1 Surgical Services and Center for Engineering in Medicine, Massachusetts General Hospital, Harvard Medical School and Shriners Hospital for Children, Boston, MA; 2 Department of Surgery, University of Florida College of Medicine, Gainesville, FL; 3 Stanford Genome Technology Center, Palo Alto, CA; 4 Department of Surgery, UMDNJ-Robert Wood Johnson Medical School, New Brunswick, NJ; 5 Department of Surgery, University of Washington, Harborview Medical Center, Seattle, WA; 6 Department of Surgery, University of Texas Medical Branch, Shriners Burns Hospital, Galveston, TX; 7 Department of Molecular Genetics and Microbiology, University of Florida College of Medicine, Gainesville, FL.
aAddress correspondence to this author at: 114 16th St., Charlestown, MA 02129. Fax: (617) 523-1684; e-mail: mtoner{at}hms.harvard.edu.
| Abstract |
|---|
|
|
|---|
Methods: We processed whole blood from hospitalized patients after burn injury and severe blunt trauma according to the microfluidic and standard macroscale leukocyte isolation protocol. Side-by-side comparison of RNA quantity, quality, and genome-wide expression patterns was used to clinically validate the microfluidic technique.
Results: When the microfluidic protocol was used for processing, sufficient amounts of total RNA were obtained for genome-wide expression analysis from 0.5 mL whole blood. We found that the leukocyte expression patterns from samples processed using the 2 protocols were concordant, and there was less variability introduced as a result of harvesting method than there existed between individuals.
Conclusions: The novel microfluidic approach achieves leukocyte isolation in <25 min, and the quality of nucleic acids and genome expression analysis is equivalent to or surpasses that obtained from macroscale approaches. Microfluidics can significantly improve the isolation of blood leukocytes for genomic analyses in the clinical setting.
| Introduction |
|---|
|
|
|---|
Improved technologies are required that are fast, reproducible, and do not activate or perturb the cell populations. Among emerging technologies, microfluidics and "lab-on-a-chip" devices have the potential to spur the development of protocols and affordable instruments for performing specific blood analyses with minimal perturbation of individual cell populations(7). Previously, we reported a microfluidic system for the enrichment and isolation of leukocytes from whole blood for genomic analysis(8)(9). The device, consisting of a serpentine channel, enables rapid depletion of erythrocytes from whole blood via continuous deionized water lysis while enriched leukocytes are readily recovered for downstream genomic analyses. We tested this device with whole blood obtained from healthy volunteers or normal whole blood treated ex vivo with a bacterial toxin(9). However, the critical clinical validation of the microfluidic technology in patient samples collected at the point-of-care has not been addressed to date.
In this study, we tested the hypothesis that the newly developed microfluidic technology, when applied for cell separation by skilled medical personnel with no previous experience with microfluidic devices, can generate samples of equal quality to those of traditional cell separation technologies. We provided 4 h of training on the use of the microfluidic device to research nurses and other personnel with previous experience in macroscale leukocyte isolation techniques. The nurses, following both the microfluidic and the macroscale protocols, processed whole blood samples from critically ill patients with severe burn or trauma injuries. The extracted RNA from these samples was processed further in parallel for microarray analysis. We show that the expression patterns from samples processed using the microfluidic and the macroscale buffy-coat procedures are similar, and there is less variability between the 2 methods than there is between individuals. To our knowledge, this is the first successful application of microfluidic techniques for processing clinical samples obtained from critically ill, hospitalized patients, and the first validation of a microfluidic technique in a multiple hospital setting.
| Materials and Methods |
|---|
|
|
|---|
|
The single-use microfluidic chips were manufactured at the BioMEMS resource center (Boston, MA) as described(9). These were assembled and shipped to 4 clinical sites (Gainesville, FL; New Brunswick, NJ; Galveston, TX; Seattle WA). Blood samples were processed at each clinical site and isolated cells were lysed in RNA isolation buffer (RLT bufferTM; Qiagen) and frozen at –80 °C. The frozen RNA samples were shipped to the sample coordination site at the University of Florida. Depending on the experiment, samples were either processed locally or sent to Stanford University for processing of RNA and hybridization to HU133 plus v2 GeneChipTM.
microfluidic device fabrication
The device (Fig. 1
) and the experimental setup have been described(9). Briefly, the microfluidic device was fabricated by casting polydimethylsiloxane (PDMS1
; Dow Corning) polymer on a resist-structured silicon wafer according to standard soft lithographic techniques. Structures in SU-8 resist (MicroChem) were produced according to the suppliers recipe using standard MEMS technology. The PDMS was mixed (10:1, wt/wt) with a cross linker, poured on top of the silicon wafer, degassed, and cured at 65 °C for 6 h. The PDMS with the replicated channels was peeled off from the master, and channel access holes were punched with a 22-gauge needle. The PDMS replica was bonded to a glass slide via oxygen plasma. Access tubing (Tygon; Small Parts) of slightly larger diameter was press-fitted into the holes.
participants
We obtained permission to obtain venous blood from healthy volunteers and patients from the institutional review boards of the participating institutions. Informed consent was obtained from 4 healthy individuals at University of Florida, 5 severely injured trauma patients at Washington University, 5 severely injured burn patients at University of Texas, and 3 healthy volunteers to be administered endotoxin at UMDNJ-Robert Wood Johnson Medical School.
blood processing
Venous blood (7 mL) was collected into VacutainerTM tubes containing EDTA (Becton Dickinson), and 0.5 mL blood was collected directly from the Vacutainer into a 1-mL syringe (Becton Dickinson) and processed according to the microfluidic protocol, whereas the remaining 6.5 mL blood was processed in parallel according to the macroscale protocol. Three healthy volunteers selected for the endotoxin study were intravenously administered NIH Clinical Center Reference Endotoxin (CC-RE-Lot 2) at a dose of 2 ng/kg body weight over a 5-min period. Blood samples were collected before endotoxin infusion (0 h) and at 3, 6, and 24 h after infusion and processed according to the microfluidic and standard bulk method in parallel. The patients clinical characteristics are shown in Table
S1 in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol54/issue5.
|
microfluidic blood processing
The experimental setup for the microfluidic technique consists of the microfluidic device and 2 syringe pumps (Fig. 1
). The device is first primed by filling with 1x PBS and connected to respective syringes without trapping air bubbles. One syringe pump (Harvard Pump 11 Plus; Harvard Apparatus) drives the whole blood at 20 µL/min through 1 inlet, and the second pump (Harvard PHD 22/2000 Syringe Pump; Harvard Apparatus) drives deionized water and 2x PBS (double the regular PBS concentration) at 600 µL/min through the other 2 inlets. The deionized water branches into 2 streams, 1 on either side of the entry stream of whole blood to focus it into a narrow stream. We optimized the flow rates such that each cell in the blood was exposed to the lysing solution for approximately 10 s before restoring the osmolarity of blood using equal volumes of 2x PBS. The enriched leukocyte population and debris of lysed erythrocytes were collected at the outlet in 50-mL Eppendorf tubes, and cell debris were removed in the supernatant by low-speed centrifugation. The cell pellets were washed in 1 mL PBS, immediately lysed in 600 µL RLT buffer (Qiagen) with 0.1% vol/vol β-mercaptoethanol, and spun through a QIAShreder column (Qiagen). The samples were immediately stored at –80 °C until shipped to the sample collection and coordination site in Florida.
conventional buffy-coat blood processing
The standard buffy-coat leukocyte isolation protocol has been described(1)(10)(11). Briefly, we centrifuged the blood sample at 400g for 10 min at room temperature and transferred the buffy-coat layer to 50-mL conical tubes (Corning Labs). We lysed residual erythrocytes with 45 mL lysing buffer (EL Buffer; Qiagen) for 15 min at 4 °C and collected the leukocyte-rich fraction by centrifugation at 400g for 10 min at 4 °C. The leukocyte pellets were washed in ice-cold PBS, immediately lysed with 2 mL RLT Buffer (RNeasy® Midi Kit; Qiagen), and homogenized by aspirating and forcing the sample repeatedly through a 10-mL syringe (Monoject) with an 18-gauge needle (Becton Dickinson). The sample was immediately stored at –80 °C until shipped to the sample collection and coordination site in Florida.
leukocyte phenotyping
We took 50 µL of the cell suspension for flow cytometry analysis before the cell-lysis step. Different subpopulations were assessed by staining with fluorescein isothiocyanate–anti-CD66b, phycoerythrin–anti-CD2, and PerCP-CY5.5–anti-CD33 antibodies (BD Biosciences). Stained cells were washed in PBS without calcium or magnesium, plus 1% BSA and 0.1% sodium azide, then pelleted, resuspended in PBS, and analyzed on a FACSCaliburTM flow cytometer (BD Biosciences), using CellQuestTM software (BD Pharmingen). We used 3-color analyses to identify granulocytes as side-scatter high, CD66b high; monocytes as side-scatter intermediate, CD33 high; and T lymphocytes as side-scatter low, CD2 high.
rna isolation, Crna synthesis, and microarray hybridization
We extracted total cellular RNA using a commercial RNA purification kit (RNeasy; Qiagen) and amplified biotinylated cRNA(12) from 1 µg total RNA. cRNA was transcribed in vitro incorporating biotinylated nucleotides using the Affymetrix High Yield RNA Transcript T7 Kit (Affymetrix), and the product was hybridized onto an Affymetrix HU133 plus v.2 GeneChipTM for 16 h at 45 °C in an Affymetrix hybridization oven. Arrays were stained and washed using an Affymetrix fluidics station and EukGEWSv4 Affymetrix protocol and analyzed on an Affymetrix scanner. We determined quality and purity of total RNA and biotinylated-synthesized cRNA by UV spectroscopy and capillary electrophoresis using the Bioanalyzer 2001 system (Agilent). An absorbance ratio (A260/A280) of 1.8 to 2.1 is indicative of highly purified RNA.
microarray data analysis
We performed low- and high-level statistics using dChip. Expression was modeled using Perfect Match Only dChip algorithms. Affymetrix Microarray Suite GeneChip Operating Software identified probe sets whose signal intensities were at or below background as "absent." Absent probe sets were discarded from the high-level analysis on all arrays.
We applied an unsupervised analysis to assess similarities and gene expression profile differences. We performed unsupervised analyses on the expression matrix by selecting probe sets whose hybridization signal varied across the data set with a CV (SD/mean ratio) >1.0. We then subjected the mean-centered and variance-normalized levels to hierarchical cluster analyses using the dChip algorithm. We calculated Pearsons correlation coefficient to assess similarity of gene expression among probe sets, with the results displayed as a dendrogram on top of the cluster image. Significance analysis of microarrays identified probe sets whose intensities differed between the 2 groups at a false discovery rate of 1% (B = 0.01).
| Results |
|---|
|
|
|---|
leukocyte composition isolated from whole blood
Complete lysis of erythrocytes was achieved within 10 s, and close to 100% of the leukocytes were recovered in the microfluidic device. We have earlier demonstrated that the microfluidic system is better than macroscale protocols for total and differential leukocyte recovery, owing to more precise control of time of exposure to the lysing conditions(9). In this study, we further examined the differential leukocyte composition by flow cytometry to evaluate the ability of the microfluidic isolation protocol to identify phenotypic changes in blood leukocyte populations secondary to in vivo endotoxin administration. Whole blood leukocytes were isolated according to the 2 protocols from 3 volunteers immediately before and at 3, 6, and 24 h after intravenous administration of bacterial endotoxin. Fig. 2
shows differential results of the leukocyte subpopulations during the 24-h time course. Comparison of the differential profiles of the subpopulations reveals that the 2 protocols generated similar results. As reported(13), endotoxin administration produces a rapid and transient leukopenia, followed by a release of mature and immature neutrophils from the bone marrow at 3 h and a sustained T lymphopenia and monocytopenia by 6 h. After 24 h, the differential subpopulation cells return to initial values, indicating that the host response to the bacterial toxin has restored homeostasis.
|
rna yield and quality
We used RNeasy Qiagen RNA purification kits to extract total RNA from both the microfluidics cassettes and the macroscale isolation procedure. We generated 26 RNA samples from each of the 2 protocols: healthy controls (n = 4) and healthy controls who received the in vivo administration of endotoxin (n = 3) at 4 time points. Results of RNA yield and quality are shown in Fig. 3
. Although there were some variations in the amount of RNA recovered among the protocols and individuals, all samples provided sufficient RNA for subsequent microarray analysis. The quality of isolated RNA was acceptable for all samples generated by both protocols, as confirmed by capillary electrophoresis using the Agilent 2100 Bioanalyzer system. Spectrometric analysis (A260/A280 ratio) of the 12 samples from the healthy volunteers administered endotoxin averaged 1.8 (SD 0.1) for both methods, indicating that the samples were free of contaminating protein.
|
gene expression analysis
We generated 52 arrays (26 for each protocol) and normalized the readings from the GeneChips by applying model-based algorithms from dChip 2006 to create an expression matrix. We identified 8766 probe sets with CVs of >1.0 across the data set. Their expression values were mean-centered and variance-normalized before hierarchical clustering. We analyzed hybridization signals from the different samples to determine the extent of gene expression differences between the 2 methods. The unsupervised hierarchical clustering with 8766 probe sets revealed grouping according to the clinical source of samples rather than according to differences in the leukocyte isolation methods (Fig. 4
). This is noticeable from the heat map of the burn and trauma patient samples, where the pattern of the individual probes and individual samples were properly ordered by hierarchical clustering. Fig. 5
shows the dendrogram from cluster analysis of the patients, reflecting the degree of similarity between the connected samples. For the samples obtained from volunteers administered endotoxin, the unsupervised analysis showed that the major node of separation tends to separate the 3- and 6-h time periods from the 0- and 24-h time points, and that within a time point there was a tendency for these individuals to subcluster over protocols (Fig. 4
). As expected, gene expression patterns from the healthy controls were subclustered with the samples from the endotoxin volunteers obtained at the 0- and 24-h time points. Pearson correlation coefficients of the gene expression profiles, estimating the degree of correlation between the isolation methods and intramethod between persons, are summarized in Table 1
. For a given patient, the correlation between the protocols was calculated and then averaged over the various individuals. For a given protocol (microfluidics or buffy coat), the correlation coefficients of the gene expression profiles between the individuals were calculated in a pairwise fashion and then averaged. The correlation coefficients for the 2 isolation protocols from the healthy controls and patients [>0.98 (0.01)] were consistently higher than the correlation between individuals for the same protocol. When the buffy coat and microfluidics protocols were compared, supervised analyses for samples obtained from volunteers administered endotoxin or from critically ill burn and trauma patients identified 76, 33, and 43 probe sets, respectively, significant at P < 0.001, which is below the number expected to exceed the significance threshold by chance. Therefore, the supervised analysis comparing the 2 protocols detected no apparent differences in gene expression profiles.
|
|
| Discussion |
|---|
|
|
|---|
Current macroscale techniques for isolating nucleic acids from whole blood and blood leukocytes use either commercial preparations such as PAXGeneTM, which introduce considerable variance in the gene expression profiles(10), or complex lysis and density centrifugation protocols that require substantial technical skill. In addition, with macroscale isolation procedures, the cells are often exposed for extended periods to known stressors including altered temperature, tonicity, and G-forces(17)(18)(19)(20)(21), all of which can activate leukocytes. There is, therefore, a compelling need for standardized, fast, and robust methods to isolate leukocytes for subsequent proteomic and genomic analyses in clinical settings.
An entirely new class of analysis systems for chemical and biological applications is emerging, enabled by microfluidic technologies(22)(23)(24), with increasing impact on the sorting, handling, and analysis of mammalian cells(24)(25)(26). The miniaturization of analytical equipment may allow several shortcomings associated with bulky and expensive instrumentation to be overcome. The new microscale devices would enable reduction in sample and reagent volumes and provide results faster, at the bedside. The technologies, derived from the electronics industry, would allow batch-fabrication, reducing costs, and resulting in disposable instruments that are used once and thrown away to prevent sample contamination. Finally, apart from having high accuracy, microfluidic-based methods should minimize the number of manual steps and provide rapid and reproducible results to compete with present technologies and be fully implemented in the clinical setting. Consequently, clinical diagnostics is one of the fastest developing areas for microfluidic applications(27). Despite the fact that microfluidics approaches hold great potential for clinical application, their validation and actual application in clinical settings are infrequent.
In this report, we critically evaluated the ability of our newly developed microfluidic technique(9) to isolate leukocytes for proteomic and genomic analyses in the clinical setting. We provided brief training at 4 clinical sites to research nurses with no prior knowledge in microfluidic techniques and compared the quality of the samples generated by the new and traditional cell separation methods. The microfluidic devices require minimal operator intervention and enable separation of leukocytes from whole blood using protocols that are faster and more efficient than standard techniques. Nurses and laboratory personnel were able to generate reliable clinical data during the brief training session using the microfluidic protocol. This rapid and single-step method does not require prior manipulation of whole blood and has paved the way for the translation of an assay from the research laboratory into the clinical setting.
The quality of RNA from all samples, after storage at –80 °C for several months, and microarray analyses was determined to be excellent. Although we observed variability in RNA yield, the amount of RNA available was always larger than the amount required for analysis. Only a fraction of the total amount of RNA isolated (1 µg total RNA) was used to prepare the microarrays. Overall, 500 µL whole blood was shown to produce sufficient quantities of RNA for genome-wide expression analysis. Hence, this microfluidic approach dramatically reduces the blood requirements compared with macroscale protocols. In the future, it may be possible to reduce this amount of blood even further, which would open the possibility for blood collection from a simple fingerprick, avoiding the need for phlebotomy and its risk of complications. These new capabilities could make microfluidic technologies extremely attractive for pediatric and neonatal patients and allow more frequent sampling from adult patients with conditions that require frequent monitoring.
Side-by-side comparison of the standard bulk and new microscale methods produced highly concordant results in healthy controls and in response to endotoxin administration. As confirmed by gene expression analysis, both isolation techniques were capable of differentiating the changes in response to endotoxin administration at 3- and 6-h time points from the 0- and 24-h time points. As expected, traumatic and burn injuries in patients induced dramatic changes in apparent gene expression. In addition, a correlation coefficient >0.98 (Table 1
) in the gene expression results for the patients by each protocol strongly suggested that there were no substantive differences between the protocols. Hence, the microfluidic technique should be readily applicable for detecting changes in gene expression that occur in response to burn and trauma injury. Because leukocytes are capable of mounting rapid responses and altering their gene expression accordingly, the ability to process blood samples quickly after collection from patients free of potential operator handling errors is extremely important. Microfluidic techniques that could be implemented as portable systems operated at the bedside would offer advantages in this area.
In summary, the 4-clinical-site comparison reveals essentially no differences between the microfluidic technique and the standard leukocyte isolation method with respect to RNA yield and quality and genome-wide expression analysis. The microfluidic technique makes the routine processing of whole blood more efficient with an automated and easy-to-use system, ideal for genome-wide expression analysis studies involving patients.
| Acknowledgments |
|---|
Financial Disclosures: None declared.
Acknowledgments: We thank Octavio Hurtado for support with microfabrication. We acknowledge Megan Knoll, Keir, Deb Benjamin, Robert Granier, and Liz Taylor for their help processing blood samples. Additional participating investigators in the Large Scale Collaborative Research Program, "Inflammation and the Host Response to Injury," are as follows: Paul E. Bankey, Timothy R. Billiar, Bernard H. Brownstein, David G. Camp II, George Casella, Irshad H. Chaudry, Mashkoor Choudhry, J. Perren Cobb, Asit De, Constance Elson, Bradley Freeman, Richard L. Gamelli, Nicole S. Gibran, Douglas L. Hayden, Brian G. Harbrecht, Jureta W. Horton, William Hubbard, Jeffrey Johnson, Matthew B. Klein, James A. Lederer, Tanya Logvinenko, John A. Mannick, Philip H. Mason, Grace P. McDonald-Smith, Bruce A. McKinley, Carol Miller-Graziano, Joseph P. Minei, Ernest E. Moore, Frederick A. Moore, Avery B. Nathens, Grant E. OKeefe, Laurence G. Rahme, Daniel G. Remick, David A. Schoenfeld, Michael B. Shapiro, Martin Schwacha, Geoffrey M. Silver, Richard D. Smith, John Storey, H. Shaw Warren, Michael A. West, and Wenzhong Xiao.