| Abstract |
|---|
|
|
|---|
Methods: We extracted 3/16-inch DBS punches with 300 µL methanol containing AA and AC stable isotope-labeled internal standards. This extract was derivatized with butanol-HCl. In parallel, we extracted SUAC from the residual filter paper with 100 µL of a 15 mmol/L hydrazine solution containing the internal standard 13C5-SUAC. We combined the derivatized aliquots in acetonitrile for MS/MS analysis of AC and AA with additional SRM experiments for SUAC (m/z 155–137) and 13C5-SUAC (m/z 160–142). Analysis time was 1.2 min.
Results: SUAC was increased in retrospectively analyzed NBS samples of 11 TYR 1 patients (length of storage, 52 months to 1 week; SUAC range, 13–81 µmol/L), with Tyr concentrations ranging from 65 to 293 µmol/L in the original NBS analysis. The mean concentration of SUAC in 13 521 control DBS was 1.25 µmol/L.
Conclusion: The inclusion of SUAC analysis into routine analysis of AC and AA allows for rapid and cost-effective screening for TYR 1 with no tangible risk of false-negative results.
| Introduction |
|---|
|
|
|---|
To remedy this situation, we developed and validated a method for the simultaneous determination of SUAC, amino acids (AA), and acylcarnitines (AC) by flow injection tandem mass spectrometry (FIA-MS/MS), which requires minor increases in reagent costs and labor but no additional equipment.
| Materials and Methods |
|---|
|
|
|---|
preparation of calibrators and controls
We prepared DBS for SUAC calibration, recovery, stability, and imprecision studies as follows: aliquots of pooled whole blood were spiked with SUAC to achieve final concentrations of 0, 5, 10, 20, 50, and 100 µmol/L, then spotted on filter paper (Whatman ProteinSaver 903) and dried overnight at room temperature. The spotted cards were transferred to a zip-lock bag with desiccator and stored at –20 °C.
samples
For the validation of this method, we analyzed 13 521 leftover newborn screening blood spots that were initially submitted to Mayo Clinics supplemental newborn screening program and were not suggestive of TYR 1, based on a previously published 2-tier screening approach(8). In addition, blood spots from the original newborn screening samples of 11 confirmed TYR 1 patients were made available by their respective physicians and screening laboratories with informed consent.
methods
Sample preparation entails a parallel work-up of eluates from the same DBS that contain either amino acids and acylcarnitines or SUAC. The analytes are subsequently recombined for ESI-MS/MS analysis as outlined in Fig. 1
.
|
We punched single 3/16-inch discs (Wallac DBS Puncher; PerkinElmer Life and Analytical Sciences) from controls and samples into wells of a flat-bottom 96-well plate (Fisher Chemical Co.). We added methanol solution containing the amino acid, carnitine, and acylcarnitine internal standards (300 µL) to each well. The plate was covered and the discs were eluted by mixing using an orbital rotator for 30 min at 120 rpm. We transferred the methanol eluates to another round-bottom 96-well plate, leaving the residual filter paper discs for subsequent elution of SUAC for 30 min at 65 °C by addition of 100 µL acetonitrile/water/formic acid solution (80:20:0.1, vol:vol:vol), which also contains 0.1% hydrazine monohydrate (15 mmol/L) and the internal standard 13C5-SUAC (0.25 µmol/L). During the extraction of the residual filter paper discs, the plate containing the methanol eluates was evaporated under heated (40 °C) nitrogen to dryness (approximately 8–12 min). We added 3 mol/L HCl in n-butanol (50 µL) to the dried residues, which were then covered and incubated for 15 min at 65 °C. After incubation, excess reagent was evaporated to dryness (approximately 5–7 min) under heated (40 °C) nitrogen, and the residue, containing butyl esters of the AA and AC, was reconstituted in 100 µL of mobile phase (acetonitrile:water:formic acid; 50:50:0.02, vol:vol:vol).
After extraction of SUAC from the leftover dried filter paper spots, the eluates were transferred to another round-bottom 96-well plate and dried under heated (40 °C) nitrogen for approximately 7 min. We removed any residual hydrazine reagent by addition of 100 µL methanol to each well, mixing, and evaporation under heated nitrogen. We transferred the mobile phase in the plate containing butylated amino acids and acylcarnitines to corresponding wells in the plate containing the SUAC-hydrazone/IS residues. This plate was covered, gently mixed by agitation, and ready for analysis by FIA-MS/MS.
ms/ms procedure
We used a triple-quadrupole MS/MS system (Applied Biosystems/MDS Sciex API 3000) operated in positive ion mode (source voltage, 5500 V). Mass calibration and resolution of both resolving quadrupoles were automatically optimized by the use of a poly(propylene)glycol solution introduced by an infusion pump. We performed method optimization for the detection of SUAC by selected reaction monitoring (SRM) by infusing a 10 µmol/L solution of SUAC and its internal standard as hydrazones at 0.6 mL/h. The instrument was optimized automatically by an internal algorithm to monitor the transitions m/z 155.0 to 137.0 and m/z 160.0 to 142.0 for SUAC and the internal standard, respectively. These SRM experiments (100-ms dwell, each experiment) were then added to precursor, neutral loss, and SRM scans for acylcarnitine and amino acid analysis in blood spots as described(9)(10)(11)(12). Sample was introduced into the atmospheric pressure ionization source by a Leap Technologies HTC PAL Autosampler and a Perkin-Elmer Micro LC pump. Autosampler injections of 10 µL per sample were made into the mobile-phase (acetonitrile:water:formic acid; 50:50:0.02, vol:vol:vol) flow of 0.025 mL/min. Analysis time was 1.2 min/injection.
safety precautions
Hydrazine is a known toxic and carcinogenic chemical. To prevent exposure to this agent and its hazardous fumes, samples are prepared by use of a robotic pipetter (Freedom Evo; Tecan Systems Inc.) equipped with Plexiglas shielding and a custom-made fume evacuation system (see Supplemental Fig. in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol54/issue4). Hydrazine containers are not opened outside of this system, and its effectiveness and safety were tested and documented by Mayo Clinics Section of Occupational Safety and Security.
| Results |
|---|
|
|
|---|
We determined intraassay imprecision by performing 5 replicate analyses of samples with 5 different SUAC concentrations, 1.1, 7.6, 26.5, 57.0, and 115.8 µmol/L. The mean CVs were 7.7%, 8.7%, 7.8%, 3.1%, and 12.3% respectively. We determined interassay imprecision by analysis of DBS with 2 different concentrations of SUAC, 6.0 and 60.7 µmol/L. Mean CVs of 192 analyses determined over 2 months were 16.7% and 15.8%, respectively.
stability
We assessed the stability of extracted and prepared specimens by analysis of 23 newborn screening samples and 4 controls enriched with SUAC at 6.4 µmol/L (n = 2) and 64 µmol/L (n = 2) before and after 24 h under ambient conditions. Prepared specimens yielded their expected concentrations within 6.2% (Table 1
).
|
method comparison for aa and ac
We conducted a comparison using 290 newborn screening samples that were analyzed with and without the SUAC modification. Bland-Altman and x-y plots were generated for those analytes where corresponding isotopically labeled internal standards are available: alanine, valine, leucine/isoleucine, methionine, phenylalanine, tyrosine, asparagine, glutamic acid, ornithine, citrulline, arginine, glycine, free carnitine, and C2, C3, C4, C5, C6, C8, C14, C16 acylcarnitines. In addition, we compared >10 000 newborn screening results obtained before and after the inclusion of SUAC analysis to our newborn screening method (Fig. 2
). AC and AA concentrations in both of these studies compared well, with no clinically significant differences (Fig. 2
). However, we noted that the modified screening method yielded a higher concentration of C12 acylcarnitine (Fig. 2B
). Whereas isolated increases of C12 acylcarnitine are not indicative of a particular disorder, we studied the cause of this observation. The molecular ion of the butylester derivatives of C12 and C16 acylcarnitine species are detected by a precursor ion scan of m/z 85 at m/z 400.5 and 456.5, respectively. The same analysis of nonderivatized C16 acylcarnitine yields a signal at m/z 400.5. We therefore concluded that the higher abundance at m/z 400.5 observed in the modified method originated from nonderivatized C16 acylcarnitine extracted from the residual DBS along with SUAC. This was proven by reconstitution of the derivatized SUAC extract in mobile phase without the addition of the extracted and derivatized amino acids and acylcarnitines in DBS spiked with either AA (n = 4) or AC (n = 4). The contribution of nonderivatized acylcarnitines, when using the modified method, is negligible for most acylcarnitine species (Fig. 2B
) and most likely notable only for C16 acylcarnitine secondary to the polarity and higher concentration of this acylcarnitine species in newborn DBS.
|
suac method comparison
We measured SUAC retrospectively in newborn screening samples of 11 patients with TYR 1 using our previously described LC-MS/MS method(8) and the modified newborn screening method. An x-y method plot yielded a slope, y-intercept, and coefficient of linear regression (R2) of 1.7, –6.0, and 0.95, respectively. The difference between these methods is likely related to different extraction, sample preparation, and analytical procedures.
suac reference range
The SUAC distribution for 13 521 random newborn screening samples is shown in Fig. 3
. The mean, median, and 99.5th percentile of SUAC values were 1.25 µmol/L, 1.21 µmol/L, and 2.65 µmol/L, respectively. The lowest SUAC values determined in leftover newborn screening samples of 11 TYR 1 patients was 13 µmol/L, a value observed in 2 cases whose samples had been in storage at room temperature for 36 and 53 months, respectively (Table 2
). As part of this retrospective study, we also compared the tyrosine concentrations determined during the initial newborn screening analysis to those measured in the retrospective analysis (Table 2
). As we have shown previously, tyrosine is typically not stable over time. Reanalyzing 3 leftover DBS of patients with TYR 1 also revealed a decrease of the measurable SUAC concentration by 35%, 21%, and 9% after storage at room temperature for 6 months, 6 months, and 1 month, respectively. Based on these observations, we set the cutoff for SUAC at 5.0 µmol/L, which allowed clear discrimination of the control population from TYR 1 patients.
|
|
| Discussion |
|---|
|
|
|---|
Sample preparation for this method includes a double extraction of the DBS sample, which raises concern as it may increase the recovery of analytes and therefore require determination of new cutoff concentrations. Comparing the results of >10 000 samples analyzed before and after implementation of the new method revealed mostly negligible increases in recovery of amino acids and acylcarnitines by the additional extraction. The only notable exception is the calculated concentration of C12 acylcarnitine, which is higher because it now represents a mixture of the butylester of C12 acylcarnitine with contribution of nonderivatized C16 acylcarnitine, the most abundant long-chain acylcarnitine species in newborn blood spots. This effect can be dealt with either by increasing the cutoff for what is considered C12 acylcarnitine or, as we have done, by avoiding setting an active cutoff value altogether, because this analyte by itself is not associated with a particular disorder. Laboratories that do not derivatize their samples would likely need to adjust their cutoff values for C16 acylcarnitine following similar validation studies as described here. Most importantly, the interpretation of newborn screening results is not affected by the new method which we implemented in May 2007. Since then, >40 000 samples were analyzed with no negative impact on newborn screening performance metrics, as the false-positive rate (0.08%), positive predictive value (42%), and detection rate (1:1688 live births) remained stable(17). So far, no newborn screening sample revealed an increased SUAC concentration. However, we assume that diagnostic sensitivity is 100% because no patient was diagnosed clinically with TYR 1 in Minnesota since June 2004.
In conclusion, we have developed and validated a new assay that successfully combines existing methods for newborn screening for disorders of fatty acid, organic acid, and amino acid metabolism including TYR 1. This method is sensitive, rapid, and cost-effective, with no increased risk of either false-positive or false-negative results and could be implemented by all laboratories already using FIA-MS/MS for newborn screening to the benefit of the population they serve, including patients with TYR 1.
| Acknowledgments |
|---|
Financial Disclosures: None declared.
Acknowledgments: The authors are deeply grateful to Shu Chaing (North Carolina Department of Health, Raleigh, NC), Sara Copeland (University of Iowa, Des Moines, IA), Kristina Cusmano-Ozog (Stanford University, Palo Alto, CA), Harry Hawkins (Michigan Department of Health, Lansing, MI), Nancy Leslie (Cincinnati Childrens Hospital, Cincinnati, OH), Fred Lorey (California Department of Health, Richmond, CA), C. Ronald Scott (University of Washington, Seattle, WA), and Jennifer Sullivan and Dwight Koeberl (Duke University Medical Center, Durham, NC) for assisting in the retrieval of original newborn screening samples of known TYR 1 patients. In addition, we appreciate the help of David McHugh in preparing Fig. 2
.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
B. W. Adam, T. H. Lim, E. M. Hall, and W. H. Hannon Preliminary Proficiency Testing Results for Succinylacetone in Dried Blood Spots for Newborn Screening for Tyrosinemia Type I Clin. Chem., December 1, 2009; 55(12): 2207 - 2213. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Dietzen, P. Rinaldo, R. J. Whitley, W. J. Rhead, W. H. Hannon, U. C. Garg, S. F. Lo, and M. J. Bennett National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines: Follow-Up Testing for Metabolic Disease Identified by Expanded Newborn Screening Using Tandem Mass Spectrometry; Executive Summary Clin. Chem., September 1, 2009; 55(9): 1615 - 1626. [Abstract] [Full Text] [PDF] |
||||
![]() |
Impact of Expanded Newborn Screening--United States, 2006 JAMA, November 19, 2008; 300(19): 2242 - 2244. [Full Text] [PDF] |
||||
![]() |
K. A. Pass and M. Morrissey Enhancing Newborn Screening for Tyrosinemia Type I Clin. Chem., April 1, 2008; 54(4): 627 - 629. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |