged the sample for 10 min, and analyzed it within 1 h at the Min-Sheng General Hospital with an Olympus AU640 chemistry analyzer. Supplementary Fig. 1 in the online Data Supplement shows the flow chart of this clinical trial.
The range of blood glucose concentrations for the 139 study participants was 1.6–26.7 mmol/L, and hematocrits met the manufacturers criteria (30%–55%). The agreement between the Rightest GM310 results and the comparison-method results (Olympus AU640) was clinically satisfactory (Table 2
; Fig. 3
).
|
|
In addition to fingertip tests, this study measured values from palm and arm sites. Although 139 measurements from the participants arms produced 5 results that failed to fall within the ±20% interval of reference values, the agreement of the values for the Rightest GM310 and the comparison method was clinically acceptable.
An evaluation of subsets of measurements for the 3 alternative sites for the comparison interval of <4.20 mmol/L showed that 26 of 26 values were within the ±0.83-mmol/L interval of comparison values specified by ISO 15197:2003. For the comparison interval of
4.20 mmol/L, 110, 110, and 108 of 113 values fell within the ±20% interval of comparison values for the finger, palm, and arm tests, respectively, as specified by the ISO 15197:2003 criteria (Table 2
). Thus, the ISO 15197:2003 requirements for SMBG meters were met. Fig. 4
shows the correlation coefficients for the comparison between the Rightest GM310 test and the comparison method for the 3 alternative sites. For the fingertip test (n = 139), y = 0.99x – 0.08 mmol/L, r = 0.989, and the SD of the residuals (Sy|x) = 0.90 mmol/L. For the palm test (n = 139), y = 0.98x + 0.14 mmol/L, r = 0.990, and Sy|x = 0.85 mmol/L. For the arm test (n = 139), y = 0.97x – 0.21 mmol/L, r = 0.990, and Sy|x = 0.85 mmol/L.
|
EGA results for the fingertip test (Fig. 4A
) were within acceptable zone A in 97.8% of the cases. In the palm and arm tests, 100% of the values were within zones A plus B (97.8% and 96.4%, respectively, within zone A; 2.2% and 3.6% within zone B; Fig. 4, B and C
). The EGA results demonstrate that Rightest SMBG meter GM310 was clinically acceptable for use with alternative test sites.
| Discussion |
|---|
|
|
|---|
The EGA results showed that 100% of the values fell within clinically acceptable zones A and B, suggesting that results obtained in alternative-site tests with the Rightest GM310 meter are clinically acceptable. According to ISO 15197:2003 criteria, the Bionime Rightest GM310 has met the requirements of SMBG systems for testing at alternative sites (fingertip, palm, and arm).
More than 50% of the values generated with the Bionime Rightest GM310 meet the alternative-site test criteria of the American Diabetes Association, which recommends that 100% of the values fall within the ±5% interval of comparison-method values(25)(26). Currently, only 15%–50% of values of most commercially available glucose meters meet this criterion(7)(22)(23)(26)(27)(28), and the performance of the Rightest meter is consistent with that of other meters.
Several complicating methodologic factors should be considered when a comparison method is used. The glucose concentration of capillary whole blood is lower than that of venous plasma(29). This difference is partly attributable to the hematocrit, or a fasting vs nonfasting state. One potential limitation of the performance of the Rightest GM310 is the hematocrit in a blood sample. Many studies have concluded that variation in the hematocrit significantly reduces the accuracy of SMBG measurements. A high hematocrit is associated with underestimation, whereas a low hematocrit causes overestimation of glucose concentrations(30)(31)(32). Therefore, hematocrits may constitute a significant clinical risk in evaluating the hypoglycemia symptoms that are encountered in the up to 20% of the neonatal population that commonly has high hematocrits(33)(34). We analyzed the effect of hematocrits in this clinical trial and obtained results similar to those of prior hematocrit studies. Fig. 2
in the online Data Supplement shows the effect of the hematocrit on glucose measurement.
In summary, the Bionime Rightest GM310 complies with ISO 15197:2003 criteria and produces acceptable glucose results for alternative test sites (fingertip, palm, or arm) that are generally equivalent.
| Acknowledgments |
|---|
Authors Disclosures of Potential Conflicts of Interest: Upon submission, all authors completed the Disclosures of Potential Conflict of Interest form. Potential conflicts of interest:
Employment or Leadership: Ming-Hsun Wu, Min-Sheng General Hospital, Taiwan; Mei-Yen Fang, Bionime Corporation; Lin-Ni Jen, Bionime Corporation; Hung-Chan Hsiao, Bionime Corporation; Andreas Muller, IMCARMED GmbH; Cheng-Teng Hsu, Bionime Corporation.
Consultant or Advisory Role: None declared.
Stock Ownership: None declared.
Honoraria: None declared.
Research Funding: This study was supported by an educational grant from the Bionime Corporation, Taiwan. The evaluation study received cooperation from the Min-Sheng General Hospital, Taiwan.
Expert Testimony: None declared.
Role of Sponsor: The funding organizations played no role in the design of study, choice of enrolled patients, review and interpretation of data, or preparation or approval of manuscript.
Acknowledgments: We would like to thank Ms. Tonia Hsiao for her helpful suggestions and the scientific skills provided by IMCARMED GmbH-Med.-Wiss. Abt., Germany. Additionally, we also appreciated Prof. Ted Knoy for his editorial assistance.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |